10.10.2020 | Knee Arthroplasty

Posterior stabilized total knee arthroplasty reproduces natural joint laxity compared to normal in kinematically aligned total knee arthroplasty: a matched pair cadaveric study

verfasst von: In Jun Koh, Christen E. Chalmers, Charles C. Lin, Soo Bin Park, Michelle H. McGarry, Thay Q. Lee

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 1/2021

Einloggen, um Zugang zu erhalten

Abstract

Purpose

As the goal of kinematic aligned (KA) total knee arthroplasty (TKA) is to preserve soft tissue tension to the native knee, many KA surgeons recommend cruciate-retaining (CR) prosthesis. However, how a posterior-stabilizing (PS) prosthesis affects the biomechanics of a KA TKA remains unclear. This cadaveric study tested the hypothesis that a PS prosthesis in KA TKA would produce biomechanics similar to CR prosthesis and KA TKA with a PS prosthesis would produce more native knee biomechanics than mechanical aligned (MA) TKA with PA prosthesis.

Methods

Fourteen cadaver knees (7 pairs) were mounted on a knee-testing system to measure knee motion during flexion. For each pair, 1 knee was assigned to KA TKA and the other to MA TKA. In the KA TKA group, the native knee, CR TKA, and PS TKA were tested sequentially. MA TKA was performed using conventional measured resection techniques with a PS prosthesis. All kinematics were measured and compared with the native knee before and after surgery.

Results

A PS prosthesis restored femoral rollback similar to a CR prosthesis. CR TKA showed less lateral rollback at knee flexion ≤ 60° than the native knee. There were no differences in soft tissue tensions among the native knee, CR, and PS prosthesis, except in varus tension at 30° of flexion. Varus tension of CR TKA was larger than those of PS TKA and the native knee after KA TKA with < 1 degree difference. Meanwhile, KA TKA achieved knee motion that was closer to the native knee than did MA TKA at ≥ 60° of flexion when using a PS prosthesis. There were no differences in soft tissue tension between KA-PS and MA-PS TKA.

Conclusions

After KA TKA, a PS prosthesis affords similar femoral rollback and soft tissue tension when compared with a CR prosthesis. A PS TKA may be a feasible strategy for patients requiring a PS prosthesis when performing KA TKA.

