Transabdominal pre-peritoneal hernia repair: risk of operation for recurrence depends on choice of both mesh and fixation device. A study from the Danish Hernia Database
Multiple methods of mesh fixation are available in laparoscopic inguinal hernia repair, as well as multiple types of mesh. No previous studies compare all methods of fixation in TAPP against each other in regards to risk of reoperation for recurrence. In addition, there is little data comparing types of mesh or the relationship between mesh and fixation method.
Methods
We compare the tissue-penetrating methods with non-penetrative as well as no fixation, and examines the interaction of fixation method and choice of mesh. Cohort was established by way of the Danish Hernia Database, identifying patients operated with TAPP from Jan. 2010 to Dec. 2022. Cox’ regression analyses were performed, with multivariate analysis correcting for significant confounding variables, yielding adjusted hazard ratios (aHR) for reoperation for each fixation method. Follow-up analyses investigated whether differences in mesh types significantly impacted the results.
Results
Among 49,029 TAPP repairs, 3.6% experienced reoperation for recurrence over a mean follow-up of 5.76 years. Tack fixation, the most common method, showed the highest reoperation rates (5.3% at 5 years). Glue, self-fixating meshes, and no fixation, had significantly lower risk in comparison (aHRs of 0.25, 0.21, and 0.51, respectively). Even after correcting for weight and pore size, some mesh types significantly impacted risk, with aHRs spanning 0.28 – 1.
Conclusion
Non-penetrative fixation methods and no fixation are associated with lower reoperation rates compared to tissue-penetrative methods, with self-fixating meshes carrying the lowest risk. In addition, we found significant differences in aHR between types of mesh.
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Background
Inguinal hernias are common, and patients are often affected in regards to discomfort, pain, and limited physical activity. Although rare in inguinal hernias, cases with incarceration and strangulation may result in serious morbidity and even mortality. Inguinal hernia repair is among the most commonly performed surgical procedures worldwide [1‐3]. It is considered a low-risk procedure with minimal short term adverse outcomes. In the long term, there is however a risk of developing chronic pain, sexual dysfunction, and recurrence [4‐7]. Due to the high number of inguinal hernia repairs carried out every year, risk of recurrence remain an important measure for health outcomes, patient satisfaction, and socio-economic considerations.
The recommended treatment for inguinal hernia is mesh repair (synthetic, biological, or combined) [8], either with open or laparoscopic approach [8‐10].
Anzeige
Mesh fixation technique in the laparoscopic transabdominal pre-peritoneal approach (TAPP) is largely dependent on surgeon or center preferences. Different methods of mesh fixation have been associated to chronic pain and hernia recurrence, although results vary especially in regards to efficacy of tack fixation compared to other methods [4, 6, 11‐23]. Most studies compare outcomes in tissue-penetrative fixation methods versus non-penetrative fixation methods, or no fixation [11, 12, 14‐16, 18‐22, 24‐31]. To our knowledge, no studies have investigated recurrence rates comparing all mesh fixation methods in a single study population.
Experimental studies show that different types of mesh have different reactions to implantation in regards to shrinkage and adhesion [32], and this reaction further varies with choice of fixation device [33]. A recent study supports this finding in a clinical setting [34]. Accordingly, it seems relevant to consider mesh type and fixation device in addition to classic factors leading to recurrence.
In the present study, we seek to investigate rates of reoperation for inguinal hernia recurrence in elective patients receiving TAPP repair, aiming to compare risk of reoperation for recurrence in no fixation versus both tissue-penetrative mesh fixation and non-penetrative mesh fixation on a thoroughly detailed patient population. We will also investigate whether choice of mesh is a significant factor in recurrence compared to choice of fixation method.
Methods
Participants and data acquirement
We obtained data from The Patient Safety Authority's Online Register and The Danish Hernia Database, a nationwide database prospectively recording detailed information on hernia operations in Denmark since 1998 with a data completeness of 95%. After any hernia repair, the surgeon reports perioperative data and patient-related factors in the database. Approximately 10,000 inguinal hernia operations per year are registered. A detailed description of methodology and limitations of the database is described elsewhere in detail [35].
