Techniques for the insertion of the proseal laryngeal mask airway: comparison of the foley airway stylet tool with the introducer tool in a prospective, randomized study

verfasst von: Mao-Kai Chen, Hung-Te Hsu, I-Cheng Lu, Chih-Kai Shih, Ya-Chun Shen, Kuang-Yi Tseng, Kuang-I Cheng

Erschienen in: BMC Anesthesiology | Ausgabe 1/2014

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

Background

Many tools have been developed to facilitate the insertion of the ProSeal laryngeal mask airway (LMA) insertion, which can be impeded by folding of its soft cuff. The aim of this study was to compare the efficiency of ProSeal LMA insertion guided by a soft, direct optical Foley Airway Stylet Tool (FAST) with the standard introducer tool (IT).

Methods

One hundred sixty patients undergoing general anesthesia using the ProSeal LMA as an airway management device were randomly allocated to either FAST-guided or IT-assisted groups. Following ProSeal LMA insertion, the glottic and esophageal openings were identified using a fiberoptic bronchoscope introduced through the airway and the drain tube. The primary outcomes were time taken to insert the ProSeal LMA and the success rate at the first attempt. Secondary end points included ease of insertion, hemodynamic response to insertion, and postoperative adverse events recorded in the recovery room and on the first postoperative morning.

Results

One hundred forty patients were included in the final analysis: 66 in the FAST-guided group and 74 in the IT-assisted group. The success rate of FAST device-guided ProSeal LMA insertion (95.7%) was broadly comparable with IT-assisted insertion (98.7%). However, the time taken to insert the ProSeal LMA was significantly longer when the FAST technique was used (p <0.001). The incidence of correct alignment of the airway tube and the drain tube did not differ significantly between the groups. There were no significant differences in ease of insertion or hemodynamic responses to insertion, except that the incidence of postoperative sore throat was significantly higher in the FAST group on the first postoperative day (22.2% compared with 6.8% in the IT group; p =0.035).

Conclusion

Both FAST-guided and IT-assisted techniques achieved correct ProSeal LMA positioning, but the IT technique was significantly quicker and less likely to cause a sore throat.

Trial registration

ClinicalTrials.gov Identifier: NCT02048657
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1471-2253-14-105) contains supplementary material, which is available to authorized users.

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

MKC: study design, data collection and writing-up draft of the manuscript. HTH: study design and conception provided. ICL: study design, editing text and tables of the manuscript. CKS: data acquisition and analysis and interpretation of data. KYT: study design, data collection, and coordination. YCS: data acquisition and primary statistical analysis. KIC: conception and design, revised and editing the manuscript in critically for important intellectual content, and final approval of the version to be published. All authors have read and approved the final version.
Abkürzungen
LMA
Laryngeal mask airway
FAST
Foley airway stylet tool
IT
Introducer tool
ETCO2
End tidal CO2.

Background

The ProSeal laryngeal mask airway (LMA; Orthofix, Maidenhead, UK) is a laryngeal mask with a modified cuff that incorporates a drainage tube to improve the quality of the seal while reducing the risk of pulmonary aspiration and gastric insufflation [13]. When the device is used for controlled ventilation, the ProSeal LMA provides a higher oropharyngeal leak pressure than the Classic LMA [4]. However, the ProSeal LMA is reported to be more difficult to insert than the Classic LMA, as its larger, softer cuff is prone to folding. The manufacturer recommends that the ProSeal LMA be inserted using either manipulation with the fingers or a curved metal introducer. Nonetheless, first attempt success rates of ProSeal LMA insertion range from 81% to 87%, which is lower than the Classic LMA [2, 4, 5].
Consequently, a variety of techniques has been developed to facilitate insertion of the ProSeal LMA, including priming the drain tube with a guiding instrument such as a suction catheter [6], a gastric tube [7], a gum elastic Bougie [8], a Flexi-Slip stylet [9] and even a fiberoptic bronchoscope [10, 11]. Most are based on blind catheter or tube insertion, and although a fiberoptic bronchoscope enables the intraoral structures to be viewed, it is too expensive and cumbersome to be used in routine practice. The Foley Airway Stylet Tool (FAST) is a portable, simple and robust battery-powered, flexible fiberoptic endoscope. It has been reported to facilitate tracheal intubation with an intubating LMA [12], and thus we hypothesized that it might also be advantageous for ProSeal LMA insertion. The aim of this study was to compare the efficiency of the FAST optical stylet technique with the standard introducer tool (IT) technique for ProSeal LMA insertion. We determined that the LMA was correctly positioned by checking its alignment with the glottic and esophageal openings using a fiberoptic bronchoscope. We evaluated both techniques in terms of success rates, insertion times, insertion difficulty, hemodynamic response to insertion and the incidence of postoperative sore throat.

