The relationship of pre-procedural Dmax based sizing to lesion level outcomes in Absorb BVS and Xience EES treated patients in the AIDA trial

verfasst von: Ruben Y. G. Tijssen, Laura S. M. Kerkmeijer, Yuki Katagiri, Robin P. Kraak, Kuniaki Takahashi, Norihiro Kogame, Ply Chichareon, Rodrigo Modolo, Taku Asano, Martina Nassif, Deborah N. Kalkman, Yohei Sotomi, Carlos Collet, Sjoerd H. Hofma, Rene J. van der Schaaf, E. Karin Arkenbout, Auke P. J. D. Weevers, Marcel A. M. Beijk, Jan J. Piek, Jan G. P. Tijssen, Jose P. Henriques, Robbert J. de Winter, Yoshinobu Onuma, Patrick W. Serruys, Joanna J. Wykrzykowska

Erschienen in: The International Journal of Cardiovascular Imaging | Ausgabe 7/2019

Abstract

Due to expansion limits of the Absorb bioresorbable scaffold a meticulous implantation with correct sizing is required. We sought to investigate the clinical outcomes based on the sizing of the device related to the maximal lumen diameter measured by quantitative coronary angiography in Absorb BVS and Xience EES treated lesions in the AIDA trial. Sizing of Absorb bioresorbable vascular scaffold (BVS) and Xience everolimus eluting stent (EES) was graded according to the definitions of device non-oversize and device oversize on pre-procedural angiography. Lesion-oriented outcomes (LOCE) (device thrombosis, TLR and TVMI) that occurred during 2 years follow-up were related to device non-oversized or oversized status. In the Absorb BVS group, LOCE occurred in 48 (7.4%) lesions in the oversized group and in 32 (8.2%) lesions in the non-oversized group (HR 0.91; 95% CI 0.58–1.42; p = 0.681), whereas TLR occurred in 34 (5.3%) lesions and in 23 lesions (5.9%), respectively (HR 0.89; 95% CI 0.52–1.51; p = 0.666). Definite scaffold thrombosis occurred in 11 (1.7%) device oversized treated lesions against 16 (4.1%) device non-oversized treated lesions (HR 0.41; 95% CI 0.19–0.89; p = 0.020). There were no differences in event rates between oversized and non-oversized groups in lesions treated with Xience EES. There was no significant difference in LOCE between oversized and non-oversized treated Absorb BVS and Xience EES treated lesions. Non-oversized Absorb BVS implantation was associated with a higher risk of scaffold thrombosis at complete 2 years follow-up. The majority of very late scaffold thrombosis occurred in properly sized devices.
Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1007/​s10554-019-01576-y) contains supplementary material, which is available to authorized users.
Ruben Y. G. Tijssen and Laura S. M. Kerkmeijer have contributed equally to this work.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
Absorb BVS
Absorb bioresorbable vascular scaffold
Dmax
Maximum diameter
LOCE
Lesion oriented composite endpoint
MACE
Major adverse cardiac events
MLD
Minimum lumen diameter
PCI
Percutaneous coronary revascularization
TLF
Target lesion failure
TLR
Target lesion revascularization
TVF
Target vessel failure
TV-MI
Target vessel myocardial infarction
QCA
Quantitative coronary angiography
Xience EES
Xience everolimus eluting stent

Introduction

The early results of Absorb everolimus-eluting bioresorbable vascular scaffold (Absorb BVS) implantation in simple coronary lesions were promising [13]. Initial enthusiasm was hampered when large randomized trials with long-term follow-up reported higher rates of target lesion failure (TLF), and scaffold thrombosis in the Absorb BVS compared to Xience everolimus eluting stent (EES) [47].
The limitations of the device, as well as implantation techniques or strategies, have been proposed as possible causes of adverse events [8]. Proper sizing of the scaffold may be of great importance acutely after scaffold implantation. The limited ability to over-expand makes the Absorb BVS prone to scaffold disruption when an undersized device is implanted and aggressively post-dilated. Furthermore, implanting an oversized device has been shown to cause more edge dissections and peri-procedural myocardial infarctions [911].
Ishibashi et al. demonstrated that implantation of an oversized Absorb BVS was associated with a higher risk of major adverse cardiac events (MACE) at 1 year [12]. Whereas, Katagiri et al. demonstrated that implantation of an undersized Absorb BVS was associated with a higher risk of MACE between 1 and 3 years [13]. This association was investigated in trials, which included only simple lesions and where quantitative assessment of the target vessel diameter by online quantitative coronary angiography (QCA) was required for enrollment in the trial [13, 14].
The Amsterdam Investigator-Initiated Absorb Strategy All-comers (AIDA) trial was an all-comers randomized controlled trial that analyzed Absorb BVS versus Xience EES in routine percutaneous coronary intervention (PCI). Although it demonstrated non-inferiority in the composite endpoint of target vessel failure (TVF) of Absorb BVS versus Xience EES at 2 years, Absorb BVS was associated with a higher risk of definite/probable device thrombosis [15]. In the current report we investigate whether proper sizing is a key factor leading to adverse events after device implantation and therefore we evaluated the relationship of maximum lumen diameter (Dmax) based device sizing on QCA, to lesion oriented outcomes in both the Absorb BVS and Xience EES arms of the AIDA trial.

