12.02.2016 | Original Research

The effect of head up tilting on bioreactance cardiac output and stroke volume readings using suprasternal transcutaneous Doppler as a control in healthy young adults

verfasst von: Jie Zhang, Lester A. H. Critchley, Daniel C. W. Lee, Kim S. Khaw, Shara W. Y. Lee

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 5/2016

Einloggen, um Zugang zu erhalten

Abstract

To compare the performance of a bioreactance cardiac output (CO) monitor (NICOM) and transcutaneous Doppler (USCOM) during head up tilting (HUT). Healthy young adult subjects, age 22 ± 1 years, 7 male and 7 female, were tilted over 3–5 s from supine to 70° HUT, 30° HUT and back to supine. Positions were held for 3 min. Simultaneous readings of NICOM and USCOM were performed 30 s into each new position. Mean blood pressure (MBP), heart rate (HR), CO and stroke volume (SV), and thoracic fluid content (TFC) were recorded. Bland–Altman, percentage changes and analysis of variance for repeated measures were used for statistical analysis. Pre-tilt NICOM CO and SV readings (6.1 ± 1.0 L/min and 113 ± 25 ml) were higher than those from USCOM (4.1 ± 0.6 L/min and 77 ± 9 ml) (P < 0.001). Bland–Altman limits of agreement for CO were wide with a percentage error of 38 %. HUT increased MBP and HR (P < 0.001). CO and SV readings decreased with HUT. However, the percentage changes in USCOM and NICOM readings did not concur (P < 0.001). Whereas USCOM provided gravitational effect proportional changes in SV readings of 23 ± 15 % (30° half tilt) and 44 ± 11 % (70° near full tilt), NICOM changes did not being 28 ± 10 and 33 ± 11 %. TFC decreased linearly with HUT. The NICOM does not provide linear changes in SV as predicted by physiology when patients are tilted. Furthermore there is a lack of agreement with USCOM measurements at baseline and during tilting.
Literatur
2.
Zurück zum Zitat Critchley LA. Impedance cardiography. The impact of new technology. Anaesthesia. 1998;53(7):677–84.CrossRefPubMed Critchley LA. Impedance cardiography. The impact of new technology. Anaesthesia. 1998;53(7):677–84.CrossRefPubMed
5.
Zurück zum Zitat Phillips RA, Hood SG, Jacobson BM, West MJ, Wan L, May CN. Pulmonary artery catheter (PAC) accuracy and efficacy compared with flow probe and transcutaneous doppler (USCOM): an ovine cardiac output validation. Crit Care Res Pract. 2012;2012:9. doi:10.1155/2012/621496. Phillips RA, Hood SG, Jacobson BM, West MJ, Wan L, May CN. Pulmonary artery catheter (PAC) accuracy and efficacy compared with flow probe and transcutaneous doppler (USCOM): an ovine cardiac output validation. Crit Care Res Pract. 2012;2012:9. doi:10.​1155/​2012/​621496.
6.
Zurück zum Zitat Huang L, Critchley LA. An assessment of two Doppler-based monitors to track cardiac output changes in anaesthetised patients undergoing major surgery. Anaesth Intensive Care. 2014;42(5):631–9.PubMed Huang L, Critchley LA. An assessment of two Doppler-based monitors to track cardiac output changes in anaesthetised patients undergoing major surgery. Anaesth Intensive Care. 2014;42(5):631–9.PubMed
8.
Zurück zum Zitat Huang L, Critchley LA, Zhang J. Major upper abdominal surgery alters the calibration of bioreactance cardiac output readings, the NICOM, when comparisons are made against suprasternal and esophageal Doppler intraoperatively. Anesth Analg. 2015;121(4):936–45. doi:10.1213/ANE.0000000000000889.CrossRefPubMed Huang L, Critchley LA, Zhang J. Major upper abdominal surgery alters the calibration of bioreactance cardiac output readings, the NICOM, when comparisons are made against suprasternal and esophageal Doppler intraoperatively. Anesth Analg. 2015;121(4):936–45. doi:10.​1213/​ANE.​0000000000000889​.CrossRefPubMed
9.
Zurück zum Zitat Sander-Jensen K, Secher NH, Astrup A, Christensen N, Giese J, Schwartz T, Warberg J, Bie P. Hypotension induced by passive head-up tilt: endocrine and circulatory mechanisms. Am J Physiol Regul Integr Comp Physiol. 1986;251(4):R742–8. Sander-Jensen K, Secher NH, Astrup A, Christensen N, Giese J, Schwartz T, Warberg J, Bie P. Hypotension induced by passive head-up tilt: endocrine and circulatory mechanisms. Am J Physiol Regul Integr Comp Physiol. 1986;251(4):R742–8.
10.
11.
Zurück zum Zitat Nidorf SM, Picard MH, Triulzi MO, Thomas JD, Newell J, King ME, Weyman AE. New perspectives in the assessment of cardiac chamber dimensions during development and adulthood. J Am Coll Cardiol. 1992;19(5):983–8.CrossRefPubMed Nidorf SM, Picard MH, Triulzi MO, Thomas JD, Newell J, King ME, Weyman AE. New perspectives in the assessment of cardiac chamber dimensions during development and adulthood. J Am Coll Cardiol. 1992;19(5):983–8.CrossRefPubMed
13.
