Open Access 11.02.2020 | Research Article

T2 mapping of the distal sciatic nerve in healthy subjects and patients suffering from lumbar disc herniation with nerve compression

verfasst von: Nico Sollmann, Dominik Weidlich, Elisabeth Klupp, Barbara Cervantes, Carl Ganter, Claus Zimmer, Ernst J. Rummeny, Thomas Baum, Jan S. Kirschke, Dimitrios C. Karampinos

Erschienen in: Magnetic Resonance Materials in Physics, Biology and Medicine | Ausgabe 5/2020

Abstract

Objective

To measure T2 values for magnetic resonance neurography (MRN) of the healthy distal sciatic nerve and compare those to T2 changes in patients with nerve compression.

Materials and methods

Twenty-one healthy subjects and five patients with sciatica due to disc herniation underwent MRN using a T2-prepared turbo spin echo (TSE) sequence of the distal sciatic nerve bilaterally. Six and one of those healthy subjects further underwent a commonly used multi-echo spin-echo (MESE) sequence and magnetic resonance spectroscopy (MRS), respectively.

Results

T2 values derived from the T2-prepared TSE sequence were 44.6 ± 3.0 ms (left) and 44.5 ± 2.6 ms (right) in healthy subjects and showed good inter-reader reliability. In patients, T2 values of 61.5 ± 6.2 ms (affected side) versus 43.3 ± 2.4 ms (unaffected side) were obtained. T2 values of MRS were in good agreement with measurements from the T2-prepared TSE, but not with those of the MESE sequence.

Discussion

A T2-prepared TSE sequence enables precise determination of T2 values of the distal sciatic nerve in agreement with MRS. A MESE sequence tends to overestimate nerve T2 compared to T2 from MRS due to the influence of residual fat on T2 quantification. Our approach may enable to quantitatively assess direct nerve affection related to nerve compression.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
2D
Two-dimensional
3D
Three-dimensional
a.u.
Arbitrary units
BMRC
British Medical Research Council
CIDP
Chronic inflammatory demyelinating polyneuropathy
CMTD
Charcot–Marie–Tooth disease
DTI
Diffusion tensor imaging
FOV
Field of view
LSP
Lumbosacral plexus
MRI
Magnetic resonance imaging
MRN
Magnetic resonance neurography
MRS
Magnetic resonance spectroscopy
MS
Multiple sclerosis
MESE
Multi-echo spin-echo
ODI
Oswestry Disability Index
PNS
Peripheral nervous system
R1
Reader 1
R2
Reader 2
RMSCV
Root mean square coefficient of variation
RMSE
Root mean square error
ROI
Region of interest
SPAIR
Spectral attenuated inversion recovery
STEAM
Stimulated echo acquisition mode
TE
Echo time
TM
Mixing time
TR
Repetition time
TSE
Turbo spin echo
VAS
Visual analogue scale
VOI
Volume of interest

Introduction

Magnetic resonance neurography (MRN) of the peripheral nervous system (PNS), the application of magnetic resonance imaging (MRI) for the depiction and study of peripheral nerves of the human body, is increasingly applied in vivo. Advances in protocols and MRI scanner techniques now allow depicting even small or highly oblique nerves originating from the brachial plexus or lumbosacral plexus (LSP) [13]. A prominent nerve originating from the LSP is the sciatic nerve, which has been routinely under investigation in studies using MRN in the past. It is composed of tibial and common peroneal components of sciatic plexus fibers (rami of the L4 to S3 nerve roots) that become enclosed in a common nerve sheath [4]. The sciatic nerve enters the gluteal region and courses inferiorly between the adductor magnus and gluteus maximus muscles to the distal one-third of the thigh where it divides into the tibial and common peroneal trunks [4].
Using qualitative MRN and structural evaluations, the sciatic nerve has shown to be affected in various pathologies, such as hamstring injury due to different causes [5], injection injury after iatrogenic intramuscular infiltrations [6], injury related to hip replacement surgeries [7], perineuriomas [8], Fabry disease [911], hereditary motor and sensory neuropathies such as Charcot–Marie–Tooth disease (CMTD) [12], or diabetic and other neuropathies [13, 14]. These studies conducted primarily visual inspections of the sciatic nerves to detect changes in signal intensities, discontinuities, or lesion loads, often together with evaluations of morphological nerve sizes and volumes [514].
In addition to such qualitative MRN, (semi-)quantitative approaches have been introduced recently. Diffusion parameters derived from diffusion tensor imaging (DTI) have been evaluated for the sciatic nerve in patients with muscular disorders such as myotonic dystrophy [15], radiculopathy [16], chronic inflammatory demyelinating polyneuropathy (CIDP) [17], CMTD [18], oxaliplatin-induced polyneuropathy [19], or diabetic neuropathy [20, 21]; however, studies have not always been successful in showing changes in diffusion parameters related to the respective pathologic conditions. In addition to that, a study relating semi-quantitatively measured magnetic transfer ratios to clinical markers found interesting correlations in patients with CMTD [22]. Further, measurements of T2 values have been achieved in addition in the same patients or diseases [1821, 23], and also in patients suffering from multiple sclerosis (MS) [24] or amyloid polyneuropathy [25]. In general, nerves enclose multiple water components with distinct relaxation times, resulting in a multi-compartment T2 signal decay [2628]. Most studies report one short T2 component (around 20 ms) and two or three longer T2 components (50–200 ms). To address these complex relaxation properties of nerve tissue, the signal decay curve can be either sampled densely with the signal decay being fully modelled, or the respective echo times (TEs) of the imaging sequence are selected in such a way that certain compartments are not visible on the signal decay curve.
Despite being increasingly used and yielding promising results among patients with primarily neuropathic diseases, evidence is still lacking for even more common conditions such as sciatica occurring due to structural pathology such as degenerative disc disease with nerve compression [29, 30]. Quantitative MRN by means of measuring T2 values is still at an incipient stage and mainly restricted to specialized centers. Hence, reference data regarding T2 values of the sciatic nerve and investigations of reliability are largely lacking. To date, the report of normal values regarding T2 of the sciatic nerve is restricted to few data of control groups of healthy or asymptomatic patients formed for comparisons to patients with diseases and focal or systemic PNS affections. A recent study explicitly focused on the assessment of normal T2 values of the sciatic nerve by investigating a cohort of 60 healthy individuals by means of MRN using a multi-echo spin-echo (MESE) sequence for T2 relaxometry at 3 Tesla [31]. This study reported on physiological T2 values in five men and five women of every decade between 20 and 80 years and additionally evaluated associations with demographic variables [31]. However, the issue of reliability has not been explicitly addressed, and it remains unclear whether the sequence used is optimal for realistic assessment of T2 values. Hence, further data on physiological T2 values and systematic evaluations of reliability are highly needed for a more profound understanding of T2 values of the sciatic nerve, thus paving the way to identify valid biomarkers in quantitative MRN of the PNS.
Against this background, the objective of the present study is (1) to provide T2 values for the healthy distal sciatic nerve and to evaluate reliability of quantitative MRN by T2 mapping in healthy subjects, (2) to compare our novel T2 mapping approach against T2 measurements in a commonly used MESE sequence (having nerve-water T2 from MRS as the reference standard), and (3) to evaluate potential alterations in T2 values of the distal sciatic nerve due to degenerative disc disease with nerve compression at the level of the LSP in patients suffering from unilateral sciatica. We hypothesize that T2 mapping using a T2-prepared turbo spin echo (TSE) sequence shows good agreement with nerve-water T2 from MRS, due to robust suppression of residual fat that can affect T2 quantification whilst being characterized by high precision. Furthermore, we propose that patients with nerve compression at the level of the LSP present with unilaterally elevated T2 values of the distal sciatic nerve when compared to the unaffected side or healthy subjects, probably due to edematous changes or swelling of nerves related to proximal compressive effects.

