The effect of variable arterial transducer level on the accuracy of pulse contour waveform-derived measurements in critically ill patients

verfasst von: Huai-wu He, Da-wei Liu, Yun Long, Xiao-ting Wang, Mei-ling Zhao, Xiao-li Lai

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 5/2016

Einloggen, um Zugang zu erhalten

Abstract

We know that a 10 cm departure from the reference level of pressure transducer position is equal to a 7.5 mmHg change of invasive hemodynamic pressure monitoring in a fluid-filled system. However, the relationship between the site level of a variable arterial pressure transducer and the pulse contour-derived parameters has yet to be established in critically ill patients. Moreover, the related quantitative analysis has never been investigated. Forty-two critically ill patients requiring PiCCO-Plus cardiac output monitoring were prospectively studied. The phlebostatic axis was defined as the zero reference level; the arterial pressure transducer was then vertically adjusted to different positions (+5, +10, +15, +20, −20, −15, −10, −5 cm) of departure from the zero reference site. The pulse contour waveform-derived parameters were recorded at each position. Elevation of the pressure transducer caused significantly positive changes in the continuous cardiac index (+CCI), stroke volume index (+SVI), and stroke volume variation (+SVV), and negative changes in the rate of left ventricular pressure rise during systole (−dP/dtmax), the systemic vascular resistance index (−SVRI), and vice versa. At the 5 cm position, the SVRI changes reached statistical significance with error. At the 10 cm position, the changes in CCI and dP/dtmax reached statistical significance with error, while the change in SVV reached statistical significance at 15 cm. The change rate of CCI was more than 5 % at the 15 cm position and approximately 10 % at the 20 cm position. On average, for every centimeter change of the transducer, there was a corresponding 0.014 L/min/m2 CCI change and 0.36 % change rate, a 1.41 mmHg/s dP/dtmax change and 0.13 % change rate, and a 25 dyne/s/cm5 SVRI change and 1.2 % change rate. The variation of arterial transducer position can result in inaccurate measurement of pulse contour waveform-derived parameters, especially when the transducer’s vertical distance is more than 10 cm from the phlebostatic axis. These findings have clinical implications for continuous hemodynamic monitoring.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Rauen CA, Makic MBF, Bridges E. Practice evidence-based practice habits: transforming research into bedside. Crit Care Nurse. 2009;29:46–59.CrossRefPubMed Rauen CA, Makic MBF, Bridges E. Practice evidence-based practice habits: transforming research into bedside. Crit Care Nurse. 2009;29:46–59.CrossRefPubMed
2.
Zurück zum Zitat Winsor T, Burch G. Phlebostatic level: reference level for venous pressure measurement in man. Proc Soc Exp Biol Med. 1945;58:165–9.CrossRef Winsor T, Burch G. Phlebostatic level: reference level for venous pressure measurement in man. Proc Soc Exp Biol Med. 1945;58:165–9.CrossRef
3.
Zurück zum Zitat McCann UG II, Schiller HJ, Carney DE, Kilpatrick J, Gatto LA, Paskanik AM, Nieman GF. Invasive arterial BP monitoring in trauma and critical care. Chest. 2001;120:1322–6.CrossRefPubMed McCann UG II, Schiller HJ, Carney DE, Kilpatrick J, Gatto LA, Paskanik AM, Nieman GF. Invasive arterial BP monitoring in trauma and critical care. Chest. 2001;120:1322–6.CrossRefPubMed
4.
Zurück zum Zitat Tagami T, Kushimoto S, Tosa R, et al. The precision of PiCCO® measurements in hypothermic post-cardiac arrest patients. Anaesthesia. 2012;67:236–43.CrossRefPubMed Tagami T, Kushimoto S, Tosa R, et al. The precision of PiCCO® measurements in hypothermic post-cardiac arrest patients. Anaesthesia. 2012;67:236–43.CrossRefPubMed
