18.08.2017 | Clinical Research | Ausgabe 12/2017

Clinical Orthopaedics and Related Research® 12/2017

Polyethylene Damage Increases With Varus Implant Alignment in Posterior-stabilized and Constrained Condylar Knee Arthroplasty

Zeitschrift:
Clinical Orthopaedics and Related Research® > Ausgabe 12/2017
Autoren:
MD Zhichang Li, PhD Christina I. Esposito, BS Chelsea N. Koch, MS Yuo-yu Lee, MD Douglas E. Padgett, PhD Timothy M. Wright
Wichtige Hinweise
One of the authors (ZL) received funding from the Beijing Joint Care Foundation.
The institution of one or more of the authors (DEP, TMW) has received, during the study period, funding from the Mary and Fred Trump Institute for Implant Analysis.
One of the authors certifies that he (TMW) or a member of his immediate family, has or may receive payments or benefits, during the study period, an amount of USD 100,001-USD 1,000,000 from Exactech Inc (Gainesville, FL, USA).
All ICMJE Conflict of Interest Forms for authors and Clinical Orthopaedics and Related Research® editors and board members are on file with the publication and can be viewed on request.
Clinical Orthopaedics and Related Research® neither advocates nor endorses the use of any treatment, drug, or device. Readers are encouraged to always seek additional information, including FDA-approval status, of any drug or device prior to clinical use.
Each author certifies that his or her institution approved the human protocol for this investigation and that all investigations were conducted in conformity with ethical principles of research.
This work was performed at the Hospital for Special Surgery (New York, NY, USA).

Abstract

Background

Implant malalignment in primary TKA has been reported to increase stresses placed on the bearing surfaces of implant components. We used a longitudinally maintained registry coupled with an implant retrieval program to consider whether preoperative, postoperative, or prerevision malalignment was associated with increased risk of revision surgery after TKA.

Questions/purposes

(1) What is the relative polyethylene damage on medial and lateral compartments of the tibial plateaus from revised TKAs? (2) Does coronal TKA alignment affect implant performance, such that TKAs aligned in varus are predisposed to experience increased polyethylene damage? (3) Does TKA alignment differ between postoperative and prerevision radiographs, and if so, what does this difference suggest about the mechanical contact load placed on a knee with a TKA?

Methods

Between 2007 and 2012, we performed 18,065 primary TKAs at our institution. By March 2016, 178 of those TKAs (1%) were revised at our center at least 2 years after primary surgery at our institution. Eighteen of those TKAs were excluded from this analysis because the tibial insert was not explanted during revision surgery, and four more were excluded because the inserts were lost or returned to the patient before the study was initiated, leaving 156 retrieved polyethylene tibial inserts (in 153 patients) revised at greater than 2 years after the primary TKA for this retrospective study. Patients who underwent revision surgery elsewhere were not considered here, since this study depended on having retrieved components. Polyethylene damage modes of burnishing, pitting, scratching, delamination, surface deformation, abrasion, and third-body debris were subjectively graded on a scale of 0 to 3 to reflect the extent and severity of each damage mode. On preoperative, postoperative, and prerevision radiographs, overall alignment, femoral alignment, and tibial alignment in the coronal plane were measured according to the protocol recommended by the Knee Society.

Results

Knees with more overall varus alignment after TKA had increased total damage on the retrieved tibial inserts (Spearman’s rank correlation coefficients of −0.3 [95% CI, −0.4 to −0.1; p = 0.001]). We also found revised TKAs tended to drift back into greater varus before revision surgery, with a mean (SD) of 3.6° ± 4.0° valgus for postoperative alignment compared with 1.7° ± 6.4° prerevision (p = 0.04).

Conclusions

Despite surgical efforts to achieve neutral mechanical alignment, remaining varus alignment places an increased contact load on the polyethylene articular surfaces. The drift toward further varus alignment postoperatively is consistent with the knee adduction moment remaining high after surgery.

Clinical Relevance

While we found a predisposition toward recurrence of the preoperative varus deformity, we did not find increased medial as opposed to lateral polyethylene damage, which may be explained by the curve-on-curve toroidal design of the articulating surfaces of the TKA implants in this study.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de. Zusätzlich können Sie eine Zeitschrift Ihrer Wahl in gedruckter Form beziehen – ohne Aufpreis.

Bis zum 22.10. bestellen und 100 € sparen!

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 12/2017

Clinical Orthopaedics and Related Research® 12/2017Zur Ausgabe

Symposium: Learning From Large-Scale Orthopaedic Databases

Editorial Comment: Learning From Large-Scale Orthopaedic Databases

  1. Das kostenlose Testabonnement läuft nach 14 Tagen automatisch und formlos aus. Dieses Abonnement kann nur einmal getestet werden.

Neu im Fachgebiet Orthopädie und Unfallchirurgie

 

 

 
 

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Orthopädie und Unfallchirurgie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise