Quantitative contrast-enhanced ultrasound of the brain on twin fetal lambs maintained by the extrauterine environment for neonatal development (EXTEND): initial experience

verfasst von: Anush Sridharan, Kendall M. Lawrence, Juan S. Martin-Saavedra, Marcus G. Davey, Alan W. Flake, Ryne A. Didier

Erschienen in: Pediatric Radiology | Ausgabe 1/2021

Einloggen, um Zugang zu erhalten

Abstract

Background

With the development of an artificial environment to support the extremely premature infant, advanced imaging techniques tested in this extrauterine system might be beneficial to evaluate the fetal brain.

Objective

We evaluated the feasibility of (a) performing contrast-enhanced ultrasound (CEUS) and (b) quantifying normal and decreased brain perfusion in fetal lambs maintained on the extrauterine environment for neonatal development (EXTEND) system.

Materials and methods

Twin premature fetal lambs (102 days of gestational age) were transferred to the EXTEND system. Twin B was subjected to sub-physiological flows (152 mL/kg/min) and oxygen delivery (15.9 mL/kg/min), while Twin A was maintained at physiological levels. We administered Lumason contrast agent into the oxygenator circuit and performed serial CEUS examinations. We quantified perfusion parameters and generated parametric maps. We also recorded hemodynamic parameters, serum blood analysis, and measurements across the oxygenator. Postmortem MRIs were performed.

Results

No significant changes in hemodynamic variables were attributable to CEUS examinations. On gray-scale images, Twin B demonstrated ventriculomegaly and progressive parenchymal volume loss culminating in hydranencephaly. By CEUS, Twin B demonstrated decreased peak enhancement and decreased overall parenchymal perfusion when compared to Twin A by perfusion parameters and parametric maps. Changes in perfusion parameters were detected immediately following blood transfusion. Postmortem MRI confirmed ultrasonographic findings in Twin B.

