Three-Dimensionally Printed Patient-Specific Surgical Plates Increase Accuracy of Oncologic Head and Neck Reconstruction Versus Conventional Surgical Plates: A Comparative Study

verfasst von: Wei-fa Yang, MDS, Wing Shan Choi, MDS, May Chun-Mei Wong, PhD, Warit Powcharoen, DDS, Wang-yong Zhu, MD, James Kit-Hon Tsoi, PhD, Marco Chow, MPhil, Ka-Wai Kwok, PhD, Yu-xiong Su, MD

Erschienen in: Annals of Surgical Oncology | Ausgabe 1/2021

Abstract

Background

Surgeons are pursuing accurate head and neck reconstruction to enhance aesthetic and functional outcomes after oncologic resection. This study aimed to investigate whether accuracy of head and neck reconstruction is improved with the use of three-dimensionally (3D)-printed patient-specific surgical plates compared with conventional plates.

Methods

In this comparative study, patients were prospectively recruited into the study group (3DJP16) with 3D-printed patient-specific surgical plates. The patients in control group with conventional surgical plates were from a historic cohort in the same unit. The primary end point of the study was the accuracy of head and neck reconstruction. The secondary end points were accuracy of osteotomy, intraoperative blood loss, total operative time, and length of hospital stay.

Results

The study recruited of 33 patients, including 17 in the study group and 16 in the control group. The patients’ baseline characteristics were similar between the two groups. The absolute distance deviation of the maxilla or mandible was 1.5 ± 0.5 mm in the study group and 2.1 ± 0.7 mm in the control group [mean difference, − 0.7 mm; 95% confidence interval (CI) − 1.1 to − 0.3; p = 0.003], showing superior accuracy of reconstruction for the patients with 3D-printed patient-specific surgical plates. Improved accuracy of reconstruction also was detected in terms of bilateral mandibular angles and bone grafts. Concerning the secondary end points, the accuracy of the osteotomy was similar in the two groups. No difference was found regarding intraoperative blood loss, total operative time, or length of hospital stay.

Conclusions

This is the first study to prove that compared with conventional plates, 3D-printed patient-specific surgical plates improve the accuracy of oncologic head and neck reconstruction.
Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1245/​s10434-020-08732-y) contains supplementary material, which is available to authorized users.
Wei-fa Yang and Wing Shan Choi have contributed equally to this work.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Together with the development of microvascular surgery, surgeons are seeking more accurate head and neck reconstruction after oncologic resection to enhance aesthetic and functional outcomes. However, a main disadvantage of autologous bone flaps is mismatch in the shape of donor bone, which needs to be cut and trimmed to fit the defects and better restore the natural appearance. Much effort has been devoted to facilitating bone manipulations, and computer-assisted surgery (CAS) has emerged in the twenty-first century as a viable option.1
With CAS, surgeons make virtual plans on a computer that will be executed in the operating room to guide the precise harvest and placement of bone segments to repair defects. Bone segments can be finely navigated for optimal restoration of the original skeleton. In previous studies, various devices were developed to navigate bone segments according to virtual plans, including cutting templates, rapid prototype skull models, and surgical navigation systems.1 However, the missing link between bone navigation and accurate reconstruction is the plate fixation procedure.2
In conventional procedure, bone segments are fixed using commercial off-the-shelf titanium plates, which should be manually bent and twisted to fit the bone anatomy (Fig. 1). The manual contouring process often is tedious and technique-demanding, and can adversely affect the precise location of bone segments.2 Even worse, the repeated bending can even lead to poor fatigue performance of surgical plates.3 The disadvantages of conventional surgical plates underscore the importance of developing patient-specific surgical plates. Compared with conventional plates, patient-specific surgical plates are designed and manufactured in a three-dimensional (3D) structure that aligns with the individual bone contour. Unlike conventional surgical plates, which need to be contoured based on bone anatomy, patient-specific surgical plates can navigate the folding and precise placement of bone segments and are expected to improve the accuracy of reconstruction.2 Meanwhile, because patient-specific surgical plates need no bending, they can be used to optimize bone reconstruction in a more efficient and standardized manner.
In our previous studies, we successfully manufactured patient-specific titanium surgical plates with high precision through selective laser melting (SLM) technology.2 As a high-tech 3D printing technology, SLM fully melts titanium powders into complete entities in a layer-by-layer manner. The SLM technology enables fabrication of patient-specific surgical plates with tailored structures and good biomechanical properties.4,5 However, evidence concerning the application of three-dimensionally (3D)-printed patient-specific surgical plates in oncologic head and neck reconstruction still is lacking. Whether 3D-printed patient-specific surgical plates can improve the accuracy of reconstruction needs to be investigated.
To investigate the accuracy of head and neck reconstruction with 3D-printed patient-specific surgical plates, the current study quantitatively compared the patient-specific surgical plates with conventional plates. We hypothesized that the 3D-printed patient-specific surgical plates could improve the accuracy of head and neck reconstruction. Proof of this would constitute the highest level of evidence.

Materials and Methods

Study Design

The current study was designed to compare the reconstruction outcomes of 3D-printed patient-specific surgical plates with the outcomes of conventional plates in computer-assisted head and neck reconstruction. Patients receiving 3D-printed patient-specific surgical plates were recruited from the 3DJP16 clinical study (ClinicalTrials.gov Identifier: NCT03057223), an ongoing single-arm, prospective clinical trial studying the application of 3D-printed patient-specific surgical plates in head and neck reconstruction at the Queen Mary Hospital in Hong Kong. The initial nine cases were reported in our previous article.2
In the current study, the accuracy of reconstruction was analyzed by comparison with a retrospective control cohort that underwent CAS using conventional surgical plates. The patients in the control group had surgery performed by the same chief surgeon in the single unit between November 2014 and November 2017. All procedures were executed strictly according to the tenets of the Helsinki Declaration. This report has been prepared to meet all recommendations in the STROBE guideline.

Participants

Patients were eligible for this study if they were older than 18 years, provided the signed informed consent form, had undergone computer-assisted maxillary or mandibular reconstruction using autologous bony free flaps after tumor resection, and agreed to comply with the follow-up procedures. Patients were excluded from the study if they were pregnant, unable to take computed tomography (CT) scans, had unstable or complicated systemic diseases that contradict the surgical process, had severe systemic diseases or conditions that compromise flap survival and the healing process, had reconstruction with nonvascular bone grafts, or had incomplete medical records including preoperative virtual plans and follow-up CT data.

