Abstract

Objectives

The aim of the study was to compare the retentive behaviors of double crowns with different designs and material compositions before and after artificial aging.

Materials and methods

Six pairs of double crowns were fabricated: telescopic crowns 0° made of high-noble metal (group A) or non-precious metal (group B), telescopic zirconia copings with secondary crowns made of electroplated gold 2° (group C), crowns with friction pins 2° made of non-precious metal (group D) or zirconia (group E), and conical crowns 6° made of high-noble metal (group F). Retention forces were assessed before and after artificial aging, and after axial and non-axial loading.

Results

Initially, specimens in group D (13.9 N), B (12.5 N), and E (12.2 N) exhibited the highest retention forces. Retention forces in groups A (9.6 N), C (7.4 N), and F (6.0 N) were statistically significantly lower than those of the other groups (p < 0.05). After artificial aging, double crowns with additional retention elements exhibited the highest retention forces. The largest retention force losses were evident in groups A (70%), B (64%), C (39%), and F (47%).

Conclusions

Double crowns with different designs and made of different materials exhibited different retention forces and different long-term retentive behavior. The highest retention force losses were evident in double crowns with more extended surface contact, such as telescopic crowns.

Clinical significance

Telescopic crowns with additional retention elements were more resistant to wearing than double crowns without additional retention elements. An additional clinical benefit might be the quick and easy possibility of enhancing retention.
Literatur
1.
Zurück zum Zitat Dexter JE (1883) The cap plate: a new appliance in mechanical dentistry. Dent Cosmos 25(7):344–350 Dexter JE (1883) The cap plate: a new appliance in mechanical dentistry. Dent Cosmos 25(7):344–350
2.
Zurück zum Zitat Wenz H-J, Kern M (2007) Long-term behavior of double crowns. Quintessenz Zahntech 33(12):1482–1494 Wenz H-J, Kern M (2007) Long-term behavior of double crowns. Quintessenz Zahntech 33(12):1482–1494
3.
Zurück zum Zitat Verma R, Joda T, Brägger U, Wittneben J-G (2013) A systematic review of the clinical performance of tooth-retained and implant-retained double crown prostheses with a follow-up of ≥ 3 years. J Prosthodont 22:2–12CrossRefPubMed Verma R, Joda T, Brägger U, Wittneben J-G (2013) A systematic review of the clinical performance of tooth-retained and implant-retained double crown prostheses with a follow-up of ≥ 3 years. J Prosthodont 22:2–12CrossRefPubMed
4.
Zurück zum Zitat ELsyad MA, Elsaadawy MG, Abdoou AM, Habib AA (2013) Effect of different implant positions on strain developed around four implants supporting a mandibular overdenture with rigid telescopic copings. Quintessence Int 44(9):679–686PubMed ELsyad MA, Elsaadawy MG, Abdoou AM, Habib AA (2013) Effect of different implant positions on strain developed around four implants supporting a mandibular overdenture with rigid telescopic copings. Quintessence Int 44(9):679–686PubMed
5.
Zurück zum Zitat Rammelsberg P, Bernhart G, Bermejo JL, Schmitter M, Schwarz S (2014) Prognosis of implants and abutment teeth under combined tooth-implant-supported and solely implant-supported double-crown-retained removable dental prostheses. Clin Oral Implants Res 25:813–818CrossRefPubMed Rammelsberg P, Bernhart G, Bermejo JL, Schmitter M, Schwarz S (2014) Prognosis of implants and abutment teeth under combined tooth-implant-supported and solely implant-supported double-crown-retained removable dental prostheses. Clin Oral Implants Res 25:813–818CrossRefPubMed
