Open Access 17.07.2018 | Original Paper

Resting Motor Threshold, MEP and TEP Variability During Daytime

verfasst von: Esther M. ter Braack, Annika A. de Goede, Michel J. A. M. van Putten

Erschienen in: Brain Topography | Ausgabe 1/2019

Abstract

Humans show a variation in physiological processes during the day. To reliably assess (changes in) cortical excitability with transcranial magnetic stimulation (TMS), it is relevant to know the natural variation in TMS readouts during the day. In case of significant daytime variations, this should be taken into account when scheduling (follow-up) measurements. This study aims to evaluate the influence of the time of day on the resting motor threshold (RMT), motor evoked potential (MEP) and TMS evoked potential (TEP) in healthy controls. TMS–EMG–EEG was recorded in 16 healthy subjects. At both motor cortices, we administered 75 pulses at an intensity of 110% RMT. Subjects were stimulated during five sessions in one day (8:00 AM, 10:30 AM, 1:00 PM, 3:30 PM and 6:00 PM) while keeping the stimulation intensity constant. We compared the TEP waveforms between the five sessions with a cluster-based permutation analysis, and the RMT and MEP amplitude with rmANOVA. In general there were no significant differences between the five sessions in the RMT, MEP amplitude or TEP. Only for the left side, N100 amplitude was larger at 3:30 PM than 10:30 AM. The standard deviation of the P30 and N100 amplitude was significantly higher between subjects within one session than within single subjects during the day. The TEP is highly reproducible during the day, with a low intra-individual variation compared to the inter-individual variation. In addition, we found no significant variation of the RMT and MEP amplitude between multiple sessions on one day.
Hinweise
Handling Editor: Carlo Miniussi.

Introduction

Humans, and animals and plants as well, show a variation in physiological processes during the day. This circadian rhythm is regulated by our biological clock, resulting in diurnal fluctuations in for instance hormone secretion, blood pressure, but also alertness (http://​www.​Nobelprize.​org). Some epilepsies show a relationship with sleep stages or the transition between sleep and wakefulness, of which juvenile myoclonic epilepsy (JME) is the most typical example where seizures occur predominantly after awakening in the morning. Besides the interaction between sleep and epilepsy, the time of day is also correlated with seizure occurrence in some focal epilepsy types (Hofstra and de Weerd 2009; Mirzoev et al. 2012; van Campen et al. 2015). This diurnal distribution of seizures is mainly evident for temporal lobe epilepsy, with a peak in seizure occurrence in the late afternoon (Durazzo et al. 2008; Hofstra et al. 2009; Pavlova et al. 2004).
Epilepsy can be characterized as a disease resulting from an imbalance between cortical excitation and inhibition. Transcranial magnetic stimulation (TMS) is a well-established technique to non-invasively activate brain areas (Barker et al. 1985), and is a promising method to assess cortical excitability, which we here define as the strength of the response of cortical neurons to an external input. The resting motor threshold (RMT) or the MEP amplitude following a neuromodulatory paradigm (paired-pulse TMS) can be used as readouts of cortical excitability. Combining TMS with EEG has also become available (Ilmoniemi and Kičić 2010; Miniussi and Thut 2010), opening novel possibilities to study cortical excitability. A single TMS pulse induces a response in the EEG, and after averaging over multiple pulses, the TMS evoked potential (TEP) is obtained. TMS–EEG could provide a more direct measure of cortical excitability than measuring the MEP (Bonato et al. 2006; Ferreri et al. 2011; Ilmoniemi and Kičić 2010), as it is not influenced by the excitability of corticospinal and spinal neurons. The TEP has been shown to change after administration of GABA-ergic drugs as well as anti-epileptic drugs (Premoli et al. 2017, 2014).
TMS is considered to be a candidate tool for a new biomarker in epilepsy (Bauer et al. 2014; Chen et al. 2008; Engel 2008; Kimiskidis 2016; Manganotti and del Felice 2013). To reliably assess (changes in) cortical excitability, it is relevant to know the natural variation in TMS readouts during the day. When there is a significant effect of time-of-day, scheduling follow-up measurements should be performed at approximately the same time in the morning or afternoon. Previous studies have shown that the RMT and MEP remains constant during the day (Doeltgen and Ridding 2010; Koski et al. 2005; Lang et al. 2011; Strutton et al. 2003). It is known that the TEP changes when different brain areas are stimulated, and also when the stimulation intensity or stimulation angle is varied (Casarotto et al. 2010). While the TEP after stimulating the motor cortex shows high repeatability comparing the first and last part of a TMS–EEG session (Casarotto et al. 2010; Kerwin et al. 2018) and is highly reproducible after 1 week when stimulating parameters are kept constant (Casarotto et al. 2010; Lioumis et al. 2009), diurnal variations of the TEP have not been systematically studied.
The variation of one component of the TEP during daytime has been studied in healthy subjects as a part of larger experimental protocols focusing on the effects of sleep deprivation. The P30 of the TEP did not differ between 9 AM and 3 PM, but was significantly higher at 9 PM compared to 9 AM in five out of six subjects (Huber et al. 2012). In a subsequent study by the same research group, the amplitude and slope of the P30 was significantly lower at 9 PM compared to 5 PM, but no difference was found between 11 AM and 5 PM (Ly et al. 2016).
The variation of the TEP using multiple measurements during daytime has not been reported before. In this study we investigated the daytime variation of the RMT, MEP amplitude and the TEP after motor cortex stimulation by measuring these responses at 2.5 h intervals between 8 AM and 6 PM in a group of healthy volunteers.

Materials and Methods

The experimental protocol was approved by the local ethics committee (Medisch Spectrum Twente, Enschede, the Netherlands) and was in accordance with the declaration of Helsinki and the guidelines for the use of TMS in clinical practice and research (Rossi et al. 2009). All subjects gave written informed consent.

Subjects

Nineteen healthy subjects participated in this study. One of these 19 subjects had a syncope at the start of the first TMS session and was excluded from the study. The remaining 18 subjects tolerated the TMS protocol well. Part of the data from these 18 subjects was previously presented, for a different objective (ter Braack et al. 2013, 2016). Another two subjects were excluded from the analysis, since one subject had a missing session for both targets, and one subject had two missing sessions for one target, both due to technical problems with the navigation system. Sixteen subjects (11 males, mean age 28 years, all right-handed) were therefore included in the analysis.

Experimental Protocol

Subjects were seated in a chair, with their hands pronated in a relaxed position. They kept their eyes open, focusing on a marked point on the wall. Subjects were asked to refrain from alcohol 2 days and from caffeinated drinks 12 h prior to measurements. Subjects were only measured if they had a good night of sleep. Each subject underwent 1 day of measurements, divided in 5 sessions. The sessions took place at 8:00 AM, 10:30 AM, 1:00 PM, 3:30 PM and 6:00 PM. In three subjects we repeated the first session a week later to confirm previous findings of a good reproducibility (Lioumis et al. 2009).