Computer-Assisted Surgery

The techniques of oncologic head and neck reconstruction using CAS in our unit have been described previously.2,6 Basically, CAS was composed of three main phases: the preoperative phase of virtual surgery and 3D printing of patient-specific devices, the intraoperative phase of precision-enhanced real surgery using patient-specific devices, and the postoperative phase of accuracy analysis.1 (Fig. 1).
In the preoperative phase, the virtual surgery was performed by surgeons using ProPlan CMF 2.0 software (Materialise, Leuven, Belgium). In ProPlan’s interactive interface, the patient’s CT data was initially segmented to construct 3D virtual models of the maxilla or mandible. Then, bone resection was performed in 3D virtual models for the en bloc removal of the tumor. Meanwhile, bone grafts were harvested from the iliac crest or fibula to repair defects and restore the normal appearance. Finally, the virtually reconstructed maxilla or mandible was used to design patient-specific devices for surgery.

3D Printing of Patient-Specific Devices

All patient-specific devices were designed in 3-Matic 13.0 (Materialise) using the “surgeon-dominated” approach.6 Cutting guides were designed by wrapping the bone surface to guide accurate bone resection. The cutting guides then were printed by Fused Deposition Manufacturing (FDM) using ULTEM 1010 or by Stereolithography using MED610 resin (Stratasys Ltd., Eden Prairie, MN, USA). Both ULTEM 1010 and MED610 are Food and Drug Administration (FDA)-cleared biocompatible materials suitable for high-temperature autoclaving.
Patient-specific surgical plates were designed by delineating a plate path on the bone surface and then placing the screw holes. Surgical plates were generated by the built-in command in 3-Matic. After that, surgical plates were fabricated by SLM using grade 2 titanium powder.

Surgical Procedures

In the current study, all the patients in both groups had undergone CAS. During the surgery, patient-specific cutting guides enabled precise oncologic resection and bone flap osteotomy. However, in the control group, the arrangement of the bone segments was manipulated according to the mutual alignment of osteotomy planes, occlusion, and mandible contour through the surgeon’s experience and judgment call during the surgery. The bone segments were stabilized using commercial titanium surgical plates (DePuy Synthes, Raynham, MA, USA), which were bent manually, then fixed by screws710 (Fig. 1).
In the study group, except for the cutting guides, the patient-specific surgical plates were designed and fabricated to custom-fit the bone contour of the reconstructed maxilla or mandible.6 The screw holes embedded in the cutting guides corresponded to those in the 3D-printed patient-specific surgical plates and guided the position of the surgical plates and bone segments, thus facilitating the folding, positioning, and fixing of the bone segments in real surgery2 (Fig. 1). The standard perioperative management was similar in the two groups. Postoperative follow-up assessment was performed in the routine manner.

Outcomes

The primary end point of the study was the accuracy of reconstruction, defined as the distance or angulation deviations of anatomic landmarks between the virtual plan and the actual surgical outcome. Various outcome parameters were established focusing on the entire maxilla or mandible, condylar head, mandibular angle, and bone grafts2,11 (Fig. 2). The postoperative skull model was repositioned to coincide with the preoperative virtually reconstructed skull based on the sides of maxilla or mandible not treated by surgery.12,13. The absolute distance deviation of the maxilla or mandible was measured by calculating the distance between pre- and postoperative models based on points, which was automatically calculated by the built-in function of the software.2,9,10
In evaluating the spatial displacement of the condylar head, an intercondylar line was created by connecting the most superior points of the bilateral condylar heads. The distance and angulation deviations of the condyle were derived by comparing the the pre- and postoperative intercondylar lines. The distance deviation was defined as the difference in length between the pre- and postoperative intercondylar lines, and the angulation deviation was the angle formed by the pre- and postoperative intercondylar lines.1418 Similarly, we connected the most posterior inferior points of bilateral mandibular angles to form the intergonial line. The pre- and postoperative intergonial lines were compared to evaluate the distance and angulation deviations of the mandibular angle.1420
In measuring the displacement of the transplanted bone grafts, the center point and central axis of each bone graft were generated. The distance deviation of the bone graft was defined as the distance between the pre- and postoperative center points, and the angulation deviation was the angle between the pre- and postoperative central axes of the bone grafts.2123
The secondary end points of the study were accuracy of osteotomy, intraoperative blood loss, total operative time, and length of hospital stay. In measuring the accuracy of skull bone resection or bone graft osteotomy, each bone segment was separately registered to the corresponding preoperative entity based on designated landmarks (Fig. 2). We initially delineated the osteotomy planes and located the center points of each osteotomy plane. Then the angle formed by the pre- and postoperative osteotomy planes was defined as the angulation deviation of osteotomy, whereas the distance deviation of osteotomy referred to the distance between the center points of the pre- and postoperative osteotomy planes.2123 Intraoperative blood loss, total operative time, and length of hospital stay also were assessed as secondary end points.
Postoperative adverse events associated with the surgical plates were summarized. All perioperative information was retrieved from the hospital’s medical database. The accuracy of both reconstruction and osteotomy was measured by two independent operators.

Statistical Analysis

Dichotomous values were depicted as counts (n) with proportion and compared using the Chi square test or Fisher’s exact test as indicated. Continuous data are expressed as mean values with standard deviation (SD) for normally distributed data and compared using the independent-samples t test. Skewed data are expressed as medians with interquartile range (IQR) or range and compared using the Mann–Whitney U test. Mean differences of accuracy are accompanied with 95% confidence intervals (CIs). All tests and reported p values are two-sided. Statistical significance is set at a p value lower than 0.05. To account for the multiple measuring of the primary end point, the adjusted p value of 0.05/7 = 0.007 is used. All statistics were calculated using SPSS Statistics (version 22.0; SPSS, IBM Corporation, Chicago, IL, USA) and GraphPad Prism 5 (GraphPad Software Inc, San Diego, CA, USA).

