J Am Acad Audiol 2005; 16(03): 172-183
DOI: 10.3766/jaaa.16.3.5
Articles
American Academy of Audiology. All rights reserved. (2005) American Academy of Audiology

Multiple Auditory Steady-State Responses to Bone-Conduction Stimuli in Adults with Normal Hearing

Susan A. Small
,
David R. Stapells
Further Information

Publication History

Publication Date:
07 August 2020 (online)

ASSR thresholds to bone-conduction stimuli were determined in 10 adults with normal hearing using mastoid placement of the bone oscillator. ASSRs to 0–50 dB HL bone-conduction stimuli and to 30–60 dB HL air-conduction stimuli were compared. The effect of alternating stimulus polarity on air- and bone-conduction ASSRs was also investigated. Stimuli were bone- and air-conduction amplitude-modulated tones (500–4000 Hz carrier frequencies, modulated at 77–101 Hz). ASSRs were recorded using the Rotman MASTER research system. Mean (1SD) bone-conduction ASSR thresholds were 22(11), 26(13), 18(8), and 18(11) dB HL for 500, 1000, 2000, and 4000 Hz, respectively. Except for a steeper slope at 500 Hz, ASSR intensity-amplitude functions for binaural bone- and air-conduction stimuli showed the same slopes; intensity-phase-delay functions were steeper at 1000 Hz for ASSRs to bone-conduction stimuli. ASSR amplitudes and phases did not differ for single- versus alternated-stimulus polarities for both bone- and air-conduction stimuli. The steeper amplitude slope for ASSRs to 500 Hz stimuli may reflect a nonauditory contribution to the ASSR.