Safety and efficacy of a two-screw cephalomedullary nail for intertrochanteric femur fracture fixation: a retrospective case series in 264 patients

verfasst von: Boris A. Zelle, Antonio J. Webb, Christopher Matson, Michael Morwood, Khang H. Dang, Samuel S. Ornell, Gabrielle Gostigian, Cody M. Ramirez, Hassan Mir

Erschienen in: Patient Safety in Surgery | Ausgabe 1/2018

Abstract

Introduction

Recent advances have led to the design of a new cephalomedullary nail, which aims to decrease the risk of failures in patients with intertrochanteric hip fractures by allowing for insertion of two interdigitating screws into the head segment. The goal of this study is to evaluate the safety and efficacy of this two-screw cephalomedullary nailing system.

Patients/participants

Patients 18 years of age and older who underwent intramedullary nailing of their intertrochanteric femoral fracture using the InterTAN nailing system (Smith and Nephew, Memphis, TN) from 2012 to 2016 were included in this retrospective study which was performed at two urban certified level-1 trauma centers and one urban certified level-3 trauma center. The study data was collected through a retrospective chart review and review of the existing radiographic studies. Primary outcome measure was mechanical hardware failure and screw cutout. Secondary outcome measures included nonunion, malunion, medical and surgical complications.

Results

A total of 264 patients were included in this analysis. Two patients (0.75%) were found to have a screw cut out requiring revision surgery. Two other revision surgeries were performed for malrotation (n = 1) and malunion (n = 1). Other implant-related complications occurred in 19 cases (7.9%), which included broken distal screws (n = 9), distal screw loosening (n = 8), and loose lag screws (n = 2). There was a total of 10 (3.8%) surgical wound complications, including four deep and six superficial infections.

Discussion

This modified cephalomedullary nail is a reliable, safe, and effective implant for management of intertrochanteric hip fractures. Surgical treatment of patients with intertrochanteric hip fractures can be performed in a safe fashion using this implant.
Abkürzungen
ATV
All-terrain vehicle
BMI
Body Mass Index
CPT
Current Procedural Terminology (CPT)
Ml
Milliliter
Mm
millimeter
OTA/AO
Orthopaedic Trauma Association/Arbeitsgemeinschaft für Osteosynthesefragen

Introduction

Intertrochanteric fractures account for a vast majority of hip fractures in the elderly population [16]. Several surgical options, such as the sliding hip compression screw, intramedullary nail, and arthroplasty, exist for the management of these fractures [7]. Due to relative ease of use and favorable clinical outcomes, cephalomedullary nailing has become one of the most common means of hip fixation in the United States [1, 8]. However, complications do arise after implantation of these nails, including mechanical failure, screw cut out, varus collapse, shortening of the femoral neck, and peri-implant femoral shaft fractures around the distal tip of the implant [9]. Recently, a new two-screw cephalomedullary nail with integrated interlocking lag and compression screws was designed to minimize these complications and to improve patient safety in surgery.
In contrast to many available chephalomedullary nailing systems, this modified cephalomedullary nail provides a fixation construct with two integrated interlocking lag and compression screws and a trapezoidal nail profile designed to optimize stability [10]. Thus, the insertion of a lag screw combined with an interdigitated compression screws may potentially minimize the risk of screw cut out from the head segment by providing immediate intraoperative linear compression, improved rotational stability, and increased bony purchase within the femoral head. The interdigitating screw insertion further allows for minimizing the risk of the reported Z-effect, which has been described as lateral and medial migration of the superior and inferior screws respectively [11].
The benefit for patient safety of this two-screw cephalomedullary nailing system requires further clinical investigations. The principal goal of this study was to examine the mechanical failure rates and to determine the safety and efficacy of this cephalomedullary nailing system. We hypothesize that the mechanical failure rates associated with this nailing system compare favorably with the results of other cephalomedullary nailing systems reported in the literature.

