Sensitivity and uniformity improvement of phased array MR images using inductive coupling and RF detuning circuits

verfasst von: Bu S. Park, Sunder S. Rajan, Brent McCright

Erschienen in: Magnetic Resonance Materials in Physics, Biology and Medicine | Ausgabe 5/2020

Einloggen, um Zugang zu erhalten

Abstract

Objective

To improve sensitivity and uniformity of MR images obtained using a phased array RF coil, an inductively coupled secondary resonator with RF detuning circuits at 300 MHz was designed.

Materials and methods

A secondary resonator having detuning circuits to turn off the resonator during the transmit mode was constructed. The secondary resonator was located at the opposite side of the four-channel phased array to improve sensitivity and uniformity of the acquired MR images. Numerical simulations along with phantom and in vivo experiments were conducted to evaluate the designed secondary resonator.

Results

The numerical simulation results of |B1+| in a transmit mode showed that magnetic field uniformity would be decreased with a secondary resonator having no detuning circuits because of unwanted interferences between the transmit birdcage coil and the secondary resonator. The standard deviation (SD) of |B1+| was decreased 57% with a secondary resonator containing detuning circuits. The sensitivity and uniformity of |B1| in the receive mode using a four-channel phased array were improved with the secondary resonator. Phantom experiments using a uniform saline phantom had 20% improvement of the mean signal intensity and 50% decrease in the SD with the secondary resonator. Mice with excess adipose tissue were imaged to demonstrate the utility of the secondary resonator.

Conclusion

The designed secondary resonator having detuning circuits improved sensitivity and uniformity of mouse MR images acquired using the four-channel phased array.
Literatur
1.
Zurück zum Zitat Vaughan JT, Garwood M, Collins CM, Liu W, DelaBarre L, Adriany G, Andersen P, Merkle H, Goebel R, Smith MB, Ugurbil K (2001) 7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head images. Magn Reson Med 46(1):24–30CrossRef Vaughan JT, Garwood M, Collins CM, Liu W, DelaBarre L, Adriany G, Andersen P, Merkle H, Goebel R, Smith MB, Ugurbil K (2001) 7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head images. Magn Reson Med 46(1):24–30CrossRef
2.
Zurück zum Zitat Park BS, Rajan SS, Guag JW, Angelone LM (2015) A novel method to decrease electric field and SAR using an external high dielectric sleeve at 3 T head MRI: numerical and experimental results. IEEE Trans Biomed Eng 62(4):1063–1069CrossRef Park BS, Rajan SS, Guag JW, Angelone LM (2015) A novel method to decrease electric field and SAR using an external high dielectric sleeve at 3 T head MRI: numerical and experimental results. IEEE Trans Biomed Eng 62(4):1063–1069CrossRef
3.
Zurück zum Zitat Webb AG, Collins CM (2010) Parallel transmit and receive technology in high-field magnetic resonance neuroimaging. Int J Imaging Syst Technol 20(1):2–13CrossRef Webb AG, Collins CM (2010) Parallel transmit and receive technology in high-field magnetic resonance neuroimaging. Int J Imaging Syst Technol 20(1):2–13CrossRef
4.
Zurück zum Zitat Collins CM, Liu W, Schreiber W, Yang QX, Smith MB (2005) Central brightening due to constructive interference with, without, and despite dielectric resonance. J Magn Reson Imaging 21(2):192–196CrossRef Collins CM, Liu W, Schreiber W, Yang QX, Smith MB (2005) Central brightening due to constructive interference with, without, and despite dielectric resonance. J Magn Reson Imaging 21(2):192–196CrossRef
5.
Zurück zum Zitat Liu W, Collins C, Smith M (2005) Calculations of B1 distribution, specific energy absorption rate, and intrinsic signal-to-noise ratio for a body-size birdcage coil loaded with different human subjects at 64 and 128 MHz. Appl Magn Reson 29(1):5–18CrossRef Liu W, Collins C, Smith M (2005) Calculations of B1 distribution, specific energy absorption rate, and intrinsic signal-to-noise ratio for a body-size birdcage coil loaded with different human subjects at 64 and 128 MHz. Appl Magn Reson 29(1):5–18CrossRef
6.
Zurück zum Zitat Park BS, Sung K, McGarrity J, Oh S-h, Cao Z, Wang Z, Collins CM (2013) Slice-selective transmit array pulses for improvement in excitation uniformity and reduction of SAR. J Electromagn Anal Appl 5:205–212 Park BS, Sung K, McGarrity J, Oh S-h, Cao Z, Wang Z, Collins CM (2013) Slice-selective transmit array pulses for improvement in excitation uniformity and reduction of SAR. J Electromagn Anal Appl 5:205–212
7.
Zurück zum Zitat Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM (1990) The NMR phased array. Magn Reson Med 16(2):192–225CrossRef Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM (1990) The NMR phased array. Magn Reson Med 16(2):192–225CrossRef
8.
Zurück zum Zitat Quick HH, Kuehl H, Kaiser G, Bosk S, Debatin JF, Ladd ME (2002) Inductively coupled stent antennas in MRI. Magn Reson Med 48(5):781–790CrossRef Quick HH, Kuehl H, Kaiser G, Bosk S, Debatin JF, Ladd ME (2002) Inductively coupled stent antennas in MRI. Magn Reson Med 48(5):781–790CrossRef
9.
Zurück zum Zitat Hoult DI, Tomanek B (2002) Use of mutually inductive coupling in probe design. Concepts Magn Reson Part B Magn Reson Eng 15(4):262–285CrossRef Hoult DI, Tomanek B (2002) Use of mutually inductive coupling in probe design. Concepts Magn Reson Part B Magn Reson Eng 15(4):262–285CrossRef
10.
Zurück zum Zitat Bilgen M (2006) Inductively-overcoupled coil design for high resolution magnetic resonance imaging. Biomed Eng Online 5(3):1–10 Bilgen M (2006) Inductively-overcoupled coil design for high resolution magnetic resonance imaging. Biomed Eng Online 5(3):1–10
11.
Zurück zum Zitat Merkle H, Murphy-Boesch J, Pv G, Wang S, Li T-Q, Koretsky AP, Duyn JH (2011) Transmit B1-field correction at 7 T using actively tuned coupled inner elements. Magn Reson Med 66(3):901–910CrossRef Merkle H, Murphy-Boesch J, Pv G, Wang S, Li T-Q, Koretsky AP, Duyn JH (2011) Transmit B1-field correction at 7 T using actively tuned coupled inner elements. Magn Reson Med 66(3):901–910CrossRef
12.
Zurück zum Zitat Bulumulla SB, Fiveland E, Park KJ, Foo TK, Hardy CJ (2015) Inductively coupled wireless RF coil arrays. Magn Reson Imaging 33(3):351–357CrossRef Bulumulla SB, Fiveland E, Park KJ, Foo TK, Hardy CJ (2015) Inductively coupled wireless RF coil arrays. Magn Reson Imaging 33(3):351–357CrossRef
13.
Zurück zum Zitat Qian C, Duan Q, Dodd S, Koretsky A, Murphy-Boesch J (2016) Sensitivity enhancement of an inductively coupled local detector using a HEMT-based current amplifier. Magn Reson Med 75(6):2573–2578CrossRef Qian C, Duan Q, Dodd S, Koretsky A, Murphy-Boesch J (2016) Sensitivity enhancement of an inductively coupled local detector using a HEMT-based current amplifier. Magn Reson Med 75(6):2573–2578CrossRef
14.
Zurück zum Zitat Park BS, Ma G, Koch WT, Rajan SS, Mastromanolis M, Lam J, Sung K, McCright B (2019) Improvement of 19F MR image uniformity in a mouse model of cellular therapy using inductive coupling. Magn Reson Mater Phy 32:15–23CrossRef Park BS, Ma G, Koch WT, Rajan SS, Mastromanolis M, Lam J, Sung K, McCright B (2019) Improvement of 19F MR image uniformity in a mouse model of cellular therapy using inductive coupling. Magn Reson Mater Phy 32:15–23CrossRef
15.
Zurück zum Zitat Raad A, Darrasse L (1992) Optimization of NMR receiver bandwidth by inductive coupling. Magn Reson Imaging 10(1):55–65CrossRef Raad A, Darrasse L (1992) Optimization of NMR receiver bandwidth by inductive coupling. Magn Reson Imaging 10(1):55–65CrossRef
16.
Zurück zum Zitat Schnall MD, Barlow C, Subramanian VH, Leigh JS (1986) Wireless implanted magnetic resonance probes for in vivo NMR. J Magn Reson 68:16l–167 Schnall MD, Barlow C, Subramanian VH, Leigh JS (1986) Wireless implanted magnetic resonance probes for in vivo NMR. J Magn Reson 68:16l–167
17.
Zurück zum Zitat Kuhns PL, Lizak MJ, Lee S-H, Conradi MS (1988) Inductive coupling and tuning in NMR probes; applications. J Magn Reson 78:69–76 Kuhns PL, Lizak MJ, Lee S-H, Conradi MS (1988) Inductive coupling and tuning in NMR probes; applications. J Magn Reson 78:69–76
18.
Zurück zum Zitat Varadkar P, Despres D, Kraman M, Lozier J, Phadke A, Nagaraju K, McCright B (2014) The protein phosphatase 2A B56γ regulatory subunit is required for heart development. Dev Dyn 243(6):778–790CrossRef Varadkar P, Despres D, Kraman M, Lozier J, Phadke A, Nagaraju K, McCright B (2014) The protein phosphatase 2A B56γ regulatory subunit is required for heart development. Dev Dyn 243(6):778–790CrossRef
19.
Zurück zum Zitat Kumar A, Bottomley PA (2008) Optimized quadrature surface coil designs. Magn Reson Med 21:41–52 Kumar A, Bottomley PA (2008) Optimized quadrature surface coil designs. Magn Reson Med 21:41–52
20.
Zurück zum Zitat Brinka WM, Börnerta P, Nehrkeb K, Webb AG (2014) Ventricular B1+ perturbation at 7 T-real effect or measurement artifact? NMR Biomed 27:617–620CrossRef Brinka WM, Börnerta P, Nehrkeb K, Webb AG (2014) Ventricular B1+ perturbation at 7 T-real effect or measurement artifact? NMR Biomed 27:617–620CrossRef
21.
Zurück zum Zitat Pruessmann KP, MarkusWeiger SMB, Boesiger P (1999) SENSE: sensitivity encoding for fast MRI. Magn Reson Med 42(5):952–962CrossRef Pruessmann KP, MarkusWeiger SMB, Boesiger P (1999) SENSE: sensitivity encoding for fast MRI. Magn Reson Med 42(5):952–962CrossRef
22.
Zurück zum Zitat Fedorov A, Beichel R, Kalpathy-Cramer J, Finet J, Fillion-Robin J-C, Pujol S, Bauer C, Jennings D, Fennessy F, Sonka M, Buatti J, Aylward S, Miller JV, Pieper S, Kikinis R (2012) 3D slicer as an image computing platform for the quantitative imaging network. Magn Reson Imaging 30(9):1323–1341CrossRef Fedorov A, Beichel R, Kalpathy-Cramer J, Finet J, Fillion-Robin J-C, Pujol S, Bauer C, Jennings D, Fennessy F, Sonka M, Buatti J, Aylward S, Miller JV, Pieper S, Kikinis R (2012) 3D slicer as an image computing platform for the quantitative imaging network. Magn Reson Imaging 30(9):1323–1341CrossRef
23.
Zurück zum Zitat Collins CM, Smith MB (2001) Signal-to-noise ratio and absorbed power as functions of main magnetic field strength, and definition of “90°” RF pulse for the head in the birdcage coil. Magn Reson Med 45(4):684–691CrossRef Collins CM, Smith MB (2001) Signal-to-noise ratio and absorbed power as functions of main magnetic field strength, and definition of “90°” RF pulse for the head in the birdcage coil. Magn Reson Med 45(4):684–691CrossRef
Metadaten
Titel
Sensitivity and uniformity improvement of phased array MR images using inductive coupling and RF detuning circuits
verfasst von
Bu S. Park
Sunder S. Rajan
Brent McCright
Publikationsdatum
24.01.2020
Verlag
Springer International Publishing
Erschienen in
Magnetic Resonance Materials in Physics, Biology and Medicine / Ausgabe 5/2020
Print ISSN: 0968-5243
Elektronische ISSN: 1352-8661
DOI
https://doi.org/10.1007/s10334-020-00827-7

Weitere Artikel der Ausgabe 5/2020

Magnetic Resonance Materials in Physics, Biology and Medicine 5/2020 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.