Erschienen in: Archives of Orthopaedic and Trauma Surgery 9/2020

09.06.2020 | Hip Arthroplasty

Surgeon-stratified cohort analysis of 1976 cementless Zweymüller total hip arthroplasties from a single hospital with 23,255 component years of follow-up

verfasst von: Samo Roškar, Vane Antolič, Blaž Mavčič

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 9/2020

Einloggen, um Zugang zu erhalten

Abstract

Introduction

The third-generation Zweymüller hip endoprosthesis has been used for decades with excellent results in arthroplasty registries, but surgeon-stratified reports on this implant are still scarce. The aim of the presented single-hospital cohort analysis of the third-generation Endoplus-Zweymüller primary total hip arthroplasty was to determine implant survival rates until the first revision/removal at 10/15/20/25 years after implantation and to find out whether implant survival depended on the operating surgeon, patient’s age and gender, operated side (right/left), season of the year and the set of implanted components (SL-PLUS femur, BICON-PLUS acetabulum or both components).

Materials and methods

The retrospective observational cohort study included 1976 consecutive patients with primary BICON/SL-PLUS Zweymüller hip endoprosthesis implanted at the same operating theater block of the University Medical Centre Ljubljana, Department of Orthopaedic Surgery (Ljubljana, Slovenia) between January 1, 1993 and May 1, 2014. Survival analyses were performed with the Kaplan–Meier method and the Cox regression analysis after minimum 5 and maximum 26 years of follow-up.

Results

At 10/15/20/25 years after implantation, the cumulative proportion of revision-free surviving Zweymüller total hip endoprostheses was 92/90/85/81% and the cumulative proportion with unremoved immobile components was 93/91/87/84%, respectively. After adjustment for age, gender and operated side, less experienced surgeons (odds ratios 2.34–5.00), season of the year at primary implantation (spring vs. winter odds ratio 1.74) and the use of BICON-PLUS acetabulum with femoral stem from another manufacturer (odds ratio 2.23) were significant risk factors of worse implant survival.

Conclusions

The study presents the largest published third-generation Zweymüller BICON/SL-PLUS total hip arthroplasty cohort from a single non-developmental hospital with 23,255 component years of observation. Study findings indicate the impact of environmental factors at primary implantation and variability between different surgeons on the long-term implant survival.
Literatur
1.
Zurück zum Zitat Zweymüller K, Litner F, Böhm G (1995) The development of the cementless hip endoprothesis: 1979–1994. In: Morscher EW (ed) Endoprosthetics. Springer, Berlin, pp 309–325CrossRef Zweymüller K, Litner F, Böhm G (1995) The development of the cementless hip endoprothesis: 1979–1994. In: Morscher EW (ed) Endoprosthetics. Springer, Berlin, pp 309–325CrossRef
2.
Zurück zum Zitat Wright C, Lambert D, Brazil D, Keggi K, Keggi J, McTighe T (2011) Clinical review of the Zweymuller femoral stem. Reconstr Rev 1(1):41–52 Wright C, Lambert D, Brazil D, Keggi K, Keggi J, McTighe T (2011) Clinical review of the Zweymuller femoral stem. Reconstr Rev 1(1):41–52
3.
Zurück zum Zitat van Oldenrijk J, Scholtes VAB, van Beers LWAH, Geerdink CH, Niers BBAM, Runne W, Bhandari M, Poolman RW, CUSTOM trail research collaborative (2017) Better early functional outcome after short stem total hip arthroplasty? A prospective blinded randomised multicentre trial comparing the collum femoris preserving stem with a Zweymuller straight cementless stem total hip replacement for the treatment of primary osteoarthritis of the hip. BMJ Open 7(10):e014522CrossRef van Oldenrijk J, Scholtes VAB, van Beers LWAH, Geerdink CH, Niers BBAM, Runne W, Bhandari M, Poolman RW, CUSTOM trail research collaborative (2017) Better early functional outcome after short stem total hip arthroplasty? A prospective blinded randomised multicentre trial comparing the collum femoris preserving stem with a Zweymuller straight cementless stem total hip replacement for the treatment of primary osteoarthritis of the hip. BMJ Open 7(10):e014522CrossRef
4.
Zurück zum Zitat Schmolders J, Amvrazis G, Pennekamp PH, Strauss AC, Fridrich MJ, Wimmer MD, Rommelspracher Y, Wirtz DC, Allny T (2017) Thirteen year follow-up of a cementless femoral stem and a threaded acetabular cup in patients younger than fifty years of age. Int Orthop 41(1):39–45CrossRef Schmolders J, Amvrazis G, Pennekamp PH, Strauss AC, Fridrich MJ, Wimmer MD, Rommelspracher Y, Wirtz DC, Allny T (2017) Thirteen year follow-up of a cementless femoral stem and a threaded acetabular cup in patients younger than fifty years of age. Int Orthop 41(1):39–45CrossRef
5.
Zurück zum Zitat Studers P, Belajevs D, Jurkevics V, Likums P (2016) Ten to fifteen-year clinical and radiographic follow-up with a third-generation cementless stem in a young patient population. Int Orthop 40(3):465–471CrossRef Studers P, Belajevs D, Jurkevics V, Likums P (2016) Ten to fifteen-year clinical and radiographic follow-up with a third-generation cementless stem in a young patient population. Int Orthop 40(3):465–471CrossRef
6.
Zurück zum Zitat Milošev I, Kovač S, Trebše R, Levašič V, Pišot V (2012) Comparison of ten-year survivorship of hip prostheses with use of conventional polyethylene, metal-on-metal, or ceramic-on-ceramic bearings. J Bone Jt Surg Am 94(19):1756–1763CrossRef Milošev I, Kovač S, Trebše R, Levašič V, Pišot V (2012) Comparison of ten-year survivorship of hip prostheses with use of conventional polyethylene, metal-on-metal, or ceramic-on-ceramic bearings. J Bone Jt Surg Am 94(19):1756–1763CrossRef
7.
Zurück zum Zitat Vogel D, Rathay A, Teufel S, Ellenrieder M, Zietz C, Sander M, Bader R (2017) Experimental analysis of insertion torques and forces of threaded and press-fit acetabular cups by means of ex vivo and in vivo measurements. Acta Bioeng Biomech 19(3):155–163PubMed Vogel D, Rathay A, Teufel S, Ellenrieder M, Zietz C, Sander M, Bader R (2017) Experimental analysis of insertion torques and forces of threaded and press-fit acetabular cups by means of ex vivo and in vivo measurements. Acta Bioeng Biomech 19(3):155–163PubMed
8.
Zurück zum Zitat Ottink K, Barnaart L, Westerbeck R, van Kampen K, Bulstra S, van Jonbergen HP (2015) Survival, clinical and radiological outcome of the Zweymüller SL/Bicon-Plus total hip arthroplasty: a 15-year follow-up study. Hip Int 25(3):204–208CrossRef Ottink K, Barnaart L, Westerbeck R, van Kampen K, Bulstra S, van Jonbergen HP (2015) Survival, clinical and radiological outcome of the Zweymüller SL/Bicon-Plus total hip arthroplasty: a 15-year follow-up study. Hip Int 25(3):204–208CrossRef
9.
