Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology 3/2021

20.01.2021 | Cataract

The impact of visual axis position on the optical quality after implantation of multifocal intraocular lenses with different asphericity values

verfasst von: Kata Miháltz, Pia Veronika Vécsei-Marlovits

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 3/2021

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To evaluate the effect of visual axis positioning on the optical performance of the Tecnis MIOL and the Diff-aA MIOL.

Methods

In this prospective, randomized comparative study, 70 eyes of 35 subjects with senile cataract were implanted with the spherical aberration–correcting diffractive, bifocal Tecnis ZLB00 IOL and 60 eyes of 30 age-matched subjects with the spherical aberration neutral, diffractive, bifocal Diffractiva IOL. Observation procedure was performed 1, 3, and 6 months postoperatively. Main outcome measures included uncorrected and corrected distance and near visual acuity, manifest refraction, ocular aberrations, and visual quality metrics with 2 mm and 4 mm pupil and the position of visual axis.

Results

At the 6-month visit, no significant difference was found in monocular and binocular uncorrected (UDVA) and corrected (CDVA) distance and near (UNVA, CNVA) visual acuity between the groups. Spherical and coma-like aberrations were similar measured with a 2-mm pupil, but with a 4-mm pupil, the SA was significantly larger (in negative direction) in the Diffractiva group. The higher-order Strehl ratio and MTF was significantly larger in the Diffractiva group measured at 2 mm entrance pupil; however, this difference disappeared by the 4-mm pupil measurements. Postoperative angle alpha distance had a significant influence on HO Strehl value.

Conclusions

The size of angle alpha is a predictive factor of image quality by multifocal IOL patients.

