Scanning laser polarimetry with variable corneal compensation and detection of glaucomatous optic neuropathy

verfasst von: Stefano Da Pozzo, Pierluigi Iacono, Roberta Marchesan, Anna Fantin, Giuseppe Ravalico

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 8/2005

Einloggen, um Zugang zu erhalten

Abstract

Background

The aim of this study was to evaluate the ability of scanning laser polarimetry (SLP) parameters provided by commercially available GDx with variable corneal compensator (VCC) to discriminate between healthy and glaucomatous eyes.

Methods

Sixty-five healthy and 59 glaucomatous age-matched patients underwent a complete ophthalmological evaluation, an achromatic automated perimetry (AAP), and SLP with GDx-VCC. One randomly selected eye from each subject was considered. All glaucomatous eyes had reproducible visual field defects. Mean values (± SD) of all SLP-VCC parameters measured in the two groups were compared. Area under receiver operating characteristics (AUROC) curve and sensitivities at predetermined specificities of ≥80% and ≥95% for each single parameter were calculated. Moreover, the nerve fiber indicator (NFI) diagnostic accuracy was evaluated calculating positive, negative, and interval likelihood ratios (LRs) at different cutoff values.

Results

All SLP parameters were significantly different between the two groups (p<0.001). The NFI showed the best AUROC curve (0.938, SE 0.02) whereas temporal, superior, nasal, inferior, temporal (TSNIT) average was second best (0.897, SE 0.03), and normalized superior area was third (0.879, SE 0.04). At fixed specificity ≥95%, sensitivities ranged from 22% to 79.7% whereas for values ≥80%, sensitivities were in the 44.1–89.8% range. At a cutoff NFI value of 30, positive LR was 17.6 (95% CI: 5.8–53.6) and negative LR was 0.19 (95% CI: 0.11–0.33). Interval LRs for NFI showed that values ≤20 or >40 were associated with large effects on posttest probability.

Conclusions

SLP-VCC allows good discrimination between healthy and glaucomatous eyes. New software-provided parameters NFI, TSNIT average, and normalized superior and inferior areas appear to be reliable in the evaluation of glaucomatous disease. In particular, after evaluation on interval LRs, the NFI showed a high diagnostic accuracy for values ≤20 or >40.
Literatur
1.
Zurück zum Zitat Bagga H, Greenfield DS, Feuer WJ, Knighton RW (2003) Scanning laser polarimetry with variable corneal compensation and optical coherence tomography in normal and glaucomatous eyes. Am J Ophthalmol 135:521–529 Bagga H, Greenfield DS, Feuer WJ, Knighton RW (2003) Scanning laser polarimetry with variable corneal compensation and optical coherence tomography in normal and glaucomatous eyes. Am J Ophthalmol 135:521–529
2.
Zurück zum Zitat Bowd C, Zangwill LM, Berry CC, Blumenthal EZ, Vasile C, Sanchez-Galeana C et al (2001) Detecting early glaucoma by assessment of retinal nerve fiber layer thickness and visual function. Invest Ophthalmol Vis Sci 42:1993–2003 Bowd C, Zangwill LM, Berry CC, Blumenthal EZ, Vasile C, Sanchez-Galeana C et al (2001) Detecting early glaucoma by assessment of retinal nerve fiber layer thickness and visual function. Invest Ophthalmol Vis Sci 42:1993–2003
3.
Zurück zum Zitat Bowd C, Zangwill LM, Weinreb RN (2003) Association between scanning laser polarimetry measurements using variable corneal polarization compensation and visual field sensitivity in glaucomatous eyes. Arch Ophthalmol 121:961–966 Bowd C, Zangwill LM, Weinreb RN (2003) Association between scanning laser polarimetry measurements using variable corneal polarization compensation and visual field sensitivity in glaucomatous eyes. Arch Ophthalmol 121:961–966
4.
Zurück zum Zitat Chen YY, Chen PP, Xu L, Ernst PK, Wang L, Mills RP (1998) Correlation of peripapillary nerve fiber layer thickness by scanning laser polarimetry with visual field defects in patients with glaucoma. J Glaucoma 7:312–316 Chen YY, Chen PP, Xu L, Ernst PK, Wang L, Mills RP (1998) Correlation of peripapillary nerve fiber layer thickness by scanning laser polarimetry with visual field defects in patients with glaucoma. J Glaucoma 7:312–316
5.
Zurück zum Zitat Choplin NT, Zhou Q, Knighton RW (2003) Effect of individualized compensation for anterior segment birefringence on retinal nerve fiber layer assessment as determined by scanning laser polarimetry. Ophthalmology 110:719–725 Choplin NT, Zhou Q, Knighton RW (2003) Effect of individualized compensation for anterior segment birefringence on retinal nerve fiber layer assessment as determined by scanning laser polarimetry. Ophthalmology 110:719–725
6.
Zurück zum Zitat Gallagher EJ (2003) Evidence-based emergency medicine/editorial: the problem with sensitivity and specificity. Ann Emerg Med 42:298–303 Gallagher EJ (2003) Evidence-based emergency medicine/editorial: the problem with sensitivity and specificity. Ann Emerg Med 42:298–303
7.
Zurück zum Zitat Greenfield DS, Knighton RW, Huang XR (2000) Effect of corneal polarization axis on assessment of retinal nerve fiber layer thickness by scanning laser polarimetry. Am J Ophthalmol 129:715–722 Greenfield DS, Knighton RW, Huang XR (2000) Effect of corneal polarization axis on assessment of retinal nerve fiber layer thickness by scanning laser polarimetry. Am J Ophthalmol 129:715–722
8.
Zurück zum Zitat Greenfield DS, Knighton RW, Feuer WJ, Schiffmann JC, Zangwill LM, Weinreb RN (2002) Correction for corneal polarization axis improves the discriminating power of scanning laser polarimetry. Am J Ophthalmol 134:27–33 Greenfield DS, Knighton RW, Feuer WJ, Schiffmann JC, Zangwill LM, Weinreb RN (2002) Correction for corneal polarization axis improves the discriminating power of scanning laser polarimetry. Am J Ophthalmol 134:27–33
9.
Zurück zum Zitat Hodapp E, Parrish RK II, Anderson DR (1993) Clinical decisions in glaucoma. Mosby Year Book, St. Louis, pp 52–61 Hodapp E, Parrish RK II, Anderson DR (1993) Clinical decisions in glaucoma. Mosby Year Book, St. Louis, pp 52–61
10.
Zurück zum Zitat Jaeschke R, Guyatt GH, Sackett DL, Evidence-Based Medicine Working Group (1994) User’s guides to the medical literature, III: how to use an article about a diagnostic test, B: what are the results and will they help me in caring for my patients? JAMA 271:703–707 Jaeschke R, Guyatt GH, Sackett DL, Evidence-Based Medicine Working Group (1994) User’s guides to the medical literature, III: how to use an article about a diagnostic test, B: what are the results and will they help me in caring for my patients? JAMA 271:703–707
11.
Zurück zum Zitat Kass MA, Heuer DK, Higginbotham EJ, Johnson CA, Keltner JL, Miller JP et al (2002) The ocular hypertension treatment study: a randomized trial determines that topical ocular hypotensive medication delays or prevents the onset of primary open-angle glaucoma. Arch Ophthalmol 120:701–713PubMed Kass MA, Heuer DK, Higginbotham EJ, Johnson CA, Keltner JL, Miller JP et al (2002) The ocular hypertension treatment study: a randomized trial determines that topical ocular hypotensive medication delays or prevents the onset of primary open-angle glaucoma. Arch Ophthalmol 120:701–713PubMed
12.
Zurück zum Zitat Knighton RW, Huang XR (2002) Linear birefringence of the central human cornea. Invest Ophthalmol Vis Sci 43:82–86 Knighton RW, Huang XR (2002) Linear birefringence of the central human cornea. Invest Ophthalmol Vis Sci 43:82–86
13.
Zurück zum Zitat Kwon YH, Hong S, Honkanen RA, Alward WL (2000) Correlation of automated visual field parameters and peripapillary nerve fiber layer thickness as measured by scanning laser polarimetry. J Glaucoma 9:281–288 Kwon YH, Hong S, Honkanen RA, Alward WL (2000) Correlation of automated visual field parameters and peripapillary nerve fiber layer thickness as measured by scanning laser polarimetry. J Glaucoma 9:281–288
14.
Zurück zum Zitat Medeiros FA, Zangwill LM, Bowd C, Bernd AS, Weinreb RN (2003) Fourier analysis of scanning laser polarimetry measurements with variable corneal compensation in glaucoma. Invest Ophthalmol Vis Sci 44:2606–2612 Medeiros FA, Zangwill LM, Bowd C, Bernd AS, Weinreb RN (2003) Fourier analysis of scanning laser polarimetry measurements with variable corneal compensation in glaucoma. Invest Ophthalmol Vis Sci 44:2606–2612
15.
Zurück zum Zitat Medeiros FA, Zangwill LM, Bowd C, Weinreb RN (2004) Comparison of the GDx VCC scanning laser polarimeter, HRT II confocal scanning laser ophthalmoscope, and stratus OCT optical coherence tomography for the detection of glaucoma. Arch Ophthalmol 122:827–837 Medeiros FA, Zangwill LM, Bowd C, Weinreb RN (2004) Comparison of the GDx VCC scanning laser polarimeter, HRT II confocal scanning laser ophthalmoscope, and stratus OCT optical coherence tomography for the detection of glaucoma. Arch Ophthalmol 122:827–837
16.
Zurück zum Zitat Pederson JE, Anderson DR (1980) The mode of progressive optic disc cupping in ocular hypertension and glaucoma. Arch Ophthalmol 98:490–495 Pederson JE, Anderson DR (1980) The mode of progressive optic disc cupping in ocular hypertension and glaucoma. Arch Ophthalmol 98:490–495
17.
Zurück zum Zitat Quigley HA, Addicks EM, Green RW (1982) Optic nerve damage in human glaucoma. III. Quantitative correlation of nerve fiber loss and visual defect in glaucoma, ischemic neuropathy, papilledema, and toxic neuropathy. Arch Ophthalmol 100:135–146PubMed Quigley HA, Addicks EM, Green RW (1982) Optic nerve damage in human glaucoma. III. Quantitative correlation of nerve fiber loss and visual defect in glaucoma, ischemic neuropathy, papilledema, and toxic neuropathy. Arch Ophthalmol 100:135–146PubMed
18.
Zurück zum Zitat Reus NJ, Colen TP, Lemij HG (2003) Visualization of localized retinal nerve fiber layer defects with the GDx with individualized and with fixed compensation of anterior segment birefringence. Ophthalmology 110:1512–1516 Reus NJ, Colen TP, Lemij HG (2003) Visualization of localized retinal nerve fiber layer defects with the GDx with individualized and with fixed compensation of anterior segment birefringence. Ophthalmology 110:1512–1516
19.
Zurück zum Zitat Reus NJ, Lemij HG (2004) Diagnostic accuracy of the GDx VCC for glaucoma. Ophthalmology 111:1860–1865 Reus NJ, Lemij HG (2004) Diagnostic accuracy of the GDx VCC for glaucoma. Ophthalmology 111:1860–1865
20.
Zurück zum Zitat Reus NJ, Lemij HG (2004) The relationship between standard automated perimetry and GDx VCC measurements. Invest Ophthalmol Vis Sci 45:840–845 Reus NJ, Lemij HG (2004) The relationship between standard automated perimetry and GDx VCC measurements. Invest Ophthalmol Vis Sci 45:840–845
21.
Zurück zum Zitat Schlottmann PG, De Cilla S, Greenfield DS, Caprioli J, Garway-Heath DF (2004) Relationship between visual field sensitivity and retinal nerve fiber layer thickness as measured by scanning laser polarimetry. Invest Ophthalmol Vis Sci 45:1823–1829 Schlottmann PG, De Cilla S, Greenfield DS, Caprioli J, Garway-Heath DF (2004) Relationship between visual field sensitivity and retinal nerve fiber layer thickness as measured by scanning laser polarimetry. Invest Ophthalmol Vis Sci 45:1823–1829
22.
Zurück zum Zitat Shields JR, Chen PP, Mills RP (2002) Topographic mapping of glaucomatous visual field defects to scanning laser polarimetry of the peripapillary nerve fiber layer. Ophthalmic Surg Lasers 33:123–126 Shields JR, Chen PP, Mills RP (2002) Topographic mapping of glaucomatous visual field defects to scanning laser polarimetry of the peripapillary nerve fiber layer. Ophthalmic Surg Lasers 33:123–126
23.
Zurück zum Zitat Simel DL, Samsa GP, Matchar DB (1991) Likelihood ratios with confidence: sample size estimation for diagnostic test studies. J Clin Epidemiol 44:763–770CrossRefPubMed Simel DL, Samsa GP, Matchar DB (1991) Likelihood ratios with confidence: sample size estimation for diagnostic test studies. J Clin Epidemiol 44:763–770CrossRefPubMed
24.
Zurück zum Zitat Sommer A, Katz J, Quigley HA, Miller NR, Robin AL, Richter RC et al (1991) Clinically detectable nerve fiber layer atrophy precedes the onset of glaucomatous field loss. Arch Ophthalmol 109:77–83 Sommer A, Katz J, Quigley HA, Miller NR, Robin AL, Richter RC et al (1991) Clinically detectable nerve fiber layer atrophy precedes the onset of glaucomatous field loss. Arch Ophthalmol 109:77–83
25.
Zurück zum Zitat Tannenbaum D, Hoffmann D, Lemij HG, Garway-Heath DF, Greenfield DS, Caprioli J (2004) Variable corneal compensation improves the discrimination between normal and glaucomatous eyes with the scanning laser polarimeter. Ophthalmology 111:259–264 Tannenbaum D, Hoffmann D, Lemij HG, Garway-Heath DF, Greenfield DS, Caprioli J (2004) Variable corneal compensation improves the discrimination between normal and glaucomatous eyes with the scanning laser polarimeter. Ophthalmology 111:259–264
26.
Zurück zum Zitat Tjon-Fo-Sang MJ, Lemij HG (1997) The sensitivity and specificity of nerve fiber layer measurements in glaucoma as determined with scanning laser polarimetry. Am J Ophthalmol 123:62–69 Tjon-Fo-Sang MJ, Lemij HG (1997) The sensitivity and specificity of nerve fiber layer measurements in glaucoma as determined with scanning laser polarimetry. Am J Ophthalmol 123:62–69
27.
Zurück zum Zitat Weinreb RN, Dreher AW, Coleman A, Quigley H, Shaw B, Reiter K (1990) Histopathologic validation of Fourier-ellipsometry measurements of retinal nerve fiber layer thickness. Arch Ophthalmol 108:557–560 Weinreb RN, Dreher AW, Coleman A, Quigley H, Shaw B, Reiter K (1990) Histopathologic validation of Fourier-ellipsometry measurements of retinal nerve fiber layer thickness. Arch Ophthalmol 108:557–560
28.
Zurück zum Zitat Weinreb RN, Shakiba S, Sample PA, Shahroknis S, van Horn S, Garden VS et al (1995) Association between quantitative nerve fiber layer measurement and visual field loss in glaucoma. Am J Ophthalmol 120:732–738 Weinreb RN, Shakiba S, Sample PA, Shahroknis S, van Horn S, Garden VS et al (1995) Association between quantitative nerve fiber layer measurement and visual field loss in glaucoma. Am J Ophthalmol 120:732–738
29.
Zurück zum Zitat Weinreb RN, Bowd C, Zangwill LM (2002) Glaucoma detection using scanning laser polarimetry with variable corneal polarization compensation. Arch Ophthalmol 120:218–224 Weinreb RN, Bowd C, Zangwill LM (2002) Glaucoma detection using scanning laser polarimetry with variable corneal polarization compensation. Arch Ophthalmol 120:218–224
30.
Zurück zum Zitat Weinreb RN, Bowd C, Greenfield DS, Zangwill LM (2002) Measurement of the magnitude and axis of corneal polarization with scanning laser polarimetry. Arch Ophthalmol 120:901–906 Weinreb RN, Bowd C, Greenfield DS, Zangwill LM (2002) Measurement of the magnitude and axis of corneal polarization with scanning laser polarimetry. Arch Ophthalmol 120:901–906
31.
Zurück zum Zitat Zangwill LM, Bowd C, Berry CC, Williams J, Blumenthal EZ, Sanchez-Galeana C et al (2001) Discriminating between normal and glaucomatous eyes using the Heidelberg retina tomograph, GDx nerve fiber analyzer, and optical coherence tomograph. Arch Ophthalmol 119:985–993 Zangwill LM, Bowd C, Berry CC, Williams J, Blumenthal EZ, Sanchez-Galeana C et al (2001) Discriminating between normal and glaucomatous eyes using the Heidelberg retina tomograph, GDx nerve fiber analyzer, and optical coherence tomograph. Arch Ophthalmol 119:985–993
32.
Zurück zum Zitat Zhou Q, Weinreb RN (2002) Individualized compensation of anterior segment birefringence during scanning laser polarimetry. Invest Ophthalmol Vis Sci 43:2221–2228 Zhou Q, Weinreb RN (2002) Individualized compensation of anterior segment birefringence during scanning laser polarimetry. Invest Ophthalmol Vis Sci 43:2221–2228
Metadaten
Titel
Scanning laser polarimetry with variable corneal compensation and detection of glaucomatous optic neuropathy
verfasst von
Stefano Da Pozzo
Pierluigi Iacono
Roberta Marchesan
Anna Fantin
Giuseppe Ravalico
Publikationsdatum
01.08.2005
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 8/2005
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-004-1118-1

Weitere Artikel der Ausgabe 8/2005

Graefe's Archive for Clinical and Experimental Ophthalmology 8/2005 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.