Zurück zum Zitat Gielis WP, van Oldenrijk J, Ten Cate N, Scholtes VAB, Geerdink CH, Poolman RW (2019) Increased persistent mid-thigh pain after short-stem compared with wedge-shaped straight-stem uncemented total hip arthroplasty at medium-term follow-up: a randomized double-blinded cross-sectional study. J Arthroplasty 35(5):912–919CrossRef Gielis WP, van Oldenrijk J, Ten Cate N, Scholtes VAB, Geerdink CH, Poolman RW (2019) Increased persistent mid-thigh pain after short-stem compared with wedge-shaped straight-stem uncemented total hip arthroplasty at medium-term follow-up: a randomized double-blinded cross-sectional study. J Arthroplasty 35(5):912–919CrossRef
10.
Zurück zum Zitat Busch VJ, Pouw MH, Laumen AM, van Susante JL, Vervest AM (2012) Long-term outcome of 73 Zweymüller total hip prostheses with a screw cup in patients under 50 years of age. Hip Int 22(3):292–295CrossRef Busch VJ, Pouw MH, Laumen AM, van Susante JL, Vervest AM (2012) Long-term outcome of 73 Zweymüller total hip prostheses with a screw cup in patients under 50 years of age. Hip Int 22(3):292–295CrossRef
11.
Zurück zum Zitat Veen EJD, Schrier JCM, van’t Reit E (2016) Outcome of the cementless Zweymüller BICON-PLUS cup and SL-PLUS stem in the very elderly individuals. Geriatr Orthop Surg Rehabil 7(2):74–80CrossRef Veen EJD, Schrier JCM, van’t Reit E (2016) Outcome of the cementless Zweymüller BICON-PLUS cup and SL-PLUS stem in the very elderly individuals. Geriatr Orthop Surg Rehabil 7(2):74–80CrossRef
12.
Zurück zum Zitat Kolb A, Grübl A, Schneckener CD, Chiari C, Kaider A, Lass R, Windhager R (2012) Cementless total hip arthroplasty with rectangular Zweymüller stem: a concise follow-up, at a minimum of twenty years, of previous reports. J Bone Jt Surg Am 94(18):1681–1684CrossRef Kolb A, Grübl A, Schneckener CD, Chiari C, Kaider A, Lass R, Windhager R (2012) Cementless total hip arthroplasty with rectangular Zweymüller stem: a concise follow-up, at a minimum of twenty years, of previous reports. J Bone Jt Surg Am 94(18):1681–1684CrossRef
13.
Zurück zum Zitat Weissinger M, Helmreich C, Pöll G (2010) Results covering 20 years use of the cement-free Zweymüller Alloclassic total endoprosthesis of the hip joint. Acta Chir Orthop Traumatol Cech 77(3):186–193PubMed Weissinger M, Helmreich C, Pöll G (2010) Results covering 20 years use of the cement-free Zweymüller Alloclassic total endoprosthesis of the hip joint. Acta Chir Orthop Traumatol Cech 77(3):186–193PubMed
14.
Zurück zum Zitat Labek G, Kovač S, Levašič V, Janda W, Zagra L (2012) The outcome of the cementless tapered SL-Plus stem: an analysis of arthroplasty register data. Int Orthop 36(6):1149–1154CrossRef Labek G, Kovač S, Levašič V, Janda W, Zagra L (2012) The outcome of the cementless tapered SL-Plus stem: an analysis of arthroplasty register data. Int Orthop 36(6):1149–1154CrossRef
15.
Zurück zum Zitat Labek G, Thaler M, Janda W, Argeiter M, Stöckl B (2011) Revisions rates after total joint replacement: cumulative results from worldwide joint register datasets. J Bone Jt Surg Br 93(3):293–297CrossRef Labek G, Thaler M, Janda W, Argeiter M, Stöckl B (2011) Revisions rates after total joint replacement: cumulative results from worldwide joint register datasets. J Bone Jt Surg Br 93(3):293–297CrossRef
16.
Zurück zum Zitat Toossi N, Adeli B, Timperley AJ, Haddad FS, Maltenfort M, Parvizi J (2013) Acetabular components in total hip arthroplasty: is there evidence that cementless fixation is better? J Bone Jt Surg Am 95(2):168–174CrossRef Toossi N, Adeli B, Timperley AJ, Haddad FS, Maltenfort M, Parvizi J (2013) Acetabular components in total hip arthroplasty: is there evidence that cementless fixation is better? J Bone Jt Surg Am 95(2):168–174CrossRef
17.
Zurück zum Zitat Pieringer H, Auersperg V, Böhler N (2006) Long-term results of the cementless Alloclassic hip arthroplasty system using a 28-mm ceramic head: with retrospective comparison to a 32-mm head. J Arthroplasty 21(7):967–974CrossRef Pieringer H, Auersperg V, Böhler N (2006) Long-term results of the cementless Alloclassic hip arthroplasty system using a 28-mm ceramic head: with retrospective comparison to a 32-mm head. J Arthroplasty 21(7):967–974CrossRef
19.
Zurück zum Zitat Cruz-Pardos A, García-Rey E, García-Cimbrelo E (2017) Total hip arthroplasty with use of the cementless Zweymüller alloclassic system: a concise follow-up, at a minimum of 25 years, of a previous report. J Bone Jt Surg Am 99(22):1927–1931CrossRef Cruz-Pardos A, García-Rey E, García-Cimbrelo E (2017) Total hip arthroplasty with use of the cementless Zweymüller alloclassic system: a concise follow-up, at a minimum of 25 years, of a previous report. J Bone Jt Surg Am 99(22):1927–1931CrossRef