Stimulation

Single biphasic TMS pulses, with pulse duration of 400 µs and inter-pulse interval of ~ 4 s, were delivered manually using a 70 mm figure-of-eight air film coil and a Magstim Rapid2 stimulator (The Magstim Company Ltd, Whitland, United Kingdom). The maximum stimulator output was 0.8 T. The coil was placed tangentially over the hot-spot of the abductor digiti minimi muscle (ADM) in the left and right hemisphere. The two targets are referred to as motor cortex left (MCL) and motor cortex right (MCR). The handle was pointing backwards and laterally at a 45° angle away from the midline. At both targets we applied 75 TMS pulses at a stimulation intensity of 110% of the RMT of the ADM hotspot. This stimulation intensity was kept constant during the day. The motor threshold was defined as the lowest stimulus intensity that produced at least five MEPs of at least 50 µV out of 10 consecutive stimuli (Rossini et al. 1994). In four of the 16 subjects, a TMS intensity of 110% RMT could not be given due to a too high threshold in session 1. In those subjects, the TMS intensity during the protocol was set to the maximum output of the stimulator (0.8 T), corresponding to 100–108% RMT. During TMS–EEG, all subjects wore protective earplugs, and noise created from the coil click was played through headphones at 95 dB to mask the sound of the TMS pulses (ter Braack et al. 2015). In addition, a thin layer of foam was placed between the coil and head of the subject to minimize bone conduction.

TMS Targeting

Positioning of the coil was achieved using a robot-navigated system (Smartmove, ANT Neuro, Enschede, Netherlands), with an accuracy of 1 mm in every direction. A headband carrying four passive reflective markers was fixed to the head of the subject and tracked by a Polaris infrared camera system (Northern Digital Inc., Waterloo, Ontario, Canada). The robot and the tracking system were registered to a common coordinate system using a calibration procedure. The robot-guided TMS coil was added to the coordinate system by registration of three reference positions on the coil using a tracking pointer. In all subjects, a 1.5 T MRI scan of the head was available. The MRI scan was used to create a subject-specific head model; this model was then registered to the subject’s head and the coordinate system by collecting three landmarks and 300 additional points on the scalp with a tracking pointer.

EEG and EMG Recording During TMS

The EEG was recorded continuously during TMS using a full-band amplifier (TMSi, Oldenzaal, Netherlands) and a TMS-compatible 64-electrode cap (ANT Neuro, Enschede, Netherlands). The EEG cap stayed in place during the whole day. Impedances were kept below 5 kΩ. The ground electrode was placed between electrode positions Fz and Fpz. We used a common average reference for the recordings. In our data, a single TMS pulse produced a magnetic stimulation artifact of 1–2 mV, lasting approximately 3 ms using the full-band amplifier. To determine the hotspot and RMT, surface electrodes were placed in a belly–tendon montage over the right and left ADM muscle. The ground electrode was placed on the upper side of the wrist. We recorded the EMG using an additional amplifier (TMSi, Oldenzaal, Netherlands) connected to the EEG amplifier, ensuring synchronized measurements. The EEG and EMG signals were low-pass filtered with an anti-aliasing filter with a cut-off frequency of 550 Hz and sampled at 2048 Hz.

Evoked Potential Analysis

EMG and EEG analysis was performed using Matlab (The Mathworks, Natick, MA, USA). To analyze the MEP, peak-to-peak amplitudes were calculated and averaged per session. Trials containing muscle pre-activation, defined as EMG activity larger than 50 µV in the 50 ms preceding a single pulse, were excluded.
TMS evoked potentials were analyzed using the common average reference. Trials were defined from 50 ms before to 300 ms after every TMS pulse, resulting in 75 trials for both targets and each session. We applied single-trial principal component analysis (PCA) to remove the first large TMS artifact, caused by the magnetic pulse, and the second TMS artifact, believed to be caused by muscle activation on the scalp. A detailed description of this PCA method can be found in a previous study (ter Braack et al. 2013). In short, we performed PCA using 40 calculated components on each individual trial, with the first component having the largest variance and the 40th component having the lowest variance. We then removed the first four components, containing the large amplitude artifacts, from the trial to obtain a signal which is almost artifact-free. After PCA, the trials were filtered with a fourth order Butterworth bandpass filter between 1 and 45 Hz and averaged per session.
To investigate the variation of the P30 and N100 amplitude during the day, we determined the standard deviation at the latency of the maximum amplitude of both components at electrode Cz. The standard deviation was determined for the response on group level for each session and then averaged over sessions, resulting in an average inter-individual variation of the P30 and N100 amplitude during the day. We also determined this standard deviation for each individual subject for the average response over all five sessions and then averaged over subjects, resulting in an average intra-individual variation of the P30 and N100 amplitude during the day.

Statistical Analysis

For all statistical analyses a p-value below 0.05 was considered statistically significant, except when a correction for multiple comparisons was applied.
RMT and mean MEP amplitude between the five sessions was compared using one-way repeated measures ANOVA with Greenhouse–Geisser correction for both left and right motor cortex stimulation. Two-way repeated measures ANOVA was used to test for differences between both hemispheres. Since RMT was occasionally above the maximum output of the stimulator, and statistical analysis of bounded variables is challenging, those subjects were excluded for RMT statistics. A total of 13 and 15 subjects were included in the RMT analysis for the left and right hemispheres, respectively.
We compared the standard deviation of the P30 and N100 on group level (five sessions) with the standard deviation of the P30 and N100 on single subject level during the day (16 subjects) using an independent t-test.
All subjects were included in the TEP statistics. To compare the reproducibility of the total TEP waveform (0–300 ms) between the five sessions, we used multiple dependent t-tests at the electrode level. A cluster-based permutation analysis (Maris and Oostenveld 2007) was applied, as implemented in FieldTrip (http://​fieldtrip.​fcdonders.​nl/​), which enables analysis of the whole waveform on all electrodes. In short, a dependent t-test comparing the TEPs from the five sessions was performed for each time sample and each EEG electrode. Only t-values with a clustering p-value < 0.05 were considered for clustering. Clustering of t-values was based on adjacent time bins and neighboring electrodes. Within each cluster, the t-values of the included electrodes were summed, and this sum was used for statistical comparison. A permutation test was performed, randomly assigning the TEPs from the 16 subjects to two different groups (for example session 1 and session 2 are now randomly shuffled) and repeating statistical testing for 1500 times. These permutation results are then combined to form a distribution of summed clusters t-values. Clusters in the original data set were considered to show a non-significant trend if < 5% of the permutations in the distribution had a cluster-level statistic larger than the statistic in the original data set, i.e. with an alpha p-value of < 0.05. Only clusters with a p-value < 0.005 were only considered statistically significant, as p-values were afterwards Bonferroni corrected for 10 comparisons (all sessions were compared to each other). The same procedure was repeated for the time-intervals 20–35 and 80–140 ms to evaluate the P30 and N100 components of the TEP in further detail.

