The usability of ventilators: a comparative evaluation of use safety and user experience—an editorial response

verfasst von: Edwin Coombs

Erschienen in: Critical Care | Ausgabe 1/2017

Hinweise
In response to the recent article in Critical Care by Morita et al. [1] entitled “The usability of ventilators: a comparative evaluation of use safety and user experience”, I agree that intensive care unit (ICU) safety and ventilator management are of paramount concern. The need for proper training and clinician education remains a constant priority for many ventilator manufacturers. To that end, the US Food and Drug Administration (FDA) now requires usability testing as part of the 510(k) clearance application [2], and I thank Morita et al. for addressing this important issue.
I would like to point out several discussion topics that I feel result in selection, but unintentional, bias to the study which may have resulted in the outcomes listed in the results of Morita et al. Selection bias occurs when researchers recruit an unrepresentative sample population. The sample population differs in some significant way from the population that generated the sample population [3]. Therefore, generalization of results should be made with caution. The test participants that were recruited were from three hospitals: Duke, Wake-Med, and UNC. These facilities currently utilize the Servo-I, PB840, and Hamilton G5 devices. Having the actual test device or immediate predecessor device of the study device provides test participants entering the testing process with advanced knowledge and experience versus that of equipment that is not used at their respective facilities.
Additionally, the study used “exploration-based training”. According to the FDA guidance on usability tests: “To the extent practicable, the content, format, and method of delivery of training given to test participants should be comparable to the training that actual users would receive”. Respiratory therapists (RTs) would typically receive training from the manufacturer or from manufacturer-certified trainers; therefore, the study should have offered such training before each test session [4]. Knowing that the study did not utilize manufacturer-trained clinical educators, the actual training used as the basis for the study must be questioned.
With these two points explained in the study design, I am of the opinion that the execution of the study with regard to recruited candidates and trainers introduces a level of bias that can easily skew the reported results [5]. I would suggest that a future study be performed with participants that have absolutely no prior experience of the study devices and that training be conducted by a certified manufacturer’s educator.
Thank you again to Morita et al. [1] for posting this study. ICU ventilator safety is of critical importance to everyone who plays a role in the delivery of care.

Acknowledgements

Not applicable.

Funding

None.

Availability of data and materials

Response letter only.

Authors’ contributions

Not applicable.

Authors’ information

EC is a Registered Respiratory Therapist with 30 years of experience and currently the Director of Marketing, Intensive Care, at Draeger, Inc.

Competing interests

EC is a full time employee of Draeger, Inc.
Not applicable.
Not applicable.

Declarations

None.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
5.
Zurück zum Zitat von Hippel PT. Mean, median and skew: correcting a textbook rule. J Statistics Educ. 2005;13(2):ww2.amstat.org/publications/jse/v13n2/vonhippel.html. von Hippel PT. Mean, median and skew: correcting a textbook rule. J Statistics Educ. 2005;13(2):ww2.amstat.org/publications/jse/v13n2/vonhippel.html.
Metadaten
Titel
The usability of ventilators: a comparative evaluation of use safety and user experience—an editorial response
verfasst von
Edwin Coombs
Publikationsdatum
01.12.2017
Verlag
BioMed Central
Erschienen in
Critical Care / Ausgabe 1/2017
Elektronische ISSN: 1364-8535
DOI
https://doi.org/10.1186/s13054-017-1751-9

Weitere Artikel der Ausgabe 1/2017

Critical Care 1/2017 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.