Materials and methods

Ethics

The study protocol was approved by the local ethics commission (registration number 408/15S) and was followed in accordance with the Declaration of Helsinki. Written informed consent was obtained from all included subjects.

Study cohort

This prospective study enrolled twenty-one healthy volunteers (14 males and 7 females, mean age 28.0 ± 3.7 years, age range 22–39 years) and five patients (3 males and 2 females, mean age 68.1 ± 11.2 years, age range 48–75 years) with unilateral sciatica due to degenerative disc disease. Inclusion criteria were (1) written informed consent, (2) age above 18 years, (3) strict unilateral sciatica matching at least one dermatome with the presence of symptoms (pain with or without additional paresthesia or impairment of motor function) for at least 4 weeks and not longer than 6 months (only for the patient group), and (4) previous anatomical MRI suggesting degenerative disc disease with disc herniation and unilateral nerve contact of degenerated disc material (only for the patient group). Exclusion criteria for both the healthy volunteers and patients were (1) pregnancy, (2) contraindications for MRI, and (3) previous surgery at the hip, thigh, or lumbar spine (except for surgical decompression without any implants).
The healthy volunteers and patients underwent MRN during a 1-day study visit. Furthermore, the patients underwent detailed clinical examinations that included assessments of coordination, sensory function, and muscle strength according to the British Medical Research Council (BMRC) scale. Additionally, the Oswestry Disability Index (ODI) was calculated [32, 33], and rating of sciatica according to a visual analogue scale (VAS) was obtained [34, 35].

Acquisition of magnetic resonance imaging

MRN was performed using a whole-body 3-Tesla MRI scanner (Ingenia, Philips Healthcare, Best, The Netherlands) using a 16-channel T/R knee coil, with the lower edge of the coil ending at the upper edge of the patella, thus covering the middle to lower thigh region. Both thighs were scanned subsequently, with the side scanned first being subject to randomization. The scanning protocol included the following sequences:
  • Two-dimensional (2D) high-resolution T2-weighted DIXON TSE sequence (acquired in all healthy volunteers and patients): field of view (FOV) = 140 × 150 × 104 mm3, acquisition voxel size = 0.3 × 0.3 × 8 mm3, TE = 60 ms, repetition time (TR) = 2500 ms, scan time = 6 min 10 s, SENSE acceleration factor = 2.0.
  • Three-dimensional (3D) T2-prepared TSE sequence (acquired in all healthy volunteers and patients): FOV = 140 × 150 × 52 mm3, acquisition voxel size = 0.7 × 0.7 × 8 mm3, TE = 30 ms, TR = 1600 ms, scan time = 3 min 30 s, fat suppression using Spectral Attenuated Inversion Recovery (SPAIR), SENSE acceleration factor = 2.5, T2 preparation applying a B1-insensitive rotation radiofrequency pulse as previously used [36, 37], T2 preparation duration = 20/40/60/80 ms. Together with the effective T2 weighting during the TSE readout, the effective TE of the sequence was 47.9 ms, 67.9 ms, 87.9 ms, and 107.9 ms.
  • 2D MESE sequence (acquired in six healthy volunteers): FOV = 160 × 160 × 38 mm3, acquisition voxel size = 0.9 × 0.8 × 3.5 mm3, 12 echoes with TE1/ΔTE = 10 ms/10 ms, TR = 2400 ms, scan time = 6 min 48 s, fat suppression using SPAIR, SENSE acceleration factor = 2.4. For this sequence, the standard vendor’s implementation was employed.
  • Stimulated Echo Acquisition Mode (STEAM) magnetic resonance spectroscopy (MRS; acquired in one healthy volunteer): volume of interest (VOI) = 5 × 5 × 10 mm3, mixing time (TM) = 11 ms, TE = 20/30/40/50/60/70/80/90/100/110/120 ms, TR = 4000 ms, scan time = 6 min 30 s, eight averages. The planning of the MRS sequence was based on the T2-weighted DIXON images and the VOI was placed in such a way that it incorporated the distal sciatic nerve.

Data processing and measurements of T2 values

T2 mapping

After exporting the data from the MRI scanner, T2 values derived from the 3D T2-prepared TSE sequence (in all healthy volunteers and all patients) and from the 2D MESE sequence (in six healthy volunteers) were calculated offline on a voxel-by-voxel basis with a combination of variable projection and golden section search assuming a mono-exponential signal decay [3841]. A mono-exponential signal decay of the reported multi-compartment T2 decay of nerve tissue was employed because of the inherently strong T2 weighting of the proposed 3D T2-prepared TSE sequence. The resulting T2 maps (one T2 map for the left and right side per healthy subject or patient and sequence) were successively opened in Horos (version 1.1.7; https://​www.​horosproject.​org) and color-coded according to a standardized scheme (Figs. 1,2) [40, 41].
First, the distal sciatic nerve was carefully identified on axial slices of the color-coded T2 maps, co-registered to uncolored anatomical images of the 2D T2-weighted DIXON TSE sequence that served as reference for enhanced structural interpretability. Then polygonal regions of interest (ROIs) were manually placed in the three consecutive slices directly above the nerve bifurcation of the sciatic nerve, thus above the split into tibial and common peroneal trunks (Figs. 1, 2) [4]. Each of these ROIs contoured the whole nerve caliber whilst inclusion of surrounding soft tissue or blood vessels was avoided. T2 values were separately extracted from the three consecutive ROIs drawn in the T2 maps derived from the 3D T2-prepared TSE sequence and the 2D MESE sequence, if available (Figs. 1,2). The approach of ROI drawing was the same in healthy subjects and patients and for both sides, respectively.
In healthy volunteers, the described approach of ROI positioning and extraction of T2 values from T2 maps of the 3D T2-prepared TSE sequence and 2D MESE sequence was performed by a medical doctor [with experience in neuroradiological imaging since 2012, reader 1 (R1)]. Furthermore, to assess inter-reader reliability, ROI positioning and extraction of T2 values from T2 maps of the 3D T2-prepared TSE sequence was also performed by a second medical doctor [with experience in neuroradiological imaging since 2014, reader 2 (R2)]. In patients, only R1 evaluated the imaging data of both sides using the T2 maps of the 3D T2-prepared TSE sequence. During analyses, readers were strictly blinded to the measurements of each other, and R1 was furthermore blinded to the side of sciatica in patients and to previous anatomical MRI that suggested degenerative disc disease with disc herniation.

