Open Access 01.12.2017 | Research article

Rapid deployment valve system shortens operative times for aortic valve replacement through right anterior minithoracotomy

verfasst von: Constanze Bening, Khaled Hamouda, Mehmet Oezkur, Christoph Schimmer, Ina Schade, Armin Gorski, Ivan Aleksic, Rainer Leyh

Erschienen in: Journal of Cardiothoracic Surgery | Ausgabe 1/2017

Abstract

Background

There is growing evidence from the literature that right anterior minithoracotomy aortic valve replacement (RAT-AVR) improves clinical outcome. However, increased cross clamp time is the strongest argument for surgeons not performing RAT-AVR. Rapid deployment aortic valve systems have the potential to decrease cross-clamp time and ease this procedure. We assessed clinical outcome of rapid deployment and conventional valves through RAT.

Methods

Sixty-eight patients (mean age 76 ± 6 years, 32% females) underwent RAT-AVR between 9/2013 and 7/2015. According to the valve type implanted the patients were divided into two groups. In 43 patients (R-group; mean age 74.1 ± 6.6 years) a rapid deployment valve system (Edwards Intuity, Edwards Lifesciences Corp; Irvine, Calif) and in 25 patients (C-group; mean age 74.2 ± 6.6 years) a conventional stented biological aortic valve was implanted.

Results

Aortic cross-clamp (42.1 ± 12 min vs. 68.3 ± 20.3 min; p < 0.001) and bypass time (80.4 ± 39.3 min vs. 106.6 ± 23.2 min; p = 0.001) were shorter in the rapid deployment group (R-group). We observed no differences in clinical outcome. Postoperative gradients (R-group: max gradient, 14.3 ± 8 mmHg vs. 15.5 ± 5 mmHg (C-group), mean gradient, 9.2 ± 1.7 mmHg (R-group) vs. 9.1 ± 2.3 mmHg (C-group) revealed no differences. However, larger prostheses were implanted in C-group (25 mm; IQR 23–27 mm vs. 23 mm; IQR 21–25; p = 0.009).

Conclusions

Our data suggest that the rapid deployment aortic valve system reduced cross clamp and bypass time in patients undergoing RAT-AVR with similar hemodynamics as with larger size stented prosthesis. However, larger studies and long-term follow-up are mandatory to confirm our findings.
Abkürzungen
AVR
Aortic valve replacement
CPB
Cardiopulmonary bypass
iEOA
Indexed effective orifice area
PPM
Patient prosthesis mismatch
RAT
Right anterior thoracotomy
RAT-AVR
Right anterior thoracotomy-aortic valve replacement
RD
Rapid deployment

Background

Rapid deployment valve technologies have been introduced recently to simplify aortic valve replacement (AVR) [16]. Numerous publications in the last 3 years reflect the increasing interest of the cardiac surgical community [16]. Due to the growing number of transcatherter aortic valve replacements a rethinking of surgical strategies for surgical AVR is inevitable to sustain surgical AVR as the gold standard for the treatment of significant aortic valve stenosis. Although minimally invasive surgical techniques minimizing surgical trauma [715] may be one approach, only 24.6% of isolated surgical AVR in Germany are performed through a minimally invasive approach [16]. Basically, two different types of surgical minimally invasive aortic valve replacement techniques have been advocated, upper hemisternotomy AVR and right minithoracotomy AVR. The majority of minimally invasive surgical AVR are performed through an upper hemisternotomy. One reason for the reluctance to adopt minimal invasive technique might be a presumed longer aortic cross clamp and cardiopulmonary bypass time with negative effects for the patient. The data from recent publications clearly demonstrate the safety and effectiveness of rapid deployment valves, showing reduced aortic cross clamp and cardiopulmonary bypass time compared to standard stented biological valves [2, 17, 18]. Today only two rapid deployment valve systems are on the market, the Perceval valve (Sorin Group; Milan; Italy) on the one hand and the Edwards Intuity rapid deployment valve system (Edwards Lifescience Corp; Irvine, Calif; U.S.A.) on the other hand. However, only few studies published data regarding rapid deployment valves for RAT-AVR and none of them reported results with the Edwards Intuity Elite rapid deployment valve sytem [4, 5]. There is no data showing that the Edwards Elite valve system may result in shorter cross clamp and cardiopulmonary bypass times compared with conventional biological aortic valves, particularly through a right anterior minithoracotomy approach.
Therefore the aim of this study was to evaluate the impact of rapid deployment aortic valve replacement on procedural times and outcome in a cohort of patients undergoing right anterior minithoracotomy aortic valve replacement with the Edwards Intuity Elite rapid deployment valve system versus conventional biological valves.

