Submillisievert imaging protocol using full reconstruction and advanced patient motion correction in 320-row area detector coronary CT angiography

verfasst von: Yuko Kawaguchi, Shinichiro Fujimoto, Kazuhisa Takamura, Etsuro Kato, Shoko Suda, Rie Matsumori, Makoto Hiki, Kanako K. Kumamaru, Hiroyuki Daida

Erschienen in: The International Journal of Cardiovascular Imaging | Ausgabe 3/2018

Einloggen, um Zugang zu erhalten

Abstract

Radiation exposure remains a concern in the use of coronary CT angiography (CCTA). Full reconstruction (Full) and reconstruction using advanced patient motion correction (APMC) could obtain a lower radiation dose using low tube current scanning in a 320-row Area Detector CT (320-ADCT). The radiation dose for an imaging protocol using Full and APMC in daily practice was estimated. A total of 209 patients who underwent CCTA in 1 rotation scanning with 100 kv and adaptive iterative dose reduction 3D in 320-ADCT were enrolled. Imaging protocols were classified into 3 groups based on estimated slow filling time: (1) slow filling time ≥ 275 msec, Full with 30% of usual tube current (N = 43)(Full30%mA) (2) 206.3 msec ≤ slow filling time < 275 msec, APMC with 50% of usual tube current (N = 48)(APMC50%mA); and (3) 137.5 msec ≤ slow filling time < 206.3 msec, Half reconstruction with usual tube current (N = 118)(Half100%mA). Radiation dose was estimated by the effective dose. The diagnostic accuracy of CCTA was compared with that of invasive coronary angiography in 28 patients. The effective doses of Full30%mA, APMC50%mA, and Half100%mA were 0.77 ± 0.31, 1.30 ± 0.85, and 1.98 ± 0.68, respectively. Of 28 patients, the sensitivity, specificity, accuracy, positive predictive value, and negative predictive value in vessel-based analyses were: Full30%mA, 66.7, 82.4, 80.0, 40.0, and 93.3%; APMC50%mA, 100.0, 80.0, 83.3, 50.05, and 100.0%; and Half100%mA, 90.9, 83.0, 86.3, 78.95, and 92.9%, respectively. An imaging protocol using Full30%mA and APMC50%mA was one of the methods how radiation dose could be reduced radiation dose maintained diagnostic accuracy compared to imaging using conventional Half100%mA.
Literatur
1.
Zurück zum Zitat Budoff MJ, Dowe D, Jollis JG, Gitter M, Sutherland J, Halamert E, Scherer M, Bellinger R, Martin A, Benton R, Delago A, Min JK (2008) Diagnostic performance of 64-multidetector row coronary computed tomographic angiography for evaluation of coronary artery stenosis in individuals without known coronary artery disease: results from the prospective multicenter ACCURACY (Assessment by Coronary Computed Tomographic Angiography of Individuals Undergoing Invasive Coronary Angiography) trial. J Am Coll Cardiol 52:1724–1732CrossRefPubMed Budoff MJ, Dowe D, Jollis JG, Gitter M, Sutherland J, Halamert E, Scherer M, Bellinger R, Martin A, Benton R, Delago A, Min JK (2008) Diagnostic performance of 64-multidetector row coronary computed tomographic angiography for evaluation of coronary artery stenosis in individuals without known coronary artery disease: results from the prospective multicenter ACCURACY (Assessment by Coronary Computed Tomographic Angiography of Individuals Undergoing Invasive Coronary Angiography) trial. J Am Coll Cardiol 52:1724–1732CrossRefPubMed
2.
Zurück zum Zitat Miller JM, Rochitte CE, Dewey M, Arbab-Zadeh A, Niinuma H, Gottlieb I, Paul N, Clouse ME, Shapiro EP, Hoe J, Lardo AC, Bush DE, de Roos A, Cox C, Brinker J, Lima JA (2008) Diagnostic performance of coronary angiography by 64-row CT. N Engl J Med 359:2324–2336CrossRefPubMed Miller JM, Rochitte CE, Dewey M, Arbab-Zadeh A, Niinuma H, Gottlieb I, Paul N, Clouse ME, Shapiro EP, Hoe J, Lardo AC, Bush DE, de Roos A, Cox C, Brinker J, Lima JA (2008) Diagnostic performance of coronary angiography by 64-row CT. N Engl J Med 359:2324–2336CrossRefPubMed
3.
