Background
Ceramic-on-ceramic bearing has been widely applied in total hip replacement for the biological inert, well wearing capacity, and low inflammatory reaction caused by wear particles [
1,
2]. The fourth-generation ceramic bearing (BIOLOX Delta, CeramTec AG, Plochingen, Germany) has been used in total hip replacement for many years. Compared with the third generation, it added zirconia (18%), chromium oxide (< 1%), and strontium oxide (< 1%) on the basis of alumina [
3]. The addition of these components reduced the risk for prosthesis fracture [
4], however, still could not get rid of squeaking [
5].
Squeaking was a high-pitched and audible sound, which was considered as a unique but unavoidable phenomenon in hard-on-hard bearing interface [
6]. It often appeared in painless or infrequent forms, but their persistence might reduce patient satisfaction, or even cause revision surgery [
7]. What is more, it also could be an early sign of pressure peaks indicating incongruency and a risk factor for component breakage [
8].
The mechanisms of squeaking were not yet fully understood; prosthesis factors have been regarded as an important contributor [
6,
9]. There has been a study comparing four kinds of femoral stems and a finding that the presence of Accolade femoral stem (Stryker, Mahwah, NJ) was significantly associated with hip squeaking [
10]. Indeed, different brands of prostheses or different types of prosthesis with the same brand had various designs and materials; we surmised that squeaking incidence might be different when different types of prostheses were used.
To our knowledge, there was no meta-analysis to determine the occurrence rate of squeaking in the fourth generation of ceramic-on-ceramic total hip replacement. We aimed to explore the occurrence rate of squeaking in the fourth generation of ceramic-on-ceramic total hip replacement. The second purpose was to determine the risk factors for squeaking. Moreover, we also attempted to observe the occurrence rate in different prosthetic brands and find the possible relationship between squeaking and prosthetic brands.
Discussion
We did the first meta-analysis to study the squeaking incidence in the fourth-generation ceramic bearing. We found that the occurrence rate of squeaking in patients who applied the fourth-generation ceramic bearing in primary total hip replacement was nearly 3%. The presence of DePuy Delta Motion cup significantly increased the incidence of squeaking, and so did the Secure-Fit femoral stem. Age, gender, BMI, and anteversion and abduction angles of acetabular cup were not found to be associated with squeaking. There was no significant difference of the squeaking incidence between less than 5-year and more than or equal to 5-year follow-up subgroups.
The fourth-generation ceramic bearing was first applied in clinical at 2000 [
28]; Yang et al. once proposed that improvement of material properties may reduce the risk for squeaking [
30], and there was indeed no report of squeak in the fourth-generation ceramic bearing in the early study [
31], but Buttaro et al. presented the first case who suffered from postoperative squeaking in 2012 [
5]. Since then, more and more researches began to declare squeaking as a complication of the fourth-generation ceramic bearing. Thus, we did this systematical study and found the occurrence rate was 3%. Previous study has shown that the occurrence rate in the third-generation ceramic bearing fluctuated at 0.7–20.9% [
32]. Stanat and Capozzi conducted the meta-analysis for the studies which mainly applied the third-generation ceramic bearing and reported the squeaking incidence was 2.4% [
33]. It seems that the fourth-generation ceramic bearing does not have a significant superiority in postoperative squeaking.
The mechanisms of postoperative squeaking have not been clearly elucidated at present; many theories have been put forward with the desire to make a comprehensive explanation, including edge loading, lubrication imbalance, rim impingement, micro-separation, and stripe wear [
6,
7]. The prosthetic factors were accepted to be an important part; the designs and materials were different in various brands of prosthesis, or in the same brand but different types of prosthesis [
9,
34]. For instance, the Stryker Trident PSL cup had an elevated rim to increase the ceramic stabilization but caused more squeaking for impingement between femoral neck and cup [
35]; the design of Delta Motion cup did not permit reinforcement by auxiliary screws, which might impact implant stability under specific conditions and be a potential risk for squeaking [
36]. Moreover, surgical techniques were connected to squeaking. Incorrect acetabular cup orientation could cause rim impingement and edge loading [
37]. The patient’s factors also counted; reports showed that younger, taller, and more active patients may be easier to suffer hip squeaking [
38]. And preoperative diagnosis of rheumatoid arthritis might be a risk factor for squeaking [
10].