Level of evidence

Therapeutic Laboratory study, I.
Literatur
1.
Zurück zum Zitat Abdel MP, Ollivier M, Parratte S, Trousdale RT, Berry DJ, Pagnano MW (2018) Effect of POSTOPERATIVE MECHANICAL AXIS ALIGNMENT ON SURVIVAL AND FUNCTIONAL OUTCOMES OF MODERN TOTAL KNEE ARTHROPLASTIES WITH CEMENT: A CONCISE FOLLOW-UP AT 20 YEARS. J Bone Joint Surg Am 100(6):472–478CrossRef Abdel MP, Ollivier M, Parratte S, Trousdale RT, Berry DJ, Pagnano MW (2018) Effect of POSTOPERATIVE MECHANICAL AXIS ALIGNMENT ON SURVIVAL AND FUNCTIONAL OUTCOMES OF MODERN TOTAL KNEE ARTHROPLASTIES WITH CEMENT: A CONCISE FOLLOW-UP AT 20 YEARS. J Bone Joint Surg Am 100(6):472–478CrossRef
2.
Zurück zum Zitat Baier C, Springorum HR, Götz J, Schaumburger J, Lüring C, Grifka J et al (2013) Comparing navigation-based in vivo knee kinematics pre- and postoperatively between a cruciate-retaining and a cruciate-substituting implant. Int Orthop 37(3):407–414CrossRef Baier C, Springorum HR, Götz J, Schaumburger J, Lüring C, Grifka J et al (2013) Comparing navigation-based in vivo knee kinematics pre- and postoperatively between a cruciate-retaining and a cruciate-substituting implant. Int Orthop 37(3):407–414CrossRef
3.
Zurück zum Zitat Bellemans J, Colyn W, Vandenneucker H, Victor J (2012) The Chitranjan Ranawat award: is neutral mechanical alignment normal for all patients? The concept of constitutional varus. Clin Orthop Relat Res 470(1):45–53CrossRef Bellemans J, Colyn W, Vandenneucker H, Victor J (2012) The Chitranjan Ranawat award: is neutral mechanical alignment normal for all patients? The concept of constitutional varus. Clin Orthop Relat Res 470(1):45–53CrossRef
4.
Zurück zum Zitat Bourne RB, Chesworth BM, Davis AM, Mahomed NN, Charron KD (2010) Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop Relat Res 468(1):57–63CrossRef Bourne RB, Chesworth BM, Davis AM, Mahomed NN, Charron KD (2010) Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop Relat Res 468(1):57–63CrossRef
5.
Zurück zum Zitat Csintalan RP, Ehsan A, McGarry MH, Fithian DF, Lee TQ (2006) Biomechanical and anatomical effects of an external rotational torque applied to the knee: a cadaveric study. Am J Sports Med 34(10):1623–1629CrossRef Csintalan RP, Ehsan A, McGarry MH, Fithian DF, Lee TQ (2006) Biomechanical and anatomical effects of an external rotational torque applied to the knee: a cadaveric study. Am J Sports Med 34(10):1623–1629CrossRef
6.
Zurück zum Zitat D'Lima DD, Patil S, Steklov N, Colwell CW Jr (2007) An ABJS Best Paper: Dynamic intraoperative ligament balancing for total knee arthroplasty. Clin Orthop Relat Res 463:208–212CrossRef D'Lima DD, Patil S, Steklov N, Colwell CW Jr (2007) An ABJS Best Paper: Dynamic intraoperative ligament balancing for total knee arthroplasty. Clin Orthop Relat Res 463:208–212CrossRef
7.
Zurück zum Zitat Dossett HG, Estrada NA, Swartz GJ, LeFevre GW, Kwasman BG (2014) A randomised controlled trial of kinematically and mechanically aligned total knee replacements: two-year clinical results. Bone Jt J 96-b(7):907–913CrossRef Dossett HG, Estrada NA, Swartz GJ, LeFevre GW, Kwasman BG (2014) A randomised controlled trial of kinematically and mechanically aligned total knee replacements: two-year clinical results. Bone Jt J 96-b(7):907–913CrossRef
8.
Zurück zum Zitat Hirschmann MT, Hess S, Behrend H, Amsler F, Leclercq V, Moser LB (2019) Phenotyping of hip-knee-ankle angle in young non-osteoarthritic knees provides better understanding of native alignment variability. Knee Surg Sports Traumatol Arthrosc 27(5):1378–1384CrossRef Hirschmann MT, Hess S, Behrend H, Amsler F, Leclercq V, Moser LB (2019) Phenotyping of hip-knee-ankle angle in young non-osteoarthritic knees provides better understanding of native alignment variability. Knee Surg Sports Traumatol Arthrosc 27(5):1378–1384CrossRef
