Primary stability of rotator cuff repair: can more suture materials yield more strength?

verfasst von: Mehmet Gülecyüz, Hannes Bortolotti, Matthias Pietschmann, Andreas Ficklscherer, Thomas Niethammer, Björn Roßbach, Peter Müller

Erschienen in: International Orthopaedics | Ausgabe 5/2016

Einloggen, um Zugang zu erhalten

Abstract

Introduction

It remains unclear if the increased biomechanical stability of the double-row repair is solely a result of the technique itself or due to the greater number of utilized materials. This study analyses the various rotator cuff (RC) repair interfaces and the influence of doubling the number of suture knots, stitch techniques and suture anchors on the primary stability of the RC reconstruction.

Materials and methods

Established cyclic testing protocols were used to test the primary stability of the knot interface with a Duncan loop and Roeder knot, the tendon-to-suture interface with a single-stitch and Mason-Allen-stitch and the anchor-to-bone interface with OPUS Magnum 2® and BioCorkscrew® anchors. After doubling the materials, the interfaces were tested again and compared to the yielded solitary values. The maximum failure loads (Fmax), clinical failure (CF), system displacement (SD) and modes of failure were recorded.

Results

Doubling the amount of knots, stitches and suture anchors improved almost all interface parameters. For example, Fmax, CF and SD of the Roeder knot could be significantly improved from 142.5 N to 256.7 N, from 82.5 N to 155.0 N and from 0.45 mm to 0.20 mm, respectively. The Mason-Allen stitch’s Fmax improved from 196.4 N to 324.9 N, and CF from 94.25 N to 139.8 N, with almost identical SD of 0.81 mm and 0.84 mm. Bio-Corkscrew® Fmax improved from 232.5 N to 317.5 N, CF from 182.5 N to 210 N, and SD from 0.85 mm to 0.64 mm.

Conclusion

Doubling the number of knots, stitches and suture anchors increases the primary stability of the various RC repair interfaces and may result in RC rerupture rate reduction. Furthermore, this study suggests that a suture anchor loaded with two or three sutures may yield the same primary stability as two or three suture anchors loaded with a single suture. This in vitro biomechanical study focuses solely on the rotator cuff interfaces at time zero; the biological healing process was therefore not analysed and requires further investigation.

