Open Access 07.05.2018 | Original Article

Titanium granules pre-treated with hydrogen peroxide inhibit growth of bacteria associated with post-operative infections in spine surgery

verfasst von: Acke Ohlin, Emma Mattsson, Matthias Mörgelin, Julia R. Davies, Gunnel Svensäter, Stéphane Corvec, Pentti Tengvall, Kristian Riesbeck

Erschienen in: European Spine Journal | Ausgabe 10/2018

Abstract

Purpose

Post-operative infections are relatively common after posterior spine surgery, and there are several observations reflecting different infection complications related to various metals implanted. Here, we selected an array of different bacterial species that are often found in infections associated with orthopaedic implants and tested for inhibition by hydrogen peroxide-treated titanium (Ti-peroxy).

Methods

To study the possibility of using Ti-peroxy as an antimicrobial prophylaxis, we developed a protocol for standardized susceptibility testing of bacteria.

Results

Importantly, we found that the resulting Ti-peroxy was highly antimicrobial against all aerobic species tested, among others, Staphylococcus aureus and Pseudomonas aeruginosa. Proteus mirabilis was slightly more resistant than, for example, Klebsiella pneumoniae and enterococci. In contrast, anaerobic bacteria Cutibacterium acnes and Parvimonas micra were equally susceptible compared to staphylococci.

Conclusions

Our findings suggest that the Ti-peroxy is a promising perioperative antimicrobial strategy that may be highly effective for prevention of post-operative infections. We therefore suggest application of hydrogen peroxide to implants prior to implantation.

Graphical abstract

These slides can be retrieved under Electronic supplementary material.
Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1007/​s00586-018-5619-8) contains supplementary material, which is available to authorized users.

Introduction

Early and late deep infection after posterior spine surgery with instrumentation is a relatively common complication, often necessitating implant removal. Factors influencing implant-related infections are multiple: the localization and extension of wound, length of surgery time, blood loss, nutritional status of patient, use of perioperative antibiotics, use of pulse lavage and quality of the operative room, e.g. ventilation conditions [1]. The rate of post-operative surgical site infections after spine surgery varies based on patient comorbidities from 0.7 to 4.3% among patients without risk factors to 2.0 to 10% in patients with obesity, diabetes and neuromuscular diseases [2]. The rate of late implant-associated deep infection after posterior spinal instrumentation is, however, significantly reduced when using a metal work made from titanium alloy compared with stainless steel [35]. In a previous study with a total of 1390 patients, the infection rate was significantly higher in patients that received stainless-steel implants [6]. Infection rate with titanium implants was 3.53% as compared to 11.56% with stainless-steel implants (p < 0.0001). In yet another investigation (n = 540), the infection rate was 1.33% compared to 4.56% for titanium and stainless-steel implants (p < 0.0001), respectively [7]. The reason for this is not fully understood.
More than 25 years ago, Tengvall et al. reported on the formation of relatively stable titanium compounds formed at Ti-implant surfaces as one explanation of the low infection rate after implantation of dental Ti-screws [8, 9]. Analytical data indicated that those compounds consisted of aqueous Ti-peroxide [10]. The formation of titanium peroxy compounds is due to a chemical reaction of Ti(III) and/or Ti(IV) ions with H2O2 being released by inflammatory cells. The tissue concentration may be up to 10 µM during inflammatory states [11]. Ti-peroxide is subsequently degraded to TiO2, H2O2, O2, and H2O. The time frame of degradation in vitro is from hours to days [12]. This process probably results in inhibition of early growth and survival of bacteria in the close vicinity of implants with this coating.
The main aim of the present study was to determine the viability of bacteria in the presence of metal granules made from commercial pure titanium (Ti) and “orthopaedic” stainless steel with or without previous treatment with hydroxyperoxide. We hypothesized that hydrogen peroxide-treated Ti granules, hereafter designated Ti-peroxy, would reduce bacterial viability. A series of different microbes, which are well known and often recognized in cultures of specimens obtained at reoperations due to deep late infections after posterior spinal instrumentation [1, 2], were included in the study and examined for growth in the presence of Ti-peroxy. Finally, scanning electron microscopy was carried out to visualize the gross appearance of Ti-peroxy-dependent inhibitory activity.