Methods

The study was approved by our local ethics committee (Kaohsiung Medical University Hospital, Kaohsiung City, Taiwan, KMHK-IRB 96023) and informed consent was obtained from all patients and got ClinicalTrials.gov Identifier (NCT02048657). We enrolled 160 adults (American Society of Anesthesiologists (ASA) physical status I-II, aged between 20 and 65 years) scheduled for elective surgery under general anesthesia using the ProSeal LMA for airway management. Exclusion criteria included an anticipated difficult airway, morbid obesity, inadequate fasting, and pre-existing sore throat or hoarseness. Patients were allocated randomly into one of two groups using a computer-generated random number table. The FAST was used to guide ProSeal LMA insertion in 80 patients (FAST group) and the ProSeal LMA was inserted with a standard IT in the remaining 80 patients (IT group) by one of two experienced anesthesiologists (Figure 1). None of the patients was aware of the insertion method used.
No premedication was administered. In the operating room, heart rate, blood pressure and arterial oxygen saturation were recorded at baseline and then every 5 min thereafter. Each patient was anesthetized in the supine position with the head resting on a 7-cm high pillow. The fluid balance deficit resulting from the overnight fast was addressed by administering lactated Ringer’s solution at one-third of the patient’s estimated circulating volume preoperatively. The size of the ProSeal LMA was selected according to the patient’s weight: size 3 for those ≤50 kg, size 4 for those >50 kg). The cuff was fully deflated and the back surface lubricated. Fentanyl 0.5 μg/kg and thiamylal 5 mg/kg were administered as induction agents, and atracurium 0.4 mg/kg given to facilitate ProSeal insertion. Following 3 minutes of mask ventilation with 60% oxygen, the ProSeal LMA was inserted using either the FAST or IT technique. Thereafter, sevoflurane was administered at an end-tidal concentration of 1.5% to 2% and end tidal CO2 (ETCO2) concentration maintained within the range 35 mmHg to 40 mmHg.The FAST device (Foley Airway Stylet Tool®, Clarus Medical, Minneapolis, MN, USA) was lightly lubricated and passed down the drainage tube until the tip of the FAST had completely emerged from its distal end (Figure 2). The FAST technique was performed thus: (1) an assistant opened the patient’s mouth; (2) the anesthesiologist held the distal portion of the stylet in the nondominant hand while holding the eyepiece of the FAST device in the dominant hand; (3) the atraumatic tip was inserted into the esophageal opening under direct vision of the laryngo-pharyngeal tissues through the patient’s mouth; (4) the anesthesiologist introduced the lubricated ProSeal LMA into the pharynx along the FAST stylet; and finally (5) the FAST device was removed while the ProSeal LMA was held in place.
In the IT group, the ProSeal LMA was inserted using the metal introducer according to the manufacturer’s instructions: (1) the patient’s mouth was opened; (2) the lubricated ProSeal LMA was inserted into the patient’s mouth using the introducer; (3) the ProSeal LMA was advanced using a one-handed technique until resistance was encountered; and (4) the introducer was removed, leaving the ProSeal in position.
Once the ProSeal LMA had been positioned in the pharynx, the cuff was inflated with air according to a pressure of 60 cmH2O. Two attempts were allowed before insertion was considered to have failed. Failed insertion was defined by any of the following criteria: (1) failure to advance the ProSeal LMA into the pharynx; (2) ProSeal LMA malposition (air leaks despite cuff inflation); or (3) ineffective ventilation (maximum expired tidal volume <6 ml/kg).
Insertion time was defined as time elapsed from opening the mouth until the ProSeal LMA was connected to the anesthetic breathing circuit. Heart rate and blood pressure were measured before and after ProSeal LMA insertion. Ease of insertion was graded as smooth, mildly resistant or requiring a second attempt. The quality of the fit of the ProSeal LMA in the glottis was evaluated using a fiberoptic bronchoscope (Olympus LF-2, Shinjuku, Tokyo, Japan) at the laryngeal aperture of the ProSeal LMA through the airway tube. The glottic view was scored using a five-point scale (Table 1) based on previous studies [2, 13, 14]. The alignment of the tip of the drain tube with the esophageal opening was evaluated using a fiberoptic bronchoscope inserted in the drain tube. The view of the esophagus was scored using a three-point scale (Table 1). Postoperative sore throat was recorded in the postoperative care unit and on the following morning. All parameters were recorded and data analyzed by an investigator blinded to the insertion technique.
Table 1
Grading of the views of the laryngeal field from the distal aperture of the airway tube and the esophagus from the drain tube using fiberoptic bronchoscopy
Viewing
Grading
Description
Laryngeal field
1
Vocal cords not seen and device functions inadequately
 
