Skip to main content
main-content

06.12.2018 | Original Article | Ausgabe 2/2019

Japanese Journal of Radiology 2/2019

Potential value of the PixelShine deep learning algorithm for increasing quality of 70 kVp+ASiR-V reconstruction pelvic arterial phase CT images

Zeitschrift:
Japanese Journal of Radiology > Ausgabe 2/2019
Autoren:
Shi-feng Tian, Ai-lian Liu, Jing-hong Liu, Yi-jun Liu, Ju-dong Pan

Abstract

Objective

To investigate the effect of a deep learning-based denoising algorithm, PixelShine (PS), on the quality of 70 kVp pelvic arterial phase CT images.

Materials and methods

A retrospective analysis was performed on arterial phase pelvic CT images from 33 patients (body-mass index ≤ 20 kg/m2) obtained with a GE Revolution CT (70 kVp tube voltage; adaptive statistical iterative reconstruction-Veo-filtered back projection, 50% blending) and designated group A. Group B images were then obtained by applying PS to group A image datasets. Subjective image quality was evaluated by two radiologists with a 5-point scoring system; the scores of the groups were compared. Image signal was assessed using CT values of the urinary bladder. CT and standard deviation (SD) values of the gluteus maximus were measured, and SD values of the gluteus maximus were used to represent image noise. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of the bladder were calculated. Image noise, SNR, and CNR of two groups were compared using paired t-tests.

Results

The subjective visual image quality scores of groups A and B, respectively, were 3.11 ± 0.30 vs. 3.82 ± 0.57; image noise was 15.79 ± 2.05 Hounsfield units (HU) vs. 11.06 ± 2.22 HU; SNRs of bladder were 0.50 ± 0.23 vs. 0.79 ± 0.39; and CNRs of bladder were 3.72 ± 0.85 vs. 5.14 ± 1.27. Group B showed better subjective image quality, lower image noise, and improved SNR and CNR, compared to group A; these differences were statistically significant (P < 0.05). The noise of group B was approximately 30% lower than that of group A; the SNR and CNR values of group B were improved by approximately 58% and 38%, respectively.

Conclusion

Using 70 kVp +ASiR-V, PS can improve the image quality of pelvic arterial phase CT images, significantly reduce the image noise, and improve the SNR and CNR.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag als Mediziner

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 2/2019

Japanese Journal of Radiology 2/2019 Zur Ausgabe
  1. Sie können e.Med Radiologie 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.

Neu im Fachgebiet Radiologie

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Radiologie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise