Open Access 01.12.2018 | Research

Torsion bottle, a very simple, reliable, and cheap tool for a basic scoliosis screening

verfasst von: Michele Romano, Matteo Mastrantonio

Erschienen in: Scoliosis and Spinal Disorders | Ausgabe 1/2018

Abstract

Background

One of the reasons that make scoliosis a disease that scares so much the parents is its specific characteristic of being difficult to detect on its onset.
The aim of this paper is to check the possible usefulness of a simple tool (the torsion bottle) that has been developed with the aim to offer an instrument for home use by parents but also for screening purposes in the low-income countries.

Methods

Study design: retrospective analysis to evaluate intra-operator reliability of the tools and inter-operator repeatability using the torsion bottle.
For the first and the second part of the study, 35 subjects were measured.
The goal of the first experiment was to evaluate the reliability of the torsion bottle to identify all individuals who experienced a thoracic or lumbar prominence equal or greater than 7°.
The secondary aim was to verify the reliability of blinded inter-operator assessments, performed with the torsion bottle by two physiotherapists on the same patients.

Results

The reliability of the assessments of the torsion bottle has been performed with the Kappa statistic to evaluate the measurement agreement.
The results have shown that the intra-operator reliability of the tool is very high between the measurements collected with the scoliometer® and those collected with the torsion bottle (kappa = 0.9278; standard error = 0.7094).
The data of the second part of the study show that the inter-operator reliability is good (kappa = 0.7988; standard error 0.1368).

Conclusion

The collected data showed that the torsion bottle revealed itself as an efficient tool to execute a basic screening to identify the presence of a prominence in a significant group of adolescents.

Background

One of the reasons that make scoliosis a disease that scares so much the parents is its specific characteristic of being difficult to detect on its onset.
Too often, it is possible to recognize the appearance of the spinal curves when the disease is already severe.
The reason is due to the typical, slow, and insidious debut of the symptoms.
Being able to identify early the first signs of misalignment of the spine is one of the most effective treatment keys.
Obviously, a frequent monitoring of the clinical signs of scoliosis is the simplest solution [13].
One of the most important clinical signs of this disease is the prominence [4]. The presence of the prominence is internationally recognized as an indisputable element of the pathology [57].
For the prominence assessment, a scoliometer® is usually used [8, 9].
The scoliometer® is a professional tool used in the medical field by physicians and physical therapists.
Obviously, in a domestic field, the scoliometer® is not available.
For this reason, a very simple tool called torsion bottle, cheap and easy to create, has been developed.
The aim of this paper is to check the possible usefulness of this simple tool (the torsion bottle) that has been developed with the aim to offer an instrument for home use by parents but also for screening purposes in the low-income countries.

Methods

Study design: retrospective analysis to evaluate intra-operator reliability of the tool and inter-operator repeatability.
The study was divided into two parts. The main aim was to compare the measurements collected with the torsion bottle and the same measurements collected with the scoliometer® (gold standard) in a population of patients accessing our clinic for the rehabilitation treatment, measured according to our standard evaluation protocol.
The secondary aim was to verify the reliability of blinded inter-operator assessments, performed with the torsion bottle by two physiotherapists on the same patients. These double-blinded measurements were taken periodically for a short time interval in our clinic to verify measurement accuracy in the everyday clinical activities.

The tool

The torsion bottle is composed of a plastic water bottle of 500 ml.
The perfect accuracy of the amount of water to be used is not an essential element. On the outer surface of the bottle, it will be marked with a series of four landmarks that will be essential for its use. These landmarks will be marked only after the bottle is partially emptied and carefully closed.
These landmarks will be adjusted only to the particular amount of water contained in that bottle. In any case, it is recommended to fill approximately half of the bottle.
The torsion bottle is used in a horizontal position as a scoliometer®.
The rationale of the preparation of the torsion bottle is the variation of the position of the little quantity of water inside the bottle, when the bottle is tilted.
The essential reference signs are:
  • Landmarks 1 and 2 which should coincide with the water surface.
  • Landmark 3 is the limit of the 7° of the bottle’s tilt, and it represents the cutoff to identify the subjects with a prominence less, equal, or greater than this value (Fig. 1).
  • Landmark 2 is marked at a distance of 80 mm with respect to landmark 1.
  • Landmark 3 is marked 10 mm from the landmark 2 in the orthogonal direction (Fig. 2).

How to use the torsion bottle

How to use the torsion bottle is as follows:
  • Forward bending of the patient’s trunk to perform an Adam test [10].
  • The torsion bottle positioned crosswise on the patient’s back with the landmark 1 positioned over the line of bony prominences and the water surface coinciding with landmarks 1 and 2 (Fig. 3).
  • Tilt the bottle until it touches the patient’s back (Fig. 4).
  • Rotate the bottle until the surface of the water coincides with the position of landmark 1 (Fig. 5).
  • Observe the position of the surface of the water with respect to landmarks 3 or 4 (Fig. 6).

Instructions for the preparation and the use of the torsion bottle

The water surface slope, with respect to the reference point, corresponds precisely to 7125°.
The following table shows the variability linked to the potential errors in the mark of the reference points (Table 1).
Table 1
Distances between reference points and relative degrees
Dots 1-2 (cm)
Dots 2-3 (cm)
Degree
8
1
7,125
8
0,7
5,001
8
0,8
5,711
8
0,9
6,419
7,9
0,7
5,064
7,9
0,8
5,782
7,9
0,9
6,499
7,8
0,7
5,128
7,8
0,8
5,856
7,8
0,9
6,582
7,7
0,7
5,194
7,7
0,8
5,932
7,7
0,9
6,667
8,1
0,7
4,939
8,1
0,8
5,641
8,1
0,9
6,340
8,2
0,7
4,879
8,2
0,8
5,572
8,2
0,9
6,263
8,3
0,7
4,821
8,3
0,8
5,505
8,3
0,9
6,189
8
1,2
8,531
For the first part of the study, 35 subjects were measured. This group included consecutive adolescents of both genders who attend a specialized clinic for the conservative treatment of scoliosis to perform their medical examinations.
The goal of the experiment was to evaluate the reliability of the torsion bottle to identify all individuals who experienced a prominence equal or greater than 7°.
The presence of a prominence higher than 7° is considered a clinical sign very likely linked to a scoliosis that requires a medical treatment [1].
The assessment was performed by the same evaluator.
The evaluator asked the patient to bend the trunk forwards to perform an Adam test.
As a first action, he performed the measurement with the torsion bottle, and in a second time, the measurement with the scoliometer® precisely performed on the same point of the trunk.
For the second part of the study, the assessors measured the prominences of 35 consecutive patients who attended the same clinic for a session of conservative treatment based on Physioterapic Scoliosis Specific Exercises (PSSE).
Measurements were collected blinded by two physiotherapists and processed by a third operator.
The first assessor performed the first evaluation.
The evaluator asked the patient to bend the trunk forwards to perform an Adam test.
The assessor used the torsion bottle to measure the prominence.
The evaluator made a small sign at the back where the measurement had been taken, to enable the second prominence assessment on the same spot (Fig. 7).
The subject was accompanied to another room to carry out a second evaluation performed by the second evaluator.
The data collected in two different files have been handed over to a third operator who performed the comparison and the statistical tests.

