Erschienen in: Acta Neurochirurgica 11/2018

29.09.2018 | Original Article - Neurosurgical technique evaluation

Telemetry in intracranial pressure monitoring: sensor survival and drift

verfasst von: Nicolas Hernandez Norager, Alexander Lilja-Cyron, Carsten Reidies Bjarkam, Sara Duus, Marianne Juhler

Erschienen in: Acta Neurochirurgica | Ausgabe 11/2018

Einloggen, um Zugang zu erhalten

Abstract

Background

Telemetric intracranial pressure (ICP) monitoring enable long-term ICP monitoring on patients during normal day activities and may accordingly be of use during evaluation and treatment of complicated ICP disorders. However, the benefits of such equipment depend strongly on the validity of the recordings and how often the telemetric sensor needs to be re-implanted. This study investigates the clinical and technical sensor survival time and drift of the telemetric ICP sensor: Raumedic Neurovent-P-tel.

Methods

Implanted telemetric ICP sensors in the period from January 2011 to December 2017 were identified, and medical records reviewed for complications, explantation reasons, and parameters relevant for determining clinical and technical sensor survival time. Explanted sensors were tested in an experimental setup to study baseline drift.

Results

In total, implantation of 119 sensors were identified. Five sensors (4.2%) were explanted due to skin damage, three (2.5%) due to wound infection, and two (1.7%) due to ethylene oxide allergy. No other complications were observed. The median clinical sensor survival time was 208 days (95% CI 150–382). The median technical sensor survival time was 556 days (95% CI 382–605). Explanted sensors had a median baseline drift of 2.5 mmHg (IQR 2.0–5.5).