5.
Zurück zum Zitat de Wilde RB, Schreuder JJ, van den Berg PC, Jansen JR. An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery. Anaesthesia. 2007;62:760–8.CrossRefPubMed de Wilde RB, Schreuder JJ, van den Berg PC, Jansen JR. An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery. Anaesthesia. 2007;62:760–8.CrossRefPubMed
6.
Zurück zum Zitat Perner A, Haase N, Wiis J, White JO, Delaney A. Central venous oxygen saturation for the diagnosis of low cardiac output in septic shock patients. Acta Anaesthesiol Scand. 2010;54:98–102.CrossRefPubMed Perner A, Haase N, Wiis J, White JO, Delaney A. Central venous oxygen saturation for the diagnosis of low cardiac output in septic shock patients. Acta Anaesthesiol Scand. 2010;54:98–102.CrossRefPubMed
7.
Zurück zum Zitat Scolletta Sabino, Bodson Laurent, Donadello Katia, Taccone Fabio S, Devigili Alessandro, Vincent Jean-Louis, De Backer Daniel. Assessment of left ventricular function by pulse wave analysis in critically ill patients. Intensive Care Med. 2013;39:1025–33.CrossRefPubMed Scolletta Sabino, Bodson Laurent, Donadello Katia, Taccone Fabio S, Devigili Alessandro, Vincent Jean-Louis, De Backer Daniel. Assessment of left ventricular function by pulse wave analysis in critically ill patients. Intensive Care Med. 2013;39:1025–33.CrossRefPubMed
8.
Zurück zum Zitat He H, Liu D, Long Y, Wang X, Yuan Y, Li X, Guo H, Cai J, Fang N. The relationship between arterial transducer level and pulse contour waveform-derived measurements. Crit Care. 2015;19:31.CrossRefPubMedPubMedCentral He H, Liu D, Long Y, Wang X, Yuan Y, Li X, Guo H, Cai J, Fang N. The relationship between arterial transducer level and pulse contour waveform-derived measurements. Crit Care. 2015;19:31.CrossRefPubMedPubMedCentral
9.
Zurück zum Zitat Knaus WA, Draper EA, Wagner DP, Zimmerman JE. APACHE II: a severity of disease classification system. Crit Care Med. 1985;13:818–8295.CrossRefPubMed Knaus WA, Draper EA, Wagner DP, Zimmerman JE. APACHE II: a severity of disease classification system. Crit Care Med. 1985;13:818–8295.CrossRefPubMed
10.
Zurück zum Zitat Kleinman B, Powell S, Kumar P, Gardner RM. The fast flush test measures the dynamic response of the entire blood pressure monitoring system. Anesthesiology. 1992;77:1215–20.CrossRefPubMed Kleinman B, Powell S, Kumar P, Gardner RM. The fast flush test measures the dynamic response of the entire blood pressure monitoring system. Anesthesiology. 1992;77:1215–20.CrossRefPubMed
11.
Zurück zum Zitat Watanabe H, Yagi S, Namiki A. Recommendation of a clinical impulse response analysis for catheter calibration-dumping coefficient and natural frequency are incomplete parameters for clinical evaluation. J Clin Monit Comput. 2006;20:37–42.CrossRefPubMed Watanabe H, Yagi S, Namiki A. Recommendation of a clinical impulse response analysis for catheter calibration-dumping coefficient and natural frequency are incomplete parameters for clinical evaluation. J Clin Monit Comput. 2006;20:37–42.CrossRefPubMed
12.
Zurück zum Zitat Roy A. Estimating correlation coefficient between two variables with repeated observations using mixed effects model. Biom J. 2006;48:286–301.CrossRefPubMed Roy A. Estimating correlation coefficient between two variables with repeated observations using mixed effects model. Biom J. 2006;48:286–301.CrossRefPubMed
13.
Zurück zum Zitat McCulloch CE, Searle SR. Generalized, linear, and mixed models. London: Wiley; 2000.CrossRef McCulloch CE, Searle SR. Generalized, linear, and mixed models. London: Wiley; 2000.CrossRef