Materials and methods

This was a retrospective study that was performed at two urban certified level-1 trauma centers and one urban certified level-3 trauma center. The study data was collected through a retrospective chart review and review of the existing radiographic studies. Patients were identified through the coding database of our institution. Approval of the study protocol was obtained from each Institutional Review Board (IRB) at the respective institutions.
Patients 18 years of age and older who underwent nail fixation of their acute intertrochanteric femoral fracture using the InterTAN cephalomedullary nail (Smith and Nephew, Memphis, TN) between 2012 and 2016 were included in this investigation. Fractures were classified using the OTA/AO fracture classification [12]. Intertrochanteric fractures treated with sliding hip compression screws, arthroplasty, or other nailing systems were excluded from this study. Patients with pathologic fractures from neoplastic disease or femoral head/neck fractures were also excluded from this study.
All patients included in this study underwent cephalomedullary nailing using the InterTAN. The surgical technique was according to widely established recommendations as described in the literature and according to manufacturer guidelines [10, 1317]. In brief, the patient is placed on a standard fracture table to allow for application of traction and appropriate fracture reduction (Fig. 1). An approximately 3–4 cm surgical incision is made approximately 3 fingerbreadths proximal to the greater trochanter. The fascia is incised and a 3.2 mm guide pin is placed on the appropriate entry point which is located on the medial face of the greater trochanter on the anteroposterior (AP) fluoroscopic view and in-line with the femoral canal on the lateral fluoroscopic view (Fig. 2). Following appropriate guide pin placement, the canal is opened with an entry reamer and a ball-tipped guide-wire is placed into the femoral canal. We recommend reaming the femoral canal for appropriate preparation of the nail insertion. Once the diameter and length of the nail has been determined, the nail is assembled with the drill guide, and advanced into the femoral canal. Following nail insertion, the appropriate depth and alignment is confirmed on both AP and lateral fluoroscopic views of the hip. The lag screw position is planned with a 3.2 mm guide pin that is inserted through the aiming jig. The lag screw position follows general guidelines with an appropriate tip-apex distance and a center/center position on both the AP and lateral fluoroscopic view (Fig. 3). Then a 7- mm drill is inserted through the aiming jig to drill for the compression screw just below the lag screw. Afterwards, an anti-rotation bar is placed into the drill hole for the compression screw in order to avoid spinning of the head segment during insertion of the lag screw (Fig. 4). The lag screw is drilled over the guide pin (Fig. 5) and the lag screw is inserted with the anti-rotation bar in place (Fig. 6). The compression screw is placed providing linear compression across the fracture site and additional stability by interdigitation with the lag screw (Fig. 7). It is recommended to release the traction during insertion of the compression screw in order to allow for appropriate linear compression. Additional stability can be achieved by insertion of a distal interlocking screw and by tightening down the proximal set screw (Fig. 8). Appropriate fracture reduction and implant position is confirmed on final AP and lateral fluoroscopic views (Fig. 9). The wounds are closed in a standard fashion. The procedure can usually be performed in a minimal invasive fashion through three relatively small incisions (Fig. 10). Additional safety features of this nailing system include a trapezoidal shape in the proximal portion providing rotational stability, press-fit in the metaphyseal region, and distribution of tensile forces. In addition, the clothespin distal tip is less rigid to decrease the stress riser, potentially reducing the incidence of periprosthetic fractures and anterior thigh pain (Fig. 11).
Patients were considered as incomplete follow-up if clinical and radiographic outcome data was not available for a minimum of 12 weeks after surgery. Technical data on the fixation construct and the mechanical failures were collected from the postoperative radiographs. Radiographic data recorded included the nail size (short vs. long), neck shaft angle, tip apex distance, and the number of distal locking screws.
The primary outcome measures were mechanical hardware failure and proximal screw cutout. Secondary outcome measures included nonunion, as defined by the need for re-operation to achieve fracture healing, malunion, varus collapse (defined as ten degrees of radiographic varus from index surgery), surgical complications, and postoperative medical complications, such as thromboembolic events, pneumonia, urinary infection, myocardial infarction, and death.

Statistical analysis

Descriptive statistics were recorded for 264 patients, who completed their minimum twelve weeks follow-up appointment. All statistical analysis was performed using Stata 14 (StataCorp, College Station, TX). Given the relatively small number of events encountered for our main outcome measure (mechanical implant failure), no comparisons between the mechanical failure group and the non-mechanical failure group was performed. All continuous variables were tested for normal distribution. Normally distributed data was reported as means with standard deviation (SD). Not normally distributed data was reported as median with range.