6.
Zurück zum Zitat Rinke S, Ziebolz D, Ratka-Krüger P, Frisch E (2014) Clinical outcome of double crown-retained mandibular removable dentures supported by a combination of residual teeth and strategic implants. J Prosthodont. https://doi.org/10.1111/jopr.12214 Rinke S, Ziebolz D, Ratka-Krüger P, Frisch E (2014) Clinical outcome of double crown-retained mandibular removable dentures supported by a combination of residual teeth and strategic implants. J Prosthodont. https://​doi.​org/​10.​1111/​jopr.​12214
7.
Zurück zum Zitat Schwindling FS, Dittmann B, Rammelsberg P (2014) Double-crown-retained removable dental prostheses: a retrospective study of survival and complications. J Prosthet Dent 112(3):488–493CrossRefPubMed Schwindling FS, Dittmann B, Rammelsberg P (2014) Double-crown-retained removable dental prostheses: a retrospective study of survival and complications. J Prosthet Dent 112(3):488–493CrossRefPubMed
8.
Zurück zum Zitat Szentpetery V, Lautenschläger C, Setz JM (2012) Frictional telescopic crowns in severely reduced dentitions: a 5-year clinical outcome study. Int J Prosthodont 25:217–220PubMed Szentpetery V, Lautenschläger C, Setz JM (2012) Frictional telescopic crowns in severely reduced dentitions: a 5-year clinical outcome study. Int J Prosthodont 25:217–220PubMed
9.
Zurück zum Zitat Wenz HJ, Hertrampf K, Lehmann KM (2001) Clinical longevity of removable partial dentures retained by telescopic crowns: outcome of the double crown with clearance fit. Int J Prosthodont 14(3):207–213PubMed Wenz HJ, Hertrampf K, Lehmann KM (2001) Clinical longevity of removable partial dentures retained by telescopic crowns: outcome of the double crown with clearance fit. Int J Prosthodont 14(3):207–213PubMed
10.
Zurück zum Zitat Wagner B, Kern M (2000) Clinical evaluation of removable partial dentures 10 years after insertion: success rates, hygienic problems and technical failures. Clin Oral Investig 4:74–80CrossRefPubMed Wagner B, Kern M (2000) Clinical evaluation of removable partial dentures 10 years after insertion: success rates, hygienic problems and technical failures. Clin Oral Investig 4:74–80CrossRefPubMed
11.
Zurück zum Zitat Behr M, Hofmann E, Rosentritt M, Lang R, Handel G (2000) Technical failure rates of double crown-retained removable partial dentures. Clin Oral Investig 4:87–90CrossRefPubMed Behr M, Hofmann E, Rosentritt M, Lang R, Handel G (2000) Technical failure rates of double crown-retained removable partial dentures. Clin Oral Investig 4:87–90CrossRefPubMed
12.
Zurück zum Zitat Igarashi Y, Goto T (1997) Ten-year follow-up study for conical crown retained dentures. Int J Prosthodont 10:149–155PubMed Igarashi Y, Goto T (1997) Ten-year follow-up study for conical crown retained dentures. Int J Prosthodont 10:149–155PubMed
13.
Zurück zum Zitat Bergman B, Ericson A, Molin M (1996) Long-term clinical results after treatment with conical crown-retained dentures. Int J Prosthodont 9:533–538PubMed Bergman B, Ericson A, Molin M (1996) Long-term clinical results after treatment with conical crown-retained dentures. Int J Prosthodont 9:533–538PubMed
14.
Zurück zum Zitat Stober T, Danner D, Bömicke W, Hassel AJ (2015) Improvement of oral health-related quality-of-life by use of different kinds of double-crown-retained removable partial dentures. Acta Odontol Scand 20:1–6 Stober T, Danner D, Bömicke W, Hassel AJ (2015) Improvement of oral health-related quality-of-life by use of different kinds of double-crown-retained removable partial dentures. Acta Odontol Scand 20:1–6