Results

The mean RMT for the first session was 79% for the left hemisphere and 78% for the right hemisphere (Table 1). The RMT normalized with respect to the first session is presented in Fig. 1. The RMT showed no differences between the left and right hemisphere (F(1.91,22.86) = 0.09, p = 0.90) and did not change significantly during the day (left hemisphere: F(1.97,23.66) = 0.41, p = 0.67; right hemisphere: F(2.08,29.11) = 1.99, p = 0.15). The MEP amplitude showed no differences between the left and right hemisphere (F(2.26,33.82) = 0.16, p = 0.88) and did not change significantly during the day (left hemisphere: F(1.94,29.11) = 1.50, p = 0.24; right hemisphere: F(2.34,35.18) = 1.94, p = 0.15). In all subjects MEPs were evoked continuously, except in one where MEPs were absent (amplitude < 50 µV) for left hemisphere stimulation at 3:30 PM and 6:00 PM. The mean MEP amplitude normalized with respect to the first session is presented in Fig. 1.
Table 1
Resting motor thresholds and used TMS intensity (left/right) for all subjects
Subject
Session 1
Session 2
Session 3
Session 4
Session 5
TMS intensity
1
87/78
89/77
92/77
95/76
93/73
96/86
2
80 /65
76/68
86/67
85/68
84/68
88/71
3
70/67
72/67
72/68
71/72
72/72
77/74
4
70/82
72/81
72/80
73/78
73/83
77/91
5
86/87
79/88
75/93
75/93
76/100
95/96
6
65/61
58/60
62/61
59/60
58/57
72/68
7a
97a/97a
95/99
93/100
95/99
96/100
100/100
8
80/84
84/83
80/81
79/86
78/86
88/93
9a,b
94a/75
100/75
> 100/85
> 100/82
> 100/90
100/83
10
88/91
89/85
85/85
87/82
89/83
97/100
11a,b,c
93a/> 100a
96/100
> 100/> 100
94/> 100
95/88
100/100
12
74/70
75/73
77/71
78/71
83/70
82/77
13
77/76
64/65
68/71
66/76
64/73
85/85
14
84/73
86/73
88/74
86/76
92/74
93/81
15
74/75
78/74
78/78
80/78
79/80
82/83
16a,b
98a/93a
> 100/93
100/93
100/93
> 100/94
100/100
Resting motor threshold and applied TMS intensity during the protocol in % of maximal stimulator output (0.8 T) for left/right hemisphere.
aProtocol intensity of 110% RMT not possible
bSubject not included in RMT analysis left hemisphere
cSubject not included in RMT analysis right hemisphere
Figure 2 shows the TEP for all five sessions at electrode Cz averaged over subjects after stimulating the left motor cortex. The response was very constant during the day, showing a similar waveform in all five sessions in each subject. We found no significant differences in the TEP between the five sessions for MCR stimulation on a group level. Also for MCL stimulation, the majority of sessions showed no significant differences, except for 10:30 AM compared to 3:30 PM (p = 0.0046). One significant positive cluster was found left occipitally around 110–120 ms, corresponding to a larger (more negative) amplitude of the N100 component at 3:30 PM. When focusing on the N100, the same positive cluster was found to be significant (10:30 AM compared to 3:30 PM for MCL, p = 0.001), located left occipitally. The electrode with the largest amplitude difference within the cluster (PO3) is shown in Fig. 3. When focusing on the P30, no additional significant clusters were found. In addition we found some non-significant trends for the N100 8:00 AM versus 3:30 PM for MCL: positive cluster left occipitally (interval 0–300 ms (p = 0.02); interval 80–140 ms (p = 0.01)), and 8:00 AM versus 10:30 AM for MCR: negative cluster left occipitally (interval 80–140 ms (p = 0.02)) and the P30 10:30 AM versus 1:00 PM for MCL: negative cluster right frontally (interval 20–35 ms (p = 0.02)). The TEP after repeating the TMS protocol a week later also closely resembles the TEP from a week earlier (see Fig. 4).
We calculated the mean TEP with standard deviation over all subjects in the five sessions. In Fig. 2 the mean TEP with standard deviation at Cz is shown for left motor stimulation for session 1. The average standard deviation of the N100 component at Cz on a group level during the day was 4.53 µV (range 3.99–5.19 µV) for MCL and 6.16 µV (range 5.40–7.02 µV) for MCR. The mean TEP with SD over five sessions for each single subject is shown in Fig. 5. The average standard deviation of the N100 component at Cz on a single subject level during the day was 2.01 µV (range 0.40–4.27 µV) for MCL and 1.35 µV (range 0.27–3.11 µV) for MCR. The average standard deviation of the P30 component at Cz on a group level during the day was 5.45 µV (range 4.36–6.98 µV) for MCL and 4.42 µV (range 4.04–4.71 µV) for MCR. The average standard deviation of the P30 component at Cz on a single subject level during the day was 2.98 µV (range 0.58–8.48 µV) for MCL and 1.14 µV (range 0.36–2.73 µV) for MCR. Both for the N100 and the P30 the standard deviation on group level was significantly higher than the standard deviation on single subject level during the day (P30 MCL: p = 0.002; P30 MCR, N100 MCL, N100 MCR: all p < 0.001).

