The John Charnley Award: Risk Factors for Cup Malpositioning: Quality Improvement Through a Joint Registry at a Tertiary Hospital

verfasst von: Mark C. Callanan, MA, Bryan Jarrett, BS, Charles R. Bragdon, PhD, David Zurakowski, PhD, Harry E. Rubash, MD, Andrew A. Freiberg, MD, Henrik Malchau, MD, PhD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 2/2011

Einloggen, um Zugang zu erhalten

Abstract

Background

Few studies have examined factors that affect acetabular cup positioning. Since cup positioning has been linked to dislocation and increased bearing surface wear, these factors affecting cup position are important considerations.

Question/purposes

We determined the percent of optimally positioned acetabular cups and whether patient and surgical factors affected acetabular component position.

Methods

We obtained postoperative AP pelvis and cross-table lateral radiographs on 2061 consecutive patients who received a THA or hip resurfacing from 2004 to 2008. One thousand nine hundred and fifty-two hips had AP pelvic radiographs with correct position of the hip center, and 1823 had both version and abduction angles measured. The AP radiograph was measured using Hip Analysis Suite™ to calculate the cup inclination and version angles, using the lateral film to determine version direction. Acceptable ranges were defined for abduction (30°–45°) and version (5°–25°).

Results

From the 1823 hips, 1144 (63%) acetabular cups were within the abduction range, 1441 (79%) were within the version range, and 917 (50%) were within the range for both. Surgical approach, surgeon volume, and obesity (body mass index > 30) independently predicted malpositioned cups. Comparison of low versus high volume surgeons, minimally invasive surgical versus posterolateral approach, and obesity versus all other body mass index groups showed a twofold (1.5–2.8), sixfold (3.5–10.7), and 1.3-fold (1.1–1.7) increased risk for malpositioned cups, respectively.

Conclusions

Factors correlated to malpositioned cups included surgical approach, surgeon volume, and body mass index with increased risk of malpositioning for minimally invasive surgical approach, low volume surgeons, and obese patients. Further analyses on patient and surgical factors’ influence on cup position at a lower volume medical center would provide a valuable comparison.