Results

Clinical Demographic Characteristics

The study included 33 patients. Between December 2016 and July 2018, 19 patients underwent oncologic head and neck reconstruction using 3D-printed patient-specific surgical plates in the 3DJP16 clinical study. Two patients were excluded from the study because the one patient had reconstruction using nonvascular bone graft and the other patient with comorbidities of renal failure and poorly controlled diabetes experienced late-stage artery thrombosis and flap failure. Therefore, the study group included 17 patients for accuracy analysis who were managed with 3D-printed patient-specific surgical plates. The 17 cases involved 3 maxillary reconstructions and 14 mandibular reconstructions.
The control group comprised 19 patients who underwent CAS using conventional surgical plates between November 2014 and November 2017. Two patients were excluded from the analysis due to loss of preoperative virtual plans, and one patient was excluded due to lack of follow-up CT data. The study included 16 patients in the control group for accuracy analysis. These 16 cases involved 3 maxillary reconstructions and 13 mandibular reconstructions (Fig. 1).
The demographics and baseline characteristics were similar between the study and control groups including the distribution of age, sex, lesion types, maxillary or mandibular defects, and types of free bony flaps including fibula or iliac crest (Table 1). Four patients in the study group underwent “double-barrel” fibula flap mandibular reconstruction with the aid of 3D-printed patient-specific surgical plates compared with no patients in the control group. Eight patients in the study group received postoperative radiotherapy (RT) or concurrent chemoradiotherapy (CRT) compared with four patients in the control group (p = 0.19). The postoperative follow-up CT data were retrieved 1.8 ± 0.8 months after surgery in the study group and 17.3 ± 11.2 months after surgery in the control group.
Table 1
Demographics and baseline characteristics of the patients
Characteristic
Patient-specific surgical plates group (n = 17)
n (%)
Control group (n = 16)
n (%)
p Value
Age (years)
55.6 ± 14.4
55.4 ± 16.0
0.97a
Sex
  
0.71b
 Male
4 (23.5)
5 (31.3)
 
 Female
13 (76.5)
11 (68.8)
 
Lesion type
  
1.00b
 Benign
3 (17.6)
4 (25.0)
 
 Malignant
12 (70.6)
11 (68.8)
 
 Otherc
2 (11.8)
1 (6.3)
 
pT3/T4 cancerd
6 (35.3)
6 (37.5)
0.90e
Defect site
  
1.00b
 Maxilla
3 (17.6)
3 (18.8)
 
 Mandible
14 (82.4)
13 (81.3)
 
Donor bone graftf
  
1.00b
 Fibula
15 (88.2)
15 (93.8)
 
 Iliac crest
2 (11.8)
1 (6.3)
 
Donor bone length (mm)
95.8 ± 38.7
88.2 ± 21.6
0.50a
No. of bone graft segments
  
0.89g
 Median
2
2
 
 Range
1–4
1–3
 
Plate thickness (mm)h
   
 Median
1.8
NA
 
 Range
0.8–2.0
NA
 
Plate width (mm)h
   
 Median
4.0
NA
 
 Range
3.0–4.5
NA
 
Postoperative RT/CRT
8 (47.1)
4 (25.0)
0.19e
NA not applicable, RT radiotherapy, CRT chemoradiotherapy
aThis p value was calculated by means of the independent-samples t test
bThis p value was calculated by means of Fisher’s exact test
cOther included secondary reconstruction and osteoradionecrosis
dPathologic tumor category according to American Joint Committee on Cancer staging for head and neck cancer, 8th edition
eThis p value was calculated by means of the Chi square test
fAll defects were reconstructed using microvascular free flap
gThis p value was calculated by means of the Mann–Whitney U test
hPlate thickness and width were based on the main body because the patient-specific surgical plate was not uniformly shaped

Primary End Point

The primary end point, accuracy of reconstruction, was derived by comparing the postoperative maxilla or mandible with the preoperative virtual models. The inter-operator agreement was good to excellent (Table S1). The mean absolute distance deviation of the maxilla or mandible was 1.5 ± 0.5 mm in the study group and 2.1 ± 0.7 mm in the control group (mean difference, − 0.7 mm; 95% CI − 1.1 to − 0.3 mm; p = 0.003), showing superior accuracy of reconstruction outcomes in the study group with 3D-printed patient-specific surgical plates (Fig. 3a; Fig. S1).
For mandibular reconstruction, we aimed to evaluate the spatial displacement of bilateral condylar heads. The distance deviation between the pre- and postoperative bilateral condylar heads was lower in the study group with 3D-printed patient-specific surgical plates (2.6 ± 3.0 mm), but did not differ significantly from the control group (5.2 ± 4.2 mm) (mean difference − 2.6 mm; 95% CI − 5.5 to 0.3 mm; p = 0.076). Similarly, no significant difference was detected in the angulation deviation of the bilateral condylar heads (mean difference − 1.7°; 95% CI − 4.5° to 1.0°; p = 0.21) (Fig. 3b).
We further analyzed the deviations of bilateral mandibular angles. The 3D-printed patient-specific surgical plates significantly increased the accuracy of reconstruction in terms of distance deviation (mean difference − 5.6 mm; 95% CI − 9.4 to − 1.8 mm; p = 0.005) and angulation deviation (mean difference − 4.0°; 95% CI − 6.6 to − 1.4; p = 0.005) of intergonial lines compared with the control group (Fig. 3c).
The dislocation of transplanted bone segments was evaluated through the center point and central axis of each bone graft. The study group had 38 transplanted bone segments eligible for accuracy analysis compared with 35 in the control group. The distance deviation of the bone grafts was 3.1 ± 1.4 mm in the study group versus 8.7 ± 4.2 mm in the control group (mean difference − 5.6 mm; 95% CI − 7.0 to − 4.2 mm; p < 0.001). The angulation deviation also was significantly less in the study group than in the control group (mean difference − 6.6°; 95% CI − 8.7 to − 4.6; p < 0.001; Figs. 3d; Fig. S2).