Discussion

In this study we applied TMS–EEG five times during the day in healthy volunteers, and measured the RMT, MEP amplitude and the TEP. We found that both RMT and MEP amplitude do not change significantly during the day. The TEP remained largely constant, except for the N100 which was more pronounced at 3:30 PM compared to 10:30 AM. The inter-individual variation of the TEP within one session is significantly larger than the intra-individual variation during the day.
The absence of significant variation of the RMT between two sessions on 1 day has been reported before (Doeltgen and Ridding 2010; Lang et al. 2011; Tamm et al. 2009). Measurements at multiple time points in a 10 h period showed that the RMT stays constant (Koski et al. 2005), and this stability of the RMT during daytime is now confirmed by our results. Practically, this means that in healthy volunteers the RMT only has to be determined at the start of a TMS session, even when this TMS session takes very long or consists of multiple measurements within a day. In our data the MEP amplitude did not change over sessions, in line with a previous report measuring the MEP over a 24 h period (Strutton et al. 2003).
The TEP had a characteristic waveform in all subjects with recognizable components at 15–30–45–60–100–180 ms. This response was very constant during the day, with only the N100 after MCL stimulation showing a significant difference between the session at 10:30 AM and 3:30 PM on a group level. This cluster was located left occipitally, just as the two other N100 clusters that turned out to be non-significant after correction. Two earlier reports did not find a significant change of the first large component of the TEP (P30) during daytime (Huber et al. 2012; Ly et al. 2016). However, these authors did not analyze any other components. We evaluated the whole TEP waveform on all electrodes, and found a significant difference in a different component, the N100. All other components, including the P30, did not show significant differences. We only found a non-significant trend for the P30, with a larger amplitude at 1:00 PM compared to 10:30 AM, which is in line with previous findings of a non-significant increase of P30 amplitude during daytime (Huber et al. 2012; Ly et al. 2016). The TEP has shown to be reproducible after 1 week (Casarotto et al. 2010; Lioumis et al. 2009), similar to our observations in three subjects.
Other TMS measures were also reported as being constant during the day, such as short interval cortical inhibition (SICI) and intra cortical facilitation (ICF) (Doeltgen and Ridding 2010; Lang et al. 2011; Pfutze et al. 2007). Both SICI and ICF are GABA-A mediated responses (Hanajima et al. 1998; Inghilleri et al. 1996; Kujirai et al. 1993). On the contrary, long interval cortical inhibition (LICI) as well as the cortical silent period (CSP) decreased in length during three sessions at 8 AM, 2 PM and 8 PM (Lang et al. 2011), suggesting that the amount of inhibition decreases during the day. Earlier studies did not find a significant change in CSP during seven sessions between 8 AM and 8 PM (Koski et al. 2005) or between an evening and a morning session (Pfutze et al. 2007). LICI is a GABA-B mediated response (McDonnell et al. 2006; Pierantozzi et al. 2004; Werhahn et al. 1999), while CSP is thought to consist of a GABA-A mediated part and a GABA-B mediated part (Inghilleri et al. 1996; Kimiskidis et al. 2006; Siebner et al. 1998; Stetkarova and Kofler 2013). The N100 has also shown to be GABA-B mediated (Premoli et al. 2014), and in our results the N100 was larger in amplitude in the afternoon compared to the morning, suggesting an increase instead of decrease in inhibition. Taking these results together, it appears that GABA-A mediated TMS responses do not fluctuate during the day, while for the GABA-B mediated LICI and N100, and the partly GABA-B mediated CSP, inconsistent results are found.
Contradicting results are reported for the time-dependency of cortical excitability measured by long-interval cortical inhibition (LICI) in JME. Excitability was decreased in the afternoon compared to the morning in drug-naïve JME patients (Badawy et al. 2009), but no difference in excitability was reported in an earlier study (Pfutze et al. 2007). In addition, a difference in excitability between morning and afternoon could not be found in a group of focal epilepsy (mainly temporal lobe epilepsy) patients (Badawy et al. 2009). This inconsistency in TMS findings in healthy subjects and epilepsy patients may be explained by differences in methodology, for example the (lack of) control of different Zeitgebers or the heterogeneity of subjects with regard to morning type and evening type. True circadian or daytime variations may only be found using a constant routine methodology (Duffy and Dijk 2002). In any case, for a TMS readout to be of clinical use, the influence of normal, daily variations should be small compared to the influence of a disease or a medication.
We further show that intra-individual variation during the day of the N100 and P30 amplitude was significantly smaller compared to the inter-individual variation within a session. This is in line with a previous report describing a large inter-individual variation in TMS measures (Koski et al. 2005). Large differences between subjects have also been described in recent publications evaluating the TEP before and after different (anti-epileptic) drugs (Premoli et al. 2017, 2014).
Our study has a few limitations. Some subjects had a very high RMT at the start and therefore it was not possible to use a relative stimulation intensity of 110% during session 1. Because we compared the TEP between sessions, and not between subjects, and the TEP is also present at stimulation intensities below RMT (Komssi et al. 2004), this most likely had no effect on the results. We used the same stimulation intensity during all five sessions, even though there were small (non-significant) changes in RMT between sessions. MEP amplitude, directly dependent on RMT, has been related to the amplitude of the N15-P30 and N100 component (Mäki and Ilmoniemi 2010b; Paus et al. 2001) and to a late response around 300 ms (Fecchio et al. 2017). However, there were no significant differences in MEP amplitude between sessions. On the one hand the constant stimulation intensity may have influenced the TEP amplitude, since the relative stimulation intensity deviated from 110% in sessions 2–5. Still, this did not result in a difference in TEP waveform during the day, probably also because the relative stimulation intensity was sometimes above and sometimes below 110%. On the other hand, keeping the stimulation intensity constant ensured that the contribution of auditory and somatosensory evoked potentials and muscle activation artifacts was similar for all sessions and did not influence our findings.
We used PCA to reduce the TMS pulse artifact and the muscle artifact from our data. Although both are not designed for time-locked data such as evoked potentials, ICA (Hamidi et al. 2010; Iwahashi et al. 2008; Korhonen et al. 2011; Rogasch et al. 2014) and PCA (Hernandez-Pavon et al. 2012; Mäki and Ilmoniemi 2010a; Rogasch et al. 2017) have been successfully used for removing artifacts from TMS–EEG. Also other techniques (Casula et al. 2017; Litvak et al. 2007; Morbidi et al. 2007) have been applied. All methods result in a reduction of TMS-related artifacts and thereby enable TEP analysis. Although we have shown that PCA is an effective method to reduce the TMS pulse artifact and the muscle artifact simultaneously (ter Braack et al. 2013), the method can possibly be further optimized. Indeed, some residual artifacts were still present, causing minor filtering effects around the time of the TMS pulse as can be seen in Figs. 2, 3, 4 and 5. We now removed the first four components in all subjects, but it may be better to decide on the number of components to be removed on a single subject level. When increasingly more PCA components are removed, the TEP components that have the same direction as the artifact also decrease in amplitude (Mäki and Ilmoniemi 2010a; ter Braack et al. 2013). Earlier research on the same dataset showed that our PCA approach reduces the TEP component around 100 ms (ter Braack et al. 2013). This is the time-frame where we found significant differences for a parieto-occipital cluster. Therefore, it has to be considered that the PCA filtering could have suppressed the TEP at areas where the TMS-artifact was more expressed, resulting in the appearance of a null-finding over e.g. motor areas. A larger reduction of the TEP over the stimulation area was also reported by other authors using PCA filtering (Mäki and Ilmoniemi 2010a). Despite of these drawbacks (Mutanen et al. 2016), PCA is still suggested as an effective method to reduce the large muscle artifacts before evaluating the TEP more closely (Rogasch et al. 2017). It is likely that a combination of different methods is needed to remove all artifacts while not affecting the TEP (Atluri et al. 2016; Rogasch et al. 2017; Wu et al. 2018).
In conclusion, we show that the RMT, MEP amplitude and TEP in healthy subjects are highly reproducible during daytime. No significant differences were found for both RMT and MEP amplitude, while only the N100 amplitude after left motor cortex stimulation was significantly larger at 3:30 PM compared to 10:30 AM. This implies that results from different sessions can be compared even though they are obtained at a different time of day. Large inter-individual differences still may cause difficulties in establishing normal ranges for TMS measures, which is important for developing clinical applications.

Acknowledgements

This research was funded by the Dutch PIDON grant, in which ANT Neuro participated as an industrial partner. This funding source played no role in the design and analysis of the study. Funding was also provided by TWIN foundation for neuromodulation.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Neurologie & Psychiatrie

Kombi-Abonnement

Mit e.Med Neurologie & Psychiatrie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

e.Med Neurologie

Kombi-Abonnement

Mit e.Med Neurologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes, den Premium-Inhalten der neurologischen Fachzeitschriften, inklusive einer gedruckten Neurologie-Zeitschrift Ihrer Wahl.

