Skip to main content
Erschienen in: Heart and Vessels 5/2014

Open Access 01.09.2014 | Case Report

Pseudo-postpacing interval of diastolic potential after entrainment pacing of remote bystander pathway in reentrant ventricular tachycardia

verfasst von: Yoshiaki Kaneko, Tadashi Nakajima, Tadanobu Irie, Osamu Igawa, Takafumi Iijima, Masaki Ota, Mio Tamura, Takashi Iizuka, Shuntaro Tamura, Akihiro Saito, Masahiko Kurabayashi

Erschienen in: Heart and Vessels | Ausgabe 5/2014

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

After entrainment pacing, the postpacing interval of a diastolic potential may be misinterpreted if the distal tip of the ablation catheter captures a remote bystander pathway adjacent to the critical isthmus of a complex reentrant circuit in a structurally diseased heart. We discuss this possible pitfall of entrainment mapping of reentrant ventricular tachycardia, observed after a healed myocardial infarction.

Introduction

Entrainment mapping is a key maneuver used to locate the critical isthmus of a reentrant circuit during catheter ablation of a reentrant ventricular tachycardia (VT). [1] The postpacing interval (PPI) of the diastolic potential (DP), reflective of the activation in the zone of slow conduction following entrainment pacing limited to the critical isthmus, is typically equal to the tachycardia cycle length (TCL) [1], whereas the PPI of a DP during entrainment pacing outside the critical isthmus is of electrophysiologic interest [2, 3]. We describe the electrophysiologic behavior and significance of the PPI after entrainment pacing of a remote bystander pathway near the reentry circuit of VT after a healed anteroseptal myocardial infarction, followed by autopsy of the heart.

Case report

A 73-year-old man with a history of healed anteroseptal myocardial infarction underwent radiofrequency (RF) catheter ablation of VT with right bundle branch block and undetermined QRS axis morphology, refractory to drug therapy. The patient had undergone implantation of a cardioverter-defibrillator 2 years earlier. A left ventriculography revealed the presence of anterior wall dyskinesia and a 20 % left ventricular ejection fraction. An area of low-amplitude bipolar electrograms, defined as ≤ 0.6 mV [4], was present on a CARTO-X-P electroanatomic map (Biosense Webster, Diamond Bar, CA, USA) of the anterior left ventricular endocardial surface, recorded during sinus rhythm (Fig. 4a). VT was reproducibly induced by programmed ventricular stimulation from the right ventricular apex. Pacing during VT at the basal edge of the low-voltage zone, where a low-amplitude DP was recorded, showed concealed entrainment, with a pacing stimulus-QRS equal to the DP-QRS interval, and a PPI of the DP equal to the TCL, indicating that the pacing site was located on the critical isthmus of the reentrant circuit (Figs. 1, 4b). Further entrainment from the same site, at a slightly longer paced cycle length and identical pacing output, caused constant fusion without pacing stimulus-QRS delay, and a PPI of the DP nearly similar to the TCL (Figs. 2, 4b). The delivery of RF energy for 21 s at that site, using a 4-mm-tip, nonirrigated catheter at a maximum power of 50 W and maximum temperature of 60 °C, terminated and eliminated the inducibility of VT (Fig. 3) [5, 6].
The patient died of incessant VT 1 year after the ablation procedure, and a postmortem examination of the heart was performed. The heart weighed 590 g. The left ventricle was open across the anterior aneurysm and the apex (Fig. 4c), along the same plane as the right oblique fluoroscopic view shown in Fig. 4b, exposing the ablation region. On visual inspection the left ventricle was not dilated, although the posterior and septal walls were hypertrophied, and a 6 cm × 5 cm × 2 mm anterior aneurysm was present, with a trabeculated and whitish endocardial surface (Fig. 4c). The site of successful ablation was at the basal margin of the aneurysm (Fig. 4b, c). The aneurysm was sectioned transmurally, perpendicular to the mitral annulus, stained with Masson trichrome, and processed for standard microscopic examination (Fig. 4d), which revealed nontransmural fibrosis within two-thirds of the aneurysmal endomyocardium, corresponding to the ablation lesion (Fig. 4e), and nonablated, surviving endomyocardium adjacent to the aneurysm (Fig. 4f).