Results

Based on the Current Procedural Terminology (CPT) code 27245, a total of 756 patients were screened for participation in this study. One hundred seventy-three patients did not meet the inclusion criteria. The specific reasons for exclusion from the study were use of another fixation construct (n = 37), treatment with a different nailing system (n = 4), non-intertrochanteric fracture (n = 107), unrelated death (n = 16), and pathological fracture (n = 9). A total of 74 patients were screen failures including duplicates, coding errors, and non-identifiable patients. Thus, a total of 509 patients were enrolled in this study. Two hundred forty-five of the remaining 509 (48%) did not complete their minimum 12-week follow-up appointment. None of the 245 patients with incomplete 12-week follow-up data was found to have any signs of mechanical hardware failure at their latest follow-up appointment. The outcome data reported herein are based on 264 patients with complete follow-up data. The demographic and clinical data of these patients are listed in Table 1 and Table 2, respectively.
Table 1
Patient demographics
Age [years]
Mean 67.9 (Range 18–98)
Gender
 Female
144 (54.5%)
 Male
120 (45.5%)
Diabetes mellitus
 No
202 (76.5%)
 Yes
60 (22.7%)
 Unknown
2 (0.8%)
Body mass index [kg/m2]
26.1 (Range 12.2 to 54.9)
Obesity
 Non-obese (BMI < 30.0 kg/m2)
205 (77.7%)
 Obese (BMI ≥ 30.0 kg/m2)
59 (22.3%)
Injury mechanism
 Ground level fall
182 (68.9%)
 Fall from height
30 (11.4%)
 Motor vehicle collision
19 (7.2%)
 Farm injury (attack by horse, bull, sheep)
4 (1.5%)
 Bicycle accident
4 (1.5%)
 Motorcycle collision
7 (2.7%)
 Gunshot injury
2 (0.8%)
 Motor vehicle versus pedestrian collision
3 (1.1%)
 Crushed injury
2 (0.8%)
 Other (golf cart, ATV, jet ski)
11 (4.2%)
Table 2
Clinical data
OTA/AO Fracture Classification
 A 1.1
45 (17%)
 A 1.2
12 (4.5%)
 A 1.3
6 (2.3%)
 A 2.1
56 (21.2%)
 A 2.2
48 (18.2%)
 A 2.3
20 (7.6%)
 A 3.1
20 (7.6%)
 A 3.2
18 (6.8%)
 A 3.3
39 (14.8%)
Time from orthopedic consultation to operative room [hours]
Mean 25.4 (Range: 1–456)
Length of hospital stay [days]
Mean 8.2 (Range: 1–55)
Length of follow up [weeks]
Mean 37 (Range: 12–186)
Operative time from skin incision [min]
Mean 95.87 (Range: 20–429)
Estimated blood loss [mL]
Mean 196.12 (Range: 5–1200)
Immediate postoperative neck-shaft angle [degrees]
Mean 127.38 (Range: 115–144)
Tip-apex distance
Mean (Range: 5.1–29.48)
The initial surgeries were performed between 2012 and 2016. 244 (92.4%) patients were operated at the two urban certified level-1 trauma centers and 20 patients (7.6%) were managed at the urban certified level-3 trauma center. All patients included in this study presented with an OTA/AO type 31-A fracture. The nail size and screw configuration was chosen based on the fracture pattern and the surgeon’s preference. Number of long nails (n = 202) used for implantation exceeded the number of short nails (n = 62). The nails of 217 patients (82.1%) included use of distal locking screws. Additionally, immediate postoperative imaging showed an average tip-apex distance of 16.2 mm [range: 5.1 mm - 29.5 mm] with a neck shaft angle of 127.4 degrees [range: 115 degrees - 144 degrees]. The mean estimated blood loss was 196.1 mL [range 5 mL – 1200 mL].