Clinical relevance

Improved primary stability by increased number of suture materials may reduce rate of RC rerupture.
Literatur
1.
Zurück zum Zitat Tashjian RZ, Henn RF, Kang L, Green A (2004) The effect of comorbidity on self-assessed function in patients with a chronic rotator cuff tear. J Bone Joint Surg Am 86-A(2):355–362PubMed Tashjian RZ, Henn RF, Kang L, Green A (2004) The effect of comorbidity on self-assessed function in patients with a chronic rotator cuff tear. J Bone Joint Surg Am 86-A(2):355–362PubMed
3.
Zurück zum Zitat Cummins CA, Murrell GA (2003) Mode of failure for rotator cuff repair with suture anchors identified at revision surgery. J Shoulder Elb Surg 12(2):128–133CrossRef Cummins CA, Murrell GA (2003) Mode of failure for rotator cuff repair with suture anchors identified at revision surgery. J Shoulder Elb Surg 12(2):128–133CrossRef
4.
Zurück zum Zitat Boileau P, Brassart N, Watkinson DJ, Carles M, Hatzidakis AM, Krishnan SG (2005) Arthroscopic repair of full-thickness tears of the supraspinatus: does the tendon really heal? J Bone Joint Surg Am 87(6):1229–1240. doi:10.2106/JBJS.D.02035 CrossRefPubMed Boileau P, Brassart N, Watkinson DJ, Carles M, Hatzidakis AM, Krishnan SG (2005) Arthroscopic repair of full-thickness tears of the supraspinatus: does the tendon really heal? J Bone Joint Surg Am 87(6):1229–1240. doi:10.​2106/​JBJS.​D.​02035 CrossRefPubMed
5.
Zurück zum Zitat Frank JB, ElAttrache NS, Dines JS, Blackburn A, Crues J, Tibone JE (2008) Repair site integrity after arthroscopic transosseous-equivalent suture-bridge rotator cuff repair. Am J Sports Med 36(8):1496–1503. doi:10.1177/0363546507313574 CrossRefPubMed Frank JB, ElAttrache NS, Dines JS, Blackburn A, Crues J, Tibone JE (2008) Repair site integrity after arthroscopic transosseous-equivalent suture-bridge rotator cuff repair. Am J Sports Med 36(8):1496–1503. doi:10.​1177/​0363546507313574​ CrossRefPubMed
6.
Zurück zum Zitat Kamath G, Galatz LM, Keener JD, Teefey S, Middleton W, Yamaguchi K (2009) Tendon integrity and functional outcome after arthroscopic repair of high-grade partial-thickness supraspinatus tears. J Bone Joint Surg Am 91(5):1055–1062. doi:10.2106/JBJS.G.00118 CrossRefPubMed Kamath G, Galatz LM, Keener JD, Teefey S, Middleton W, Yamaguchi K (2009) Tendon integrity and functional outcome after arthroscopic repair of high-grade partial-thickness supraspinatus tears. J Bone Joint Surg Am 91(5):1055–1062. doi:10.​2106/​JBJS.​G.​00118 CrossRefPubMed
7.
Zurück zum Zitat Sugaya H, Maeda K, Matsuki K, Moriishi J (2007) Repair integrity and functional outcome after arthroscopic double-row rotator cuff repair. A prospective outcome study. J Bone Joint Surg Am 89(5):953–960. doi:10.2106/JBJS.F.00512 CrossRefPubMed Sugaya H, Maeda K, Matsuki K, Moriishi J (2007) Repair integrity and functional outcome after arthroscopic double-row rotator cuff repair. A prospective outcome study. J Bone Joint Surg Am 89(5):953–960. doi:10.​2106/​JBJS.​F.​00512 CrossRefPubMed
8.
Zurück zum Zitat Zumstein MA, Jost B, Hempel J, Hodler J, Gerber C (2008) The clinical and structural long-term results of open repair of massive tears of the rotator cuff. J Bone Joint Surg Am 90(11):2423–2431. doi:10.2106/JBJS.G.00677 CrossRefPubMed Zumstein MA, Jost B, Hempel J, Hodler J, Gerber C (2008) The clinical and structural long-term results of open repair of massive tears of the rotator cuff. J Bone Joint Surg Am 90(11):2423–2431. doi:10.​2106/​JBJS.​G.​00677 CrossRefPubMed
9.
Zurück zum Zitat Meier SW, Meier JD (2006) Rotator cuff repair: the effect of double-row fixation on three-dimensional repair site. J Shoulder Elb Surg 15(6):691–696CrossRef Meier SW, Meier JD (2006) Rotator cuff repair: the effect of double-row fixation on three-dimensional repair site. J Shoulder Elb Surg 15(6):691–696CrossRef
