Skip to main content
Erschienen in: Acta Diabetologica 4/2016

21.03.2016 | Original Article

Prior knowledge of blood glucose meter download improves the accuracy of verbal self-reported blood glucose in teenagers with type I diabetes at ski camp

verfasst von: Annika Sjoeholm, Andrew Gray, Jenny Rayns, Paul A. Tomlinson, Benjamin J. Wheeler

Erschienen in: Acta Diabetologica | Ausgabe 4/2016

Einloggen, um Zugang zu erhalten

Abstract

Aims

Despite advances in diabetes management, self-monitoring of blood glucose (SMBG) remains fundamental. A number of studies, principally in adults, have confirmed that logbook entries and verbal SMBG reports are prone to common errors. In the context of an adolescent diabetes camp, the accuracy of verbally reported SMBG is crucial for guiding safe therapeutic management, and negating the risk of exercise-induced hypoglycemia. We aimed to assess whether awareness of a planned meter download at the completion of a diabetes camp would improve the overall accuracy of verbally reported SMBG.

Methods

Adolescents with type one diabetes (n = 26) attended a 3-day ski camp in 2014. Verbally reported SMBG values were recorded by camp supervisors at multiple time points throughout the camp. The intervention involved ensuring that all participants (at camp commencement) were aware of a planned meter download and SMBG review at camp conclusion. These data were then compared with historical camp data from 2012, collected using identical methodology, in which participants (n = 20) were unaware of the planned meter download. For analysis, blood glucose (BGL) data were classified as: matching, phantom (verbal SMBG value with no corresponding meter download value), and over- or underestimate (verbally reported value >/< meter downloaded value).

Results

Dual data regarding verbal SMBG and meter downloads were obtained on 550 instances of BGL testing during the 2014 camp (the intervention group). This was compared to dual data for 396 historical tests from the 2012 control group. For the intervention group, the overall error rate was 4.7 %, over 34 % of participants. There was a statistically significant improvement in accuracy compared to historical nonintervention data, in which there was an error rate of 14.1 % over 70 % of participants (p < 0.001). There was also a decrease in phantom readings to 2 %, from 8.6 % in 2012 (p < 0.001).