Level of Evidence

Level II, prognostic study. See Guidelines for Authors for a complete description of levels of evidence.
Literatur
1.
Zurück zum Zitat Ali Khan MA, Brakenbury PH, IS R. Dislocation following total hip replacement. J Bone Joint Surg Br. 1981;63:214–218.PubMed Ali Khan MA, Brakenbury PH, IS R. Dislocation following total hip replacement. J Bone Joint Surg Br. 1981;63:214–218.PubMed
2.
Zurück zum Zitat Arthursson AJ, Furnes O, Espehaug B, Havelin LI, Soreide JA. Prosthesis survival after total hip arthroplasty—does surgical approach matter? Analysis of 19,304 Charnley and 6,002 Exeter primary total hip arthroplasties reported to the Norwegian Arthroplasty Register. Acta Orthop. 2007;78:719–729.CrossRefPubMed Arthursson AJ, Furnes O, Espehaug B, Havelin LI, Soreide JA. Prosthesis survival after total hip arthroplasty—does surgical approach matter? Analysis of 19,304 Charnley and 6,002 Exeter primary total hip arthroplasties reported to the Norwegian Arthroplasty Register. Acta Orthop. 2007;78:719–729.CrossRefPubMed
3.
Zurück zum Zitat Bartz RL, Noble PC, Kadakia NR, Tullos HS. The effect of femoral component head size on posterior dislocation of the artificial hip joint. J Bone Joint Surg Am. 2000;82:1300–1307.PubMed Bartz RL, Noble PC, Kadakia NR, Tullos HS. The effect of femoral component head size on posterior dislocation of the artificial hip joint. J Bone Joint Surg Am. 2000;82:1300–1307.PubMed
4.
Zurück zum Zitat Biedermann R, Tonin A, Krismer M, Rachbauer F, Eibl G, Stockl B. Reducing the risk of dislocation after total hip arthroplasty: the effect of orientation of the acetabular component. J Bone Joint Surg Br. 2005;87:762–769.CrossRefPubMed Biedermann R, Tonin A, Krismer M, Rachbauer F, Eibl G, Stockl B. Reducing the risk of dislocation after total hip arthroplasty: the effect of orientation of the acetabular component. J Bone Joint Surg Br. 2005;87:762–769.CrossRefPubMed
5.
Zurück zum Zitat Bosker BH, Verheyen CC, Horstmann WG, Tulp NJ. Poor accuracy of freehand cup positioning during total hip arthroplasty. Arch Orthop Trauma Surg. 2007;127:375–379.CrossRefPubMed Bosker BH, Verheyen CC, Horstmann WG, Tulp NJ. Poor accuracy of freehand cup positioning during total hip arthroplasty. Arch Orthop Trauma Surg. 2007;127:375–379.CrossRefPubMed
6.
Zurück zum Zitat Bradford L, Kurland R, Sankaran M, Kim H, Pruitt LA, Ries MD. Early failure due to osteolysis associated with contemporary highly cross-linked ultra-high molecular weight polyethylene. A case report. J Bone Joint Surg Am. 2004;86:1051–1056. Bradford L, Kurland R, Sankaran M, Kim H, Pruitt LA, Ries MD. Early failure due to osteolysis associated with contemporary highly cross-linked ultra-high molecular weight polyethylene. A case report. J Bone Joint Surg Am. 2004;86:1051–1056.
7.
Zurück zum Zitat Bragdon CR, Martell JM, Greene ME, Estok DM II, Thanner J, Karrholm J, Harris WH, Malchau H. Comparison of femoral head penetration using RSA and the Martell method. Clin Orthop Relat Res. 2006;448:52–57.CrossRefPubMed Bragdon CR, Martell JM, Greene ME, Estok DM II, Thanner J, Karrholm J, Harris WH, Malchau H. Comparison of femoral head penetration using RSA and the Martell method. Clin Orthop Relat Res. 2006;448:52–57.CrossRefPubMed
8.
Zurück zum Zitat Broers H, Jansing N. How precise is navigation for minimally invasive surgery? Int Orthop. 2007;31 Suppl 1:S39–S42.CrossRefPubMed Broers H, Jansing N. How precise is navigation for minimally invasive surgery? Int Orthop. 2007;31 Suppl 1:S39–S42.CrossRefPubMed