2
Vocal cords not seen but device functions adequately
 
3
Vocal cords and anterior epiglottis
 
4
Vocal cords and posterior epiglottis
 
5
Only vocal cords visible
Esophagus
A
Sealed orifice of esophagus
 
B
Crescent shape opening of esophagus
 
C
Full opening of esophagus
Study design was informed by the findings of a previous report of Bougie-guided LMA insertion time [8], and sample size estimated on the basis of one control per experimental subject, a difference in mean insertion time of 10 seconds between groups, normally distributed data and a standard deviation of 17 seconds. At least 62 subjects were needed in each group based on a type I error 0.05 and a power of 0.9. We allocated 80 patients to each group to take into account the possibility of surgical problems and failed ProSeal LMA insertion requiring patients to be excluded from analysis of total insertion time to find out reasons. We compared patient characteristics, insertion time, insertion attempts, adverse effects and complications between the groups using the two-sample t-test (for numerical variables) and the Chi-square test (for categorical variables). Statistical significance was set at a p value <0.05.

Results

From June 2008 to Dec 2009, fourteen patients in the FAST group were excluded from final analysis (seven owing to changes in the surgical plan, four withdraw consent, and three owing to failed insertion). Of the three failed insertions, two in the FAST group were as a consequence of the cuff folding in the oropharynx and one owing to an unexpectedly blurred view through the FAST device during insertion. Six patients in the IT group were excluded from final analysis (three owing to a change in surgical plan, two withdraw consent, and one owing to a failed insertion caused by folding back of the ProSeal LMA and air leak). Therefore, 69 patients in the FAST group and 75 patients in IT the group were included in the primary analysis (Figure 1). Sixty-six patients in the FAST group and 74 patients in the IT group were included in secondary analysis for ProSeal LMA insertion time spent and adverse events. There were no differences between the groups in terms of age, sex, height, weight, body mass index, ASA physical status or size of ProSeal LMA used (Table 2).
Table 2
Patient characteristics, ASA classification, LMA size, and duration and specialty of surgery
 
FAST group (n = 69)
IT group (n = 75)
P values
Male/female
16/53
28/47
0.066
Age (yr)#
41 (20-65)
43 (22-65)
0.452
Height (cm)
159.5 (8.4)
160.7 (8.0)
0.360
Weight (kg)
61.4 (13.3)
62.6 (11.2)
0.573
BMI
24.0 (4.0)
24.2 (3.5)
0.863
ASA I/ II
35/34
31/44
0.259
LMA size NO:3/NO:4
26/43
23/52
0.375
Duration of surgery (min)
117 (53)
120 (51)
0.717
Type of surgery
   
Otorhinolaryngology
3
1
0.182
Gynecology
29
26
Orthopedic
21
29
General
16
15
Urology
0
4
#Median [95% confidence interval]; BMI: body mass index; ASA, American Society of Anesthesiologists; SD: standard deviation.
All data are shown as mean (SD) unless otherwise stated.
The hemodynamic responses to ProSeal LMA insertion were broadly comparable between the groups: there were no significance differences in heart rate or mean arterial pressure (Table 3).
Table 3
Hemodynamic responses to ProSeal laryngeal mask airway insertion
 
FAST group
-Tool group
P value
Number of patients
66
74
 
Heart rate
   
  Before
80.8 (14.1)
79.1 (15.6)
0.97
  After
77.8 (12.8)
75.5 (14.7)
0.33
Mean arterial pressure
   
  Before
99.2 (13.9)
99.1 (15.4)
0.82
  After
101.1 (15.4)
98.3 (17.8)
0.27
The success rate of ProSeal LMA insertion was 95.7% in the FAST group and 98.7% in the IT group (p =0.26), and insertion took significantly longer when the FAST technique was used (p <0.001; Table 4). The ability to visualize the glottic structures and esophageal opening was also not significantly different between the groups (Table 4). The incidence of sore throat was higher in the FAST group than the IT group on the next postoperative day (p =0.035).
Table 4
ProSeal LMA insertion parameters and incidence of postoperative sore throat
 