9.
Zurück zum Zitat Hirschmann MT, Karlsson J, Becker R (2018) Hot topic: alignment in total knee arthroplasty-systematic versus more individualised alignment strategies. Knee Surg Sports Traumatol Arthrosc 26(6):1587–1588CrossRef Hirschmann MT, Karlsson J, Becker R (2018) Hot topic: alignment in total knee arthroplasty-systematic versus more individualised alignment strategies. Knee Surg Sports Traumatol Arthrosc 26(6):1587–1588CrossRef
10.
Zurück zum Zitat Howell SM, Papadopoulos S, Kuznik KT, Hull ML (2013) Accurate alignment and high function after kinematically aligned TKA performed with generic instruments. Knee Surg Sports Traumatol Arthrosc 21(10):2271–2280CrossRef Howell SM, Papadopoulos S, Kuznik KT, Hull ML (2013) Accurate alignment and high function after kinematically aligned TKA performed with generic instruments. Knee Surg Sports Traumatol Arthrosc 21(10):2271–2280CrossRef
12.
Zurück zum Zitat Hutt JR, LeBlanc MA, Masse V, Lavigne M, Vendittoli PA (2016) Kinematic TKA using navigation: surgical technique and initial results. Orthop Traumatol Surg Res 102(1):99–104CrossRef Hutt JR, LeBlanc MA, Masse V, Lavigne M, Vendittoli PA (2016) Kinematic TKA using navigation: surgical technique and initial results. Orthop Traumatol Surg Res 102(1):99–104CrossRef
13.
Zurück zum Zitat Kim K, Kim J, Lee D, Lim S, Eom J (2019) The accuracy of alignment determined by patient-specific instrumentation system in total knee arthroplasty. Knee Surg Relat Res 31(1):19–24CrossRef Kim K, Kim J, Lee D, Lim S, Eom J (2019) The accuracy of alignment determined by patient-specific instrumentation system in total knee arthroplasty. Knee Surg Relat Res 31(1):19–24CrossRef
14.
Zurück zum Zitat Kim SH, Park YB, Song MK, Lim JW, Lee HJ (2018) Reliability and validity of the femorotibial mechanical axis angle in primary total knee arthroplasty: navigation versus weight bearing or supine whole leg radiographs. Knee Surg Relat Res 30(4):326–333CrossRef Kim SH, Park YB, Song MK, Lim JW, Lee HJ (2018) Reliability and validity of the femorotibial mechanical axis angle in primary total knee arthroplasty: navigation versus weight bearing or supine whole leg radiographs. Knee Surg Relat Res 30(4):326–333CrossRef
15.
Zurück zum Zitat Koh IJ, Kwak DS, Kim TK, Park IJ, In Y (2014) How effective is multiple needle puncturing for medial soft tissue balancing during total knee arthroplasty? A cadaveric study. J Arthroplasty 29(12):2478–2483CrossRef Koh IJ, Kwak DS, Kim TK, Park IJ, In Y (2014) How effective is multiple needle puncturing for medial soft tissue balancing during total knee arthroplasty? A cadaveric study. J Arthroplasty 29(12):2478–2483CrossRef
16.
Zurück zum Zitat Koh IJ, Lin CC, Patel NA, Chalmers CE, Maniglio M, Han SB et al (2019) Kinematically aligned total knee arthroplasty reproduces more native rollback and laxity than mechanically aligned total knee arthroplasty: a matched pair cadaveric study. Orthop Traumatol Surg Res 105(4):605–611CrossRef Koh IJ, Lin CC, Patel NA, Chalmers CE, Maniglio M, Han SB et al (2019) Kinematically aligned total knee arthroplasty reproduces more native rollback and laxity than mechanically aligned total knee arthroplasty: a matched pair cadaveric study. Orthop Traumatol Surg Res 105(4):605–611CrossRef
17.