Discussion

The morphology, amplitude, and timing of the local ventricular electrograms and DP relative to the QRS complex recorded from the ablation catheter during VT are similar in Figs. 1 and 2, confirming that the tip of the ablation catheter was in the same position relative to the critical isthmus during both episodes of entrainment pacing. Since, as shown in Figs. 1 and 2, the 90-ms DP-QRS interval represents <30 % of the 380-ms TCL, the pacing site was located near the exit of the critical isthmus (Fig. 5b) [1]. However, in contrast to Fig. 1, no latency of the pacing stimulus-QRS complex associated with a short, 40-ms interval between the pacing stimulus and the local ventricular electrogram is visible during entrainment pacing in Fig. 2, suggesting that, unlike the capture limited to the critical isthmus in Fig. 1, pacing in Fig. 2 captured the ventricular myocardium outside the zone of slow conduction. When entrainment pacing captured the ventricular myocardium outside the zone of slow conduction, the orthodromic wavefront reached the recording site of the DP in the critical isthmus without colliding with the antidromic wavefront (Fig. 5c). Therefore, in contrast to the PPI of the DP shown in Fig. 1, that shown in Fig. 2 is not a true PPI, generally defined as the interval between the last pacing stimulus and the subsequent, directly captured electrogram [1]. The true PPI in Fig. 2 corresponded to the interval between the last pacing stimulus and the captured ventricular electrogram, obviously longer than the TCL, suggesting that pacing in Fig. 2 captured a remote bystander pathway (Fig. 5c) [7]. Consequently, the PPI of the DP in Fig. 2 corresponds to the interval between the pacing site in the remote bystander pathway and the DP recorded in the critical isthmus (pseudo-PPI) (Fig. 5c). Furthermore, that the PPI of the DP is nearly equal to the TCL in Fig. 2 does not mean that the pacing site was on the critical isthmus. Although the interval between the pacing stimulus and the DP recorded at the proximal electrode of the ablation catheter during entrainment pacing was similar in Figs. 1 and 2, it is not clear that pacing in Fig. 2 also captured the critical isthmus. Even in the presence of simultaneous capture of the essential and remote bystander pathways, neither the orthodromic nor the antidromic wavefronts originating from pacing at the critical isthmus would affect the orthodromic wavefront created by pacing at the remote bystander pathway, for the following reasons. When entrainment pacing captures the critical isthmus and the remote bystander simultaneously (as illustrated in Fig. 5d), the Nth antidromic wavefront originating from pacing at the remote bystander collides constantly, probably outside the critical isthmus, with the Nth orthodromic wavefront created by pacing at the critical isthmus, representing atypical constant fusion (Fig. 5d). Furthermore, the Nth antidromic wavefront originating from pacing at the critical isthmus collided within the critical isthmus, with the N − 1th orthodromic wavefront created by pacing at the remote bystander (Fig. 5d). Finally, the Nth orthodromic wavefront created by pacing at the remote bystander pathway reached the recording site of the DP in the critical isthmus, without colliding with the antidromic wavefront (Fig. 5c). Consequently, the PPI corresponds to the interval between the pacing site in the remote bystander and the DP recording in the critical isthmus (Fig. 5c).
This was probably due to simultaneous contact of the distal tip of the ablation catheter with the critical isthmus and the neighboring remote bystander pathway, along with a slight change in the contact with the endocardium associated with a minimal change in the paced cycle length, without a change in the pacing output/pulse width of 5 V/0.5 ms. The pathologic-fluoroscopic correlation suggested that the critical isthmus and the bystander pathway were consistent with the ablated, basal margin of an aneurysm and surviving endomyocardium adjacent to the aneurysm, respectively, both of which could be simultaneously in contact with the tip of the ablation catheter (Fig. 4).
In summary, entrainment pacing capturing the remote bystander pathway simultaneously with a recording of the DP represented a pseudo-PPI of the DP. This phenomenon, which may be observed in complex reentrant circuits with a critical isthmus located near the remote bystander, is not rare. The pseudo-PPI of the DP can be misinterpreted as the true PPI, especially when constant fusion causing minimal QRS fusion and simultaneous recording of the DP is misidentified as concealed entrainment. When entrainment pacing captures the remote bystander pathway simultaneously with a recording of a DP, causing constant fusion, the PPI of the DP does not represent the true PPI. In some circumstances, the pseudo-PPI might be shorter than the tachycardia cycle length [2, 3]. The absence of QRS fusion, i.e., concealed entrainment, must be confirmed during entrainment pacing to measure the true PPI of the DP.

