Single-needle electroporation and interstitial electrochemotherapy: in vivo safety and efficacy evaluation of a new system

Zeitschrift:
European Radiology
Autoren:
Federico Pedersoli, Andreas Ritter, Markus Zimmermann, Maximilian Schulze-Hagen, Martin Liebl, Ebba Dethlefsen, Saskia von Stillfried, Joachim Pfeffer, Christiane K. Kuhl, Philipp Bruners, Peter Isfort
Wichtige Hinweise
Philipp Bruners and Peter Isfort contributed equally to this work.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Abstract

Objectives

We conducted an in vivo trial to investigate the safety and efficacy of a newly developed system for the application of a combined therapy consisting of irreversible electroporation (IRE) and electrochemotherapy (IRECT) in the liver. The system is conceived as a single-needle multitined applicator with expandable electrodes that allow interstitial injection of fluids, e.g., chemotherapy.

Methods

Experiments were conducted in ten domestic pigs. The applicator was placed in different liver lobes under CT guidance. In one lobe, the applicator was used for conventional IRE (1500 V, 120 pulses, pulse length 100 μs). In the other lobe, the same procedure was performed preceded by the injection of a doxorubicin mixture through the expandable electrodes (IRECT). Contrast-enhanced CT and MRI were performed on days 1, 3, and 7 after the procedure. Accordingly, three animals were sacrificed on days 1, 3, and 7 after the imaging and ablation volumes were evaluated histopathologically. Related t test was used to compare the groups.

Results

Technical success was achieved in 9/10 experiments. One animal deceased during the intervention because of ventricular fibrillation. Follow-up CT 1 and 3 days after intervention showed a significant (p < 0.05) difference in the ablation volumes of IRECT vs IRE, respectively, of 4.47 ± 1.78 ml vs 2.51 ± 0.93 ml and of 3.39 ± 1.05 vs 1.53 ± 0.78 ml.

Conclusions

IRECT using the newly developed system proved to be effective and provided significantly larger ablation volumes compared with IRE alone. However, ECG triggering is a necessary prerequisite to allow a safe application of the system.

Key Points

• Working on the geometry of the IRE applicator in terms of expandable electrodes may overcome the current limitations of IRE resulting from the placement of multiple electrodes.
• Efficacy of IRE ablations can be enhanced by the interstitial application of chemotherapy in the periphery of ablation areas.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de. Zusätzlich können Sie eine Zeitschrift Ihrer Wahl in gedruckter Form beziehen – ohne Aufpreis.

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel
  1. Sie können e.Med Radiologie 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.

Neu im Fachgebiet Radiologie

Meistgelesene Bücher aus der Radiologie

2016 | Buch

Medizinische Fremdkörper in der Bildgebung

Thorax, Abdomen, Gefäße und Kinder

Dieses einzigartige Buch enthält ca. 1.600 hochwertige radiologische Abbildungen und Fotos iatrogen eingebrachter Fremdmaterialien im Röntgenbild und CT.

Herausgeber:
Dr. med. Daniela Kildal

2011 | Buch

Atlas Klinische Neuroradiologie des Gehirns

Radiologie lebt von Bildern! Der vorliegende Atlas trägt dieser Tatsache Rechnung. Sie finden zu jedem Krankheitsbild des Gehirns Referenzbilder zum Abgleichen mit eigenen Befunden.

Autoren:
Priv.-Doz. Dr. med. Jennifer Linn, Prof. Dr. med. Martin Wiesmann, Prof. Dr. med. Hartmut Brückmann

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Radiologie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise