Potential of remote monitoring to prevent sensing and detection failures in implantable cardioverter defibrillators

verfasst von: Tobias Franz Götz, MD, Jochen Proff, Tobias Timmel, Clemens Jilek, Klaus Tiemann, Thorsten Lewalter

Erschienen in: Herzschrittmachertherapie + Elektrophysiologie | Ausgabe 1/2022

Einloggen, um Zugang zu erhalten

Abstract

Background

Sensing malfunction and misinterpretation of intracardiac electrograms (IEGMs) in patients with implantable cardioverter defibrillators (ICDs) may lead to inadequate device activity such as inappropriate shock delivery or unnecessary mode-switching. Remote monitoring has the potential for early detection of sensing malfunction or misclassification and may thus prevent adverse device activity. Therefore, the authors analyzed the amount, nature, and distribution of misclassification in current ICD and cardiac resynchronization therapy defibrillator technology using the device transmissions of the IN-TIME study population.

Methods

All transmitted tachyarrhythmic episodes in the 664 IN-TIME patients, comprising 2214 device-classified atrial fibrillation (DC-AF) episodes lasting ≥ 30 s and 1330 device-classified ventricular tachycardia or fibrillation (DC-VT/VF) episodes, were manually analyzed by two experienced cardiologists.

Results

After evaluation of all DC-VT/VF episodes, a total of 300 VT/VF events (23.1%) were false-positive, with supraventricular tachycardia being the most frequent cause (51.7%), followed by atrial fibrillation (21.3%) and T‑wave oversensing (21.0%). A total of 15 patients with false-positive DC-VT/VF received inappropriate shocks. According to the inclusion criteria, 616 IEGMs with DC-AF were assessed. A total of 19.7% were false-positive AF episodes and R‑wave oversensing was the most common reason (55.9%).

Conclusions

Remote monitoring offers the opportunity of early detection of signal misclassification and thus early prevention of adverse device reaction, such as inappropriate shock delivery or mode-switching with intermittent loss of atrioventricular synchrony, by correcting the underlying causes.
Literatur
1.
Zurück zum Zitat Boriani G, Da Costa A, Quesada A et al (2017) Effects of remote monitoring on clinical outcomes and use of healthcare resources in heart failure patients with biventricular defibrillators: results of the MORE-CARE multicentre randomized controlled trial. Eur J Heart Fail 19:416–425CrossRef Boriani G, Da Costa A, Quesada A et al (2017) Effects of remote monitoring on clinical outcomes and use of healthcare resources in heart failure patients with biventricular defibrillators: results of the MORE-CARE multicentre randomized controlled trial. Eur J Heart Fail 19:416–425CrossRef
2.
Zurück zum Zitat Burri H, Senouf D (2009) Remote monitoring and follow-up of pacemakers and implantable cardioverter defibrillators. Europace 11:701–709CrossRef Burri H, Senouf D (2009) Remote monitoring and follow-up of pacemakers and implantable cardioverter defibrillators. Europace 11:701–709CrossRef
3.
Zurück zum Zitat Cheng A, Landman SR, Stadler RW (2012) Reasons for loss of cardiac resynchronization therapy pacing: insights from 32 844 patients. Circ Arrhythm Electrophysiol 5:884–888CrossRef Cheng A, Landman SR, Stadler RW (2012) Reasons for loss of cardiac resynchronization therapy pacing: insights from 32 844 patients. Circ Arrhythm Electrophysiol 5:884–888CrossRef
4.
Zurück zum Zitat Daubert JP, Zareba W, Cannom DS et al (2008) Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact. J Am Coll Cardiol 51:1357–1365CrossRef Daubert JP, Zareba W, Cannom DS et al (2008) Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact. J Am Coll Cardiol 51:1357–1365CrossRef
5.
Zurück zum Zitat Gasparini M, Proclemer A, Klersy C et al (2013) Effect of long-detection interval vs standard-detection interval for implantable cardioverter-defibrillators on antitachycardia pacing and shock delivery: the ADVANCE III randomized clinical trial. JAMA 309:1903–1911CrossRef Gasparini M, Proclemer A, Klersy C et al (2013) Effect of long-detection interval vs standard-detection interval for implantable cardioverter-defibrillators on antitachycardia pacing and shock delivery: the ADVANCE III randomized clinical trial. JAMA 309:1903–1911CrossRef
6.
Zurück zum Zitat Hawkins NM, Virani SA, Sperrin M et al (2016) Predicting heart failure decompensation using cardiac implantable electronic devices: a review of practices and challenges. Eur J Heart Fail 18:977–986CrossRef Hawkins NM, Virani SA, Sperrin M et al (2016) Predicting heart failure decompensation using cardiac implantable electronic devices: a review of practices and challenges. Eur J Heart Fail 18:977–986CrossRef
7.
Zurück zum Zitat Hayes DL, Boehmer JP, Day JD et al (2011) Cardiac resynchronization therapy and the relationship of percent biventricular pacing to symptoms and survival. Heart Rhythm 8:1469–1475CrossRef Hayes DL, Boehmer JP, Day JD et al (2011) Cardiac resynchronization therapy and the relationship of percent biventricular pacing to symptoms and survival. Heart Rhythm 8:1469–1475CrossRef
8.
Zurück zum Zitat Healey JS, Connolly SJ, Gold MR et al (2012) Subclinical atrial fibrillation and the risk of stroke. N Engl J Med 366:120–129CrossRef Healey JS, Connolly SJ, Gold MR et al (2012) Subclinical atrial fibrillation and the risk of stroke. N Engl J Med 366:120–129CrossRef
9.
Zurück zum Zitat Hindricks G, Taborsky M, Glikson M et al (2014) Implant-based multiparameter telemonitoring of patients with heart failure (IN-TIME): a randomised controlled trial. Lancet 384:583–590CrossRef Hindricks G, Taborsky M, Glikson M et al (2014) Implant-based multiparameter telemonitoring of patients with heart failure (IN-TIME): a randomised controlled trial. Lancet 384:583–590CrossRef
10.
Zurück zum Zitat Jung W, Rillig A, Birkemeyer R et al (2008) Advances in remote monitoring of implantable pacemakers, cardioverter defibrillators and cardiac resynchronization therapy systems. J Interv Card Electrophysiol 23:73–85CrossRef Jung W, Rillig A, Birkemeyer R et al (2008) Advances in remote monitoring of implantable pacemakers, cardioverter defibrillators and cardiac resynchronization therapy systems. J Interv Card Electrophysiol 23:73–85CrossRef
11.