Conflict of interest

The authors have no conflict of interest to disclose.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Literatur
1.
Zurück zum Zitat Stevenson WG, Khan H, Sager P, Saxon LA, Middlekauff HR, Natterson PD, Wiener I (1993) Identification of reentry circuit sites during catheter mapping and radiofrequency ablation of ventricular tachycardia late after myocardial infarction. Circulation 88(4 Pt 1):1647–1670PubMedCrossRef Stevenson WG, Khan H, Sager P, Saxon LA, Middlekauff HR, Natterson PD, Wiener I (1993) Identification of reentry circuit sites during catheter mapping and radiofrequency ablation of ventricular tachycardia late after myocardial infarction. Circulation 88(4 Pt 1):1647–1670PubMedCrossRef
2.
Zurück zum Zitat Kaneko Y, Nakajima T, Irie T, Kato T, Iijima T, Kurabayashi M (2012) Putative mechanism of a post pacing interval paradoxically shorter than the tachycardia cycle length. J Cardiovasc Electrophysiol 23:666–668PubMedCrossRef Kaneko Y, Nakajima T, Irie T, Kato T, Iijima T, Kurabayashi M (2012) Putative mechanism of a post pacing interval paradoxically shorter than the tachycardia cycle length. J Cardiovasc Electrophysiol 23:666–668PubMedCrossRef
3.
Zurück zum Zitat Tung S, Soejima K, Maisel WH, Suzuki M, Epstein L, Stevenson WG (2003) Recognition of far-field electrograms during entrainment mapping of ventricular tachycardia. J Am Coll Cardiol 42:110–115PubMedCrossRef Tung S, Soejima K, Maisel WH, Suzuki M, Epstein L, Stevenson WG (2003) Recognition of far-field electrograms during entrainment mapping of ventricular tachycardia. J Am Coll Cardiol 42:110–115PubMedCrossRef
4.
Zurück zum Zitat Yoshida K, Sekiguchi Y, Tanoue K, Endo M, Suzuki A, Kanemoto M, Yamasaki H, Yamauchi Y, Takahashi A, Kuga K, Yamaguchi I, Aonuma K (2008) Feasibility of targeting catheter ablation to the markedly low-voltage area surrounding infarct scars in patients with post-infarction ventricular tachycardia. Circ J 72:1112–1119PubMedCrossRef Yoshida K, Sekiguchi Y, Tanoue K, Endo M, Suzuki A, Kanemoto M, Yamasaki H, Yamauchi Y, Takahashi A, Kuga K, Yamaguchi I, Aonuma K (2008) Feasibility of targeting catheter ablation to the markedly low-voltage area surrounding infarct scars in patients with post-infarction ventricular tachycardia. Circ J 72:1112–1119PubMedCrossRef
5.
Zurück zum Zitat Suenari K, Nakano Y, Hirai Y, Ogi H, Oda N, Makita Y, Ueda S, Kajihara K, Tokuyama T, Motoda C, Fujiwara M, Chayama K, Kihara Y (2013) Left atrial thickness under the catheter ablation lines in patients with paroxysmal atrial fibrillation: insights from 64-slice multidetector computed tomography. Heart Vessel 28:360–368CrossRef Suenari K, Nakano Y, Hirai Y, Ogi H, Oda N, Makita Y, Ueda S, Kajihara K, Tokuyama T, Motoda C, Fujiwara M, Chayama K, Kihara Y (2013) Left atrial thickness under the catheter ablation lines in patients with paroxysmal atrial fibrillation: insights from 64-slice multidetector computed tomography. Heart Vessel 28:360–368CrossRef
6.
Zurück zum Zitat Park HE, Chang SA, Kim JH, Oh IY, Choi EK, Oh S (2013) Left ventricular dyssynchrony in pre-excitation syndrome: effect of accessory pathway location and reversibility after ablation therapy. Heart Vessels 28:199–207PubMedCrossRef Park HE, Chang SA, Kim JH, Oh IY, Choi EK, Oh S (2013) Left ventricular dyssynchrony in pre-excitation syndrome: effect of accessory pathway location and reversibility after ablation therapy. Heart Vessels 28:199–207PubMedCrossRef
7.
Zurück zum Zitat Kocovic DZ, Harada T, Friedman PL, Stevenson WG (1999) Characteristics of electrograms recorded at reentry circuit sites and bystanders during ventricular tachycardia after myocardial infarction. J Am Coll Cardiol 34:381–388PubMedCrossRef Kocovic DZ, Harada T, Friedman PL, Stevenson WG (1999) Characteristics of electrograms recorded at reentry circuit sites and bystanders during ventricular tachycardia after myocardial infarction. J Am Coll Cardiol 34:381–388PubMedCrossRef
Metadaten
Titel
Pseudo-postpacing interval of diastolic potential after entrainment pacing of remote bystander pathway in reentrant ventricular tachycardia
verfasst von
Yoshiaki Kaneko
Tadashi Nakajima
Tadanobu Irie
Osamu Igawa
Takafumi Iijima
Masaki Ota
Mio Tamura
Takashi Iizuka
Shuntaro Tamura
Akihiro Saito
Masahiko Kurabayashi
Publikationsdatum
01.09.2014
Verlag
Springer Japan
Erschienen in
Heart and Vessels / Ausgabe 5/2014
Print ISSN: 0910-8327
Elektronische ISSN: 1615-2573
DOI
https://doi.org/10.1007/s00380-013-0411-5

Weitere Artikel der Ausgabe 5/2014

Heart and Vessels 5/2014 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.