Zurück zum Zitat Meijboom WB, Meijs MF, Schuijf JD, Cramer MJ, Mollet NR, van Mieghem CA, Nieman K, van Werkhoven JM, Pundziute G, Weustink AC, de Vos AM, Pugliese F, Rensing B, Jukema JW, Bax JJ, Prokop M, Doevendans PA, Hunink MG, Krestin GP, Feyter PJ (2008) Diagnostic accuracy of 64-slice computed tomography coronary angiography: a prospective, multicenter, multivendor study. J Am Coll Cardiol 52:2135–2144CrossRefPubMed Meijboom WB, Meijs MF, Schuijf JD, Cramer MJ, Mollet NR, van Mieghem CA, Nieman K, van Werkhoven JM, Pundziute G, Weustink AC, de Vos AM, Pugliese F, Rensing B, Jukema JW, Bax JJ, Prokop M, Doevendans PA, Hunink MG, Krestin GP, Feyter PJ (2008) Diagnostic accuracy of 64-slice computed tomography coronary angiography: a prospective, multicenter, multivendor study. J Am Coll Cardiol 52:2135–2144CrossRefPubMed
4.
Zurück zum Zitat Einstein AJ, Henzlova MJ, Rajagopalan S (2007) Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA 29:317–323CrossRef Einstein AJ, Henzlova MJ, Rajagopalan S (2007) Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA 29:317–323CrossRef
5.
Zurück zum Zitat Richardson DB, Cardis E, Daniels RD, Gilies M, O’Hagan JA, Hamra GB, Haylock R, Laurier D, Leuraud K, Moissonnier M, Schubauer-Berigan MK, Thierry-Chef I, Kesminiene A (2015) Risk of cancer from occupational exposure to ionising radiation: retrospective cohort study of workers in France, the United Kingdom, and the United States (INWORKS). BMJ 351:h5359CrossRefPubMedPubMedCentral Richardson DB, Cardis E, Daniels RD, Gilies M, O’Hagan JA, Hamra GB, Haylock R, Laurier D, Leuraud K, Moissonnier M, Schubauer-Berigan MK, Thierry-Chef I, Kesminiene A (2015) Risk of cancer from occupational exposure to ionising radiation: retrospective cohort study of workers in France, the United Kingdom, and the United States (INWORKS). BMJ 351:h5359CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Takayanagi T, Sano T, Kondo T, Amanuma M, Ishizaka K, Sekine T, Matsutani H, Morita H, Arai T, Takase S (2015) Clinical usefulness of low tube current scanning with full reconstruction and automatic patient motion correction (APMC) reconstruction in a prospective ECG-gated coronary CT angiography using 320-row area detector CT. Nihon Hoshasen Gijutsu Gakkai Zasshi 71:237–245CrossRefPubMed Takayanagi T, Sano T, Kondo T, Amanuma M, Ishizaka K, Sekine T, Matsutani H, Morita H, Arai T, Takase S (2015) Clinical usefulness of low tube current scanning with full reconstruction and automatic patient motion correction (APMC) reconstruction in a prospective ECG-gated coronary CT angiography using 320-row area detector CT. Nihon Hoshasen Gijutsu Gakkai Zasshi 71:237–245CrossRefPubMed
7.
Zurück zum Zitat Takayanagi T, Sekine T, Kondo T, Amanuma M, Sano T, Ishizaka K, Matsutani H, Morita H, Arai T, Takase S (2016) The minimum values of (RR-PQ) for high quality coronary ct angiogram in half, automatic patient’s motion correction (APMC) and full reconstruction using 320-ADCT (0.275 s/r). Nihon Hoshasen Gijutsu Gakkai Zasshi 72:496–502CrossRefPubMed Takayanagi T, Sekine T, Kondo T, Amanuma M, Sano T, Ishizaka K, Matsutani H, Morita H, Arai T, Takase S (2016) The minimum values of (RR-PQ) for high quality coronary ct angiogram in half, automatic patient’s motion correction (APMC) and full reconstruction using 320-ADCT (0.275 s/r). Nihon Hoshasen Gijutsu Gakkai Zasshi 72:496–502CrossRefPubMed
8.