Zurück zum Zitat Koh IJ, Park IJ, Lin CC, Patel NA, Chalmers CE, Maniglio M et al (2019) Kinematically aligned total knee arthroplasty reproduces native patellofemoral biomechanics during deep knee flexion. Knee Surg Sports Traumatol Arthrosc 27(5):1520–1528CrossRef Koh IJ, Park IJ, Lin CC, Patel NA, Chalmers CE, Maniglio M et al (2019) Kinematically aligned total knee arthroplasty reproduces native patellofemoral biomechanics during deep knee flexion. Knee Surg Sports Traumatol Arthrosc 27(5):1520–1528CrossRef
18.
Zurück zum Zitat Maderbacher G, Keshmiri A, Krieg B, Greimel F, Grifka J, Baier C (2019) Kinematic component alignment in total knee arthroplasty leads to better restoration of natural tibiofemoral kinematics compared to mechanic alignment. Knee Surg Sports Traumatol Arthrosc 27(5):1427–1433CrossRef Maderbacher G, Keshmiri A, Krieg B, Greimel F, Grifka J, Baier C (2019) Kinematic component alignment in total knee arthroplasty leads to better restoration of natural tibiofemoral kinematics compared to mechanic alignment. Knee Surg Sports Traumatol Arthrosc 27(5):1427–1433CrossRef
19.
Zurück zum Zitat McEwen P, Balendra G, Doma K (2019) Medial and lateral gap laxity differential in computer-assisted kinematic total knee arthroplasty. Bone Jt J 101-B(3):331–339CrossRef McEwen P, Balendra G, Doma K (2019) Medial and lateral gap laxity differential in computer-assisted kinematic total knee arthroplasty. Bone Jt J 101-B(3):331–339CrossRef
20.
Zurück zum Zitat Murakami K, Hamai S, Okazaki K, Ikebe S, Nakahara H, Higaki H et al (2017) Kinematic analysis of stair climbing in rotating platform cruciate-retaining and posterior-stabilized mobile-bearing total knee arthroplasties. Arch Orthop Trauma Surg 137(5):701–711CrossRef Murakami K, Hamai S, Okazaki K, Ikebe S, Nakahara H, Higaki H et al (2017) Kinematic analysis of stair climbing in rotating platform cruciate-retaining and posterior-stabilized mobile-bearing total knee arthroplasties. Arch Orthop Trauma Surg 137(5):701–711CrossRef
21.
Zurück zum Zitat Nakamura S, Ito H, Yoshitomi H, Kuriyama S, Komistek RD, Matsuda S (2015) Analysis of the flexion gap on in vivo knee kinematics using fluoroscopy. J Arthroplasty 30(7):1237–1242CrossRef Nakamura S, Ito H, Yoshitomi H, Kuriyama S, Komistek RD, Matsuda S (2015) Analysis of the flexion gap on in vivo knee kinematics using fluoroscopy. J Arthroplasty 30(7):1237–1242CrossRef
22.
Zurück zum Zitat Nam D, Nunley RM, Barrack RL (2014) Patient dissatisfaction following total knee replacement: a growing concern? Bone Jt J 96-B(11 Supple A):96–100CrossRef Nam D, Nunley RM, Barrack RL (2014) Patient dissatisfaction following total knee replacement: a growing concern? Bone Jt J 96-B(11 Supple A):96–100CrossRef
23.
Zurück zum Zitat Nedopil AJ, Singh AK, Howell SM, Hull ML (2018) Does calipered kinematically aligned TKA restore native left to right symmetry of the lower limb and improve function? J Arthroplasty 33(2):398–406CrossRef Nedopil AJ, Singh AK, Howell SM, Hull ML (2018) Does calipered kinematically aligned TKA restore native left to right symmetry of the lower limb and improve function? J Arthroplasty 33(2):398–406CrossRef
25.
Zurück zum Zitat Park JK, Seon JK, Cho KJ, Lee NH, Song EK (2018) Is immediate postoperative mechanical axis associated with the revision rate of primary total knee arthroplasty? A 10-year follow-up study. Clin Orthop Surg 10(2):167–173CrossRef Park JK, Seon JK, Cho KJ, Lee NH, Song EK (2018) Is immediate postoperative mechanical axis associated with the revision rate of primary total knee arthroplasty? A 10-year follow-up study. Clin Orthop Surg 10(2):167–173CrossRef
26.
