Radiation exposure using the O-arm® surgical imaging system

verfasst von: Nicolas Pitteloud, Axel Gamulin, Christophe Barea, Jerome Damet, Guillaume Racloz, Marta Sans-Merce

Erschienen in: European Spine Journal | Ausgabe 3/2017

Einloggen, um Zugang zu erhalten

Abstract

Purpose

This study was conducted to characterise the O-arm® surgical imaging system in terms of patient organ doses and medical staff occupational exposure during three-dimensional thoracic spine and pelvic examinations.

Methods

An anthropomorphic phantom was used to evaluate absorbed organ doses during a three-dimensional thoracic spine scan and a three-dimensional pelvic scan with the O-arm®. Staff occupational exposure was evaluated by constructing an ambient dose cartography of the operating theatre during a three-dimensional pelvic scan as well as using an anthropomorphic phantom to simulate the O-arm® operator.

Results

Patient organ doses ranged from 30 ± 4 μGy to 20.0 ± 3.0 mGy and 4 ± 1 μGy to 6.7 ± 1.0 mGy for a three-dimensional thoracic spine and pelvic examination, respectively. For a single three-dimensional acquisition, the maximum ambient equivalent dose at 2 m from the iso-centre was 11 ± 1 μSv.

Conclusion

Doses delivered to the patient during a three-dimensional thoracic spine image acquisition were found to be significant with the O-arm®, but lower than those observed with a standard computed tomography examination. The detailed dose cartography allows for the optimisation of medical staff positioning within the operating theatre while imaging with the O-arm®.
Literatur
1.
Zurück zum Zitat Medtronic (2007) O-arm user manual, MN Medtronic (2007) O-arm user manual, MN
2.
Zurück zum Zitat Jin M, Liu Z, Liu X, Yan H, Han X, Qiu Y, Zhu Z (2015) Does intraoperative navigation improve the accuracy of pedicle screw placement in the apical region of dystrophic scoliosis secondary to neurofibromatosis type I: comparison between O-arm navigation and free-hand technique. Eur Spine J [Epub ahead of print] Jin M, Liu Z, Liu X, Yan H, Han X, Qiu Y, Zhu Z (2015) Does intraoperative navigation improve the accuracy of pedicle screw placement in the apical region of dystrophic scoliosis secondary to neurofibromatosis type I: comparison between O-arm navigation and free-hand technique. Eur Spine J [Epub ahead of print]
3.
Zurück zum Zitat Van de Kelft E, Costa F, Van der Planken D, Schils F (2012) A prospective multicenter registry on the accuracy of pedicle screw placement in the thoracic, lumbar and sacral level with the use of the O-arm imaging system and StealthStation navigation. Spine (Phila Pa 1976) 37:E1580–E1587CrossRef Van de Kelft E, Costa F, Van der Planken D, Schils F (2012) A prospective multicenter registry on the accuracy of pedicle screw placement in the thoracic, lumbar and sacral level with the use of the O-arm imaging system and StealthStation navigation. Spine (Phila Pa 1976) 37:E1580–E1587CrossRef
4.
Zurück zum Zitat Beir VII (2006) Health risks from exposure to low levels of ionizing radiation, National Research Council of the National Academies, Washington D.C Beir VII (2006) Health risks from exposure to low levels of ionizing radiation, National Research Council of the National Academies, Washington D.C
5.
Zurück zum Zitat Park MS, Lee KM, Lee B, Min E, Kim Y, Jeon S, Huh Y, Lee K (2012) Comparison of operator radiation exposure between C-arm and O-arm fluoroscopy for orthopaedic surgery. Radiat Prot Dosimetry 148:431–438CrossRefPubMed Park MS, Lee KM, Lee B, Min E, Kim Y, Jeon S, Huh Y, Lee K (2012) Comparison of operator radiation exposure between C-arm and O-arm fluoroscopy for orthopaedic surgery. Radiat Prot Dosimetry 148:431–438CrossRefPubMed
6.