9.
Zurück zum Zitat Colwell CWJ. Instability after total hip arthroplasty. Current Orthop Prac. 2009;20:8–14.CrossRef Colwell CWJ. Instability after total hip arthroplasty. Current Orthop Prac. 2009;20:8–14.CrossRef
10.
Zurück zum Zitat Conroy JL, Whitehouse SL, Graves SE, Pratt NL, Ryan P, Crawford RW. Risk factors for revision for early dislocation in total hip arthroplasty. J Arthroplasty. 2008;23:867–872.CrossRefPubMed Conroy JL, Whitehouse SL, Graves SE, Pratt NL, Ryan P, Crawford RW. Risk factors for revision for early dislocation in total hip arthroplasty. J Arthroplasty. 2008;23:867–872.CrossRefPubMed
11.
Zurück zum Zitat De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K. Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br. 2008;90:1291–1297.CrossRefPubMed De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K. Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br. 2008;90:1291–1297.CrossRefPubMed
12.
Zurück zum Zitat Dorr LD. Acetabular cup position: the imperative of getting it right. Orthopedics. 2008;31:898–899.CrossRefPubMed Dorr LD. Acetabular cup position: the imperative of getting it right. Orthopedics. 2008;31:898–899.CrossRefPubMed
13.
Zurück zum Zitat Enocson ATJ, Tornkvist H, Lapidus L. Dislocation of hemiarthroplasty after femoral neck fracture: better outcome after the anterolateral approach in a prospective cohort study on 739 consecutive hips. Acta Orthop. 2008;79:211–217.CrossRefPubMed Enocson ATJ, Tornkvist H, Lapidus L. Dislocation of hemiarthroplasty after femoral neck fracture: better outcome after the anterolateral approach in a prospective cohort study on 739 consecutive hips. Acta Orthop. 2008;79:211–217.CrossRefPubMed
14.
Zurück zum Zitat Gallo J, Havranek V, Zapletalova J. Risk factors for accelerated polyethylene wear and osteolysis in ABG I total hip arthroplasty. Int Orthop. 2010;34:19–26.CrossRefPubMed Gallo J, Havranek V, Zapletalova J. Risk factors for accelerated polyethylene wear and osteolysis in ABG I total hip arthroplasty. Int Orthop. 2010;34:19–26.CrossRefPubMed
15.
Zurück zum Zitat Ghelman B, Kepler CK, Lyman S, Della Valle AG. CT outperforms radiography for determination of acetabular cup version after THA. Clin Orthop Relat Res. 2009;467:2362–2370.CrossRefPubMed Ghelman B, Kepler CK, Lyman S, Della Valle AG. CT outperforms radiography for determination of acetabular cup version after THA. Clin Orthop Relat Res. 2009;467:2362–2370.CrossRefPubMed
16.
Zurück zum Zitat Holt G, Murnaghan C, Reilly J, Meek RM. The biology of aseptic osteolysis. Clin Orthop Relat Res. 2007;460:240–252.PubMed Holt G, Murnaghan C, Reilly J, Meek RM. The biology of aseptic osteolysis. Clin Orthop Relat Res. 2007;460:240–252.PubMed
17.
Zurück zum Zitat Hui AJ, McCalden RW, Martell JM, MacDonald SJ, Bourne RB, Rorabeck CH. Validation of two and three-dimensional radiographic techniques for measuring polyethylene wear after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:505–511.PubMed Hui AJ, McCalden RW, Martell JM, MacDonald SJ, Bourne RB, Rorabeck CH. Validation of two and three-dimensional radiographic techniques for measuring polyethylene wear after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:505–511.PubMed
18.