Trial Registration

Trial registration number and date of registration: NCT04274088, 14.02.2020.
Literatur
1.
Zurück zum Zitat Berrio E, Tabernero J, Artal P (2010) Optical aberrations and alignment of the eye with age. J Vis 10:34CrossRef Berrio E, Tabernero J, Artal P (2010) Optical aberrations and alignment of the eye with age. J Vis 10:34CrossRef
2.
Zurück zum Zitat Marcos S, Rosales P, Llorente L, Barbero S, Jiménez-Alfaro I (2008) Balance of corneal horizontal coma by internal optics in eyes with intraocular artificial lenses: evidence of a passive mechanism. Vis Res 48:70–79CrossRef Marcos S, Rosales P, Llorente L, Barbero S, Jiménez-Alfaro I (2008) Balance of corneal horizontal coma by internal optics in eyes with intraocular artificial lenses: evidence of a passive mechanism. Vis Res 48:70–79CrossRef
3.
Zurück zum Zitat Yoon SY, Song IS, Kim JY, Kim MJ, Tchah H (2013) Bilateral mix-and-match versus unilateral multifocal intraocular lens implantation: long-term comparison. J Cataract Refract Surg 39:1682–1690CrossRef Yoon SY, Song IS, Kim JY, Kim MJ, Tchah H (2013) Bilateral mix-and-match versus unilateral multifocal intraocular lens implantation: long-term comparison. J Cataract Refract Surg 39:1682–1690CrossRef
4.
Zurück zum Zitat Rabsilber TM, Rudalevicius P, Jasinskas V, Holzer MP, Auffarth GU (2013) Influence of +3.00 D and +4.00 D near addition on functional outcomes of a refractive multifocal intraocular lens model. J Cataract Refract Surg 39:350–357CrossRef Rabsilber TM, Rudalevicius P, Jasinskas V, Holzer MP, Auffarth GU (2013) Influence of +3.00 D and +4.00 D near addition on functional outcomes of a refractive multifocal intraocular lens model. J Cataract Refract Surg 39:350–357CrossRef
5.
Zurück zum Zitat Prakash G, Agarwal A, Prakash DR, Kumar DA, Agarwal A, Jacob S (2011) Role of angle kappa in patient dissatisfaction with refractive-design multifocal intraocular lenses. J Cataract Refract Surg 37:1739–1740CrossRef Prakash G, Agarwal A, Prakash DR, Kumar DA, Agarwal A, Jacob S (2011) Role of angle kappa in patient dissatisfaction with refractive-design multifocal intraocular lenses. J Cataract Refract Surg 37:1739–1740CrossRef
6.
Zurück zum Zitat Prakash G, Prakash DR, Agarwal A, Kumar DA, Agarwal A, Jacob S (2011) Predictive factor and kappa angle analysis for visual satisfactions in patients with multifocal IOL implantation. Eye 25:1187–1193CrossRef Prakash G, Prakash DR, Agarwal A, Kumar DA, Agarwal A, Jacob S (2011) Predictive factor and kappa angle analysis for visual satisfactions in patients with multifocal IOL implantation. Eye 25:1187–1193CrossRef
7.
Zurück zum Zitat Kim JW, Eom Y, Chung HW (2020) Factors for good near and distance visual outcomes of multifocal intraocular lens with inferior segmental near add. Graefes Arch Clin Exp Ophthalmol 258:1735–1743CrossRef Kim JW, Eom Y, Chung HW (2020) Factors for good near and distance visual outcomes of multifocal intraocular lens with inferior segmental near add. Graefes Arch Clin Exp Ophthalmol 258:1735–1743CrossRef
9.
Zurück zum Zitat Eom Y, Yoo E, Kang SY, Kim HM, Song JS (2013) Change in efficiency of aspheric intraocular lenses based on pupil diameter. Am J Ophthalmol 155:492–498CrossRef Eom Y, Yoo E, Kang SY, Kim HM, Song JS (2013) Change in efficiency of aspheric intraocular lenses based on pupil diameter. Am J Ophthalmol 155:492–498CrossRef
10.
Zurück zum Zitat Eppig T, Scholz K, Langenbucher A (2008) Assessing the optical performance of multifocal (diffractive) intraocular lenses. Ophthalmic Physiol Opt 28:467–474CrossRef Eppig T, Scholz K, Langenbucher A (2008) Assessing the optical performance of multifocal (diffractive) intraocular lenses. Ophthalmic Physiol Opt 28:467–474CrossRef
11.
Zurück zum Zitat Kumar DA, Agarwal A, Prakash G, Jacob S, Saravanan Y, Agarwal A (2011) Evaluation of intraocular lens tilt with anterior segment optical coherence tomography. Am J Ophthalmol 151:406–412CrossRef Kumar DA, Agarwal A, Prakash G, Jacob S, Saravanan Y, Agarwal A (2011) Evaluation of intraocular lens tilt with anterior segment optical coherence tomography. Am J Ophthalmol 151:406–412CrossRef
12.
Zurück zum Zitat Piñero DP, Sánchez-Pérez PJ, Alió JL (2011) Repeatability of measurements obtained with a ray tracing aberrometer. Optom Vis Sci 88:1099–1105CrossRef Piñero DP, Sánchez-Pérez PJ, Alió JL (2011) Repeatability of measurements obtained with a ray tracing aberrometer. Optom Vis Sci 88:1099–1105CrossRef