10.
Zurück zum Zitat Brady PC, Arrigoni P, Burkhart SS (2006) Evaluation of residual rotator cuff defects after in vivo single- versus double-row rotator cuff repairs. Arthroscopy 22(10):1070–1075CrossRefPubMed Brady PC, Arrigoni P, Burkhart SS (2006) Evaluation of residual rotator cuff defects after in vivo single- versus double-row rotator cuff repairs. Arthroscopy 22(10):1070–1075CrossRefPubMed
11.
Zurück zum Zitat Kim DH, Elattrache NS, Tibone JE, Jun BJ, DeLaMora SN, Kvitne RS, Lee TQ (2006) Biomechanical comparison of a single-row versus double-row suture anchor technique for rotator cuff repair. Am J Sports Med 34(3):407–414CrossRefPubMed Kim DH, Elattrache NS, Tibone JE, Jun BJ, DeLaMora SN, Kvitne RS, Lee TQ (2006) Biomechanical comparison of a single-row versus double-row suture anchor technique for rotator cuff repair. Am J Sports Med 34(3):407–414CrossRefPubMed
12.
Zurück zum Zitat Wall LB, Keener JD, Brophy RH (2009) Double-row vs single-row rotator cuff repair: a review of the biomechanical evidence. J Shoulder Elb Surg 18(6):933–941CrossRef Wall LB, Keener JD, Brophy RH (2009) Double-row vs single-row rotator cuff repair: a review of the biomechanical evidence. J Shoulder Elb Surg 18(6):933–941CrossRef
14.
Zurück zum Zitat Chen M, Xu W, Dong Q, Huang Q, Xie Z, Mao Y (2013) Outcomes of single-row versus double-row arthroscopic rotator cuff repair: a systematic review and meta-analysis of current evidence. Arthroscopy 29(8):1437–1449CrossRefPubMed Chen M, Xu W, Dong Q, Huang Q, Xie Z, Mao Y (2013) Outcomes of single-row versus double-row arthroscopic rotator cuff repair: a systematic review and meta-analysis of current evidence. Arthroscopy 29(8):1437–1449CrossRefPubMed
15.
Zurück zum Zitat Lapner PL, Sabri E, Rakhra K, McRae S, Leiter J, Bell K, Macdonald P (2012) A multicenter randomized controlled trial comparing single-row with double-row fixation in arthroscopic rotator cuff repair. J Bone Joint Surg Am 94(14):1249–1257CrossRefPubMed Lapner PL, Sabri E, Rakhra K, McRae S, Leiter J, Bell K, Macdonald P (2012) A multicenter randomized controlled trial comparing single-row with double-row fixation in arthroscopic rotator cuff repair. J Bone Joint Surg Am 94(14):1249–1257CrossRefPubMed
16.
Zurück zum Zitat McCormick F, Gupta A, Bruce B, Harris J, Abrams G, Wilson H, Hussey K, Cole BJ (2014) Single-row, double-row, and transosseous equivalent techniques for isolated supraspinatus tendon tears with minimal atrophy: a retrospective comparative outcome and radiographic analysis at minimum 2-year followup. Int J Shoulder Surg 8(1):15–20. doi:10.4103/0973-6042.131850 CrossRefPubMedPubMedCentral McCormick F, Gupta A, Bruce B, Harris J, Abrams G, Wilson H, Hussey K, Cole BJ (2014) Single-row, double-row, and transosseous equivalent techniques for isolated supraspinatus tendon tears with minimal atrophy: a retrospective comparative outcome and radiographic analysis at minimum 2-year followup. Int J Shoulder Surg 8(1):15–20. doi:10.​4103/​0973-6042.​131850 CrossRefPubMedPubMedCentral
17.
Zurück zum Zitat Millett PJ, Warth RJ, Dornan GJ, Lee JT, Spiegl UJ (2014) Clinical and structural outcomes after arthroscopic single-row versus double-row rotator cuff repair: a systematic review and meta-analysis of level I randomized clinical trials. J Shoulder Elb Surg 23(4):586–597CrossRef Millett PJ, Warth RJ, Dornan GJ, Lee JT, Spiegl UJ (2014) Clinical and structural outcomes after arthroscopic single-row versus double-row rotator cuff repair: a systematic review and meta-analysis of level I randomized clinical trials. J Shoulder Elb Surg 23(4):586–597CrossRef