Zurück zum Zitat Kaufman ES, Israel CW, Nair GM et al (2012) Positive predictive value of device-detected atrial high-rate episodes at different rates and durations: an analysis from ASSERT. Heart Rhythm 9:1241–1246CrossRef Kaufman ES, Israel CW, Nair GM et al (2012) Positive predictive value of device-detected atrial high-rate episodes at different rates and durations: an analysis from ASSERT. Heart Rhythm 9:1241–1246CrossRef
12.
Zurück zum Zitat Leclercq C, Kass DA (2002) Retiming the failing heart: principles and current clinical status of cardiac resynchronization. J Am Coll Cardiol 39:194–201CrossRef Leclercq C, Kass DA (2002) Retiming the failing heart: principles and current clinical status of cardiac resynchronization. J Am Coll Cardiol 39:194–201CrossRef
14.
Zurück zum Zitat Lunati M, Centres CC, Proclemer A et al (2016) Reduction of inappropriate anti-tachycardia pacing therapies and shocks by a novel suite of detection algorithms in heart failure patients with cardiac resynchronization therapy defibrillators: a historical comparison of a prospective database. Europace 18:1391–1398CrossRef Lunati M, Centres CC, Proclemer A et al (2016) Reduction of inappropriate anti-tachycardia pacing therapies and shocks by a novel suite of detection algorithms in heart failure patients with cardiac resynchronization therapy defibrillators: a historical comparison of a prospective database. Europace 18:1391–1398CrossRef
15.
Zurück zum Zitat Morgan JM, Kitt S, Gill J et al (2017) Remote management of heart failure using implantable electronic devices. Eur Heart J 38:2352–2360CrossRef Morgan JM, Kitt S, Gill J et al (2017) Remote management of heart failure using implantable electronic devices. Eur Heart J 38:2352–2360CrossRef
16.
Zurück zum Zitat Moss AJ, Schuger C, Beck CA et al (2012) Reduction in inappropriate therapy and mortality through ICD programming. N Engl J Med 367:2275–2283CrossRef Moss AJ, Schuger C, Beck CA et al (2012) Reduction in inappropriate therapy and mortality through ICD programming. N Engl J Med 367:2275–2283CrossRef
17.
Zurück zum Zitat Nishimura RA, Hayes DL, Holmes DR et al (1995) Mechanism of hemodynamic improvement by dual-chamber pacing for severe left ventricular dysfunction: an acute Doppler and catheterization hemodynamic study. J Am Coll Cardiol 25:281–288CrossRef Nishimura RA, Hayes DL, Holmes DR et al (1995) Mechanism of hemodynamic improvement by dual-chamber pacing for severe left ventricular dysfunction: an acute Doppler and catheterization hemodynamic study. J Am Coll Cardiol 25:281–288CrossRef
18.
Zurück zum Zitat Pollak WM, Simmons JD, Interian A Jr et al (2001) Clinical utility of intraatrial pacemaker stored electrograms to diagnose atrial fibrillation and flutter. Pacing Clin Electrophysiol 24:424–429CrossRef Pollak WM, Simmons JD, Interian A Jr et al (2001) Clinical utility of intraatrial pacemaker stored electrograms to diagnose atrial fibrillation and flutter. Pacing Clin Electrophysiol 24:424–429CrossRef
19.
Zurück zum Zitat Poole JE, Johnson GW, Hellkamp AS et al (2008) Prognostic importance of defibrillator shocks in patients with heart failure. N Engl J Med 359:1009–1017CrossRef Poole JE, Johnson GW, Hellkamp AS et al (2008) Prognostic importance of defibrillator shocks in patients with heart failure. N Engl J Med 359:1009–1017CrossRef
20.
Zurück zum Zitat Ricci RP, Pignalberi C, Landolina M et al (2014) Ventricular rate monitoring as a tool to predict and prevent atrial fibrillation-related inappropriate shocks in heart failure patients treated with cardiac resynchronisation therapy defibrillators. Heart 100:848–854CrossRef Ricci RP, Pignalberi C, Landolina M et al (2014) Ventricular rate monitoring as a tool to predict and prevent atrial fibrillation-related inappropriate shocks in heart failure patients treated with cardiac resynchronisation therapy defibrillators. Heart 100:848–854CrossRef
21.
Zurück zum Zitat Ruiz-Granell R, Dovellini EV, Dompnier A et al (2019) Algorithm-based reduction of inappropriate defibrillator shock: results of the inappropriate shock reduction wIth PARAD+ rhythm discrimination—implantable cardioverter defibrillator study. Heart Rhythm 16:1429–1435CrossRef Ruiz-Granell R, Dovellini EV, Dompnier A et al (2019) Algorithm-based reduction of inappropriate defibrillator shock: results of the inappropriate shock reduction wIth PARAD+ rhythm discrimination—implantable cardioverter defibrillator study. Heart Rhythm 16:1429–1435CrossRef
22.