Material and methods

Bacterial strains and culture conditions

The different bacterial species examined in the present study are outlined in Table 1. A collection of Cutibacterium acnes isolated from patients with spine and knee joint infections after surgery was also included in our study [13]. This chosen set of both Gram-negative and Gram–positive bacteria was grown according to standard microbiological culture conditions. Briefly, Staphylococcus aureus, S. epidermidis, Corynebacterium renale, C. pilosum, Enterococcus faecalis, Bacillus badius, B. cereus, Escherichia coli, Pseudomonas aeruginosa, Enterobacter cloacae, Proteus mirabilis and, finally, Klebsiella pneumoniae were grown on chocolate agar plates (Oxoid, Waltham, MA). In contrast, the anaerobic species Cutibacterium acnes (previously Propionibacterium acnes) and Parvimonas micra (previously Peptostreptococcus micros) were grown on FAA plates (LabM, Heywood, UK). Aerobic bacteria were grown at 37 °C in 5% CO2, whereas the anaerobes were incubated in anaerobic jars at 37 °C.
Table 1
Bacterial strains used in the present study
Gram-positive
aerobic bacteria
Methicillin-susceptible Staphylococcus aureus (MSSA)
S. aureus ATCCa 25923
S. aureus ATCC BAA 1026
Methicillin-resistant S. aureus (MRSA)
S. aureus KR889b
S. aureus KR890
Staphylococcus epidermidis
S. epidermidis CCUGc 23118
S. epidermidis CCUG 3709
Corynebacterium spp.
C. renale CCUG 27542
C. pilosum CCUG 27193
Enterococcus faecalis
E. faecalis KR891
E. faecalis KR892
Bacillus spp.
B. badius KR893
B. cereus KR894
Gram-negative
aerobic bacteria
Escherichia coli
E. coli KR575
E. coli KR576
Pseudomonas aeruginosa
P. aeruginosa KR826
P. aeruginosa KR827
Enterobacter cloacae
E. cloacae KR835
E. cloacae KR836
Proteus mirabilis
P. mirabilis KR837
P. mirabilis KR838
Klebsiella pneumoniae
K. pneumoniae KR895
K. pneumoniae KR896
Gram-positive
anaerobic bacteria
Cutibacterium acnes d
C. acnes KR897 (skin)
C. acnes CCUG 1794 (skin)
Cutibacterium acnes PA 2004-8219, CC28e (spine) [13]
Cutibacterium acnes PA 2006-3267, CC36 (spine) [13]
Cutibacterium acnes PA 2004-10708, CC18 (spine) [13]
Cutibacterium acnes PA Ntes CC18 (knee joint) [13]
Parvimonas micra f
P. micra KR892 [14]
P. micra KR929
aATCC: American Type Culture Collection, LGC Standards, Teddington, UK
bKR; Kristian Riesbeck laboratory, Clinical Microbiology, Malmö, Sweden
cCulture Collection University of Gothenburg, Sweden
dPreviously designated Propionibacterium acnes
eCC; clonal complex as described [13]
fPreviously designated Peptostreptococcus micros

Preparation of Ti-peroxy

Sterilized granules (2 ml) consisting of commercially available titanium (size 0.5–1.0 mm) (Koch-Light, Haverhill, UK), chrome 50 Ln (size 0.5–1.0 mm) or chrome/nickel 50 (size 0.5–1.0 mm) (kindly provided by Pometon, Maeme [Venice], Italy) were included in the study. The chrome 50 Ln and chrome/nickel 50 granules are comparable with stainless steel in orthopaedic implants. Granules were incubated with 5 ml 30% H2O2 for 15 min at room temperature. Thereafter, granules were washed with 50 ml 0.9% NaCl and dispensed into deep-bottom 96-well plates (Whatman, Maidstone, UK). Granules (40 µl wet volume) were added to each well corresponding to approximately 86 µg Ti, 188 µg Cr 50 Ln or 270 µg Cr/Ni 50.