Methods

The ethics committee of the University Hospital Wuerzburg approved the present study. Due to the retrospective character of the study the ethics committee did not require written consent of the patients.
This was a retrospective study of prospectively collected data from consecutive patients scheduled for elective RAT-AVR at the Department of Thoracic and Cardiovascular Surgery in the University hospital Wuerzburg between 09/2013 and 07/2015.
Patients scheduled for elective surgical aortic valve replacement underwent 64 slice computed tomography (Somatom Definition ATS, Siemens, Germany) without contrast enhancement to evaluate the patient’s suitability for RAT-AVR. Basically, Glaubers criteria were used [7] to select patients for right anterior thoracotomy aortic valve replacement. The RD-valve system was the primary choice for AVR in these patients. Contraindications for implantation of an Edwards intuity RP valve system were a bicuspid aortic valve, non-calcified aortic valve pathology and an aortic valve annulus >27 mm in diameter. The later due to the maximum diameter of 27 mm provided by the company for this valve type. However, in some patients the adequately sized RD-valve was not available and a conventional stented biological valve was implanted instead.
From 09/2013 to 07/2015, 68 patients received a RAT-AVR for isolated aortic valve stenosis. According to valve type implanted the patients were divided into two groups. In one group (C-Group) a conventional stented biological valve was implanted. In the rapid deployment group (R-Group) the Edwards Intuity rapid deployment valve system was implanted. Pre-, perioperative and postoperative data were collected from patient records recorded. In addition to clinical variables, pre-and postoperative transthoracic echocardiographic parameters (mean-maximum valvular gradients and paravalvular leakage) were obtained from standardized echocardiography’s performed within the clinical routine. Patient-prosthesis mismatch (PPM) and an indexed effective orifice area (iEOA) < 0.65 cm2/m2 was considered as severe PPM and an iEOA <0.85 cm2/m2 as moderate PPM [19, 20]. Patient’s preoperative categorical and continuous data are depicted in Table 1.
Table 1
Patient’s characteristic
Variable
All pts.(n = 68)
C-Group-(n = 25)
R-Group(n = 43)
p-value
Age (y)
75.9 (SD 5.7)
74.2 (SD 6.6)
74.1 (SD 6.6)
0.058
BSA (m2)
1.89 (SD 0.17)
1.93 (SD 0.13)
1.86 (SD 0.18)
0.11
Female gender; n (%)
24 (35.3)
8 (32.0)
16 (37.7)
>0.5
COPD; n (%)
7 (10.3)
2 (8)
5 (11.6)
>0.5
DM; n (%)
26 (38.2)
5 (20)
21 (48.8)
0.022
Creatinine (mg/dl)
1.10 (0.86–1.4)
1.01 (0.7–1.5)
1.20 (1.0–1.4)
0.31
CKD-EPI (ml/min/1,73 m2)
64.5 (SD 21.2)
69.8 (SD 27.0)
61.5 (SD 16.5)
0.18
Chronic AF; n (%)
15 (22.1)
7 (28)
8 (18.6)
0.38
PAVD; n (%)
9 (13.2)
3 (12)
6 (14)
>0.5
Stroke; n (%)
4(5.9)
2(8.0)
2(4.7)
>0.5
Art.hypertension; n(%)
63(94)
21(87.5)
42(97.7)
0.13
Pul. hypertension; n (%)
24(35.3)
7(28.0)
17(39.5)
0.43
NYHA class III -IV; n (%)
40 (58.8)
14 (56)
26 (60.5)
>0.5
LVEF (%)
54 (SD 12.8)
54.2 (SD 14.0)
53.9 (SD 12)
0.25
EuroSCORE II (%)
5.3 (3.7–8.9)
5.0 (3.1–9.1)
5.7 (3.9–9.0)
0.29
Preoperative EOA, cm2
0.78 (SD 0.22)
0.75 (SD 0.25)
0.79 (SD 0.19)
>0.5
Preoperative mean aortic valve gradient, mmHG
49.6 (SD 11.6)
48.8 (SD 12.7)
50.1 (SD 10.9)
>0.5
Data presented as mean ± standard deviation (SD), median and IQR, or number of observations (n) with proportions (%)
BSA body surface area, COPD Chronic obstructive pulmonary disease, DM diabetes mellitus, CKD-EPI Chronic Kidney Disease Epidemiology Collaboration, AF atrial fibrillation, NYHA New York Heart Association Classification, LVEF left ventricular ejection fraction, PAVD peripheral arterial vascular disease; pulmonary hypertension: mean pulmonary pressure > 30 mmHG; EOA effective orifice area