In our study, the presence of DePuy Delta Motion cup significantly increased squeak. Delta Motion was a pre-assembled cup with large-diameter femoral heads, which was engineered to maximize head-neck ratios and reduce the risk for impingement, thus decreased the possibility of squeak [
20]. However, it did not achieve the intended purpose. In fact, the technology advancement of the fourth-generation ceramic reduced the thickness of bearing and promoted the application of large-diameter femoral heads. Large-diameter femoral heads could reduce joint instability and give patients a similar sense of movement to healthy subjects [
39,
40], which has become increasingly popular in hip replacement. However, it might cause micro-separation during movement because of the small opening angle between femoral head and acetabular cup [
41]. Moreover, if the acetabular cup was too vertical, large femoral head would significantly increase the load on its edge [
42]. Micro-separation and edge loading are two critical mechanisms for squeaking.
Moreover, we know that squeak originates from irregular vibrations, which are resulted from the combined effects of initial pulse, vibration propagation, and amplification [
43]. The DePuy Delta Motion cup with larger femoral head might significantly increase the articulation surface and frictional moment at the inadequately lubricated condition, thus inducing vibration [
44]. And they might increase the prosthesis mass, which could reduce the natural frequency of vibration, increase the amplitude, and further amplify the initial vibration [
15]. In addition, McDonnell pointed out that soft tissue laxity and a wider range of movement may provide supportive environment for squeaking among patients applying Delta Motion [
20]. Combination of the above effects might be responsible for the significant higher incidence of squeak in Delta Motion cup.
Femoral stem design and metallurgical technology also have a great impact on squeak. Studies have reported that the Stryker Accolade femoral stem was prone to generate squeak [
10,
33]. In the fourth-generation ceramic bearing, only one study reported the application of Stryker femoral stem (Secure-Fit); it had the highest incidence of squeak in all the used stems, and this may be related to the unique design of Stryker femoral stem (high rim and short neck) [
9]. But Secure-Fit femoral stem had a relatively wider neck with thick long stem compared to the Accolade femoral stem [
6,
15], and it was made of Ti-6Al-4V, which has lower tensile properties and flexibility than Ti-12Mo-6Zr-2Fe (Accolade femoral stem), so it was not conducive to adhere to the femur, resulting abrasion and lubrication imbalance and then causing squeak [
35,
45]. Therefore, we speculated it was highly possible that the high incidence was attributed to the combinative use of Delta Motion cup, but the specific link need to be further explored.
In addition, the occurrence rate of squeak with the presence of Summit stem was high, but the applied patients’ number was small; our conclusion still required further verification.
We also conducted regression analysis for age, gender, BMI, acetabular cup abduction and anteversion angle; these factors have no significance for squeak, and the results were consistent with previous studies [
33,
46]. Lee [
46] once pointed out that abduction angle of acetabular cup was a risk factor for squeak, but his research was limited to the Asian population, different from ours.
Furthermore, dislocation, exposure types, and surgical indications might have effects on squeaking. Dislocation often had similar risk factors for squeaking, including improperly placed prosthesis, unreasonable prosthesis design, and soft tissue laxity. Excessive or insufficient anteversion of acetabular cup were important reasons for dislocation [
47], which might not only cause an increasing load on the cup edge, but also lead to rim impingement. In our included studies, posterolateral, posterior, direct lateral, and anterolateral approaches were applied in the surgery. Different surgical approaches had different effects on soft tissue and affect joint stability [
48], which might induce squeaking. As for surgical indications, study has shown that rheumatoid arthritis was related with squeaking [
10]. Different indications had different surgical requirements, and patients’ basic conditions were also different, which might have potential effects on squeaking. Our research was limited to the factors such as data amounts and the designs of the study; it was difficult to make further assessment. In the future, clinical data based on large sample sizes and multi-centers can help to establish the specific effects of these factors on squeaking.
The high heterogeneity was a limitation of our meta-analysis. We found the high heterogeneity mainly distributed in studies applied the DePuy Pinnacle cup and the DePuy Corail femoral stem. These two prostheses are currently most widely used; studies from different regions and populations were included in our analysis, while studies that applied other types of prostheses almost came from the same countries and populations. Thus, we supposed the high heterogeneity may be due to this. Our research still has some other limitations. Firstly, studies included were mainly on the issue of squeaking, and the data was all retrieved from the published literatures; these might cause selection or publication bias, and strictly designed and high-quality RCTs are needed in the future. Secondly, some available articles applied a variety of brands of prostheses but did not report the corresponding brands of prostheses with squeaking; removing these literatures during subgroup analysis might have an influence on results. Thirdly, the reported studies about squeaking generated in the fourth-generation ceramic bearing is limited; our findings require to be further supported by larger number of cases and more long-term follow-up studies.