20.
Zurück zum Zitat Jost PW, Khair MM, Chen DX, Wright TM, Kelly AM, Rodeo SA (2012) Suture number determines strength of rotator cuff repair. J Bone Joint Surg Am 94(14):e100CrossRefPubMed Jost PW, Khair MM, Chen DX, Wright TM, Kelly AM, Rodeo SA (2012) Suture number determines strength of rotator cuff repair. J Bone Joint Surg Am 94(14):e100CrossRefPubMed
21.
Zurück zum Zitat Lo IK, Burkhart SS, Chan KC, Athanasiou K (2004) Arthroscopic knots: determining the optimal balance of loop security and knot security. Arthroscopy 20(5):489–502CrossRefPubMed Lo IK, Burkhart SS, Chan KC, Athanasiou K (2004) Arthroscopic knots: determining the optimal balance of loop security and knot security. Arthroscopy 20(5):489–502CrossRefPubMed
22.
Zurück zum Zitat Pietschmann MF, Sadoghi P, Hauser E, Scharpf A, Gulecyuz MF, Schroder C, Jansson V, Muller PE (2011) Influence of testing conditions on primary stability of arthroscopic knot tying for rotator cuff repair: slippery when wet? Arthroscopy 27(12):1628–1636CrossRefPubMed Pietschmann MF, Sadoghi P, Hauser E, Scharpf A, Gulecyuz MF, Schroder C, Jansson V, Muller PE (2011) Influence of testing conditions on primary stability of arthroscopic knot tying for rotator cuff repair: slippery when wet? Arthroscopy 27(12):1628–1636CrossRefPubMed
23.
Zurück zum Zitat Baumgarten KM, Wright RW (2004) Arthroscopic knot tying: an instruction manual. Lippincott Raven, New York Baumgarten KM, Wright RW (2004) Arthroscopic knot tying: an instruction manual. Lippincott Raven, New York
24.
Zurück zum Zitat Ma CB, MacGillivray JD, Clabeaux J, Lee S, Otis JC (2004) Biomechanical evaluation of arthroscopic rotator cuff stitches. J Bone Joint Surg Am 86-A(6):1211–1216PubMed Ma CB, MacGillivray JD, Clabeaux J, Lee S, Otis JC (2004) Biomechanical evaluation of arthroscopic rotator cuff stitches. J Bone Joint Surg Am 86-A(6):1211–1216PubMed
25.
Zurück zum Zitat Pietschmann MF, Frohlich V, Ficklscherer A, Gulecyuz MF, Wegener B, Jansson V, Muller PE (2009) Suture anchor fixation strength in osteopenic versus non-osteopenic bone for rotator cuff repair. Arch Orthop Trauma Surg 129(3):373–379. doi:10.1007/s00402-008-0689-4 CrossRefPubMed Pietschmann MF, Frohlich V, Ficklscherer A, Gulecyuz MF, Wegener B, Jansson V, Muller PE (2009) Suture anchor fixation strength in osteopenic versus non-osteopenic bone for rotator cuff repair. Arch Orthop Trauma Surg 129(3):373–379. doi:10.​1007/​s00402-008-0689-4 CrossRefPubMed
26.
Zurück zum Zitat Pietschmann MF, Gulecyuz MF, Fieseler S, Hentschel M, Rossbach B, Jansson V, Muller PE (2010) Biomechanical stability of knotless suture anchors used in rotator cuff repair in healthy and osteopenic bone. Arthroscopy 26(8):1035–1044CrossRefPubMed Pietschmann MF, Gulecyuz MF, Fieseler S, Hentschel M, Rossbach B, Jansson V, Muller PE (2010) Biomechanical stability of knotless suture anchors used in rotator cuff repair in healthy and osteopenic bone. Arthroscopy 26(8):1035–1044CrossRefPubMed
27.
Zurück zum Zitat Millett PJ, Warth RJ, Dornan GJ, Lee JT, Spiegl UJ (2014) Clinical and structural outcomes after arthroscopic single-row versus double-row rotator cuff repair: a systematic review and meta-analysis of level I randomized clinical trials. J Shoulder Elb Surg 23(4):586–597. doi:10.1016/j.jse.2013.10.006 CrossRef Millett PJ, Warth RJ, Dornan GJ, Lee JT, Spiegl UJ (2014) Clinical and structural outcomes after arthroscopic single-row versus double-row rotator cuff repair: a systematic review and meta-analysis of level I randomized clinical trials. J Shoulder Elb Surg 23(4):586–597. doi:10.​1016/​j.​jse.​2013.​10.​006 CrossRef