Surgical procedure

After induction of general anesthesia RAT-AVR was performed through a 5–6 cm skin incision placed above the third intercostal space in all patients, the fourth rib was separated from the sternum and carefully translocated caudally. The right mammary artery was severed and clipped in all patients. After dissection of mediastinal fat tissue the pericardium was opened longitudinally and multiple stay sutures were placed. A soft tissue retractor (Alexis Wound Protector, Applied Medical, USA), together with a small rib retractor (MRP-1, Fehling, Germany) was inserted to increase visualization of the surgical field. Direct ascending aortic cannulation with a flexible cannula (Easyflow Duo Cannulae, Sorin Group, Italy) was performed in 58 patients; in the remaining 10 patients peripheral arterial cannulation was performed. Indications for peripheral arterial cannulation were difficulties to visualize the area where to place the aortic cannula. Percutaneous venous cannulation (RAP femoral venous cannula, Sorin Group Italy) was used in all patients. Transesophageal echocardiogram was used in all patients to verify the correct position of the venous cannula in the vena cava superior and of the aortic cannula in the descending aorta, respectively.
Surgery was performed at moderate hypothermia (32 °C), standard myocardial protection was used. Before cross-clamping the aorta, a left atrial vent was placed in the right upper pulmonary vein in a usual fashion. After administration of cardioplegia a standard hockey stick incision towards the non coronary sinus extended below the sino tubular junction was performed to expose the native aortic valve and thorough decalcification of the aortic annulus was performed. At this point and after sizing the native aortic annulus the decision was made whether to implant a conventional stented biological valve or an Edwards Intuity rapid deployment valve. For implantation of a conventional standard valve pledgetted mattress sutures were placed in a standard fashion and the appropriate valve was positioned in a supraannular position. The following conventional valves were implanted: Trifecta (St, Jude Medical, St. Paul, MN, USA; n = 4), or Perimount (Edwards Lifesciences; n = 21).
For implantation of an Edwards Intuity rapid deployment valve a meticulous sizing technique has to be performed to omit the risk of malposition of the valve, since oversizing will result in valve luxation and undersizing in paravalvular leakage. Three guiding sutures were placed through the nadir of each sinus and through the corresponding part of the valve sewing ring. The valve system was lowered into the aortic annulus and proper seating of the valve was ensured. The guiding sutures were snared and the balloon catheter inflated for 10 s to deploy the stent frame. After deployment the delivery system and valve holder were removed carefully and the guiding sutures tied. In all patients the aortotomy was closed in a typical double layer fashion with a pledgetted suture starting at each corner of the aortotomy. After weaning from cardiopulmonary bypass standard transoesophegal echocardiographic examination of the implanted valve was performed to rule out any paravalvular leakage. Postoperative anticoagulation was in accordance with recent Guidelines for Management of Patients with Valvular Disease [21].

Statistical analysis

Statistical analysis was performed using SPSS, version 23. Patient demographics are presented as mean ± standard deviation, medians and interquartile ranges and number of observations with proportions (%), as appropriate. Differences across group were assessed by t-Test, χ2-test/Fisher’s exact test, and Mann-Whitney-U-Test respectively. Two sided p-values < 0.05 were considered statistically significant.

Results

Patient characteristics are shown in Table 1. Demographic data were comparable, the incidence of diabetes mellitus was higher in the R-group . Surgical data are depicted in Table 2. We observed shorter CPB-time (p = 0.001) and shorter aortic cross clamp-time (p < 0.001) for patients receiving Edwards Intuity rapid deployment valve (R-group). Aortic cross clamp time was reduced by 38% and CPB-time by 25%. A trend towards a shorter operative time was found (p = 0.071).
Table 2
Intraoperative characteristics
Variable
All pts.(n = 68)
C-Group-(n = 25)
R-Group(n = 43)
p-value
OP-time, min.
182 (167–207)
189 (169–219)
180 (152–197)
0.071
CPB-time, min.
90 (SD 36.4)
106.6 (SD 23.2)
80.4 (SD 39.3)
0.001
X-clamp time, min.
51.7 (SD 20.2)
68.3 (SD 20.3)
42.1 (SD 12.6)
<0.001
Conversion, n(%)
2 (2.9)
0 (0)
2 (4.6)
>0.5
Valve Diameter, mm
23 (21–25)
25 (23–27)
23 (21–25)
0.009
Data presented as mean ± standard deviation (SD), median and IQR, or number of observations (n) with proportions (%)
OP operation, CPB cardiopulmonary bypass
x-clamp, aortic cross clamp
Conversion rate for the entire patient cohort was 2.9% (2/68), with no differences between groups. One patient developed severe intraoperative bleeding due to an aortic root perforation. The patient required an aortic root replacement plus bypass surgery to the left coronary system due to laceration of the left main stem. The second patient became asystolic after central arterial cannulation, external chest compression was necessary, and the percutaneous femoral venous cannula was rapidly introduced to establish cardiopulmonary bypass. After uneventful AVR, severe hemorrhage developed. A full sternotomy revealed a laceration of the right ventricle caused by the percutaneous venous cannula as the bleeding source.
Biological valves with significantly larger diameter were implanted in the standard group (S-Group) (p = 0.009), this was due to fact that in the majority of these patients either a bicuspid aortic valve or a valve diameter >27 mm was detected during operation.
No differences in postoperative outcome were observed (Table 3). New onset atrial fibrillation had a low incidence in our patient cohort 8.8% (6/68). The same is true for atrioventricular block III° requiring pacemaker implantation (1.5%). However, three patients with temporary atrioventricular block III° were detected in the Edwards Intuity rapid deployment valve group (R-group), and one needed a permanent pacemaker implantation.
Table 3
Postoperative characteristics
Variable
All pts.(n = 68)
C-Group-(n = 25)
R-Group(n = 43)
p-value
Assisted ventilation (h)
11 (8–15)
12 (10–17)
11 (8–14)
0.14
Reoperation, n(%)
0 (0)
0 (0)
0 (0)
>0.5
Transfusion, n(%)
38 (57.4)
12 (48.0)
26 (60.5)
>0.5
RBC, units
0 (0–2)
0 (0–2)
0 (0–2)
>0.5
New–AF, n (%)
6 (8.8)
3 (12)
3 (7)
>0.5
Temp. AVB III°,n(%)
3 (4.4)
0 (0)
3 (7.0)
0.29
Perm. AVB III°,n(%)
1 (1.5)
0 (0)
1 (2.3)
>0.5
TND, n(%)
1(1.5)
1(4.0)
0 (0)
0.37
ICU stay, (d)
1(1–3)
1(1–3)
1(1–3)
>0.5
30d mortality, n(%)
2(2.9)
0(0)
2 (4.7)
>0.5
Data presented as mean ± standard deviation (SD), median and interquartile range (IQR), or number of observations (n) with proportions (%)
RBC red blood cell, New-AF new onset atrial fibrillation, Temp AVB III°, temporary atrioventricular block third degree; perm. AVB III° permanent atrioventricular block third degree, TND temporary neurological dysfunction, ICU intensive care unit
Overall 30 day mortality was 2.9% (2/68) for the entire patient cohort with no differences between groups. One patient who needed conversion to full sternotomy for aortic root replacement died on the 5th postoperative day due to multiorgan failure, a second patient died on postoperative day 5 on the normal ward after an initial uneventful postoperative course. Since postoperative echocardiography on the same day had shown normal left and right ventricular function, low transprosthetic mean gradient and no paravalvular leakage it seems unlikely that the implanted valve caused this in-hospital death. However, we have no autopsy results on this patient.
Hemodynamic data showed no difference between groups for resting transvalvular mean (R group: 9.2 ± 1.7 mmHG, vs. 9.1 ± 2.3 mmHg (C-group), p > 0.5) and maximal gradients (R-group: 14.3 ± 8 mmHG, vs. 15.5 ± 5.0 mmHg, C-group, p = 0.30) respectively. In C-group significantly larger prostheses were implanted (Table 2), thus similar hemodynamics were found for the Rapid deployment as with larger stented conventional bioprosthesis. In neither group paravalvular leakage or PPM were detected.

Discussion

We demonstrated a reduced aortic cross clamp-time and CPB time with the Edwards’s Intuity rapid deployment valve system compared with conventional stented biological valves for AVR through right anterior thoracotomy. To the best of our knowledge there are no data about aortic cross clamp and CPB time for the Edwards Intuity valve system in right minithoracotomy AVR. The mean aortic cross clamp time for the Edwards Intuity rapid deployment valve system in our study with 42.1 ± 12.6 min and CPB time of 80.4 ± 39.3 min are comparable to isolated AVR with the same valve type in the setting of either hemi- or full- sternotomy AVR [3, 11]. Only Schlömicher and coworker demonstrated a markedly reduced aortic cross clamp time for the Edwards Intuity rapid deployment valve (26 ± 7 min.) in hemisternotomy AVR [2]. A randomized comparison between patients undergoing minimally invasive AVR with rapid deployment valve systems and standard stented biological valves has not been reported yet. The only randomized study compared minimally invasive rapid deployment versus conventional full sternotomy aortic valve replacement; in this study a significantly reduced aortic cross clamp time was observed for rapid deployment AVR but no difference in clinical outcome [3].
Advantages for right mini-thoracotomy AVR compared to full-sternotomy AVR have been published [9, 11, 22]. The benefits included shorter mechanical ventilation time, shorter intensive care unit stay and hospital length of stay, lower incidence of blood transfusion and postoperative atrial fibrillation. However, consistent disadvantages were increased aortic cross-clamp and cardiopulmonary bypass time. A recent Bayesian network meta-analysis comparing ministernotomy or minithoracotomy for minimalinvasive AVR revealed longer aortic cross clamp and CPB time for minithoracotomy AVR with similar post-operative outcomes compared to ministernotomy AVR [14]. It has been shown that aortic crossclamp time > 60 min and prolonged cardiopulmonary bypass time are independent predictors of mortality and morbidity in low- and high-risk cardiac patients [23, 24]. In minithoracotomy AVR exposure and implantation of the prosthesis is more challenging than for full sternotomy or hemisternotomy AVR, which could be the reason for longer aortic cross clamp and CPB time. Therefore right minithoracotomy AVR is generally considered a challenging approach reserved for very experienced surgeons. New technologies, like sutureless or rapid deployment valves facilitate right minithoracotomy AVR due to a faster and easier implantation technique. Glauber and coworkers were able to reduce aortic cross clamp for minithoracotomy AVR by 28 min when using a sutureless Perceval S prosthesis (Sorin Biomedica) [4, 9]. In this study we reduced the aortic cross clamp time and cardiopulmonary bypass time by 26 min each using the Edwards Intuity rapid deployment valve.
Valve hemodynamics at discharge showed equally good results for all valve types with low mean pressure gradients at. However, in the patients with standard stented bioprosthesis larger valves were implanted. Therefore, it can be assumed that there might be a hemodynamic advantage for the Edwards intuity rapid deployment compared to conventional stented valves. The specific design of the rapid deployment valve could explain this result. Basically, the Edwards intuity valve is a stented bovine pericardial bioprosthesis attached to a balloon expandable, cloth covered skirt frame at the inflow part of the valve. After inflation of the balloon the skirt frame will expand and create a reversed funnel thereby widening the left ventricular outflow tract. This could result in less turbulence with subsequently better hemodynamic. Our results are supported by Borger and coworkers who showed lower mean gradients after 3 months and lower incidence of PPM for the Edwards Intuity valve compared to standard stented valves, although the implanted valve size revealed no differences [3].
Overall complication rate in this study was remarkably low. Neither PPM nor paravalvular leakage was detected in this series independent from the implanted valve type. The incidence of pacemaker implantation for new atrioventricuar block III° was 1.5% (1/68) for the entire patient cohort and 2.3% (1/43) in the rapid deployment group (R-group). It is noteworthy that after implantation of the Edwards intuity valve a 7%(3/43) incidence of temporary atrioventricuar block III° occurred but did notin the conventional group. Whether balloons expansion of the subvalvular skirt frame has any negative impact on the conduction system has not been evaluated in detail yet.
Low complication rates, good hemodynamics and the non-sternal splitting minimally invasive approach of minithoracotomy AVR in combination with new rapid deployment or sutureless valves AVR justifies that this technique should be added to the armamentarium in the treatment of aortic valve stenosis. Furthermore, there are data indicating that the combination of rapid deployment or sutureless valves with minimally invasive AVR approaches show better clinical results in high risk patients compared with transapical transcatheter aortic valve implantation [25, 26] In view of the increasing competition with transcatheter approaches for patients with an intermediate and high risk profile this technique deserves comparison with transfemoral transcatheter aortic valve implantation.