With regards to the primary outcome measure of mechanical hardware failure and screw cutout, we encountered a total of two screw cut outs among the 264 patients (0.75%). In one case of screw cut out, the implant was a long nail with one distal interlocking screw. The patient had an immediate post-operative neck-shaft angle of 136 degrees and a tip-apex distance of 6.8 mm. Upon follow up, non-union and varus collapse of 14 degrees was recorded. The patient underwent removal of hardware and total hip arthroplasty. The other screw cut out occurred in a patient with pre-existing avascular necrosis of the femoral head and, subsequently, underwent hardware removal and total hip arthroplasty at approximately 6 weeks postoperative. Regarding the nail construct, the immediate postoperative tip-apex distance was 13.3 mm with a neck shaft angle of 131 degrees. The implant was a short nail with one distal locking screw.
Other implant-related complications occurred in 19 cases (7.9%), which included broken distal screws (n = 9), distal screw loosening (n = 8), and loose lag screws (n = 2). Two other revision surgeries were performed for malrotation (n = 1) and malunion (n = 1). In addition, removal of symptomatic hardware was required in three patients (1.1%).
As for other secondary outcomes, two patients (0.75%) presented with delayed unions, both of whom were treated conservatively with either a bone stimulator and/or vitamin D supplementation. They subsequently went to bony union at their final follow-up. There were 30 Postoperative medical complications (11.4%), including acute renal injury (n = 5), urinary tract infections (n = 13), respiratory distress (n = 5), deep vein thrombosis (n = 1), pulmonary embolism (n = 1), pneumonia (n = 4), and myocardial infarctions (n = 1).
There was a total of 10 (3.8%) surgical postoperative complications, including four deep wound infections and six superficial wound complications. Three patients with deep wound infections were treated successfully with operative irrigation and debridements in addition to intravenous antibiotics while one patient had to undergo hardware removal. The six patients with superficial wound infections were successfully treated with oral antibiotics resulting in resolution of their symptoms. Data pertaining to complications is listed in Table 3.
Table 3
Complications
Mechanical hardware failure
 Screw cutout
2 (0.75%)
 Broken distal screws
9 (3.4%)
 Distal screw loosening
8 (3.0%)
 Loose lag screws
2 (0.75%)
 Delayed union
2 (0.75%)
Postoperative medical complications
 Acute renal injury
5 (1.9%)
 Urinary tract infection
13 (4.9%)
 Respiratory distress
5 (1.9%)
 Deep vein thrombosis
1 (0.38%)
 Pulmonary embolism
1 (0.38%)
 Myocardial infarction
1 (0.38%)
Postoperative surgical complications
 Superficial wound infection
6 (2.3%)
 Deep wound infection
4 (1.5%)
Revision surgery
 Malrotation
1 (0.38%)
 Malunion
1 (0.38%)
 Symptomatic hardware removal
3 (1.1%)