Secondary End Points

We also evaluated the accuracy of osteotomy at both recipient and donor sites. The inter-operator agreement was fair to good (Table S1). No significant difference in accuracy of the osteotomy was detected between the study and control groups (Fig. 4; Fig. S3). As shown in Fig. 4a, the distance deviation of the maxilla or mandible resection was 3.2 ± 1.3 mm in the patients with 3D-printed patient-specific surgical plates and 3.3 ± 2.9 mm in the control group (p = 0.78). The angulation deviations did not differ significantly between the two groups (p = 0.45). As shown in Fig. 4b, the osteotomy of the bone grafts did not differ significantly between the study and control groups in terms of either distance deviation (p = 0.17) or angulation deviation (p = 0.10).
The intraoperative blood loss in the study group was similar to that in control group (p = 0.31) (Table 2). Likewise, the total operative time was comparable between the two groups (p = 0.99). The hospital stay was 15.1 ± 3.6 days in the 3D-printed patient-specific surgical plates group compared with 17.0 ± 5.7 days in the control group (p = 0.25). No major postoperative adverse events at recipient sites were identified in either group. Due to the small number of participants, statistical comparisons were not performed. Two patients in the control group underwent plate removal due to plate exposure compared with no patients in the study group.
Table 2
Intraoperative blood loss, total operative time, hospital stay, and postoperative adverse events at the recipient site
Characteristic
Patient-specific surgical plates group (n = 17)
Control group (n = 16)
p value
Estimated blood loss (ml)
  
0.31a
 Median
800
900
 
 IQR
500–1000
500–1300
 
Total operative time (min)
  
0.99a
 Median
510
496
 
 IQR
453–602.5
446.8–630.5
 
Mean hospital stay (days)
15.1 ± 3.6
17.0 ± 5.7
0.25b
Adverse events: n (%)
   
 Wound infection
2 (11.8)
3 (18.8)
 
 Plate breakage
1 (5.9)
0
 
 Plate removal
0
2 (12.5)
 
 Bone mal-/non-union
0
0
 
IQR interquartile range
aThis p value was calculated by means of the Mann–Whitney U test
bThis p value was calculated by means of the independent-samples t test
Bony union was achieved for all the bone segments in both groups, as shown on an orthopantomogram during the follow-up assessment. The occlusal functions were satisfactory for most of the patients with 3D-printed patient-specific surgical plates. For the patients with indications, dental implants were placed in the second stage to restore missing teeth, as illustrated in Fig. 5.