FAST group
IT group
P value
Insertion time
17.4 ± 6.1
12.6 ± 4.7
<0.001***
Insertion condition
   
  Smooth
62 (89.9%)
67 (89.4%)
 
  Mild resistance
2 (2.9%)
1 (1.3%)
0.413
  Two attempts
2 (2.9%)
6 (8.0%)
 
  Failed insertion#
3 (4.3%)
1 (1.3%)
 
Viewing glottic field
   
   1
0 (0%)
0 (0%)
0.201
   2
4 (6.1%)
2 (2.7%)
 
   3
42 (63.6%)
57 (77.0%)
 
   4
20 (30.3%)
15 (20.3%)
 
   5
0 (0%)
0 (0%)
 
Viewing esophagus
   
   A
59 (89.4%)
69 (93.2%)
0.435
   B
6 (9.1%)
3 (4.1%)
 
   C
1 (1.5%)
2 (2.7%)
 
Sore throat in PACU
   
  None
44 (66.7%)
56 (75.7%)
0.377
  Mild
20 (30.1.1%)
18 (24.3%)
 
  Moderate
1 (1.5%)
0 (0%)
 
  Severe
1 (1.5%)
0 (0%)
 
Sore throat on the first
   
Postoperative morning
   
  None
52 (78.8%)
69 (93.2%)
0.035*
  Mild
11 (16.7%)
3 (4.1%)
 
  Moderate
3 (4.5%)
2 (2.7%)
 
  Severe
0 (0%)
0 (0%)
 
PACU, post-anesthesia care unit; #: Due to failed insertion, ProSeal LMA post-insertion assessments were excluded. *: P< 0.05; ***: P< 0.001.

Discussion

We found that the ProSeal LMA can be properly positioned using either the FAST-guided technique or IT techniques; fiberoptic bronchoscopy showed that the mask aligned correctly with the glottis and esophagus in the majority of cases. The success rates of ProSeal LMA insertion by each technique were broadly comparable, and both exceeded 95%. Nevertheless, the FAST technique took significantly longer and the incidence of sore throat on the second postoperative day was significantly higher than the traditional IT technique. We can therefore draw the conclusion that the direct vision FAST device does not appear to have any advantages over the traditional introducer for ProSeal LMA insertion.
A popular method of facilitating ProSeal LMA insertion is to use a gum-elastic Bougie in the drainage tube together with a laryngoscope, which allows the tip to be correctly positioned in the esophagus. This technique is recommended as a second-line backup means of directly visualizing the upper airway [8]. The Bougie provides sufficient rigidity to guide the cuff directly into the pharynx without folding [8, 15, 16]. However, without laryngoscopic assistance, blind insertion is reported to have caused pharyngeal wall perforation [17].
The FAST device is a lighted fiberoptic malleable stylet that functions as an “optical Bougie” and has several advantages that make it suitable for airway management: it is flexible and portable, has an atraumatic tip, and can be used easily by a single operator. These advantages ensure that the drain tube aligns with the esophageal opening during ProSeal LMA insertion. The device has been reported to aid endotracheal intubation through the intubating LMA with a success rate of 90% (27 out of 30 patients) at the first attempt and 96.7% after subsequent attempts [12]. For ProSeal LMA insertion, the FAST device provides direct vision to guide passage of the airway and prevent cuff folding or impaction during manipulation. Our findings did not support the hypothesis that the FAST device was superior to the traditional introducer-assisted technique, but it may still be useful for some instances of ProSeal LMA insertion.