Literatur
Zurück zum Zitat Atluri S, Frehlich M, Mei Y, Garcia Dominguez L, Rogasch NC, Wong W et al (2016) TMSEEG: a MATLAB-based graphical user interface for processing electrophysiological signals during transcranial magnetic stimulation. Front Neural Circuits 10:78CrossRefPubMedPubMedCentral Atluri S, Frehlich M, Mei Y, Garcia Dominguez L, Rogasch NC, Wong W et al (2016) TMSEEG: a MATLAB-based graphical user interface for processing electrophysiological signals during transcranial magnetic stimulation. Front Neural Circuits 10:78CrossRefPubMedPubMedCentral
Zurück zum Zitat Badawy RAB, Macdonell RAL, Jackson GD, Berkovic SF (2009) Why do seizures in generalized epilepsy often occur in the morning? Neurology 73:218–222CrossRefPubMed Badawy RAB, Macdonell RAL, Jackson GD, Berkovic SF (2009) Why do seizures in generalized epilepsy often occur in the morning? Neurology 73:218–222CrossRefPubMed
Zurück zum Zitat Barker AT, Jalinous R, Freeston IL (1985) Non-invasive magnetic stimulation of human motor cortex. Lancet 1:1106–1107CrossRefPubMed Barker AT, Jalinous R, Freeston IL (1985) Non-invasive magnetic stimulation of human motor cortex. Lancet 1:1106–1107CrossRefPubMed
Zurück zum Zitat Bauer PR, Kalitzin S, Zijlmans M, Sander JW, Visser GH (2014) Cortical excitability as a potential clinical marker of epilepsy: a review of the clinical application of transcranial magnetic stimulation. Int J Neural Syst 24:e1430001CrossRef Bauer PR, Kalitzin S, Zijlmans M, Sander JW, Visser GH (2014) Cortical excitability as a potential clinical marker of epilepsy: a review of the clinical application of transcranial magnetic stimulation. Int J Neural Syst 24:e1430001CrossRef
Zurück zum Zitat Bonato C, Miniussi C, Rossini PM (2006) Transcranial magnetic stimulation and cortical evoked potentials: a TMS/EEG co-registration study. Clin Neurophysiol 117:1699–1707CrossRefPubMed Bonato C, Miniussi C, Rossini PM (2006) Transcranial magnetic stimulation and cortical evoked potentials: a TMS/EEG co-registration study. Clin Neurophysiol 117:1699–1707CrossRefPubMed
Zurück zum Zitat Casarotto S, Lauro LJR, Bellina V, Casali AG, Rosanova M, Pigorini A et al (2010) EEG responses to TMS are sensitive to changes in the perturbation parameters and repeatable over time. PLoS ONE 5:e10281CrossRefPubMedPubMedCentral Casarotto S, Lauro LJR, Bellina V, Casali AG, Rosanova M, Pigorini A et al (2010) EEG responses to TMS are sensitive to changes in the perturbation parameters and repeatable over time. PLoS ONE 5:e10281CrossRefPubMedPubMedCentral
Zurück zum Zitat Casula EP, Bertoldo A, Tarantino V, Maiella M, Koch G, Rothwell JC et al (2017) TMS-evoked long-lasting artefacts: a new adaptive algorithm for EEG signal correction. Clin Neurophysiol 128:1563–1574CrossRefPubMed Casula EP, Bertoldo A, Tarantino V, Maiella M, Koch G, Rothwell JC et al (2017) TMS-evoked long-lasting artefacts: a new adaptive algorithm for EEG signal correction. Clin Neurophysiol 128:1563–1574CrossRefPubMed
Zurück zum Zitat Chen R, Cros D, Curra A, Lazzaro VD, Lefaucheur J-P, Magistris MR et al (2008) The clinical diagnostic utility of transcranial magnetic stimulation: report of an IFCN committee. Clin Neurophysiol 119:504–532CrossRefPubMed Chen R, Cros D, Curra A, Lazzaro VD, Lefaucheur J-P, Magistris MR et al (2008) The clinical diagnostic utility of transcranial magnetic stimulation: report of an IFCN committee. Clin Neurophysiol 119:504–532CrossRefPubMed
Zurück zum Zitat Doeltgen SH, Ridding MC (2010) Behavioural exposure and sleep do not modify corticospinal and intracortical excitability in the human motor system. Clin Neurophysiol 121:448–452CrossRefPubMed Doeltgen SH, Ridding MC (2010) Behavioural exposure and sleep do not modify corticospinal and intracortical excitability in the human motor system. Clin Neurophysiol 121:448–452CrossRefPubMed
Zurück zum Zitat Duffy JF, Dijk DJ (2002) Getting through to circadian oscillators: why use constant routines? J Biol Rhythm 17:4–13CrossRef Duffy JF, Dijk DJ (2002) Getting through to circadian oscillators: why use constant routines? J Biol Rhythm 17:4–13CrossRef
Zurück zum Zitat Durazzo TS, Spencer SS, Duckrow RB, Novotny EJ, Spencer DD, Zaveri HP (2008) Temporal distributions of seizure occurrence from various epileptogenic regions. Neurology 70:1265–1271CrossRefPubMed Durazzo TS, Spencer SS, Duckrow RB, Novotny EJ, Spencer DD, Zaveri HP (2008) Temporal distributions of seizure occurrence from various epileptogenic regions. Neurology 70:1265–1271CrossRefPubMed
Zurück zum Zitat Engel J Jr (2008) Progress in epilepsy: reducing the treatment gap and the promise of biomarkers. Curr Opin Neurol 21:50–54CrossRef Engel J Jr (2008) Progress in epilepsy: reducing the treatment gap and the promise of biomarkers. Curr Opin Neurol 21:50–54CrossRef
Zurück zum Zitat Fecchio M, Pigorini A, Comanducci A, Sarasso S, Casarotto S, Premoli I et al (2017) The spectral features of EEG responses to transcranial magnetic stimulation of the primary motor cortex depend on the amplitude of the motor evoked potentials. PLoS ONE 12:e0184910CrossRefPubMedPubMedCentral Fecchio M, Pigorini A, Comanducci A, Sarasso S, Casarotto S, Premoli I et al (2017) The spectral features of EEG responses to transcranial magnetic stimulation of the primary motor cortex depend on the amplitude of the motor evoked potentials. PLoS ONE 12:e0184910CrossRefPubMedPubMedCentral
Zurück zum Zitat Ferreri F, Pasqualetti P, Määttä S, Ponzo D, Ferrarelli F, Tononi G et al (2011) Human brain connectivity during single and paired pulse transcranial magnetic stimulation. Neuroimage 54:90–102CrossRefPubMed Ferreri F, Pasqualetti P, Määttä S, Ponzo D, Ferrarelli F, Tononi G et al (2011) Human brain connectivity during single and paired pulse transcranial magnetic stimulation. Neuroimage 54:90–102CrossRefPubMed
Zurück zum Zitat Hamidi M, Slagter HA, Tononi G, Postle BR (2010) Brain responses evoked by high-frequency repetitive transcranial magnetic stimulation: an event-related potential study. Brain Stimul 3:2–14CrossRefPubMedPubMedCentral Hamidi M, Slagter HA, Tononi G, Postle BR (2010) Brain responses evoked by high-frequency repetitive transcranial magnetic stimulation: an event-related potential study. Brain Stimul 3:2–14CrossRefPubMedPubMedCentral
Zurück zum Zitat Hanajima R, Ugawa Y, Terao Y, Sakai K, Furubayashi T, Machii K, Kanazawa I (1998) Paired-pulse magnetic stimulation of the human motor cortex—differences among I-waves. J Physiol 509:607–618CrossRefPubMedPubMedCentral Hanajima R, Ugawa Y, Terao Y, Sakai K, Furubayashi T, Machii K, Kanazawa I (1998) Paired-pulse magnetic stimulation of the human motor cortex—differences among I-waves. J Physiol 509:607–618CrossRefPubMedPubMedCentral
