01.03.2020 | Original Article

Superior performance of 18F-fluorocholine digital PET/CT in the detection of parathyroid adenomas

verfasst von: Diego Alfonso López-Mora, Marina Sizova, Montserrat Estorch, Albert Flotats, Valle Camacho, Alejandro Fernández, Safae Abouzian, Francisco Fuentes-Ocampo, José Ignacio Pérez Garcia, Ana Isabel Chico Ballesteros, Joan Duch, Anna Domènech, Antonio Moral Duarte, Ignasi Carrió

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 3/2020

Einloggen, um Zugang zu erhalten

Abstract

Objective

To compare detectability of hyperfunctioning parathyroid tissue (HPT) by digital and analog 18F-fluorocholine PET/CT in patients with primary hyperparathyroidism and negative/inconclusive 99mTc-MIBI scintigraphy-SPECT/CT.

Materials and methods

Thirty-three patients with primary hyperparathyroidism and negative/inconclusive 99mTc-MIBI scintigraphy-SPECT/CT were prospectively included. All patients accepted to be scanned by digital and analog PET/CT in the same imaging session after a single injection of 18F-fluorocholine. Three nuclear medicine physicians evaluated the digital and analog PET/CT datasets to assess the detection rate of HPT. Maximum standard uptake values (SUVmax) of HPT and locoregional lymph nodes were measured in both systems.

Results

HPT was detected in 30/33 patients by the digital system, whereas it was detected in 22/33 patients by the analog system (p < 0.01). Moreover, in 21 of these 33 patients, both systems detected one focal 18F-fluorocholine uptake, and in one patient the digital system detected two foci. Histopathology demonstrated HPT in 32 patients and it was inconclusive in one patient. The digital PET/CT detected HPT in 29 of the 32 patients, and the analog system in 22 of the 32 (p < 0.01). All HPT suspected lesions resected and detected only by the digital system (n = 8) were < 10 mm (7.5 ± 1.3 mm), while those detected by both systems (n = 22) were > 10 mm (13 ± 3.8 mm). SUVmax of HPT lesions was significantly higher than SUVmax of locoregional lymph node independently of the PET/CT system used (4.5 ± 1.9 vs. 2.9 ± 1.3, p < 0.0001).

