Tomosynthesis implementation with adaptive online calibration on clinical C-arm systems

verfasst von: Khanlian Chung, Lothar R. Schad, Frank G. Zöllner

Erschienen in: International Journal of Computer Assisted Radiology and Surgery | Ausgabe 10/2018

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Cone beam computed tomography (CBCT) systems offer physicians crucial 3D and 2D imaging capabilities during interventions. However, certain medical applications only require very specific information from the CBCTs (e.g., determination of the position of high-contrast objects). In diagnostics, tomosynthesis techniques can be used in these cases to minimize dose exposure. Therefore, integrating such techniques on CBCT systems could also be beneficial for interventions. In this paper, we investigate the performance of our implementation of circular tomosynthesis on a CBCT device.

Methods

The tomosynthesis scan trajectory is realized with step-and-shoot on a clinical C-arm device. The online calibration algorithm uses conventionally acquired 3D CBCT of the scanned object as prior knowledge to correct the imaging geometries. The online calibration algorithm was compared to an offline calibration to test its performance. A ball bearing phantom was used to evaluate the reconstructions with respect to geometric distortions. The evaluation was done for three different scenarios to test the robustness of our tomosynthesis implementation against object deviations (e.g., pen) and different object positioning.

Results

The circular tomosynthesis was tested on a ball bearing and an anthropomorphic phantom. The results show that the calibration is robust against isocenter shifts and object deviations in the CBCT. All reconstructions used 100 projections and displayed limited angle artifacts. The accuracy of the positions and shapes of high-contrast objects were, however, determined precisely. (The maximal center position deviation is 0.31 mm.)

Conclusion

For medical procedures that primarily determine the precise position of high-contrast objects, circular tomosynthesis could offer an approach to reduce dose exposure.
Literatur
7.
Zurück zum Zitat Langan DA, Claus BEH, Al Assad O, Trousset Y, Riddell C, Avignon G, Solomon SB, Lai H, Wang X (2015) Interventional C-arm tomosynthesis for vascular imaging: initial results. In: Hoeschen C, Kontos D, Flohr TG (eds). SPIE, p 94125N Langan DA, Claus BEH, Al Assad O, Trousset Y, Riddell C, Avignon G, Solomon SB, Lai H, Wang X (2015) Interventional C-arm tomosynthesis for vascular imaging: initial results. In: Hoeschen C, Kontos D, Flohr TG (eds). SPIE, p 94125N
8.
Zurück zum Zitat Claus BE, Langan DA, Al Assad O, Wang X (2015) Circular tomosynthesis for neuro perfusion imaging on an interventional C-arm. In: Hoeschen C, Kontos D, Flohr TG (eds). SPIE, p 94122A Claus BE, Langan DA, Al Assad O, Wang X (2015) Circular tomosynthesis for neuro perfusion imaging on an interventional C-arm. In: Hoeschen C, Kontos D, Flohr TG (eds). SPIE, p 94122A
9.
Zurück zum Zitat Nett BE, Zambelli J, Riddell C, Belanger B, Chen G-H (2007) Circular tomosynthesis implemented with a clinical interventional flat-panel based C-Arm: initial performance study. In: Proceeding SPIE 6510, Medical imaging 2007: physics of medical imaging, 65101N. https://doi.org/10.1117/12.713789 Nett BE, Zambelli J, Riddell C, Belanger B, Chen G-H (2007) Circular tomosynthesis implemented with a clinical interventional flat-panel based C-Arm: initial performance study. In: Proceeding SPIE 6510, Medical imaging 2007: physics of medical imaging, 65101N. https://​doi.​org/​10.​1117/​12.​713789
10.
Zurück zum Zitat Nett B, Tang J, Leng S, Chen G-H (2008) Tomosynthesis via total variation minimization reconstruction and prior image constrained compressed sensing (PICCS) on a C-arm System. In: Proceedings of SPIE—the international society for optical engineering, vol 6913: nihpa92672. https://doi.org/10.1117/12.771294 Nett B, Tang J, Leng S, Chen G-H (2008) Tomosynthesis via total variation minimization reconstruction and prior image constrained compressed sensing (PICCS) on a C-arm System. In: Proceedings of SPIE—the international society for optical engineering, vol 6913: nihpa92672. https://​doi.​org/​10.​1117/​12.​771294
12.
Zurück zum Zitat Chung K, Karstensen L, Siegfarth M, Schad LR, Zöllner FG (2017) Investigation of scatter radiation of non-circular X-ray scan trajectories. Fully3D 2017 Chung K, Karstensen L, Siegfarth M, Schad LR, Zöllner FG (2017) Investigation of scatter radiation of non-circular X-ray scan trajectories. Fully3D 2017
13.
Zurück zum Zitat Solomon EG, Wilfley BP, van Lysel MS, Joseph AW, Heanue JA (1999) Scanning-beam digital x-ray (SBDX) system for cardiac angiography. In: Proceeding SPIE 3659, Medical imaging 1999: physics of medical imaging. https://doi.org/10.1117/12.349499 Solomon EG, Wilfley BP, van Lysel MS, Joseph AW, Heanue JA (1999) Scanning-beam digital x-ray (SBDX) system for cardiac angiography. In: Proceeding SPIE 3659, Medical imaging 1999: physics of medical imaging. https://​doi.​org/​10.​1117/​12.​349499
15.
Zurück zum Zitat Slagowski JM, Dunkerley DAP, Hatt CR, Speidel MA (2016) A geometric calibration method for inverse geometry computed tomography using P-matrices. In: Proceedings of SPIE—the international society for optical engineering, p 9783. https://doi.org/10.1117/12.2216565 Slagowski JM, Dunkerley DAP, Hatt CR, Speidel MA (2016) A geometric calibration method for inverse geometry computed tomography using P-matrices. In: Proceedings of SPIE—the international society for optical engineering, p 9783. https://​doi.​org/​10.​1117/​12.​2216565
26.
Zurück zum Zitat Schnurr A-K, Chung K, Schad LR, Zöllner FG (2018) Abstract. Deep residual learning for limited angle artefact correction. In: Maier A, Deserno TM, Handels H et al (eds) Bildverarbeitung für die Medizin 2018. Springer, Berlin, p 280CrossRef Schnurr A-K, Chung K, Schad LR, Zöllner FG (2018) Abstract. Deep residual learning for limited angle artefact correction. In: Maier A, Deserno TM, Handels H et al (eds) Bildverarbeitung für die Medizin 2018. Springer, Berlin, p 280CrossRef
Metadaten
Titel
Tomosynthesis implementation with adaptive online calibration on clinical C-arm systems
verfasst von
Khanlian Chung
Lothar R. Schad
Frank G. Zöllner
Publikationsdatum
08.05.2018
Verlag
Springer International Publishing
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 10/2018
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-018-1782-y

Weitere Artikel der Ausgabe 10/2018

International Journal of Computer Assisted Radiology and Surgery 10/2018 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.