01.01.2004 | Original Article | Ausgabe 1/2004

European Journal of Applied Physiology 1/2004

Reliability and validity of the Biodex system 3 pro isokinetic dynamometer velocity, torque and position measurements

Zeitschrift:
European Journal of Applied Physiology > Ausgabe 1/2004
Autoren:
Joshua M. Drouin, Tamara C. Valovich-mcLeod, Sandra J. Shultz, Bruce M. Gansneder, David H. Perrin
Wichtige Hinweise
Disclosure
The Biodex dynamometer used for this investigation was donated to the laboratory by Biodex Medical Systems. The authors have no commercial or proprietary interest in this device.

Abstract

This study quantitatively assessed the mechanical reliability and validity of position, torque and velocity measurements of the Biodex System 3 isokinetic dynamometer. Trial-to-trial and day-to-day reliability were assessed during three trials on two separate days. To assess instrument validity, measurement of each variable using the Biodex System 3 dynamometer was compared to a criterion measure of position, torque and velocity. Position was assessed at 5° increments across the available range of motion of the dynamometer. Torque measures were assessed isometrically by hanging six different calibrated weights from the lever arm. Velocity was assessed (30°/s to 500°/s) across a 70° arc of motion by manually accelerating the weighted lever arm. With the exception of a systematic decrease in velocity at speeds of 300°/s and higher, the Biodex System 3 performed with acceptable mechanical reliability and validity on all variables tested.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de. Zusätzlich können Sie eine Zeitschrift Ihrer Wahl in gedruckter Form beziehen – ohne Aufpreis.

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 1/2004

European Journal of Applied Physiology 1/2004 Zur Ausgabe
  1. Sie können e.Med Allgemeinmedizin 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.

Neu im Fachgebiet Arbeitsmedizin