Analysis of the inhibitory effect of Ti-peroxy on bacteria

To prepare bacteria, overnight cultures of E. coli and P. aeruginosa were resuspended in Luria–Bertani (LB) broth (Becton–Dickinson, Franklin Lakes, NJ), whereas the remaining species were resuspended in Brain Heart Infusion (BHI) broth (Becton–Dickinson, Stockholm, Sweden). P. acnes and P. micra were diluted in BHI + 10% foetal bovine serum (FBS) (Gibco Life Technologies, Warrington, UK) and FA broth (LabM, Heywood, UK), respectively. After resuspension, optical density (OD)600 was set to 0.1 and 0.5 for aerobes and the anaerobes including Corynebacterium spp., respectively. Thereafter, a tenfold dilution series was made, and 800 µl of the bacterial suspensions was added to plates containing various granules. Microtiter plates were sealed with adhesive gas-permeable seals (Thermo Scientific, Waltham, MA) and incubated at 37 °C, 5% CO2 shaking at 200 rpm. Anaerobes were incubated in anaerobic jars, 37 °C, shaking at 150 rpm. After 24 h incubation or 7 d for anaerobes, 100 µl/well was transferred to microtiter plates to measure absorbance at OD600 nm. In separate experiments, the kinetics of antibacterial activity of hydrogen peroxide-treated Ti granules was determined. Granules were preincubated with H2O2 and thereafter stored in a closed container before addition to bacterial cultures as described above.

Scanning electron microscopy

For scanning electron microscopy, specimens were fixed overnight at room temperature with 2.5% glutaraldehyde in cacodylate buffer, washed with cacodylate buffer and dehydrated with an ascending ethanol series from 50% (v/v) to absolute ethanol. The specimens were dried in a critical point dryer with carbon dioxide and absolute ethanol as an intermediate solvent. Samples were mounted on aluminium holders, sputtered with 20 nm palladium/gold and examined in a Philips/FEI XL 30 FESEM scanning electron microscope using an Everhart–Thornley secondary electron detector. Image processing was carried out with the Scandium software (Emsis, Münster, Germany) for simple image acquiring and auto-storage into the Scandium database. All electron microscopic work was performed at the Core Facility for Integrated Microscopy, Panum Institute (University of Copenhagen, Denmark).

Results

Titanium peroxy efficiently inhibits bacterial growth

We chose a large panel of different bacterial species that are commonly found in patients who have undergone spine surgery complicated by deep infection (Table 1) [1]. Two isolates of each species except for Cutibacterium acnes (n = 6) were included in the study. Titanium granules were treated with hydrogen peroxide for 15 min followed by washing with saline resulting in a Ti-peroxide. To analyse the inhibitory effect on growth, bacteria were incubated with the Ti-peroxide, and, in addition, hydrogen peroxide-treated Cr 50 Ln or Cr/Ni was included as controls. The effect of untreated Ti (without previous exposure of hydrogen peroxide) was also analysed. All granules were supplemented to microtiter plates followed by addition of bacteria in solution.
Hydrogen peroxide-treated Ti granules, i.e. the Ti-peroxy inhibited a large number of bacteria in our experimental set-up (Table 2). Bacterial growth was significantly inhibited in the interval 6 × 104 to 108 CFU/ml upon exposure to the Ti-peroxy for 24 h (aerobes) or 7 days (anaerobes). In contrast, Cr 50 Ln and Cr/Ni that were preincubated with H2O2 did not interfere with bacterial growth, and with these apparently inert granules, no inhibition was observed as compared to culture media without any granules added. In parallel, untreated Ti granules did not interfere with bacterial growth (results not shown). Proteus mirabilis was the most resistant species, and here more than 1000-fold fewer bacteria (6 × 104 CFU/ml) were inhibited compared to, for example, enterococci (108 CFU/ml). Regarding other Gram-negative species, we did not find any large differences in Ti-peroxy-dependent growth inhibition. In parallel, anaerobes including C. acnes and Parvimonas micra were both efficiently inhibited by the Ti-peroxy, but not by Cr 50 Ln, Cr/Ni or untreated Ti. A series of different clinical C. acnes isolated from orthopaedic patients belonging to different clonal complexes (CC) [13] was also tested, but no difference was observed in Ti-peroxy-dependent growth inhibition as compared to C. acnes isolates derived from the skin.
Table 2
Titanium peroxy has an effect on all bacterial species tested
Bacterial species
Number of bacteria (CFU/ml) inhibited by hydrogen peroxide-treated granules consisting of
Ti (Ti-peroxy)
Cr 50 Ln
Cr/Ni 50
MSSAa
1 × 107
0
0
MRSAa
1 × 107
0
0
Staphylococcus epidermidis a
1 × 107
0
0
Corynebacterium spp.b
6 × 106
0
0
Enterococcus faecalis a
1 × 108
0
0
Bacillus spp.a
1 × 107
0
0
Escherichia coli a
1 × 107
0
0
Pseudomonas aeruginosa a
1 × 106
0
0
Enterobacter cloacae a
1 × 106
0
0
Proteus mirabilis a
6 × 104
0
0
Klebsiella pneumoniae a
6 × 107
0
0
Cutibacterium acnes b
1 × 106
0
0
Parvimonas micra b
1 × 107
0
0
Diluted bacteria were transferred to microtiter plates with granules that had been pre-treated with H2O2 followed by washing with NaCl. After 24 h or 7 d incubation for anaerobes, the number (colony-forming units: CFU) of surviving bacteria was estimated by determining absorbance at OD600. Values indicate the highest bacterial growth observed based upon 2–3 observations for each bacterial species
aAll aerobic bacterial species were diluted in microtiter plates starting at 108 bacteria/ml
bAll anaerobic bacterial species including Corynebacterium spp. were diluted in microtiter plates starting at 5 × 108 bacteria/ml
To determine the stability of the Ti-peroxy, titanium was pre-treated with hydrogen peroxide according to the standard protocol, and thereafter, the resulting Ti-peroxy was added to bacterial cultures after different time points. The antibacterial activity of the Ti-peroxy persisted for at least 24 h, but at later time points, the activity decreased, and when the Ti-peroxy had been stored for more than 72 h at room temperature, no inhibitory effect was observed (data not shown). However, when the Ti-peroxy was stored at − 18 °C, the antibacterial activity was preserved for at least 90 days. Taken together, although differences in the inhibitory effect by the Ti-peroxy exist between various bacterial species tested in our hands, we conclude that the Ti-peroxy has a powerful effect upon most bacterial species found in post-operative infections after spine surgery.

