Shear wave elastography for liver fibrosis in chronic hepatitis B: Adapting the cut-offs to alanine aminotransferase levels improves accuracy

verfasst von: Jie Zeng, Jian Zheng, Jie-Yang Jin, Yong-Jiang Mao, Huan-Yi Guo, Ming-De Lu, Hai-Rong Zheng, Rong-Qin Zheng

Erschienen in: European Radiology | Ausgabe 2/2019

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To determine and validate alanine aminotransferase (ALT)-adapted dual cut-offs of liver stiffness measurements (LSMs) for assessing liver fibrosis with two-dimensional shear wave elastography (2D-SWE) in patients with chronic hepatitis B (CHB) infection.

Methods

Patients with CHB infection who underwent liver biopsy to assess liver fibrosis were consecutively included. 2D-SWE confirmation thresholds with a positive likelihood ratio ≥10 and 2D-SWE exclusion thresholds with a negative likelihood ratio ≤0.1 were identified to rule in or rule out significant fibrosis and cirrhosis, respectively.

Results

The first 515 patients (index cohort) and the next 421 patients (validation cohort) were included in the final analysis. The low and high cut-offs to rule out and rule in patients with significant fibrosis (≥ F2) were 5.4 kPa and 9.0 kPa, respectively, in patients with ALT levels ≤ 2 × the upper limit of normal (ULN) and 7.1 kPa and 11.2 kPa in patients with ALT levels > 2 × ULN. For cirrhosis (F4), the corresponding values were 8.1 kPa and 12.3 kPa in patients with ALT levels ≤ 2 × ULN and 11.9 kPa and 24.7 kPa in patients with ALT levels > 2 × ULN. The dual cut-off values showed an overall accuracy of more than 90% for diagnosis of the presence or absence of significant fibrosis and cirrhosis in the index and validation cohorts. There were no significant differences in the accuracy values between the cohorts (all p>0.05).

Conclusion

The ALT-adapted dual cut-offs of LSMs showed high accuracy for diagnosis of the presence or absence of significant fibrosis and cirrhosis in patients with CHB infection.

