13.01.2017 | Original Article

Technical characterization of dialysis fluid flow and mass transfer rate in dialyzers with various filtration coefficients using dimensionless correlation equation

verfasst von: Makoto Fukuda, Kengo Yoshimura, Koki Namekawa, Kiyotaka Sakai

Erschienen in: Journal of Artificial Organs | Ausgabe 2/2017

Einloggen, um Zugang zu erhalten

Abstract

The objective of the present study is to evaluate the effect of filtration coefficient and internal filtration on dialysis fluid flow and mass transfer coefficient in dialyzers using dimensionless mass transfer correlation equations. Aqueous solution of vitamin B12 clearances were obtained for REXEED-15L as a low flux dialyzer, and APS-15EA and APS-15UA as high flux dialyzers. All the other design specifications were identical for these dialyzers except for filtration coefficient. The overall mass transfer coefficient was calculated, moreover, the exponents of Reynolds number (Re) and film mass transfer coefficient of the dialysis-side fluid (k D) for each flow rate were derived from the Wilson plot and dimensionless correlation equation. The exponents of Re were 0.4 for the low flux dialyzer whereas 0.5 for the high flux dialyzers. Dialysis fluid of the low flux dialyzer was close to laminar flow because of its low filtration coefficient. On the other hand, dialysis fluid of the high flux dialyzers was assumed to be orthogonal flow. Higher filtration coefficient was associated with higher k D influenced by mass transfer rate through diffusion and internal filtration. Higher filtration coefficient of dialyzers and internal filtration affect orthogonal flow of dialysis fluid.
Literatur
1.
Zurück zum Zitat Annual Report, Fresenius Medical Care, 2015;1–256. Annual Report, Fresenius Medical Care, 2015;1–256.
2.
Zurück zum Zitat Saito A. Definition of high-Performance membranes—from the clinical point of view. Contributions to nephrology. Karger. 2011;173:1–10. Saito A. Definition of high-Performance membranes—from the clinical point of view. Contributions to nephrology. Karger. 2011;173:1–10.
3.
Zurück zum Zitat Yamashita AC. Commercially available high-performance membrane dialyzers and several special kinds on the Japanese market. (as of April 30, 2011). Contributions to nephrology. Karger. 2011;173:191–5. Yamashita AC. Commercially available high-performance membrane dialyzers and several special kinds on the Japanese market. (as of April 30, 2011). Contributions to nephrology. Karger. 2011;173:191–5.
4.
Zurück zum Zitat Yamamoto K, Matsukawa H, Yakushiji T, Fukuda M, Hiyoshi T, Sakai K. Technical evaluation of dialysate flow in a newly designed dialyzer. ASAIO J. 2007;53:36–40.CrossRefPubMed Yamamoto K, Matsukawa H, Yakushiji T, Fukuda M, Hiyoshi T, Sakai K. Technical evaluation of dialysate flow in a newly designed dialyzer. ASAIO J. 2007;53:36–40.CrossRefPubMed
5.
Zurück zum Zitat Fukuda M, Hosoya N, Kanamori T, Sakai K, Nishikido J, Watanabe T, et al. Determination of optimal fiber density of convection and high performance dialyzers. Artif Organs Today. 1992;2:205–14. Fukuda M, Hosoya N, Kanamori T, Sakai K, Nishikido J, Watanabe T, et al. Determination of optimal fiber density of convection and high performance dialyzers. Artif Organs Today. 1992;2:205–14.
6.
Zurück zum Zitat Fukuda M, Miyazaki M, Uezumi S, Yoshida M. Design and assessment of the new APS dialyzer (APS-SA series). J Artif Organs. 2006;9:192–8.CrossRefPubMed Fukuda M, Miyazaki M, Uezumi S, Yoshida M. Design and assessment of the new APS dialyzer (APS-SA series). J Artif Organs. 2006;9:192–8.CrossRefPubMed
