Sub-millisievert CT colonography: effect of knowledge-based iterative reconstruction on the detection of colonic polyps

verfasst von: Hyo-Jin Kang, Se Hyung Kim, Cheong-Il Shin, Ijin Joo, Hwaseong Ryu, Sang Gyun Kim, Jong Pil Im, Joon Koo Han

Erschienen in: European Radiology | Ausgabe 12/2018

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To assess the feasibility of ultra-low dose computed tomography colonography (CTC) using knowledge-based iterative reconstruction (IR) and to determine its effect on polyp detection.

Methods

Forty-nine prospectively-enrolled patients underwent ultra-low dose CTC in the supine (100 kVp/20 mAs) and prone positions (80 kVp/20 mAs), followed by same-day colonoscopy. Thereafter, images were reconstructed using filtered back projection (FBP) and knowledge-based IR (IMR; Philips Healthcare, Best, Netherlands) algorithms. Effective radiation dose of CTC was recorded. Pooled per-polyp sensitivity and positive predictive value of three radiologists was analysed and compared between FBP and IMR. Image quality was assessed on a five-point scale and image noise was recorded using standard deviations.

Results

Mean effective radiation dose of ultra-low dose CTC was 0.90 ± 0.06 mSv. Eighty-nine polyps were detected on colonoscopy (mean, 8.5 ± 4.7 mm). The pooled per-polyp sensitivity for polyps 6.0-9.9 mm (n = 22) on CTC reconstructed with IMR (36/66, 54.5%) was not significantly different with that using FBP algorithm (34/66, 51.5%) (p = 0.414). For polyps ≥10 mm (n = 35), however, the pooled per-polyp sensitivity on CTC with IMR (73/105, 69.5%) was significantly higher than that with FBP (55/105, 52.4%) (p < 0.001). In particular, the difference of per-polyp sensitivity was statistically significant in intermediate (p = 0.014) and novice (p = 0.003) reviewers. Furthermore, mean image noise of IMR (8.4 ± 6.2 HU) was significantly lower than that of FBP (37.5 ± 13.9 HU) (p < 0.001) and image quality with IMR was significantly better than with FBP in all evaluated segments in all reviewers (all ps < 0.001).

Conclusions

Sub-mSv CTC reconstructed with IMR was feasible for the detection of clinically significant polyps, demonstrating 70% per-polyp sensitivity of polyps ≥10 mm, while allowing significant noise reduction and improvement in image quality compared with FBP reconstruction.

Key Points

• Sub-mSv CTC using IMR demonstrated 70% per-polyp sensitivity for polyps ≥10 mm.
• CTC using IMR significantly outperformed CTC reconstructed with FBP.
• IMR allows significantly more noise reduction and improvement in image quality than FBP.
Literatur
1.
Zurück zum Zitat Johnson CD, Chen M-H, Toledano AY et al (2008) Accuracy of CT colonography for detection of large adenomas and cancers. N Engl J Med 359:1207–1217CrossRef Johnson CD, Chen M-H, Toledano AY et al (2008) Accuracy of CT colonography for detection of large adenomas and cancers. N Engl J Med 359:1207–1217CrossRef
2.
Zurück zum Zitat Levin B, Lieberman DA, McFarland B et al (2008) Screening and surveillance for the early detection of colorectal cancer and adenomatous polyps, 2008: a joint guideline from the American Cancer Society, the US Multi-Society Task Force on Colorectal Cancer, and the American College of Radiology. CA Cancer J Clin 58:130–160CrossRef Levin B, Lieberman DA, McFarland B et al (2008) Screening and surveillance for the early detection of colorectal cancer and adenomatous polyps, 2008: a joint guideline from the American Cancer Society, the US Multi-Society Task Force on Colorectal Cancer, and the American College of Radiology. CA Cancer J Clin 58:130–160CrossRef
3.
Zurück zum Zitat Park SH, Yee J, Kim SH et al (2007) Fundamental elements for successful performance of CT colonography (virtual colonoscopy). Korean J Radiol 8:264–275CrossRef Park SH, Yee J, Kim SH et al (2007) Fundamental elements for successful performance of CT colonography (virtual colonoscopy). Korean J Radiol 8:264–275CrossRef
4.
Zurück zum Zitat Chung SY, Park SH, Lee SS et al (2012) Comparison between CT colonography and double-contrast barium enema for colonic evaluation in patients with renal insufficiency. Korean J Radiol 13:290–299CrossRef Chung SY, Park SH, Lee SS et al (2012) Comparison between CT colonography and double-contrast barium enema for colonic evaluation in patients with renal insufficiency. Korean J Radiol 13:290–299CrossRef
5.
Zurück zum Zitat Angtuaco TL, Banaad-Omiotek GD, Howden CW (2001) Differing attitudes toward virtual and conventional colonoscopy for colorectal cancer screening: surveys among primary care physicians and potential patients. Am J Gastroenterol 96:887–893CrossRef Angtuaco TL, Banaad-Omiotek GD, Howden CW (2001) Differing attitudes toward virtual and conventional colonoscopy for colorectal cancer screening: surveys among primary care physicians and potential patients. Am J Gastroenterol 96:887–893CrossRef
6.
Zurück zum Zitat Gluecker TM, Johnson CD, Harmsen WS et al (2003) Colorectal cancer screening with CT colonography, colonoscopy, and double-contrast barium enema examination: prospective assessment of patient perceptions and preferences. Radiology 227:378–384CrossRef Gluecker TM, Johnson CD, Harmsen WS et al (2003) Colorectal cancer screening with CT colonography, colonoscopy, and double-contrast barium enema examination: prospective assessment of patient perceptions and preferences. Radiology 227:378–384CrossRef
7.
Zurück zum Zitat Iannaccone R, Catalano C, Mangiapane F et al (2005) Colorectal polyps: detection with low-dose multi-detector row helical CT colonography versus two sequential colonoscopies. Radiology 237:927–937CrossRef Iannaccone R, Catalano C, Mangiapane F et al (2005) Colorectal polyps: detection with low-dose multi-detector row helical CT colonography versus two sequential colonoscopies. Radiology 237:927–937CrossRef
8.
Zurück zum Zitat Cohnen M, Vogt C, Beck A et al (2004) Feasibility of MDCT Colonography in ultra-low-dose technique in the detection of colorectal lesions: comparison with high-resolution video colonoscopy. AJR Am J Roentgenol 183:1355–1359CrossRef Cohnen M, Vogt C, Beck A et al (2004) Feasibility of MDCT Colonography in ultra-low-dose technique in the detection of colorectal lesions: comparison with high-resolution video colonoscopy. AJR Am J Roentgenol 183:1355–1359CrossRef
9.
Zurück zum Zitat Shin C-I, Kim SH, Lee ES et al (2014) Ultra-low peak voltage CT colonography: effect of iterative reconstruction algorithms on performance of radiologists who use anthropomorphic colonic phantoms. Radiology 273:759–771CrossRef Shin C-I, Kim SH, Lee ES et al (2014) Ultra-low peak voltage CT colonography: effect of iterative reconstruction algorithms on performance of radiologists who use anthropomorphic colonic phantoms. Radiology 273:759–771CrossRef
10.