Zurück zum Zitat Tatsugami F, Matsuki M, Nakai G, Inada Y, Kanazawa S, Takeda Y, Morita H, Takada H, Yoshikawa S, Fukumura K, Narumi Y (2012) The effect of adaptive iterative dose reduction on image quality in 320-detector row CT coronary angiography. Br J Radiol 85:e378-382CrossRef Tatsugami F, Matsuki M, Nakai G, Inada Y, Kanazawa S, Takeda Y, Morita H, Takada H, Yoshikawa S, Fukumura K, Narumi Y (2012) The effect of adaptive iterative dose reduction on image quality in 320-detector row CT coronary angiography. Br J Radiol 85:e378-382CrossRef
9.
Zurück zum Zitat Williams MC, Weir NW, Mirsadraee S, Milalr F, Baird A, Minns F, Uren NG, McKillop G, Bull RK, van Beek EJ, Reid JH, Newby DE (2013) Iterative reconstruction and individualized automatic tube current selection reduce radiation dose while maintaining image quality in 320-multidetector computed tomography coronary angiography. Clin Radiol 68:e570-577 Williams MC, Weir NW, Mirsadraee S, Milalr F, Baird A, Minns F, Uren NG, McKillop G, Bull RK, van Beek EJ, Reid JH, Newby DE (2013) Iterative reconstruction and individualized automatic tube current selection reduce radiation dose while maintaining image quality in 320-multidetector computed tomography coronary angiography. Clin Radiol 68:e570-577
10.
Zurück zum Zitat Pontone G, Andreini D, Bartorelli AL, Bertella E, Mushtaq S, Foti C, Formenti A, Chiappa L, Annoni A, Cortinovis S, Baggiano A, Conte E, Bovis F, Veglia F, Ballerini G, Agostoni P, Fiorentini C, Pepi M (2012) Feasibility and diagnostic accuracy of a low radiation exposure protocol for prospective ECG-triggering coronary MDCT angiography. Clin Radiol 67:207–215CrossRefPubMed Pontone G, Andreini D, Bartorelli AL, Bertella E, Mushtaq S, Foti C, Formenti A, Chiappa L, Annoni A, Cortinovis S, Baggiano A, Conte E, Bovis F, Veglia F, Ballerini G, Agostoni P, Fiorentini C, Pepi M (2012) Feasibility and diagnostic accuracy of a low radiation exposure protocol for prospective ECG-triggering coronary MDCT angiography. Clin Radiol 67:207–215CrossRefPubMed
11.
Zurück zum Zitat Layritz C, Schmid J, Achenbach S, Ulzheimer S, Wuest W, May M, Ropers D, Klinghammer L, Daniel WG, Pflederer T, Lell M (2014) Acccuracy of prospectively ECG-triggered very low-dose coronary dual-source T angiography using iterative reconstruction for the detection of coronary artery stenosis: comparison with invasive catheterization. Eur Heart J Cardiovasc Imaging 15:1238–1245CrossRefPubMed Layritz C, Schmid J, Achenbach S, Ulzheimer S, Wuest W, May M, Ropers D, Klinghammer L, Daniel WG, Pflederer T, Lell M (2014) Acccuracy of prospectively ECG-triggered very low-dose coronary dual-source T angiography using iterative reconstruction for the detection of coronary artery stenosis: comparison with invasive catheterization. Eur Heart J Cardiovasc Imaging 15:1238–1245CrossRefPubMed
12.
Zurück zum Zitat Hou Y, Ma Y, Fan W, Wang Y, Yu M, Vembar M, Guo Q (2014) Diagnostic accuracy of low-dose 256-slice multi-detector coronary CT angiography using iterative reconstruction in patients with suspected coronary artery disease. Eur Radiol 24:3–11CrossRefPubMed Hou Y, Ma Y, Fan W, Wang Y, Yu M, Vembar M, Guo Q (2014) Diagnostic accuracy of low-dose 256-slice multi-detector coronary CT angiography using iterative reconstruction in patients with suspected coronary artery disease. Eur Radiol 24:3–11CrossRefPubMed
13.