Conclusion

In most cases, the ICP sensor provides reliable measurements beyond the approved implantation time of 90 days. Thus, the sensor should not be routinely removed after this period, if ICP monitoring is still indicated. However, some sensors showed technical malfunction prior to the CE-approval, underlining that caution should always be taken when analyzing telemetric ICP curves.
Literatur
1.
Zurück zum Zitat Andresen M, Juhler M (2014) Intracranial pressure following complete removal of a small demarcated brain tumor: a model for normal intracranial pressure in humans. J Neurosurg 121(4):797–801CrossRef Andresen M, Juhler M (2014) Intracranial pressure following complete removal of a small demarcated brain tumor: a model for normal intracranial pressure in humans. J Neurosurg 121(4):797–801CrossRef
2.
Zurück zum Zitat Antes S, Tschan CA, Kunze G, Ewert L, Zimmer A, Halfmann A, Oertel J (2014) Clinical and radiological findings in long-term intracranial pressure monitoring. Acta Neurochir 156(5):1009–1019CrossRef Antes S, Tschan CA, Kunze G, Ewert L, Zimmer A, Halfmann A, Oertel J (2014) Clinical and radiological findings in long-term intracranial pressure monitoring. Acta Neurochir 156(5):1009–1019CrossRef
3.
Zurück zum Zitat Antes S, Tschan CA, Heckelmann M, Breuskin D, Oertel J (2016) Telemetric intracranial pressure monitoring with the raumedic Neurovent P-tel. World Neurosurg 91:133–148CrossRef Antes S, Tschan CA, Heckelmann M, Breuskin D, Oertel J (2016) Telemetric intracranial pressure monitoring with the raumedic Neurovent P-tel. World Neurosurg 91:133–148CrossRef
4.
Zurück zum Zitat Barber JM, Pringle CJ, Raffalli-Ebezant H, Pathmanaban O, Ramirez R, Kamaly-Asl ID (2017) Telemetric intra-cranial pressure monitoring: clinical and financial considerations. Br J Neurosurg 31(3):300–306CrossRef Barber JM, Pringle CJ, Raffalli-Ebezant H, Pathmanaban O, Ramirez R, Kamaly-Asl ID (2017) Telemetric intra-cranial pressure monitoring: clinical and financial considerations. Br J Neurosurg 31(3):300–306CrossRef
5.
Zurück zum Zitat Bousquet J, Michel FÇB (1991) Allergy to formaldehyde and ethylene-oxide. Clin Rev Allergy 9(3–4):357–370PubMed Bousquet J, Michel FÇB (1991) Allergy to formaldehyde and ethylene-oxide. Clin Rev Allergy 9(3–4):357–370PubMed
6.
Zurück zum Zitat Czosnyka Z, Czosnyka M (2017) Long-term monitoring of intracranial pressure in normal pressure hydrocephalus and other CSF disorders. Acta Neurochir 159(10):1979–1980CrossRef Czosnyka Z, Czosnyka M (2017) Long-term monitoring of intracranial pressure in normal pressure hydrocephalus and other CSF disorders. Acta Neurochir 159(10):1979–1980CrossRef
8.
Zurück zum Zitat Freimann FB, Sprung C, Chopra SS, Vajkoczy P, Wolf S (2013) Large-scale referencing of the telemetric Neurovent-P-tel intracranial pressure sensor in a porcine model. Pediatr Neurosurg 49(1):29–32CrossRef Freimann FB, Sprung C, Chopra SS, Vajkoczy P, Wolf S (2013) Large-scale referencing of the telemetric Neurovent-P-tel intracranial pressure sensor in a porcine model. Pediatr Neurosurg 49(1):29–32CrossRef
9.
Zurück zum Zitat Fried HI, Nathan BR, Rowe AS, Zabramski JM, Andaluz N, Bhimraj A, Guanci MMK, Seder DB, Singh JM (2016) The insertion and management of external ventricular drains: an evidence-based consensus statement: a statement for healthcare professionals from the neurocritical care society. Neurocrit Care 24(1):61–81CrossRef Fried HI, Nathan BR, Rowe AS, Zabramski JM, Andaluz N, Bhimraj A, Guanci MMK, Seder DB, Singh JM (2016) The insertion and management of external ventricular drains: an evidence-based consensus statement: a statement for healthcare professionals from the neurocritical care society. Neurocrit Care 24(1):61–81CrossRef
10.
Zurück zum Zitat Jiang G (2010) Design challenges of implantable pressure monitoring system. Front Neurosci 4(FEB):2–5PubMedCentral Jiang G (2010) Design challenges of implantable pressure monitoring system. Front Neurosci 4(FEB):2–5PubMedCentral
11.
Zurück zum Zitat Kiefer M, Antes S, Leonhardt S, Schmitt M, Orakcioglu B, Sakowitz OW, Eymann R (2012) Telemetric ICP measurement with the first CE-approved device: data from animal experiments and initial clinical experiences. Acta Neurochir (Suppl):111–116 Kiefer M, Antes S, Leonhardt S, Schmitt M, Orakcioglu B, Sakowitz OW, Eymann R (2012) Telemetric ICP measurement with the first CE-approved device: data from animal experiments and initial clinical experiences. Acta Neurochir (Suppl):111–116
12.
Zurück zum Zitat Kotzar G, Freas M, Abel P, Fleischman A, Roy S, Zorman C, Moran JM, Melzak J (2002) Evaluation of MEMS materials of construction for implantable medical devices. Biomaterials 23(13):2737–2750CrossRef Kotzar G, Freas M, Abel P, Fleischman A, Roy S, Zorman C, Moran JM, Melzak J (2002) Evaluation of MEMS materials of construction for implantable medical devices. Biomaterials 23(13):2737–2750CrossRef
13.