Limitation of the study

Several facts could have influenced our results.
1.
The fact that this is not a randomized study and the number of patients evaluated is relatively small should lead to a cautious interpretation of these data. However, the reduction in aortic cross clamp time and CPB time was striking and highly significant. The effectiveness and time saving effect of the Edwards Intuity rapid deployment valve system implanted through a right anterior mini-thoracotomy have been proven.
 
2.
A selection bias cannot be ruled out entirely. Contraindication for placing the Edwards RD-valve system was a bicuspid aortic valve, and an annulus size > 27 mm. Furthermore, lack of availability ofadequately sized valves precluded implantation of RD-valve system in several patients, and this could have influenced our result. However, difficulties in placing interrupted mattress sutures in the aortic annulus are not influenced by the size of the annulus, but rather by difficulties in proper exposition of the native annulus through a right anterior mini-thoracotomy. Therefore we believe that our results reflect daily surgical practice when performing right anterior mini-thoracotomy AVR.
 

Conclusion

Our preliminary data indicate that the implantation of Edwards Intuity RD-valve system via a right mini-thoracotomy is safe and reproducible. This valve system reduced aortic cross clamp and CPB time compared with conventional stented biological valves thereby supporting further spreading of this innovative surgical approach for isolated surgical AVR. These results could accelerate a rethinking in surgical strategies for non-sternal surgical AVR. For final judgment of this RD-valve system for RAT-AVR larger studies comparing different RD-valve systems and conventional stented biological valves are mandatory. Furthermore, long-term follow up is necessary to evaluate the longevity of the valve and to prove the good hemodynamics in the long run.

Acknowledgment

Not applicable.

Funding

The study was funded by the Department of Cardiothoracic- and Vascular Surgery of the University Hospital Wuerzburg.

Availability of data and materials

Please contact author for data requests.

Authors’ contribution

CB: Writing and drafting the manuscript. KH: Performing the operation with RD intuity valve, drafting the manuscript. MO: Statitical analysis,drafting the manuscript. CS: Drafting the manuscript, editing the discussion chapter. IS: Drafting the manuscript, analyzing the results. AG: Performing the operation with RD intuity valve, discussing the results. IA: Performing the operation, drafting the manuscript. RL: Performing the operation with RD intuity valve, drafting the manuscript, discussion of results, approval of final version. All authors read and approved the final manuscript.