Bacteria are lysed by Ti-peroxy as revealed by scanning electron microscopy

To further shed light upon the inhibitory activity of Ti-peroxy on bacterial growth, P. aeruginosa and methicillin-susceptible S. aureus (MSSA) representing Gram-negative and Gram-positive species, respectively, were subjected to analysis of the ultrastructure by scanning electron microscopy. Interestingly, bacteria, as exemplified by P. aeruginosa, were relatively smoothly attached to the surface of Ti (Fig. 1b, c). After 30 min incubation, bacteria started to lyse (Fig. 1d, g), and after further 2.5 h, the majority of both P. aeruginosa and S. aureus were destroyed and severely damaged (Fig. 1e, h, respectively). The Ti-peroxide was thus very efficient in bacterial killing.

Discussion

In this study, we examined Ti granules preincubated with H2O2 forming a Ti-peroxy that was added to both catalase-positive and -negative bacteria of common species that cause deep infections around spinal implants. Catalase is a well-known virulence factor and defence mechanism aimed at neutralizing the oxidative burst consisting of H202. All bacteria were hence inhibited albeit with various efficiencies. Sterile but non-H2O2-preincubated Ti granules and also “orthopaedic” stainless-steel granules with or without H2O2 preincubation were included as controls. Importantly, an inhibition and decreased viability of bacteria were only observed in experiments with H2O2-preincubated Ti granules (Table 2).
The antibacterial effect of UV treatment of TiO2 is well known [15]. In parallel, RF oxygen plasma treatment of Ti results in an antibacterial effect [16]. The mechanism underlying these bactericidal effects is probably generation and degradation of Ti-peroxy compounds into H2O2 and oxygen radicals resulting in destruction of bacterial membranes via oxidative processes [15]. The kinetics of these chemical reactions is difficult to determine, and probably the effect of these radical–chemical reactions is present for only a day or two. This presumption is supported by previous degradation experiments with aged and non-aged Ti-peroxy compounds [12]. The Ti-peroxide FT-Raman signal remained detectable in aqueous 3-month-old and dried compounds and in freeze-dried fresh compounds for 3 years. The inhibitory effects on different bacterial species upon incubation with fresh H2O2-treated Ti granules in the present study are therefore reasonable.
The relatively short period of time with antimicrobial activity may nevertheless be enough to inhibit any hibernating single-cell bacteria before a protective protein coating and/or biofilm is formed. Furthermore, it can be postulated that in the case of deposition of bacteria in the vicinity of implants due to bacteraemia, the biological reaction with recruited inflammatory cells makes a similar formation of Ti-peroxide compound(s) on the surface of Ti implants, possibly enhancing or prolonging the antibacterial effect that would be highly beneficial for the patient.
One interesting observation is that the Ti-peroxy was also effective on catalase-positive species, indicating that the membrane destruction mechanisms are complex and cannot easily be eradicated by catalase. However, since E. faecalis is catalase negative, it was also 10–100-fold more susceptible as compared to the other bacterial species examined (Table 2).
When we have encountered skin perforation due to progressive deformity or trauma in a clinical setting, we have been able to prevent infections by partial implant removal and local H2O2 treatment in conjunction with parental treatment with antibiotics over a short period of time. This has been successful with Ti-alloy in relation to stainless steel or CrCo implants (Ohlin et al. unpublished data 2007–2014). Based on these observations, we now routinely expose our titanium implants to H2O2 prior to implantation.
Studies published aiming at predicting risks of unfavourable results in spine surgery, all lack information of which alloy the metal work used was made from and therefore they are not informative in this regard [17, 18]. On the contrary, there is one clinical series published comprising a limited number of patients in which revision due to late infected stainless steel to a titanium-based implants in one surgery was used, with full success in 10/10 cases, supporting our hypothesis in a clinical setting [4].
In conclusion, despite relatively large differences existing between the various species tested in this study, we postulate that the Ti-peroxy has a very efficient effect upon most bacterial species irrespective of origin. It should also be emphasized that most likely only single bacteria occur in wounds further increasing the likelihood of preventive antibacterial measures by the Ti-peroxy.