13.
Zurück zum Zitat Jun I, Choi YJ, Kim EK, Seo KY, Kim TI (2012) Internal spherical aberration by ray tracing-type aberrometry in multifocal pseudophakic eyes. Eye 26:1243–1248CrossRef Jun I, Choi YJ, Kim EK, Seo KY, Kim TI (2012) Internal spherical aberration by ray tracing-type aberrometry in multifocal pseudophakic eyes. Eye 26:1243–1248CrossRef
14.
Zurück zum Zitat Lombardo M, Lombardo G (2010) Wave aberration of human eyes and new descriptors of image optical quality and visual performance. J Cataract Refract Surg 36:313–331CrossRef Lombardo M, Lombardo G (2010) Wave aberration of human eyes and new descriptors of image optical quality and visual performance. J Cataract Refract Surg 36:313–331CrossRef
15.
Zurück zum Zitat Grzybowski A, Kanclerz P, Muzyka-Woźniak M (2019) Methods for evaluating quality of life and vision in patients undergoing lens refractive surgery. Graefes Arch Clin Exp Ophthalmol 257(6):1091–1099 Grzybowski A, Kanclerz P, Muzyka-Woźniak M (2019) Methods for evaluating quality of life and vision in patients undergoing lens refractive surgery. Graefes Arch Clin Exp Ophthalmol 257(6):1091–1099
16.
Zurück zum Zitat Labuz G, Reus NJ, van den Berg TJ (2016) Comparison of ocular straylight after implantation of multifocal intraocular lenses. J Cataract Refract Surg 42:618–625CrossRef Labuz G, Reus NJ, van den Berg TJ (2016) Comparison of ocular straylight after implantation of multifocal intraocular lenses. J Cataract Refract Surg 42:618–625CrossRef
17.
Zurück zum Zitat Puell MC, Pérez-Carrasco MJ, Hurtado-Ceña FJ, Álvarez-Rementería L (2015) Disk halo size measured in individuals with monofocal versus diffractive multifocal intraocular lenses. J Cataract Refract Surg 41:2417–2423CrossRef Puell MC, Pérez-Carrasco MJ, Hurtado-Ceña FJ, Álvarez-Rementería L (2015) Disk halo size measured in individuals with monofocal versus diffractive multifocal intraocular lenses. J Cataract Refract Surg 41:2417–2423CrossRef
19.
Zurück zum Zitat Ravikumar S, Bradley A, Thibos LN (2014) Chromatic aberration and polychromatic image quality with diffractive multifocal intraocular lenses. J Cataract Refract Surg 40:1192–1204CrossRef Ravikumar S, Bradley A, Thibos LN (2014) Chromatic aberration and polychromatic image quality with diffractive multifocal intraocular lenses. J Cataract Refract Surg 40:1192–1204CrossRef
20.
Zurück zum Zitat Vega F, Alba-Bueno F, Millán MS, Varón C, Gil MA, Buil JA (2015) Halo and through-focus performance of four diffractive multifocal intraocular lenses. Invest Ophthalmol Vis Sci 56:3967–3975CrossRef Vega F, Alba-Bueno F, Millán MS, Varón C, Gil MA, Buil JA (2015) Halo and through-focus performance of four diffractive multifocal intraocular lenses. Invest Ophthalmol Vis Sci 56:3967–3975CrossRef
21.
Zurück zum Zitat Santhiago MR, Wilson SE, Netto MV, Ghanen RC, Monteiro ML, Bechara SJ, Espana EM, Mello GR, Kara N Jr (2012) Modulation transfer function and optical quality after bilateral implantation of a +3.00 D versus a +4.00 D multifocal intraocular lens. J Cataract Refract Surg 38:215–220CrossRef Santhiago MR, Wilson SE, Netto MV, Ghanen RC, Monteiro ML, Bechara SJ, Espana EM, Mello GR, Kara N Jr (2012) Modulation transfer function and optical quality after bilateral implantation of a +3.00 D versus a +4.00 D multifocal intraocular lens. J Cataract Refract Surg 38:215–220CrossRef
22.
Zurück zum Zitat Castillo-Gómez A, Carmona-González D, Martínez-de-la-Casa JM, Palomino-Bautista C, García-Feijoo J (2009) Evaluation of image quality after implantation of 2 diffractive multifocal intraocular lens models. J Cataract Refract Surg 35:1244–1250CrossRef Castillo-Gómez A, Carmona-González D, Martínez-de-la-Casa JM, Palomino-Bautista C, García-Feijoo J (2009) Evaluation of image quality after implantation of 2 diffractive multifocal intraocular lens models. J Cataract Refract Surg 35:1244–1250CrossRef
23.
Zurück zum Zitat Alba-Bueno F, Vega F, Millán MS (2014) Halos and multifocal intraocular lenses: origin and interpretation. Arch Soc Esp Oftalmol 89:397–404CrossRef Alba-Bueno F, Vega F, Millán MS (2014) Halos and multifocal intraocular lenses: origin and interpretation. Arch Soc Esp Oftalmol 89:397–404CrossRef
24.
