The modular cardiac rhythm management system: the EMPOWER leadless pacemaker and the EMBLEM subcutaneous ICD

verfasst von: F. V. Y. Tjong, MD, PhD, B. E. Koop, PhD

Erschienen in: Herzschrittmachertherapie + Elektrophysiologie | Ausgabe 4/2018

Abstract

Cardiac implantable electronic devices have been successfully treating patients with brady- and tachyarrhythmias for decades. However, there are still significant complications related to this therapy modality, many related to the transvenous lead. Paradigm-shifting technologies, such as the subcutaneous implantable cardioverter–defibrillator (S-ICD) and leadless cardiac pacemakers (LCP), have emerged to address these complications. The novel modular cardiac rhythm management (mCRM) system, consisting of a communicating antitachycardia pacing-enabled LCP and S‑ICD, is the first system to integrate wireless intrabody communication between devices to allow for coordination of leadless pacing and defibrillator therapy delivery. In this review, the design and concept of the mCRM system are presented and available evidence is summarized.

Introduction

For many decades, cardiac electrical implantable devices (CIED) have been the cornerstone therapy for patients with brady- and tachyarrhythmias [14]. Despite many years of innovation and device development, these systems are still associated with a significant amount of acute and chronic complications, of which many are related to the “weakest link” of these systems: the transvenous lead [57]. Paradigm shifting technologies have been introduced to eliminate the need for a transvenous lead. First, in 2008, the subcutaneous implantable cardioverter defibrillator (S-ICD) was introduced [8], a device placed fully subcutaneously, limiting intracardiac hardware. Improvements in battery technology and advanced electrical circuitry brought about leadless pacemakers, miniaturized self-contained intracardiac single-chamber pacemakers delivered in the right ventricle through a percutaneous transfemoral catheter based approach [9, 10]. Two systems have been clinically available now for several years and showed clinical efficacy and safety [913]. However, existing leadless pacing systems and the S‑ICD are only available for patients requiring single-chamber right ventricular (RV) pacing or shock-only defibrillation therapy, respectively. Combined use of both types of devices could bring the potential benefits of this leadless approach to a larger patient population. Adding antitachycardia pacing (ATP) to this combined system would enable even larger patient groups to potentially benefit [14]. These combined device systems require safe and reliable device–device communication and enable a novel treatment concept of modular cardiac rhythm management (mCRM) therapy, in which device therapy could be further personalized to the current and future patients’ needs. The first mCRM system is presented in this review and consists of a communicating and the EMPOWER™ leadless cardiac pacemaker (LCP; both Boston Scientific), which are shown in Fig. 1. To date, this is the first system consisting of communicating independent cardiac devices for synchronized therapy application.

