01.12.2017 | Research article | Ausgabe 1/2017 Open Access

BMC Ophthalmology 1/2017

The effect of ocular biometric factors on the accuracy of various IOL power calculation formulas

Zeitschrift:
BMC Ophthalmology > Ausgabe 1/2017
Autoren:
Jinho Jeong, Han Song, Jimmy K. Lee, Roy S. Chuck, Ji-Won Kwon

Abstract

Background

To evaluate how differences in ocular biometry affects the Hoffer Q, Holladay 1, SRK/T, and Haigis intraocular lens power calculation formulae predictions.

Methods

This study was performed on 91 eyes of 91 patients who underwent uneventful cataract surgery. Ocular biometry values were measured using the IOL Master 500, and intraocular lens (IOL) power was calculated using the Haigis, Hoffer Q, Holladay 1, and SRK/T formulas. We calculated the expected difference (ED) of each 3rd generation formula from the Haigis formula by subtracting the predicted refraction of the Haigis formula from the predicted refraction of each 3rd generation formula. Post-operative anterior chamber depth (ACD) was measured at 1 month after surgery using the IOL master. We calculated errors of each formula by subtracting predicted from manifest refraction at post-operative 1 month. Correlation analysis was performed between ocular biometry values, formula expectation values, formula errors and absolute formula errors.

Results

Multiple regression analysis revealed that preoperative ACD was the only significant factor for ED prediction in all of the 3rd generation formulas. For mean errors, axial length and post-operative 1-month change of ACD (delta ACD) correlated significantly with the errors in all 3rd generation formulas, but not with errors of the Haigis formula. Median absolute error (MedAE) of the formulas were 0.40 D for the Hoffer Q formula, 0.37 D for the Holladay formula, 0.34 D for the SRK/T formula, and 0.41 D for the Haigis formula. The MAE of the formulas were 0.50 ± 0.47 D for the Hoffer Q formula, 0.50 ± 0.50 D for the Holladay formula, 0.47 ± 0.51 D for the SRK/T formula, and 0.50 ± 0.47 D for the Haigis formula.

Conclusion

Regarding ED between the third generation and Haigis formulas, preoperative ACD demonstrated the greatest influence. Calculating mean absolute errors of the formulas, all IOL formulas showed excellent and comparable accuracy. Post-operative change (delta) of ACD correlated significantly with errors of third generation formulas according to simulated ACD.
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 1/2017

BMC Ophthalmology 1/2017 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

08.03.2021 | Video plus | Ausgabe 4/2021

„Simple limbal epithelial transplantation“ (SLET)

Eine einfache Technik zur Behandlung der kompletten, einseitigen Limbusstammzellinsuffizienz. Videobeitrag

10.02.2021 | Katarakt | Leitthema | Ausgabe 3/2021

Therapie der postoperativen Endophthalmitis

Operieren oder nur injizieren?