Zurück zum Zitat Hernandez-Pavon JC, Metsomaa J, Mutanen T, Stenroos M, Mäki H, Ilmoniemi RJ, Sarvas J (2012) Uncovering neural independent components from highly artifactual TMS-evoked EEG data. J Neurosci Methods 209:144–157CrossRefPubMed Hernandez-Pavon JC, Metsomaa J, Mutanen T, Stenroos M, Mäki H, Ilmoniemi RJ, Sarvas J (2012) Uncovering neural independent components from highly artifactual TMS-evoked EEG data. J Neurosci Methods 209:144–157CrossRefPubMed
Zurück zum Zitat Hofstra WA, de Weerd AW (2009) The circadian rhythm and its interaction with human epilepsy: a review of literature. Sleep Med Rev 13:413–420CrossRef Hofstra WA, de Weerd AW (2009) The circadian rhythm and its interaction with human epilepsy: a review of literature. Sleep Med Rev 13:413–420CrossRef
Zurück zum Zitat Hofstra WA, Spetgens WP, Leijten FS, van Rijen PC, Gosselaar P, van der Palen J, de Weerd AW (2009) Diurnal rhythms in seizures detected by intracranial electrocorticographic monitoring: an observational study. Epilepsy Behav 14:617–621CrossRefPubMed Hofstra WA, Spetgens WP, Leijten FS, van Rijen PC, Gosselaar P, van der Palen J, de Weerd AW (2009) Diurnal rhythms in seizures detected by intracranial electrocorticographic monitoring: an observational study. Epilepsy Behav 14:617–621CrossRefPubMed
Zurück zum Zitat Huber R, Maki H, Rosanova M, Casarotto S, Canali P, Casali AG et al (2012) Human cortical excitability increases with time awake. Cereb Cortex 23:1–7CrossRef Huber R, Maki H, Rosanova M, Casarotto S, Canali P, Casali AG et al (2012) Human cortical excitability increases with time awake. Cereb Cortex 23:1–7CrossRef
Zurück zum Zitat Ilmoniemi RJ, Kičić D (2010) Methodology for combined TMS and EEG. Brain Topogr 22:233–248CrossRefPubMed Ilmoniemi RJ, Kičić D (2010) Methodology for combined TMS and EEG. Brain Topogr 22:233–248CrossRefPubMed
Zurück zum Zitat Inghilleri M, Berardelli A, Marchetti P, Manfredi M (1996) Effects of diazepam, baclofen and thiopental on the silent period evoked by transcranial magnetic stimulation in humans. Exp Brain Res 109:467–472CrossRefPubMed Inghilleri M, Berardelli A, Marchetti P, Manfredi M (1996) Effects of diazepam, baclofen and thiopental on the silent period evoked by transcranial magnetic stimulation in humans. Exp Brain Res 109:467–472CrossRefPubMed
Zurück zum Zitat Iwahashi M, Arimatsu T, Ueno S, Iramina K (2008) Differences in evoked EEG by transcranial magnetic stimulation at various stimulus points on the head. Conf Proc IEEE Eng Med Biol Soc 2008:2570–2573PubMed Iwahashi M, Arimatsu T, Ueno S, Iramina K (2008) Differences in evoked EEG by transcranial magnetic stimulation at various stimulus points on the head. Conf Proc IEEE Eng Med Biol Soc 2008:2570–2573PubMed
Zurück zum Zitat Kerwin LJ, Keller CJ, Wu W, Narayan M, Etkin A (2018) Test-retest reliability of transcranial magnetic stimulation EEG evoked potentials. Brain Stimul 11:536–544CrossRefPubMed Kerwin LJ, Keller CJ, Wu W, Narayan M, Etkin A (2018) Test-retest reliability of transcranial magnetic stimulation EEG evoked potentials. Brain Stimul 11:536–544CrossRefPubMed
Zurück zum Zitat Kimiskidis VK (2016) TMS coupled with EEG biomarker of the future. Revue Neurologique 172:123–126CrossRefPubMed Kimiskidis VK (2016) TMS coupled with EEG biomarker of the future. Revue Neurologique 172:123–126CrossRefPubMed
Zurück zum Zitat Kimiskidis VK, Papagiannopoulos S, Kazis DA, Sotirakoglou K, Vasiliadis G, Zara F et al (2006) Lorazepam-induced effects on silent period and corticomotor excitability. Exp Brain Res 173:603–611CrossRefPubMed Kimiskidis VK, Papagiannopoulos S, Kazis DA, Sotirakoglou K, Vasiliadis G, Zara F et al (2006) Lorazepam-induced effects on silent period and corticomotor excitability. Exp Brain Res 173:603–611CrossRefPubMed
Zurück zum Zitat Komssi S, Kähkönen S, Ilmoniemi RJ (2004) The effect of stimulus intensity on brain responses evoked by transcranial magnetic stimulation. Hum Brain Mapp 21:154–164CrossRefPubMed Komssi S, Kähkönen S, Ilmoniemi RJ (2004) The effect of stimulus intensity on brain responses evoked by transcranial magnetic stimulation. Hum Brain Mapp 21:154–164CrossRefPubMed
Zurück zum Zitat Korhonen RJ, Hernandez-Pavon JC, Metsomaa J, Mäki H, Ilmoniemi RJ, Sarvas J (2011) Removal of large muscle artifacts from transcranial magnetic stimulation-evoked EEG by independent component analysis. Med Biol Eng Comput 49:397–407CrossRefPubMed Korhonen RJ, Hernandez-Pavon JC, Metsomaa J, Mäki H, Ilmoniemi RJ, Sarvas J (2011) Removal of large muscle artifacts from transcranial magnetic stimulation-evoked EEG by independent component analysis. Med Biol Eng Comput 49:397–407CrossRefPubMed
Zurück zum Zitat Koski L, Schrader LM, Wu AD, Stern JM (2005) Normative data on changes in transcranial magnetic stimulation measures over a ten hour period. Clin Neurophysiol 116:2099–2109CrossRefPubMed Koski L, Schrader LM, Wu AD, Stern JM (2005) Normative data on changes in transcranial magnetic stimulation measures over a ten hour period. Clin Neurophysiol 116:2099–2109CrossRefPubMed
Zurück zum Zitat Kujirai T, Caramia MD, Rothwell JC, Day BL, Thompson PD, Ferbert A et al (1993) Corticocortical inhibition in human motor cortex. J Physiol 471:501–519CrossRefPubMedPubMedCentral Kujirai T, Caramia MD, Rothwell JC, Day BL, Thompson PD, Ferbert A et al (1993) Corticocortical inhibition in human motor cortex. J Physiol 471:501–519CrossRefPubMedPubMedCentral
Zurück zum Zitat Lang N, Rothkegel H, Reiber H, Hasan A, Sueske E, Tergau F et al (2011) Circadian modulation of GABA-mediated cortical inhibition. Cereb Cortex 21:2299–2306CrossRefPubMed Lang N, Rothkegel H, Reiber H, Hasan A, Sueske E, Tergau F et al (2011) Circadian modulation of GABA-mediated cortical inhibition. Cereb Cortex 21:2299–2306CrossRefPubMed
Zurück zum Zitat Lioumis P, Kičić D, Savolainen P, Mäkelä JP, Kähkönen S (2009) Reproducibility of TMS-evoked EEG responses. Hum Brain Mapp 30:1387–1396CrossRefPubMed Lioumis P, Kičić D, Savolainen P, Mäkelä JP, Kähkönen S (2009) Reproducibility of TMS-evoked EEG responses. Hum Brain Mapp 30:1387–1396CrossRefPubMed
Zurück zum Zitat Litvak V, Komssi S, Scherg M, Hoechstetter K, Classen J, Zaaroor M et al (2007) Artifact correction and source analysis of early electroencephalographic responses evoked by transcranial magnetic stimulation over primary motor cortex. Neuroimage 37:56–70CrossRefPubMed Litvak V, Komssi S, Scherg M, Hoechstetter K, Classen J, Zaaroor M et al (2007) Artifact correction and source analysis of early electroencephalographic responses evoked by transcranial magnetic stimulation over primary motor cortex. Neuroimage 37:56–70CrossRefPubMed