Zurück zum Zitat Cattermole G, Leung P, Tang C, Smith B, Graham C, Rainer T. A new method to score the quality of USCOM scans. Hong Kong J Emerg Med. 2009;16(4):288. Cattermole G, Leung P, Tang C, Smith B, Graham C, Rainer T. A new method to score the quality of USCOM scans. Hong Kong J Emerg Med. 2009;16(4):288.
14.
Zurück zum Zitat Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8(2):135–60.CrossRefPubMed Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8(2):135–60.CrossRefPubMed
16.
Zurück zum Zitat Critchley LA, Conway F, Anderson PJ, Tomlinson B, Critchley JA. Non-invasive continuous arterial pressure, heart rate and stroke volume measurements during graded head-up tilt in normal man. Clin Auton Res. 1997;7(2):97–101.CrossRefPubMed Critchley LA, Conway F, Anderson PJ, Tomlinson B, Critchley JA. Non-invasive continuous arterial pressure, heart rate and stroke volume measurements during graded head-up tilt in normal man. Clin Auton Res. 1997;7(2):97–101.CrossRefPubMed
18.
Zurück zum Zitat Raval NY, Squara P, Cleman M, Yalamanchili K, Winklmaier M, Burkhoff D. Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique. J Clin Monit Comput. 2008;22(2):113–9. doi:10.1007/s10877-008-9112-5.CrossRefPubMed Raval NY, Squara P, Cleman M, Yalamanchili K, Winklmaier M, Burkhoff D. Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique. J Clin Monit Comput. 2008;22(2):113–9. doi:10.​1007/​s10877-008-9112-5.CrossRefPubMed
22.
Zurück zum Zitat Meng L, Tran NP, Alexander BS, Laning K, Chen G, Kain ZN, Cannesson M. The impact of phenylephrine, ephedrine, and increased preload on third-generation Vigileo-FloTrac and esophageal doppler cardiac output measurements. Anesth Analg. 2011;113(4):751–7. doi:10.1213/ANE.0b013e31822649fb.CrossRefPubMed Meng L, Tran NP, Alexander BS, Laning K, Chen G, Kain ZN, Cannesson M. The impact of phenylephrine, ephedrine, and increased preload on third-generation Vigileo-FloTrac and esophageal doppler cardiac output measurements. Anesth Analg. 2011;113(4):751–7. doi:10.​1213/​ANE.​0b013e31822649fb​.CrossRefPubMed
23.
Zurück zum Zitat Matthieu B, Karine N-G, Vincent C, Alain V, François CJ, Philippe R, François S. Cardiac output measurement in patients undergoing liver transplantation: pulmonary artery catheter versus uncalibrated arterial pressure waveform analysis. Anesth Analg. 2008;106(5):1480–6.CrossRef Matthieu B, Karine N-G, Vincent C, Alain V, François CJ, Philippe R, François S. Cardiac output measurement in patients undergoing liver transplantation: pulmonary artery catheter versus uncalibrated arterial pressure waveform analysis. Anesth Analg. 2008;106(5):1480–6.CrossRef
24.
Zurück zum Zitat Critchley LA, Peng ZY, Fok BS, James AE. The effect of peripheral resistance on impedance cardiography measurements in the anesthetized dog. Anesth Analg. 2005;100(6):1708–12.CrossRefPubMed Critchley LA, Peng ZY, Fok BS, James AE. The effect of peripheral resistance on impedance cardiography measurements in the anesthetized dog. Anesth Analg. 2005;100(6):1708–12.CrossRefPubMed
25.
26.
Zurück zum Zitat Marik PE, Levitov A, Young A, Andrews L. The use of bioreactance and carotid Doppler to determine volume responsiveness and blood flow redistribution following passive leg raising in hemodynamically unstable patients. Chest. 2013;143(2):364–70. doi:10.1378/chest.12-1274.CrossRefPubMed Marik PE, Levitov A, Young A, Andrews L. The use of bioreactance and carotid Doppler to determine volume responsiveness and blood flow redistribution following passive leg raising in hemodynamically unstable patients. Chest. 2013;143(2):364–70. doi:10.​1378/​chest.​12-1274.CrossRefPubMed
28.
Zurück zum Zitat Kupersztych-Hagege E, Teboul JL, Artigas A, Talbot A, Sabatier C, Richard C, Monnet X. Bioreactance is not reliable for estimating cardiac output and the effects of passive leg raising in critically ill patients. Br J Anaesth. 2013;111(6):961–6. doi:10.1093/bja/aet282.CrossRefPubMed Kupersztych-Hagege E, Teboul JL, Artigas A, Talbot A, Sabatier C, Richard C, Monnet X. Bioreactance is not reliable for estimating cardiac output and the effects of passive leg raising in critically ill patients. Br J Anaesth. 2013;111(6):961–6. doi:10.​1093/​bja/​aet282.CrossRefPubMed
29.
Metadaten
Titel
The effect of head up tilting on bioreactance cardiac output and stroke volume readings using suprasternal transcutaneous Doppler as a control in healthy young adults
verfasst von
Jie Zhang
Lester A. H. Critchley
Daniel C. W. Lee
Kim S. Khaw
Shara W. Y. Lee
Publikationsdatum
12.02.2016
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 5/2016
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-016-9835-7

Weitere Artikel der Ausgabe 5/2016

Journal of Clinical Monitoring and Computing 5/2016 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.