Methods

Study design AIDA trial

The AIDA compared Absorb BVS (Abbott Vascular, Santa Clara, CA, USA) with Xience EES (Abbott Vascular, Santa Clara, CA, USA) in routine PCI. The study design [16], the preliminary safety report [6], and the 2-year results [15] have been published previously.
Briefly, between August 2013 and December 2015, 1845 consecutive patients with coronary artery disease undergoing PCI with one or more target lesions suitable for drug-eluting stent implantation were included in AIDA. Key exclusion criteria were target lesions longer than 70 mm, a visually estimated reference vessel diameter of < 2.5 mm of > 4.0 mm, treatment of a true bifurcation lesion with a priori planned two device strategy, and treatment of in-stent restenosis. Patients were randomized after successful pre-dilatation of the first lesion. Device sizing was based on visual assessment by the operator. Online QCA, or periprocedural testing of cardiac biomarkers, were not mandatory. Device implantation strategy was planned according to the instructions for use of the implanted device.

Design of the current analysis

QCA analysis

The current analysis includes all study lesions treated with at least one study device of which baseline angiogram suited for offline QCA analysis was available. QCA assessment of device sizing relied on the angiographic diameter function curve of the pre-treatment vessel segment consisting of three non-ambiguous luminal dimensions; namely, the minimal lumen diameter (MLD) and the Dmax in the proximal segment (proximal Dmax) and in the distal segment of interest (distal Dmax) with respect to the MLD.
Pre- and post-procedural measurements were performed in either (1) multiple matched views or (2) a single matched view. If no matched views were available, measurements were done within the view with visually the highest stenosis grade. Proximal and distal Dmax were measured in (a) pre-procedural projection(s). In case of a total occlusion, the MLD was considered to be 0 mm; the proximal Dmax was measured in a pre-procedural projection; and the distal Dmax was considered similar to the proximal Dmax. Offline QCA analyses were initially performed on the post-procedural angiograms within the Absorb BVS arm, and were performed by seven experienced readers [6]. We later added 5 experienced readers to our academic Corelab, in order to complete pre-procedural measures on the Absorb BVS arm, and complete the pre-procedural and post-procedural measures within the Xience EES arm. All QCA analyses were performed using Cardiovascular Angiography Analyses System [CAAS], version 5.11 (Pie Medical Imaging, Maastricht, the Netherlands). All QCA readers were blinded for clinical events, and were supervised by one cardiologist [YO], who is an expert in this field.

Definitions

The study population in both arms was stratified by the difference between both the proximal and distal angiographic Dmax and the nominal diameter of the implanted device. The selection of device size was considered “oversized” when the proximal and distal Dmax of the treated lesion were smaller than the nominal size of the device. Lesions with either a proximal or a distal Dmax or both Dmax larger than the nominal size of the device constituted the “non-oversized group”. When a patient received 2 or 3 overlapping devices in a long lesion, the nominal size of the proximally implanted device was compared with the proximal Dmax, whereas the nominal size of the distally implanted device was compared with the distal Dmax. Devices were considered properly sized when the ratio of the proximal or distal Dmax minus the device diameter fell within the range of -0.5 to + 0.5 mm.

Lesion oriented endpoints

The lesion-oriented composite outcome (LOCE) of this analysis included definite device thrombosis, target lesion revascularization (TLR) and target vessel myocardial infarction (TVMI). All myocardial infarctions were defined by the Third Universal Myocardial Infarction definitions and other events were defined according to the Academic Research Consortium definitions. All reported events were adjudicated by an independent Clinical Events Committee (Cardialysis BV., Rotterdam, the Netherlands).

Statistical analysis

This report provides information on the lesion-oriented outcomes subdivided by their device non-oversize or oversize status. All analyses were conducted using the per-protocol population. Descriptive statistics were used to present the data. Groups were compared using Chi square test and student t-test. Kaplan–Meier method, log-rank test and Cox proportional hazards models were used for time-to-event analysis. Landmark analyses were performed at 31 days and 365 days after index procedure. A 2-sided p value < 0.05 was considered significant. The prognostic value for LOCE of device oversize was assessed in a univariate Cox proportional-hazard analysis. In a multivariate Cox regression analysis we identified other statistically significant predictors of LOCE. Next we performed a multivariate regression analysis for device oversize with adjustment for other statistically significant predictors of LOCE.
The entry criterium for the multivariable analysis was set at p < 0.1. Device oversize was forced into the multivariate model. All statistical tests were performed using SPSS software, version 23.0 (IBM SPSS Statistics, IBM Chicago, IL, USA).