Spectroscopy

The processing of the MRS data was performed with in-house software and included zero-order phasing, Gaussian apodization, and frequency alignment of single acquisitions. Peak area quantification was achieved considering eight fat peaks and was considered separately for each TE. The T2 fitting was only performed on the water peak signal and was carried out twice, once considering all available TEs and once only considering the matched TEs with the 3D T2-prepared TSE sequence.

Statistical analyses

Statistics and generation of graphs were performed using GraphPad Prism (version 6.0; GraphPad Software Inc., La Jolla, CA, USA) and SPSS (version 23.0; IBM, Chicago, IL, USA). A p value < 0.05 was considered statistically significant.
In healthy volunteers, the measured T2 values of the ROIs of the three consecutive slices were averaged, which was achieved separately for both sides, the measurements of both readers, and for values of the T2 maps as derived from the 3D T2-prepared TSE sequence or 2D MESE sequence. Differences between sides were assessed using Wilcoxon matched-pairs signed rank tests, which was achieved separately for the measurements of both readers and measurements derived from the T2 maps of the 3D T2-prepared TSE sequence and the 2D MESE sequence. Moreover, Wilcoxon matched-pairs signed rank tests were conducted to compare T2 values of the 3D T2-prepared TSE sequence and the 2D MESE sequence. To assess inter-reader reliability for T2 mapping derived from the 3D T2-prepared TSE sequence, we calculated the root mean square error (RMSE) and the root mean square coefficient of variation (RMSCV) based on the values obtained from R1 and R2, again separately considering T2 values of the left and right sides [42]. Since MRS was only acquired in one healthy subject, no statistical testing was performed for related T2 values.
In patients, the measured T2 values by R1 derived from the ROIs of the three consecutive slices were averaged, which was achieved separately for the affected side and the contralateral side with respect to the side of disc herniation and unilateral nerve contact. The T2 values of the affected side were compared to the left-sided and right-sided values obtained from the healthy subjects derived from evaluations of R1 according to T2 mapping from the 3D T2-prepared TSE sequence using Mann–Whitney tests.

Results

Healthy subjects

MRN including T2 mapping of the distal sciatic nerve with the 3D T2-prepared TSE sequence was successfully achieved bilaterally in all enrolled healthy volunteers, with the distal sciatic nerve being clearly interpretable in all subjects. Thus, forty-two T2 values (twenty-one T2 values per side as the result of averaging measurements obtained from ROIs in three consecutive slices) derived from the 3D T2-prepared TSE sequence were available for analyses. Furthermore, six volunteers also underwent bilateral scanning of the distal sciatic nerve with the 2D MESE sequence, leading to twelve T2 values (six T2 values per side as the result of averaging measurements obtained from ROIs in three consecutive slices) derived from this sequence. One volunteer also underwent MRS bilaterally using STEAM.

T2 mapping—3D T2-prepared TSE sequence

The T2 values derived from the 3D T2-prepared TSE sequence were 44.6 ± 3.0 ms (R1) and 45.7 ± 3.3 ms (R2) for the left and 44.5 ± 2.6 ms (R1) and 45.7 ± 3.0 ms (R2) for the right distal sciatic nerve (Table 1, Fig. 3). There was no statistically significant difference between sides (R1 p = 0.87, R2 p = 0.92; Table 1, Fig. 3). Furthermore, there was high inter-reader agreement regarding measurements of T2 values (Table 1).
Table 1
T2 values and inter-reader reliability
 
R1
R2
RMSE
RMSCV
T2 value (in ms; mean ± standard deviation)
    
 Left side
44.6 ± 3.0
45.7 ± 3.3
0.29
0.67%
 Right side
44.5 ± 2.6
45.7 ± 3.0
0.34
0.73%
p value
0.87
0.92
  
T2 values (in ms) as mean ± standard deviation as measured in T2 maps of the three-dimensional (3D) T2-prepared turbo spin echo (TSE) sequence for the left- and right-sided distal sciatic nerves of twenty-one healthy subjects are depicted. Measurements were performed by two readers (R1 and R2), with evaluations of inter-reader reliability presented as root mean square error (RMSE) and root mean square coefficient of variation (RMSCV). There were no statistically significant differences in T2 values between sides among the healthy subjects according to the evaluations of both readers (p > 0.05)
Figure 4 shows the T2-weighted raw images of the 3D T2-prepared TSE sequence and the corresponding nerve signal decay curves. A homogenous and consistent fat suppression can be observed for all effective TEs. The nerve signal decay shows a mono-exponential course.

T2 mapping—2D MESE sequence

The T2 values derived from the 2D MESE sequence were 63.7 ± 3.8 ms for the left and 63.4 ± 3.1 ms for the right distal sciatic nerve (R1; Fig. 3). There was no statistically significant difference between sides (R1 p = 0.56; Fig. 3). Thus, in comparison to T2 values derived from the 3D T2-prepared TSE sequence, T2 values of the 2D MESE sequence were clearly higher for both the left and right sides (R1 p = 0.03 each; Fig. 3).
Figure 4 shows the T2-weighted raw images of the 2D MESE sequence for the TEs matching the effective TEs of the 3D T2-prepared TSE sequence. The signal of the fat is increasing with increasing TEs. In the corresponding signal decay curves, stimulated echoes and an increasing fat component at longer TEs are observed. The T2 value obtained by fitting all TEs and by fitting the matched TEs is elevated compared to the values derived from the 3D T2-prepared TSE and MRS sequences.

Spectroscopy

The T2 values derived from MRS were 48.1 ms (37.9 ms unmatched) for the left and 43.6 ms (40.7 ms unmatched) for the right distal sciatic nerve for the volunteer examined (R1).
Figure 4 shows the comparison of MRS with the results from the T2 mapping derived from the 3D T2-prepared TSE sequence and the 2D MESE sequence, respectively. The signal decay appears to be multi-compartmental. However, at longer TEs (corresponding to the effective TEs of the 3D T2-prepared TSE sequence), a mono-exponential decay can be observed. The T2 values obtained by the 2D MESE sequence are considerably higher compared to the T2 values derived from the 3D T2-prepared TSE sequence and when compared to MRS measurements when the signal decay is fitted for the matched TE range.

