28.03.2019 | Original Article • KNEE - ARTHROPLASTY

Rotational component alignment in patient-specific total knee arthroplasty compared with conventional cutting instrument

verfasst von: Nuttawut Chanalithichai, Nattapol Tammachote, Chane Jitapunkul, Supakit Kanitnate

Erschienen in: European Journal of Orthopaedic Surgery & Traumatology | Ausgabe 6/2019

Einloggen, um Zugang zu erhalten

Abstract

Background

Although many studies investigated the accuracy of customized cutting block (CCB), the data on rotational alignment are still lacking. The study aimed to assess whether CCB improved the component rotational position compared with conventional cutting instrument (CCI) using computed tomography scanning.

Methods

Eighty-six of 102 total knee arthroplasties from the previous randomized study were analyzed. The outcomes were rotational position of the femoral and tibial components, frequency of outliers and intra-class correlation coefficient.

Results

The mean femoral component rotation was not different between CCB versus CCI: 0.9° ± 0.8° versus 1.1° ± 1.1° (P = 0.29). Both groups had similar outlier frequencies: 2% (CCB) versus 2% (CCI) (P = 0.74). CCB had nearly 1° less mean tibial component deviation compared with CCI (P < 0.001): (1) dorsal tangent reference (DTR): 0.7° ± 0.8° versus 1.5° ± 1.0°, and (2) tibial trans-epicondylar reference (TTR): 0.5° ± 0.9° versus 1.4° ± 1.1°. Outlier frequencies were similar: (1) DTR: 0% CCB versus 5% CCI (P = 0.24), and (2) TTR: 5% in CCB versus 12% CCI (P = 0.20). Measurements based on tibial tubercle showed that CCB had ~ 1.4° less mean tibial component deviation compared with CCI: 0.3° ± 1.4° versus 1.7° ± 1.6° (P < 0.001) with a corresponding, less frequency of outliers: 0% versus 19% (P = 0.002). However, there was poor intra-observer reproducibility (0.61).