We did not use a size 5 ProSeal LMA in patients weighing more than 80 kg as some patients’ passive mouth opening was not sufficient to accommodate a mask of that size. We prefer to use a smaller ProSeal LMA in such patients, which in our experience is easier to insert. Six patients in FAST group and eight patients in the IT group weighed >80 kg, respectively, and in all cases insertion was successful at the first attempt. Brimacombe and Keller have reported the successful use of a size 4 ProSeal LMA in patients within a height range of 150–193 cm and weight rang of 40–115 kg [2]. A variety of means of selecting LMA size on the basis of sex [18, 19], height [20], or an algorithm [21, 22] have been described. In apneic anesthetized adults weighing <100 kg, the optimal LMA size is reported to be size 5 in 63% and size 4 in 36% [20]. Determining the optimal size LMA for an individual adult is complex, and the best strategies for selecting either size 4 or 5 LMA in adults should be investigated in more detail.
There are several possible explanations for the folding of the cuff in the oropharynx that we observed in two patients in the FAST group despite guidance with the flexible stylet. First, the need for the operator to use two hands may complicate insertion; the dominant hand holds the eyepiece firmly but overly rapid withdrawal of the stylet by the nondominant hand might displace the LMA from its correct position. Second, an inadequate length of insertion of the stylet might easily dislodge the tip from the esophagus. Third, secretions may obscure the visual field and impede advancement of the device. Fourth, the unidirectional tip could obscure the visual field causing the tip to impact on the pharyngeal wall. Neither the FAST device nor the introducer tool can completely eliminate folding of the cuff during ProSeal LMA insertion.
Sore throat may be caused by friction between the ProSeal LMA cuff and oropharyngeal tissues during placement and removal, high cuff inflation pressure, forceful LMA advancement, or advancement of the FAST stylet tip. When injury occurs, it is usually manifest as a minor complaint such as dry mouth or sore throat [23]. In this study, the cuff of the ProSeal LMA was inflated to 60 cmH2O. The incidence of sore throat on the second postoperative day was 21.2% in the FAST group and 6.8% in the IT group, which is similar to that reported by other investigators [24, 25]. Although the exact mechanisms of ProSeal LMA-induced sore throat have yet to be identified, the finding that sore throat is also common after upper gastrointestinal endoscopy suggests that instrumentation of the airway is more likely to be responsible than pressure from the cuff once it is in situ[26]. This may also help explain the increased incidence of sore throat after FAST-guided ProSeal LMA insertion that we observed.
Our study had some limitations. First, all ProSeal LMA insertions were undertaken by experts (>1,000 cases) who clearly could not be blinded to the insertion technique, so our findings may not apply to inexperienced users. Second, the use of muscle relaxants meant that there was minimal resistance in the oropharynx and larynx during ProSeal LMA insertion, and only mild airway trauma was encountered. This may not be the case in spontaneously breathing patients. Third, patients with anticipated difficult airways were excluded, so our findings may only apply to patients with normal airways. Fourth, we found that the unidirectional tip of the FAST device limited our ability to direct the LMA during insertion, likely causing insertion failure in three cases. The incidence of failed insertion is likely to be higher in patients with anticipated difficult intubation.