Zurück zum Zitat Jasty M, Goetz DD, Bragdon CR, Lee KR, Hanson AE, Elder JR, Harris WH. Wear of polyethylene acetabular components in total hip arthroplasty. An analysis of one hundred and twenty-eight components retrieved at autopsy or revision operations. J Bone Joint Surg Am. 1997;79:349–358.CrossRefPubMed Jasty M, Goetz DD, Bragdon CR, Lee KR, Hanson AE, Elder JR, Harris WH. Wear of polyethylene acetabular components in total hip arthroplasty. An analysis of one hundred and twenty-eight components retrieved at autopsy or revision operations. J Bone Joint Surg Am. 1997;79:349–358.CrossRefPubMed
19.
Zurück zum Zitat Jolles BM, Genoud P, Hoffmeyer P. Computer-assisted cup placement techniques in total hip arthroplasty improve accuracy of placement. Clin Orthop Relat Res. 2004;426:174–179.CrossRefPubMed Jolles BM, Genoud P, Hoffmeyer P. Computer-assisted cup placement techniques in total hip arthroplasty improve accuracy of placement. Clin Orthop Relat Res. 2004;426:174–179.CrossRefPubMed
20.
Zurück zum Zitat Kelley SS, Lachiewicz PF, Hickman JM, Paterno SM. Relationship of femoral head and acetabular size to the prevalence of dislocation. Clin Orthop Relat Res. 1998;355:163–170.CrossRefPubMed Kelley SS, Lachiewicz PF, Hickman JM, Paterno SM. Relationship of femoral head and acetabular size to the prevalence of dislocation. Clin Orthop Relat Res. 1998;355:163–170.CrossRefPubMed
21.
Zurück zum Zitat Kennedy JG, Rogers WB, Soffe KE, Sullivan RJ, Griffen DG, Sheehan LJ. Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration. J Arthroplasty. 1998;13:530–534.CrossRefPubMed Kennedy JG, Rogers WB, Soffe KE, Sullivan RJ, Griffen DG, Sheehan LJ. Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration. J Arthroplasty. 1998;13:530–534.CrossRefPubMed
22.
Zurück zum Zitat Kluess D, Martin H, Mittelmeier W, Schmitz KP, Bader R. Influence of femoral head size on impingement, dislocation and stress distribution in total hip replacement. Med Eng Phys. 2007;29:465–471.CrossRefPubMed Kluess D, Martin H, Mittelmeier W, Schmitz KP, Bader R. Influence of femoral head size on impingement, dislocation and stress distribution in total hip replacement. Med Eng Phys. 2007;29:465–471.CrossRefPubMed
23.
Zurück zum Zitat Kummer FJ, Shah S, Lyer S, DiCesare PE. The effect of acetabular cup orientations on limiting hip rotation. J Arthroplasty. 1999;14:509–513.CrossRefPubMed Kummer FJ, Shah S, Lyer S, DiCesare PE. The effect of acetabular cup orientations on limiting hip rotation. J Arthroplasty. 1999;14:509–513.CrossRefPubMed
24.
Zurück zum Zitat Leichtle U, Gosselke N, Wirth CJ, Rudert M. Radiologic evaluation of cup placement variation in conventional total hip arthroplasty [in German]. Rofo. 2007;179:46–52.PubMed Leichtle U, Gosselke N, Wirth CJ, Rudert M. Radiologic evaluation of cup placement variation in conventional total hip arthroplasty [in German]. Rofo. 2007;179:46–52.PubMed
25.
Zurück zum Zitat Leslie IJ, Williams S, Isaac G, Ingham E, Fisher J. High cup angle and microseparation increase the wear of hip surface replacements. Clin Orthop Relat Res. 2009;467:2259–2265.CrossRefPubMed Leslie IJ, Williams S, Isaac G, Ingham E, Fisher J. High cup angle and microseparation increase the wear of hip surface replacements. Clin Orthop Relat Res. 2009;467:2259–2265.CrossRefPubMed
26.
Zurück zum Zitat Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–220.PubMed Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–220.PubMed
27.
Zurück zum Zitat Lubbeke A, Moons KG, Garavaglia G, Hoffmeyer P. Outcomes of obese and nonobese patients undergoing revision total hip arthroplasty. Arthritis Rheum. 2008;59:738–745.CrossRefPubMed Lubbeke A, Moons KG, Garavaglia G, Hoffmeyer P. Outcomes of obese and nonobese patients undergoing revision total hip arthroplasty. Arthritis Rheum. 2008;59:738–745.CrossRefPubMed