Intrabody communication and the modular cardiac rhythm management system

Beginning in the 1990s, researchers became interested in the possibilities of intrabody communication involving implanted devices in various forms and using various communication modalities [1517]. It later became clear that one of the most energy efficient modalities for communication between implanted devices was galvanic coupling, or communication using the conductive properties of the body [1820]. Through galvanic coupling, one implanted device transmits low-energy electrical signals, such as via alternating current, which is detected by two electrodes on the receiving device. The advantages of galvanic coupling over other communication modalities such as radiofrequency (RF) communication include not only lower energy requirements but also lower signal losses in tissue. The disadvantages of galvanic coupling include comparatively lower rates of data transmission and the potential for larger null spaces due to relative physical orientation of device electrodes.
Due to the power efficiency and less lossy nature in human tissue, galvanic coupling, or “conducted” communication, was selected as the communication modality to be used for device–device communication in the mCRM system. A secondary design goal was to allow existing implanted S‑ICDs to be upgraded via firmware, with no change to the hardware within the system, to enable the ability to communicate. The advantage of this strategy is to allow access to this new technology, the modular addition of intracardiac therapy, for patients with an EMBLEM platform S‑ICD system without the need to explant and replace the S‑ICD pulse generator. The compromise is to utilize only unidirectional conducted communication from S‑ICD to LCP, which allows access to ATP therapy but limits additional features that could come from this pairing to future generations of S‑ICD pulse generator where bidirectional communication would be possible.
To accomplish the objective of adding conducted communication ability to the S‑ICD system without updating its hardware, existing circuitry in the S‑ICD pulse generator, normally utilized to check shock circuit integrity, was repurposed to send bursts of approximately 25 kHz pulses from the shock coil of the S‑ICD electrode to the S‑ICD pulse generator can. These pulses are low in voltage amplitude (<5 V emitted) and are sent in a proprietary pattern that can be recognized by the LCP and is distinct from various levels of signal noise that may be sensed by the LCP. The pattern of pulses constituting a full communication message from S‑ICD requires less than 40 ms in duration, though to provide redundancy two messages are sent in series to help mitigate risk of failed receipt by the LCP. Two full communication messages back-to-back require less than 135 ms in duration, which is short enough to fit within a standard pacemaker cardiac sensing refractory window. S‑ICD communication messages are sent coincident with the R‑wave of the cardiac cycle to avoid oversensing of these communications on the LCP’s cardiac sensing channel. Fig. 2 depicts the S‑ICD-to-LCP communication details.
Since the vector used by the S‑ICD to communicate electrical signals is from the shock coil to the pulse generator can, and the LCP senses these conducted signals via its cathode and anode, the spatial orientation of the LCP within the S‑ICD communication vector can impact the ability of the LCP to sense the transmitted signals from S‑ICD. Theoretically, the highest communication signal amplitude sensed by the LCP is when its main axis is aligned with the S‑ICD communication vector as shown in Fig. 2. Conversely, the lowest communication signal amplitude sensed by the LCP is when its main axis is perpendicular to the S‑ICD communication vector, or when the LCP is approximately parallel with the S‑ICD shock coil when the implanted system is viewed in an anterior–posterior view under fluoroscopy. The orientation of the LCP relative to the S‑ICD communication vector in humans is expected to have a favorable position in relation to the communication vector, with a preferred device deployment in the right ventricular apicoseptal region. Implantation of the LCP in the free wall of the right ventricle is not recommended because of the less favorable device orientation and the higher potential risk of cardiac perforation. A previous report evaluated the device orientation of other leadless pacemakers in humans (e. g., [21]). However, if the EMPOWER LCP were to be in an unfavorable orientation, the combination of S‑ICD communication message redundancy and the motion of the heart during the cardiac cycle is intended to mitigate the possibility of the S‑ICD communication not being received by the LCP.
In terms of tachyarrhythmia and bradyarrhythmia sensing and therapy, the S‑ICD and LCP elements of the mCRM system act independently from each other except for antitachycardia pacing (ATP) therapy. For ATP, S‑ICD employs the same sensing and detection criteria to declare a tachyarrhythmia episode as those used in market-approved A209 and A219 model S‑ICDs. However, the same model S‑ICD with firmware upgrade and mCRM communication feature activated can request the EMPOWER LCP to perform ATP therapy according to programmed parameters. These programmable parameters include the number of ATP attempts within a programmed S‑ICD tachyarrhythmia rate zone, scheme of ATP (Burst or Scan), ATP coupling/burst interval, and number of pulses per ATP attempt. Depending on the rate zone of the S‑ICD in which tachyarrhythmia is occurring, ATP therapy can be requested of the LCP, including multiple times, before moving to charging for shock therapy, or ATP therapy can be requested and applied in parallel to charging for shock. After the ATP communication request by the S‑ICD there is a programmed delay of about 4 s before the S‑ICD starts sensing the intrinsic rhythm to allow ATP therapy to be applied by the LCP and ensure ATP is not included in the subsequent tachydetection by the S‑ICD.