28.
Zurück zum Zitat Burks RT, Crim J, Brown N, Fink B, Greis PE (2009) A prospective randomized clinical trial comparing arthroscopic single- and double-row rotator cuff repair: magnetic resonance imaging and early clinical evaluation. Am J Sports Med 37(4):674–682CrossRefPubMed Burks RT, Crim J, Brown N, Fink B, Greis PE (2009) A prospective randomized clinical trial comparing arthroscopic single- and double-row rotator cuff repair: magnetic resonance imaging and early clinical evaluation. Am J Sports Med 37(4):674–682CrossRefPubMed
29.
Zurück zum Zitat Charousset C, Grimberg J, Duranthon LD, Bellaiche L, Petrover D (2007) Can a double-row anchorage technique improve tendon healing in arthroscopic rotator cuff repair?: a prospective, nonrandomized, comparative study of double-row and single-row anchorage techniques with computed tomographic arthrography tendon healing assessment. Am J Sports Med 35(8):1247–1253. doi:10.1177/0363546507301661 CrossRefPubMed Charousset C, Grimberg J, Duranthon LD, Bellaiche L, Petrover D (2007) Can a double-row anchorage technique improve tendon healing in arthroscopic rotator cuff repair?: a prospective, nonrandomized, comparative study of double-row and single-row anchorage techniques with computed tomographic arthrography tendon healing assessment. Am J Sports Med 35(8):1247–1253. doi:10.​1177/​0363546507301661​ CrossRefPubMed
30.
Zurück zum Zitat Franceschi F, Ruzzini L, Longo UG, Martina FM, Zobel BB, Maffulli N, Denaro V (2007) Equivalent clinical results of arthroscopic single-row and double-row suture anchor repair for rotator cuff tears: a randomized controlled trial. Am J Sports Med 35(8):1254–1260. doi:10.1177/0363546507302218 CrossRefPubMed Franceschi F, Ruzzini L, Longo UG, Martina FM, Zobel BB, Maffulli N, Denaro V (2007) Equivalent clinical results of arthroscopic single-row and double-row suture anchor repair for rotator cuff tears: a randomized controlled trial. Am J Sports Med 35(8):1254–1260. doi:10.​1177/​0363546507302218​ CrossRefPubMed
32.
Zurück zum Zitat Bisson L, Zivaljevic N, Sanders S, Pula D (2015) A cost analysis of single-row versus double-row and suture bridge rotator cuff repair methods. Knee Surg Sports Traumatol Arthrosc 23(2):487–493. doi:10.1007/s00167-012-2338-2 Bisson L, Zivaljevic N, Sanders S, Pula D (2015) A cost analysis of single-row versus double-row and suture bridge rotator cuff repair methods. Knee Surg Sports Traumatol Arthrosc 23(2):487–493. doi:10.​1007/​s00167-012-2338-2
33.
Zurück zum Zitat Genuario JW, Donegan RP, Hamman D, Bell JE, Boublik M, Schlegel T, Tosteson AN (2012) The cost-effectiveness of single-row compared with double-row arthroscopic rotator cuff repair. J Bone Joint Surg Am 94(15):1369–1377. doi:10.2106/JBJS.J.01876 CrossRefPubMed Genuario JW, Donegan RP, Hamman D, Bell JE, Boublik M, Schlegel T, Tosteson AN (2012) The cost-effectiveness of single-row compared with double-row arthroscopic rotator cuff repair. J Bone Joint Surg Am 94(15):1369–1377. doi:10.​2106/​JBJS.​J.​01876 CrossRefPubMed
Metadaten
Titel
Primary stability of rotator cuff repair: can more suture materials yield more strength?
verfasst von
Mehmet Gülecyüz
Hannes Bortolotti
Matthias Pietschmann
Andreas Ficklscherer
Thomas Niethammer
Björn Roßbach
Peter Müller
Publikationsdatum
07.10.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
International Orthopaedics / Ausgabe 5/2016
Print ISSN: 0341-2695
Elektronische ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-015-3012-3

Weitere Artikel der Ausgabe 5/2016

International Orthopaedics 5/2016 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.