Patients

Five patients underwent MRN with T2 mapping of the distal sciatic nerve bilaterally using the 3D T2-prepared TSE sequence. These patients initially presented in the outpatient clinic with unilateral sciatica due to degenerative disc disease with disc herniation and unilateral nerve contact of degenerated disc material according to previous anatomical MRI (Fig. 5). Neuroforaminal to extra-neuroforaminal disc herniations with nerve contact affecting L4 or L5 were present in these patients (Fig. 5). A mean ODI score of 35.2 ± 10.7 was registered, with a mean pain rating according to the VAS of 7.4 ± 1.1 points. None of the patients suffered from a manifest hemiparesis according to the BMRC scale. Two patients reported previous surgery for decompression at a different level than the currently affected level 37 and 17 years ago, respectively.
When discriminating between T2 values of the affected and non-affected side derived from the 3D T2-prepared TSE sequence for the distal sciatic nerve, patients showed T2 values of 61.5 ± 6.2 ms for the affected side and 43.3 ± 2.4 ms for the contralateral, unaffected side (R1). When comparing the T2 values of the affected side against the T2 values obtained in the healthy subjects, significant differences were observed for both the comparison against left-sided and right-sided normal T2 values (R1 p < 0.0001 each).

Discussion

We successfully applied MRN with T2 mapping in a cohort of twenty-one healthy subjects, reporting on normal T2 values for the distal sciatic nerve of approximately 45 ms at 3 Tesla when considering a 3D T2-prepared TSE sequence. A more commonly used 2D MESE sequence showed higher T2 values for the distal sciatic nerve of around 63 ms; however, data derived from MRS were in accordance with the T2 values derived from the 3D T2-prepared TSE sequence. Patients showed significantly elevated T2 values unilaterally for the distal sciatic nerve when compared to healthy subjects applying the 3D T2-prepared TSE sequence.
The sciatic nerve has regularly been under investigation by means of MRN. However, until recently, primarily qualitative MRN has been carried out, with the signal of the sciatic nerve or its morphology being altered in the context of different pathologic conditions such as traumatic or iatrogenic injury and tumors such as perineuriomas [58], Fabry disease [911], CMTD [12], or diabetic and other neuropathies [13, 14]. Importantly, (semi-)quantitative MRN might be more objective and potentially more robust and reproducible when compared to mere qualitative MRN, with DTI, for instance, being capable of contributing with quantitatively assessable changes by means of diffusion parameters that have already been evaluated among patients with muscular disorders such as myotonic dystrophy [15], radiculopathy [16], CIDP [17], CMTD [18], or different kinds of neuropathies [1921]. Further, measurements of T2 values have been obtained recently in the same diseases and further chronic systemic diseases, such as MS or amyloid polyneuropathy [1821, 2325]. Although the approach of (semi-)quantitative MRN using DTI or evaluations of T2 is on the increase, normal values from healthy subjects and analyses of reliability are still largely missing, thus hampering clear identification of pathologic values due to the lack of references.
To the authors’ knowledge, normal T2 values of the sciatic nerve have been primarily gathered from few control groups of healthy subjects or asymptomatic patients that were formed for comparisons to diseased cohorts with symptoms, with only one recent study being available to date that aimed on the distinct exploration of normal T2 values of the sciatic nerve [31]. This study investigated 60 healthy individuals by applying a MESE sequence for T2 relaxometry at 3 Tesla and demonstrated physiological T2 values in 5 men and 5 women of every decade between 20 and 80 years [31]. The study reported mean T2 values of 70.85 ± 4.91 ms [31]. These values are, however, clearly higher in comparison to the T2 values obtained with the 3D T2-prepared TSE sequence in the present study, which were 44.6 ± 3.0 ms for the left and 44.5 ± 2.6 ms for the right distal sciatic nerve (according to measurements of R1). The 2D MESE sequence, which was conducted in addition to the 3D T2-prepared TSE sequence for T2 measurements in six subjects, showed T2 values that were comparable to those reported in the previous study, accounting for 63.7 ± 3.8 ms for the left and 63.4 ± 3.1 ms for the right distal sciatic nerve. However, the T2 values detected by the 3D T2-prepared TSE sequence are in close agreement with the T2 values obtained by MRS that has been acquired in one volunteer and can be seen as the reference acquisition regarding T2 measurements.
A potential reason for the considerable discrepancy between the results derived from T2 mapping with the 2D MESE sequence and the 3D T2-prepared TSE sequence as well as MRS could be the influence of residual fat on the T2 quantification. Peripheral nerves are surrounded by fat and also partially consist of fat. The fat suppression technique of SPAIR does not suppress the olefinic fat peak and the residual fat signal could thus potentially lead to an overestimation of the pure nerve signal. Indeed, previous work in muscle tissue with fatty infiltration has shown that the employed 3D T2-prepared TSE sequence is less affected by the influence of fat and, thus, shows good agreement with the method of MRS for this purpose [43]. Other, minor factors contributing to discrepancies between T2 values of our investigation and the previous study might be represented by differences in placing the coil and defining the ROIs. However, previous research on other nerves of the PNS (e.g., tibial nerve or median nerve) among healthy subjects using different coils and ways of T2 extraction has also arrived at T2 values between 40 and 50 ms on average at 3 Tesla [44, 45]. Overall, this suggests that normal T2 values might be comparable between different nerves of the PNS in different body parts, with the chosen sequence probably being the most essential contributor to discrepancies when comparing values between studies.
So far, the focus of the small body of literature using (semi-)quantitative MRN for evaluations of the sciatic nerve was on systemic diseases with components affecting also the PNS or rather rarely observed diseases [15, 17, 18, 20, 21, 2325]. However, other conditions, such as sciatica resulting from degenerative disc disease with nerve compression due to disc herniation, are much more common [29, 30]. Recently, MRN with T2 mapping has been shown to be able to detect elevated T2 values at the level of the LSP in compressed nerves when compared to the contralateral side or a control level without pathology, which was described even in the absence of clear qualitative, visually assessed signal alterations in affected nerves [41]. The present study extends these findings by showing more distant alterations of nerves by means of T2 values at the level of the distal sciatic nerve. Thus, while clinically used anatomical MRI at the LSP may detect structural changes and disc herniation, quantitative MRN using T2 mapping of the distal sciatic nerve might be capable of detecting direct nerve affection in a timely fashion. Follow-up studies may need to investigate further the alterations of T2 values in sciatica due to degenerative disc disease with disc herniation and unilateral nerve contact to explore how to best use this additional quantitative information. It seems possible that the magnitude of T2 alterations due to such conditions may show changes over time, which might even allow to discriminate between irreversibly affected nerves, e.g. in patients not responding well to later decompressive surgery, and nerves that are not lastingly affected.
As the study’s limitations, the size of the study cohort that was comparatively small has to be mentioned. Although forty-two T2 values derived from the 3D T2-prepared TSE sequence were obtained as a result of the segmentation of both the left-sided and right-sided distal sciatic nerves in healthy subjects, upcoming studies using data derived from larger cohorts are highly needed to provide normative data. Furthermore, factors such as sex, age, or body mass index should be considered in future studies as they may have an impact on T2 measurements. Moreover, we have only reported on a case series of patients, with upcoming studies having to confirm the results observed in the small number of patients of this investigation. However, the finding of differences in T2 values between sides in the patients and when comparing T2 values between healthy subjects and patients still points at the clinical relevance of quantitative MRN using T2 mapping. The multi-compartment signal decay of peripheral nerves should also be considered as a potential shortcoming regarding determination of T2 values, especially with respect to the desirable comparability of MRI sequences of different vendors or centers. In this study, a sequence with strong initial T2 weighting was employed that mitigates contributions of short T2 components on the signal decay curves. With the used 3D T2-prepared TSE sequence a very good agreement with MRS measurements was observed. This indicates that this sequence is well suited in the presented context but other sequences might have to be adapted to achieve a similar fat suppression and TE to deliver comparable results. Another limitation regarding the MRS measurements is the potential contribution of surrounding water-containing tissues. We tried to avoid muscle tissue surrounding the nerves when placing the VOI. However, contributions of the surrounding tissue due to partial volume averaging cannot be fully excluded but should be minor compared to the contribution of the nerves.
In conclusion, this is one of the first studies to apply quantitative MRN and to present T2 values of the distal sciatic nerve in healthy subjects. Our results suggest that T2 mapping derived from a 3D T2-prepared TSE sequence is highly reliable and, importantly, shows lower values than a commonly used 2D MESE sequence whilst being in agreement with the nerve water T2 from MRS. The influence of residual fat on the T2 quantification that is not entirely suppressed in a commonly used 2D MESE sequence may account for this discrepancy and should be considered during T2 mapping of the PNS. Furthermore, T2 mapping using a 3D T2-prepared TSE sequence might allow for evaluations of quantitative alterations of the distal sciatic nerve in patients with unilateral sciatica due to degenerative disc disease with unilateral nerve contact of degenerated disc material, which may allow T2 values of the distal sciatic nerve to become biomarkers of direct nerve affection in patients with degenerative changes at the level of the LSP.