Conclusions

CCB did not improve femoral component rotational alignment compared with CCI nor affect outlier frequency, but it marginally improved the accuracy of tibial rotational alignment. The tibial tubercle reference point had poor intra-observer reproducibility.
Literatur
3.
Zurück zum Zitat Ritter MA, Faris PM, Keating EM, Meding JB (1994) Postoperative alignment of total knee replacement. Its effect on survival. Clin Orthop Relat Res 299:153–156 Ritter MA, Faris PM, Keating EM, Meding JB (1994) Postoperative alignment of total knee replacement. Its effect on survival. Clin Orthop Relat Res 299:153–156
4.
Zurück zum Zitat Bargren JH, Blaha JD, Freeman MA (1983) Alignment in total knee arthroplasty. Correlated biomechanical and clinical observations. Clin Orthop Relat Res 173:178–183 Bargren JH, Blaha JD, Freeman MA (1983) Alignment in total knee arthroplasty. Correlated biomechanical and clinical observations. Clin Orthop Relat Res 173:178–183
5.
Zurück zum Zitat Lotke PA, Ecker ML (1977) Influence of positioning of prosthesis in total knee replacement. J Bone Joint Surg Am 59(1):77–79CrossRefPubMed Lotke PA, Ecker ML (1977) Influence of positioning of prosthesis in total knee replacement. J Bone Joint Surg Am 59(1):77–79CrossRefPubMed
6.
Zurück zum Zitat Akagi M, Matsusue Y, Mata T, Asada Y, Horiguchi M, Iida H, Nakamura T (1999) Effect of rotational alignment on patellar tracking in total knee arthroplasty. Clin Orthop Relat Res 366:155–163CrossRef Akagi M, Matsusue Y, Mata T, Asada Y, Horiguchi M, Iida H, Nakamura T (1999) Effect of rotational alignment on patellar tracking in total knee arthroplasty. Clin Orthop Relat Res 366:155–163CrossRef
7.
Zurück zum Zitat Berger RA, Crossett LS, Jacobs JJ, Rubash HE (1998) Malrotation causing patellofemoral complications after total knee arthroplasty. Clin Orthop Relat Res 356:144–153CrossRef Berger RA, Crossett LS, Jacobs JJ, Rubash HE (1998) Malrotation causing patellofemoral complications after total knee arthroplasty. Clin Orthop Relat Res 356:144–153CrossRef
9.
Zurück zum Zitat Lombardi AV Jr, Berend KR, Adams JB (2008) Patient-specific approach in total knee arthroplasty. Orthopedics 31(9):927–930CrossRefPubMed Lombardi AV Jr, Berend KR, Adams JB (2008) Patient-specific approach in total knee arthroplasty. Orthopedics 31(9):927–930CrossRefPubMed
11.
Zurück zum Zitat Nam D, McArthur BA, Cross MB, Pearle AD, Mayman DJ, Haas SB (2012) Patient-specific instrumentation in total knee arthroplasty: a review. J Knee Surg 25(3):213–219CrossRefPubMed Nam D, McArthur BA, Cross MB, Pearle AD, Mayman DJ, Haas SB (2012) Patient-specific instrumentation in total knee arthroplasty: a review. J Knee Surg 25(3):213–219CrossRefPubMed
24.
Zurück zum Zitat Berger RA, Rubash HE, Seel MJ, Thompson WH, Crossett LS (1993) Determining the rotational alignment of the femoral component in total knee arthroplasty using the epicondylar axis. Clin Orthop Relat Res 286:40–47 Berger RA, Rubash HE, Seel MJ, Thompson WH, Crossett LS (1993) Determining the rotational alignment of the femoral component in total knee arthroplasty using the epicondylar axis. Clin Orthop Relat Res 286:40–47
25.
Zurück zum Zitat Barrack RL, Schrader T, Bertot AJ, Wolfe MW, Myers L (2001) Component rotation and anterior knee pain after total knee arthroplasty. Clin Orthop Relat Res 392:46–55CrossRef Barrack RL, Schrader T, Bertot AJ, Wolfe MW, Myers L (2001) Component rotation and anterior knee pain after total knee arthroplasty. Clin Orthop Relat Res 392:46–55CrossRef
28.
Zurück zum Zitat Luring C, Perlick L, Bathis H, Tingart M, Grifka J (2007) The effect of femoral component rotation on patellar tracking in total knee arthroplasty. Orthopedics 30(11):965–967PubMed Luring C, Perlick L, Bathis H, Tingart M, Grifka J (2007) The effect of femoral component rotation on patellar tracking in total knee arthroplasty. Orthopedics 30(11):965–967PubMed
30.
Zurück zum Zitat Rhoads DD, Noble PC, Reuben JD, Mahoney OM, Tullos HS (1990) The effect of femoral component position on patellar tracking after total knee arthroplasty. Clin Orthop Relat Res 260:43–51CrossRef Rhoads DD, Noble PC, Reuben JD, Mahoney OM, Tullos HS (1990) The effect of femoral component position on patellar tracking after total knee arthroplasty. Clin Orthop Relat Res 260:43–51CrossRef
Metadaten
Titel
Rotational component alignment in patient-specific total knee arthroplasty compared with conventional cutting instrument
verfasst von
Nuttawut Chanalithichai
Nattapol Tammachote
Chane Jitapunkul
Supakit Kanitnate
Publikationsdatum
28.03.2019
Verlag
Springer Paris
Erschienen in
European Journal of Orthopaedic Surgery & Traumatology / Ausgabe 6/2019
Print ISSN: 1633-8065
Elektronische ISSN: 1432-1068
DOI
https://doi.org/10.1007/s00590-019-02423-5

Weitere Artikel der Ausgabe 6/2019

European Journal of Orthopaedic Surgery & Traumatology 6/2019 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.