Zurück zum Zitat Ly JQ, Gaggioni G, Chellappa SL, Papachilleos S, Brzozowski A, Borsu C et al (2016) Circadian regulation of human cortical excitability. Nat Commun 7:e11828CrossRef Ly JQ, Gaggioni G, Chellappa SL, Papachilleos S, Brzozowski A, Borsu C et al (2016) Circadian regulation of human cortical excitability. Nat Commun 7:e11828CrossRef
Zurück zum Zitat Mäki H, Ilmoniemi RJ (2010a) Projecting out muscle artifacts from TMS-evoked EEG. Neuroimage 54:2706–2710CrossRefPubMed Mäki H, Ilmoniemi RJ (2010a) Projecting out muscle artifacts from TMS-evoked EEG. Neuroimage 54:2706–2710CrossRefPubMed
Zurück zum Zitat Mäki H, Ilmoniemi RJ (2010b) The relationship between peripheral and early cortical activation induced by transcranial magnetic stimulation. Neurosci Lett 478:24–28CrossRefPubMed Mäki H, Ilmoniemi RJ (2010b) The relationship between peripheral and early cortical activation induced by transcranial magnetic stimulation. Neurosci Lett 478:24–28CrossRefPubMed
Zurück zum Zitat Manganotti P, del Felice A (2013) New perspectives in transcranial magnetic stimulation: epilepsy, consciousness and the perturbational approach. Behav Neurol 27:155–167CrossRefPubMedPubMedCentral Manganotti P, del Felice A (2013) New perspectives in transcranial magnetic stimulation: epilepsy, consciousness and the perturbational approach. Behav Neurol 27:155–167CrossRefPubMedPubMedCentral
Zurück zum Zitat Maris E, Oostenveld R (2007) Nonparametric statistical testing of EEG- and MEG-data. J Neurosci Methods 164:177–190CrossRefPubMed Maris E, Oostenveld R (2007) Nonparametric statistical testing of EEG- and MEG-data. J Neurosci Methods 164:177–190CrossRefPubMed
Zurück zum Zitat McDonnell MN, Orekhov Y, Ziemann U (2006) The role of GABA(B) receptors in intracortical inhibition in the human motor cortex. Exp Brain Res 173:86–93CrossRefPubMed McDonnell MN, Orekhov Y, Ziemann U (2006) The role of GABA(B) receptors in intracortical inhibition in the human motor cortex. Exp Brain Res 173:86–93CrossRefPubMed
Zurück zum Zitat Miniussi C, Thut G (2010) Combining TMS and EEG offers new prospects in cognitive neuroscience. Brain Topogr 22:249–256CrossRefPubMed Miniussi C, Thut G (2010) Combining TMS and EEG offers new prospects in cognitive neuroscience. Brain Topogr 22:249–256CrossRefPubMed
Zurück zum Zitat Mirzoev A, Bercovici E, Stewart LS, Cortez MA, Snead OC 3rd, Desrocher M (2012) Circadian profiles of focal epileptic seizures: a need for reappraisal. Seizure 21:412–416CrossRefPubMed Mirzoev A, Bercovici E, Stewart LS, Cortez MA, Snead OC 3rd, Desrocher M (2012) Circadian profiles of focal epileptic seizures: a need for reappraisal. Seizure 21:412–416CrossRefPubMed
Zurück zum Zitat Morbidi F, Garulli A, Prattichizzo D, Rizzo C, Manganotti P, Rossi S (2007) Off-line removal of TMS-induced artifacts on human electroencephalography by Kalman filter. J Neurosci Methods 162:293–302CrossRefPubMed Morbidi F, Garulli A, Prattichizzo D, Rizzo C, Manganotti P, Rossi S (2007) Off-line removal of TMS-induced artifacts on human electroencephalography by Kalman filter. J Neurosci Methods 162:293–302CrossRefPubMed
Zurück zum Zitat Mutanen TP, Kukkonen M, Nieminen JO, Stenroos M, Sarvas J, Ilmoniemi RJ (2016) Recovering TMS-evoked EEG responses masked by muscle artifacts. Neuroimage 139:157–166CrossRefPubMed Mutanen TP, Kukkonen M, Nieminen JO, Stenroos M, Sarvas J, Ilmoniemi RJ (2016) Recovering TMS-evoked EEG responses masked by muscle artifacts. Neuroimage 139:157–166CrossRefPubMed
Zurück zum Zitat Paus T, Sipila PK, Strafella AP (2001) Synchronization of neuronal activity in the human primary motor cortex by transcranial magnetic stimulation: an EEG study. J Neurophysiol 86:1983–1990CrossRefPubMed Paus T, Sipila PK, Strafella AP (2001) Synchronization of neuronal activity in the human primary motor cortex by transcranial magnetic stimulation: an EEG study. J Neurophysiol 86:1983–1990CrossRefPubMed
Zurück zum Zitat Pavlova MK, Shea SA, Bromfield EB (2004) Day/night patterns of focal seizures. Epilepsy Behav 5:44–49CrossRefPubMed Pavlova MK, Shea SA, Bromfield EB (2004) Day/night patterns of focal seizures. Epilepsy Behav 5:44–49CrossRefPubMed
Zurück zum Zitat Pfutze M, Reis J, Haag A, John D, Hattemer K, Oertel WH et al (2007) Lack of differences of motorcortical excitability in the morning as compared to the evening in juvenile myoclonic epilepsy—a study using transcranial magnetic stimulation. Epilepsy Res 74:239–242CrossRefPubMed Pfutze M, Reis J, Haag A, John D, Hattemer K, Oertel WH et al (2007) Lack of differences of motorcortical excitability in the morning as compared to the evening in juvenile myoclonic epilepsy—a study using transcranial magnetic stimulation. Epilepsy Res 74:239–242CrossRefPubMed
Zurück zum Zitat Pierantozzi M, Marciani MG, Palmieri MG, Brusa L, Galati S, Caramia MD et al (2004) Effect of Vigabatrin on motor responses to transcranial magnetic stimulation: an effective tool to investigate in vivo GABAergic cortical inhibition in humans. Brain Res 1028:1–8CrossRefPubMed Pierantozzi M, Marciani MG, Palmieri MG, Brusa L, Galati S, Caramia MD et al (2004) Effect of Vigabatrin on motor responses to transcranial magnetic stimulation: an effective tool to investigate in vivo GABAergic cortical inhibition in humans. Brain Res 1028:1–8CrossRefPubMed
Zurück zum Zitat Premoli I, Castellanos N, Rivolta D, Belardinelli P, Bajo R, Zipser C et al (2014) TMS-EEG signatures of GABAergic neurotransmission in the human cortex. J Neurosci 34:5603–5612CrossRefPubMedPubMedCentral Premoli I, Castellanos N, Rivolta D, Belardinelli P, Bajo R, Zipser C et al (2014) TMS-EEG signatures of GABAergic neurotransmission in the human cortex. J Neurosci 34:5603–5612CrossRefPubMedPubMedCentral
Zurück zum Zitat Premoli I, Biondi A, Carlesso S, Rivolta D, Richardson MP (2017) Lamotrigine and levetiracetam exert a similar modulation of TMS-evoked EEG potentials. Epilepsia 58:42–50CrossRefPubMed Premoli I, Biondi A, Carlesso S, Rivolta D, Richardson MP (2017) Lamotrigine and levetiracetam exert a similar modulation of TMS-evoked EEG potentials. Epilepsia 58:42–50CrossRefPubMed