Data from acute and chronic animal implants

Data regarding successful S‑ICD/LCP device–device communication and coordination of ATP during simulated ventricular tachycardia in an ovine model was first reported in 2016 [22]. A larger animal study was published later with more extensive evaluations of the technology [23]. In the latter study, acute and chronic animal implants of the mCRM system were completed to test the pacing and sensing performance of the EMPOWER LCP, as well as the communication and system performance of the S‑ICD and LCP together. Data from acute implants in multiple animal species (canine, ovine, and swine) and up to 90 days of chronic implant in canines were reported, with LCP pacing and sensing data as shown in Table 1.
Table 1
Acute and 3‑month EMPOWER LCP performance showing favorable pacing and sensing performance over the course of the study. Values are n (%) or mean ± standard deviation
 
Acute Performance (n = 40)
3-Month Performance (n = 23)
Baseline
Canine
Ovine (n = 8)
Swine (n = 5)
Canine (n = 27)
7 Days
28 Days
90 Days
p Value
Implant Success
8 (100)
5 (100)
26 (96)
LCP Position
RV Apex
8 (100)
4(80)
12 (46)
RV Apical Septum
0
0
14 (54)
RV Outflow Tract
0
1 (20)
0
LCP Electrical Performance
(n= 8)
(n= 5)
(n= 26)
(n= 23)
(n= 23)
(n= 23)
 
Pacing Threshold, V at 0.5 ms
1.10 ± 0.81
0.53 ± 0.49
0.37 ± 0.19
0.56 ± 0.37b
0.54 ± 0.30b
0.72 ± 0.45b,c
<0.001
R-Wave Amplitude, mV
6.6 ± 1.4
28.3 ± 5.8
25.8 ± 5.1d
26.3 ± 6.8d
25.0 ± 9.4d
23.3 ± 9.4
<0.001
Impedance, Ω
665 ± 225
753 ± 118
826 ± 171d
785 ± 129d
827 ± 105d
728 ± 141c
0.04
LCP Post Shock Performance
(n= 8)
(n= 2)
(n= 1)
  
(n= 7)
 
Pre- to Post Shock Change in Pacing Threshold, V at 0.5 ms
0.0 ± 0.5
0.1 ± 0.1
0
0.0 ± 0.1
Pre- to Post-Shock Change in Impedance, W
18 ± 49
26 ± 40
5
−48 ± 58
Table reprinted from Tjong et al. [23] with permission from the publisher
LCP leadless cardiac pacemaker, RV right ventricle
ap value between baseline and 90 days, for pacing threshold and impedance tested with the Student t test, for R wave calculated with a linear regression analysis
bPacing threshold data from 3 animals were excluded because the LCP prototypes did not have steroid-eluting electrode
cImpedance and pacing threshold data from 1 animal were excluded due to suspected device malfunction at day 90
dR-wave and impedance data from 7 animals at baseline, 9 animals at day 7, and 7 animals at day 28 were excluded due to programmer software malfunction
Data obtained on voltage threshold for successful communication from S‑ICD to LCP was also obtained in canines for three postures in the same 3‑month study and is shown in Fig. 3. The data were taken with respect to the voltage amplitude setting in the S‑ICD pulse generator used to generate the communication pulses. This setting is not equal to the voltage emitted by the device, which is approximately 30–75% lower than the voltage amplitude setting (depending on the setting used).
These results are important for understanding the chronic stability of device–device communication for the mCRM system in vivo. Specifically, the results show that, on average, the communication threshold was low (signal margin for successful device–device communication was high) and stable throughout the 3‑month study. Further, any differences in communication threshold between the three postures tested were small and not of practical significance relative to the signal margin that exists if the S‑ICD was programmed to its nominal amplitude setting. Indeed, there is little incentive to program the system to a communication setting lower than the nominal value as there is expected to be very low impact to S‑ICD longevity (days or weeks depending on the number of ambulatory ATP episodes) to send communications during the device lifetime, regardless of amplitude setting.
Using data from the same 3‑month animal study, an analysis of the impact of EMPOWER LCP spatial orientation on communication threshold was conducted and reported in 2018 [24]. Included in the same evaluation was a retrospective analysis of patients from a single center implanted with either an S‑ICD system or a commercially available leadless pacemaker. This analysis was performed to utilize the data from the animal study to determine the favorability of the LCP spatial orientation relative to the S‑ICD communication vector in human anatomy. An example of the communication vector between S‑ICD and LCP and the orientation of the LCP within this vector is shown in Fig. 4. The findings from the animal data showed that a when the LCP is approximately parallel with the S‑ICD shock coil in an anterior–posterior view under fluoroscopy, the communication threshold trends higher but the effect was moderate. It is hypothesized that the practice of sending two communication messages in series from the S‑ICD helps alleviate the possibility of communication failure. Additionally, the human body is not a homogenous conductive medium; organs such as the lungs likely cause a divergence of electric field orientation created by the S‑ICD communication pulses which may allow a high tolerance for many LCP spatial orientations S‑ICD relative to the communication vector. Overall, the angle (α, Fig. 4) of the LCP axis relative to the S‑ICD shock coil in canines was found to be 25° on average, while in humans the mean angle was 56°. This result may predict an even higher signal margin for communication when the mCRM system is evaluated in humans in the near future, though it is acknowledged there are other important factors such as torso size which can impact communication success.

