Pre-clinical study on a telemetric gastric sensor for recognition of acute upper gastrointestinal bleeding: the “HemoPill monitor”
verfasst von:
Sebastian Schostek, Melanie Zimmermann, Jan Keller, Mario Fode, Michael Melbert, Ruediger L. Prosst, Thomas Gottwald, Marc O. Schurr
Erschienen in:
Surgical Endoscopy
|
Ausgabe 2/2020
Einloggen, um Zugang zu erhalten
Abstract
Background
Acute upper gastrointestinal bleeding is a life-threatening medical condition with a relevant risk of re-bleeding even after initial endoscopic hemostasis. The implantable HemoPill monitor contains a novel telemetric sensor to optically detect blood in the stomach allowing the surveillance of high-risk patients for re-bleedings.
Methods
In this pre-clinical porcine study, bleeding has been simulated by injecting porcine blood into the stomach of a pig through an implanted catheter using a syringe pump. The effect of the sensor position in the stomach, the gastric food content, and the bleeding intensity was investigated.
Results
Sensitivity and specificity of the sensor reached more than 87.5% when the sensor was positioned close to the source of bleeding. Solid food had a higher negative impact on sensitivity than liquid food but a positive impact on specificity. A heavy bleeding was more likely to be detected by the sensor but was also associated with a lower likelihood for true-negative results than weaker bleedings.
Conclusions
The study clearly demonstrated the capability of the HemoPill sensor prototype to detect clinically relevant bleedings with high sensitivity and specificity (> 80%) when the sensor was positioned close to the bleeding site. The sensors proved to be robust against artefact effects from stomach content. These are favorable findings that underline the potential benefit for the use of the HemoPill sensor in monitoring patients with a risk of re-bleeding in the upper gastrointestinal tract.