Key Points

• The ALT-adapted dual cut-off values of LSMs showed high accuracy for diagnosis of the presence or absence of significant fibrosis and cirrhosis.
• ALT levels did not influence the overall diagnostic accuracy for predicting significant fibrosis and cirrhosis.
• The ALT-adapted dual cut-offs in patients with ALT levels > 2 × ULN were markedly higher than those in patients with ALT levels2 × ULN.
Literatur
1.
Zurück zum Zitat European Association for the Study of The Liver (2012) EASL clinical practice guidelines: management of chronic hepatitis B virus infection. J Hepatol 57:167–185CrossRef European Association for the Study of The Liver (2012) EASL clinical practice guidelines: management of chronic hepatitis B virus infection. J Hepatol 57:167–185CrossRef
2.
Zurück zum Zitat Sarin SK, Kumar M, Lau GK et al (2016) Asian-Pacific clinical practice guidelines on the management of hepatitis B: a 2015 update. Hepatol Int 10:1–98CrossRef Sarin SK, Kumar M, Lau GK et al (2016) Asian-Pacific clinical practice guidelines on the management of hepatitis B: a 2015 update. Hepatol Int 10:1–98CrossRef
3.
Zurück zum Zitat Shiina T, Nightingale KR, Palmeri ML et al (2015) WFUMB guidelines and recommendations for clinical use of ultrasound elastography: Part 1: basic principles and terminology. Ultrasound Med Biol 41:1126–1147CrossRef Shiina T, Nightingale KR, Palmeri ML et al (2015) WFUMB guidelines and recommendations for clinical use of ultrasound elastography: Part 1: basic principles and terminology. Ultrasound Med Biol 41:1126–1147CrossRef
4.
Zurück zum Zitat Bercoff J, Tanter M, Fink M (2004) Supersonic shear imaging: a new technique for soft tissue elasticity mapping. IEEE Trans Ultrason Ferroelectr Freq Control 51:396–409CrossRef Bercoff J, Tanter M, Fink M (2004) Supersonic shear imaging: a new technique for soft tissue elasticity mapping. IEEE Trans Ultrason Ferroelectr Freq Control 51:396–409CrossRef
5.
Zurück zum Zitat Samir AE, Dhyani M, Vij A et al (2015) Shear-wave elastography for the estimation of liver fibrosis in chronic liver disease: determining accuracy and ideal site for measurement. Radiology 274:888–896CrossRef Samir AE, Dhyani M, Vij A et al (2015) Shear-wave elastography for the estimation of liver fibrosis in chronic liver disease: determining accuracy and ideal site for measurement. Radiology 274:888–896CrossRef
6.
Zurück zum Zitat Kennedy P, Wagner M, Castéra L et al (2018) Quantitative elastography methods in liver disease: current evidence and future directions. Radiology 286:738–763CrossRef Kennedy P, Wagner M, Castéra L et al (2018) Quantitative elastography methods in liver disease: current evidence and future directions. Radiology 286:738–763CrossRef
7.
Zurück zum Zitat Herrmann E, de Lédinghen V, Cassinotto C, et al (2018) Assessment of biopsy-proven liver fibrosis by two-dimensional shear wave elastography: an individual patient data-based meta-analysis. Hepatology 67:260-272 Herrmann E, de Lédinghen V, Cassinotto C, et al (2018) Assessment of biopsy-proven liver fibrosis by two-dimensional shear wave elastography: an individual patient data-based meta-analysis. Hepatology 67:260-272
8.
Zurück zum Zitat Leung VY, Shen J, Wong VW et al (2013) Quantitative elastography of liver fibrosis and spleen stiffness in chronic hepatitis B carriers: comparison of shear-wave elastography and transient elastography with liver biopsy correlation. Radiology 269:910–918CrossRef Leung VY, Shen J, Wong VW et al (2013) Quantitative elastography of liver fibrosis and spleen stiffness in chronic hepatitis B carriers: comparison of shear-wave elastography and transient elastography with liver biopsy correlation. Radiology 269:910–918CrossRef
9.
Zurück zum Zitat Grgurevic I, Puljiz Z, Brnic D et al (2015) Liver and spleen stiffness and their ratio assessed by real-time two dimensional-shear wave elastography in patients with liver fibrosis and cirrhosis due to chronic viral hepatitis. Eur Radiol 25:3214–3221CrossRef Grgurevic I, Puljiz Z, Brnic D et al (2015) Liver and spleen stiffness and their ratio assessed by real-time two dimensional-shear wave elastography in patients with liver fibrosis and cirrhosis due to chronic viral hepatitis. Eur Radiol 25:3214–3221CrossRef
10.
Zurück zum Zitat Barr RG, Ferraioli G, Palmeri ML et al (2015) Elastography assessment of liver fibrosis: society of radiologists in ultrasound consensus conference statement. Radiology 276:845–886CrossRef Barr RG, Ferraioli G, Palmeri ML et al (2015) Elastography assessment of liver fibrosis: society of radiologists in ultrasound consensus conference statement. Radiology 276:845–886CrossRef
11.
Zurück zum Zitat Dietrich CF, Bamber J, Berzigotti A et al (2017) EFSUMB guidelines and recommendations on the clinical use of liver ultrasound elastography, update 2017 (long version). Ultraschall Med 38:e16-e47 Dietrich CF, Bamber J, Berzigotti A et al (2017) EFSUMB guidelines and recommendations on the clinical use of liver ultrasound elastography, update 2017 (long version). Ultraschall Med 38:e16-e47
12.
Zurück zum Zitat Mueller S, Englert S, Seitz HK et al (2015) Inflammation-adapted liver stiffness values for improved fibrosis staging in patients with hepatitis C virus and alcoholic liver disease. Liver Int 35:2514–2521CrossRef Mueller S, Englert S, Seitz HK et al (2015) Inflammation-adapted liver stiffness values for improved fibrosis staging in patients with hepatitis C virus and alcoholic liver disease. Liver Int 35:2514–2521CrossRef
13.
Zurück zum Zitat Zhuang Y, Ding H, Zhang Y et al (2017) Two-dimensional shear-wave elastography performance in the noninvasive evaluation of liver fibrosis in patients with chronic hepatitis B: comparison with serum fibrosis indexes. Radiology 283:873–882CrossRef Zhuang Y, Ding H, Zhang Y et al (2017) Two-dimensional shear-wave elastography performance in the noninvasive evaluation of liver fibrosis in patients with chronic hepatitis B: comparison with serum fibrosis indexes. Radiology 283:873–882CrossRef