Results

In the AIDA trial, 924 patients were randomized to treatment with Absorb BVS and 921 patients to treatment with Xience EES. A total of 2446 lesions were treated; 1237 lesions within the Absorb BVS arm and 1209 within the Xience EES arm. We excluded 89 lesions that did not receive any study device. Another 179 lesion were not analyzable, and were therefore excluded. Pre-procedural Dmax assessment was available in 2152 lesions (87.9%). The full study flowchart is shown in Fig. 1.
Baseline and angiographic characteristics of the oversized and non-oversized groups for both study devices are displayed in Tables 1 and 2. In 653 (62.0%) of Absorb BVS treated lesions, and 648 (59.0%) of Xience EES treated lesions an oversized device was implanted. Oversized devices were more frequently implanted in the left coronary artery, and in lesions with a TIMI flow of 0/1. In both study devices pre-procedural reference vessel diameter (RVD), both proximal and distal Dmax and minimal lumen diameter (MLD), were significantly smaller in the oversized groups compared with the non-oversized groups. Moreover, pre- and post-dilatation were less frequently performed in the Absorb BVS oversized group compared with the non-oversized groups.
Table 1
Shown are the clinical and angiographic characteristics of Absorb BVS treated lesions, stratified by oversized/non-oversized device
https://static-content.springer.com/image/art%3A10.1007%2Fs10554-019-01576-y/MediaObjects/10554_2019_1576_Tab1_HTML.png
Plus-minus values are means ± SD
BVS bioresorbable vascular scaffold, (N)STEMI (non)ST-elevation myocardial infarction, RCA right coronary artery, LAD left anterior descending coronary artery, RCX ramus circumflex, TIMI thrombolysis in myocardial infarction, Dmax maximum diameter
Table 2
Shown are the clinical and angiographic characteristics of Xience EES treated lesions, stratified by oversized/non-oversized device
Xience EES treated lesions
Device oversize
Device non-oversize
P Value
(n = 648)
(n = 450)
STEMI
151 (23.3%)
82 (18.2%)
0.010
NSTEMI
26 (4.0%)
22 (4.9%)
0.502
Lesion location
 RCA
178 (27.5%)
143 (31.8%)
0.138
 LAD
304 (46.9%)
172 (38.2%)
0.004
 RCX
161 (24.8%)
130 (28.9%)
0.144
 LM
2 (0.3%)
4 (0.9%)
0.234
Bypass graft
3 (0.5%)
1(0.2%)
0.648
TIMI 0 or 1
109 (16.8%)
36 (8.1%)
0.001
Calcification
170 (26.2%)
128
0.448
Thrombus present
102 (15.7%)
42 (71.6%)
0.002
Pre-procedural
 Predilatation performed
607 (93.7%)
406 (90.2%)
0.039
 Reference vessel diameter
2.31 ± 0.57
2.79 ± 0.59
< 0.001
 Proximal Dmax
2.58 ± 0.50
3.24 ± 0.65
< 0.001
 Distal Dmax
2.50 ± 0.49
3.04 ± 0.65
< 0.001
 Minimum lumen diameter
0.94 ± 0.37
1.09 ± 0.46
< 0.001
 Diameter stenosis
68.77 ± 18.67
66.06 ± 16.85
< 0.001
Post-procedural
 Post-dilatation performed
295 (45.5%)
243(54.0%)
0.008
 Reference vessel diameter
2.54 ± 0.50
2.76 ± 0.57
< 0.001
 Minimal lumen diameter
1.88 ± 0.48
2.06 ± 0.51
< 0.001
 Diameter stenosis (%)
26.23 ± 11.68
25.30 ± 11.11
0.189
Plus-minus values are means ± SD
EES everolimus eluting stent, (N)STEMI (non)ST-elevation myocardial infarction, RCA right coronary artery, LAD left anterior descending coronary artery, RCX ramus circumflex, TIMI thrombolysis in myocardial infarction, Dmax maximum diameter

Lesion oriented outcomes

Lesion-oriented outcomes are shown in Table 3. LOCE occurred in 48 (7.4%) lesions in the Absorb BVS group oversized group and in 32 (8.2%) lesions in the non-oversized group (HR 0.91; 95%CI 0.58–1.42; p = 0.681).
Table 3
Shown are the lesion oriented outcomes at 2 years of Absorb BVS and Xience EES treated lesions
https://static-content.springer.com/image/art%3A10.1007%2Fs10554-019-01576-y/MediaObjects/10554_2019_1576_Tab3_HTML.png
BVS bioresorbable vascular scaffold, EES everolimus eluting stent, CI confidence interval
aInteraction for device oversize and device modality
bAdjusted for age, male sex and Syntax score for Absorb BVS, and adjusted for hypertension, hypercholesterolemia, any diabetes mellitus, and moderate to severe calcification for Xience EES
LOCE occurred in 35 (5.5%) in the Xience EES group oversized group and in 22 (5.0%) lesions in the non-oversized group (HR 1.11; 95% CI 0.65–1.89; p = 0.703). Kaplan–Meier curves for LOCE at 2 years after Absorb BVS or Xience EES implantation are shown in Figs. 2, 3, respectively.
Within the Absorb BVS arm TLR occurred in 34 (5.3%) lesions in the device oversized group and in 23 lesions (5.9%) in the device non-oversized group (HR 0.89; 95% CI 0.52–1.51; p = 0.666). TLR not caused by scaffold thrombosis occurred at higher, but non-significantly different, rates in the device oversized group (HR 1.62; 95% CI 0.75–3.49; p = 0.211). Definite scaffold thrombosis occurred in 11 (1.7%) device oversized treated lesions against 16 (4.1%) device non-oversized treated lesions (HR 0.41; 95% CI 0.19–0.89; p = 0.020) (Fig. 3). Landmark analyses of definite scaffold thrombosis until 31-days, from 31 days to 1 year, and from 1 to 2 years after scaffold implantation are shown in Fig. 4a–c. Definite scaffold thrombosis from 1 to 2 years occurred less frequently in oversized Absorb BVS treated lesions (HR 0.25; 95% CI 0.07–0.98; p = 0.031). Figure 5a, b show explorative analysis of the distribution of proximal and distal Dmax minus the nominal device size in Absorb BVS treated lesions (A) and Xience EES treated lesions (B) with or without definite device thrombosis. The majority of very late scaffold thrombosis cases occurred in properly sized scaffolds that fell within the definitions of the Instructions for Use of the Absorb BVS and the Dmax assessment. Supplementary Figs. 1 and 2 show the ROC curves of the relative difference between the Dmax and the device diameter and LOCE (1) and definite device thrombosis (2) at 2 years after Absorb BVS or Xience EES implantation.
The predictors of LOCE at 2 years after Absorb BVS or Xience EES implantation are shown in supplementary table 1 and 2, respectively.