Zurück zum Zitat Flicek KT, Hara AK, Silva AC et al (2010) Reducing the radiation dose for CT colonography using adaptive statistical iterative reconstruction: a pilot study. AJR Am J Roentgenol 195:126–131CrossRef Flicek KT, Hara AK, Silva AC et al (2010) Reducing the radiation dose for CT colonography using adaptive statistical iterative reconstruction: a pilot study. AJR Am J Roentgenol 195:126–131CrossRef
11.
Zurück zum Zitat Fletcher JG, Grant KL, Fidler JL et al (2012) Validation of dual-source single-tube reconstruction as a method to obtain half-dose images to evaluate radiation dose and noise reduction: phantom and human assessment using CT colonography and sinogram-affirmed iterative reconstruction (SAFIRE). J Comput Assist Tomogr 36:560–569CrossRef Fletcher JG, Grant KL, Fidler JL et al (2012) Validation of dual-source single-tube reconstruction as a method to obtain half-dose images to evaluate radiation dose and noise reduction: phantom and human assessment using CT colonography and sinogram-affirmed iterative reconstruction (SAFIRE). J Comput Assist Tomogr 36:560–569CrossRef
12.
Zurück zum Zitat Nagata K, Fujiwara M, Kanazawa H et al (2015) Evaluation of dose reduction and image quality in CT colonography: comparison of low-dose CT with iterative reconstruction and routine-dose CT with filtered back projection. Eur Radiol 25:221–229CrossRef Nagata K, Fujiwara M, Kanazawa H et al (2015) Evaluation of dose reduction and image quality in CT colonography: comparison of low-dose CT with iterative reconstruction and routine-dose CT with filtered back projection. Eur Radiol 25:221–229CrossRef
13.
Zurück zum Zitat Willemink MJ, de Jong PA, Leiner T et al (2013) Iterative reconstruction techniques for computed tomography Part 1: technical principles. Eur Radiol 23:1623–1631CrossRef Willemink MJ, de Jong PA, Leiner T et al (2013) Iterative reconstruction techniques for computed tomography Part 1: technical principles. Eur Radiol 23:1623–1631CrossRef
14.
Zurück zum Zitat Yoon MA, Kim SH, Lee JM et al (2012) Adaptive statistical iterative reconstruction and Veo: assessment of image quality and diagnostic performance in CT colonography at various radiation doses. J Comput Assist Tomogr 36:596–601CrossRef Yoon MA, Kim SH, Lee JM et al (2012) Adaptive statistical iterative reconstruction and Veo: assessment of image quality and diagnostic performance in CT colonography at various radiation doses. J Comput Assist Tomogr 36:596–601CrossRef
15.
Zurück zum Zitat Mehta D, Thompson R, Morton T, Dhanantwari A, Shefer E (2013) Iterative model reconstruction: simultaneously lowered computed tomography radiation dose and improved image quality. Med Phys Int J 2:147–155 Mehta D, Thompson R, Morton T, Dhanantwari A, Shefer E (2013) Iterative model reconstruction: simultaneously lowered computed tomography radiation dose and improved image quality. Med Phys Int J 2:147–155
16.
Zurück zum Zitat Bongartz G, Golding S, Jurik A et al (2004) European guidelines for multislice computed tomography. European Commission, Luxembourg Bongartz G, Golding S, Jurik A et al (2004) European guidelines for multislice computed tomography. European Commission, Luxembourg
17.
Zurück zum Zitat Kim SH, Lee JM, Shin CI et al (2008) Effects of spatial resolution and tube current on computer-aided detection of polyps on CT colonographic images: phantom study. Radiology 248:492–503CrossRef Kim SH, Lee JM, Shin CI et al (2008) Effects of spatial resolution and tube current on computer-aided detection of polyps on CT colonographic images: phantom study. Radiology 248:492–503CrossRef
18.
Zurück zum Zitat Shin CI, Kim SH, Im JP et al (2016) One-mSv CT colonography: Effect of different iterative reconstruction algorithms on radiologists' performance. Eur J Radiol 85:641–648CrossRef Shin CI, Kim SH, Im JP et al (2016) One-mSv CT colonography: Effect of different iterative reconstruction algorithms on radiologists' performance. Eur J Radiol 85:641–648CrossRef
19.
Zurück zum Zitat Lambert L, Ourednicek P, Briza J et al (2016) Sub-milliSievert ultralow-dose CT colonography with iterative model reconstruction technique. Peer J 4:e1883CrossRef Lambert L, Ourednicek P, Briza J et al (2016) Sub-milliSievert ultralow-dose CT colonography with iterative model reconstruction technique. Peer J 4:e1883CrossRef
20.