Discussion

The incidence of hip fractures, such as intertrochanteric femoral fractures, is expected to double in the next 25 years due to the higher life expectancy of the population [1821]. Because most of these patients will be elderly, operative management should consist of a stable construct performed in a timely manner to decrease both surgical and medical complications. Currently, several fixation techniques for these fractures consist of intramedullary nails, dynamic hip screws, or arthroplasty [7, 22]. The results of our retrospective study confirmed our hypothesis that this innovative two-screw cephalomedullary nail is a safe and reliable nailing system for the treatment of patients with intertrochanteric femoral fractures. In our series of 264 patients, we observed only two screw cut outs (0.75%).
Our mechanical failures of this study must be interpreted in the context of the patient demographics. One of the patients with screw cut out was elderly with pre-existing avascular necrosis of the femoral head while the other had a nonunion in the setting of acceptable tip-apex distance of 13.3 mm with a neck shaft angle of 131 degrees [1315, 17]. Possible reasons for mechanical failure can be from surgical technique, position of lag screws, or tip apex distance. Hopp et al. [23] found mechanical failure to be correlated with the positioning of the lag screw but suggested that the surgeon’s techniques (closed reduction, positioning of lag screw), not implant configuration, is crucial in achieving successful outcome. Also, lag screw placement, specifically as it pertains to tip apex distance, has proven to be a meaningful calculation when predicting complications [1315, 17]. Our study showed two possible preventable revision surgeries for malrotation and malunion, which can be the result of subpar surgical technique. These findings may represent the inherent risks of any intramedullary nail procedure, irrespective of the safety and effectiveness of the two-screw cephalomedullary nailing system itself. Our study did however have two loosening of lag screws, one of which required hardware removal. We also encountered nine broken distal screws and eight distal screw loosenings, whereby none of which were associated with any adverse outcomes. Overall, our findings are in line with other reports of this two-screw cephalomedullary nail found in the literature. Ruecker et al. [9] had two varus malalignment due to poor reduction in their study of 100 patients. Also, Jong-Won Kim et al. [24] encountered three complications requiring reoperation: two varus collapse with cut out and one case of periprosthetic fracture.
Several biomechanical and clinical data reported in the literature has shown favorable outcomes with the two-screw cephalomedullary nail construct [4, 6, 25]. Ruecker et al. [9] looked at 100 patients with stable and unstable intertrochanteric fractures treated with this two-screw cephalomedullary nail construct and found no loss of reduction, non-unions, or implant failures. Compared with the Synthes Proximal Femoral Nail Anti-rotation (PFNA), the two-screw cephalomedullary nail had significantly less complications, such as screw cut out, femoral shaft fracture distal to the tip of the nail, fracture collapse, and revisions [4, 25]. Also, the Stryker Gamma nail had a higher incidence of cut out and femoral shaft fracture compared to this two-screw cephalomedullary nail [6]. However, other studies reported equivalent outcomes. Jong-Won Kim et al. [24] found that only 45 out of their 100 patients (60.8%) with intertrochanteric femur fractures treated with the two-screw cephalomedullary nail recovered pre-fracture functional status. Also, Hopp et al. [23] found no differences between the two-screw nail and the Gamma3 nail in terms of mechanical failures and functional outcomes determined by the Harris Hip Score at 6 months postoperatively. Erez et al. [26] found rates of complications between the two-screw cephalomedullary nail and the Gamma nail to be similar.
Strengths of this study include its multicenter design. Also, our study looked at the clinical and radiographic outcomes in a large sample size with 264 patients treated with the two-screw cephalomedullary device. Limitations of our study include the retrospective design. Moreover, the patients were managed by different surgeons and according to different surgical and postoperative treatment protocols. Also, the study did not have a comparison group treated with a different nailing system. Finally, we encountered a loss of follow-up at 12 weeks of 48%, which is inherent for this patient population, but may potentially bias the study results. However, we would like to emphasize that none of the patients lost to follow-up had demonstrated signs of mechanical failure during their last postoperative visit. In addition, we carefully suggest that in case of any significant postoperative problems, a return to our trauma centers would have been likely for most of these patients.
Future studies on this cephalomedullary device should be conducted as multicenter trials given the low rate of mechanical failure, which necessitates a large sample size to capture and assess the effects of the implant. Given our results and the current status of the literature, we speculate that future studies will confirm low rates of hardware failure and screw cut out and will support this two-screw cephalomedullary nail as a safe and effective implant for the treatment of intertrochanteric fractures.
In conclusion, our multicenter study found that this two-screw cephalomedullary nail system had a low cut out and mechanical failure rate. Cephalomedullary nailing of intertrochanteric hip fractures using this improved system can be considered a safe and effective treatment method. Our findings confirmed that the two-screw cephalomedullary nail provides appropriate patient safety.

Acknowledgements

The publication costs for this article (APC) were covered in full by a grant from the Colorado Physician Insurance Company (www.​copic.​com). COPIC had no influence on authorship or scientific content of this article.

Funding

The publication costs for this article were covered in full by a grant from the Colorado Physician Insurance Company (www.​copic.​com) to Philip F. Stahel, MD. COPIC had no influence on authorship or scientific content of this article. There were no additional sources of funding obtained for this study.

Availability of data and materials

The datasets generated and/or analyzed during the current study are not publicly available due the presence of personal health information but available from the corresponding author on reasonable request.
This study was approved by the Institutional Review Boards of University of Texas Health at San Antonio, Methodist Hospital San Antonio, and University of South Florida.
There is no information with individual personal data so no consent for publication was needed.

Competing interests

BAZ has received speaker fees from AO North America, research grants from DepuySynthes and the Orthopaedic Trauma Association, and educational grant support from Smith and Nephew Inc. He serves on the editorial board of International Orthopaedics, Journal of Orthopaedic Trauma, and Patients Safety in Surgery. None of these activities are related to the content of this article. The remaining authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Forte ML, Virnig BA, Eberly LE, et al. Provider factors associated with intramedullary nail use for intertrochanteric hip fractures. J Bone Joint Surg Am. 2010;92:1105–14.CrossRef Forte ML, Virnig BA, Eberly LE, et al. Provider factors associated with intramedullary nail use for intertrochanteric hip fractures. J Bone Joint Surg Am. 2010;92:1105–14.CrossRef
2.
Zurück zum Zitat Kokoroghiannis C, Aktselis I, Deligeorgis A, et al. Evolving concepts of stability and intramedullary fixation of intertrochanteric fractures– a review. Injury. 2012;43:686–93.CrossRef Kokoroghiannis C, Aktselis I, Deligeorgis A, et al. Evolving concepts of stability and intramedullary fixation of intertrochanteric fractures– a review. Injury. 2012;43:686–93.CrossRef
3.
Zurück zum Zitat Ma KL, Wang X, Luan FJ, et al. Proximal femoral nails antirotation, gamma nails, and dynamic hip screws for fixation of intertrochanteric fractures of femur: a meta-analysis. Orthop Traumatol Surg Res. 2014;100(8):859–66.CrossRef Ma KL, Wang X, Luan FJ, et al. Proximal femoral nails antirotation, gamma nails, and dynamic hip screws for fixation of intertrochanteric fractures of femur: a meta-analysis. Orthop Traumatol Surg Res. 2014;100(8):859–66.CrossRef
4.
Zurück zum Zitat Nherera L, Trueman P, Horner A, Watson T, Johnstone AJ. Comparison of a twin interlocking derotation and compression screw cephalomedullary nail (InterTAN) with a single screw derotation cephalomedullary nail (proximal femoral nail antirotation): a systematic review and meta-analysis for intertrochanteric fractures. J Orthop Surg Res. 2018;13:46. https://doi.org/10.1186/s13018-018-0749-6.CrossRefPubMedPubMedCentral Nherera L, Trueman P, Horner A, Watson T, Johnstone AJ. Comparison of a twin interlocking derotation and compression screw cephalomedullary nail (InterTAN) with a single screw derotation cephalomedullary nail (proximal femoral nail antirotation): a systematic review and meta-analysis for intertrochanteric fractures. J Orthop Surg Res. 2018;13:46. https://​doi.​org/​10.​1186/​s13018-018-0749-6.CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat Swart E, Makhni EC, Macaulay W, et al. Cost-effectiveness analysis of fixation options for intertrochanteric hip fractures. J Bone Joint Surg Am. 2014;96:1612–20.CrossRef Swart E, Makhni EC, Macaulay W, et al. Cost-effectiveness analysis of fixation options for intertrochanteric hip fractures. J Bone Joint Surg Am. 2014;96:1612–20.CrossRef
6.
Zurück zum Zitat Wu D, Ren G, Peng C, Zheng X, Mao F, Zhang Y. InterTan nail versus Gamma3 nail for intramedullary nailing of unstable trochanteric fractures. Diagn Pathol. 2014;9:191.CrossRef Wu D, Ren G, Peng C, Zheng X, Mao F, Zhang Y. InterTan nail versus Gamma3 nail for intramedullary nailing of unstable trochanteric fractures. Diagn Pathol. 2014;9:191.CrossRef
7.
Zurück zum Zitat Lorich DG, Geller DS, Nielson JH. Osteoporotic pertrochanteric hip fractures: management and current controversies. Instr Course Lect. 2004;53:441–54.PubMed Lorich DG, Geller DS, Nielson JH. Osteoporotic pertrochanteric hip fractures: management and current controversies. Instr Course Lect. 2004;53:441–54.PubMed
8.
Zurück zum Zitat Galli M, Ciriello V, Bocchino L, Gangemi NM, Peruzzi M, Marzetti E. Clinical and functional outcomes of internal fixation with intertrochanteric antegrade nail in older patients with proximal extracapsular femoral fractures. Eur J Trauma Emerg Surg. 2014 Aug;40(4):495–500.CrossRef Galli M, Ciriello V, Bocchino L, Gangemi NM, Peruzzi M, Marzetti E. Clinical and functional outcomes of internal fixation with intertrochanteric antegrade nail in older patients with proximal extracapsular femoral fractures. Eur J Trauma Emerg Surg. 2014 Aug;40(4):495–500.CrossRef
9.
Zurück zum Zitat Ruecker AH, Rupprecht M, Gruber M, Gebauer M, Barvencik F, Briem D, Rueger JM. The treatment of intertrochanteric fractures: results using an intramedullary nail with integrated cephalocervical screws and linear compression. J Orthop Trauma. 2009 Jan;23(1):22–30.CrossRef Ruecker AH, Rupprecht M, Gruber M, Gebauer M, Barvencik F, Briem D, Rueger JM. The treatment of intertrochanteric fractures: results using an intramedullary nail with integrated cephalocervical screws and linear compression. J Orthop Trauma. 2009 Jan;23(1):22–30.CrossRef
10.
Zurück zum Zitat Smith & Nephew TRIGEN INTERTAN Intertrochanteric Antegrade Nail [Surgical Technique]. (2012). Smith & Nephew, Inc., Cordova, TN. Smith & Nephew TRIGEN INTERTAN Intertrochanteric Antegrade Nail [Surgical Technique]. (2012). Smith & Nephew, Inc., Cordova, TN.
11.
Zurück zum Zitat Strauss EJ, Kummer FJ, Koval KJ, Egol KA. The “Z-effect” phenomenon defined: a laboratory study. J Orthop Res. 2007;25(12):1568–73.CrossRef Strauss EJ, Kummer FJ, Koval KJ, Egol KA. The “Z-effect” phenomenon defined: a laboratory study. J Orthop Res. 2007;25(12):1568–73.CrossRef
12.
Zurück zum Zitat Marsh JL, Slongo TF, Agel J, Broderick JS, Creevey W, DeCoster TA, Prokuski L, Sirkin MS, Ziran B, Henley B, Audigé L. Fracture and dislocation classification compendium - 2007: Orthopaedic trauma association classification, database and outcomes committee. J Orthop Trauma. 2007;21(10 Suppl):S1–133.CrossRef Marsh JL, Slongo TF, Agel J, Broderick JS, Creevey W, DeCoster TA, Prokuski L, Sirkin MS, Ziran B, Henley B, Audigé L. Fracture and dislocation classification compendium - 2007: Orthopaedic trauma association classification, database and outcomes committee. J Orthop Trauma. 2007;21(10 Suppl):S1–133.CrossRef
14.
Zurück zum Zitat Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM. The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am. 1995;77-A:1058–64.CrossRef Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM. The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am. 1995;77-A:1058–64.CrossRef
15.
Zurück zum Zitat Caruso G, Bonomo M, Valpiani G, et al. A six-year retrospective analysis of cut-out risk predictors in cephalomedullary nailing for pertrochanteric fractures: can the tip-apex distance (TAD) still be considered the best parameter? Bone Joint Res. 2017;6:481–8.CrossRef Caruso G, Bonomo M, Valpiani G, et al. A six-year retrospective analysis of cut-out risk predictors in cephalomedullary nailing for pertrochanteric fractures: can the tip-apex distance (TAD) still be considered the best parameter? Bone Joint Res. 2017;6:481–8.CrossRef
16.
Zurück zum Zitat Cleveland M, Bosworth DM, Thompson FR, Wilson HJ Jr, Ishizuka T. A ten-year analysis of intertrochanteric fractures of the femur. J Bone Joint Surg Am. 1959;41-A:1399–408.CrossRef Cleveland M, Bosworth DM, Thompson FR, Wilson HJ Jr, Ishizuka T. A ten-year analysis of intertrochanteric fractures of the femur. J Bone Joint Surg Am. 1959;41-A:1399–408.CrossRef