Zurück zum Zitat Planckaert C, Larose G, Ranger P, Lacelle M, Fuentes A, Hagemeister N (2018) Total knee arthroplasty with unexplained pain: new insights from kinematics. Arch Orthop Trauma Surg 138(4):553–561CrossRef Planckaert C, Larose G, Ranger P, Lacelle M, Fuentes A, Hagemeister N (2018) Total knee arthroplasty with unexplained pain: new insights from kinematics. Arch Orthop Trauma Surg 138(4):553–561CrossRef
27.
Zurück zum Zitat Roth JD, Howell SM, Hull ML (2019) Analysis of differences in laxities and neutral positions from native after kinematically aligned TKA using cruciate retaining implants. J Orthop Res 37(2):358–369CrossRef Roth JD, Howell SM, Hull ML (2019) Analysis of differences in laxities and neutral positions from native after kinematically aligned TKA using cruciate retaining implants. J Orthop Res 37(2):358–369CrossRef
28.
Zurück zum Zitat Song SJ, Park CH, Bae DK (2019) What to know for selecting cruciate-retaining or posterior-stabilized total knee arthroplasty. Clin Orthop Surg 11(2):142–150CrossRef Song SJ, Park CH, Bae DK (2019) What to know for selecting cruciate-retaining or posterior-stabilized total knee arthroplasty. Clin Orthop Surg 11(2):142–150CrossRef
29.
Zurück zum Zitat Thienpont E, Schwab PE, Cornu O, Bellemans J, Victor J (2017) Bone morphotypes of the varus and valgus knee. Arch Orthop Trauma Surg 137(3):393–400CrossRef Thienpont E, Schwab PE, Cornu O, Bellemans J, Victor J (2017) Bone morphotypes of the varus and valgus knee. Arch Orthop Trauma Surg 137(3):393–400CrossRef
30.
Zurück zum Zitat Tsukiyama H, Kuriyama S, Kobayashi M, Nakamura S, Furu M, Ito H et al (2017) Medial rather than lateral knee instability correlates with inferior patient satisfaction and knee function after total knee arthroplasty. Knee 24(6):1478–1484CrossRef Tsukiyama H, Kuriyama S, Kobayashi M, Nakamura S, Furu M, Ito H et al (2017) Medial rather than lateral knee instability correlates with inferior patient satisfaction and knee function after total knee arthroplasty. Knee 24(6):1478–1484CrossRef
31.
Zurück zum Zitat Tuecking LR, Savov P, Windhagen H, Ettinger M (2020) Current evidence base for kinematic alignment. Orthopade 49(7):584–592CrossRef Tuecking LR, Savov P, Windhagen H, Ettinger M (2020) Current evidence base for kinematic alignment. Orthopade 49(7):584–592CrossRef
32.
Zurück zum Zitat Wickiewicz TL, Roy RR, Powell PL, Edgerton VR (1983) Muscle architecture of the human lower limb. Clin Orthop Relat Res 179:275–283CrossRef Wickiewicz TL, Roy RR, Powell PL, Edgerton VR (1983) Muscle architecture of the human lower limb. Clin Orthop Relat Res 179:275–283CrossRef
33.
Zurück zum Zitat Yoshiya S, Matsui N, Komistek RD, Dennis DA, Mahfouz M, Kurosaka M (2005) In vivo kinematic comparison of posterior cruciate-retaining and posterior stabilized total knee arthroplasties under passive and weight-bearing conditions. J Arthroplasty 20(6):777–783CrossRef Yoshiya S, Matsui N, Komistek RD, Dennis DA, Mahfouz M, Kurosaka M (2005) In vivo kinematic comparison of posterior cruciate-retaining and posterior stabilized total knee arthroplasties under passive and weight-bearing conditions. J Arthroplasty 20(6):777–783CrossRef
Metadaten
Titel
Posterior stabilized total knee arthroplasty reproduces natural joint laxity compared to normal in kinematically aligned total knee arthroplasty: a matched pair cadaveric study
verfasst von
In Jun Koh
Christen E. Chalmers
Charles C. Lin
Soo Bin Park
Michelle H. McGarry
Thay Q. Lee
Publikationsdatum
10.10.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 1/2021
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-020-03624-y

Weitere Artikel der Ausgabe 1/2021

Archives of Orthopaedic and Trauma Surgery 1/2021 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.