Skip to main content
Erschienen in: European Spine Journal 12/2015

01.12.2015 | Original Article

The accuracy of 3D fluoroscopy-navigated screw insertion in the upper and subaxial cervical spine

verfasst von: J. Bredow, J. Oppermann, B. Kraus, P. Schiller, G. Schiffer, R. Sobottke, P. Eysel, T. Koy

Erschienen in: European Spine Journal | Ausgabe 12/2015

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Due to better primary stability and repositioning options, pedicle screws are increasingly used during posterior stabilization of the cervical spine. However, the serious risks generally associated with the insertion of screws in the cervical spine remain. The purpose of this study is to examine the accuracy of pedicle screw insertion with the use of 3D fluoroscopy navigation systems, also accounting for various spine levels.

Methods

Data of 64 patients were collected during and after screw implantation (axial and subaxial) in the cervical spine. 207 screws were implanted from C1 to C7 and analyzed for placement accuracy according to postoperative CT scans and following the modified Gertzbein and Robbins classification.

Results

The accuracy of most of the inserted screws was assessed as grade 2 according to the modified Gertzbein and Robbins classification. 93.9 % of the screws implanted at C1 or C2, and 78.51 % of the screws implanted at levels C3–C7 showed placement accuracy grade 2 or better, indicating pedicle wall perforation of <2 mm. Overall, seven complications were observed. In three cases, the vertebral artery was affected, leading to one fatality. Surgical revision was necessary once because of Magerl screw misplacement and three times due to impaired wound healing. No radicular symptoms resulted from screw malposition.

