Segmental and global left ventricular function assessment using gated SPECT with a semiconductor Cadmium Zinc Telluride (CZT) camera: Phantom study and clinical validation vs cardiac magnetic resonance

Zeitschrift:
Journal of Nuclear Cardiology > Ausgabe 4/2014
Autoren:
MD Alban Bailliez, MD Tanguy Blaire, MD, PhD Frédéric Mouquet, MD R. Legghe, MD B. Etienne, MD Damien Legallois, MD, PhD Denis Agostini, MD, PhD Alain Manrique

Abstract

Background

We evaluated gated-SPECT using a Cadmium-Zinc-Telluride (CZT) camera for assessing global and regional left ventricular (LV) function.

Methods

A phantom study evaluated the accuracy of wall thickening assessment using systolic count increase on both Anger and CZT (Discovery 530NMc) cameras. The refillable phantom simulated variable myocardial wall thicknesses. The apparent count increase (%CI) was compared to the thickness increase (%Th). CZT gated-SPECT was compared to cardiac magnetic resonance (CMR) in 27 patients. Global and regional LV function (wall thickening and motion) were quantified and compared between SPECT and CMR data.

Results

In the phantom study using a 5-mm object, the regression between %CI and %Th was significantly closer to the line of identity (y = x) with the CZT (R 2 = 0.9955) than the Anger (R 2 = 0.9995, P = .03). There was a weaker correlation for larger objects (P = .003). In patients, there was a high concordance between CZT and CMR for ESV, EDV, and LVEF (all CCC >0.80, P < .001). CZT underestimated %CI and wall motion (WM) compared to CMR (P < .001). The agreement to CMR was better for WM than wall thickening.

Conclusion

The Discovery 530NMc provided accurate measurements of global LV function but underestimated regional wall thickening, especially in patients with increased wall thickness.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt e.Med zum Sonderpreis bestellen!

Sichern Sie sich jetzt Ihr e.Med-Abo und sparen Sie 50 %!

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 4/2014

Journal of Nuclear Cardiology 4/2014 Zur Ausgabe
  1. Sie können e.Med Innere Medizin 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.

  2. Sie können e.Med Radiologie 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.