Discussion

Modular therapy is the idea that one can tailor implantable medical device therapy to the specific needs of the patient throughout their lifetime so that the patient receives only the device(s) they need, only when needed. Medical devices inherently involve risks to the patient related to implantation and to continuing operation in the body. Ideally for the patient the benefits of the implanted devices far outweigh the risks. However, some patients who would be eligible to receive an implanted S‑ICD system instead receive a transvenous ICD system despite having no immediate need for pacing therapy or ATP. This may be a decision made by the physician with a view to the possibility the patient would need these therapies in the future. The downside of this decision is the patient receives a system that has disadvantages compared to S‑ICD: Intracardiac leads which may fail, infection pathways to the heart, and complications that occur acutely or chronically due to implantation within the heart.
With the advent of the mCRM system and wireless communication capabilities of implants, there will be an alternative approach: Implant the S‑ICD system in eligible patients now and implant the EMPOWER LCP later only if the patient develops a therapeutic need for the device. The number of patients who develop such a need are expected to be low [25], meaning more patients can benefit from a less invasive device therapy strategy. As further components of this modular system are developed in the future, further therapies will be available for application in a modular fashion as the patient’s disease state progresses, heralding an era of truly personalized device medicine.

Limitations and next steps

The main limitations in understanding the safety and effectiveness of the mCRM system is that only animal data exist currently. Though animal evaluations in multiple models can serve as a useful surrogate when human evaluation is not available, there remain aspects of any animal model which cannot fully replicate human anatomy and environment. Human clinical studies of the EMPOWER LCP as a standalone device and the mCRM system in humans are expected to commence in 2019. Future iterations of the mCRM system are expected to include leadless devices for pacing other chambers of the heart to provide dual chamber pacing therapy or cardiac resynchronization therapy, as standalone devices and in coordination with a co-implanted S‑ICD. Future iterations of S‑ICD system are also expected to allow bidirectional communication with other leadless devices in the mCRM system, as well as new algorithms to allow sensing verification, new discrimination capabilities and new features via device–device communication.

Practical conclusion

  • Complications with the transvenous lead have been overcome using S‑ICD and LCP.
  • A modular cardiac rhythm management (mCRM) system, consisting of a communicating antitachycardia pacing-enabled EMPOWER™ LCP and EMBLEM™ platform S‑ICD system, allows for coordination of leadless pacing and defibrillator therapy delivery.
  • Firmware can be upgraded in implanted S‑ICDs without explantation or replacement of the pulse generator.
  • Acute and 3‑month data regarding device–device communication and coordination of ATP during simulated ventricular tachycardia in animal studies are promising.
  • Advances include the possibility to implant the S‑ICD system in eligible patients now and implant the EMPOWER LCP later only if the patient develops a therapeutic need for the device.
  • Human clinical studies with the EMPOWER LCP and the mCRM system are expected in 2019.

Compliance with ethical guidelines

Conflict of interest

F.V.Y. Tjong reports consulting fees from St. Jude Medical, Boston Scientific. B.E. Koop is an employee of Boston Scientific.
This article does not contain any studies with human participants or animals performed by any of the authors.
Open Access. This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

Herzschrittmachertherapie + Elektrophysiologie

Print-Titel

  • Themenbezogene Hefte zu aktuellen Fragen der Elektrophysiologie sowie der Diagnostik und Therapie von Herzrhythmusstörungen

  • Organ der Deutschen Gesellschaft für Kardiologie - Herz- und Kreislaufforschung e.V.
  • Journal affiliated to the European Heart Rhythm Association 
  • Partnerjournal der AG Rhythmologie der ÖKG

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Tracy CM, Epstein AE, Darbar D et al (2012) 2012 ACCF/AHA/HRS focused update of the 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: a report of the American college of cardiology foundation/American heart association task force on practice guidelines. J Am Coll Cardiol 60:1297–1313CrossRef Tracy CM, Epstein AE, Darbar D et al (2012) 2012 ACCF/AHA/HRS focused update of the 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: a report of the American college of cardiology foundation/American heart association task force on practice guidelines. J Am Coll Cardiol 60:1297–1313CrossRef
2.
Zurück zum Zitat Moss AJ, Hall WJ, Cannom DS et al (1996) Improved survival with an implanted Defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. Multicenter automatic defibrillator implantation trial investigators. N Engl J Med 335:1933–1940CrossRef Moss AJ, Hall WJ, Cannom DS et al (1996) Improved survival with an implanted Defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. Multicenter automatic defibrillator implantation trial investigators. N Engl J Med 335:1933–1940CrossRef
3.
Zurück zum Zitat Moss AJ, Zareba W, Hall WJ et al (2002) Multicenter automatic defibrillator implantation trial II. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med 346:877–883CrossRef Moss AJ, Zareba W, Hall WJ et al (2002) Multicenter automatic defibrillator implantation trial II. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med 346:877–883CrossRef
4.
Zurück zum Zitat Bardy GH, Lee KL, Mark DB et al (2005) Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 352:225–237CrossRef Bardy GH, Lee KL, Mark DB et al (2005) Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 352:225–237CrossRef
5.
Zurück zum Zitat Kirkfeldt RE, Johansen JB, Nohr EA, Jørgensen OD, Nielsen JC (2014) Complications after cardiac Implantable electronic device Implantations: an analysis of a complete, nationwide cohort in Denmark. Eur Heart J 35:1186–1194CrossRef Kirkfeldt RE, Johansen JB, Nohr EA, Jørgensen OD, Nielsen JC (2014) Complications after cardiac Implantable electronic device Implantations: an analysis of a complete, nationwide cohort in Denmark. Eur Heart J 35:1186–1194CrossRef
6.
Zurück zum Zitat Van der Heijden AC, Borleffs CJ, Buiten MS et al (2015) The clinical course of patients with Implantable Cardioverter-Defibrillators: extended experience on clinical outcome, device replacements, and device-related complications. Heart Rhythm 12:1169–1176CrossRef Van der Heijden AC, Borleffs CJ, Buiten MS et al (2015) The clinical course of patients with Implantable Cardioverter-Defibrillators: extended experience on clinical outcome, device replacements, and device-related complications. Heart Rhythm 12:1169–1176CrossRef
7.
Zurück zum Zitat Udo EO, Zuithoff NP, van Hemel NM et al (2012) Incidence and predictors of short- and long-term complications in pacemaker therapy: the FOLLOWPACE study. Heart Rhythm 9:728–735CrossRef Udo EO, Zuithoff NP, van Hemel NM et al (2012) Incidence and predictors of short- and long-term complications in pacemaker therapy: the FOLLOWPACE study. Heart Rhythm 9:728–735CrossRef