Zurück zum Zitat Stehli J, Fuchs TA, Bull S, Clerc OF, Possner M, Buech RR, Gaemperli O, Kaufmann PA (2014) Accuracy of coronary CT angiography using a submillisievert fraction of radiation exposure. J Am Coll Cardiol 64:772–780CrossRefPubMed Stehli J, Fuchs TA, Bull S, Clerc OF, Possner M, Buech RR, Gaemperli O, Kaufmann PA (2014) Accuracy of coronary CT angiography using a submillisievert fraction of radiation exposure. J Am Coll Cardiol 64:772–780CrossRefPubMed
14.
Zurück zum Zitat Oda S, Utsunomiya D, Funama Y, Yonenaga K, Namimoto T, Nakaura T, Yamashita Y (2012) A hybrid iterative redonstruction algorithm that improves the image quality of low- tube-voltage coronary CT angiography. AJR Am J Roentgenol 198:1126–1131CrossRefPubMed Oda S, Utsunomiya D, Funama Y, Yonenaga K, Namimoto T, Nakaura T, Yamashita Y (2012) A hybrid iterative redonstruction algorithm that improves the image quality of low- tube-voltage coronary CT angiography. AJR Am J Roentgenol 198:1126–1131CrossRefPubMed
15.
Zurück zum Zitat Zhang LJ, Wang Y, Schoepf UJ, Meiner FG, Bayer RR 2nd, Qi L, Cao J, Zhou CS, Zhao YE, Li X, Gong JB, Jin Z, Lu GM (2016) Image quality, radiation dose, and diagnostic accuracy of prospectively ECG-triggered high-pitch coronary CT angiography at 80 kVp in a clinical setting: comparison with invasive coronary angiography. Eur Radiol 26:797–806CrossRefPubMed Zhang LJ, Wang Y, Schoepf UJ, Meiner FG, Bayer RR 2nd, Qi L, Cao J, Zhou CS, Zhao YE, Li X, Gong JB, Jin Z, Lu GM (2016) Image quality, radiation dose, and diagnostic accuracy of prospectively ECG-triggered high-pitch coronary CT angiography at 80 kVp in a clinical setting: comparison with invasive coronary angiography. Eur Radiol 26:797–806CrossRefPubMed
16.
Zurück zum Zitat Sano T, Kondo T, Matsutani H, Morita H, Arai T, Sekine T, Takase S, Oida A, Fukuzawa H, Kodama T, Kondo M, Orihira T, Yamada N, Jagat N (2009) Significance of PQ interval in acquisition of coronary multidetector row computed tomography. J Cardiol 54:441–451CrossRefPubMed Sano T, Kondo T, Matsutani H, Morita H, Arai T, Sekine T, Takase S, Oida A, Fukuzawa H, Kodama T, Kondo M, Orihira T, Yamada N, Jagat N (2009) Significance of PQ interval in acquisition of coronary multidetector row computed tomography. J Cardiol 54:441–451CrossRefPubMed
17.
Zurück zum Zitat Hoe J, Toh KH (2009) First experience with 320-row multidetector CT coronary angiography scanning with prospective electrocardiogram gating to reduce radiation dose. J Cardiovasc Comput Tomogr 3:257–261CrossRefPubMed Hoe J, Toh KH (2009) First experience with 320-row multidetector CT coronary angiography scanning with prospective electrocardiogram gating to reduce radiation dose. J Cardiovasc Comput Tomogr 3:257–261CrossRefPubMed
18.
Zurück zum Zitat Mori S, Nishizawa K, Ohno M, Endo M (2006) Conversion factor for CT dosimetry to assess patient dose using a 256-slice CT scanner. Br J Radiol 79:888–892CrossRefPubMed Mori S, Nishizawa K, Ohno M, Endo M (2006) Conversion factor for CT dosimetry to assess patient dose using a 256-slice CT scanner. Br J Radiol 79:888–892CrossRefPubMed
19.
Zurück zum Zitat Rybicki FJ, Otero HJ, Steigner ML, Vorobiof G, Nallamshetty L, Mitsouras D, Ersoy H, Mather RT, Judy PF, Cai T, Coyner K, Schultz K, Whitmore AG, Di Carli MF (2008) Initial evaluation of coronary images from 320-detector row computed tomography. Int J Cardiovasc Imaging 24:535–546CrossRefPubMed Rybicki FJ, Otero HJ, Steigner ML, Vorobiof G, Nallamshetty L, Mitsouras D, Ersoy H, Mather RT, Judy PF, Cai T, Coyner K, Schultz K, Whitmore AG, Di Carli MF (2008) Initial evaluation of coronary images from 320-detector row computed tomography. Int J Cardiovasc Imaging 24:535–546CrossRefPubMed
20.
