Skip to main content
Erschienen in: Magnetic Resonance Materials in Physics, Biology and Medicine 3/2019

07.02.2019 | Research Article

Realization of 19F MRI oximetry method using perfluorodecalin

verfasst von: Mikhail V. Gulyaev, Aleksandra V. Kuznetsova, Denis N. Silachev, Tatyana I. Danilina, Lev L. Gervits, Yury A. Pirogov

Erschienen in: Magnetic Resonance Materials in Physics, Biology and Medicine | Ausgabe 3/2019

Einloggen, um Zugang zu erhalten

Abstract

Objective

To identify the technical aspects of the potential use of clinically approved perfluorodecalin (PFD, C10F18) for 19F magnetic resonance imaging (MRI) oximetry method at high magnetic field 7.05 T.

Materials and methods

19F T1 measurements were made on a set of PFD samples with different oxygen contents (0%, 21%, and 100%) at room (21 °C) and body temperature (37 °C). In vivo MRI studies were carried out on one healthy rat and two rats with C6 brain glioma.

Results

The selective excitation of the magnetically equivalent 19F nuclei of CF2 groups of trans-isomer of PFD, which give a doublet at a frequency of about − 140 ppm (in relation the chemical shift of trifluoroacetic acid, which is − 76.55 ppm) should be done for correct implementation of 19F MRI oximetry method. The amount of PFD equal to 30 μl is the optimal for obtaining reliable data on the measured T1 values. In this case, the standard deviation of T1 does not exceed 5%. In vivo MRI studies showed that the values of the partial pressure of oxygen (pO2) decrease from normal values of about 38 mmHg (healthy brain) to almost 0 mmHg at the last stage of tumor growth.