Discussion

Our previous prospective single-arm pilot study explored the feasibility of using 3D-printed patient-specific surgical plates in head and neck reconstruction.2 Based on this, the current study further demonstrated that compared with conventional surgical plates, 3D-printed patient-specific surgical plates reduced the distance deviations of reconstructed maxilla or mandible, distance and angulation deviations of bilateral mandibular angles, and transplanted bone segments, thus leading to enhanced reconstruction accuracy. To the best of our knowledge, this is the first study to investigate the accuracy of outcomes from 3D-printed patient-specific surgical plates versus conventional surgical plates in head and neck reconstruction.
Computer technology enables CAS to study the accuracy of head and neck reconstruction by comparing postoperative reconstruction outcomes with preoperative virtual plans. However, no standard parameters exist to date.11 Different studies have tended to use different measurements, which severely limits the interpretation and comparison of results.11,24
The current study used a series of parameters to evaluate the accuracy of reconstruction systematically. The absolute distance deviation of the maxilla or mandible is the parameter most commonly used to evaluate the overall accuracy of reconstruction due to its simplicity and intuitive form for comparison.2
Considering the crucial functional roles of the mandible, we also assessed the displacement of condylar heads and mandibular angles, which are the most commonly evaluated landmarks according to the literature.11 Because the accuracy of condylar heads is affected bilaterally, we used the intercondylar line to connect the bilateral condylar heads to evaluate the overall accuracy of condylar heads. Meanwhile, it is simpler and more straightforward to understand and compare different studies by using the intercondylar line.2
Similarly, bilateral mandibular angles also were connected and evaluated using the intergonial line. In the Yu et al.25 study of mandibular reconstruction using CAS and navigation, the mean distance deviations of the condyle and gonion were respectively 9.3 ± 2.6 mm and 7.3 ± 2.5 mm, higher than those in our study. In the study of Bao et al.13 that analyzed pre-bent surgical plates, the distance deviation of the condyle was 6.0 ± 1.5 mm, and the distance deviation of the gonion was 5.8 ± 1.2 mm, comparable with those in our control group. However, if the pre-bent surgical plates were used together with screw-predesignated surgical guides, the distance deviations of the condyle and gonion were reduced to 2.6 ± 0.8 mm and 3.2 ± 0.8 mm, comparable with those in our study group, confirming that the screw-drilling guidance is important to enhancing the accuracy of reconstruction.13
In evaluating the accuracy of bone reconstruction, we delineated and compared the center point and central axis of each bone graft, similar to the method in the Schepers et al. study using CNC-milled patient-specific surgical plates. In the seven cases studied by Schepers et al.,26 the distance deviation of fibular bone grafts was 2.9 ± 1.7 mm, and the angulation deviation was 4.3 ± 3.2°, approximating our results for our study group.26 The inter-operator agreement for reconstruction measurements in this study was at least good, indicating the high reliability of the measured parameters.27
With the advancement of microvascular surgery, the survival rates for vascularized bone flaps have been ultimately improved, and surgeons are pursuing more accurate reconstruction results for enhancement of aesthetic and functional outcomes. It is well-known that bone deviations may severely affect factors such as aesthetics, occlusion, and condylar positions, highlighting the clinical significance of improved reconstruction accuracy. However, it is a concerning question whether the quantitative accuracy is clinically relevant or not.
On the one hand, the 3D-printed patient-specific plates improved the accuracy of reconstruction by one-third in the entire maxilla or mandible, by one half in the distance deviation of the bilateral condylar heads, and by nearly three-fourths in the distance deviation of the bilateral mandibular angles compared with conventional plates. The improved accuracy definitely is clinically significant for enhancement of aesthetics, occlusion, and condylar functions. Based on the increased accuracy, we currently have started simultaneous dental implantation during surgery, which further pushes forward this exciting new frontier.
On the other hand, the small inaccuracy with hard tissue can be covered by soft tissue in aesthetics. Meanwhile, the mobile mandible also can accommodate slight malocclusion and minor joint dislocation. Knowledge of the minimal clinically important difference (MCID) in head and neck reconstruction still is lacking.1 Therefore, the relationship between quantitative accuracy and clinical outcomes remains to be elucidated, and the clinical relevance of our results cannot be overinterpreted.
The lack of MCID in head and neck reconstruction further highlights the importance of the current quantitative study. The measured outcomes in the current study may provide a good reference for future studies to clarify the clinically relevant accuracy in head and neck reconstruction.
In addition to the clinical significance, the improved accuracy endowed by 3D-printed patient-specific surgical plates is particularly important for the continuing progress of head and neck reconstruction because it enhances the predictability and repeatability of accurate head and neck reconstruction to ensure quality control of surgery, especially for young surgeons and residents.6 For example, the 3D-printed patient-specific surgical plates can cope with more challenging cases.2830
In the current study, more patients in the study group had reconstruction with the “double-barrel” fibula flaps, which usually are too complicated to be arranged accurately by conventional plates considering the multiple bone segments.31 However, with the 3D-printed patient-specific surgical plates, all bone segments can be easily arranged and secured. This underscores the importance of technological advancements of patient-specific surgical plates in head and neck reconstruction.
The accuracy of osteotomy also was evaluated. The basic principle of cutting guides is to direct the exact position and direction of osteotomy, so we delineated the osteotomy planes and located corresponding center points for accuracy analysis.2123 In the study by Maesschalck et al.32, the distance deviations of osteotomy were 2.3 ± 1.0 mm in the mandible and 1.9 ± 1.1 mm in the fibula, comparable with our results. As expected, we found no significant difference in osteotomy between the two groups because both groups used cutting guides for osteotomy, confirming that the improved accuracy of reconstruction is induced by the 3D-printed patient-specific surgical plates.
From our experience, application of 3D-printed patient-specific surgical plates facilitated the surgical manipulation and reduced the operative time, as reported in other studies.28,33,34 However, in the current study, these parameters did not differ significantly between the study and control groups. However, these results need to be interpreted with caution. These end points are significantly affected by multiple confounding factors, especially comorbidities of the patients, extension of the tumors, and complexity of the surgery.35
For further verification of these outcomes, more patients with well-controlled confounding factors should be recruited. Furthermore, our previous study indicated that CAS saves time and increases efficiency compared with conventional freehand surgery.36 Therefore, the marginal increased efficiency of the 3D-printed patient-specific surgical plates may not have been well reflected, especially considering that the number of patients was small in this study.
Another issue calling for discussion is the cost-effectiveness, mechanical strength, and accessibility of the 3D-printed patient-specific surgical plates.1 Because medical treatment in our Hong Kong public hospital is supported primarily by the government, the current study had no cost-effectiveness analysis. Due to the high cost of CAS and 3D-printing technology in other countries or areas, the wide application of 3D-printed patient-specific surgical plates may be restricted. In our study, the additive manufacturing of titanium was performed using the SLM technology. According to our knowledge, the yield strength, tensile strength, elastic modulus, and Poisson’s ratio of melted titanium plates are similar to those of forged plates. However, the elongation properties are poorer, which should be confirmed and improved in additional studies.
Concerning the accessibility of SLM, with our proposed “surgeon-dominated” workflow, surgeons are encouraged to design patient-specific surgical plates, whereas expert engineers are responsible for the optimization and production.6 The titanium SLM is explosive and should be performed with special equipment under an argon atmosphere, which cannot be completed without industrial support.6 However, the enhanced reconstruction accuracy, as shown in the current study, supports the rationale for applying 3D-printed patient-specific surgical plates in head and neck reconstruction, which likely would promote the development and generalization of this new technology and reduce the cost in the near future.
Certain limitations of the present study need to be addressed. First, our study was conducted by the same chief surgeon, thus precluding the potential biases incurred with multiple surgical teams that may have varied surgical experience and preferences.37 Second, the follow-up periods differed between the study and control groups. In evaluating the accuracy of reconstruction, we recommend that follow-up CT scans be collected within 3 months after surgery, as in the study group.24 However, with more accurate restoration of anatomic structures using CAS, we predict that only minor changes would occur in the long term, which should not jeopardize the main results in this study. The longitudinal changes in maxilla or mandible reconstruction should be investigated in future studies for a comparison of both the short- and long-term outcomes.
Third, although it was reasonable to investigate the accuracy of head and neck reconstruction by including both mandible and maxilla patients, we acknowledge that the sample was too small for a compelling subgroup analysis. However, our results still were sufficient to show the increased accuracy provided by patient-specific surgical plates. Especially considering the novelty and relatively the few studies analyzing patient-specific plates, our study may promote more research and provide a good reference for future studies.
In conclusion, this is the first study to prove that compared with conventional plates, 3D-printed patient-specific surgical plates can precisely transfer a virtual surgical plan to real surgery, leading to enhanced accuracy of oncologic head and neck reconstruction. This quantitative study constitutes the highest level of evidence to date. The increased prevalence 3D-printed patient-specific surgical plates used in clinical practice should lead to well-designed prospective clinical trials to confirm our results in the near future.

Acknowledgment

This study was supported by the Health and Medical Research Fund (project no.: 05161626), Food and Health Bureau, Hong Kong. The authors thank Chun-yu Zhang and Xian-shuai Chen (Guangzhou Janus Biotechnology Co., Ltd., Guangzhou, China), and Prof. Ruxu Du (Shien-Ming Wu School of Intelligent Engineering, Guangzhou International Campus of South China University of Technology, Guangzhou, China) for their support in the 3D printing of titanium.

DISCLOSURES

All authors declare that they have no conflict of interest.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Anhänge