Conclusions

Digital PET/CT offers superior performance over analog system in patients with suspected HPT and previous negative/inconclusive imaging examinations, particularly in sub-centimeter lesions. SUVmax can help in the differentiation between HTP and locoregional lymph nodes.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Cordellat IM. Hyperparathyroidism: primary or secondary disease? Rheumatol Clin. 2012;8:287–91. Cordellat IM. Hyperparathyroidism: primary or secondary disease? Rheumatol Clin. 2012;8:287–91.
2.
Zurück zum Zitat Wilhelm SM, Wang TS, Ruan DT, et al. The American Association of Endocrine Surgeons guidelines for definitive management of primary hyperparathyroidism. JAMA Surg. 2016;151:959–68.CrossRef Wilhelm SM, Wang TS, Ruan DT, et al. The American Association of Endocrine Surgeons guidelines for definitive management of primary hyperparathyroidism. JAMA Surg. 2016;151:959–68.CrossRef
3.
Zurück zum Zitat Beheshti M, Hehenwarter L, Paymani Z, et al. 18F-Fluorocholine PET/CT in the assessment of primary hyperparathyroidism compared with 99mTc-MIBI or 99mTc-tetrofosmin SPECT/CT: a prospective dual-centre study in 100 patients. Eur J Nucl Med Mol Imaging. 2018;45:1762–71.CrossRef Beheshti M, Hehenwarter L, Paymani Z, et al. 18F-Fluorocholine PET/CT in the assessment of primary hyperparathyroidism compared with 99mTc-MIBI or 99mTc-tetrofosmin SPECT/CT: a prospective dual-centre study in 100 patients. Eur J Nucl Med Mol Imaging. 2018;45:1762–71.CrossRef
4.
Zurück zum Zitat Treglia G, Piccardo A, Imperiale A, et al. Diagnostic performance of choline PET for detection of hyperfunctioning parathyroid glands in hyperparathyroidism: a systematic review and meta-analysis. Eur J Nucl Med Mol Imaging. 2019;46:751–65.CrossRef Treglia G, Piccardo A, Imperiale A, et al. Diagnostic performance of choline PET for detection of hyperfunctioning parathyroid glands in hyperparathyroidism: a systematic review and meta-analysis. Eur J Nucl Med Mol Imaging. 2019;46:751–65.CrossRef
5.
Zurück zum Zitat Huber GF, Hüllner M, Schmid C, et al. Benefit of 18F-fluorocholine PET imaging in parathyroid surgery. Eur Radiol. 2018;28:2700–7.CrossRef Huber GF, Hüllner M, Schmid C, et al. Benefit of 18F-fluorocholine PET imaging in parathyroid surgery. Eur Radiol. 2018;28:2700–7.CrossRef
6.
Zurück zum Zitat Thanseer N, Bhadada SK, Sood A, et al. Comparative effectiveness of ultrasonography, 99mTc-sestamibi, and 18F-fluorocholine PET/CT in detecting parathyroid adenomas in patients with primary hyperparathyroidism. Clin Nucl Med. 2017;42:e491–7.CrossRef Thanseer N, Bhadada SK, Sood A, et al. Comparative effectiveness of ultrasonography, 99mTc-sestamibi, and 18F-fluorocholine PET/CT in detecting parathyroid adenomas in patients with primary hyperparathyroidism. Clin Nucl Med. 2017;42:e491–7.CrossRef
7.
Zurück zum Zitat Bossert I, Chytiris S, Hodolic M, et al. PETC/CT with 18F-choline localizes hyperfunctioning parathyroid adenomas equally well in normocalcemic hyperparathyroidism as in overt hyperparathyroidism. J Endocrinol Investig. 2019;42:419–26.CrossRef Bossert I, Chytiris S, Hodolic M, et al. PETC/CT with 18F-choline localizes hyperfunctioning parathyroid adenomas equally well in normocalcemic hyperparathyroidism as in overt hyperparathyroidism. J Endocrinol Investig. 2019;42:419–26.CrossRef
8.
Zurück zum Zitat Nguyen NC, Vercher-Conejero JL, Sattar A, et al. Image quality and diagnostic performance of a digital PET prototype in patients with oncologic diseases: initial experience and comparison with analog PET. J Nucl Med. 2015;56:1378–85.CrossRef Nguyen NC, Vercher-Conejero JL, Sattar A, et al. Image quality and diagnostic performance of a digital PET prototype in patients with oncologic diseases: initial experience and comparison with analog PET. J Nucl Med. 2015;56:1378–85.CrossRef
9.
Zurück zum Zitat Frach T, Prescher G, Degenhardt C, et al. The digital silicon photomultiplier: principle of operation and intrinsic detector performance. IEEE Nucl Sci Symp Conf Rec. 2009:1959–65. Frach T, Prescher G, Degenhardt C, et al. The digital silicon photomultiplier: principle of operation and intrinsic detector performance. IEEE Nucl Sci Symp Conf Rec. 2009:1959–65.
10.
Zurück zum Zitat Degenhardt C, Prescher G, Frach T, et al. The digital silicon photomultiplier: a novel sensor for the detection of scintillation light. IEEE Nucl Sci Symp Conf Rec. 2009:2383–6. Degenhardt C, Prescher G, Frach T, et al. The digital silicon photomultiplier: a novel sensor for the detection of scintillation light. IEEE Nucl Sci Symp Conf Rec. 2009:2383–6.
11.
Zurück zum Zitat Degenhardt C, Rodrigues P, Trindade A, et al. Performance evaluation of a prototype positron emission tomography scanner using digital photon counters (DPC). IEEE Nucl Sci Symp Conf Rec. 2012:2820–4. Degenhardt C, Rodrigues P, Trindade A, et al. Performance evaluation of a prototype positron emission tomography scanner using digital photon counters (DPC). IEEE Nucl Sci Symp Conf Rec. 2012:2820–4.
12.