Results

The following table shows the detail of the measurements taken for the first part of the study (Table 2):
  • Column 1: gender of the subject
  • Column 2: the result of the measurement performed with the torsion bottle and the positive or negative hypothesis of the presence of a prominence equal to or greater than 7°
  • Column 3: the measurement of the angle trunk rotation performed with the scoliometer®
  • Column 4: the agreement of the measurement performed with the torsion bottle and the scoliometer®
Table 2
Concordance between Torsion Bottle and Scoliometer assessments
Gender
Torsion Bottle Assessment
Scoliometer
Assessment
Concordance
Female
Positive
13
YES
Female
Positive
8
YES
Female
Positive
11
YES
Female
Positive
7
YES
Female
Negative
5
YES
Female
Positive
10
YES
Female
Positive
10
YES
Female
Positive
7
YES
Male
Negative
3
YES
Female
Positive
8
YES
Female
Positive
8
YES
Female
Positive
7
YES
Female
Positive
11
YES
Female
Positive
8
YES
Female
Negative
3
YES
Female
Negative
8
NO
Female
Positive
7
YES
Female
Positive
9
YES
Male
Positive
12
YES
Male
Positive
9
YES
Female
Positive
8
YES
Female
Positive
8
YES
Female
Positive
8
YES
Female
Negative
4
YES
Female
Positive
7
YES
Female
Positive
10
YES
Female
Negative
5
YES
Female
Positive
10
YES
Female
Negative
5
YES
Male
Negative
6
YES
Female
Negative
6
YES
Female
Positive
8
YES
Female
Positive
7
YES
Female
Negative
5
YES
Female
Positive
11
YES
Table 3 shows the details of the measurements of the second part of the study:
  • Column 1: gender of the subject
  • Column 2: results of the first assessor
  • Column 3: results of the second assessor
Table 3
Agreement between Torsion Bottle assessment performed by blinded assessors
Gender
Assessor 1
Assessorr 2
Female
Positive
Positive
Female
Positive
Positive
Female
Positive
Positive
Female
Positive
Positive
Female
Positive
Positive
Female
Negative
Negative
Female
Positive
Positive
Male
Positive
Positive
Female
Positive
Positive
Female
Positive
Positive
Female
Positive
Positive
Male
Negative
Negative
Male
Positive
Positive
Male
Positive
Negative
Female
Positive
Positive
Female
Positive
Positive
Female
Positive
Positive
Female
Positive
Positive
Female
Positive
Positive
Female
Positive
Positive
Female
Negative
Negative
Female
Positive
Positive
Female
Positive
Positive
Female
Positive
Positive
Female
Positive
Positive
Female
Positive
Positive
Female
Positive
Positive
Male
Negative
Negative
Female
Positive
Positive
Female
Positive
Positive
Female
Positive
Positive
Female
Negative
Positive
Female
Positive
Positive
Female
Negative
Negative
Female
Positive
Positive
The reliability of the assessments of the torsion bottle has been performed with the Kappa statistic to evaluate the measurement agreement.
The results have shown that the intra-operator reliability of the tool is very high between the measurements collected with the scoliometer® and those collected with the torsion bottle (kappa = 0.9278; standard error = 0.7094).
The data of the second part of the study show that the inter-operator reliability is good (kappa = 0.7988; standard error 0.1368).

Discussion

The collected data showed that the torsion bottle revealed itself as an efficient tool to execute a basic screening to identify the presence of a prominence in a significant group of adolescents.
The basic scoliosis screening is a very simple process but is not performed regularly because, in many countries, the health policy of this disease is based on the concept of “wait and see.” Unfortunately, this does not help in prevention but too often leads to surgical intervention, when the severity of the disease is serious.
This torsion bottle has not been conceived for a professional use because the torsion bottle does not quantify the degree of the prominence, unlike the scoliometer®.
It is considered more useful for a domestic use after having trained the parents. This makes it possible to achieve greater awareness and more frequent monitoring in situations of young patients at risk of evolution.

Conclusion

The torsion bottle is useful to perform a pre-investigation of the presence of a prominence, permitting a simplified assessment of its value thus directing to a specialist for a medical assessment.
Another use of the torsion bottle may be provided for screening to be carried out in countries where due to the low-income problems it is difficult to find a scoliometer® or where the presence of health professionals is not so widely distributed.
The water plastic bottles do not have the same shape in all countries. In some countries, it is typical to find this small bottle with a hollow in the upper part (Fig. 1).
This hollow is useful because it avoids a contact with the bony prominences of the spinous processes. When such a tool is not available and it is necessary to use a bottle with a straight profile, it is important to pay attention to the spinous processes to avoid that the contact with the bottle falses the measurement.

Acknowledgements

Not applicable

Funding

No funding was received.

Availability of data and materials

Please contact the author for data requests.