Conclusion

The study showed the feasibility of the successful application of PFD for 19F MRI oximetry method.
Literatur
1.
Zurück zum Zitat Dardzinski BJ, Sotak CH (1994) Rapid tissue oxygen tension mapping using 19F inversion-recovery echo-planar imaging of perfluoro-15-crown-5-ether. Magn Reson Med 32(1):88–97CrossRefPubMed Dardzinski BJ, Sotak CH (1994) Rapid tissue oxygen tension mapping using 19F inversion-recovery echo-planar imaging of perfluoro-15-crown-5-ether. Magn Reson Med 32(1):88–97CrossRefPubMed
2.
Zurück zum Zitat Mignion L, Magat J, Schakman O, Marbaix E, Gallez B, Jordan BF (2013) Hexafluorobenzene in comparison with perfluoro-15-crown-5-ether for repeated monitoring of oxygenation using 19F MRI in a mouse model. Magn Reson Med 69(1):248–254CrossRefPubMed Mignion L, Magat J, Schakman O, Marbaix E, Gallez B, Jordan BF (2013) Hexafluorobenzene in comparison with perfluoro-15-crown-5-ether for repeated monitoring of oxygenation using 19F MRI in a mouse model. Magn Reson Med 69(1):248–254CrossRefPubMed
3.
Zurück zum Zitat Maevsky EI, Gervits LL (2008) Perfluorocarbon-based blood substitute—PERFTORAN Russian experience. Chim Oggi 26(3):34–37 Maevsky EI, Gervits LL (2008) Perfluorocarbon-based blood substitute—PERFTORAN Russian experience. Chim Oggi 26(3):34–37
4.
Zurück zum Zitat Parhami P, Fung BM (1983) Fluorine-19 relaxation study of perfluoro chemicals as oxygen carriers. J Phys Chem 87:1928–1931CrossRef Parhami P, Fung BM (1983) Fluorine-19 relaxation study of perfluoro chemicals as oxygen carriers. J Phys Chem 87:1928–1931CrossRef
5.
Zurück zum Zitat Jacoby C, Temmea S, Mayenfelsb F et al (2014) Probing different perfluorocarbons for in vivo inflammation imaging by 19F MRI: image reconstruction, biological half-lives and sensitivity. NMR Biomed 27(3):261–271CrossRefPubMed Jacoby C, Temmea S, Mayenfelsb F et al (2014) Probing different perfluorocarbons for in vivo inflammation imaging by 19F MRI: image reconstruction, biological half-lives and sensitivity. NMR Biomed 27(3):261–271CrossRefPubMed
6.
Zurück zum Zitat Kodibagkar VD, Wang X, Mason RP (2008) Physical principles of quantitative nuclear magnetic resonance oximetry. Front Biosci 13:1371–1384CrossRefPubMed Kodibagkar VD, Wang X, Mason RP (2008) Physical principles of quantitative nuclear magnetic resonance oximetry. Front Biosci 13:1371–1384CrossRefPubMed
7.
Zurück zum Zitat Grobben B, DeDeyn PP, Slegers H (2002) Rat C6 gliomas experimental model system for the study of glioblastoma growth and invasion. Cell Tissue Res 310(3):257–270CrossRefPubMed Grobben B, DeDeyn PP, Slegers H (2002) Rat C6 gliomas experimental model system for the study of glioblastoma growth and invasion. Cell Tissue Res 310(3):257–270CrossRefPubMed
8.
Zurück zum Zitat Dietrich O, Raya JG, Reeder SB, Reiser MF, Schoenberg SO (2007) Measurement of signal-to-noise ratios in MR images: Influence of multichannel coils, parallel imaging, and reconstruction filters. J Magn Reson Imaging 26:375–385CrossRefPubMed Dietrich O, Raya JG, Reeder SB, Reiser MF, Schoenberg SO (2007) Measurement of signal-to-noise ratios in MR images: Influence of multichannel coils, parallel imaging, and reconstruction filters. J Magn Reson Imaging 26:375–385CrossRefPubMed
9.
Zurück zum Zitat Einstein SA, Weegman BP, Firpo MT, Papas KK, Garwood M (2016) Development and validation of noninvasive magnetic resonance relaxometry for the in vivo assessment of tissue-engineered graft oxygenation. Tissue Eng Part C Methods 22(11):1009–1017CrossRefPubMedPubMedCentral Einstein SA, Weegman BP, Firpo MT, Papas KK, Garwood M (2016) Development and validation of noninvasive magnetic resonance relaxometry for the in vivo assessment of tissue-engineered graft oxygenation. Tissue Eng Part C Methods 22(11):1009–1017CrossRefPubMedPubMedCentral
10.
Zurück zum Zitat Carreau A, Hafny-Rahbi BE, Matejuk A, Grillon C, Kieda C (2011) Why is the partial oxygen pressure of human tissues a crucial parameter? Small molecules and hypoxia. J Cell Mol Med 15:1239–1253CrossRefPubMedPubMedCentral Carreau A, Hafny-Rahbi BE, Matejuk A, Grillon C, Kieda C (2011) Why is the partial oxygen pressure of human tissues a crucial parameter? Small molecules and hypoxia. J Cell Mol Med 15:1239–1253CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat Liu S, Shah SJ, Wilmes LJ et al (2011) Quantitative tissue oxygen measurement in multiple organs using 19F MRI in a rat model. Magn Reson Med 66:1722–1730CrossRefPubMedPubMedCentral Liu S, Shah SJ, Wilmes LJ et al (2011) Quantitative tissue oxygen measurement in multiple organs using 19F MRI in a rat model. Magn Reson Med 66:1722–1730CrossRefPubMedPubMedCentral
Metadaten
Titel
Realization of 19F MRI oximetry method using perfluorodecalin
verfasst von
Mikhail V. Gulyaev
Aleksandra V. Kuznetsova
Denis N. Silachev
Tatyana I. Danilina
Lev L. Gervits
Yury A. Pirogov
Publikationsdatum
07.02.2019
Verlag
Springer International Publishing
Erschienen in
Magnetic Resonance Materials in Physics, Biology and Medicine / Ausgabe 3/2019
Print ISSN: 0968-5243
Elektronische ISSN: 1352-8661
DOI
https://doi.org/10.1007/s10334-019-00739-1

Weitere Artikel der Ausgabe 3/2019

Magnetic Resonance Materials in Physics, Biology and Medicine 3/2019 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.