14.
Zurück zum Zitat Wesseling KH, de Wit B, Weber JAP, Smith NT. A simple device for the continuous measurement of cardiac output. Adv Cardiovasc Phys. 1983;5:16–52. Wesseling KH, de Wit B, Weber JAP, Smith NT. A simple device for the continuous measurement of cardiac output. Adv Cardiovasc Phys. 1983;5:16–52.
15.
Zurück zum Zitat Wesseling KH, Jansen JRC, Settels JJ, Schreuder JJ. Computation of aortic flow from pressure in humans using a nonlinear three-element model. J Appl Physiol. 1984;74(5):2566–73. Wesseling KH, Jansen JRC, Settels JJ, Schreuder JJ. Computation of aortic flow from pressure in humans using a nonlinear three-element model. J Appl Physiol. 1984;74(5):2566–73.
16.
Zurück zum Zitat Wesseling, KH, Smith, NT, Nichols, WW et al. Beat to beat cardiac output from the arterial pressure pulse contour. In: Boerhave, SJ, Course on measurement in anaesthesia. Leiden: University of Leiden Press; 1974. pp. 150–64. Wesseling, KH, Smith, NT, Nichols, WW et al. Beat to beat cardiac output from the arterial pressure pulse contour. In: Boerhave, SJ, Course on measurement in anaesthesia. Leiden: University of Leiden Press; 1974. pp. 150–64.
17.
Zurück zum Zitat Godje O, Hoke K, Goetz AE, Felbinger TW, Reuter DA, Reichart B, Friedl R, Hannekum A, Pfeiffer UJ. Reliability of a new algorithm for continuous cardiac output determination by pulse-contour analysis during hemodynamic instability. Crit Care Med. 2002;30:52–8.CrossRefPubMed Godje O, Hoke K, Goetz AE, Felbinger TW, Reuter DA, Reichart B, Friedl R, Hannekum A, Pfeiffer UJ. Reliability of a new algorithm for continuous cardiac output determination by pulse-contour analysis during hemodynamic instability. Crit Care Med. 2002;30:52–8.CrossRefPubMed
18.
Zurück zum Zitat Monnet X, Anguel N, Naudin B, Jabot J, Richard C, Teboul JL. Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices. Crit Care. 2010;14(3):R109.CrossRefPubMedPubMedCentral Monnet X, Anguel N, Naudin B, Jabot J, Richard C, Teboul JL. Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices. Crit Care. 2010;14(3):R109.CrossRefPubMedPubMedCentral
19.
Zurück zum Zitat Parry T, Hirsch N, Fauvel N. Comparison of direct blood pressure measurement at the radial and dorsalis pedis arteries during surgery in the horizontal and reverse Trendelenburg positions. Anaesthesia. 1995;50:553–5.CrossRefPubMed Parry T, Hirsch N, Fauvel N. Comparison of direct blood pressure measurement at the radial and dorsalis pedis arteries during surgery in the horizontal and reverse Trendelenburg positions. Anaesthesia. 1995;50:553–5.CrossRefPubMed
20.
Zurück zum Zitat Chulay M, Holland S. Where should the transducer be leveled for radial or femoral arterial pressure monitoring? Crit Care Nurs. 1996;16:103–7. Chulay M, Holland S. Where should the transducer be leveled for radial or femoral arterial pressure monitoring? Crit Care Nurs. 1996;16:103–7.
21.
Zurück zum Zitat Ahrens TS, Taylor LA. Technical considerations in obtaining hemodynamic waveform values. In: Darovic GO, editor. Hemodynamic waveform analysis. Philadelphia: W.B. Saunders press; 1992. p. 222. Ahrens TS, Taylor LA. Technical considerations in obtaining hemodynamic waveform values. In: Darovic GO, editor. Hemodynamic waveform analysis. Philadelphia: W.B. Saunders press; 1992. p. 222.
22.
Zurück zum Zitat Courtois M, Fattal PG, Kovacs SJ, et al. Anatomically and physiologically based reference level for measurement of intracardiac pressures. Circulation. 1995;92:1994–2000.CrossRefPubMed Courtois M, Fattal PG, Kovacs SJ, et al. Anatomically and physiologically based reference level for measurement of intracardiac pressures. Circulation. 1995;92:1994–2000.CrossRefPubMed
23.