7.
Zurück zum Zitat Kunikata S, Fukuda M, Yamamoto K, Yagi Y, Matsuda M, Sakai K. Technical characterization of dialysis fluid flow of newly-developed dialyzers using mass transfer correlation equations. ASAIO J. 2009;55:231–5.CrossRefPubMed Kunikata S, Fukuda M, Yamamoto K, Yagi Y, Matsuda M, Sakai K. Technical characterization of dialysis fluid flow of newly-developed dialyzers using mass transfer correlation equations. ASAIO J. 2009;55:231–5.CrossRefPubMed
8.
Zurück zum Zitat Yamamoto K, Matsuda M, Hirano A, Takizawa N, Iwashima S, Yakushiji T, Fukuda M, Miyasaka T, Sakai K. Computational evaluation of dialysis fluid flow in dialyzers with variously designed jackets. Artif Organs. 2009;33:481–6.CrossRefPubMed Yamamoto K, Matsuda M, Hirano A, Takizawa N, Iwashima S, Yakushiji T, Fukuda M, Miyasaka T, Sakai K. Computational evaluation of dialysis fluid flow in dialyzers with variously designed jackets. Artif Organs. 2009;33:481–6.CrossRefPubMed
9.
Zurück zum Zitat Takesawa S, Terasawa M, Sakagami M, Kobayashi T, Hidai H, Sakai K. Nondestructive evaluation by X-ray computed tomography of dialysate flow pattern in capillary dialyzers. ASAIO Trans. 1988;34:794–9.PubMed Takesawa S, Terasawa M, Sakagami M, Kobayashi T, Hidai H, Sakai K. Nondestructive evaluation by X-ray computed tomography of dialysate flow pattern in capillary dialyzers. ASAIO Trans. 1988;34:794–9.PubMed
10.
Zurück zum Zitat Poh CK, Hardy PA, Liao Z, Huang Z, Clark WR, Gao D. Effect of spacer yarns on the dialysate flow distribution of hemodialyzers: a magnetic resonance imaging study. ASAIO J. 2003;49:440–8.PubMed Poh CK, Hardy PA, Liao Z, Huang Z, Clark WR, Gao D. Effect of spacer yarns on the dialysate flow distribution of hemodialyzers: a magnetic resonance imaging study. ASAIO J. 2003;49:440–8.PubMed
11.
Zurück zum Zitat Colburn AP. A method of correlating forced convection heat transfer data and a comparison with fluid friction. Trans Am Inst Chem Eng. 1933;29:174–210. Colburn AP. A method of correlating forced convection heat transfer data and a comparison with fluid friction. Trans Am Inst Chem Eng. 1933;29:174–210.
12.
Zurück zum Zitat Wickramasinghe SR, Semmens MJ, Cussler EL. Hollow fiber modules made with hollow fiber fabric. J Membr Sci. 1993;84:1–14.CrossRef Wickramasinghe SR, Semmens MJ, Cussler EL. Hollow fiber modules made with hollow fiber fabric. J Membr Sci. 1993;84:1–14.CrossRef
13.
Zurück zum Zitat Wickramasinghe SR, Semmens MJ, Cussler EL. Mass transfer in various hollow fiber geometries. J Membr Sci. 1992;69:235–50.CrossRef Wickramasinghe SR, Semmens MJ, Cussler EL. Mass transfer in various hollow fiber geometries. J Membr Sci. 1992;69:235–50.CrossRef
14.
Zurück zum Zitat Fukuda M, Namekawa K, Sakai K. Identical dependence of dialysate-side mass transfer coefficient on Reynolds number using dimensionless correlation based on the mass transfer model in newly developed dialyzers and a downsized dialyzer. Adv Biomed Eng. 2016;5:118–23.CrossRef Fukuda M, Namekawa K, Sakai K. Identical dependence of dialysate-side mass transfer coefficient on Reynolds number using dimensionless correlation based on the mass transfer model in newly developed dialyzers and a downsized dialyzer. Adv Biomed Eng. 2016;5:118–23.CrossRef
15.