Acknowledgement

Open Access funding provided by Projekt DEAL.

Compliance with ethical standards

Conflict of interest

Jan S. Kirschke received speaker honoraria from Philips Healthcare. Dimitrios C. Karampinos received grant support from Philips Healthcare. Nico Sollmann, Dominik Weidlich, Elisabeth Klupp, Barbara Cervantes, Carl Ganter, Claus Zimmer, Ernst J. Rummeny, and Thomas Baum declare that they have no conflict of interest.

Ethical approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
Informed consent was obtained from all subjects included in the study.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

© Springer Medizin

Bis 11. April 2024 bestellen und im ersten Jahr 50 % sparen!

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

© Springer Medizin

Bis 11. April 2024 bestellen und im ersten Jahr 50 % sparen!

Literatur
1.
Zurück zum Zitat Muniz Neto FJ, Kihara Filho EN, Miranda FC, Rosemberg LA, Santos DCB, Taneja AK (2018) Demystifying MR neurography of the lumbosacral plexus: from protocols to pathologies. Biomed Res Int 2018:9608947CrossRef Muniz Neto FJ, Kihara Filho EN, Miranda FC, Rosemberg LA, Santos DCB, Taneja AK (2018) Demystifying MR neurography of the lumbosacral plexus: from protocols to pathologies. Biomed Res Int 2018:9608947CrossRef
2.
Zurück zum Zitat Chhabra A, Madhuranthakam AJ, Andreisek G (2018) Magnetic resonance neurography: current perspectives and literature review. Eur Radiol 28(2):698–707CrossRef Chhabra A, Madhuranthakam AJ, Andreisek G (2018) Magnetic resonance neurography: current perspectives and literature review. Eur Radiol 28(2):698–707CrossRef
3.
Zurück zum Zitat Chhabra A, Andreisek G, Soldatos T, Wang KC, Flammang AJ, Belzberg AJ, Carrino JA (2011) MR neurography: past, present, and future. AJR Am J Roentgenol 197(3):583–591CrossRef Chhabra A, Andreisek G, Soldatos T, Wang KC, Flammang AJ, Belzberg AJ, Carrino JA (2011) MR neurography: past, present, and future. AJR Am J Roentgenol 197(3):583–591CrossRef
4.
Zurück zum Zitat Soldatos T, Andreisek G, Thawait GK, Guggenberger R, Williams EH, Carrino JA, Chhabra A (2013) High-resolution 3-T MR neurography of the lumbosacral plexus. Radiographics 33(4):967–987CrossRef Soldatos T, Andreisek G, Thawait GK, Guggenberger R, Williams EH, Carrino JA, Chhabra A (2013) High-resolution 3-T MR neurography of the lumbosacral plexus. Radiographics 33(4):967–987CrossRef
5.
Zurück zum Zitat Bucknor MD, Steinbach LS, Saloner D, Chin CT (2014) Magnetic resonance neurography evaluation of chronic extraspinal sciatica after remote proximal hamstring injury: a preliminary retrospective analysis. J Neurosurg 121(2):408–414CrossRef Bucknor MD, Steinbach LS, Saloner D, Chin CT (2014) Magnetic resonance neurography evaluation of chronic extraspinal sciatica after remote proximal hamstring injury: a preliminary retrospective analysis. J Neurosurg 121(2):408–414CrossRef
6.
Zurück zum Zitat Pham M, Wessig C, Brinkhoff J, Reiners K, Stoll G, Bendszus M (2011) MR neurography of sciatic nerve injection injury. J Neurol 258(6):1120–1125CrossRef Pham M, Wessig C, Brinkhoff J, Reiners K, Stoll G, Bendszus M (2011) MR neurography of sciatic nerve injection injury. J Neurol 258(6):1120–1125CrossRef
7.
Zurück zum Zitat Wolf M, Baumer P, Pedro M, Dombert T, Staub F, Heiland S, Bendszus M, Pham M (2014) Sciatic nerve injury related to hip replacement surgery: imaging detection by MR neurography despite susceptibility artifacts. PLoS ONE 9(2):e89154CrossRef Wolf M, Baumer P, Pedro M, Dombert T, Staub F, Heiland S, Bendszus M, Pham M (2014) Sciatic nerve injury related to hip replacement surgery: imaging detection by MR neurography despite susceptibility artifacts. PLoS ONE 9(2):e89154CrossRef
8.
Zurück zum Zitat Leon Cejas L, Binaghi D, Socolovsky M, Dubrovsky A, Pirra L, Marchesoni C, Pardal A, Monges S, Peretti G, Taratuto AL, Lubinieki F, Reisin R (2018) Intraneural perineuriomas: diagnostic value of magnetic resonance neurography. J Peripher Nerv Syst 23(1):23–28CrossRef Leon Cejas L, Binaghi D, Socolovsky M, Dubrovsky A, Pirra L, Marchesoni C, Pardal A, Monges S, Peretti G, Taratuto AL, Lubinieki F, Reisin R (2018) Intraneural perineuriomas: diagnostic value of magnetic resonance neurography. J Peripher Nerv Syst 23(1):23–28CrossRef
9.
Zurück zum Zitat Godel T, Baumer P, Pham M, Kohn A, Muschol N, Kronlage M, Kollmer J, Heiland S, Bendszus M, Mautner VF (2017) Human dorsal root ganglion in vivo morphometry and perfusion in Fabry painful neuropathy. Neurology 89(12):1274–1282CrossRef Godel T, Baumer P, Pham M, Kohn A, Muschol N, Kronlage M, Kollmer J, Heiland S, Bendszus M, Mautner VF (2017) Human dorsal root ganglion in vivo morphometry and perfusion in Fabry painful neuropathy. Neurology 89(12):1274–1282CrossRef
11.