Competing interests

Rainer Leyh reports lecture fees from Edwards Lifesciences (Edwards Lifescience Corp; Irvine, Calif), all other authors have nothing to disclose.
All authors of the manuscript have read and agreed to its content and are accountable for all aspects of the accuracy and integrity of the manuscript in accordance with ICMJE criteria. That the article is original, has not already been published in a journal, and is not currently under consideration by another journal.
Because of analyzing retrospectively the intraoperative data of patients treated with RD Intuity valve we consulted the ethic’s committee: There is no need of an approval of the ethic’s committee. This statement can be provided.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Santarpino G, Pfeiffer S, Concistré G, Grossmann I, Hinzmann M, Fischlein T. The Perceval S aortic valve has the potential of shortening surgical time:does it also result in improved outcome? Ann Thorac Surg. 2013;96:77–82.CrossRefPubMed Santarpino G, Pfeiffer S, Concistré G, Grossmann I, Hinzmann M, Fischlein T. The Perceval S aortic valve has the potential of shortening surgical time:does it also result in improved outcome? Ann Thorac Surg. 2013;96:77–82.CrossRefPubMed
2.
Zurück zum Zitat Schlömmicher M, Haldenwang PL, Moustafine V, Bechtel M, Strauch JT. Minimal access rapid deployment aortic valve replacement: initial single- center experience and 12-months outcomes. J Thorac Cardiovasc Surg. 2015;149:434–40.CrossRef Schlömmicher M, Haldenwang PL, Moustafine V, Bechtel M, Strauch JT. Minimal access rapid deployment aortic valve replacement: initial single- center experience and 12-months outcomes. J Thorac Cardiovasc Surg. 2015;149:434–40.CrossRef
3.
Zurück zum Zitat Borger MA, Moustafine V, Conradi L, et al. A randomized multicenter trial of minimally invasive rapid deployment versus conventional full sternotomy aortic valve replacement. Ann Thorac Surg. 2015;99:17–25.CrossRefPubMed Borger MA, Moustafine V, Conradi L, et al. A randomized multicenter trial of minimally invasive rapid deployment versus conventional full sternotomy aortic valve replacement. Ann Thorac Surg. 2015;99:17–25.CrossRefPubMed
4.
Zurück zum Zitat Gilmanov D, Miceli A, Bevilacqua S, et al. Sutureless implantation of the perceval s aortic valve prosthesis through right anterior minithoracotomy. Ann Thorac Surg. 2013;96:2101–8.CrossRefPubMed Gilmanov D, Miceli A, Bevilacqua S, et al. Sutureless implantation of the perceval s aortic valve prosthesis through right anterior minithoracotomy. Ann Thorac Surg. 2013;96:2101–8.CrossRefPubMed
5.
Zurück zum Zitat Vola M, Albertini A, Campisi S, et al. Right anterior minithoracotomy aortic valve replacement with a sutureless bioprosthesis: Early outcomes and 1-year follow-up from 2 European centers. J Thorac Cardiovasc Surg. 2015;149:1052–7.CrossRefPubMed Vola M, Albertini A, Campisi S, et al. Right anterior minithoracotomy aortic valve replacement with a sutureless bioprosthesis: Early outcomes and 1-year follow-up from 2 European centers. J Thorac Cardiovasc Surg. 2015;149:1052–7.CrossRefPubMed
6.
Zurück zum Zitat Kocher AA, Laufer G, Haverich A, et al. One-year experience of the surgical treatment of aortic stenosis with a next generation surgical aortic valve (TRITON) trial: a prospective multicenter study of rapid-deployment aortic valve replacement with the Edwards intuity valve system. J Thorac Cardiovasc Surg. 2013;1:110–5.CrossRef Kocher AA, Laufer G, Haverich A, et al. One-year experience of the surgical treatment of aortic stenosis with a next generation surgical aortic valve (TRITON) trial: a prospective multicenter study of rapid-deployment aortic valve replacement with the Edwards intuity valve system. J Thorac Cardiovasc Surg. 2013;1:110–5.CrossRef
7.
Zurück zum Zitat Glauber M, Miceli A, Bevilacqua S, Farneti PA. Minmally invasive aortic valve replacement via right anterior minithoracotomy: early outcomes and midterm follow-up. J Thoracic Cardiovasc Surg. 2011;142:1577–9.CrossRef Glauber M, Miceli A, Bevilacqua S, Farneti PA. Minmally invasive aortic valve replacement via right anterior minithoracotomy: early outcomes and midterm follow-up. J Thoracic Cardiovasc Surg. 2011;142:1577–9.CrossRef
8.
Zurück zum Zitat Lamelas J. Minimal invasive aortic valve replacement: the “Miami Method”. Ann Cardiothorac Surg. 2015;4:71–7.PubMedPubMedCentral Lamelas J. Minimal invasive aortic valve replacement: the “Miami Method”. Ann Cardiothorac Surg. 2015;4:71–7.PubMedPubMedCentral
9.
Zurück zum Zitat Glauber M, Micelli I, Gilmannow D, et al. Right anterior minithoracotomy versus conventional aortic valve replacement: a propensity score matched study. J Thorac Cardiovasc Surg. 2013;5:1222–6.CrossRef Glauber M, Micelli I, Gilmannow D, et al. Right anterior minithoracotomy versus conventional aortic valve replacement: a propensity score matched study. J Thorac Cardiovasc Surg. 2013;5:1222–6.CrossRef