Zurück zum Zitat Yamaguchi T, Negishi K, Ono T, Torii H, Dogru M, Yamaguchi K, Ohnuma K, Tsubota K (2009) Feasibility of spherical aberration correction with aspheric intraocular lenses in cataract surgery based on individual pupil diameter. J Cataract Refract Surg 35:1725–1733CrossRef Yamaguchi T, Negishi K, Ono T, Torii H, Dogru M, Yamaguchi K, Ohnuma K, Tsubota K (2009) Feasibility of spherical aberration correction with aspheric intraocular lenses in cataract surgery based on individual pupil diameter. J Cataract Refract Surg 35:1725–1733CrossRef
25.
Zurück zum Zitat Liu JP, Zhang F, Zhao JY, Ma LW, Zhang JS (2013) Visual function and higher order aberration after implantation of aspheric and spherical multifocal intraocular lenses: a meta-analysis. Int J Ophthalmol 6:690–695PubMedPubMedCentral Liu JP, Zhang F, Zhao JY, Ma LW, Zhang JS (2013) Visual function and higher order aberration after implantation of aspheric and spherical multifocal intraocular lenses: a meta-analysis. Int J Ophthalmol 6:690–695PubMedPubMedCentral
26.
Zurück zum Zitat Ortiz D, Alió JL, Bernabéu G, Pongo V (2008) Optical performance of monofocal and multifocal intraocular lenses in the human eye. J Cataract Refract Surg 34:755–762CrossRef Ortiz D, Alió JL, Bernabéu G, Pongo V (2008) Optical performance of monofocal and multifocal intraocular lenses in the human eye. J Cataract Refract Surg 34:755–762CrossRef
27.
Zurück zum Zitat Alio JL, Plaza-Puche AB, Férnandez-Buenaga R, Pikkel J, Maldonado M (2017) Multifocal intraocular lenses: an overview. Surv Ophthalmol 62:611–634CrossRef Alio JL, Plaza-Puche AB, Férnandez-Buenaga R, Pikkel J, Maldonado M (2017) Multifocal intraocular lenses: an overview. Surv Ophthalmol 62:611–634CrossRef
28.
Zurück zum Zitat Liu JW, Haw WW (2014) Optimizing outcomes of multifocal intraocular lenses. Curr Opin Ophthalmol 25:44–48CrossRef Liu JW, Haw WW (2014) Optimizing outcomes of multifocal intraocular lenses. Curr Opin Ophthalmol 25:44–48CrossRef
29.
Zurück zum Zitat Ortiz C, Esteve-Taboada JJ, Belda-Salmerón L, Monsálvez-Romín D, Domínguez-Vicent (2016) Effect of decentration on the optical quality of two intraocular lenses. Optom Vis Sci 93:1552–1559CrossRef Ortiz C, Esteve-Taboada JJ, Belda-Salmerón L, Monsálvez-Romín D, Domínguez-Vicent (2016) Effect of decentration on the optical quality of two intraocular lenses. Optom Vis Sci 93:1552–1559CrossRef
30.
Zurück zum Zitat Karhanová M, Pluháček F, Mlčák P, Vláčil O, Šín M, Marešová K (2015) The importance of angle kappa evaluation for implantation of diffractive multifocal intra-ocular lenses using pseudophakic eye model. Acta Ophthalmol 93:e123–e128CrossRef Karhanová M, Pluháček F, Mlčák P, Vláčil O, Šín M, Marešová K (2015) The importance of angle kappa evaluation for implantation of diffractive multifocal intra-ocular lenses using pseudophakic eye model. Acta Ophthalmol 93:e123–e128CrossRef
31.
Zurück zum Zitat Tchah H, Nam K, Yoo A (2017) Predictive factors for photic phenomena after refractive, rotationally asymmetric, multifocal intraocular lens implantation. Int J Ophthalmol 10:241–245PubMedPubMedCentral Tchah H, Nam K, Yoo A (2017) Predictive factors for photic phenomena after refractive, rotationally asymmetric, multifocal intraocular lens implantation. Int J Ophthalmol 10:241–245PubMedPubMedCentral
32.
Zurück zum Zitat Fischinger I, Seiler TG, Schmidinger G, Seiler T (2015) Pupil centroid shift: marketing tool or essential clinical parameter? Ophthalmologe 112:661–664CrossRef Fischinger I, Seiler TG, Schmidinger G, Seiler T (2015) Pupil centroid shift: marketing tool or essential clinical parameter? Ophthalmologe 112:661–664CrossRef
33.
Zurück zum Zitat Guillon M, Dumbleton K, Theodoratos P, Gobbe M, Wooley CB, Moody K (2016) The effects of age, refractive status, and luminance on pupil size. Optom Vis Sci 93:1093–1100CrossRef Guillon M, Dumbleton K, Theodoratos P, Gobbe M, Wooley CB, Moody K (2016) The effects of age, refractive status, and luminance on pupil size. Optom Vis Sci 93:1093–1100CrossRef
Metadaten
Titel
The impact of visual axis position on the optical quality after implantation of multifocal intraocular lenses with different asphericity values
verfasst von
Kata Miháltz
Pia Veronika Vécsei-Marlovits
Publikationsdatum
20.01.2021
Verlag
Springer Berlin Heidelberg
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 3/2021
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-020-05052-5

Weitere Artikel der Ausgabe 3/2021

Graefe's Archive for Clinical and Experimental Ophthalmology 3/2021 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.