Zurück zum Zitat Godel T, Kohn A, Muschol N, Kronlage M, Schwarz D, Kollmer J, Heiland S, Bendszus M, Mautner VF, Baumer P (2018) Dorsal root ganglia in vivo morphometry and perfusion in female patients with Fabry disease. J Neurol 265(11):2723–2729CrossRef Godel T, Kohn A, Muschol N, Kronlage M, Schwarz D, Kollmer J, Heiland S, Bendszus M, Mautner VF, Baumer P (2018) Dorsal root ganglia in vivo morphometry and perfusion in female patients with Fabry disease. J Neurol 265(11):2723–2729CrossRef
12.
Zurück zum Zitat Chhabra A, Carrino JA, Farahani SJ, Thawait GK, Sumner CJ, Wadhwa V, Chaudhary V, Lloyd TE (2016) Whole-body MR neurography: prospective feasibility study in polyneuropathy and Charcot–Marie–Tooth disease. J Magn Reson Imaging 44(6):1513–1521CrossRef Chhabra A, Carrino JA, Farahani SJ, Thawait GK, Sumner CJ, Wadhwa V, Chaudhary V, Lloyd TE (2016) Whole-body MR neurography: prospective feasibility study in polyneuropathy and Charcot–Marie–Tooth disease. J Magn Reson Imaging 44(6):1513–1521CrossRef
13.
Zurück zum Zitat Jende JME, Groener JB, Oikonomou D, Heiland S, Kopf S, Pham M, Nawroth P, Bendszus M, Kurz FT (2018) Diabetic neuropathy differs between type 1 and type 2 diabetes: insights from magnetic resonance neurography. Ann Neurol 83(3):588–598CrossRef Jende JME, Groener JB, Oikonomou D, Heiland S, Kopf S, Pham M, Nawroth P, Bendszus M, Kurz FT (2018) Diabetic neuropathy differs between type 1 and type 2 diabetes: insights from magnetic resonance neurography. Ann Neurol 83(3):588–598CrossRef
14.
Zurück zum Zitat Chhabra A, Chalian M, Soldatos T, Andreisek G, Faridian-Aragh N, Williams E, Belzberg AJ, Carrino JA (2012) 3-T high-resolution MR neurography of sciatic neuropathy. AJR Am J Roentgenol 198(4):W357–364CrossRef Chhabra A, Chalian M, Soldatos T, Andreisek G, Faridian-Aragh N, Williams E, Belzberg AJ, Carrino JA (2012) 3-T high-resolution MR neurography of sciatic neuropathy. AJR Am J Roentgenol 198(4):W357–364CrossRef
15.
Zurück zum Zitat Keller S, Wang ZJ, Golsari A, Kim AC, Kooijman H, Adam G, Yamamura J (2018) Feasibility of peripheral nerve MR neurography using diffusion tensor imaging adapted to skeletal muscle disease. Acta Radiol 59(5):560–568CrossRef Keller S, Wang ZJ, Golsari A, Kim AC, Kooijman H, Adam G, Yamamura J (2018) Feasibility of peripheral nerve MR neurography using diffusion tensor imaging adapted to skeletal muscle disease. Acta Radiol 59(5):560–568CrossRef
16.
Zurück zum Zitat Chhabra A, Farahani SJ, Thawait GK, Wadhwa V, Belzberg AJ, Carrino JA (2016) Incremental value of magnetic resonance neurography of lumbosacral plexus over non-contributory lumbar spine magnetic resonance imaging in radiculopathy: a prospective study. World J Radiol 8(1):109–116CrossRef Chhabra A, Farahani SJ, Thawait GK, Wadhwa V, Belzberg AJ, Carrino JA (2016) Incremental value of magnetic resonance neurography of lumbosacral plexus over non-contributory lumbar spine magnetic resonance imaging in radiculopathy: a prospective study. World J Radiol 8(1):109–116CrossRef
17.
Zurück zum Zitat Kronlage M, Pitarokoili K, Schwarz D, Godel T, Heiland S, Yoon MS, Bendszus M, Baumer P (2017) Diffusion tensor imaging in chronic inflammatory demyelinating polyneuropathy: diagnostic accuracy and correlation with electrophysiology. Invest Radiol 52(11):701–707CrossRef Kronlage M, Pitarokoili K, Schwarz D, Godel T, Heiland S, Yoon MS, Bendszus M, Baumer P (2017) Diffusion tensor imaging in chronic inflammatory demyelinating polyneuropathy: diagnostic accuracy and correlation with electrophysiology. Invest Radiol 52(11):701–707CrossRef
18.
Zurück zum Zitat Vaeggemose M, Vaeth S, Pham M, Ringgaard S, Jensen UB, Tankisi H, Ejskjaer N, Heiland S, Andersen H (2017) Magnetic resonance neurography and diffusion tensor imaging of the peripheral nerves in patients with Charcot–Marie–Tooth Type 1A. Muscle Nerve 56(6):E78–E84CrossRef Vaeggemose M, Vaeth S, Pham M, Ringgaard S, Jensen UB, Tankisi H, Ejskjaer N, Heiland S, Andersen H (2017) Magnetic resonance neurography and diffusion tensor imaging of the peripheral nerves in patients with Charcot–Marie–Tooth Type 1A. Muscle Nerve 56(6):E78–E84CrossRef
19.
Zurück zum Zitat Apostolidis L, Schwarz D, Xia A, Weiler M, Heckel A, Godel T, Heiland S, Schlemmer HP, Jager D, Bendszus M, Baumer P (2017) Dorsal root ganglia hypertrophy as in vivo correlate of oxaliplatin-induced polyneuropathy. PLoS ONE 12(8):e0183845CrossRef Apostolidis L, Schwarz D, Xia A, Weiler M, Heckel A, Godel T, Heiland S, Schlemmer HP, Jager D, Bendszus M, Baumer P (2017) Dorsal root ganglia hypertrophy as in vivo correlate of oxaliplatin-induced polyneuropathy. PLoS ONE 12(8):e0183845CrossRef
20.
Zurück zum Zitat Vaeggemose M, Pham M, Ringgaard S, Tankisi H, Ejskjaer N, Heiland S, Poulsen PL, Andersen H (2017) Diffusion tensor imaging MR neurography for the detection of polyneuropathy in type 1 diabetes. J Magn Reson Imaging 45(4):1125–1134CrossRef Vaeggemose M, Pham M, Ringgaard S, Tankisi H, Ejskjaer N, Heiland S, Poulsen PL, Andersen H (2017) Diffusion tensor imaging MR neurography for the detection of polyneuropathy in type 1 diabetes. J Magn Reson Imaging 45(4):1125–1134CrossRef