Conclusion

In this preliminary experience, we show that CEUS of the brain is feasible in fetal lambs maintained on the EXTEND system and that changes in perfusion can be quantified, which is promising for the application of CEUS in this extrauterine system supporting the premature infant.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Reitmeir R, Eyding J, Oertel MF et al (2017) Is ultrasound perfusion imaging capable of detecting mismatch? A proof-of-concept study in acute stroke patients. J Cereb Blood Flow Metab 37:1517–1526CrossRef Reitmeir R, Eyding J, Oertel MF et al (2017) Is ultrasound perfusion imaging capable of detecting mismatch? A proof-of-concept study in acute stroke patients. J Cereb Blood Flow Metab 37:1517–1526CrossRef
2.
Zurück zum Zitat Vinke EJ, Kortenbout AJ, Eyding J et al (2017) Potential of contrast-enhanced ultrasound as a bedside monitoring technique in cerebral perfusion: a systematic review. Ultrasound Med Biol 43:2751–2757CrossRef Vinke EJ, Kortenbout AJ, Eyding J et al (2017) Potential of contrast-enhanced ultrasound as a bedside monitoring technique in cerebral perfusion: a systematic review. Ultrasound Med Biol 43:2751–2757CrossRef
3.
Zurück zum Zitat Hwang M, Riggs BJ, Saade-Lemus S, Huisman TAGM (2018) Bedside contrast-enhanced ultrasound diagnosing cessation of cerebral circulation in a neonate: a novel bedside diagnostic tool. Neuroradiol J 31:578–580CrossRef Hwang M, Riggs BJ, Saade-Lemus S, Huisman TAGM (2018) Bedside contrast-enhanced ultrasound diagnosing cessation of cerebral circulation in a neonate: a novel bedside diagnostic tool. Neuroradiol J 31:578–580CrossRef
4.
Zurück zum Zitat Benjamin JL, Dennis R, White S et al (2019) Improved diagnostic sensitivity of bowel disease of prematurity on contrast-enhanced ultrasound. J Ultrasound Med 39:1031–1036CrossRef Benjamin JL, Dennis R, White S et al (2019) Improved diagnostic sensitivity of bowel disease of prematurity on contrast-enhanced ultrasound. J Ultrasound Med 39:1031–1036CrossRef
5.
Zurück zum Zitat Thimm MA, Cuffari C, Garcia A et al (2019) Contrast-enhanced ultrasound and shear wave elastography evaluation of Crohn’s disease activity in three adolescent patients. Pediatr Gastroenterol Hepatol Nutr 22:282–290CrossRef Thimm MA, Cuffari C, Garcia A et al (2019) Contrast-enhanced ultrasound and shear wave elastography evaluation of Crohn’s disease activity in three adolescent patients. Pediatr Gastroenterol Hepatol Nutr 22:282–290CrossRef
6.
Zurück zum Zitat Hwang M, Thimm MA, Guerrerio AL (2018) Detection of cavernous transformation of the portal vein by contrast-enhanced ultrasound. J Ultrasound 21:153–157CrossRef Hwang M, Thimm MA, Guerrerio AL (2018) Detection of cavernous transformation of the portal vein by contrast-enhanced ultrasound. J Ultrasound 21:153–157CrossRef
7.
Zurück zum Zitat Hwang M, Sridharan A, Darge K et al (2019) Novel quantitative contrast-enhanced ultrasound detection of hypoxic ischemic injury in neonates and infants: pilot study 1. J Ultrasound Med 38:2025–2038CrossRef Hwang M, Sridharan A, Darge K et al (2019) Novel quantitative contrast-enhanced ultrasound detection of hypoxic ischemic injury in neonates and infants: pilot study 1. J Ultrasound Med 38:2025–2038CrossRef
8.
Zurück zum Zitat Hwang M, De Jong RM, Herman S et al (2017) Novel contrast-enhanced ultrasound evaluation in neonatal hypoxic ischemic injury: clinical application and future directions. J Ultrasound Med 36:2379–2386CrossRef Hwang M, De Jong RM, Herman S et al (2017) Novel contrast-enhanced ultrasound evaluation in neonatal hypoxic ischemic injury: clinical application and future directions. J Ultrasound Med 36:2379–2386CrossRef
9.
Zurück zum Zitat Errico C, Osmanski BF, Pezet S et al (2016) Transcranial functional ultrasound imaging of the brain using microbubble-enhanced ultrasensitive Doppler. Neuroimage 124:752–761CrossRef Errico C, Osmanski BF, Pezet S et al (2016) Transcranial functional ultrasound imaging of the brain using microbubble-enhanced ultrasensitive Doppler. Neuroimage 124:752–761CrossRef
10.
Zurück zum Zitat Wang X, Liu B, Liu J et al (2017) Evaluation of cerebral perfusion by contrast-enhanced ultrasound at low mechanical index in cerebral ischemia rat model. Int Angiol 36:545–552PubMed Wang X, Liu B, Liu J et al (2017) Evaluation of cerebral perfusion by contrast-enhanced ultrasound at low mechanical index in cerebral ischemia rat model. Int Angiol 36:545–552PubMed
11.
Zurück zum Zitat Lin WH, Fan CH, Ting CY et al (2013) Dynamic perfusion assessment by contrast-enhanced ultrasound in blood-brain barrier disruption. Conf Proc IEEE Eng Med Biol Soc 2013:1152–1155 Lin WH, Fan CH, Ting CY et al (2013) Dynamic perfusion assessment by contrast-enhanced ultrasound in blood-brain barrier disruption. Conf Proc IEEE Eng Med Biol Soc 2013:1152–1155
12.