Zurück zum Zitat Lange J, Karellas A, Street J, Eck JC, Lapinsky A, Connolly PJ, Dipaola CP (2013) Estimating the effective radiation dose imparted by intraoperative cone-beam computed tomography in thoracolumbar spinal surgery. Spine (Phila Pa 1976) 38:E306–E312CrossRef Lange J, Karellas A, Street J, Eck JC, Lapinsky A, Connolly PJ, Dipaola CP (2013) Estimating the effective radiation dose imparted by intraoperative cone-beam computed tomography in thoracolumbar spinal surgery. Spine (Phila Pa 1976) 38:E306–E312CrossRef
7.
Zurück zum Zitat Zhang J, Weir V, Fajardo L, Lin J, Hsiung H, Ritenour ER (2009) Dosimetric characterization of a cone-beam O-arm imaging system. J Xray Sci Technol 17:305–317PubMed Zhang J, Weir V, Fajardo L, Lin J, Hsiung H, Ritenour ER (2009) Dosimetric characterization of a cone-beam O-arm imaging system. J Xray Sci Technol 17:305–317PubMed
8.
Zurück zum Zitat CIRS (2012) ATOM dosimetry phantoms, VA CIRS (2012) ATOM dosimetry phantoms, VA
9.
Zurück zum Zitat ICRP (1975) Report on the task group on reference man. Publication 23, Ottawa ICRP (1975) Report on the task group on reference man. Publication 23, Ottawa
10.
Zurück zum Zitat ICRU (1992) Phantoms and computational models in therapy, diagnosis and protection. Publication 48, MD ICRU (1992) Phantoms and computational models in therapy, diagnosis and protection. Publication 48, MD
11.
Zurück zum Zitat JCGM-W (2008) Guide to the expression of uncertainty in measurement. International Organization for Standardization, Geneva JCGM-W (2008) Guide to the expression of uncertainty in measurement. International Organization for Standardization, Geneva
12.
Zurück zum Zitat APVL (2011) Manuel d’utilisation AT1123 V1. Saint-Cyr-sur-Loire, France APVL (2011) Manuel d’utilisation AT1123 V1. Saint-Cyr-sur-Loire, France
13.
Zurück zum Zitat Hadelsberg UP, Harel R (2012) Hazards of ionizing radiation and its impact on spine surgery. World Neurosurg [Epub ahead of print] Hadelsberg UP, Harel R (2012) Hazards of ionizing radiation and its impact on spine surgery. World Neurosurg [Epub ahead of print]
14.
Zurück zum Zitat Qureshi S, Lu Y, McAnany S, Baird E (2014) Three-dimensional intraoperative imaging modalities in orthopaedic surgery: a narrative review. J Am Acad Orthop Surg 22:800–809CrossRefPubMed Qureshi S, Lu Y, McAnany S, Baird E (2014) Three-dimensional intraoperative imaging modalities in orthopaedic surgery: a narrative review. J Am Acad Orthop Surg 22:800–809CrossRefPubMed
15.
Zurück zum Zitat Söderberg M, Abul-Kasim K, Ohlin A, Gunnarsson M (2011) Estimation of organ and effective dose to the patient during spinal surgery with a cone-beam O-arm system. Proc SPIE 7961:79613G1-G6 Söderberg M, Abul-Kasim K, Ohlin A, Gunnarsson M (2011) Estimation of organ and effective dose to the patient during spinal surgery with a cone-beam O-arm system. Proc SPIE 7961:79613G1-G6
16.
Zurück zum Zitat Zhang D, Li X, Gao Y, Xu X, Liu B (2013) A method to acquire CT organ dose map using OSL dosimeters and atom anthropomorphic phantoms. Med Phys 40:081918CrossRefPubMedPubMedCentral Zhang D, Li X, Gao Y, Xu X, Liu B (2013) A method to acquire CT organ dose map using OSL dosimeters and atom anthropomorphic phantoms. Med Phys 40:081918CrossRefPubMedPubMedCentral
17.
Zurück zum Zitat ICRP (2007) Radiological protection in medicine. Publication 105, Ottawa ICRP (2007) Radiological protection in medicine. Publication 105, Ottawa
Metadaten
Titel
Radiation exposure using the O-arm® surgical imaging system
verfasst von
Nicolas Pitteloud
Axel Gamulin
Christophe Barea
Jerome Damet
Guillaume Racloz
Marta Sans-Merce
Publikationsdatum
21.09.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
European Spine Journal / Ausgabe 3/2017
Print ISSN: 0940-6719
Elektronische ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-016-4773-0

Weitere Artikel der Ausgabe 3/2017

European Spine Journal 3/2017 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.