Conclusion

Both the FAST device and IT technique can achieve accurate ProSeal LMA positioning with similar success rates. However, for inexperienced users the FAST device may not hold any advantages over the standard introducer-guided technique. Introducer-tool assisted ProSeal LMA insertion is quicker and is associated with a lower incidence of sore throat.

Acknowledgment

We acknowledge Steve Tredrea for the English revision of the manuscript.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.
The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.
The Creative Commons Public Domain Dedication waiver (https://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

MKC: study design, data collection and writing-up draft of the manuscript. HTH: study design and conception provided. ICL: study design, editing text and tables of the manuscript. CKS: data acquisition and analysis and interpretation of data. KYT: study design, data collection, and coordination. YCS: data acquisition and primary statistical analysis. KIC: conception and design, revised and editing the manuscript in critically for important intellectual content, and final approval of the version to be published. All authors have read and approved the final version.
Anhänge

Authors’ original submitted files for images

Below are the links to the authors’ original submitted files for images.
Literatur
1.
Zurück zum Zitat Brain AI, Verghese C, Strube PJ: The LMA ‘ProSeal’–a laryngeal mask with an oesophageal vent. Br J Anaesth. 2000, 84 (5): 650-654.CrossRefPubMed Brain AI, Verghese C, Strube PJ: The LMA ‘ProSeal’–a laryngeal mask with an oesophageal vent. Br J Anaesth. 2000, 84 (5): 650-654.CrossRefPubMed
2.
Zurück zum Zitat Brimacombe J, Keller C: The ProSeal laryngeal mask airway: a randomized, crossover study with the standard laryngeal mask airway in paralyzed, anesthetized patients. Anesthesiology. 2000, 93 (1): 104-109.CrossRefPubMed Brimacombe J, Keller C: The ProSeal laryngeal mask airway: a randomized, crossover study with the standard laryngeal mask airway in paralyzed, anesthetized patients. Anesthesiology. 2000, 93 (1): 104-109.CrossRefPubMed
3.
Zurück zum Zitat Keller C, Brimacombe J, Kleinsasser A, Loeckinger A: Does the ProSeal laryngeal mask airway prevent aspiration of regurgitated fluid?. Anesth Analg. 2000, 91 (4): 1017-1020.CrossRefPubMed Keller C, Brimacombe J, Kleinsasser A, Loeckinger A: Does the ProSeal laryngeal mask airway prevent aspiration of regurgitated fluid?. Anesth Analg. 2000, 91 (4): 1017-1020.CrossRefPubMed
4.
Zurück zum Zitat Lu PP, Brimacombe J, Yang C, Shyr M: ProSeal versus the Classic laryngeal mask airway for positive pressure ventilation during laparoscopic cholecystectomy. Br J Anaesth. 2002, 88 (6): 824-827.CrossRefPubMed Lu PP, Brimacombe J, Yang C, Shyr M: ProSeal versus the Classic laryngeal mask airway for positive pressure ventilation during laparoscopic cholecystectomy. Br J Anaesth. 2002, 88 (6): 824-827.CrossRefPubMed
5.
Zurück zum Zitat Cook TM, Nolan JP, Verghese C, Strube PJ, Lees M, Millar JM, Baskett PJ: Randomized crossover comparison of the proseal with the classic laryngeal mask airway in unparalysed anaesthetized patients. Br J Anaesth. 2002, 88 (4): 527-533.CrossRefPubMed Cook TM, Nolan JP, Verghese C, Strube PJ, Lees M, Millar JM, Baskett PJ: Randomized crossover comparison of the proseal with the classic laryngeal mask airway in unparalysed anaesthetized patients. Br J Anaesth. 2002, 88 (4): 527-533.CrossRefPubMed
6.
Zurück zum Zitat Garcia-Aguado R, Vinoles J, Brimacombe J, Vivo M, Lopez-Estudillo R, Ayala G: Suction catheter guided insertion of the ProSeal laryngeal mask airway is superior to the digital technique. Can J Anaesth. 2006, 53 (4): 398-403.CrossRefPubMed Garcia-Aguado R, Vinoles J, Brimacombe J, Vivo M, Lopez-Estudillo R, Ayala G: Suction catheter guided insertion of the ProSeal laryngeal mask airway is superior to the digital technique. Can J Anaesth. 2006, 53 (4): 398-403.CrossRefPubMed
7.
Zurück zum Zitat Drolet P, Girard M: An aid to correct positioning of the ProSeal laryngeal mask. Can J Anaesth. 2001, 48 (7): 718-719.CrossRefPubMed Drolet P, Girard M: An aid to correct positioning of the ProSeal laryngeal mask. Can J Anaesth. 2001, 48 (7): 718-719.CrossRefPubMed