Zurück zum Zitat López-Mora DA, Flotats A, Fuentes-Ocampo F, et al. Comparison of image quality and lesion detection between digital and analog PET/CT. Eur J Nucl Med Mol Imaging. 2019;46:1383–90.CrossRef López-Mora DA, Flotats A, Fuentes-Ocampo F, et al. Comparison of image quality and lesion detection between digital and analog PET/CT. Eur J Nucl Med Mol Imaging. 2019;46:1383–90.CrossRef
13.
Zurück zum Zitat Rausch I, Ruiz A, Valverde-Pascual I, et al. Performance evaluation of the Vereos PET/CT system according to the NEMA NU2–2012 standard. J Nucl Med. 2019;60:561–7.CrossRef Rausch I, Ruiz A, Valverde-Pascual I, et al. Performance evaluation of the Vereos PET/CT system according to the NEMA NU2–2012 standard. J Nucl Med. 2019;60:561–7.CrossRef
14.
Zurück zum Zitat Surti S, Kuhn A, Werner ME, et al. Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities. J Nucl Med. 2007;48:471–80.PubMed Surti S, Kuhn A, Werner ME, et al. Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities. J Nucl Med. 2007;48:471–80.PubMed
15.
Zurück zum Zitat Rausch I, Cal-González J, Dapra D, et al. Performance evaluation of the biograph mCT flow PET/CT system according to the NEMA NU2-2012 standard. EJNMMI Phys. 2015;2:26.CrossRef Rausch I, Cal-González J, Dapra D, et al. Performance evaluation of the biograph mCT flow PET/CT system according to the NEMA NU2-2012 standard. EJNMMI Phys. 2015;2:26.CrossRef
16.
Zurück zum Zitat López-Mora DA, Estorch M, Fuentes-Ocampo F, et al. Digital PET/CT vs. analogue PET/CT in a parathyroid gland study with 18F-fluorocholine. Rev Esp Med Nucl Imagen Mol. 2019;38:121–2.PubMed López-Mora DA, Estorch M, Fuentes-Ocampo F, et al. Digital PET/CT vs. analogue PET/CT in a parathyroid gland study with 18F-fluorocholine. Rev Esp Med Nucl Imagen Mol. 2019;38:121–2.PubMed
17.
Zurück zum Zitat Broos WAM, Wondergem M, Knol RJJ, et al. Parathyroid imaging with 18F-fluorocholine PET/CT as a first-line imaging modality in primary hyperparathyroidism: a retrospective cohort study. EJNMMI Res. 2019;9:72.CrossRef Broos WAM, Wondergem M, Knol RJJ, et al. Parathyroid imaging with 18F-fluorocholine PET/CT as a first-line imaging modality in primary hyperparathyroidism: a retrospective cohort study. EJNMMI Res. 2019;9:72.CrossRef
18.
Zurück zum Zitat Grimaldi S, Young J, Kamenicky P, et al. Challenging pre-surgical localization of hyperfunctioning parathyroid glands in primary hyperparathyroidism: the added value of (18)F-fluorocholine PET/CT. Eur J Nucl Med Mol Imaging. 2018;45:1772–80.CrossRef Grimaldi S, Young J, Kamenicky P, et al. Challenging pre-surgical localization of hyperfunctioning parathyroid glands in primary hyperparathyroidism: the added value of (18)F-fluorocholine PET/CT. Eur J Nucl Med Mol Imaging. 2018;45:1772–80.CrossRef
19.
Zurück zum Zitat Amadou C, Bera G, Ezziane M, et al. 18F-Fluorocholine PET/CT and parathyroid 4D computed tomography for primary hyperparathyroidism: the challenge of reoperative patients. World J Surg. 2019;43:1232–42.CrossRef Amadou C, Bera G, Ezziane M, et al. 18F-Fluorocholine PET/CT and parathyroid 4D computed tomography for primary hyperparathyroidism: the challenge of reoperative patients. World J Surg. 2019;43:1232–42.CrossRef
20.
Zurück zum Zitat Khafif A, Masalha M, Landsberg R, et al. The role of F18-fluorocholine positron emission tomography/magnetic resonance imaging in localizing parathyroid adenomas. Eur Arch Otorhinolaryngol. 2019;276:1509–16.CrossRef Khafif A, Masalha M, Landsberg R, et al. The role of F18-fluorocholine positron emission tomography/magnetic resonance imaging in localizing parathyroid adenomas. Eur Arch Otorhinolaryngol. 2019;276:1509–16.CrossRef
21.
Zurück zum Zitat Prabhu M, Kumari G, Damle NA, et al. Assessment of the role of early dynamic PET/CT with 18F-fluorocholine in detection of parathyroid lesions in patients with primary hyperparathyroidism. Nucl Med Commun. 2018;39:1190–6.CrossRef Prabhu M, Kumari G, Damle NA, et al. Assessment of the role of early dynamic PET/CT with 18F-fluorocholine in detection of parathyroid lesions in patients with primary hyperparathyroidism. Nucl Med Commun. 2018;39:1190–6.CrossRef
22.
Zurück zum Zitat Fuentes-Ocampo F, López-Mora DA, Flotats A, et al. Digital vs. analog PET/CT: intra-subject comparison of the SUVmax in target lesions and reference regions. Eur J Nucl Med Mol Imaging. 2019;46:1745–50.CrossRef Fuentes-Ocampo F, López-Mora DA, Flotats A, et al. Digital vs. analog PET/CT: intra-subject comparison of the SUVmax in target lesions and reference regions. Eur J Nucl Med Mol Imaging. 2019;46:1745–50.CrossRef
Metadaten
Titel
Superior performance of 18F-fluorocholine digital PET/CT in the detection of parathyroid adenomas
verfasst von
Diego Alfonso López-Mora
Marina Sizova
Montserrat Estorch
Albert Flotats
Valle Camacho
Alejandro Fernández
Safae Abouzian
Francisco Fuentes-Ocampo
José Ignacio Pérez Garcia
Ana Isabel Chico Ballesteros
Joan Duch
Anna Domènech
Antonio Moral Duarte
Ignasi Carrió
Publikationsdatum
01.03.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 3/2020
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-020-04680-7

Weitere Artikel der Ausgabe 3/2020

European Journal of Nuclear Medicine and Molecular Imaging 3/2020 Zur Ausgabe