Zurück zum Zitat Brunner C, Isabel C, Martin A et al (2017) Mapping the dynamics of brain perfusion using functional ultrasound in a rat model of transient middle cerebral artery occlusion. J Cereb Blood Flow Metab 37:263–276CrossRef Brunner C, Isabel C, Martin A et al (2017) Mapping the dynamics of brain perfusion using functional ultrasound in a rat model of transient middle cerebral artery occlusion. J Cereb Blood Flow Metab 37:263–276CrossRef
13.
Zurück zum Zitat De Lange C, Brabrand K, Emblem KE et al (2011) Cerebral perfusion in perinatal hypoxia and resuscitation assessed by transcranial contrast-enhanced ultrasound and 3 T MRI in newborn pigs. Investig Radiol 46:686–696CrossRef De Lange C, Brabrand K, Emblem KE et al (2011) Cerebral perfusion in perinatal hypoxia and resuscitation assessed by transcranial contrast-enhanced ultrasound and 3 T MRI in newborn pigs. Investig Radiol 46:686–696CrossRef
14.
Zurück zum Zitat Deng D, Dan G, Tao J et al (2015) Conventional and contrast-enhanced ultrasound assessment of craniocerebral gunshot wounds. Genet Mol Res 14:3345–3354CrossRef Deng D, Dan G, Tao J et al (2015) Conventional and contrast-enhanced ultrasound assessment of craniocerebral gunshot wounds. Genet Mol Res 14:3345–3354CrossRef
15.
Zurück zum Zitat Bryner B, Gray B, Perkins E et al (2015) An extracorporeal artificial placenta supports extremely premature lambs for 1 week. J Pediatr Surg 50:44–49CrossRef Bryner B, Gray B, Perkins E et al (2015) An extracorporeal artificial placenta supports extremely premature lambs for 1 week. J Pediatr Surg 50:44–49CrossRef
16.
Zurück zum Zitat Partridge EA, Davey MG, Hornick MA et al (2017) An extra-uterine system to physiologically support the extreme premature lamb. Nat Commun 8:15112CrossRef Partridge EA, Davey MG, Hornick MA et al (2017) An extra-uterine system to physiologically support the extreme premature lamb. Nat Commun 8:15112CrossRef
17.
Zurück zum Zitat Partridge EA, Davey MG, Hornick MA, Flake AW (2017) An extrauterine environment for neonatal development: extending fetal physiology beyond the womb. Semin Fetal Neonatal Med 22:404–409CrossRef Partridge EA, Davey MG, Hornick MA, Flake AW (2017) An extrauterine environment for neonatal development: extending fetal physiology beyond the womb. Semin Fetal Neonatal Med 22:404–409CrossRef
18.
Zurück zum Zitat Didier RA, Sridharan A, Lawrence K et al (2019) Contrast-enhanced ultrasound in extracorporeal support: in vitro studies and initial experience and safety data in the extreme premature fetal lamb maintained by the extrauterine environment for neonatal development. J Ultrasound Med 38:1971–1978CrossRef Didier RA, Sridharan A, Lawrence K et al (2019) Contrast-enhanced ultrasound in extracorporeal support: in vitro studies and initial experience and safety data in the extreme premature fetal lamb maintained by the extrauterine environment for neonatal development. J Ultrasound Med 38:1971–1978CrossRef
19.
Zurück zum Zitat Kilkenny C, Browne WJ, Cuthill IC et al (2010) Improving bioscience research reporting: the ARRIVE guidelines for reporting animal research. PLoS Biol 8:e1000412CrossRef Kilkenny C, Browne WJ, Cuthill IC et al (2010) Improving bioscience research reporting: the ARRIVE guidelines for reporting animal research. PLoS Biol 8:e1000412CrossRef
20.
Zurück zum Zitat Bilotta F, Robba C, Santoro A et al (2016) Contrast-enhanced ultrasound imaging in detection of changes in cerebral perfusion. Ultrasound Med Biol 42:2708–2716CrossRef Bilotta F, Robba C, Santoro A et al (2016) Contrast-enhanced ultrasound imaging in detection of changes in cerebral perfusion. Ultrasound Med Biol 42:2708–2716CrossRef
21.
Zurück zum Zitat Vinke EJ, Eyding J, de Korte C et al (2018) Quantification of macrocirculation and microcirculation in brain using ultrasound perfusion imaging. Acta Neurochir Suppl 126:115–120CrossRef Vinke EJ, Eyding J, de Korte C et al (2018) Quantification of macrocirculation and microcirculation in brain using ultrasound perfusion imaging. Acta Neurochir Suppl 126:115–120CrossRef
22.
Zurück zum Zitat Kastler A, Manzoni P, Chapuy S et al (2014) Transfontanellar contrast enhanced ultrasound in infants: initial experience. J Neuroradiol 41:251–258CrossRef Kastler A, Manzoni P, Chapuy S et al (2014) Transfontanellar contrast enhanced ultrasound in infants: initial experience. J Neuroradiol 41:251–258CrossRef
Metadaten
Titel
Quantitative contrast-enhanced ultrasound of the brain on twin fetal lambs maintained by the extrauterine environment for neonatal development (EXTEND): initial experience
verfasst von
Anush Sridharan
Kendall M. Lawrence
Juan S. Martin-Saavedra
Marcus G. Davey
Alan W. Flake
Ryne A. Didier
Publikationsdatum
01.09.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
Pediatric Radiology / Ausgabe 1/2021
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-020-04797-9

Weitere Artikel der Ausgabe 1/2021

Pediatric Radiology 1/2021 Zur Ausgabe

Hermes

Hermes

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.