Electronic supplementary material

Below is the link to the electronic supplementary material.
Literatur
1.
Zurück zum Zitat Powcharoen W, Yang WF, Yan Li K, Zhu W, Su YX. Computer-assisted versus conventional freehand mandibular reconstruction with fibula free flap: a systematic review and meta-analysis. Plast Reconstr Surg. 2019;144:1417–28.CrossRef Powcharoen W, Yang WF, Yan Li K, Zhu W, Su YX. Computer-assisted versus conventional freehand mandibular reconstruction with fibula free flap: a systematic review and meta-analysis. Plast Reconstr Surg. 2019;144:1417–28.CrossRef
2.
Zurück zum Zitat Yang WF, Choi WS, Leung YY, et al. Three-dimensional printing of patient-specific surgical plates in head and neck reconstruction: a prospective pilot study. Oral Oncol. 2018;78:31–6.CrossRef Yang WF, Choi WS, Leung YY, et al. Three-dimensional printing of patient-specific surgical plates in head and neck reconstruction: a prospective pilot study. Oral Oncol. 2018;78:31–6.CrossRef
4.
Zurück zum Zitat Shaoki A, Xu JY, Sun HP, et al. Osseointegration of three-dimensional designed titanium implants manufactured by selective laser melting. Biofabrication. 2016;8:045014.CrossRef Shaoki A, Xu JY, Sun HP, et al. Osseointegration of three-dimensional designed titanium implants manufactured by selective laser melting. Biofabrication. 2016;8:045014.CrossRef
5.
Zurück zum Zitat Xu JY, Chen XS, Zhang CY, Liu Y, Wang J, Deng FL. Improved bioactivity of selective laser melting titanium: surface modification with micro-/nano-textured hierarchical topography and bone regeneration performance evaluation. Mater Sci Eng C Mater Biol Appl. 2016;68:229–40.CrossRef Xu JY, Chen XS, Zhang CY, Liu Y, Wang J, Deng FL. Improved bioactivity of selective laser melting titanium: surface modification with micro-/nano-textured hierarchical topography and bone regeneration performance evaluation. Mater Sci Eng C Mater Biol Appl. 2016;68:229–40.CrossRef
6.
Zurück zum Zitat Yang WF, Zhang CY, Choi WS, et al. A novel “surgeon-dominated” approach to the design of 3D-printed patient-specific surgical plates in mandibular reconstruction: a proof-of-concept study. Int J Oral Maxillofac Surg. 2020;49:13–21.CrossRef Yang WF, Zhang CY, Choi WS, et al. A novel “surgeon-dominated” approach to the design of 3D-printed patient-specific surgical plates in mandibular reconstruction: a proof-of-concept study. Int J Oral Maxillofac Surg. 2020;49:13–21.CrossRef
7.
Zurück zum Zitat Zheng GS, Su YX, Liao GQ, et al. Mandible reconstruction assisted by preoperative virtual surgical simulation. Oral Surg Oral Med Oral Pathol Oral Radiol. 2012;113:604–11.CrossRef Zheng GS, Su YX, Liao GQ, et al. Mandible reconstruction assisted by preoperative virtual surgical simulation. Oral Surg Oral Med Oral Pathol Oral Radiol. 2012;113:604–11.CrossRef
8.
Zurück zum Zitat Zheng GS, Su YX, Liao GQ, et al. Mandible reconstruction assisted by preoperative simulation and transferring templates: cadaveric study of accuracy. J Oral Maxillofac Surg. 2012;70:1480–5.CrossRef Zheng GS, Su YX, Liao GQ, et al. Mandible reconstruction assisted by preoperative simulation and transferring templates: cadaveric study of accuracy. J Oral Maxillofac Surg. 2012;70:1480–5.CrossRef
9.
Zurück zum Zitat Zheng GS, Su YX, Liao GQ, Liu HC, Zhang SE, Liang LZ. Mandibular reconstruction assisted by preoperative simulation and accurate transferring templates: preliminary report of clinical application. J Oral Maxillofac Surg. 2013;71:1613–8.CrossRef Zheng GS, Su YX, Liao GQ, Liu HC, Zhang SE, Liang LZ. Mandibular reconstruction assisted by preoperative simulation and accurate transferring templates: preliminary report of clinical application. J Oral Maxillofac Surg. 2013;71:1613–8.CrossRef
10.
Zurück zum Zitat Zheng GS, Wang L, Su YX, Liao GQ, Zhang SE, Lao XM. Maxillary reconstruction assisted by preoperative planning and accurate surgical templates. Oral Surg Oral Med Oral Pathol Oral Radiol. 2016;121:233–8.CrossRef Zheng GS, Wang L, Su YX, Liao GQ, Zhang SE, Lao XM. Maxillary reconstruction assisted by preoperative planning and accurate surgical templates. Oral Surg Oral Med Oral Pathol Oral Radiol. 2016;121:233–8.CrossRef
11.
Zurück zum Zitat van Baar GJC, Forouzanfar T, Liberton N, Winters HAH, Leusink FKJ. Accuracy of computer-assisted surgery in mandibular reconstruction: a systematic review. Oral Oncol. 2018;84:52–60.CrossRef van Baar GJC, Forouzanfar T, Liberton N, Winters HAH, Leusink FKJ. Accuracy of computer-assisted surgery in mandibular reconstruction: a systematic review. Oral Oncol. 2018;84:52–60.CrossRef
12.
Zurück zum Zitat Ayoub N, Ghassemi A, Rana M, et al. Evaluation of computer-assisted mandibular reconstruction with vascularized iliac crest bone graft compared to conventional surgery: a randomized prospective clinical trial. Trials. 2014;15:114.CrossRef Ayoub N, Ghassemi A, Rana M, et al. Evaluation of computer-assisted mandibular reconstruction with vascularized iliac crest bone graft compared to conventional surgery: a randomized prospective clinical trial. Trials. 2014;15:114.CrossRef
13.
Zurück zum Zitat Bao T, He J, Yu C, et al. Utilization of a pre-bent plate-positioning surgical guide system in precise mandibular reconstruction with a free fibula flap. Oral Oncol. 2017;75:133–9.CrossRef Bao T, He J, Yu C, et al. Utilization of a pre-bent plate-positioning surgical guide system in precise mandibular reconstruction with a free fibula flap. Oral Oncol. 2017;75:133–9.CrossRef
14.
