10.01.2020 | Original Article

Portable imageless navigation system and surgeon’s estimate for accurate evaluation of acetabular cup orientation during total hip arthroplasty in supine position

verfasst von: Ryohei Takada, Tetsuya Jinno, Kazumasa Miyatake, Masanobu Hirao, Toshitaka Yoshii, Atsushi Okawa

Erschienen in: European Journal of Orthopaedic Surgery & Traumatology | Ausgabe 4/2020

Einloggen, um Zugang zu erhalten

Abstract

Background

This prospective study aimed to clarify whether this novel device can evaluate the cup orientation during total hip arthroplasty (THA) more closely to that measured in postoperative computed tomography (CT) compared to the surgeon’s estimate using a manual goniometer.

Methods

We prospectively performed 30 cementless THAs via the anterolateral approach in supine position between October 2018 and July 2019, wherein cup orientation was evaluated by both a portable imageless navigation system (HipAlign) and a manual goniometer during surgeries. Primary outcome was the absolute estimate error [the absolute value of the difference between cup angles measured by postoperative CT and those measured by HipAlign (group H) or surgeon’s estimate using the manual goniometer (group S) during surgery]. The number of outliers of the absolute estimate error (> 10°) in each group was also estimated.

Results

The absolute estimate error of cup inclination in groups H and S was 3.3° ± 2.7° and 3.0° ± 2.5°, respectively (p = 0.51), whereas that of cup anteversion was 3.8° ± 3.4° and 6.0° ± 3.7°, respectively (p = 0.0008). The number of outliers of the estimate error in groups H and S was one case (3.3%) and six cases (20.0%), respectively (p = 0.04). In all six outlier cases, surgeons underestimated cup anteversion during surgeries.

Conclusions

This portable imageless navigation system was a useful method, especially for avoiding incorrect cup anteversion. Underestimation of cup anteversion during THA in the supine position with the conventional alignment assisting device should be given attention.
Literatur
Metadaten
Titel
Portable imageless navigation system and surgeon’s estimate for accurate evaluation of acetabular cup orientation during total hip arthroplasty in supine position
verfasst von
Ryohei Takada
Tetsuya Jinno
Kazumasa Miyatake
Masanobu Hirao
Toshitaka Yoshii
Atsushi Okawa
Publikationsdatum
10.01.2020
Verlag
Springer Paris
Erschienen in
European Journal of Orthopaedic Surgery & Traumatology / Ausgabe 4/2020
Print ISSN: 1633-8065
Elektronische ISSN: 1432-1068
DOI
https://doi.org/10.1007/s00590-020-02625-2

Weitere Artikel der Ausgabe 4/2020

European Journal of Orthopaedic Surgery & Traumatology 4/2020 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.