17.
Zurück zum Zitat Kane P, Vopat B, Heard W, et al. Is tip apex distance as important as we think? A biomechanical study examining optimal lag-screw placement. Clin Orthop Relat Res. 2014;472:2492–8.CrossRef Kane P, Vopat B, Heard W, et al. Is tip apex distance as important as we think? A biomechanical study examining optimal lag-screw placement. Clin Orthop Relat Res. 2014;472:2492–8.CrossRef
18.
Zurück zum Zitat Burge R, Dawson-Hughes B, Solomon DH, Wong JB, King A, Tosteson A. Incidence and economic burden of osteoporosis-related fractures in the United States, 2005-2025. J Bone Miner Res. 2007;22:465–75.CrossRef Burge R, Dawson-Hughes B, Solomon DH, Wong JB, King A, Tosteson A. Incidence and economic burden of osteoporosis-related fractures in the United States, 2005-2025. J Bone Miner Res. 2007;22:465–75.CrossRef
19.
Zurück zum Zitat Cheng SY, Levy AR, Lefaivre KA, Guy P, Kuramoto L, Sobolev B. Geographic trends in incidence of hip fractures: a comprehensive literature review. Osteoporos Int. 2011 Oct;22(10):2575–86.CrossRef Cheng SY, Levy AR, Lefaivre KA, Guy P, Kuramoto L, Sobolev B. Geographic trends in incidence of hip fractures: a comprehensive literature review. Osteoporos Int. 2011 Oct;22(10):2575–86.CrossRef
20.
Zurück zum Zitat Dhanwal DK, Dennison EM, Harvey NC, Cooper C. Epidemiology of hip fracture: worldwide geographic variation. Indian J Orthop. 2011;45:15.CrossRef Dhanwal DK, Dennison EM, Harvey NC, Cooper C. Epidemiology of hip fracture: worldwide geographic variation. Indian J Orthop. 2011;45:15.CrossRef
21.
22.
Zurück zum Zitat Chirodian N, Arch B, Parker MJ. Sliding hip screw fixation of trochanteric hip fractures: outcome of 1024 procedures. Injury. 2005;36(6):793–800.CrossRef Chirodian N, Arch B, Parker MJ. Sliding hip screw fixation of trochanteric hip fractures: outcome of 1024 procedures. Injury. 2005;36(6):793–800.CrossRef
23.
Zurück zum Zitat Hopp S, et al. Does the implant make the difference? - prospective comparison of two different proximal femur nails. Acta Orthop Belg. 2016;82:319–31.PubMed Hopp S, et al. Does the implant make the difference? - prospective comparison of two different proximal femur nails. Acta Orthop Belg. 2016;82:319–31.PubMed
26.
Zurück zum Zitat Erez O, Dougherty PJ. Early complications associated with cephalomedullary nail for intertrochanteric hip fractures. J Trauma Acute Care Surg. 2012;72:E101–5.CrossRef Erez O, Dougherty PJ. Early complications associated with cephalomedullary nail for intertrochanteric hip fractures. J Trauma Acute Care Surg. 2012;72:E101–5.CrossRef
Metadaten
Titel
Safety and efficacy of a two-screw cephalomedullary nail for intertrochanteric femur fracture fixation: a retrospective case series in 264 patients
verfasst von
Boris A. Zelle
Antonio J. Webb
Christopher Matson
Michael Morwood
Khang H. Dang
Samuel S. Ornell
Gabrielle Gostigian
Cody M. Ramirez
Hassan Mir
Publikationsdatum
01.12.2018
Verlag
BioMed Central
Erschienen in
Patient Safety in Surgery / Ausgabe 1/2018
Elektronische ISSN: 1754-9493
DOI
https://doi.org/10.1186/s13037-018-0177-x

Weitere Artikel der Ausgabe 1/2018

Patient Safety in Surgery 1/2018 Zur Ausgabe

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

CME: 2 Punkte

Prof. Dr. med. Gregor Antoniadis Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

CME: 2 Punkte

Dr. med. Benjamin Meyknecht, PD Dr. med. Oliver Pieske Das Webinar S2e-Leitlinie „Distale Radiusfraktur“ beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

CME: 2 Punkte

Dr. med. Mihailo Andric
Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.