Acknowledgements

This work was supported by grants from the Anna and Edwin Berger Foundation, Medical Research Council (Grant Number K2015-57X-03163-43-4, www.​vr.​se) and Skåne County Council’s Research and Development Foundation. We thank Mrs. Allessandro Breda (Pomerton, Maerne [Venice], Italy) for providing us with control granules.

Compliance with ethical standards

Conflict of interest

The authors declare that they have no competing interests.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Orthopädie & Unfallchirurgie

Kombi-Abonnement

Mit e.Med Orthopädie & Unfallchirurgie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

Anhänge

Electronic supplementary material

Below is the link to the Electronic supplementary material.
Literatur
1.
Zurück zum Zitat Abdul-Jabbar A et al (2013) Surgical site infections in spine surgery. Spine 38:E1425–E1431CrossRef Abdul-Jabbar A et al (2013) Surgical site infections in spine surgery. Spine 38:E1425–E1431CrossRef
2.
Zurück zum Zitat Boody B et al (2015) Surgical site infections in spinal surgery. J Spinal Disord Tech 28:352–362CrossRef Boody B et al (2015) Surgical site infections in spinal surgery. J Spinal Disord Tech 28:352–362CrossRef
3.
Zurück zum Zitat Clark CE, Shufflebarger HL (1999) Late-developing infection in instrumented idiopathic scoliosis. Spine 24:1909–1912CrossRef Clark CE, Shufflebarger HL (1999) Late-developing infection in instrumented idiopathic scoliosis. Spine 24:1909–1912CrossRef
4.
Zurück zum Zitat Muschik M, Luck W, Schlenzka D (2004) Implant removal for late-developing infection after instrumented posterior spinal fusion for scoliosis: reinstrumentation reduces loss of correction. A retrospective analysis of 45 cases. Eur Spine J 13:645–651CrossRef Muschik M, Luck W, Schlenzka D (2004) Implant removal for late-developing infection after instrumented posterior spinal fusion for scoliosis: reinstrumentation reduces loss of correction. A retrospective analysis of 45 cases. Eur Spine J 13:645–651CrossRef
5.
Zurück zum Zitat Wåxnäs T, Hasserius R, Ohlin A (2012) Implant associated late developing infection after posterior spinal surgery, presented at SRS 47th annual meeting, Chicago, IL Wåxnäs T, Hasserius R, Ohlin A (2012) Implant associated late developing infection after posterior spinal surgery, presented at SRS 47th annual meeting, Chicago, IL
6.
Zurück zum Zitat LaGreca J et al (2014) Bacteriology and risk factors for development of late (greater than one year) deep infection following spinal fusion with instrumentation. Spine Deform 3:186–190CrossRef LaGreca J et al (2014) Bacteriology and risk factors for development of late (greater than one year) deep infection following spinal fusion with instrumentation. Spine Deform 3:186–190CrossRef
7.
Zurück zum Zitat Di Silvestre M, Bakaloudis G, Lolli F, Giacomini S (2011) Late-developing infection following posterior fusion for adolescent idiopathic scoliosis. Eur Spine J Suppl 1:S121–S127CrossRef Di Silvestre M, Bakaloudis G, Lolli F, Giacomini S (2011) Late-developing infection following posterior fusion for adolescent idiopathic scoliosis. Eur Spine J Suppl 1:S121–S127CrossRef
8.
Zurück zum Zitat Tengvall P, Lundström I (1992) Interaction between hydrogen peroxide and titanium: a possible role in the biocompatibility of titanium. Clin Mater 9:115–134CrossRef Tengvall P, Lundström I (1992) Interaction between hydrogen peroxide and titanium: a possible role in the biocompatibility of titanium. Clin Mater 9:115–134CrossRef
9.