28.
Zurück zum Zitat Mallory TH, Lombardi AV Jr, Fada RA, Herrington SM, Eberle RW. Dislocation after total hip arthroplasty using the anterolateral abductor split approach. Clin Orthop Relat Res. 1999;358:166–172.CrossRefPubMed Mallory TH, Lombardi AV Jr, Fada RA, Herrington SM, Eberle RW. Dislocation after total hip arthroplasty using the anterolateral abductor split approach. Clin Orthop Relat Res. 1999;358:166–172.CrossRefPubMed
29.
Zurück zum Zitat Malviya A, Holland JP. Pseudotumours associated with metal-on-metal hip resurfacing: 10-year Newcastle experience. Acta Orthop Belgica. 2009;75:477–483. Malviya A, Holland JP. Pseudotumours associated with metal-on-metal hip resurfacing: 10-year Newcastle experience. Acta Orthop Belgica. 2009;75:477–483.
30.
Zurück zum Zitat Manley M, Ong K, Lau E, Kurtz SM. Effect of volume on total hip arthroplasty revision rates in the United States Medicare population. J Bone Joint Surg. 2008;90:2446–2451.CrossRefPubMed Manley M, Ong K, Lau E, Kurtz SM. Effect of volume on total hip arthroplasty revision rates in the United States Medicare population. J Bone Joint Surg. 2008;90:2446–2451.CrossRefPubMed
31.
Zurück zum Zitat Marx A, von Knoch M, Pfortner J, Wiese M, Saxler G. Misinterpretation of cup anteversion in total hip arthroplasty using planar radiography. Arch Orthop Trauma Surg. 2006;126:487–492.CrossRefPubMed Marx A, von Knoch M, Pfortner J, Wiese M, Saxler G. Misinterpretation of cup anteversion in total hip arthroplasty using planar radiography. Arch Orthop Trauma Surg. 2006;126:487–492.CrossRefPubMed
32.
Zurück zum Zitat Morlock MM, Bishop N, Zustin J, Hahn M, Ruther W, Amling M. Modes of implant failure after hip resurfacing: morphological and wear analysis of 267 retrieval specimens. J Bone Joint Surg Am. 2008;90 Suppl 3:89–95.CrossRefPubMed Morlock MM, Bishop N, Zustin J, Hahn M, Ruther W, Amling M. Modes of implant failure after hip resurfacing: morphological and wear analysis of 267 retrieval specimens. J Bone Joint Surg Am. 2008;90 Suppl 3:89–95.CrossRefPubMed
33.
Zurück zum Zitat Morrey BF. Difficult complications after hip joint replacement: dislocation. Clin Orthop Relat Res. 1997;344:179–187.CrossRefPubMed Morrey BF. Difficult complications after hip joint replacement: dislocation. Clin Orthop Relat Res. 1997;344:179–187.CrossRefPubMed
34.
Zurück zum Zitat Morrey BF. Instability after total hip arthroplasty. Orthop Clin North Am. 1992;23:237–248.PubMed Morrey BF. Instability after total hip arthroplasty. Orthop Clin North Am. 1992;23:237–248.PubMed
35.
Zurück zum Zitat Muratoglu OK, Bragdon CR, O’Connor D, Perinchief RS, Estok DM II, Jasty M, Harris WH. Larger diameter femoral heads used in conjunction with a highly cross-linked ultra-high molecular weight polyethylene: a new concept. J Arthroplasty. 2001;16(8 Suppl 1):24–30. Muratoglu OK, Bragdon CR, O’Connor D, Perinchief RS, Estok DM II, Jasty M, Harris WH. Larger diameter femoral heads used in conjunction with a highly cross-linked ultra-high molecular weight polyethylene: a new concept. J Arthroplasty. 2001;16(8 Suppl 1):24–30.
36.
Zurück zum Zitat Myers GJ, Morgan D, McBryde CW, O’Dwyer K. Does surgical approach influence component positioning with Birmingham Hip Resurfacing? Int Orthop. 2009;33:59–63.CrossRefPubMed Myers GJ, Morgan D, McBryde CW, O’Dwyer K. Does surgical approach influence component positioning with Birmingham Hip Resurfacing? Int Orthop. 2009;33:59–63.CrossRefPubMed