Zurück zum Zitat Lambert L, Ourednicek P, Jahoda J et al (2015) Model-based vs hybrid iterative reconstruction technique in ultralow-dose submillisievert CT colonography. Br J Radiol 88:20140667CrossRef Lambert L, Ourednicek P, Jahoda J et al (2015) Model-based vs hybrid iterative reconstruction technique in ultralow-dose submillisievert CT colonography. Br J Radiol 88:20140667CrossRef
21.
Zurück zum Zitat Lubner MG, Pooler BD, Kitchin DR et al (2015) Sub-milliSievert (sub-mSv) CT colonography: a prospective comparison of image quality and polyp conspicuity at reduced-dose versus standard-dose imaging. Eur Radiol 25:2089–2102CrossRef Lubner MG, Pooler BD, Kitchin DR et al (2015) Sub-milliSievert (sub-mSv) CT colonography: a prospective comparison of image quality and polyp conspicuity at reduced-dose versus standard-dose imaging. Eur Radiol 25:2089–2102CrossRef
22.
Zurück zum Zitat Yamamura S, Oda S, Imuta M et al (2016) Reducing the radiation dose for CT colonography: effect of low tube voltage and iterative reconstruction. Acad Radiol 23:155–162CrossRef Yamamura S, Oda S, Imuta M et al (2016) Reducing the radiation dose for CT colonography: effect of low tube voltage and iterative reconstruction. Acad Radiol 23:155–162CrossRef
24.
Zurück zum Zitat de Haan MC, van Gelder RE, Graser A et al (2011) Diagnostic value of CT-colonography as compared to colonoscopy in an asymptomatic screening population: a meta-analysis. Eur Radiol 21:1747–1763CrossRef de Haan MC, van Gelder RE, Graser A et al (2011) Diagnostic value of CT-colonography as compared to colonoscopy in an asymptomatic screening population: a meta-analysis. Eur Radiol 21:1747–1763CrossRef
25.
Zurück zum Zitat Lin JS, Piper MA, Perdue LA et al (2016) Screening for Colorectal Cancer: Updated Evidence Report and Systematic Review for the US Preventive Services Task Force. JAMA 315:2576–2594CrossRef Lin JS, Piper MA, Perdue LA et al (2016) Screening for Colorectal Cancer: Updated Evidence Report and Systematic Review for the US Preventive Services Task Force. JAMA 315:2576–2594CrossRef
26.
Zurück zum Zitat Sosna J, Sella T, Sy O et al (2008) Critical analysis of the performance of double-contrast barium enema for detecting colorectal polyps > or = 6 mm in the era of CT colonography. AJR Am J Roentgenol 190:374–385CrossRef Sosna J, Sella T, Sy O et al (2008) Critical analysis of the performance of double-contrast barium enema for detecting colorectal polyps > or = 6 mm in the era of CT colonography. AJR Am J Roentgenol 190:374–385CrossRef
27.
Zurück zum Zitat Chang W, Lee JM, Lee K et al (2013) Assessment of a model-based, iterative reconstruction algorithm (MBIR) regarding image quality and dose reduction in liver computed tomography. Invest Radiol 48:598–606CrossRef Chang W, Lee JM, Lee K et al (2013) Assessment of a model-based, iterative reconstruction algorithm (MBIR) regarding image quality and dose reduction in liver computed tomography. Invest Radiol 48:598–606CrossRef
28.
Zurück zum Zitat Lv P, Liu J, Zhang R et al (2015) Combined use of automatic tube voltage selection and current modulation with iterative reconstruction for CT evaluation of small hypervascular hepatocellular carcinomas: effect on lesion conspicuity and image quality. Korean J Radiol 16:531–540CrossRef Lv P, Liu J, Zhang R et al (2015) Combined use of automatic tube voltage selection and current modulation with iterative reconstruction for CT evaluation of small hypervascular hepatocellular carcinomas: effect on lesion conspicuity and image quality. Korean J Radiol 16:531–540CrossRef
29.
Zurück zum Zitat Song JS, Choi EJ, Kim EY et al (2015) Attenuation-based automatic kilovoltage selection and sinogram-affirmed iterative reconstruction: effects on radiation exposure and image quality of portal-phase liver CT. Korean J Radiol 16:69–79CrossRef Song JS, Choi EJ, Kim EY et al (2015) Attenuation-based automatic kilovoltage selection and sinogram-affirmed iterative reconstruction: effects on radiation exposure and image quality of portal-phase liver CT. Korean J Radiol 16:69–79CrossRef
Metadaten
Titel
Sub-millisievert CT colonography: effect of knowledge-based iterative reconstruction on the detection of colonic polyps
verfasst von
Hyo-Jin Kang
Se Hyung Kim
Cheong-Il Shin
Ijin Joo
Hwaseong Ryu
Sang Gyun Kim
Jong Pil Im
Joon Koo Han
Publikationsdatum
08.06.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 12/2018
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5545-5

Weitere Artikel der Ausgabe 12/2018

European Radiology 12/2018 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.