Zurück zum Zitat Tengvall P, Hörnsten EG, Elwing H, Lundström I (1990) Bactericidal properties of a titanium-peroxy gel obtained from metallic titanium and hydrogen peroxide. J Biomed Mater Res 24:319–330CrossRef Tengvall P, Hörnsten EG, Elwing H, Lundström I (1990) Bactericidal properties of a titanium-peroxy gel obtained from metallic titanium and hydrogen peroxide. J Biomed Mater Res 24:319–330CrossRef
10.
Zurück zum Zitat Tengvall P et al (1990) Degradation of Ti-peroxy gels made from metallic titanium and hydrogen peroxide. J Colloid Interface Sci 139:575–580CrossRef Tengvall P et al (1990) Degradation of Ti-peroxy gels made from metallic titanium and hydrogen peroxide. J Colloid Interface Sci 139:575–580CrossRef
11.
Zurück zum Zitat Mueller S, Arnhold J (1995) Fast and sensitive chemiluminiscence determination of H2O2 concentration in stimulated human neutrophils. J Biolumin 10:229–237CrossRef Mueller S, Arnhold J (1995) Fast and sensitive chemiluminiscence determination of H2O2 concentration in stimulated human neutrophils. J Biolumin 10:229–237CrossRef
12.
Zurück zum Zitat Tengvall P, Vikinge TP, Lundström I, Lieberg B (1993) FT-Raman spectroscopic studies of the degradation of titanium peroxy gels made from metallic titanium and hydrogen peroxide. J Colloid Interface Sci 160:10–15CrossRef Tengvall P, Vikinge TP, Lundström I, Lieberg B (1993) FT-Raman spectroscopic studies of the degradation of titanium peroxy gels made from metallic titanium and hydrogen peroxide. J Colloid Interface Sci 160:10–15CrossRef
13.
Zurück zum Zitat Aubin GG et al (2017) Interaction of cutibacterium (formerly propionibacterium) acnes with bone cells: a step toward understanding bone and joint infection development. Sci Rep 7:42918CrossRef Aubin GG et al (2017) Interaction of cutibacterium (formerly propionibacterium) acnes with bone cells: a step toward understanding bone and joint infection development. Sci Rep 7:42918CrossRef
14.
Zurück zum Zitat Riesbeck K, Sanzén L (1999) Destructive knee joint infection caused by Peptostreptococcus micros: importance of early microbiological diagnosis. J Clin Microbiol 37:2737–2739PubMedPubMedCentral Riesbeck K, Sanzén L (1999) Destructive knee joint infection caused by Peptostreptococcus micros: importance of early microbiological diagnosis. J Clin Microbiol 37:2737–2739PubMedPubMedCentral
15.
Zurück zum Zitat Foster HA, Ditta IB, Varghese S, Steele A (2011) Photocatalytic disinfection using titanium dioxide: spectrum and mechanism of antimicrobial activity. Appl Microbiol Biotechnol 90:1847–1868CrossRef Foster HA, Ditta IB, Varghese S, Steele A (2011) Photocatalytic disinfection using titanium dioxide: spectrum and mechanism of antimicrobial activity. Appl Microbiol Biotechnol 90:1847–1868CrossRef
16.
Zurück zum Zitat Monetta T, Belluci F (2014) Strong and durable antibacterial effect of titanium treated in RF oxygen plasma: preliminary results. Plasma Chem Process 34:1247–1256CrossRef Monetta T, Belluci F (2014) Strong and durable antibacterial effect of titanium treated in RF oxygen plasma: preliminary results. Plasma Chem Process 34:1247–1256CrossRef
Metadaten
Titel
Titanium granules pre-treated with hydrogen peroxide inhibit growth of bacteria associated with post-operative infections in spine surgery
verfasst von
Acke Ohlin
Emma Mattsson
Matthias Mörgelin
Julia R. Davies
Gunnel Svensäter
Stéphane Corvec
Pentti Tengvall
Kristian Riesbeck
Publikationsdatum
07.05.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
European Spine Journal / Ausgabe 10/2018
Print ISSN: 0940-6719
Elektronische ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-018-5619-8