8.
Zurück zum Zitat Bardy GH, Smith WM, Hood MA et al (2010) An entirely subcutaneous implantable cardioverter-defibrillator. N Engl J Med 363:36–44CrossRef Bardy GH, Smith WM, Hood MA et al (2010) An entirely subcutaneous implantable cardioverter-defibrillator. N Engl J Med 363:36–44CrossRef
9.
Zurück zum Zitat Reddy VY, Exner DV, Cantillon DJ et al (2015) LEADLESS II study investigators percutaneous implantation of an entirely intracardiac leadless pacemaker. N Engl J Med 373:1125–1135CrossRef Reddy VY, Exner DV, Cantillon DJ et al (2015) LEADLESS II study investigators percutaneous implantation of an entirely intracardiac leadless pacemaker. N Engl J Med 373:1125–1135CrossRef
10.
Zurück zum Zitat Reynolds D, Duray GZ, Omar R et al (2016) Micra transcatheter pacing study group A leadless intracardiac transcatheter pacing system. N Engl J Med 374:533–541CrossRef Reynolds D, Duray GZ, Omar R et al (2016) Micra transcatheter pacing study group A leadless intracardiac transcatheter pacing system. N Engl J Med 374:533–541CrossRef
11.
Zurück zum Zitat Burke MC, Gold MR, Knight BP et al (2015) Safety and efficacy of the totally subcutaneous implantable defibrillator: 2‑year results from a pooled analysis of the IDE study and EFFORTLESS registry. J Am Coll Cardiol 65:1605–1615CrossRef Burke MC, Gold MR, Knight BP et al (2015) Safety and efficacy of the totally subcutaneous implantable defibrillator: 2‑year results from a pooled analysis of the IDE study and EFFORTLESS registry. J Am Coll Cardiol 65:1605–1615CrossRef
12.
Zurück zum Zitat Boersma LV, Barr CS, Burke MC et al (2017) Performance of the subcutaneous implantable cardioverter-defibrillator in patients with a primary prevention indication with and without a reduced ejection fraction versus patients with a secondary prevention indication. Heart Rhythm 14(3):367–375CrossRef Boersma LV, Barr CS, Burke MC et al (2017) Performance of the subcutaneous implantable cardioverter-defibrillator in patients with a primary prevention indication with and without a reduced ejection fraction versus patients with a secondary prevention indication. Heart Rhythm 14(3):367–375CrossRef
14.
Zurück zum Zitat Kleemann T, Strauss M, Kouraki K, Zahn R (2015) Clinical course and prognostic relevance of antitachycardia pacing-terminated ventricular tachyarrhythmias in implantable cardioverter-defibrillator patients. Europace 17:1068–1075CrossRef Kleemann T, Strauss M, Kouraki K, Zahn R (2015) Clinical course and prognostic relevance of antitachycardia pacing-terminated ventricular tachyarrhythmias in implantable cardioverter-defibrillator patients. Europace 17:1068–1075CrossRef
15.
Zurück zum Zitat Zimmerman TG (1995) Personal Area Networks (PAN): Near-Field Intra-Body Communication. PhD thesis. Massachusetts Institute of Technology, MA, USA Zimmerman TG (1995) Personal Area Networks (PAN): Near-Field Intra-Body Communication. PhD thesis. Massachusetts Institute of Technology, MA, USA
16.
Zurück zum Zitat Lindsey DP, McKee EL, Hull ML, Howell SM (1998) A new technique for transmission of signals from implantable transducers. Ieee Trans Biomed Eng 45(5):614–619CrossRef Lindsey DP, McKee EL, Hull ML, Howell SM (1998) A new technique for transmission of signals from implantable transducers. Ieee Trans Biomed Eng 45(5):614–619CrossRef
17.
Zurück zum Zitat Hachisuka K, Nakata A, Takeda T, Terauchi Y, Shiba K, Sasaki K, Hosaka H, Itao K (2003) Development and performance analysis of an intra-body communication device. The 12th International Conference on Transducers, Solid State Sensors, Actuators, and Microsystems, vol 2, pp 1722–1725 Hachisuka K, Nakata A, Takeda T, Terauchi Y, Shiba K, Sasaki K, Hosaka H, Itao K (2003) Development and performance analysis of an intra-body communication device. The 12th International Conference on Transducers, Solid State Sensors, Actuators, and Microsystems, vol 2, pp 1722–1725
18.
Zurück zum Zitat Ferguson JE, Redish AD (2011) Wireless communication with implanted medical devices using the conductive properties of the body. Expert Rev Med Devices 8(4):427–433CrossRef Ferguson JE, Redish AD (2011) Wireless communication with implanted medical devices using the conductive properties of the body. Expert Rev Med Devices 8(4):427–433CrossRef
19.
Zurück zum Zitat TruptiLotlikar AI, Tilak D, Komal R, Prajakata S (2012) Comparison of approaches to intra-body communication. Indian J Comp Sci Eng 3(5):731–736 TruptiLotlikar AI, Tilak D, Komal R, Prajakata S (2012) Comparison of approaches to intra-body communication. Indian J Comp Sci Eng 3(5):731–736
20.
Zurück zum Zitat Zhang S, Qin Y, Kuang J, Mak PU, Pun SH, Vai MI, Liu Y (2014) Development and prospect of Implantable intra-body communication technology. J Comput 9(2):463–473 Zhang S, Qin Y, Kuang J, Mak PU, Pun SH, Vai MI, Liu Y (2014) Development and prospect of Implantable intra-body communication technology. J Comput 9(2):463–473
21.
Zurück zum Zitat Ahmed FZ, Cunnington C, Motwani M, Zaidi AM (2017) Totally Leadless Dual-Device Implantation for Combined Spontaneous Ventricular Tachycardia Defibrillation and Pacemaker Function: A First Report. Can J Cardiol 33:1066e5–1066e7CrossRef Ahmed FZ, Cunnington C, Motwani M, Zaidi AM (2017) Totally Leadless Dual-Device Implantation for Combined Spontaneous Ventricular Tachycardia Defibrillation and Pacemaker Function: A First Report. Can J Cardiol 33:1066e5–1066e7CrossRef
22.
Zurück zum Zitat Tjong FVY, Brouwer TF, Kooiman KM, Smeding L, Koop B, Soltis B, Shuros A, Wilde AAM, Burke MC, Knops RE (2016) Communicating Antitachycardia pacing-enabled leadless pacemaker and subcutaneous implantable defibrillator. J Am Coll Cardiol 67(15):1865–1866CrossRef Tjong FVY, Brouwer TF, Kooiman KM, Smeding L, Koop B, Soltis B, Shuros A, Wilde AAM, Burke MC, Knops RE (2016) Communicating Antitachycardia pacing-enabled leadless pacemaker and subcutaneous implantable defibrillator. J Am Coll Cardiol 67(15):1865–1866CrossRef
23.
Zurück zum Zitat Tjong FVY, Brouwer TF, Koop B, Soltis B, Shuros A, Schmidt B, Swackhamer B, Quast AFBE, Wilde AAM, Burke MC, Knops RE (2017) Acute and 3‑month performance of a communicating leadless antitachycardia pacemaker and subcutaneous implantable defibrillator. J Am Coll Cardiol 3(13):1487–1498 Tjong FVY, Brouwer TF, Koop B, Soltis B, Shuros A, Schmidt B, Swackhamer B, Quast AFBE, Wilde AAM, Burke MC, Knops RE (2017) Acute and 3‑month performance of a communicating leadless antitachycardia pacemaker and subcutaneous implantable defibrillator. J Am Coll Cardiol 3(13):1487–1498
25.
Zurück zum Zitat Boersma L, Barr C, Knops R, Theuns D, Eckardt L, Neuzil P, Scholten M, Hood M, Kuschyk J, Jones P, Duffy E, Husby M, Stein K, Lambiase PD, Investigator Group (2017) Implant and midterm outcomes of the subcutaneous implantable cardioverter-defibrillator registry: the EFFORTLESS study. J Am Coll Cardiol 70(7):830–841CrossRef Boersma L, Barr C, Knops R, Theuns D, Eckardt L, Neuzil P, Scholten M, Hood M, Kuschyk J, Jones P, Duffy E, Husby M, Stein K, Lambiase PD, Investigator Group (2017) Implant and midterm outcomes of the subcutaneous implantable cardioverter-defibrillator registry: the EFFORTLESS study. J Am Coll Cardiol 70(7):830–841CrossRef
Metadaten
Titel
The modular cardiac rhythm management system: the EMPOWER leadless pacemaker and the EMBLEM subcutaneous ICD
verfasst von
F. V. Y. Tjong, MD, PhD
B. E. Koop, PhD
Publikationsdatum
31.10.2018
Verlag
Springer Medizin
Erschienen in
Herzschrittmachertherapie + Elektrophysiologie / Ausgabe 4/2018