8.
Zurück zum Zitat Brimacombe J, Keller C: Gum elastic bougie-guided insertion of the ProSeal Laryngeal Mask Airway. Anaesth Intensive Care. 2004, 32 (5): 681-684.PubMed Brimacombe J, Keller C: Gum elastic bougie-guided insertion of the ProSeal Laryngeal Mask Airway. Anaesth Intensive Care. 2004, 32 (5): 681-684.PubMed
9.
Zurück zum Zitat Chen HS, Yang SC, Chien CF, Spielberger J, Hung KC, Chung KC: Insertion of the ProSeal laryngeal mask airway is more successful with the Flexi-Slip stylet than with the introducer. Can J Anaesth. 2011, 58 (7): 617-623.CrossRefPubMed Chen HS, Yang SC, Chien CF, Spielberger J, Hung KC, Chung KC: Insertion of the ProSeal laryngeal mask airway is more successful with the Flexi-Slip stylet than with the introducer. Can J Anaesth. 2011, 58 (7): 617-623.CrossRefPubMed
10.
Zurück zum Zitat Brimacombe J, Keller C: Awake fibreoptic-guided insertion of the ProSeal Laryngeal Mask Airway. Anaesthesia. 2002, 57 (7): 719-CrossRefPubMed Brimacombe J, Keller C: Awake fibreoptic-guided insertion of the ProSeal Laryngeal Mask Airway. Anaesthesia. 2002, 57 (7): 719-CrossRefPubMed
11.
Zurück zum Zitat Chiu PC, Cheng KI, Tseng KY, Shih CK, Chen MK: Fibreoptic bronchoscopy to facilitate ProSeal laryngeal mask airway insertion in a patient with ankylosing spondylitis. Anaesthesia. 2011, 66 (2): 138-139.CrossRefPubMed Chiu PC, Cheng KI, Tseng KY, Shih CK, Chen MK: Fibreoptic bronchoscopy to facilitate ProSeal laryngeal mask airway insertion in a patient with ankylosing spondylitis. Anaesthesia. 2011, 66 (2): 138-139.CrossRefPubMed
12.
Zurück zum Zitat Heuer JF, Crozier TA, Braun U, Neumann P, Hilgers R, Quintel M, Timmermann A: The use of the Foley Airway Stylet Tool(R) to guide tracheal intubations through an intubating laryngeal mask airway. Minerva Anestesiol. 2011, 77 (11): 1037-1042.PubMed Heuer JF, Crozier TA, Braun U, Neumann P, Hilgers R, Quintel M, Timmermann A: The use of the Foley Airway Stylet Tool(R) to guide tracheal intubations through an intubating laryngeal mask airway. Minerva Anestesiol. 2011, 77 (11): 1037-1042.PubMed
13.
Zurück zum Zitat Kapila A, Addy EV, Verghese C, Brain AI: The intubating laryngeal mask airway: an initial assessment of performance. Br J Anaesth. 1997, 79 (6): 710-713.CrossRefPubMed Kapila A, Addy EV, Verghese C, Brain AI: The intubating laryngeal mask airway: an initial assessment of performance. Br J Anaesth. 1997, 79 (6): 710-713.CrossRefPubMed
14.
Zurück zum Zitat Teoh CY, Lim FS: The Proseal laryngeal mask airway in children: a comparison between two insertion techniques. Paediatr Anaesth. 2008, 18 (2): 119-124.PubMed Teoh CY, Lim FS: The Proseal laryngeal mask airway in children: a comparison between two insertion techniques. Paediatr Anaesth. 2008, 18 (2): 119-124.PubMed
15.
Zurück zum Zitat Howath A, Brimacombe J, Keller C: Gum-elastic bougie-guided insertion of the ProSeal laryngeal mask airway: a new technique. Anaesth Intensive Care. 2002, 30 (5): 624-627.PubMed Howath A, Brimacombe J, Keller C: Gum-elastic bougie-guided insertion of the ProSeal laryngeal mask airway: a new technique. Anaesth Intensive Care. 2002, 30 (5): 624-627.PubMed
16.
Zurück zum Zitat Taneja S, Agarwalt M, Dali JS, Agrawal G: Ease of Proseal Laryngeal Mask Airway insertion and its fibreoptic view after placement using Gum Elastic Bougie: a comparison with conventional techniques. Anaesth Intensive Care. 2009, 37 (3): 435-440.PubMed Taneja S, Agarwalt M, Dali JS, Agrawal G: Ease of Proseal Laryngeal Mask Airway insertion and its fibreoptic view after placement using Gum Elastic Bougie: a comparison with conventional techniques. Anaesth Intensive Care. 2009, 37 (3): 435-440.PubMed
17.
Zurück zum Zitat Kadry M, Popat M: Pharyngeal wall perforation–an unusual complication of blind intubation with a gum elastic bougie. Anaesthesia. 1999, 54 (4): 404-405.CrossRefPubMed Kadry M, Popat M: Pharyngeal wall perforation–an unusual complication of blind intubation with a gum elastic bougie. Anaesthesia. 1999, 54 (4): 404-405.CrossRefPubMed
18.