20.
Zurück zum Zitat Pisecky L, Hipmair G, Schauer B, Böhler N (2019) 30-Years of experience with the cementless implanted Alloclassic CSF screw cup total hip arthroplasty system—an ultra-long-term follow-up. J Orthop 16(2):182–186CrossRef Pisecky L, Hipmair G, Schauer B, Böhler N (2019) 30-Years of experience with the cementless implanted Alloclassic CSF screw cup total hip arthroplasty system—an ultra-long-term follow-up. J Orthop 16(2):182–186CrossRef
21.
Zurück zum Zitat Makarewich CA, Anderson MB, Gililland JM, Pelt CE, Peters CL (2018) Ten-year survivorship of primary total hip arthroplasty in patients 30 years of age or younger. Bone Jt J 100(7):867–874CrossRef Makarewich CA, Anderson MB, Gililland JM, Pelt CE, Peters CL (2018) Ten-year survivorship of primary total hip arthroplasty in patients 30 years of age or younger. Bone Jt J 100(7):867–874CrossRef
22.
Zurück zum Zitat Parkinson B, Armit D, McEwen P, Lorimer M, Harris IA (2018) Is climate associated with revision for prosthetic joint infection after primary TKA? Clin Orthop Relat Res 476(6):1200–1204CrossRef Parkinson B, Armit D, McEwen P, Lorimer M, Harris IA (2018) Is climate associated with revision for prosthetic joint infection after primary TKA? Clin Orthop Relat Res 476(6):1200–1204CrossRef
23.
Zurück zum Zitat Kane P, Chen C, Post Z, Radcliff K, Orozco F, Ong A (2014) Seasonality of infection rates after total joint arthroplasty. Orthopedics 37(2):e182–e186CrossRef Kane P, Chen C, Post Z, Radcliff K, Orozco F, Ong A (2014) Seasonality of infection rates after total joint arthroplasty. Orthopedics 37(2):e182–e186CrossRef
24.
Zurück zum Zitat Malik AT, Azmat SK, Ali A et al (2018) Seasonal influence on postoperative complications after total knee arthroplasty. Knee Surg Relat Res 30(1):42–49CrossRef Malik AT, Azmat SK, Ali A et al (2018) Seasonal influence on postoperative complications after total knee arthroplasty. Knee Surg Relat Res 30(1):42–49CrossRef
25.
Zurück zum Zitat Hasegawa Y, Amano T (2015) Surgical skills training for primary total hip arthroplasty. Nagoya J Med Sci 77(1–2):51–57PubMedPubMedCentral Hasegawa Y, Amano T (2015) Surgical skills training for primary total hip arthroplasty. Nagoya J Med Sci 77(1–2):51–57PubMedPubMedCentral
26.
Zurück zum Zitat Wilson MD, Dowsey MM, Spelman T, Choong PF (2016) Impact of surgical experience on outcomes in total joint arthroplasties. ANZ J Surg 86(12):967–972CrossRef Wilson MD, Dowsey MM, Spelman T, Choong PF (2016) Impact of surgical experience on outcomes in total joint arthroplasties. ANZ J Surg 86(12):967–972CrossRef
27.
Zurück zum Zitat Jolbäck P, Rolfson O, Mohaddes M, Nemes S, Kärrholm J, Garellick G, Lindahl H (2018) Does surgeon experience affect patient-reported outcomes 1 year after primary total hip arthroplasty? Acta Orthop 89(3):265–271CrossRef Jolbäck P, Rolfson O, Mohaddes M, Nemes S, Kärrholm J, Garellick G, Lindahl H (2018) Does surgeon experience affect patient-reported outcomes 1 year after primary total hip arthroplasty? Acta Orthop 89(3):265–271CrossRef
28.
Zurück zum Zitat Cossec CL, Colas S, Zureik M (2017) Relative impact of hospital and surgeon procedure volumes on primary total hip arthroplasty revision: a nationwide cohort study in France. Arthroplast Today 3(3):176–182CrossRef Cossec CL, Colas S, Zureik M (2017) Relative impact of hospital and surgeon procedure volumes on primary total hip arthroplasty revision: a nationwide cohort study in France. Arthroplast Today 3(3):176–182CrossRef
29.
Zurück zum Zitat Bohl DD, Nolte MT, Ong K, Lau E, Calkins TE, Della Valle CJ (2019) Computer-assisted navigation is associated with reductions in the rates of dislocation and acetabular component revision following primary total hip arthroplasty. J Bone Jt Surg Am 101(3):250–256CrossRef Bohl DD, Nolte MT, Ong K, Lau E, Calkins TE, Della Valle CJ (2019) Computer-assisted navigation is associated with reductions in the rates of dislocation and acetabular component revision following primary total hip arthroplasty. J Bone Jt Surg Am 101(3):250–256CrossRef
Metadaten
Titel
Surgeon-stratified cohort analysis of 1976 cementless Zweymüller total hip arthroplasties from a single hospital with 23,255 component years of follow-up
verfasst von
Samo Roškar
Vane Antolič
Blaž Mavčič
Publikationsdatum
09.06.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 9/2020
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-020-03517-0

Weitere Artikel der Ausgabe 9/2020

Archives of Orthopaedic and Trauma Surgery 9/2020 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.