10.
Zurück zum Zitat Gilmanov D, Miceli A, Ferrarini M, et al. Aortic valve replacement through right anterior minithoracotomy: can sutureless technology improve clinical outcomes? Ann Thorac Surg. 2014;98:1585–92.CrossRefPubMed Gilmanov D, Miceli A, Ferrarini M, et al. Aortic valve replacement through right anterior minithoracotomy: can sutureless technology improve clinical outcomes? Ann Thorac Surg. 2014;98:1585–92.CrossRefPubMed
11.
Zurück zum Zitat Bowdish ME, Hui DS, Cleveland JD, et al. A comparison of aortic valve replacement via an anterior right minithoracotomy with standard sternotomy: a propensity score analysis of 492 patients. Eur J Cardiothorac Surg. 2016;49(2):456–63.CrossRefPubMed Bowdish ME, Hui DS, Cleveland JD, et al. A comparison of aortic valve replacement via an anterior right minithoracotomy with standard sternotomy: a propensity score analysis of 492 patients. Eur J Cardiothorac Surg. 2016;49(2):456–63.CrossRefPubMed
12.
Zurück zum Zitat Miceli A, Murzi M, Gilmanov D, et al. Minimally invasive aortic valve replacement using right minithoracotomy is associated with better outcomes than ministernotomy. J Thorac Cardiovasc Surg. 2014;148:133–7.CrossRefPubMed Miceli A, Murzi M, Gilmanov D, et al. Minimally invasive aortic valve replacement using right minithoracotomy is associated with better outcomes than ministernotomy. J Thorac Cardiovasc Surg. 2014;148:133–7.CrossRefPubMed
13.
Zurück zum Zitat Mikus E, Turci S, Calvi S, Ricci M, Dozza L, Del Giglio M. Aortic valve replacement through right minithoracotomy: is it really biologically minimally invasive? Ann Thorac Surg. 2015;99:826–30.CrossRefPubMed Mikus E, Turci S, Calvi S, Ricci M, Dozza L, Del Giglio M. Aortic valve replacement through right minithoracotomy: is it really biologically minimally invasive? Ann Thorac Surg. 2015;99:826–30.CrossRefPubMed
14.
Zurück zum Zitat Phan K, Xie A, Tsai YC, Black D, Di Eusanio M, Yan TD. Ministernotomy or minithoracotomy for minimally invasive aortic valve replacement: a Bayesian network meta-analysis. Ann Cardiothorac Surg. 2015;4:3–14.PubMedPubMedCentral Phan K, Xie A, Tsai YC, Black D, Di Eusanio M, Yan TD. Ministernotomy or minithoracotomy for minimally invasive aortic valve replacement: a Bayesian network meta-analysis. Ann Cardiothorac Surg. 2015;4:3–14.PubMedPubMedCentral
15.
Zurück zum Zitat Ghanta RK, Lapar DJ, Kern JA, et al. Minimally invasive aortic valve replacement provides equivalent outcomes at reduced cost compared with conventional aortic valve replacement: A real-world multi-institutional analysis. J Thorac Cardiovasc Surg. 2015;149:1060–5.CrossRefPubMedPubMedCentral Ghanta RK, Lapar DJ, Kern JA, et al. Minimally invasive aortic valve replacement provides equivalent outcomes at reduced cost compared with conventional aortic valve replacement: A real-world multi-institutional analysis. J Thorac Cardiovasc Surg. 2015;149:1060–5.CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Funkat A, Beckmann A, Lewandowski J, et al. Cardiac surgery in Germany during 2013: A report on behalf of the German Society for Thoracic and Cardiovascular Surgery. Thorac Cardiovasc Surg. 2014;62:380–92.CrossRefPubMed Funkat A, Beckmann A, Lewandowski J, et al. Cardiac surgery in Germany during 2013: A report on behalf of the German Society for Thoracic and Cardiovascular Surgery. Thorac Cardiovasc Surg. 2014;62:380–92.CrossRefPubMed
17.
Zurück zum Zitat Miceli A, Santarpino G, Pfeiffer S, et al. Minimally invasive aortic valve replacement with Perceval S sutureless valve: early outcomes and one-year survival from two European centers. J Thorac Cardiovasc Surg. 2014;148:2838–43.CrossRefPubMed Miceli A, Santarpino G, Pfeiffer S, et al. Minimally invasive aortic valve replacement with Perceval S sutureless valve: early outcomes and one-year survival from two European centers. J Thorac Cardiovasc Surg. 2014;148:2838–43.CrossRefPubMed
18.
Zurück zum Zitat Dalén M, Biancari F, Rubino AS, et al. Ministernotomy versus full sternotomy aortic valve replacement with sutureless bioprosthesis: a multicenter study. Ann Thorac Surg. 2015;99:524–31.CrossRefPubMed Dalén M, Biancari F, Rubino AS, et al. Ministernotomy versus full sternotomy aortic valve replacement with sutureless bioprosthesis: a multicenter study. Ann Thorac Surg. 2015;99:524–31.CrossRefPubMed
19.