Zurück zum Zitat Rogasch NC, Thomson RH, Farzan F, Fitzgibbon BM, Bailey NW, Hernandez-Pavon JC et al (2014) Removing artefacts from TMS-EEG recordings using independent component analysis: importance for assessing prefrontal and motor cortex network properties. Neuroimage 101:425–439CrossRefPubMed Rogasch NC, Thomson RH, Farzan F, Fitzgibbon BM, Bailey NW, Hernandez-Pavon JC et al (2014) Removing artefacts from TMS-EEG recordings using independent component analysis: importance for assessing prefrontal and motor cortex network properties. Neuroimage 101:425–439CrossRefPubMed
Zurück zum Zitat Rogasch NC, Sullivan C, Thomson RH, Rose NS, Bailey NW, Fitzgerald PB et al (2017) Analysing concurrent transcranial magnetic stimulation and electroencephalographic data: a review and introduction to the open-source TESA software. Neuroimage 147:934–951CrossRefPubMed Rogasch NC, Sullivan C, Thomson RH, Rose NS, Bailey NW, Fitzgerald PB et al (2017) Analysing concurrent transcranial magnetic stimulation and electroencephalographic data: a review and introduction to the open-source TESA software. Neuroimage 147:934–951CrossRefPubMed
Zurück zum Zitat Rossi S, Hallett M, Rossini PM, Pascual-Leone A (2009) Safety of TMSCG. Safety, ethical considerations, and application guidelines for the use of transcranial magnetic stimulation in clinical practice and research. Clin Neurophysiol 120:2008–2039CrossRefPubMedPubMedCentral Rossi S, Hallett M, Rossini PM, Pascual-Leone A (2009) Safety of TMSCG. Safety, ethical considerations, and application guidelines for the use of transcranial magnetic stimulation in clinical practice and research. Clin Neurophysiol 120:2008–2039CrossRefPubMedPubMedCentral
Zurück zum Zitat Rossini PM, Barker AT, Berardelli A, Caramia MD, Caruso G, Cracco RQ et al (1994) Non-invasive electrical and magnetic stimulation of the brain, spinal cord and roots: basic principles and procedures for routine clinical application. Report of an IFCN committee. Electroencephalogr Clin Neurophysiol 91:79–92CrossRefPubMed Rossini PM, Barker AT, Berardelli A, Caramia MD, Caruso G, Cracco RQ et al (1994) Non-invasive electrical and magnetic stimulation of the brain, spinal cord and roots: basic principles and procedures for routine clinical application. Report of an IFCN committee. Electroencephalogr Clin Neurophysiol 91:79–92CrossRefPubMed
Zurück zum Zitat Siebner HR, Dressnandt J, Auer C, Conrad B (1998) Continuous intrathecal baclofen infusion induced a marked increase of the transcranially evoked silent period. Muscle Nerve 21:1209–1212CrossRefPubMed Siebner HR, Dressnandt J, Auer C, Conrad B (1998) Continuous intrathecal baclofen infusion induced a marked increase of the transcranially evoked silent period. Muscle Nerve 21:1209–1212CrossRefPubMed
Zurück zum Zitat Stetkarova I, Kofler M (2013) Differential effect of baclofen on cortical and spinal inhibitory circuits. Clin Neurophysiol 124:339–345CrossRefPubMed Stetkarova I, Kofler M (2013) Differential effect of baclofen on cortical and spinal inhibitory circuits. Clin Neurophysiol 124:339–345CrossRefPubMed
Zurück zum Zitat Strutton P, Catley M, Davey N (2003) Stability of corticospinal excitability and grip force in intrinsic hand muscles in man over a 24-h period. Physiol Behav 79:679–682CrossRefPubMed Strutton P, Catley M, Davey N (2003) Stability of corticospinal excitability and grip force in intrinsic hand muscles in man over a 24-h period. Physiol Behav 79:679–682CrossRefPubMed
Zurück zum Zitat Tamm AS, Lagerquist O, Ley AL, Collins DF (2009) Chronotype influences diurnal variations in the excitability of the human motor cortex and the ability to generate torque during a maximum voluntary contraction. J Biol Rhythms 24:211–224CrossRefPubMed Tamm AS, Lagerquist O, Ley AL, Collins DF (2009) Chronotype influences diurnal variations in the excitability of the human motor cortex and the ability to generate torque during a maximum voluntary contraction. J Biol Rhythms 24:211–224CrossRefPubMed
Zurück zum Zitat ter Braack EM, de Jonge B, van Putten MJ (2013) Reduction of TMS induced artifacts in EEG using principal component analysis. IEEE Trans Neural Syst Rehabil Eng 21:376–382CrossRefPubMed ter Braack EM, de Jonge B, van Putten MJ (2013) Reduction of TMS induced artifacts in EEG using principal component analysis. IEEE Trans Neural Syst Rehabil Eng 21:376–382CrossRefPubMed
Zurück zum Zitat ter Braack EM, de Vos CC, van Putten MJ (2015) Masking the auditory evoked potential in TMS–EEG: a comparison of various methods. Brain Topogr 28:520–528CrossRefPubMed ter Braack EM, de Vos CC, van Putten MJ (2015) Masking the auditory evoked potential in TMS–EEG: a comparison of various methods. Brain Topogr 28:520–528CrossRefPubMed
Zurück zum Zitat ter Braack EM, Koopman AW, van Putten MJ (2016) Early TMS evoked potentials in epilepsy: a pilot study. Clin Neurophysiol 127:3025–3032CrossRefPubMed ter Braack EM, Koopman AW, van Putten MJ (2016) Early TMS evoked potentials in epilepsy: a pilot study. Clin Neurophysiol 127:3025–3032CrossRefPubMed
Zurück zum Zitat van Campen JS, Valentijn FA, Jansen FE, Joels M, Braun KP (2015) Seizure occurrence and the circadian rhythm of cortisol: a systematic review. Epilepsy Behav 47:132–137CrossRef van Campen JS, Valentijn FA, Jansen FE, Joels M, Braun KP (2015) Seizure occurrence and the circadian rhythm of cortisol: a systematic review. Epilepsy Behav 47:132–137CrossRef
Zurück zum Zitat Werhahn KJ, Kunesch E, Noachtar S, Benecke R, Classen J (1999) Differential effects on motorcortical inhibition induced by blockade of GABA uptake in humans. J Physiol 517:591–597CrossRefPubMedPubMedCentral Werhahn KJ, Kunesch E, Noachtar S, Benecke R, Classen J (1999) Differential effects on motorcortical inhibition induced by blockade of GABA uptake in humans. J Physiol 517:591–597CrossRefPubMedPubMedCentral
Zurück zum Zitat Wu W, Keller CJ, Rogasch NC, Longwell P, Shpigel E, Rolle CE, Etkin A (2018) ARTIST: a fully automated artifact rejection algorithm for single-pulse TMS-EEG data. Hum Brain Mapp 39:1607–1625CrossRefPubMed Wu W, Keller CJ, Rogasch NC, Longwell P, Shpigel E, Rolle CE, Etkin A (2018) ARTIST: a fully automated artifact rejection algorithm for single-pulse TMS-EEG data. Hum Brain Mapp 39:1607–1625CrossRefPubMed
Metadaten
Titel
Resting Motor Threshold, MEP and TEP Variability During Daytime
verfasst von
Esther M. ter Braack
Annika A. de Goede
Michel J. A. M. van Putten
Publikationsdatum
17.07.2018
Verlag
Springer US
Erschienen in
Brain Topography / Ausgabe 1/2019
Print ISSN: 0896-0267
Elektronische ISSN: 1573-6792
DOI
https://doi.org/10.1007/s10548-018-0662-7

Weitere Artikel der Ausgabe 1/2019

Brain Topography 1/2019 Zur Ausgabe

Leitlinien kompakt für die Neurologie

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Neurologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.