Discussion

The main findings of this AIDA sub study are:
1.
We found no difference in Lesion-oriented Composite Endpoint between oversized and non-oversized Absorb BVS treated lesions.
 
2.
Definite Scaffold Thrombosis and Target Vessel Myocardial Infarction occurred more frequently in non-oversized Absorb BVS treated lesions, especially between 1 and 2 year follow-up.
 
3.
Most cases of definite scaffold thrombosis occurred in properly sized Absorb BVS according to the Instructions for Use of Absorb BVS and Dmax assessment.
 
4.
We found no difference in lesion-oriented outcomes between oversized and non-oversized Xience EES treated lesions.
 
Scaffold thrombosis is an unpredictable severe complication, which usually results in myocardial infarction, or even cardiac death. Several studies have investigated whether Absorb BVS optimized implantation techniques are able to resolve the scaffold thrombosis issue. The results of these studies are, however, equivocal. Correct scaffold sizing remains a challenge in routine PCI because appropriate device selection relies on the device to coronary diameter ratio and its potential mismatch [12, 17]. Correct scaffold sizing is particularly difficult in long tapered lesions. It is known that the avoidance of a scaffold to coronary diameter mismatch is associated with less frequent ischemia driven TLR, whereas scaffold oversizing may be associated with higher rates of MACE at 1 year follow-up [12]. Whether this theory also applies to metallic stents remains unknown.
In the current analysis we investigated the relationship between Dmax based device sizing and lesion-oriented outcomes in both the Absorb BVS as the Xience EES treated arm of the AIDA trial. Dmax based device sizing is a lesion specific characteristic, and therefore we present specific lesion-oriented outcomes. In contrast with the findings of Ishibashi et al, our analysis showed that oversized Absorb BVS were not associated with higher rates of lesion oriented events. This may be due to the difference in MI definition in the two studies, namely in AIDA trial there was no routine post-procedure marker measurement and only clinically significant post-procedural myocardial infarctions were captured. However, from 1 to 2 year follow-up, similarly to Katagiri et al. we found significant increase in the rates of device thrombosis in non-oversized Absorb BVS treated lesions. Furthermore, most cases of very late scaffold thrombosis occurred in properly sized devices [13]. This later finding may implicate the complex interaction between vessel wall and the device resorption and dismantling process rather that any initial procedural technique as a major factor in very late stent thrombosis cases.
Our study is the first to investigate the effect of Dmax based device sizing in Xience EES treated lesions. We found no difference in event rates between oversized and non-oversized Xience EES treated lesions. Definite stent thrombosis is unpredictable and does not depend on the Xience EES / coronary artery diameter mismatch. Moreover, all three acute stent thrombosis occurred in properly sized devices, in patients taking dual antiplatelet therapy, suggesting that other factors than implantation technique play a role in the occurrence. These results can potentially be attributed to the better device characteristics of the Xience EES such as thinner struts, better expansion limits, higher tensile and radial strength as compared to Absorb BVS.
The insights of this AIDA sub-study can be useful for the development of the next generation coronary bioresorbable scaffolds, with wider expansion limits and better tensile and radial strength, potentially resolving the scaffold sizing issues and difficulties. In addition, QCA or intravascular imaging should be strongly considered in forthcoming trials of new generation bioresorbable coronary scaffolds.

Limitations

The current analysis has several limitations. First, in AIDA, device sizing based on online QCA was not mandatory, and therefore incorrect device sizing is likely to have occurred more frequently. Second, routine intravascular imaging has not been performed and therefore the study does not provide mechanistic insights in the occurrence of events due to a device / artery diameter mismatch. Third, the role of lesion morphology and typology stratification as a prognostic facture for device failure has not been analyzed in the current explorative analysis. The gold standard for lesion morphology and typology stratification is intracoronary imaging. The AIDA population reflects routine PCI, and since intracoronary imaging is not a part of routine PCI (used in less than 10% of the cases), it was not possible to explore the exact role of these factors in this study. Fourth, as with all post-hoc analysis, this AIDA sub-study is subject to under powering.

Conclusions

In this AIDA trial sub-study, we found no significant difference in LOCE between oversized and non-oversized treated Absorb BVS and Xience EES treated lesions. However, non-oversized Absorb BVS implantation was associated with a higher risk of scaffold thrombosis at complete 2 years follow-up, in particular between 1 and 2 years follow-up.

Acknowledgements

The AIDA QCA study team is immensely grateful to Joelle Elias, Ivo van Dongen, Ineke Verhulst, Esther Scheunhage, Monika van ‘t Hul Gorlewska, Hanny Boutkan, Marie-Angèle Morel, Han Blomer, Andrew Kingsley, Jeannette Schoep, Mirjam van Zeijst, Sander IJsselmuiden, Floris Kauer, Selina de Vlieger, Esther Salentijn, and all interventional cardiology research fellows of the Amsterdam UMC–AMC for their continuous support and assistance during the conduct of the AIDA QCA analyses.