14.
Zurück zum Zitat Ferraioli G, Tinelli C, Dal Bello B et al (2012) Accuracy of real-time shear wave elastography for assessing liver fibrosis in chronic hepatitis C: a pilot study. Hepatology 56:2125–2133CrossRef Ferraioli G, Tinelli C, Dal Bello B et al (2012) Accuracy of real-time shear wave elastography for assessing liver fibrosis in chronic hepatitis C: a pilot study. Hepatology 56:2125–2133CrossRef
15.
Zurück zum Zitat Zheng J, Guo H, Zeng J et al (2015) Two-dimensional shear-wave elastography and conventional US: the optimal evaluation of liver fibrosis and cirrhosis. Radiology 275:290–300CrossRef Zheng J, Guo H, Zeng J et al (2015) Two-dimensional shear-wave elastography and conventional US: the optimal evaluation of liver fibrosis and cirrhosis. Radiology 275:290–300CrossRef
16.
Zurück zum Zitat Xiao G, Yang J, Yan L (2015) Comparison of diagnostic accuracy of aspartate aminotransferase to platelet ratio index and fibrosis-4 index for detecting liver fibrosis in adult patients with chronic hepatitis B virus infection: a systemic review and meta-analysis. Hepatology 61:292–302CrossRef Xiao G, Yang J, Yan L (2015) Comparison of diagnostic accuracy of aspartate aminotransferase to platelet ratio index and fibrosis-4 index for detecting liver fibrosis in adult patients with chronic hepatitis B virus infection: a systemic review and meta-analysis. Hepatology 61:292–302CrossRef
17.
Zurück zum Zitat Houot M, Ngo Y, Munteanu M et al (2016) Systematic review with meta-analysis: direct comparisons of biomarkers for the diagnosis of fibrosis in chronic hepatitis C and B. Aliment Pharmacol Ther 43:16–29CrossRef Houot M, Ngo Y, Munteanu M et al (2016) Systematic review with meta-analysis: direct comparisons of biomarkers for the diagnosis of fibrosis in chronic hepatitis C and B. Aliment Pharmacol Ther 43:16–29CrossRef
18.
Zurück zum Zitat Wai CT, Greenson JK, Fontana RJ et al (2003) A simple noninvasive index can predict both significant fibrosis and cirrhosis in patients with chronic hepatitis C. Hepatology 38:518–526CrossRef Wai CT, Greenson JK, Fontana RJ et al (2003) A simple noninvasive index can predict both significant fibrosis and cirrhosis in patients with chronic hepatitis C. Hepatology 38:518–526CrossRef
19.
Zurück zum Zitat Bedossa P, Poynard T (1996) An algorithm for grading of activity in chronic hepatitis C. Hepatology 24:289–293CrossRef Bedossa P, Poynard T (1996) An algorithm for grading of activity in chronic hepatitis C. Hepatology 24:289–293CrossRef
20.
Zurück zum Zitat Deeks JJ, Altman DG (2004) Diagnostic tests 4: likelihood ratios. BMJ 329:168–169CrossRef Deeks JJ, Altman DG (2004) Diagnostic tests 4: likelihood ratios. BMJ 329:168–169CrossRef
21.
Zurück zum Zitat Black ER, Panzer RJ, Mayewski RJ et al (1999) Characteristics of diagnostic tests and principles for their use in quantitative decision making in diagnostic strategies for common medical problems. In: Black ER, Bordley DR, Tape TG, Panzer RJ (eds) Diagnostic strategies of common medical problems. American College of Physicians, Philadelphia, pp 1–17 Black ER, Panzer RJ, Mayewski RJ et al (1999) Characteristics of diagnostic tests and principles for their use in quantitative decision making in diagnostic strategies for common medical problems. In: Black ER, Bordley DR, Tape TG, Panzer RJ (eds) Diagnostic strategies of common medical problems. American College of Physicians, Philadelphia, pp 1–17
22.
Zurück zum Zitat Lijmer JG, Mol BW, Heisterkamp S et al (1999) Empirical evidence of design related bias in studies of diagnostic tests. JAMA 282:1061–1066CrossRef Lijmer JG, Mol BW, Heisterkamp S et al (1999) Empirical evidence of design related bias in studies of diagnostic tests. JAMA 282:1061–1066CrossRef
23.
Zurück zum Zitat DeLong ER, DeLong DM, Clarke-Pearson DL (1988) Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics 44:837–845CrossRef DeLong ER, DeLong DM, Clarke-Pearson DL (1988) Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics 44:837–845CrossRef
24.
Zurück zum Zitat Ferraioli G, Filice C, Castera L et al (2015) WFUMB guidelines and recommendations for clinical use of ultrasound elastography: Part 3: liver. Ultrasound Med Biol 41:1161–1179CrossRef Ferraioli G, Filice C, Castera L et al (2015) WFUMB guidelines and recommendations for clinical use of ultrasound elastography: Part 3: liver. Ultrasound Med Biol 41:1161–1179CrossRef
25.
Zurück zum Zitat Zeng J, Liu GJ, Huang ZP et al (2014) Diagnostic accuracy of two-dimensional shear wave elastography for the non-invasive staging of hepatic fibrosis in chronic hepatitis B: a cohort study with internal validation. Eur Radiol 24:2572–2581CrossRef Zeng J, Liu GJ, Huang ZP et al (2014) Diagnostic accuracy of two-dimensional shear wave elastography for the non-invasive staging of hepatic fibrosis in chronic hepatitis B: a cohort study with internal validation. Eur Radiol 24:2572–2581CrossRef
26.
Zurück zum Zitat Rockey DC, Caldwell SH, Goodman ZD et al (2009) Liver biopsy. Hepatology 49:1017–1044CrossRef Rockey DC, Caldwell SH, Goodman ZD et al (2009) Liver biopsy. Hepatology 49:1017–1044CrossRef
Metadaten
Titel
Shear wave elastography for liver fibrosis in chronic hepatitis B: Adapting the cut-offs to alanine aminotransferase levels improves accuracy
verfasst von
Jie Zeng
Jian Zheng
Jie-Yang Jin
Yong-Jiang Mao
Huan-Yi Guo
Ming-De Lu
Hai-Rong Zheng
Rong-Qin Zheng
Publikationsdatum
23.07.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 2/2019
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5621-x

Weitere Artikel der Ausgabe 2/2019

European Radiology 2/2019 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.