Zurück zum Zitat Whitman G, Howaniak D, Verga T. Comparison of cardiac output measurements in 20-degree supine and 20-degree right and left lateral recumbent positions. Heart Lung. 1982;11:256–7. Whitman G, Howaniak D, Verga T. Comparison of cardiac output measurements in 20-degree supine and 20-degree right and left lateral recumbent positions. Heart Lung. 1982;11:256–7.
24.
Zurück zum Zitat Dobbin K, Wallace S, Ahlberg J, Chulay M. Pulmonary artery pressure measurement in patients with elevated pressures: effect of backrest elevation and method of measurement. Am J Crit Care. 1992;1:61–9.PubMed Dobbin K, Wallace S, Ahlberg J, Chulay M. Pulmonary artery pressure measurement in patients with elevated pressures: effect of backrest elevation and method of measurement. Am J Crit Care. 1992;1:61–9.PubMed
25.
Zurück zum Zitat Wilson AE, Bermingham-Mitchell K, Wells N, Zachary K. Effect of backrest position on hemodynamic and right ventricular measurements in critically ill adults. Am J Crit Care. 1996;5:264–70.PubMed Wilson AE, Bermingham-Mitchell K, Wells N, Zachary K. Effect of backrest position on hemodynamic and right ventricular measurements in critically ill adults. Am J Crit Care. 1996;5:264–70.PubMed
26.
Zurück zum Zitat Giuliano KK, Scott SS, Brown V, Olson M. Backrest angle and cardiac output measurement in critically ill patients. Nurs Res. 2003;52:242–8.CrossRefPubMed Giuliano KK, Scott SS, Brown V, Olson M. Backrest angle and cardiac output measurement in critically ill patients. Nurs Res. 2003;52:242–8.CrossRefPubMed
27.
Zurück zum Zitat Grose BL, Woods SL, Laurent DJ. Effect of backrest position on cardiac output measured by the thermodilution method in acutely ill patients. Heart Lung. 1981;10:661–5.PubMed Grose BL, Woods SL, Laurent DJ. Effect of backrest position on cardiac output measured by the thermodilution method in acutely ill patients. Heart Lung. 1981;10:661–5.PubMed
28.
Zurück zum Zitat Monnet X, Bataille A, Magalhaes E, Barrois J, Le Corre M, Gosset C, Guerin L, Richard C, Teboul JL. End-tidal carbon dioxide is better than arterial pressure for predicting volume responsiveness by the passive leg raising test. Intensive Care Med. 2013;39:93–100.CrossRefPubMed Monnet X, Bataille A, Magalhaes E, Barrois J, Le Corre M, Gosset C, Guerin L, Richard C, Teboul JL. End-tidal carbon dioxide is better than arterial pressure for predicting volume responsiveness by the passive leg raising test. Intensive Care Med. 2013;39:93–100.CrossRefPubMed
29.
Zurück zum Zitat Jabot J, Teboul JL, Richard C, Monnet X. Passive leg raising for predicting fluid responsiveness: importance of the postural change. Intensive Care Med. 2009;35:85–90.CrossRefPubMed Jabot J, Teboul JL, Richard C, Monnet X. Passive leg raising for predicting fluid responsiveness: importance of the postural change. Intensive Care Med. 2009;35:85–90.CrossRefPubMed
30.
Zurück zum Zitat He HW, Liu DW. Passive leg raising: influence of blood pressure transducer site. Intensive Care Med. 2013;39(9):1668.CrossRefPubMed He HW, Liu DW. Passive leg raising: influence of blood pressure transducer site. Intensive Care Med. 2013;39(9):1668.CrossRefPubMed
Metadaten
Titel
The effect of variable arterial transducer level on the accuracy of pulse contour waveform-derived measurements in critically ill patients
verfasst von
Huai-wu He
Da-wei Liu
Yun Long
Xiao-ting Wang
Mei-ling Zhao
Xiao-li Lai
Publikationsdatum
23.08.2015
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 5/2016
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-015-9756-x

Weitere Artikel der Ausgabe 5/2016

Journal of Clinical Monitoring and Computing 5/2016 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.