21.
Zurück zum Zitat Vaeggemose M, Pham M, Ringgaard S, Tankisi H, Ejskjaer N, Heiland S, Poulsen PL, Andersen H (2017) Magnetic resonance neurography visualizes abnormalities in sciatic and tibial nerves in patients with type 1 diabetes and neuropathy. Diabetes 66(7):1779–1788CrossRef Vaeggemose M, Pham M, Ringgaard S, Tankisi H, Ejskjaer N, Heiland S, Poulsen PL, Andersen H (2017) Magnetic resonance neurography visualizes abnormalities in sciatic and tibial nerves in patients with type 1 diabetes and neuropathy. Diabetes 66(7):1779–1788CrossRef
22.
Zurück zum Zitat Dortch RD, Dethrage LM, Gore JC, Smith SA, Li J (2014) Proximal nerve magnetization transfer MRI relates to disability in Charcot–Marie–Tooth diseases. Neurology 83(17):1545–1553CrossRef Dortch RD, Dethrage LM, Gore JC, Smith SA, Li J (2014) Proximal nerve magnetization transfer MRI relates to disability in Charcot–Marie–Tooth diseases. Neurology 83(17):1545–1553CrossRef
23.
Zurück zum Zitat Kronlage M, Baumer P, Pitarokoili K, Schwarz D, Schwehr V, Godel T, Heiland S, Gold R, Bendszus M, Yoon MS (2017) Large coverage MR neurography in CIDP: diagnostic accuracy and electrophysiological correlation. J Neurol 264(7):1434–1443CrossRef Kronlage M, Baumer P, Pitarokoili K, Schwarz D, Schwehr V, Godel T, Heiland S, Gold R, Bendszus M, Yoon MS (2017) Large coverage MR neurography in CIDP: diagnostic accuracy and electrophysiological correlation. J Neurol 264(7):1434–1443CrossRef
24.
Zurück zum Zitat Jende JME, Hauck GH, Diem R, Weiler M, Heiland S, Wildemann B, Korporal-Kuhnke M, Wick W, Hayes JM, Pfaff J, Pham M, Bendszus M, Kollmer J (2017) Peripheral nerve involvement in multiple sclerosis: Demonstration by magnetic resonance neurography. Ann Neurol 82(5):676–685CrossRef Jende JME, Hauck GH, Diem R, Weiler M, Heiland S, Wildemann B, Korporal-Kuhnke M, Wick W, Hayes JM, Pfaff J, Pham M, Bendszus M, Kollmer J (2017) Peripheral nerve involvement in multiple sclerosis: Demonstration by magnetic resonance neurography. Ann Neurol 82(5):676–685CrossRef
25.
Zurück zum Zitat Kollmer J, Hund E, Hornung B, Hegenbart U, Schonland SO, Kimmich C, Kristen AV, Purrucker J, Rocken C, Heiland S, Bendszus M, Pham M (2015) In vivo detection of nerve injury in familial amyloid polyneuropathy by magnetic resonance neurography. Brain 138(Pt 3):549–562CrossRef Kollmer J, Hund E, Hornung B, Hegenbart U, Schonland SO, Kimmich C, Kristen AV, Purrucker J, Rocken C, Heiland S, Bendszus M, Pham M (2015) In vivo detection of nerve injury in familial amyloid polyneuropathy by magnetic resonance neurography. Brain 138(Pt 3):549–562CrossRef
26.
Zurück zum Zitat Webb S, Munro CA, Midha R, Stanisz GJ (2003) Is multicomponent T2 a good measure of myelin content in peripheral nerve? Magn Reson Med 49(4):638–645CrossRef Webb S, Munro CA, Midha R, Stanisz GJ (2003) Is multicomponent T2 a good measure of myelin content in peripheral nerve? Magn Reson Med 49(4):638–645CrossRef
27.
Zurück zum Zitat Gambarota G, Mekle R, Mlynarik V, Krueger G (2009) NMR properties of human median nerve at 3 T: proton density, T1, T2, and magnetization transfer. J Magn Reson Imaging 29(4):982–986CrossRef Gambarota G, Mekle R, Mlynarik V, Krueger G (2009) NMR properties of human median nerve at 3 T: proton density, T1, T2, and magnetization transfer. J Magn Reson Imaging 29(4):982–986CrossRef
28.
Zurück zum Zitat Does MD, Snyder RE (1995) T2 relaxation of peripheral nerve measured in vivo. Magn Reson Imaging 13(4):575–580CrossRef Does MD, Snyder RE (1995) T2 relaxation of peripheral nerve measured in vivo. Magn Reson Imaging 13(4):575–580CrossRef
29.
Zurück zum Zitat Konstantinou K, Dunn KM (2008) Sciatica: review of epidemiological studies and prevalence estimates. Spine 33(22):2464–2472CrossRef Konstantinou K, Dunn KM (2008) Sciatica: review of epidemiological studies and prevalence estimates. Spine 33(22):2464–2472CrossRef
30.
Zurück zum Zitat Van Boxem K, Cheng J, Patijn J, van Kleef M, Lataster A, Mekhail N, Van Zundert J (2010) 11. Lumbosacral radicular pain. Pain Pract 10(4):339–358CrossRef Van Boxem K, Cheng J, Patijn J, van Kleef M, Lataster A, Mekhail N, Van Zundert J (2010) 11. Lumbosacral radicular pain. Pain Pract 10(4):339–358CrossRef
32.
Zurück zum Zitat Fairbank JC, Pynsent PB (2000) The Oswestry disability index. Spine 25(22):2940–2952 (discussion 2952)CrossRef Fairbank JC, Pynsent PB (2000) The Oswestry disability index. Spine 25(22):2940–2952 (discussion 2952)CrossRef
33.
Zurück zum Zitat Fairbank J (1995) Use of Oswestry disability index (ODI). Spine 20(13):1535–1537CrossRef Fairbank J (1995) Use of Oswestry disability index (ODI). Spine 20(13):1535–1537CrossRef