Zurück zum Zitat Foley BD, Thayer WP, Honeybrook A, McKenna S, Press S. Mandibular reconstruction using computer-aided design and computer-aided manufacturing: an analysis of surgical results. J Oral Maxillofac Surg. 2013;71:e111–9.CrossRef Foley BD, Thayer WP, Honeybrook A, McKenna S, Press S. Mandibular reconstruction using computer-aided design and computer-aided manufacturing: an analysis of surgical results. J Oral Maxillofac Surg. 2013;71:e111–9.CrossRef
15.
Zurück zum Zitat Huang JW, Shan XF, Lu XG, Cai ZG. Preliminary clinic study on computer-assisted mandibular reconstruction: the positive role of surgical navigation technique. Maxillofac Plast Reconstr Surg. 2015;37:20.CrossRef Huang JW, Shan XF, Lu XG, Cai ZG. Preliminary clinic study on computer-assisted mandibular reconstruction: the positive role of surgical navigation technique. Maxillofac Plast Reconstr Surg. 2015;37:20.CrossRef
16.
Zurück zum Zitat Wilde F, Winter K, Kletsch K, Lorenz K, Schramm A. Mandible reconstruction using patient-specific pre-bent reconstruction plates: comparison of standard and transfer key methods. Int J Comput Assist Radiol Surg. 2015;10:129–40.CrossRef Wilde F, Winter K, Kletsch K, Lorenz K, Schramm A. Mandible reconstruction using patient-specific pre-bent reconstruction plates: comparison of standard and transfer key methods. Int J Comput Assist Radiol Surg. 2015;10:129–40.CrossRef
17.
Zurück zum Zitat Zhang L, Liu Z, Li B, Yu H, Shen SG, Wang X. Evaluation of computer-assisted mandibular reconstruction with vascularized fibular flap compared to conventional surgery. Oral Surg Oral Med Oral Pathol Oral Radiol. 2016;121:139–48.CrossRef Zhang L, Liu Z, Li B, Yu H, Shen SG, Wang X. Evaluation of computer-assisted mandibular reconstruction with vascularized fibular flap compared to conventional surgery. Oral Surg Oral Med Oral Pathol Oral Radiol. 2016;121:139–48.CrossRef
18.
Zurück zum Zitat Zheng L, Lv X, Zhang J, Liu S, Zhang J, Zhang Y. Translating computer-aided design and surgical planning into successful mandibular reconstruction using a vascularized iliac-crest flap. J Oral Maxillofac Surg. 2018;76:886–93.CrossRef Zheng L, Lv X, Zhang J, Liu S, Zhang J, Zhang Y. Translating computer-aided design and surgical planning into successful mandibular reconstruction using a vascularized iliac-crest flap. J Oral Maxillofac Surg. 2018;76:886–93.CrossRef
19.
Zurück zum Zitat Kraeima J, Glas HH, Witjes MJH, Schepman KP. Patient-specific pre-contouring of osteosynthesis plates for mandibular reconstruction: using a three-dimensional key printed solution. J Craniomaxillofac Surg. 2018;46:1037–40.CrossRef Kraeima J, Glas HH, Witjes MJH, Schepman KP. Patient-specific pre-contouring of osteosynthesis plates for mandibular reconstruction: using a three-dimensional key printed solution. J Craniomaxillofac Surg. 2018;46:1037–40.CrossRef
20.
Zurück zum Zitat Mascha F, Winter K, Pietzka S, Heufelder M, Schramm A, Wilde F. Accuracy of computer-assisted mandibular reconstructions using patient-specific implants in combination with CAD/CAM fabricated transfer keys. J Craniomaxillofac Surg. 2017;45:1884–97.CrossRef Mascha F, Winter K, Pietzka S, Heufelder M, Schramm A, Wilde F. Accuracy of computer-assisted mandibular reconstructions using patient-specific implants in combination with CAD/CAM fabricated transfer keys. J Craniomaxillofac Surg. 2017;45:1884–97.CrossRef
21.
Zurück zum Zitat Hanken H, Schablowsky C, Smeets R, et al. Virtual planning of complex head and neck reconstruction results in satisfactory match between real outcomes and virtual models. Clin Oral Investig. 2015;19:647–56.CrossRef Hanken H, Schablowsky C, Smeets R, et al. Virtual planning of complex head and neck reconstruction results in satisfactory match between real outcomes and virtual models. Clin Oral Investig. 2015;19:647–56.CrossRef
22.
Zurück zum Zitat Roser SM, Ramachandra S, Blair H, et al. The accuracy of virtual surgical planning in free fibula mandibular reconstruction: comparison of planned and final results. J Oral Maxillofac Surg. 2010;68:2824–32.CrossRef Roser SM, Ramachandra S, Blair H, et al. The accuracy of virtual surgical planning in free fibula mandibular reconstruction: comparison of planned and final results. J Oral Maxillofac Surg. 2010;68:2824–32.CrossRef
23.
Zurück zum Zitat Schepers RH, Kraeima J, Vissink A, et al. Accuracy of secondary maxillofacial reconstruction with prefabricated fibula grafts using 3D planning and guided reconstruction. J Craniomaxillofac Surg. 2016;44:392–9.CrossRef Schepers RH, Kraeima J, Vissink A, et al. Accuracy of secondary maxillofacial reconstruction with prefabricated fibula grafts using 3D planning and guided reconstruction. J Craniomaxillofac Surg. 2016;44:392–9.CrossRef
24.
Zurück zum Zitat van Baar GJ, Liberton NP, Forouzanfar T, Winters HA, Leusink FK. Accuracy of computer-assisted surgery in mandibular reconstruction: a postoperative evaluation guideline. Oral Oncology. 2019;88:1–8.CrossRef van Baar GJ, Liberton NP, Forouzanfar T, Winters HA, Leusink FK. Accuracy of computer-assisted surgery in mandibular reconstruction: a postoperative evaluation guideline. Oral Oncology. 2019;88:1–8.CrossRef
25.
Zurück zum Zitat Yu Y, Zhang WB, Liu XJ, Guo CB, Yu GY, Peng X. Three-dimensional accuracy of virtual planning and surgical navigation for mandibular reconstruction with free fibula flap. J Oral Maxillofac Surg. 2016;74:1503.e1501–10. Yu Y, Zhang WB, Liu XJ, Guo CB, Yu GY, Peng X. Three-dimensional accuracy of virtual planning and surgical navigation for mandibular reconstruction with free fibula flap. J Oral Maxillofac Surg. 2016;74:1503.e1501–10.
26.