Compliance with ethical standards

Conflict of interest

The Amsterdam UMC Heart Center received an unrestricted educational research grant from Abbott Vascular for the AIDA trial. The Research Department of the cardiology division of the Medical Center Leeuwarden received non-study related unrestricted educational research grants from Abbott Vascular. J.J. Wykrzykowska receives consultancy fees and research grants from Abbott Vascular. J.P.S Henriques receives research grants from Abbott Vascular. J.J. Piek is a member of the Medical Advisory Board of Abbott Vascular. J.G.P. Tijssen served on the DSMB of the early ABSORB trials, including ABSORB II. The other co-authors declare that they have no conflicts of interest.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

Anhänge

Electronic supplementary material

Below is the link to the electronic supplementary material.
Literatur
1.
Zurück zum Zitat Serruys PW, Onuma Y, Garcia-Garcia HM, Muramatsu T, van Geuns RJ, de Bruyne B, Dudek D, Thuesen L, Smits PC, Chevalier B, McClean D, Koolen J, Windecker S, Whitbourn R, Meredith I, Dorange C, Veldhof S, Hebert KM, Rapoza R, Ormiston JA (2014) Dynamics of vessel wall changes following the implantation of the absorb everolimus-eluting bioresorbable vascular scaffold: a multi-imaging modality study at 6, 12, 24 and 36 months. EuroIntervention 9(11):1271–1284CrossRefPubMed Serruys PW, Onuma Y, Garcia-Garcia HM, Muramatsu T, van Geuns RJ, de Bruyne B, Dudek D, Thuesen L, Smits PC, Chevalier B, McClean D, Koolen J, Windecker S, Whitbourn R, Meredith I, Dorange C, Veldhof S, Hebert KM, Rapoza R, Ormiston JA (2014) Dynamics of vessel wall changes following the implantation of the absorb everolimus-eluting bioresorbable vascular scaffold: a multi-imaging modality study at 6, 12, 24 and 36 months. EuroIntervention 9(11):1271–1284CrossRefPubMed
2.
Zurück zum Zitat Onuma Y, Dudek D, Thuesen L, Webster M, Nieman K, Garcia-Garcia HM, Ormiston JA, Serruys PW (2013) Five-year clinical and functional multislice computed tomography angiographic results after coronary implantation of the fully resorbable polymeric everolimus-eluting scaffold in patients with de novo coronary artery disease: the ABSORB cohort A trial. JACC Cardiovasc Interv 6(10):999–1009CrossRefPubMed Onuma Y, Dudek D, Thuesen L, Webster M, Nieman K, Garcia-Garcia HM, Ormiston JA, Serruys PW (2013) Five-year clinical and functional multislice computed tomography angiographic results after coronary implantation of the fully resorbable polymeric everolimus-eluting scaffold in patients with de novo coronary artery disease: the ABSORB cohort A trial. JACC Cardiovasc Interv 6(10):999–1009CrossRefPubMed
3.
Zurück zum Zitat Abizaid A, Ribamar Costa J Jr, Bartorelli AL, Whitbourn R, van Geuns RJ, Chevalier B, Patel T, Seth A, Stuteville M, Dorange C, Cheong WF, Sudhir K, Serruys PW (2015) investigators AE. The ABSORB EXTEND study: preliminary report of the twelve-month clinical outcomes in the first 512 patients enrolled. EuroIntervention 10(12):1396–1401CrossRefPubMed Abizaid A, Ribamar Costa J Jr, Bartorelli AL, Whitbourn R, van Geuns RJ, Chevalier B, Patel T, Seth A, Stuteville M, Dorange C, Cheong WF, Sudhir K, Serruys PW (2015) investigators AE. The ABSORB EXTEND study: preliminary report of the twelve-month clinical outcomes in the first 512 patients enrolled. EuroIntervention 10(12):1396–1401CrossRefPubMed
4.
Zurück zum Zitat Serruys PW, Chevalier B, Sotomi Y, Cequier A, Carrie D, Piek JJ, Van Boven AJ, Dominici M, Dudek D, McClean D, Helqvist S, Haude M, Reith S, de Sousa Almeida M, Campo G, Iniguez A, Sabate M, Windecker S, Onuma Y (2016) Comparison of an everolimus-eluting bioresorbable scaffold with an everolimus-eluting metallic stent for the treatment of coronary artery stenosis (ABSORB II): a 3 year, randomised, controlled, single-blind, multicentre clinical trial. Lancet 388(10059):2479–2491CrossRefPubMed Serruys PW, Chevalier B, Sotomi Y, Cequier A, Carrie D, Piek JJ, Van Boven AJ, Dominici M, Dudek D, McClean D, Helqvist S, Haude M, Reith S, de Sousa Almeida M, Campo G, Iniguez A, Sabate M, Windecker S, Onuma Y (2016) Comparison of an everolimus-eluting bioresorbable scaffold with an everolimus-eluting metallic stent for the treatment of coronary artery stenosis (ABSORB II): a 3 year, randomised, controlled, single-blind, multicentre clinical trial. Lancet 388(10059):2479–2491CrossRefPubMed