37.
Zurück zum Zitat Najarian BC, Kilgore JE, Markel DC. Evaluation of component positioning in primary total hip arthroplasty using an imageless navigation device compared with traditional methods. J Arthroplasty. 2009;24:15–21.CrossRefPubMed Najarian BC, Kilgore JE, Markel DC. Evaluation of component positioning in primary total hip arthroplasty using an imageless navigation device compared with traditional methods. J Arthroplasty. 2009;24:15–21.CrossRefPubMed
38.
Zurück zum Zitat Nevelos JE, Ingham E, Doyle C, Nevelos AB, Fisher J. The influence of acetabular cup angle on the wear of “BIOLOX Forte” alumina ceramic bearing couples in a hip joint simulator. J Mater Sci. 2001;12:141–144. Nevelos JE, Ingham E, Doyle C, Nevelos AB, Fisher J. The influence of acetabular cup angle on the wear of “BIOLOX Forte” alumina ceramic bearing couples in a hip joint simulator. J Mater Sci. 2001;12:141–144.
39.
Zurück zum Zitat Newington DP, Bannister GC, Fordyce M. Primary total hip replacement in patients over 80 years of age. J Bone Joint Surg Br. 1990;72:450–452.PubMed Newington DP, Bannister GC, Fordyce M. Primary total hip replacement in patients over 80 years of age. J Bone Joint Surg Br. 1990;72:450–452.PubMed
40.
Zurück zum Zitat Nogler M, Mayr E, Krismer M, Thaler M. Reduced variability in cup positioning: the direct anterior surgical approach using navigation. Acta orthop. 2008;79:789–793.CrossRefPubMed Nogler M, Mayr E, Krismer M, Thaler M. Reduced variability in cup positioning: the direct anterior surgical approach using navigation. Acta orthop. 2008;79:789–793.CrossRefPubMed
41.
Zurück zum Zitat Pandit H, Glyn-Jones S, McLardy-Smith P, Gundle R, Whitwell D, Gibbons CL, Ostlere S, Athanasou N, Gill HS, Murray DW. Pseudotumours associated with metal-on-metal hip resurfacings. J Bone Joint Surg Br. 2008;90:847–851.CrossRefPubMed Pandit H, Glyn-Jones S, McLardy-Smith P, Gundle R, Whitwell D, Gibbons CL, Ostlere S, Athanasou N, Gill HS, Murray DW. Pseudotumours associated with metal-on-metal hip resurfacings. J Bone Joint Surg Br. 2008;90:847–851.CrossRefPubMed
42.
Zurück zum Zitat Pedersen DR, Callaghan JJ, Brown TD. Activity-dependence of the “safe zone” for impingement versus dislocation avoidance. Med Eng Phys. 2005;27:323–328.CrossRefPubMed Pedersen DR, Callaghan JJ, Brown TD. Activity-dependence of the “safe zone” for impingement versus dislocation avoidance. Med Eng Phys. 2005;27:323–328.CrossRefPubMed
43.
Zurück zum Zitat Penney GP, Edwards PJ, Hipwell JH, Slomczykowski M, Revie I, Hawkes DJ. Postoperative calculation of acetabular cup position using 2-D-3-D registration. IEEE Trans Biomed Eng. 2007;54:1342–1348.CrossRefPubMed Penney GP, Edwards PJ, Hipwell JH, Slomczykowski M, Revie I, Hawkes DJ. Postoperative calculation of acetabular cup position using 2-D-3-D registration. IEEE Trans Biomed Eng. 2007;54:1342–1348.CrossRefPubMed
44.
Zurück zum Zitat Pirard E, De Lint JA. Anteversion of the acetabular component in obese patients. Hip Int. 2007;17:99–103.PubMed Pirard E, De Lint JA. Anteversion of the acetabular component in obese patients. Hip Int. 2007;17:99–103.PubMed
45.