34.
Zurück zum Zitat Langley GB, Sheppeard H (1985) The visual analogue scale: its use in pain measurement. Rheumatol Int 5(4):145–148CrossRef Langley GB, Sheppeard H (1985) The visual analogue scale: its use in pain measurement. Rheumatol Int 5(4):145–148CrossRef
35.
Zurück zum Zitat Price DD, McGrath PA, Rafii A, Buckingham B (1983) The validation of visual analogue scales as ratio scale measures for chronic and experimental pain. Pain 17(1):45–56CrossRef Price DD, McGrath PA, Rafii A, Buckingham B (1983) The validation of visual analogue scales as ratio scale measures for chronic and experimental pain. Pain 17(1):45–56CrossRef
36.
Zurück zum Zitat Weidlich D, Schlaeger S, Kooijman H, Bornert P, Kirschke JS, Rummeny EJ, Haase A, Karampinos DC (2017) T2 mapping with magnetization-prepared 3D TSE based on a modified BIR-4 T2 preparation. NMR Biomed 30(11). Weidlich D, Schlaeger S, Kooijman H, Bornert P, Kirschke JS, Rummeny EJ, Haase A, Karampinos DC (2017) T2 mapping with magnetization-prepared 3D TSE based on a modified BIR-4 T2 preparation. NMR Biomed 30(11).
37.
Zurück zum Zitat Klupp E, Weidlich D, Schlaeger S, Baum T, Cervantes B, Deschauer M, Kooijman H, Rummeny EJ, Zimmer C, Kirschke JS, Karampinos DC (2017) B1-insensitive T2 mapping of healthy thigh muscles using a T2-prepared 3D TSE sequence. PLoS ONE 12(2):e0171337CrossRef Klupp E, Weidlich D, Schlaeger S, Baum T, Cervantes B, Deschauer M, Kooijman H, Rummeny EJ, Zimmer C, Kirschke JS, Karampinos DC (2017) B1-insensitive T2 mapping of healthy thigh muscles using a T2-prepared 3D TSE sequence. PLoS ONE 12(2):e0171337CrossRef
38.
Zurück zum Zitat Golub GH, Pereyra V (1973) Differentiation of pseudo-inverses and nonlinear least-squares problems whose variables separate. SIAM J Numer Anal 10:413–432CrossRef Golub GH, Pereyra V (1973) Differentiation of pseudo-inverses and nonlinear least-squares problems whose variables separate. SIAM J Numer Anal 10:413–432CrossRef
39.
Zurück zum Zitat Kiefer J (1953) Sequential minimax search for a maximum. Proc Am Math Soc 4:502–506CrossRef Kiefer J (1953) Sequential minimax search for a maximum. Proc Am Math Soc 4:502–506CrossRef
42.
Zurück zum Zitat Gluer CC, Blake G, Lu Y, Blunt BA, Jergas M, Genant HK (1995) Accurate assessment of precision errors: how to measure the reproducibility of bone densitometry techniques. Osteoporos Int 5(4):262–270CrossRef Gluer CC, Blake G, Lu Y, Blunt BA, Jergas M, Genant HK (1995) Accurate assessment of precision errors: how to measure the reproducibility of bone densitometry techniques. Osteoporos Int 5(4):262–270CrossRef
43.
Zurück zum Zitat Schlaeger S, Weidlich D, Klupp E, Montagnese F, Deschauer M, Schoser B, Bublitz S, Zimmer C, Rummeny EJ, Kirschke JS, Karampinos DC (2018) Water T2 mapping in fatty infiltrated thigh muscles of patients with neuromuscular diseases using a T2-prepared 3D TSE with SPAIR. In: Proceedings 26. annual meeting international society for magnetic resonance in medicine (ISMRM). Schlaeger S, Weidlich D, Klupp E, Montagnese F, Deschauer M, Schoser B, Bublitz S, Zimmer C, Rummeny EJ, Kirschke JS, Karampinos DC (2018) Water T2 mapping in fatty infiltrated thigh muscles of patients with neuromuscular diseases using a T2-prepared 3D TSE with SPAIR. In: Proceedings 26. annual meeting international society for magnetic resonance in medicine (ISMRM).
44.
Zurück zum Zitat Wang D, Wang C, Duan X, Yang Z, Bai Z, Hu H, Yan L, Shen J (2018) MR T2 value of the tibial nerve can be used as a potential non-invasive and quantitative biomarker for the diagnosis of diabetic peripheral neuropathy. Eur Radiol 28(3):1234–1241CrossRef Wang D, Wang C, Duan X, Yang Z, Bai Z, Hu H, Yan L, Shen J (2018) MR T2 value of the tibial nerve can be used as a potential non-invasive and quantitative biomarker for the diagnosis of diabetic peripheral neuropathy. Eur Radiol 28(3):1234–1241CrossRef
45.
Zurück zum Zitat Cha JG, Han JK, Im SB, Kang SJ (2014) Median nerve T2 assessment in the wrist joints: preliminary study in patients with carpal tunnel syndrome and healthy volunteers. J Magn Reson Imaging 40(4):789–795CrossRef Cha JG, Han JK, Im SB, Kang SJ (2014) Median nerve T2 assessment in the wrist joints: preliminary study in patients with carpal tunnel syndrome and healthy volunteers. J Magn Reson Imaging 40(4):789–795CrossRef
Metadaten
Titel
T2 mapping of the distal sciatic nerve in healthy subjects and patients suffering from lumbar disc herniation with nerve compression
verfasst von
Nico Sollmann
Dominik Weidlich
Elisabeth Klupp
Barbara Cervantes
Carl Ganter
Claus Zimmer
Ernst J. Rummeny
Thomas Baum
Jan S. Kirschke
Dimitrios C. Karampinos
Publikationsdatum
11.02.2020
Verlag
Springer International Publishing
Erschienen in
Magnetic Resonance Materials in Physics, Biology and Medicine / Ausgabe 5/2020
Print ISSN: 0968-5243
Elektronische ISSN: 1352-8661