5.
Zurück zum Zitat Ellis SG, Kereiakes DJ, Metzger DC, Caputo RP, Rizik DG, Teirstein PS, Litt MR, Kini A, Kabour A, Marx SO, Popma JJ, McGreevy R, Zhang Z, Simonton C, Stone GW, Investigators AI (2015) Everolimus-eluting bioresorbable scaffolds for coronary artery disease. N Engl J Med 373(20):1905–1915CrossRefPubMed Ellis SG, Kereiakes DJ, Metzger DC, Caputo RP, Rizik DG, Teirstein PS, Litt MR, Kini A, Kabour A, Marx SO, Popma JJ, McGreevy R, Zhang Z, Simonton C, Stone GW, Investigators AI (2015) Everolimus-eluting bioresorbable scaffolds for coronary artery disease. N Engl J Med 373(20):1905–1915CrossRefPubMed
6.
Zurück zum Zitat Wykrzykowska JJ, Kraak RP, Hofma SH, van der Schaaf RJ, Arkenbout EK, AJ IJ, Elias J, van Dongen IM, Tijssen RYG, Koch KT, Baan J Jr, Vis MM, de Winter RJ, Piek JJ, Tijssen JGP, Henriques JPS, Investigators A (2017) Bioresorbable scaffolds versus metallic stents in routine PCI. N Engl J Med 376(24):2319–2328CrossRefPubMed Wykrzykowska JJ, Kraak RP, Hofma SH, van der Schaaf RJ, Arkenbout EK, AJ IJ, Elias J, van Dongen IM, Tijssen RYG, Koch KT, Baan J Jr, Vis MM, de Winter RJ, Piek JJ, Tijssen JGP, Henriques JPS, Investigators A (2017) Bioresorbable scaffolds versus metallic stents in routine PCI. N Engl J Med 376(24):2319–2328CrossRefPubMed
7.
Zurück zum Zitat Kereiakes DJ, Ellis SG, Metzger C, Caputo RP, Rizik DG, Teirstein PS, Litt MR, Kini A, Kabour A, Marx SO, Popma JJ, McGreevy R, Zhang Z, Simonton C, Stone GW, Investigators AI (2017) 3-year clinical outcomes with everolimus-eluting bioresorbable coronary scaffolds: the ABSORB III trial. J Am Coll Cardiol 70(23):2852–2862CrossRefPubMed Kereiakes DJ, Ellis SG, Metzger C, Caputo RP, Rizik DG, Teirstein PS, Litt MR, Kini A, Kabour A, Marx SO, Popma JJ, McGreevy R, Zhang Z, Simonton C, Stone GW, Investigators AI (2017) 3-year clinical outcomes with everolimus-eluting bioresorbable coronary scaffolds: the ABSORB III trial. J Am Coll Cardiol 70(23):2852–2862CrossRefPubMed
8.
Zurück zum Zitat Serruys PW, Onuma Y (2017) Dmax for sizing, PSP-1, PSP-2, PSP-3 or OCT guidance: interventionalist’s jargon or indispensable implantation techniques for short- and long-term outcomes of Absorb BRS? EuroIntervention 12(17):2047–56CrossRefPubMed Serruys PW, Onuma Y (2017) Dmax for sizing, PSP-1, PSP-2, PSP-3 or OCT guidance: interventionalist’s jargon or indispensable implantation techniques for short- and long-term outcomes of Absorb BRS? EuroIntervention 12(17):2047–56CrossRefPubMed
9.
Zurück zum Zitat Onuma Y, Serruys PW, Muramatsu T, Nakatani S, van Geuns RJ, de Bruyne B, Dudek D, Christiansen E, Smits PC, Chevalier B, McClean D, Koolen J, Windecker S, Whitbourn R, Meredith I, Garcia-Garcia HM, Veldhof S, Rapoza R, Ormiston JA (2014) Incidence and imaging outcomes of acute scaffold disruption and late structural discontinuity after implantation of the absorb Everolimus-Eluting fully bioresorbable vascular scaffold: optical coherence tomography assessment in the ABSORB cohort B Trial (A Clinical Evaluation of the Bioabsorbable Everolimus Eluting Coronary Stent System in the Treatment of Patients With De Novo Native Coronary Artery Lesions). JACC Cardiovasc Interv 7(12):1400–1411CrossRefPubMed Onuma Y, Serruys PW, Muramatsu T, Nakatani S, van Geuns RJ, de Bruyne B, Dudek D, Christiansen E, Smits PC, Chevalier B, McClean D, Koolen J, Windecker S, Whitbourn R, Meredith I, Garcia-Garcia HM, Veldhof S, Rapoza R, Ormiston JA (2014) Incidence and imaging outcomes of acute scaffold disruption and late structural discontinuity after implantation of the absorb Everolimus-Eluting fully bioresorbable vascular scaffold: optical coherence tomography assessment in the ABSORB cohort B Trial (A Clinical Evaluation of the Bioabsorbable Everolimus Eluting Coronary Stent System in the Treatment of Patients With De Novo Native Coronary Artery Lesions). JACC Cardiovasc Interv 7(12):1400–1411CrossRefPubMed
10.