15.
Zurück zum Zitat Eisenmann E, Mokabberi A, Walter MH, Freesmeyer WB (2004) Improving the fit of implant-supported superstructures using the spark erosion technique. Int J Oral Maxillofac Implants 19(6):810–818PubMed Eisenmann E, Mokabberi A, Walter MH, Freesmeyer WB (2004) Improving the fit of implant-supported superstructures using the spark erosion technique. Int J Oral Maxillofac Implants 19(6):810–818PubMed
16.
Zurück zum Zitat Rübeling G (1999) New techniques in spark erosion: the solution to an accurately fitting screw-retained implant restoration. Quintessence Int 30(1):37–48 Rübeling G (1999) New techniques in spark erosion: the solution to an accurately fitting screw-retained implant restoration. Quintessence Int 30(1):37–48
17.
Zurück zum Zitat Weber H, Frank G (1993) Spark erosion procedure: a method for extensive combined fixed and removable prosthodontic care. J Prosthet Dent 69(2):222–227CrossRefPubMed Weber H, Frank G (1993) Spark erosion procedure: a method for extensive combined fixed and removable prosthodontic care. J Prosthet Dent 69(2):222–227CrossRefPubMed
18.
Zurück zum Zitat Bayer S, Zuziak W, Kraus D, Keilig L, Stark H, Enkling N (2011) Conical crowns with electroplated gold copings: retention force changes caused by wear and combined off-axial load. Clin Oral Implants Res 22:323–329CrossRefPubMed Bayer S, Zuziak W, Kraus D, Keilig L, Stark H, Enkling N (2011) Conical crowns with electroplated gold copings: retention force changes caused by wear and combined off-axial load. Clin Oral Implants Res 22:323–329CrossRefPubMed
19.
Zurück zum Zitat Ohkawa S, Okane H, Nagasawa T, Tsuru H (1990) Changes in retention of various telescope crown assemblies over long term use. J Prosthet Dent 64:153–158CrossRefPubMed Ohkawa S, Okane H, Nagasawa T, Tsuru H (1990) Changes in retention of various telescope crown assemblies over long term use. J Prosthet Dent 64:153–158CrossRefPubMed
20.
Zurück zum Zitat Alsabeeha NHM, Payne AGT, Swain MV (2009) Attachment systems for mandibular two-implant overdentures: a review of in vitro investigations on retention and wear features. Int J Prosthodont 22(5):429–440PubMed Alsabeeha NHM, Payne AGT, Swain MV (2009) Attachment systems for mandibular two-implant overdentures: a review of in vitro investigations on retention and wear features. Int J Prosthodont 22(5):429–440PubMed
21.
Zurück zum Zitat Chung KH, Chung CY, Cagna DR, Cronin RJ (2004) Retention characteristics of attachment systems for implant overdentures. J Prosthodont 13(4):221–226CrossRefPubMed Chung KH, Chung CY, Cagna DR, Cronin RJ (2004) Retention characteristics of attachment systems for implant overdentures. J Prosthodont 13(4):221–226CrossRefPubMed
22.
Zurück zum Zitat Sarnat AE (1983) The efficiency of cobalt samarium (Co5Sm) magnets as retention units for overdentures. J Dent 11(4):324–333CrossRefPubMed Sarnat AE (1983) The efficiency of cobalt samarium (Co5Sm) magnets as retention units for overdentures. J Dent 11(4):324–333CrossRefPubMed
23.
Zurück zum Zitat Besimo CH, Graber G, Flühler M (1996) Retention force changes in implant-supported titanium telescope crowns over long-term use in vitro. J Oral Rehabil 23:372–378CrossRefPubMed Besimo CH, Graber G, Flühler M (1996) Retention force changes in implant-supported titanium telescope crowns over long-term use in vitro. J Oral Rehabil 23:372–378CrossRefPubMed
24.
Zurück zum Zitat Beuer F, Edelhoff D, Gernet W, Naumann M (2010) Parameters affecting retentive force of electroformed double-crown systems. Clin Oral Investig 14:129–135CrossRefPubMed Beuer F, Edelhoff D, Gernet W, Naumann M (2010) Parameters affecting retentive force of electroformed double-crown systems. Clin Oral Investig 14:129–135CrossRefPubMed