Zurück zum Zitat ICRP Publication 102 (2007) Managing patient dose in multi-detector computed tomography (MDCT). International Commission on Radiological Protection. Elsevier, Amsterdam ICRP Publication 102 (2007) Managing patient dose in multi-detector computed tomography (MDCT). International Commission on Radiological Protection. Elsevier, Amsterdam
21.
Zurück zum Zitat Matsutani H, Sano T, Kondo T, Morita H, Arai T, Sekine T, Takase S, Oida A, Fukazawa H, Suguta M, Kondo M, Kodama T, Orihara T, Yamada N, Tsuyuki M, Narula J (2008) ECG-edit function in multidetector-row computed tomography coronary arteriography for patients with arrhythmias. Circ J 72:1071–1078CrossRefPubMed Matsutani H, Sano T, Kondo T, Morita H, Arai T, Sekine T, Takase S, Oida A, Fukazawa H, Suguta M, Kondo M, Kodama T, Orihara T, Yamada N, Tsuyuki M, Narula J (2008) ECG-edit function in multidetector-row computed tomography coronary arteriography for patients with arrhythmias. Circ J 72:1071–1078CrossRefPubMed
22.
Zurück zum Zitat Gosling O, Loader R, Venables P, Rowles N, Morgan-Hughes G, Roobottom C (2010) Cardiac CT: are we underestimating the dose? A radiation dose study utilizing the 2007 ICRP tissue weighting factors and a cardiac specific scan volume. Clin Radiol 65:1013–1017CrossRefPubMed Gosling O, Loader R, Venables P, Rowles N, Morgan-Hughes G, Roobottom C (2010) Cardiac CT: are we underestimating the dose? A radiation dose study utilizing the 2007 ICRP tissue weighting factors and a cardiac specific scan volume. Clin Radiol 65:1013–1017CrossRefPubMed
23.
Zurück zum Zitat Tanaka R, Yoshioka K, Muranaka K, Chiba T, Ueda T, Sasaki T, Fusazaki T, Ehara S (2013) Improved evaluation of calsified segments on coronary CT angiography: a feasibility study of coronary carcium subtraction. Int J Cardiovasc Imaging 29:75–81CrossRefPubMed Tanaka R, Yoshioka K, Muranaka K, Chiba T, Ueda T, Sasaki T, Fusazaki T, Ehara S (2013) Improved evaluation of calsified segments on coronary CT angiography: a feasibility study of coronary carcium subtraction. Int J Cardiovasc Imaging 29:75–81CrossRefPubMed
24.
Zurück zum Zitat Yoshioka K, Tanaka R, Muranaka K, Sasaki T, Ueda T, Chiba T, Takeda K, Sugawara T (2015) Subtraction coronary CT angiography using second-generation 320-detector row CT. Int J Cardiovasc Imaging 31:51–58CrossRefPubMed Yoshioka K, Tanaka R, Muranaka K, Sasaki T, Ueda T, Chiba T, Takeda K, Sugawara T (2015) Subtraction coronary CT angiography using second-generation 320-detector row CT. Int J Cardiovasc Imaging 31:51–58CrossRefPubMed
Metadaten
Titel
Submillisievert imaging protocol using full reconstruction and advanced patient motion correction in 320-row area detector coronary CT angiography
verfasst von
Yuko Kawaguchi
Shinichiro Fujimoto
Kazuhisa Takamura
Etsuro Kato
Shoko Suda
Rie Matsumori
Makoto Hiki
Kanako K. Kumamaru
Hiroyuki Daida
Publikationsdatum
12.09.2017
Verlag
Springer Netherlands
Erschienen in
The International Journal of Cardiovascular Imaging / Ausgabe 3/2018
Print ISSN: 1569-5794
Elektronische ISSN: 1875-8312
DOI
https://doi.org/10.1007/s10554-017-1237-5

Weitere Artikel der Ausgabe 3/2018

The International Journal of Cardiovascular Imaging 3/2018 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.