Zurück zum Zitat Schepers RH, Raghoebar GM, Vissink A, et al. Accuracy of fibula reconstruction using patient-specific CAD/CAM reconstruction plates and dental implants: a new modality for functional reconstruction of mandibular defects. J Craniomaxillofac Surg. 2015;43:649–57.CrossRef Schepers RH, Raghoebar GM, Vissink A, et al. Accuracy of fibula reconstruction using patient-specific CAD/CAM reconstruction plates and dental implants: a new modality for functional reconstruction of mandibular defects. J Craniomaxillofac Surg. 2015;43:649–57.CrossRef
27.
Zurück zum Zitat Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med. 2016;15:155–63.CrossRef Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med. 2016;15:155–63.CrossRef
28.
Zurück zum Zitat Wilde F, Hanken H, Probst F, Schramm A, Heiland M, Cornelius CP. Multicenter study on the use of patient-specific CAD/CAM reconstruction plates for mandibular reconstruction. Int J Comput Assist Radiol Surg. 2015;10:2035–51.CrossRef Wilde F, Hanken H, Probst F, Schramm A, Heiland M, Cornelius CP. Multicenter study on the use of patient-specific CAD/CAM reconstruction plates for mandibular reconstruction. Int J Comput Assist Radiol Surg. 2015;10:2035–51.CrossRef
29.
Zurück zum Zitat Ciocca L, Mazzoni S, Fantini M, et al. A CAD/CAM-prototyped anatomical condylar prosthesis connected to a custom-made bone plate to support a fibula free flap. Med Biol Eng Comput. 2012;50:743–9.CrossRef Ciocca L, Mazzoni S, Fantini M, et al. A CAD/CAM-prototyped anatomical condylar prosthesis connected to a custom-made bone plate to support a fibula free flap. Med Biol Eng Comput. 2012;50:743–9.CrossRef
30.
Zurück zum Zitat Ciocca L, Mazzoni S, Fantini M, Persiani F, Marchetti C, Scotti R. CAD/CAM guided secondary mandibular reconstruction of a discontinuity defect after ablative cancer surgery. J Craniomaxillofac Surg. 2012;40:e511–5.CrossRef Ciocca L, Mazzoni S, Fantini M, Persiani F, Marchetti C, Scotti R. CAD/CAM guided secondary mandibular reconstruction of a discontinuity defect after ablative cancer surgery. J Craniomaxillofac Surg. 2012;40:e511–5.CrossRef
32.
Zurück zum Zitat De Maesschalck T, Courvoisier DS, Scolozzi P. Computer-assisted versus traditional freehand technique in fibular free-flap mandibular reconstruction: a morphological comparative study. Eur Arch Otorhinolaryngol. 2017;274:517–26.CrossRef De Maesschalck T, Courvoisier DS, Scolozzi P. Computer-assisted versus traditional freehand technique in fibular free-flap mandibular reconstruction: a morphological comparative study. Eur Arch Otorhinolaryngol. 2017;274:517–26.CrossRef
33.
Zurück zum Zitat Ciocca L, Marchetti C, Mazzoni S, et al. Accuracy of fibular sectioning and insertion into a rapid-prototyped bone plate, for mandibular reconstruction using CAD-CAM technology. J Craniomaxillofac Surg. 2015;43:28–33.CrossRef Ciocca L, Marchetti C, Mazzoni S, et al. Accuracy of fibular sectioning and insertion into a rapid-prototyped bone plate, for mandibular reconstruction using CAD-CAM technology. J Craniomaxillofac Surg. 2015;43:28–33.CrossRef
34.
Zurück zum Zitat Heufelder M, Wilde F, Pietzka S, et al. Clinical accuracy of waferless maxillary positioning using customized surgical guides and patient-specific osteosynthesis in bimaxillary orthognathic surgery. J Craniomaxillofac Surg. 2017;45:1578–85.CrossRef Heufelder M, Wilde F, Pietzka S, et al. Clinical accuracy of waferless maxillary positioning using customized surgical guides and patient-specific osteosynthesis in bimaxillary orthognathic surgery. J Craniomaxillofac Surg. 2017;45:1578–85.CrossRef
35.
Zurück zum Zitat Pollei TR, Barrs DM, Hinni ML, Bansberg SF, Walter LC. Operative time and cost of resident surgical experience: effect of instituting an otolaryngology residency program. Otolaryngol Head Neck Surg. 2013;148:912–8.CrossRef Pollei TR, Barrs DM, Hinni ML, Bansberg SF, Walter LC. Operative time and cost of resident surgical experience: effect of instituting an otolaryngology residency program. Otolaryngol Head Neck Surg. 2013;148:912–8.CrossRef
36.
Zurück zum Zitat Toto JM, Chang EI, Agag R, Devarajan K, Patel SA, Topham NS. Improved operative efficiency of free fibula flap mandible reconstruction with patient-specific, computer-guided preoperative planning. Head Neck. 2015;37:1660–4.CrossRef Toto JM, Chang EI, Agag R, Devarajan K, Patel SA, Topham NS. Improved operative efficiency of free fibula flap mandible reconstruction with patient-specific, computer-guided preoperative planning. Head Neck. 2015;37:1660–4.CrossRef
37.
Zurück zum Zitat Parekh DJ, Reis IM, Castle EP, et al. Robot-assisted radical cystectomy versus open radical cystectomy in patients with bladder cancer (RAZOR): an open-label, randomised, phase 3, non-inferiority trial. Lancet. 2018;391:2525–36.CrossRef Parekh DJ, Reis IM, Castle EP, et al. Robot-assisted radical cystectomy versus open radical cystectomy in patients with bladder cancer (RAZOR): an open-label, randomised, phase 3, non-inferiority trial. Lancet. 2018;391:2525–36.CrossRef
Metadaten
Titel
Three-Dimensionally Printed Patient-Specific Surgical Plates Increase Accuracy of Oncologic Head and Neck Reconstruction Versus Conventional Surgical Plates: A Comparative Study
verfasst von
Wei-fa Yang, MDS
Wing Shan Choi, MDS
May Chun-Mei Wong, PhD
Warit Powcharoen, DDS
Wang-yong Zhu, MD
James Kit-Hon Tsoi, PhD
Marco Chow, MPhil
Ka-Wai Kwok, PhD
Yu-xiong Su, MD
Publikationsdatum
22.06.2020
Verlag
Springer International Publishing
Erschienen in
Annals of Surgical Oncology / Ausgabe 1/2021
Print ISSN: 1068-9265
Elektronische ISSN: 1534-4681
DOI
https://doi.org/10.1245/s10434-020-08732-y

Weitere Artikel der Ausgabe 1/2021

Annals of Surgical Oncology 1/2021 Zur Ausgabe

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.