Zurück zum Zitat Zoghbi WA, Chambers JB, Dumesnil JG, et al. Recommendations for evaluation of prosthetic valves with echocardiography and doppler ultrasound. J Am Soc Echocardiogr. 2009;22:975–1014.CrossRefPubMed Zoghbi WA, Chambers JB, Dumesnil JG, et al. Recommendations for evaluation of prosthetic valves with echocardiography and doppler ultrasound. J Am Soc Echocardiogr. 2009;22:975–1014.CrossRefPubMed
20.
Zurück zum Zitat Bleiziffer S, Ali A, Hettich IM, et al. Impact of the indexed effective orifice area on mid-term cardiac-related mortality after aortic valve replacement. Heart. 2010;96:865–71.CrossRefPubMed Bleiziffer S, Ali A, Hettich IM, et al. Impact of the indexed effective orifice area on mid-term cardiac-related mortality after aortic valve replacement. Heart. 2010;96:865–71.CrossRefPubMed
21.
Zurück zum Zitat Butchart EG, Gohlke-Bärwolf C, Antunes MJ, Working Groups on Valvular Heart Disease, Thrombosis, and Cardiac Rehabilitation and Exercise Physiology, European Society of Cardiology, et al. Recommendations for the management of patients after heart valve surgery. Eur Heart J. 2005;26:2463–71.CrossRefPubMed Butchart EG, Gohlke-Bärwolf C, Antunes MJ, Working Groups on Valvular Heart Disease, Thrombosis, and Cardiac Rehabilitation and Exercise Physiology, European Society of Cardiology, et al. Recommendations for the management of patients after heart valve surgery. Eur Heart J. 2005;26:2463–71.CrossRefPubMed
22.
Zurück zum Zitat Sharony R, Grossi EA, Saunders PC, et al. Propensity score analysis of a six-year experience with minimally invasive isolated aortic valve replacement. J Heart Valve Dis. 2004;13:887–93.PubMed Sharony R, Grossi EA, Saunders PC, et al. Propensity score analysis of a six-year experience with minimally invasive isolated aortic valve replacement. J Heart Valve Dis. 2004;13:887–93.PubMed
23.
Zurück zum Zitat Salis S, Mazzanti VV, Merli G, et al. Cardiopulmonary bypass duration is an independent predictor for morbidity and maortality after cardiac surgery. J Cardiothorac Vasc Anesth. 2008;22:814–22.CrossRefPubMed Salis S, Mazzanti VV, Merli G, et al. Cardiopulmonary bypass duration is an independent predictor for morbidity and maortality after cardiac surgery. J Cardiothorac Vasc Anesth. 2008;22:814–22.CrossRefPubMed
24.
Zurück zum Zitat Al-Sarraf N, Thalib L, Hughes A, et al. Cross-clamp time is an independent predictor of mortality and morbidity in low- and high-risk cardiac patients. Int J Surg. 2011;9:104–9.CrossRefPubMed Al-Sarraf N, Thalib L, Hughes A, et al. Cross-clamp time is an independent predictor of mortality and morbidity in low- and high-risk cardiac patients. Int J Surg. 2011;9:104–9.CrossRefPubMed
25.
Zurück zum Zitat D’Onofrio A, Rizzoli G, Messina A, et al. Conventional surgery, sutureless valves, and transapical aortic valve replacement: what is the best option for patients with aortic valve stenosis? A multicenter, propensity-matched analysis. J Thorac Cardiovasc Surg. 2013;146:1065–70.CrossRefPubMed D’Onofrio A, Rizzoli G, Messina A, et al. Conventional surgery, sutureless valves, and transapical aortic valve replacement: what is the best option for patients with aortic valve stenosis? A multicenter, propensity-matched analysis. J Thorac Cardiovasc Surg. 2013;146:1065–70.CrossRefPubMed
26.
Zurück zum Zitat D’Onofrio A, Messina A, Lorusso R, et al. Sutureless aortic valve replacement as an alternative treatment for patients belonging to the “gray zone” between transcatheter aortic valve implantation and conventional surgery: a propensity-matched, multicenter analysis. J Thorac Cardiovasc Surg. 2012;144:1010–6.CrossRefPubMed D’Onofrio A, Messina A, Lorusso R, et al. Sutureless aortic valve replacement as an alternative treatment for patients belonging to the “gray zone” between transcatheter aortic valve implantation and conventional surgery: a propensity-matched, multicenter analysis. J Thorac Cardiovasc Surg. 2012;144:1010–6.CrossRefPubMed
Metadaten
Titel
Rapid deployment valve system shortens operative times for aortic valve replacement through right anterior minithoracotomy
verfasst von
Constanze Bening
Khaled Hamouda
Mehmet Oezkur
Christoph Schimmer
Ina Schade
Armin Gorski
Ivan Aleksic
Rainer Leyh
Publikationsdatum
01.12.2017
Verlag
BioMed Central
Erschienen in
Journal of Cardiothoracic Surgery / Ausgabe 1/2017
Elektronische ISSN: 1749-8090
DOI
https://doi.org/10.1186/s13019-017-0598-0

Weitere Artikel der Ausgabe 1/2017

Journal of Cardiothoracic Surgery 1/2017 Zur Ausgabe

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

CME: 2 Punkte

Prof. Dr. med. Gregor Antoniadis Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

CME: 2 Punkte

Dr. med. Benjamin Meyknecht, PD Dr. med. Oliver Pieske Das Webinar S2e-Leitlinie „Distale Radiusfraktur“ beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

CME: 2 Punkte

Dr. med. Mihailo Andric
Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.