25.
Zurück zum Zitat Gurbulak AG, Kilic K, Eroglu Z, Gercekcioglu E, Kesim B (2013) Evaluation of the retention force of double crowns used in combination with a galvanoforming and casting fabrication technique. J Prosthodont 22:63–68CrossRefPubMed Gurbulak AG, Kilic K, Eroglu Z, Gercekcioglu E, Kesim B (2013) Evaluation of the retention force of double crowns used in combination with a galvanoforming and casting fabrication technique. J Prosthodont 22:63–68CrossRefPubMed
26.
Zurück zum Zitat Engels J, Schubert O, Güth JF, Hoffmann M, Jauernig CH, Erdelt K, Stimmelmayr M, Beuer F (2013) Wear behavior of different double-crown system. Clin Oral Investig 17:503–510CrossRefPubMed Engels J, Schubert O, Güth JF, Hoffmann M, Jauernig CH, Erdelt K, Stimmelmayr M, Beuer F (2013) Wear behavior of different double-crown system. Clin Oral Investig 17:503–510CrossRefPubMed
27.
Zurück zum Zitat Sakai Y, Takahashi H, Iwasaki N, Igarashi Y (2011) Effects of surface roughness and tapered angle of cone crown telescopic system on retentive force. Dent Mater J 30(5):635–641CrossRefPubMed Sakai Y, Takahashi H, Iwasaki N, Igarashi Y (2011) Effects of surface roughness and tapered angle of cone crown telescopic system on retentive force. Dent Mater J 30(5):635–641CrossRefPubMed
28.
Zurück zum Zitat Güngör MA, Artunc C, Sonugelen M (2004) Parameters affecting retentive force of conus crowns. J Oral Rehabil 31:271–277CrossRefPubMed Güngör MA, Artunc C, Sonugelen M (2004) Parameters affecting retentive force of conus crowns. J Oral Rehabil 31:271–277CrossRefPubMed
29.
Zurück zum Zitat Minagi S, Natsuaki N, Nishigawa G, Sato T (1999) New telescopic crown design for removable partial dentures. J Prosthet Dent 8:684–688CrossRef Minagi S, Natsuaki N, Nishigawa G, Sato T (1999) New telescopic crown design for removable partial dentures. J Prosthet Dent 8:684–688CrossRef
30.
Zurück zum Zitat Weigl P, Hahn L, Lauer HC (2000) Advanced biomaterials used for a new telescopic retainer for removable dentures: Ceramics vs. electroplated gold copings: Part I. In vitro tribology effects. J Biomed Mater Res 53:320–336CrossRefPubMed Weigl P, Hahn L, Lauer HC (2000) Advanced biomaterials used for a new telescopic retainer for removable dentures: Ceramics vs. electroplated gold copings: Part I. In vitro tribology effects. J Biomed Mater Res 53:320–336CrossRefPubMed
31.
Zurück zum Zitat Dillschneider T, Nothdurft FP, Abed-Rabbo M, Mitov G, Pospiech PR (2009) In vitro-investigations on the wear behavior of different double crown systems. Dent Mater 25:e20CrossRef Dillschneider T, Nothdurft FP, Abed-Rabbo M, Mitov G, Pospiech PR (2009) In vitro-investigations on the wear behavior of different double crown systems. Dent Mater 25:e20CrossRef
32.
Zurück zum Zitat Turp I, Bozdağ E, Sünbüloğlu E, Kahruman C, Yusufoğlu I, Bayraktar G (2014) Retention and surface changes of zirconia primary crowns with secondary crowns different materials. Clin Oral Investig 18:2023–2035CrossRefPubMed Turp I, Bozdağ E, Sünbüloğlu E, Kahruman C, Yusufoğlu I, Bayraktar G (2014) Retention and surface changes of zirconia primary crowns with secondary crowns different materials. Clin Oral Investig 18:2023–2035CrossRefPubMed
33.
Zurück zum Zitat Dabrowa T, Dobrowolska A, Wieleba W (2013) The role of friction in the mechanism of retaining the partial removable dentures with double crown systems. Acta of Bioeng Biomech 15(4):43–48 Dabrowa T, Dobrowolska A, Wieleba W (2013) The role of friction in the mechanism of retaining the partial removable dentures with double crown systems. Acta of Bioeng Biomech 15(4):43–48