Zurück zum Zitat Suwannasom P, Sotomi Y, Ishibashi Y, Cavalcante R, Albuquerque FN, Macaya C, Ormiston JA, Hill J, Lang IM, Egred M, Fajadet J, Lesiak M, Tijssen JG, Wykrzykowska JJ, de Winter RJ, Chevalier B, Serruys PW, Onuma Y (2016) The impact of post-procedural asymmetry, expansion, and eccentricity of bioresorbable everolimus-eluting scaffold and metallic everolimus-eluting stent on clinical outcomes in the ABSORB II trial. JACC Cardiovasc Interv 9(12):1231–1242CrossRefPubMed Suwannasom P, Sotomi Y, Ishibashi Y, Cavalcante R, Albuquerque FN, Macaya C, Ormiston JA, Hill J, Lang IM, Egred M, Fajadet J, Lesiak M, Tijssen JG, Wykrzykowska JJ, de Winter RJ, Chevalier B, Serruys PW, Onuma Y (2016) The impact of post-procedural asymmetry, expansion, and eccentricity of bioresorbable everolimus-eluting scaffold and metallic everolimus-eluting stent on clinical outcomes in the ABSORB II trial. JACC Cardiovasc Interv 9(12):1231–1242CrossRefPubMed
11.
Zurück zum Zitat Farooq V, Gomez-Lara J, Brugaletta S, Gogas BD, Garcia-Garcia HM, Onuma Y, van Geuns RJ, Bartorelli A, Whitbourn R, Abizaid A, Serruys PW (2012) Proximal and distal maximal luminal diameters as a guide to appropriate deployment of the ABSORB everolimus-eluting bioresorbable vascular scaffold: a sub-study of the ABSORB Cohort B and the on-going ABSORB EXTEND Single Arm Study. Catheter Cardiovasc Interv 79(6):880–888CrossRefPubMed Farooq V, Gomez-Lara J, Brugaletta S, Gogas BD, Garcia-Garcia HM, Onuma Y, van Geuns RJ, Bartorelli A, Whitbourn R, Abizaid A, Serruys PW (2012) Proximal and distal maximal luminal diameters as a guide to appropriate deployment of the ABSORB everolimus-eluting bioresorbable vascular scaffold: a sub-study of the ABSORB Cohort B and the on-going ABSORB EXTEND Single Arm Study. Catheter Cardiovasc Interv 79(6):880–888CrossRefPubMed
12.
Zurück zum Zitat Ishibashi Y, Nakatani S, Sotomi Y, Suwannasom P, Grundeken MJ, Garcia-Garcia HM, Bartorelli AL, Whitbourn R, Chevalier B, Abizaid A, Ormiston JA, Rapoza RJ, Veldhof S, Onuma Y, Serruys PW (2015) Relation Between bioresorbable scaffold sizing using QCA-Dmax and clinical outcomes at 1 Year in 1,232 patients from 3 study cohorts (ABSORB Cohort B, ABSORB EXTEND, and ABSORB II). JACC Cardiovasc Interv 8(13):1715–1726CrossRefPubMed Ishibashi Y, Nakatani S, Sotomi Y, Suwannasom P, Grundeken MJ, Garcia-Garcia HM, Bartorelli AL, Whitbourn R, Chevalier B, Abizaid A, Ormiston JA, Rapoza RJ, Veldhof S, Onuma Y, Serruys PW (2015) Relation Between bioresorbable scaffold sizing using QCA-Dmax and clinical outcomes at 1 Year in 1,232 patients from 3 study cohorts (ABSORB Cohort B, ABSORB EXTEND, and ABSORB II). JACC Cardiovasc Interv 8(13):1715–1726CrossRefPubMed
13.
Zurück zum Zitat Katagiri Y, Onuma Y, Asano T, Chichareon P, Collet C, Miyazaki Y, Piek JJ, Wykrzykowska JJ, Abizaid A, Ormiston JA, Chevalier B, Serruys PW (2018) Relation between Bioresorbable Scaffold Sizing Using QCA-Dmax and Long-term Clinical Outcomes in 1,232 Patients from 3 Study Cohorts (ABSORB Cohort B, ABSORB EXTEND, and ABSORB II). EuroIntervention 14:e1057–1066CrossRefPubMed Katagiri Y, Onuma Y, Asano T, Chichareon P, Collet C, Miyazaki Y, Piek JJ, Wykrzykowska JJ, Abizaid A, Ormiston JA, Chevalier B, Serruys PW (2018) Relation between Bioresorbable Scaffold Sizing Using QCA-Dmax and Long-term Clinical Outcomes in 1,232 Patients from 3 Study Cohorts (ABSORB Cohort B, ABSORB EXTEND, and ABSORB II). EuroIntervention 14:e1057–1066CrossRefPubMed
14.
Zurück zum Zitat Diletti R, Serruys PW, Farooq V, Sudhir K, Dorange C, Miquel-Hebert K, Veldhof S, Rapoza R, Onuma Y, Garcia-Garcia HM, Chevalier B (2012) ABSORB II randomized controlled trial: a clinical evaluation to compare the safety, efficacy, and performance of the Absorb everolimus-eluting bioresorbable vascular scaffold system against the XIENCE everolimus-eluting coronary stent system in the treatment of subjects with ischemic heart disease caused by de novo native coronary artery lesions: rationale and study design. Am Heart J 164(5):654–663CrossRefPubMed Diletti R, Serruys PW, Farooq V, Sudhir K, Dorange C, Miquel-Hebert K, Veldhof S, Rapoza R, Onuma Y, Garcia-Garcia HM, Chevalier B (2012) ABSORB II randomized controlled trial: a clinical evaluation to compare the safety, efficacy, and performance of the Absorb everolimus-eluting bioresorbable vascular scaffold system against the XIENCE everolimus-eluting coronary stent system in the treatment of subjects with ischemic heart disease caused by de novo native coronary artery lesions: rationale and study design. Am Heart J 164(5):654–663CrossRefPubMed