Zurück zum Zitat Saxon LA, Hayes DL, Gilliam FR et al (2010) Long-term outcome after ICD and CRT implantation and influence of remote device follow-up: the ALTITUDE survival study. Circulation 122:2359–2367CrossRef Saxon LA, Hayes DL, Gilliam FR et al (2010) Long-term outcome after ICD and CRT implantation and influence of remote device follow-up: the ALTITUDE survival study. Circulation 122:2359–2367CrossRef
23.
Zurück zum Zitat Slotwiner D, Varma N, Akar JG et al (2015) HRS expert consensus statement on remote interrogation and monitoring for cardiovascular implantable electronic devices. Heart Rhythm 12:e69–e100CrossRef Slotwiner D, Varma N, Akar JG et al (2015) HRS expert consensus statement on remote interrogation and monitoring for cardiovascular implantable electronic devices. Heart Rhythm 12:e69–e100CrossRef
24.
Zurück zum Zitat Sweeney MO, Wathen MS, Volosin K et al (2005) Appropriate and inappropriate ventricular therapies, quality of life, and mortality among primary and secondary prevention implantable cardioverter defibrillator patients: results from the Pacing Fast VT REduces Shock ThErapies (PainFREE Rx II) trial. Circulation 111:2898–2905CrossRef Sweeney MO, Wathen MS, Volosin K et al (2005) Appropriate and inappropriate ventricular therapies, quality of life, and mortality among primary and secondary prevention implantable cardioverter defibrillator patients: results from the Pacing Fast VT REduces Shock ThErapies (PainFREE Rx II) trial. Circulation 111:2898–2905CrossRef
25.
Zurück zum Zitat Swerdlow CD, Asirvatham SJ, Ellenbogen KA et al (2015) Troubleshooting implantable cardioverter-defibrillator sensing problems II. Circ Arrhythm Electrophysiol 8:212–220CrossRef Swerdlow CD, Asirvatham SJ, Ellenbogen KA et al (2015) Troubleshooting implantable cardioverter-defibrillator sensing problems II. Circ Arrhythm Electrophysiol 8:212–220CrossRef
26.
Zurück zum Zitat Swerdlow CD, Asirvatham SJ, Ellenbogen KA et al (2014) Troubleshooting implanted cardioverter defibrillator sensing problems I. Circ Arrhythm Electrophysiol 7:1237–1261CrossRef Swerdlow CD, Asirvatham SJ, Ellenbogen KA et al (2014) Troubleshooting implanted cardioverter defibrillator sensing problems I. Circ Arrhythm Electrophysiol 7:1237–1261CrossRef
27.
Zurück zum Zitat Van Rees JB, Borleffs CJW, De Bie MK et al (2011) Inappropriate implantable cardioverter-defibrillator shocks: incidence, predictors, and impact on mortality. J Am Coll Cardiol 57:556–562CrossRef Van Rees JB, Borleffs CJW, De Bie MK et al (2011) Inappropriate implantable cardioverter-defibrillator shocks: incidence, predictors, and impact on mortality. J Am Coll Cardiol 57:556–562CrossRef
28.
Zurück zum Zitat Varma N, Wilkoff B (2011) Device features for managing patients with heart failure. Heart Fail Clin 7:215–225CrossRef Varma N, Wilkoff B (2011) Device features for managing patients with heart failure. Heart Fail Clin 7:215–225CrossRef
29.
Zurück zum Zitat Wilkoff BL, DAVID Trial Investigators (2003) The Dual Chamber and VVI Implantable Defibrillator (DAVID) trial: rationale, design, results, clinical implications and lessons for future trials. Card Electrophysiol Rev 7:468–472CrossRef Wilkoff BL, DAVID Trial Investigators (2003) The Dual Chamber and VVI Implantable Defibrillator (DAVID) trial: rationale, design, results, clinical implications and lessons for future trials. Card Electrophysiol Rev 7:468–472CrossRef
30.
Zurück zum Zitat Wilkoff BL, Ousdigian KT, Sterns LD et al (2006) A comparison of empiric to physician-tailored programming of implantable cardioverter-defibrillators: results from the prospective randomized multicenter EMPIRIC trial. J Am Coll Cardiol 48:330–339CrossRef Wilkoff BL, Ousdigian KT, Sterns LD et al (2006) A comparison of empiric to physician-tailored programming of implantable cardioverter-defibrillators: results from the prospective randomized multicenter EMPIRIC trial. J Am Coll Cardiol 48:330–339CrossRef
Metadaten
Titel
Potential of remote monitoring to prevent sensing and detection failures in implantable cardioverter defibrillators
verfasst von
Tobias Franz Götz, MD
Jochen Proff
Tobias Timmel
Clemens Jilek
Klaus Tiemann
Thorsten Lewalter
Publikationsdatum
01.09.2021
Verlag
Springer Medizin
Erschienen in
Herzschrittmachertherapie + Elektrophysiologie / Ausgabe 1/2022
Print ISSN: 0938-7412
Elektronische ISSN: 1435-1544
DOI
https://doi.org/10.1007/s00399-021-00802-2

Weitere Artikel der Ausgabe 1/2022

Herzschrittmachertherapie + Elektrophysiologie 1/2022 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.