35.
Zurück zum Zitat Körber KH (1983) Dynamischer Mechanismus von Parodontium und Gewebestrukturen unter herausnehmbarem Zahnersatz. Dtsch Zahnärztl Z 38:975–985PubMed Körber KH (1983) Dynamischer Mechanismus von Parodontium und Gewebestrukturen unter herausnehmbarem Zahnersatz. Dtsch Zahnärztl Z 38:975–985PubMed
36.
Zurück zum Zitat Behr M, Rosentritt M, Taubenhansl P, Kolbeck C, Handel G (2005) Fracture resistance of fiber-reinforced composite restorations with different framework design. Acta Odontol Scand 63:153–157CrossRefPubMed Behr M, Rosentritt M, Taubenhansl P, Kolbeck C, Handel G (2005) Fracture resistance of fiber-reinforced composite restorations with different framework design. Acta Odontol Scand 63:153–157CrossRefPubMed
37.
Zurück zum Zitat Heintze SD, Cavalleri A, Zellweger G, Büchler A, Zappini G (2008) Fracture frequency of all-ceramic crowns during dynamic loading in a chewing simulator using different loading and luting protocols. Dent Mater 24:1352–1361CrossRefPubMed Heintze SD, Cavalleri A, Zellweger G, Büchler A, Zappini G (2008) Fracture frequency of all-ceramic crowns during dynamic loading in a chewing simulator using different loading and luting protocols. Dent Mater 24:1352–1361CrossRefPubMed
38.
Zurück zum Zitat Nejatidanesh F, Momeni G, Savabi O (2009) Flexural strength of interim resin materials for fixed prosthodontics. J Prosthodont 18:507–511CrossRefPubMed Nejatidanesh F, Momeni G, Savabi O (2009) Flexural strength of interim resin materials for fixed prosthodontics. J Prosthodont 18:507–511CrossRefPubMed
39.
Zurück zum Zitat Stappert CF, Chitmongkolsuk S, Silva NR, Att W, Strub JR (2008) Effect of mouth-motion fatigue and thermal cycling on the marginal accuracy of partial coverage restorations made of various dental materials. Dent Mater 24:1248–1257CrossRefPubMed Stappert CF, Chitmongkolsuk S, Silva NR, Att W, Strub JR (2008) Effect of mouth-motion fatigue and thermal cycling on the marginal accuracy of partial coverage restorations made of various dental materials. Dent Mater 24:1248–1257CrossRefPubMed
40.
Zurück zum Zitat Bayer S, Stark H, Gölz L, Keilig L, Kraus D, Hansen A, Enkling N (2012) Telescopic crowns: extra-oral and intra-oral retention force measurement—in vitro/in vivo correlation. Gerodontology 29:e340–e347CrossRefPubMed Bayer S, Stark H, Gölz L, Keilig L, Kraus D, Hansen A, Enkling N (2012) Telescopic crowns: extra-oral and intra-oral retention force measurement—in vitro/in vivo correlation. Gerodontology 29:e340–e347CrossRefPubMed
42.
Zurück zum Zitat Bayer S, Stark H, Gölz L, Keilig L, Kraus D, Hansen A, Enkling N (2012) Clinical retention force development of double crowns. Clin Oral Investig 16:407–411CrossRefPubMed Bayer S, Stark H, Gölz L, Keilig L, Kraus D, Hansen A, Enkling N (2012) Clinical retention force development of double crowns. Clin Oral Investig 16:407–411CrossRefPubMed
Metadaten
Titel
Retention force of removable partial dentures with different double crowns
verfasst von
Christin Arnold
Jeremias Hey
Jürgen M. Setz
Arne F. Boeckler
Ramona Schweyen
Publikationsdatum
03.11.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
Clinical Oral Investigations / Ausgabe 4/2018
Print ISSN: 1432-6981
Elektronische ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-017-2224-x

Weitere Artikel der Ausgabe 4/2018

Clinical Oral Investigations 4/2018 Zur Ausgabe

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.