Zurück zum Zitat Brimacombe J, Keller C: Laryngeal mask airway size selection in males and females: ease of insertion, oropharyngeal leak pressure, pharyngeal mucosal pressures and anatomical position. Br J Anaesth. 1999, 82 (5): 703-707.CrossRefPubMed Brimacombe J, Keller C: Laryngeal mask airway size selection in males and females: ease of insertion, oropharyngeal leak pressure, pharyngeal mucosal pressures and anatomical position. Br J Anaesth. 1999, 82 (5): 703-707.CrossRefPubMed
19.
Zurück zum Zitat Kihara S, Brimacombe J: Sex-based ProSeal laryngeal mask airway size selection: a randomized crossover study of anesthetized, paralyzed male and female adult patients. Anesth Analg. 2003, 97 (1): 280-284. table of contentsCrossRefPubMed Kihara S, Brimacombe J: Sex-based ProSeal laryngeal mask airway size selection: a randomized crossover study of anesthetized, paralyzed male and female adult patients. Anesth Analg. 2003, 97 (1): 280-284. table of contentsCrossRefPubMed
20.
Zurück zum Zitat Berry AM, Brimacombe JR, McManus KF, Goldblatt M: An evaluation of the factors influencing selection of the optimal size of laryngeal mask airway in normal adults. Anaesthesia. 1998, 53 (6): 565-570.CrossRefPubMed Berry AM, Brimacombe JR, McManus KF, Goldblatt M: An evaluation of the factors influencing selection of the optimal size of laryngeal mask airway in normal adults. Anaesthesia. 1998, 53 (6): 565-570.CrossRefPubMed
21.
Zurück zum Zitat Kagawa T, Obara H: An easy formula to remember the laryngeal mask airway size-patient weight relationship. Anesthesiology. 2000, 92 (2): 631-632.CrossRefPubMed Kagawa T, Obara H: An easy formula to remember the laryngeal mask airway size-patient weight relationship. Anesthesiology. 2000, 92 (2): 631-632.CrossRefPubMed
22.
Zurück zum Zitat Hirai S, Nakamura R, Maekawa N: Alternative formula for Laryngeal Mask Airway trade mark size selection. Anesthesiology. 2004, 100 (2): 460-CrossRefPubMed Hirai S, Nakamura R, Maekawa N: Alternative formula for Laryngeal Mask Airway trade mark size selection. Anesthesiology. 2004, 100 (2): 460-CrossRefPubMed
23.
Zurück zum Zitat McHardy FE, Chung F: Postoperative sore throat: cause, prevention and treatment. Anaesthesia. 1999, 54 (5): 444-453.CrossRefPubMed McHardy FE, Chung F: Postoperative sore throat: cause, prevention and treatment. Anaesthesia. 1999, 54 (5): 444-453.CrossRefPubMed
24.
Zurück zum Zitat Vasanth Karthik R, Ranganathan P, Kulkarni AP, Sharma KS: Does cuff pressure monitoring reduce postoperative pharyngolaryngeal adverse events after LMA-ProSeal insertion? A parallel group randomised trial. J Anesth. 2014, 28 (5): 662-667.CrossRefPubMed Vasanth Karthik R, Ranganathan P, Kulkarni AP, Sharma KS: Does cuff pressure monitoring reduce postoperative pharyngolaryngeal adverse events after LMA-ProSeal insertion? A parallel group randomised trial. J Anesth. 2014, 28 (5): 662-667.CrossRefPubMed
25.
Zurück zum Zitat Roth H, Genzwuerker HV, Rothhaas A, Finteis T, Schmeck J: The ProSeal laryngeal mask airway and the laryngeal tube Suction for ventilation in gynaecological patients undergoing laparoscopic surgery. Eur J Anaesthesiol. 2005, 22 (2): 117-122.CrossRefPubMed Roth H, Genzwuerker HV, Rothhaas A, Finteis T, Schmeck J: The ProSeal laryngeal mask airway and the laryngeal tube Suction for ventilation in gynaecological patients undergoing laparoscopic surgery. Eur J Anaesthesiol. 2005, 22 (2): 117-122.CrossRefPubMed
26.
Zurück zum Zitat Evans LT, Saberi S, Kim HM, Elta GH, Schoenfeld P: Pharyngeal anesthesia during sedated EGDs: is “the spray” beneficial? A meta-analysis and systematic review. Gastrointest Endosc. 2006, 63 (6): 761-766.CrossRefPubMed Evans LT, Saberi S, Kim HM, Elta GH, Schoenfeld P: Pharyngeal anesthesia during sedated EGDs: is “the spray” beneficial? A meta-analysis and systematic review. Gastrointest Endosc. 2006, 63 (6): 761-766.CrossRefPubMed
Metadaten
Titel
Techniques for the insertion of the proseal laryngeal mask airway: comparison of the foley airway stylet tool with the introducer tool in a prospective, randomized study
verfasst von
Mao-Kai Chen
Hung-Te Hsu
I-Cheng Lu
Chih-Kai Shih
Ya-Chun Shen
Kuang-Yi Tseng
Kuang-I Cheng
Publikationsdatum
01.12.2014
Verlag
BioMed Central
Erschienen in
BMC Anesthesiology / Ausgabe 1/2014
Elektronische ISSN: 1471-2253
DOI
https://doi.org/10.1186/1471-2253-14-105

Weitere Artikel der Ausgabe 1/2014

BMC Anesthesiology 1/2014 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.