Conclusion

Axial and subaxial screws can be inserted with a high grade of accuracy using 3D fluoroscopy-based navigation systems. Nevertheless, while this useful innovation helps to minimize the risks of misplacement, the surgery is still a challenge, as arising complications remain severe.
Literatur
1.
Zurück zum Zitat Abumi K, Itoh H, Taneichi H, Kaneda K (1994) Transpedicular screw fixation for traumatic lesions of the middle and lower cervical spine: description of the techniques and preliminary report. J Spinal Disord 7(1):19–28CrossRefPubMed Abumi K, Itoh H, Taneichi H, Kaneda K (1994) Transpedicular screw fixation for traumatic lesions of the middle and lower cervical spine: description of the techniques and preliminary report. J Spinal Disord 7(1):19–28CrossRefPubMed
2.
Zurück zum Zitat Abumi K, Kaneda K (1997) Pedicle screw fixation for nontraumatic lesions of the cervical spine. Spine 22(16):1853–1863CrossRefPubMed Abumi K, Kaneda K (1997) Pedicle screw fixation for nontraumatic lesions of the cervical spine. Spine 22(16):1853–1863CrossRefPubMed
3.
Zurück zum Zitat Abumi K, Shono Y, Ito M, Taneichi H, Kotani Y, Kaneda K (2000) Complications of pedicle screw fixation in reconstructive surgery of the cervical spine. Spine 25(8):962–969CrossRefPubMed Abumi K, Shono Y, Ito M, Taneichi H, Kotani Y, Kaneda K (2000) Complications of pedicle screw fixation in reconstructive surgery of the cervical spine. Spine 25(8):962–969CrossRefPubMed
5.
Zurück zum Zitat Blauth M, Richter M, Lange U (1999) Trans-articular screw fixation of C1/C2 in atlanto-axial instability. Comparison between percutaneous and open procedures. Der Orthopade 28(8):651–661PubMed Blauth M, Richter M, Lange U (1999) Trans-articular screw fixation of C1/C2 in atlanto-axial instability. Comparison between percutaneous and open procedures. Der Orthopade 28(8):651–661PubMed
7.
Zurück zum Zitat Ebraheim NA, Xu R, Knight T, Yeasting RA (1997) Morphometric evaluation of lower cervical pedicle and its projection. Spine 22(1):1–6CrossRefPubMed Ebraheim NA, Xu R, Knight T, Yeasting RA (1997) Morphometric evaluation of lower cervical pedicle and its projection. Spine 22(1):1–6CrossRefPubMed
9.
Zurück zum Zitat Gelalis ID, Paschos NK, Pakos EE, Politis AN, Arnaoutoglou CM, Karageorgos AC, Ploumis A, Xenakis TA (2012) Accuracy of pedicle screw placement: a systematic review of prospective in vivo studies comparing free hand, fluoroscopy guidance and navigation techniques. Eur Spine J 21(2):247–255. doi:10.1007/s00586-011-2011-3 PubMedCentralCrossRefPubMed Gelalis ID, Paschos NK, Pakos EE, Politis AN, Arnaoutoglou CM, Karageorgos AC, Ploumis A, Xenakis TA (2012) Accuracy of pedicle screw placement: a systematic review of prospective in vivo studies comparing free hand, fluoroscopy guidance and navigation techniques. Eur Spine J 21(2):247–255. doi:10.​1007/​s00586-011-2011-3 PubMedCentralCrossRefPubMed
10.
Zurück zum Zitat Gertzbein SD, Robbins SE (1990) Accuracy of pedicular screw placement in vivo. Spine 15(1):11–14CrossRefPubMed Gertzbein SD, Robbins SE (1990) Accuracy of pedicular screw placement in vivo. Spine 15(1):11–14CrossRefPubMed
11.
Zurück zum Zitat Glossop ND, Hu RW, Randle JA (1996) Computer-aided pedicle screw placement using frameless stereotaxis. Spine 21(17):2026–2034CrossRefPubMed Glossop ND, Hu RW, Randle JA (1996) Computer-aided pedicle screw placement using frameless stereotaxis. Spine 21(17):2026–2034CrossRefPubMed
12.
Zurück zum Zitat Haid RW Jr (2001) C1–C2 transarticular screw fixation: technical aspects. Neurosurgery 49(1):71–74PubMed Haid RW Jr (2001) C1–C2 transarticular screw fixation: technical aspects. Neurosurgery 49(1):71–74PubMed
13.
Zurück zum Zitat Ishikawa Y, Kanemura T, Yoshida G, Ito Z, Muramoto A, Ohno S (2010) Clinical accuracy of three-dimensional fluoroscopy-based computer-assisted cervical pedicle screw placement: a retrospective comparative study of conventional versus computer-assisted cervical pedicle screw placement. J Neurosurg Spine 13(5):606–611. doi:10.3171/2010.5.SPINE09993 CrossRefPubMed Ishikawa Y, Kanemura T, Yoshida G, Ito Z, Muramoto A, Ohno S (2010) Clinical accuracy of three-dimensional fluoroscopy-based computer-assisted cervical pedicle screw placement: a retrospective comparative study of conventional versus computer-assisted cervical pedicle screw placement. J Neurosurg Spine 13(5):606–611. doi:10.​3171/​2010.​5.​SPINE09993 CrossRefPubMed
14.
Zurück zum Zitat Ishikawa Y, Kanemura T, Yoshida G, Matsumoto A, Ito Z, Tauchi R, Muramoto A, Ohno S, Nishimura Y (2011) Intraoperative, full-rotation, three-dimensional image (O-arm)-based navigation system for cervical pedicle screw insertion. J Neurosurg Spine 15(5):472–478. doi:10.3171/2011.6.SPINE10809 CrossRefPubMed Ishikawa Y, Kanemura T, Yoshida G, Matsumoto A, Ito Z, Tauchi R, Muramoto A, Ohno S, Nishimura Y (2011) Intraoperative, full-rotation, three-dimensional image (O-arm)-based navigation system for cervical pedicle screw insertion. J Neurosurg Spine 15(5):472–478. doi:10.​3171/​2011.​6.​SPINE10809 CrossRefPubMed
15.
16.
Zurück zum Zitat Karaikovic EE, Kunakornsawat S, Daubs MD, Madsen TW, Gaines RW Jr (2000) Surgical anatomy of the cervical pedicles: landmarks for posterior cervical pedicle entrance localization. J Spinal Disord 13(1):63–72CrossRefPubMed Karaikovic EE, Kunakornsawat S, Daubs MD, Madsen TW, Gaines RW Jr (2000) Surgical anatomy of the cervical pedicles: landmarks for posterior cervical pedicle entrance localization. J Spinal Disord 13(1):63–72CrossRefPubMed
19.
Zurück zum Zitat Liu YJ, Tian W, Liu B, Li Q, Hu L, Li ZY, Yuan Q, Lu YW, Sun YZ (2010) Comparison of the clinical accuracy of cervical (C2–C7) pedicle screw insertion assisted by fluoroscopy, computed tomography-based navigation, and intraoperative three-dimensional C-arm navigation. Chin Med J 123(21):2995–2998PubMed Liu YJ, Tian W, Liu B, Li Q, Hu L, Li ZY, Yuan Q, Lu YW, Sun YZ (2010) Comparison of the clinical accuracy of cervical (C2–C7) pedicle screw insertion assisted by fluoroscopy, computed tomography-based navigation, and intraoperative three-dimensional C-arm navigation. Chin Med J 123(21):2995–2998PubMed
20.
Zurück zum Zitat Ludwig SC, Kramer DL, Vaccaro AR, Albert TJ (1999) Transpedicle screw fixation of the cervical spine. Clin Orthop Relat Res 359:77–88CrossRefPubMed Ludwig SC, Kramer DL, Vaccaro AR, Albert TJ (1999) Transpedicle screw fixation of the cervical spine. Clin Orthop Relat Res 359:77–88CrossRefPubMed
21.
Zurück zum Zitat Luther N, Iorgulescu JB, Geannette C, Gebhard H, Saleh T, Tsiouris AJ, Hartl R (2013) Comparison of navigated versus non-navigated pedicle screw placement in 260 patients and 1434 screws: screw accuracy, screw size, and the complexity of surgery. J Spinal Disord Tech. doi:10.1097/BSD.0b013e31828af33e Luther N, Iorgulescu JB, Geannette C, Gebhard H, Saleh T, Tsiouris AJ, Hartl R (2013) Comparison of navigated versus non-navigated pedicle screw placement in 260 patients and 1434 screws: screw accuracy, screw size, and the complexity of surgery. J Spinal Disord Tech. doi:10.​1097/​BSD.​0b013e31828af33e​
23.
Zurück zum Zitat Nakashima H, Yukawa Y, Imagama S, Kanemura T, Kamiya M, Yanase M, Ito K, Machino M, Yoshida G, Ishikawa Y, Matsuyama Y, Ishiguro N, Kato F (2012) Complications of cervical pedicle screw fixation for nontraumatic lesions: a multicenter study of 84 patients. J Neurosurg Spine 16(3):238–247. doi:10.3171/2011.11.SPINE11102 CrossRefPubMed Nakashima H, Yukawa Y, Imagama S, Kanemura T, Kamiya M, Yanase M, Ito K, Machino M, Yoshida G, Ishikawa Y, Matsuyama Y, Ishiguro N, Kato F (2012) Complications of cervical pedicle screw fixation for nontraumatic lesions: a multicenter study of 84 patients. J Neurosurg Spine 16(3):238–247. doi:10.​3171/​2011.​11.​SPINE11102 CrossRefPubMed
24.
Zurück zum Zitat Panjabi MM, Shin EK, Chen NC, Wang JL (2000) Internal morphology of human cervical pedicles. Spine 25(10):1197–1205CrossRefPubMed Panjabi MM, Shin EK, Chen NC, Wang JL (2000) Internal morphology of human cervical pedicles. Spine 25(10):1197–1205CrossRefPubMed
25.
Zurück zum Zitat Richter M, Cakir B, Schmidt R (2005) Cervical pedicle screws: conventional versus computer-assisted placement of cannulated screws. Spine 30(20):2280–2287CrossRefPubMed Richter M, Cakir B, Schmidt R (2005) Cervical pedicle screws: conventional versus computer-assisted placement of cannulated screws. Spine 30(20):2280–2287CrossRefPubMed
27.
Zurück zum Zitat Richter M, Schmidt R, Claes L, Puhl W, Wilke HJ (2002) Posterior atlantoaxial fixation: biomechanical in vitro comparison of six different techniques. Spine 27(16):1724–1732CrossRefPubMed Richter M, Schmidt R, Claes L, Puhl W, Wilke HJ (2002) Posterior atlantoaxial fixation: biomechanical in vitro comparison of six different techniques. Spine 27(16):1724–1732CrossRefPubMed
28.
Zurück zum Zitat Roy-Camille R, Saillant G, Laville C, Benazet JP (1992) Treatment of lower cervical spinal injuries—C3 to C7. Spine 17(10 Suppl):S442–S446CrossRefPubMed Roy-Camille R, Saillant G, Laville C, Benazet JP (1992) Treatment of lower cervical spinal injuries—C3 to C7. Spine 17(10 Suppl):S442–S446CrossRefPubMed
29.
Zurück zum Zitat Sakamoto T, Neo M, Nakamura T (2004) Transpedicular screw placement evaluated by axial computed tomography of the cervical pedicle. Spine 29(22):2510–2514 discussion 2515 CrossRefPubMed Sakamoto T, Neo M, Nakamura T (2004) Transpedicular screw placement evaluated by axial computed tomography of the cervical pedicle. Spine 29(22):2510–2514 discussion 2515 CrossRefPubMed
30.
Zurück zum Zitat Schmidt R, Wilke HJ, Claes L, Puhl W, Richter M (2003) Pedicle screws enhance primary stability in multilevel cervical corpectomies: biomechanical in vitro comparison of different implants including constrained and nonconstrained posterior instrumentations. Spine 28(16):1821–1828. doi:10.1097/01.BRS.0000083287.23521.48 CrossRefPubMed Schmidt R, Wilke HJ, Claes L, Puhl W, Richter M (2003) Pedicle screws enhance primary stability in multilevel cervical corpectomies: biomechanical in vitro comparison of different implants including constrained and nonconstrained posterior instrumentations. Spine 28(16):1821–1828. doi:10.​1097/​01.​BRS.​0000083287.​23521.​48 CrossRefPubMed
32.
Zurück zum Zitat Wang Y, Xie J, Yang Z, Zhao Z, Zhang Y, Li T, Liu L (2013) Computed tomography assessment of lateral pedicle wall perforation by free-hand subaxial cervical pedicle screw placement. Arch Orthop Trauma Surg 133(7):901–909. doi:10.1007/s00402-013-1752-3 CrossRefPubMed Wang Y, Xie J, Yang Z, Zhao Z, Zhang Y, Li T, Liu L (2013) Computed tomography assessment of lateral pedicle wall perforation by free-hand subaxial cervical pedicle screw placement. Arch Orthop Trauma Surg 133(7):901–909. doi:10.​1007/​s00402-013-1752-3 CrossRefPubMed
33.
Zurück zum Zitat Weinstein JN, Spratt KF, Spengler D, Brick C, Reid S (1988) Spinal pedicle fixation: reliability and validity of roentgenogram-based assessment and surgical factors on successful screw placement. Spine 13(9):1012–1018CrossRefPubMed Weinstein JN, Spratt KF, Spengler D, Brick C, Reid S (1988) Spinal pedicle fixation: reliability and validity of roentgenogram-based assessment and surgical factors on successful screw placement. Spine 13(9):1012–1018CrossRefPubMed
34.
Zurück zum Zitat Xu R, Ebraheim NA, Skie M (2008) Pedicle screw fixation in the cervical spine. Am J Orthop 37(8):403–408 discussion 408 PubMed Xu R, Ebraheim NA, Skie M (2008) Pedicle screw fixation in the cervical spine. Am J Orthop 37(8):403–408 discussion 408 PubMed
35.
Zurück zum Zitat Yukawa Y, Kato F, Ito K, Nakashima H, Machino M (2009) Anterior cervical pedicle screw and plate fixation using fluoroscope-assisted pedicle axis view imaging: a preliminary report of a new cervical reconstruction technique. Eur Spine J 18(6):911–916. doi:10.1007/s00586-009-0949-1 PubMedCentralCrossRefPubMed Yukawa Y, Kato F, Ito K, Nakashima H, Machino M (2009) Anterior cervical pedicle screw and plate fixation using fluoroscope-assisted pedicle axis view imaging: a preliminary report of a new cervical reconstruction technique. Eur Spine J 18(6):911–916. doi:10.​1007/​s00586-009-0949-1 PubMedCentralCrossRefPubMed
Metadaten
Titel
The accuracy of 3D fluoroscopy-navigated screw insertion in the upper and subaxial cervical spine
verfasst von
J. Bredow
J. Oppermann
B. Kraus
P. Schiller
G. Schiffer
R. Sobottke
P. Eysel
T. Koy
Publikationsdatum
01.12.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
European Spine Journal / Ausgabe 12/2015
Print ISSN: 0940-6719
Elektronische ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-015-3974-2