Conclusions

This study demonstrates an improvement in accuracy and reliability of verbally reported SMBG, following a simple intervention of ensuring participants were aware of a meter download at the completion of camp. This intervention could be easily incorporated into adolescent diabetes camp safety protocols and may provide an easy, low-cost way of improving verbally reported SMBG accuracy and therefore safety on camp.
Literatur
1.
Zurück zum Zitat The Diabetes Control and Complications Trial Research Group (1993) The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med 329(14):977–986. doi:10.1056/nejm199309303291401 CrossRef The Diabetes Control and Complications Trial Research Group (1993) The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med 329(14):977–986. doi:10.​1056/​nejm199309303291​401 CrossRef
2.
Zurück zum Zitat Nathan DM, Cleary PA, Backlund JY et al (2005) Intensive diabetes treatment and cardiovascular disease in patients with type 1 diabetes. N Engl J Med 353(25):2643–2653CrossRefPubMed Nathan DM, Cleary PA, Backlund JY et al (2005) Intensive diabetes treatment and cardiovascular disease in patients with type 1 diabetes. N Engl J Med 353(25):2643–2653CrossRefPubMed
3.
Zurück zum Zitat Clarke SF, Foster JR (2012) A history of blood glucose meters and their role in self-monitoring of diabetes mellitus. Br J Biomed Sci 69(2):83–93PubMed Clarke SF, Foster JR (2012) A history of blood glucose meters and their role in self-monitoring of diabetes mellitus. Br J Biomed Sci 69(2):83–93PubMed
5.
Zurück zum Zitat Gonder-Frederick LA, Julian DM, Cox DJ, Clarke WL, Carter WR (1988) Self-measurement of blood glucose: accuracy of self-reported data and adherence to recommended regimen. Diabetes Care 11(7):579–5856CrossRefPubMed Gonder-Frederick LA, Julian DM, Cox DJ, Clarke WL, Carter WR (1988) Self-measurement of blood glucose: accuracy of self-reported data and adherence to recommended regimen. Diabetes Care 11(7):579–5856CrossRefPubMed
6.
Zurück zum Zitat Kalergis M, Nadeau J, Pacaud D, Yared Z, Yale JF (2006) Accuracy and reliability of reporting self-monitoring of blood glucose results in adults with type 1 and type 2 diabetes. Can J Diabetes 30(3):241–247CrossRef Kalergis M, Nadeau J, Pacaud D, Yared Z, Yale JF (2006) Accuracy and reliability of reporting self-monitoring of blood glucose results in adults with type 1 and type 2 diabetes. Can J Diabetes 30(3):241–247CrossRef
8.
Zurück zum Zitat Chae M, Reith DM, Tomlinson PA, Rayns J, Wheeler BJ (2014) Accuracy of verbal self-reported blood glucose in teenagers with type I diabetes at diabetes ski camp. J Diabetes Metab Dis 13(1):14. doi:10.1186/2251-6581-13-14 CrossRef Chae M, Reith DM, Tomlinson PA, Rayns J, Wheeler BJ (2014) Accuracy of verbal self-reported blood glucose in teenagers with type I diabetes at diabetes ski camp. J Diabetes Metab Dis 13(1):14. doi:10.​1186/​2251-6581-13-14 CrossRef
9.
Zurück zum Zitat Mazze RS, Pasmantier R, Ann Murphy J, Shamoon H (1985) Self-monitoring of capillary blood glucose: changing the performance of individuals with diabetes. Diabetes Care 8(3):207–213CrossRefPubMed Mazze RS, Pasmantier R, Ann Murphy J, Shamoon H (1985) Self-monitoring of capillary blood glucose: changing the performance of individuals with diabetes. Diabetes Care 8(3):207–213CrossRefPubMed
11.
Zurück zum Zitat Salmond C, Crampton P, Atkinson J (2007) NZDep2006 index of deprivation. University of Otago, Wellington Salmond C, Crampton P, Atkinson J (2007) NZDep2006 index of deprivation. University of Otago, Wellington
12.
Zurück zum Zitat Miller AR, Nebesio TD, DiMeglio LA (2011) Insulin dose changes in children attending a residential diabetes camp. Diabet Med 28(4):480–486CrossRefPubMed Miller AR, Nebesio TD, DiMeglio LA (2011) Insulin dose changes in children attending a residential diabetes camp. Diabet Med 28(4):480–486CrossRefPubMed
13.
Zurück zum Zitat Schütt M, Kern W, Krause U et al (2006) Is the frequency of self-monitoring of blood glucose related to long-term metabolic control? Multicenter analysis including 24,500 patients from 191 centers in Germany and Austria. Exp Clin Endocrinol Diabetes 114(7):384–388CrossRefPubMed Schütt M, Kern W, Krause U et al (2006) Is the frequency of self-monitoring of blood glucose related to long-term metabolic control? Multicenter analysis including 24,500 patients from 191 centers in Germany and Austria. Exp Clin Endocrinol Diabetes 114(7):384–388CrossRefPubMed
Metadaten
Titel
Prior knowledge of blood glucose meter download improves the accuracy of verbal self-reported blood glucose in teenagers with type I diabetes at ski camp
verfasst von
Annika Sjoeholm
Andrew Gray
Jenny Rayns
Paul A. Tomlinson
Benjamin J. Wheeler
Publikationsdatum
21.03.2016
Verlag
Springer Milan
Erschienen in
Acta Diabetologica / Ausgabe 4/2016
Print ISSN: 0940-5429
Elektronische ISSN: 1432-5233
DOI
https://doi.org/10.1007/s00592-016-0855-z