15.
Zurück zum Zitat Tijssen RYG, Kraak RP, Hofma SH, van der Schaaf RJ, Arkenbout K, Weevers A, Elias J, van Dongen IM, Koch KT, Baan J Jr, Vis M, de Winter RJ, Piek JJ, Tijssen JGP, Henriques JPS, Wykrzykowska JJ. Complete two-year follow-up with formal non-inferiority testing on primary outcomes of the AIDA trial comparing the Absorb bioresorbable scaffold with the XIENCE drug-eluting metallic stent in routine PCI. EuroIntervention. 2018 Tijssen RYG, Kraak RP, Hofma SH, van der Schaaf RJ, Arkenbout K, Weevers A, Elias J, van Dongen IM, Koch KT, Baan J Jr, Vis M, de Winter RJ, Piek JJ, Tijssen JGP, Henriques JPS, Wykrzykowska JJ. Complete two-year follow-up with formal non-inferiority testing on primary outcomes of the AIDA trial comparing the Absorb bioresorbable scaffold with the XIENCE drug-eluting metallic stent in routine PCI. EuroIntervention. 2018
16.
Zurück zum Zitat Woudstra P, Grundeken MJ, Kraak RP, Hassell ME, Arkenbout EK, Baan J Jr, Vis MM, Koch KT, Tijssen JG, Piek JJ, de Winter RJ, Henriques JP, Wykrzykowska JJ (2014) Amsterdam Investigator-initiateD Absorb strategy all-comers trial (AIDA trial): a clinical evaluation comparing the efficacy and performance of ABSORB everolimus-eluting bioresorbable vascular scaffold strategy vs the XIENCE family (XIENCE PRIME or XIENCE Xpedition) everolimus-eluting coronary stent strategy in the treatment of coronary lesions in consecutive all-comers: rationale and study design. Am Heart J 167(2):133–140CrossRefPubMed Woudstra P, Grundeken MJ, Kraak RP, Hassell ME, Arkenbout EK, Baan J Jr, Vis MM, Koch KT, Tijssen JG, Piek JJ, de Winter RJ, Henriques JP, Wykrzykowska JJ (2014) Amsterdam Investigator-initiateD Absorb strategy all-comers trial (AIDA trial): a clinical evaluation comparing the efficacy and performance of ABSORB everolimus-eluting bioresorbable vascular scaffold strategy vs the XIENCE family (XIENCE PRIME or XIENCE Xpedition) everolimus-eluting coronary stent strategy in the treatment of coronary lesions in consecutive all-comers: rationale and study design. Am Heart J 167(2):133–140CrossRefPubMed
17.
Zurück zum Zitat Tijssen RYG, Kraak RP, Elias J, van Dongen IM, Kalkman DN, Nassif M, Sotomi Y, Asano T, Katagiri Y, Collet C, Piek JJ, Henriques JPS, de Winter RJ, Tijssen JGP, Onuma Y, Serruys PW, Wykrzykowska JJ (2018) Implantation techniques (predilatation, sizing, and post-dilatation) and the incidence of scaffold thrombosis and revascularisation in lesions treated with an everolimus-eluting bioresorbable vascular scaffold: insights from the AIDA trial. EuroIntervention 14(4):e434–e442CrossRefPubMed Tijssen RYG, Kraak RP, Elias J, van Dongen IM, Kalkman DN, Nassif M, Sotomi Y, Asano T, Katagiri Y, Collet C, Piek JJ, Henriques JPS, de Winter RJ, Tijssen JGP, Onuma Y, Serruys PW, Wykrzykowska JJ (2018) Implantation techniques (predilatation, sizing, and post-dilatation) and the incidence of scaffold thrombosis and revascularisation in lesions treated with an everolimus-eluting bioresorbable vascular scaffold: insights from the AIDA trial. EuroIntervention 14(4):e434–e442CrossRefPubMed
Metadaten
Titel
The relationship of pre-procedural Dmax based sizing to lesion level outcomes in Absorb BVS and Xience EES treated patients in the AIDA trial
verfasst von
Ruben Y. G. Tijssen
Laura S. M. Kerkmeijer
Yuki Katagiri
Robin P. Kraak
Kuniaki Takahashi
Norihiro Kogame
Ply Chichareon
Rodrigo Modolo
Taku Asano
Martina Nassif
Deborah N. Kalkman
Yohei Sotomi
Carlos Collet
Sjoerd H. Hofma
Rene J. van der Schaaf
E. Karin Arkenbout
Auke P. J. D. Weevers
Marcel A. M. Beijk
Jan J. Piek
Jan G. P. Tijssen
Jose P. Henriques
Robbert J. de Winter
Yoshinobu Onuma
Patrick W. Serruys
Joanna J. Wykrzykowska
Publikationsdatum
25.03.2019
Verlag
Springer Netherlands
Erschienen in
The International Journal of Cardiovascular Imaging / Ausgabe 7/2019
Print ISSN: 1569-5794
Elektronische ISSN: 1875-8312
DOI
https://doi.org/10.1007/s10554-019-01576-y

Weitere Artikel der Ausgabe 7/2019

The International Journal of Cardiovascular Imaging 7/2019 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.