Zurück zum Zitat Lilja A, Andresen M, Hadi A, Christoffersen D, Juhler M (2014) Clinical experience with telemetric intracranial pressure monitoring in a Danish neurosurgical center. Clin Neurol Neurosurg 120:36–40CrossRef Lilja A, Andresen M, Hadi A, Christoffersen D, Juhler M (2014) Clinical experience with telemetric intracranial pressure monitoring in a Danish neurosurgical center. Clin Neurol Neurosurg 120:36–40CrossRef
14.
Zurück zum Zitat Lilja-Cyron A, Kelsen J, Andresen M, Fugleholm K, Juhler M (2018) Feasibility of telemetric intracranial pressure monitoring in the neuro intensive care unit. J Neurotrauma 35(14):1578–1586CrossRef Lilja-Cyron A, Kelsen J, Andresen M, Fugleholm K, Juhler M (2018) Feasibility of telemetric intracranial pressure monitoring in the neuro intensive care unit. J Neurotrauma 35(14):1578–1586CrossRef
15.
Zurück zum Zitat Lozier AP, Sciacca RR, Romagnoli MF, Connolly ES, McComb JG, Cohen AR, Rock JP (2002) Ventriculostomy-related infections: a critical review of the literature. Neurosurgery 51(1):170–182CrossRef Lozier AP, Sciacca RR, Romagnoli MF, Connolly ES, McComb JG, Cohen AR, Rock JP (2002) Ventriculostomy-related infections: a critical review of the literature. Neurosurgery 51(1):170–182CrossRef
16.
Zurück zum Zitat Raboel PH, Bartek J, Andresen M, Bellander BM, Romner B (2012) Intracranial pressure monitoring: invasive versus non-invasive methods-a review. Crit Care Res Pract 2012:3–7 Raboel PH, Bartek J, Andresen M, Bellander BM, Romner B (2012) Intracranial pressure monitoring: invasive versus non-invasive methods-a review. Crit Care Res Pract 2012:3–7
17.
Zurück zum Zitat Sæhle T, Eide PK (2015) Intracranial pressure monitoring in pediatric and adult patients with hydrocephalus and tentative shunt failure: a single-center experience over 10 years in 146 patients. J Neurosurg 122(5):1076–1086CrossRef Sæhle T, Eide PK (2015) Intracranial pressure monitoring in pediatric and adult patients with hydrocephalus and tentative shunt failure: a single-center experience over 10 years in 146 patients. J Neurosurg 122(5):1076–1086CrossRef
18.
Zurück zum Zitat Schuhmann MU, Sood S, McAllister JP, Jaeger M, Ham SD, Czosnyka Z, Czosnyka M (2008) Value of overnight monitoring of intracranial pressure in hydrocephalic children. Pediatr Neurosurg 44(4):269–279CrossRef Schuhmann MU, Sood S, McAllister JP, Jaeger M, Ham SD, Czosnyka Z, Czosnyka M (2008) Value of overnight monitoring of intracranial pressure in hydrocephalic children. Pediatr Neurosurg 44(4):269–279CrossRef
19.
Zurück zum Zitat Welschehold S, Schmalhausen E, Dodier P, Vulcu S, Oertel J, Wagner W, Tschan CA (2012) First clinical results with a new telemetric intracranial pressure-monitoring system. Oper Neurosurg 70(1):44–49CrossRef Welschehold S, Schmalhausen E, Dodier P, Vulcu S, Oertel J, Wagner W, Tschan CA (2012) First clinical results with a new telemetric intracranial pressure-monitoring system. Oper Neurosurg 70(1):44–49CrossRef
20.
Zurück zum Zitat Wiegand C, Richards P (2007) Measurement of intracranial pressure in children: a critical review of current methods. Dev Med Child Neurol 49(12):935–941CrossRef Wiegand C, Richards P (2007) Measurement of intracranial pressure in children: a critical review of current methods. Dev Med Child Neurol 49(12):935–941CrossRef
21.
Zurück zum Zitat Yu L, Kim BJ, Meng E (2014) Chronically implanted pressure sensors: challenges and state of the field. Sensors (Switzerland) 14(11):20620–20644CrossRef Yu L, Kim BJ, Meng E (2014) Chronically implanted pressure sensors: challenges and state of the field. Sensors (Switzerland) 14(11):20620–20644CrossRef
22.
Zurück zum Zitat Zacchetti L, Magnoni S, Di Corte F, Zanier ER, Stocchetti N (2015) Accuracy of intracranial pressure monitoring: systematic review and meta-analysis. Crit Care 19(1):1–8CrossRef Zacchetti L, Magnoni S, Di Corte F, Zanier ER, Stocchetti N (2015) Accuracy of intracranial pressure monitoring: systematic review and meta-analysis. Crit Care 19(1):1–8CrossRef
23.
Zurück zum Zitat Zhang X, Medow JE, Iskandar BJ, Wang F, Shokoueinejad M, Koueik J, Webster JG (2017) Invasive and noninvasive means of measuring intracranial pressure: a review. Physiol Meas 38(8):R143–R182CrossRef Zhang X, Medow JE, Iskandar BJ, Wang F, Shokoueinejad M, Koueik J, Webster JG (2017) Invasive and noninvasive means of measuring intracranial pressure: a review. Physiol Meas 38(8):R143–R182CrossRef
24.
Zurück zum Zitat Zhong J, Dujovny M, Park HK, Perez E, Perlin AR, Diaz FG (2003) Advances in ICP monitoring techniques. Neurol Res 25(4):339–350CrossRef Zhong J, Dujovny M, Park HK, Perez E, Perlin AR, Diaz FG (2003) Advances in ICP monitoring techniques. Neurol Res 25(4):339–350CrossRef
Metadaten
Titel
Telemetry in intracranial pressure monitoring: sensor survival and drift
verfasst von
Nicolas Hernandez Norager
Alexander Lilja-Cyron
Carsten Reidies Bjarkam
Sara Duus
Marianne Juhler
Publikationsdatum
29.09.2018
Verlag
Springer Vienna
Erschienen in
Acta Neurochirurgica / Ausgabe 11/2018
Print ISSN: 0001-6268
Elektronische ISSN: 0942-0940
DOI
https://doi.org/10.1007/s00701-018-3691-9