Zurück zum Zitat Reize P, Geiger EV, Suckel A, Rudert M, Wulker N. Influence of surgical experience on accuracy of acetabular cup positioning in total hip arthroplasty. Am J Orthop (Belle Mead NJ). 2008;37:360–363. Reize P, Geiger EV, Suckel A, Rudert M, Wulker N. Influence of surgical experience on accuracy of acetabular cup positioning in total hip arthroplasty. Am J Orthop (Belle Mead NJ). 2008;37:360–363.
46.
Zurück zum Zitat Sadr Azodi O, Adami J, Lindstrom D, Eriksson KO, Wladis A, Bellocco R. High body mass index is associated with increased risk of implant dislocation following primary total hip replacement: 2,106 patients followed for up to 8 years. Acta Orthop. 2008;79:141–147.CrossRefPubMed Sadr Azodi O, Adami J, Lindstrom D, Eriksson KO, Wladis A, Bellocco R. High body mass index is associated with increased risk of implant dislocation following primary total hip replacement: 2,106 patients followed for up to 8 years. Acta Orthop. 2008;79:141–147.CrossRefPubMed
47.
Zurück zum Zitat Sanchez-Sotelo J, Berry DJ. Epidemiology of instability after total hip replacement. Orthop Clin North Am. 2001;32:543–552.CrossRefPubMed Sanchez-Sotelo J, Berry DJ. Epidemiology of instability after total hip replacement. Orthop Clin North Am. 2001;32:543–552.CrossRefPubMed
48.
Zurück zum Zitat Saxler G, Marx A, Vandevelde D, Langlotz U, Tannast M, Wiese M, Michaelis U, Kemper G, Grutzner PA, Steffen R, von Knoch M, Holland-Letz T, Bernsmann K. The accuracy of free-hand cup positioning—a CT based measurement of cup placement in 105 total hip arthroplasties. Int Orthop. 2004;28:198–201.CrossRefPubMed Saxler G, Marx A, Vandevelde D, Langlotz U, Tannast M, Wiese M, Michaelis U, Kemper G, Grutzner PA, Steffen R, von Knoch M, Holland-Letz T, Bernsmann K. The accuracy of free-hand cup positioning—a CT based measurement of cup placement in 105 total hip arthroplasties. Int Orthop. 2004;28:198–201.CrossRefPubMed
49.
Zurück zum Zitat Shervin N, Rubash HE, Katz JN. Orthopaedic procedure volume and patient outcomes: a systematic literature review. Clin Orthop Relat Res. 2007;457:35–41.CrossRefPubMed Shervin N, Rubash HE, Katz JN. Orthopaedic procedure volume and patient outcomes: a systematic literature review. Clin Orthop Relat Res. 2007;457:35–41.CrossRefPubMed
50.
Zurück zum Zitat Shon WY, Baldini T, Peterson MG, Wright TM, Salvati EA. Impingement in total hip arthroplasty a study of retrieved acetabular components. J Arthroplasty. 2005;20:427–435.CrossRefPubMed Shon WY, Baldini T, Peterson MG, Wright TM, Salvati EA. Impingement in total hip arthroplasty a study of retrieved acetabular components. J Arthroplasty. 2005;20:427–435.CrossRefPubMed
51.
Zurück zum Zitat Shon WY, Gupta S, Biswal S, Hur CY, Jajodia N, Hong SJ, Myung JS. Validation of a simple radiographic method to determine variations in pelvic and acetabular cup sagittal plane alignment after total hip arthroplasty. Skeletal Radiol. 2008;37:1119–1127.CrossRefPubMed Shon WY, Gupta S, Biswal S, Hur CY, Jajodia N, Hong SJ, Myung JS. Validation of a simple radiographic method to determine variations in pelvic and acetabular cup sagittal plane alignment after total hip arthroplasty. Skeletal Radiol. 2008;37:1119–1127.CrossRefPubMed
52.
Zurück zum Zitat Solomon DH, Losina E, Baron JA, Fossel AH, Guadagnoli E, Lingard EA, Miner A, Phillips CB, Katz JN. Contribution of hospital characteristics to the volume-outcome relationship: dislocation and infection following total hip replacement surgery. Arthritis Rheum. 2002;46:2436–2444.CrossRefPubMed Solomon DH, Losina E, Baron JA, Fossel AH, Guadagnoli E, Lingard EA, Miner A, Phillips CB, Katz JN. Contribution of hospital characteristics to the volume-outcome relationship: dislocation and infection following total hip replacement surgery. Arthritis Rheum. 2002;46:2436–2444.CrossRefPubMed