Zurück zum Zitat Kanamori T, Mizoguchi K. Solute-removal enhancement caused by local convective flow in a hemodialyzer. J Artif Organs. 2012;15:305–10.CrossRefPubMed Kanamori T, Mizoguchi K. Solute-removal enhancement caused by local convective flow in a hemodialyzer. J Artif Organs. 2012;15:305–10.CrossRefPubMed
16.
Zurück zum Zitat Sato T, Saito A, Suzuki S, et al. Guidelines for performance evaluation and category of blood purification devices (in Japanese). J Jpn Soc Dial Ther. 1996;29:1231–45.CrossRef Sato T, Saito A, Suzuki S, et al. Guidelines for performance evaluation and category of blood purification devices (in Japanese). J Jpn Soc Dial Ther. 1996;29:1231–45.CrossRef
17.
Zurück zum Zitat Sakai K. Determination of pore size and pore size distribution. 2. Dialysis membranes. J Membr Sci. 1994;96:91–130.CrossRef Sakai K. Determination of pore size and pore size distribution. 2. Dialysis membranes. J Membr Sci. 1994;96:91–130.CrossRef
18.
Zurück zum Zitat Morti S, Shao J, Zydney AL. Importance of asymmetric structure in determining mass transport characteristics of hollow fiber hemodialyzers. J Membr Sci. 2003;224:39–49.CrossRef Morti S, Shao J, Zydney AL. Importance of asymmetric structure in determining mass transport characteristics of hollow fiber hemodialyzers. J Membr Sci. 2003;224:39–49.CrossRef
19.
Zurück zum Zitat Wilson EE. A basis for the rational design of heat transfer apparatus. Trans Am Soc Mech Eng. 1915;37:47–82. Wilson EE. A basis for the rational design of heat transfer apparatus. Trans Am Soc Mech Eng. 1915;37:47–82.
20.
Zurück zum Zitat Leypoldt JK, Kamerath CD, Gilson JF, Friederichs G. Dialyzer clearances and mass transfer-area coefficients for small solutes at low dialysate flow rates. ASAIO J. 2006;404–9. Leypoldt JK, Kamerath CD, Gilson JF, Friederichs G. Dialyzer clearances and mass transfer-area coefficients for small solutes at low dialysate flow rates. ASAIO J. 2006;404–9.
21.
Zurück zum Zitat Catapano G, Hornscheidt R, Wodetzki A, Baurmeister U. Turbulent flow technique for the estimation of oxygen diffusive permeability of membranes for the oxygenation of blood and other cell suspensions. J Membr Sci. 2004;230:131–9.CrossRef Catapano G, Hornscheidt R, Wodetzki A, Baurmeister U. Turbulent flow technique for the estimation of oxygen diffusive permeability of membranes for the oxygenation of blood and other cell suspensions. J Membr Sci. 2004;230:131–9.CrossRef
22.
Zurück zum Zitat Mineshima M, Ishimori I, Sakiyama R Validity of internal filtration-enhanced hemodialysis as a new hemodiafiltration therapy. Blood Purif. 2009;27:33–7.CrossRefPubMed Mineshima M, Ishimori I, Sakiyama R Validity of internal filtration-enhanced hemodialysis as a new hemodiafiltration therapy. Blood Purif. 2009;27:33–7.CrossRefPubMed
23.
Zurück zum Zitat Sekino M. Effect of ultrafiltration in hemodiafiltration system. Kagakukougaku Ronbunsyu. 2012;38:155–60 (in Japanese). Sekino M. Effect of ultrafiltration in hemodiafiltration system. Kagakukougaku Ronbunsyu. 2012;38:155–60 (in Japanese).
Metadaten
Titel
Technical characterization of dialysis fluid flow and mass transfer rate in dialyzers with various filtration coefficients using dimensionless correlation equation
verfasst von
Makoto Fukuda
Kengo Yoshimura
Koki Namekawa
Kiyotaka Sakai
Publikationsdatum
13.01.2017
Verlag
Springer Japan
Erschienen in
Journal of Artificial Organs / Ausgabe 2/2017
Print ISSN: 1434-7229
Elektronische ISSN: 1619-0904
DOI
https://doi.org/10.1007/s10047-016-0942-5

Weitere Artikel der Ausgabe 2/2017

Journal of Artificial Organs 2/2017 Zur Ausgabe

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.