Motivation for the publication

This paper deserves to be published for a variety of reasons. A very basic tool for assessing the presence of a prominence has been experimented. The prominence is one of the most important clinical signs for the identification of a scoliosis. This data allows to do an objective screening and monitoring of the evolution of the prominence even in a familiar habit and the use of an objective evaluation tool in countries where the gold standard instrument is unavailable.
Ethical approval was obtained from Comitato Etico-Milano-Area B with an approval reference: 20,121,528.
We have obtained consent to publish from the legal parent of all the participants with individual patient data.

Competing interests

Michele Romano is the stockholder of ISICO (Italian Scoliosis Spine Institute). The other author declares that he has no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
3.
Zurück zum Zitat Negrini S, et al. Italian guidelines on rehabilitation treatment of adolescents with scoliosis or other spinal deformities. Eura Medicophys. 2005;41(2):183–201.PubMed Negrini S, et al. Italian guidelines on rehabilitation treatment of adolescents with scoliosis or other spinal deformities. Eura Medicophys. 2005;41(2):183–201.PubMed
4.
Zurück zum Zitat Zaina F, et al. Clinical evaluation of scoliosis during growth: description and reliability. Stud Health Technol Inform. 2008;135:125–38.PubMed Zaina F, et al. Clinical evaluation of scoliosis during growth: description and reliability. Stud Health Technol Inform. 2008;135:125–38.PubMed
5.
Zurück zum Zitat Parent EC, et al. Identifying the best surface topography parameters for detecting idiopathic scoliosis curve progression. Stud Health Technol Inform. 2010;158:78–82.PubMed Parent EC, et al. Identifying the best surface topography parameters for detecting idiopathic scoliosis curve progression. Stud Health Technol Inform. 2010;158:78–82.PubMed
6.
Zurück zum Zitat Aulisa AG, et al. Correlation between hump dimensions and curve severity in idiopathic scoliosis before and after conservative treatment Spine (Phila Pa 1976). 2018;43(2):114-119. Aulisa AG, et al. Correlation between hump dimensions and curve severity in idiopathic scoliosis before and after conservative treatment Spine (Phila Pa 1976). 2018;43(2):114-119.
7.
Zurück zum Zitat Ferraro C, et al. Hump height in idiopathic scoliosis measured using a humpmeter in growing subjects: relationship between the hump height and the Cobb angle and the effect of age on the hump height. Eur J Phys Rehabil Med. 2017;53(3):377-389. Ferraro C, et al. Hump height in idiopathic scoliosis measured using a humpmeter in growing subjects: relationship between the hump height and the Cobb angle and the effect of age on the hump height. Eur J Phys Rehabil Med. 2017;53(3):377-389.
8.
Zurück zum Zitat Amendt LE, et al. Validity and reliability testing of the Scoliometer®. Phys Ther. 1990;70(2):108–17.CrossRefPubMed Amendt LE, et al. Validity and reliability testing of the Scoliometer®. Phys Ther. 1990;70(2):108–17.CrossRefPubMed
9.
Zurück zum Zitat Murrell GA, et al. An assessment of the reliability of the Scoliometer®. Spine (Phila Pa 1976). 1993;18(6):709–12.CrossRef Murrell GA, et al. An assessment of the reliability of the Scoliometer®. Spine (Phila Pa 1976). 1993;18(6):709–12.CrossRef
10.
Zurück zum Zitat Côté P, Kreitz B, Cassidy JD, Dzus AK, Martel J. A study of the diagnostic accuracy and reliability of the scoliometer® and Adam’s forward bend test. Spine 1998; 1;23(7):796-802. Côté P, Kreitz B, Cassidy JD, Dzus AK, Martel J. A study of the diagnostic accuracy and reliability of the scoliometer® and Adam’s forward bend test. Spine 1998; 1;23(7):796-802.
Metadaten
Titel
Torsion bottle, a very simple, reliable, and cheap tool for a basic scoliosis screening
verfasst von
Michele Romano
Matteo Mastrantonio
Publikationsdatum
01.12.2018
Verlag
BioMed Central
Erschienen in
Scoliosis and Spinal Disorders / Ausgabe 1/2018
Elektronische ISSN: 2397-1789
DOI
https://doi.org/10.1186/s13013-018-0150-6

Weitere Artikel der Ausgabe 1/2018

Scoliosis and Spinal Disorders 1/2018 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.