53.
Zurück zum Zitat Todkar M. Obesity does not necessarily affect the accuracy of acetabular cup implantation in total hip replacement. Acta Orthop Belg. 2008;74:206–209.PubMed Todkar M. Obesity does not necessarily affect the accuracy of acetabular cup implantation in total hip replacement. Acta Orthop Belg. 2008;74:206–209.PubMed
54.
Zurück zum Zitat Widmer KH, Zurfluh B. Compliant positioning of total hip components for optimal range of motion. J Orthop Res. 2004;22:815–821.CrossRefPubMed Widmer KH, Zurfluh B. Compliant positioning of total hip components for optimal range of motion. J Orthop Res. 2004;22:815–821.CrossRefPubMed
55.
Zurück zum Zitat Wixson RL, MacDonald MA. Total hip arthroplasty through a minimal posterior approach using imageless computer-assisted hip navigation. J Arthroplasty. 2005;20(7 Suppl 3):51–56.CrossRefPubMed Wixson RL, MacDonald MA. Total hip arthroplasty through a minimal posterior approach using imageless computer-assisted hip navigation. J Arthroplasty. 2005;20(7 Suppl 3):51–56.CrossRefPubMed
56.
Zurück zum Zitat Yamaguchi M, Akisue T, Bauer TW, Hashimoto Y. The spatial location of impingement in total hip arthroplasty. J Arthroplasty. 2000;15:305–313.CrossRefPubMed Yamaguchi M, Akisue T, Bauer TW, Hashimoto Y. The spatial location of impingement in total hip arthroplasty. J Arthroplasty. 2000;15:305–313.CrossRefPubMed
57.
Zurück zum Zitat Zheng G, Steppacher S, Zhang X, Tannast M. Precise estimation of postoperative cup alignment from single standard X-ray radiograph with gonadal shielding. Med Image Comput Comput Assist Interv Int Conf Med Image Comput Comput Assist Interv. 2007;10(Pt 2):951–959. Zheng G, Steppacher S, Zhang X, Tannast M. Precise estimation of postoperative cup alignment from single standard X-ray radiograph with gonadal shielding. Med Image Comput Comput Assist Interv Int Conf Med Image Comput Comput Assist Interv. 2007;10(Pt 2):951–959.
Metadaten
Titel
The John Charnley Award: Risk Factors for Cup Malpositioning: Quality Improvement Through a Joint Registry at a Tertiary Hospital
verfasst von
Mark C. Callanan, MA
Bryan Jarrett, BS
Charles R. Bragdon, PhD
David Zurakowski, PhD
Harry E. Rubash, MD
Andrew A. Freiberg, MD
Henrik Malchau, MD, PhD
Publikationsdatum
01.02.2011
Verlag
Springer-Verlag
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 2/2011
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-010-1487-1

Weitere Artikel der Ausgabe 2/2011

Clinical Orthopaedics and Related Research® 2/2011 Zur Ausgabe

Symposium: Papers Presented at the Hip Society Meetings 2010

Tranexamic Acid Reduces Allogenic Transfusion in Revision Hip Arthroplasty

Symposium: Papers Presented at the Hip Society Meetings 2010

Bisphosphonate Remains Highly Localized After Elution From Porous Implants

Symposium: Papers Presented at the Hip Society Meetings 2010

Ceramic-on-Ceramic Total Hip Arthroplasty: Incidence of Instability and Noise

Symposium: Papers Presented at the Hip Society Meetings 2010

Wear versus Thickness and Other Features of 5-Mrad Crosslinked UHMWPE Acetabular Liners

Symposium: Papers Presented at the Hip Society Meetings 2010

Modular Femoral Sleeve and Stem Implant Provides Long-term Total Hip Survivorship

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.