Anzeige
The inclusion criteriae were age ≥ 18 years and primary elective operation for inguinal or femoral hernia using TAPP between January 2010 and December 2022.
Exclusion criteriae were open approach, perioperative conversion to alternative approach, such as Lichtenstein's procedure, robot-assisted or totally extraperitoneal repair, findings of alternative pathology such as tumor, undescended testicle, femoral artery aneurism etc., and signs of strangulation. Data entries with unclear or defective reporting were also excluded from analysis.
Endpoints of interest
Primary endpoint was the reoperation rate and risk of reoperation due to recurrence after glue fixation of mesh, self-adhesive mesh fixation, no fixation, or tissue-penetrating fixation in the form of clips, sutures and tacks, indicated by 5-year reoperation rate and hazard ratio. Recurrences were defined by the findings of an ipsilateral reoperation following previous TAPP repair. For each patient, multiple recurrences in the same side were not counted separately.
Previous recommendations on choice of fixation method have depended on hernia size and medial hernia [8]. We included analyses on larger and smaller defects for each fixation method, as well as for medial and pantaloon hernias. In these analyses, hernia size or type were appropriately disincluded from multivariate correction.
Secondary endpoints were differences in risk of reoperation between the various mesh types utilised in Denmark during the study period, with a follow-up analysis of whether these differences significantly impact the results of the primary endpoint. The impact of density and pore size of mesh for risk of reoperation was also investigated, as was the impact of absorbable tacks versus non-absorbable.
Statistical analysis
All statistical analyses were done using STATA (StataCorp LP, College Station, TX) with a threshold of significance of 5%.
Prior to analysis of the endpoints, potential confounding variables in patient characteristics were identified. Differences of continuous variables (age, hernia size, mesh size) were examined using student's t-test subsequent to testing for normal distribution and equal variance for continuous data by histograms and Q-Q plots. For data not following a normal distribution, medians were used instead of means with a non-parametric Mann–Whitney U test. To evaluate potential difference in reoperation dependent on center caseload, centres were pooled into groups depending on their annual procedural volume as noted by the Danish Hernia Database. Groups were as follows: low volume (1–99 cases/year), medium volume (100–249 cases/year) and high volume (exceeding 250 cases/year). Categorical characteristics (right/left side surgery, supervised procedure, sex, hernia specification, sliding hernia, center volume) were tested using χ2 test. Statistically significant differences in variables were identified and factored in subsequent multivariate analyses.
Cox proportional hazards regression analyses were used to assess the crude hazard ratios (HR) of reoperation relative to method of mesh fixation or lack thereof. Patient characteristics with significant associations were included for further analysis in a multivariate model correcting for the other independent variables, computing adjusted hazard ratios (aHR). To show the unadjusted cumulative reoperation risk by either fixation method or mesh use, Kaplan–Meier plots were used.
Anzeige
To investigate whether choice of mesh had a significant effect on risk of reoperation, we repeated cox regressions for meshes rather than fixation method, focusing on meshes with at least 5 years observation period and at least 50 includable patients for analysis feasibility. To analyse whether choice of mesh could significantly alter the risk of reoperation for a method of fixation, we repeated this analysis for each fixation group separately.
The largest subgroup was chosen as reference for all regression analyses.
Difference in use of absorbable or permanent tacks between the groups of reoperation or no reoperation was tested using χ2 test, and combinations of high or low weight meshes (± 50 g/m2) with small or large pore size (± 1 mm) was also compared using multivariate Cox proportional hazards regression analyses.
Results
Participation and patient characteristics
A total of 129,271 hernia repairs were identified during the trial period (156 months) as having received mesh implantation in relation to primary inguinal and femoral hernia repair. Of these, 49,029 had TAPP repair and was included in the study (Fig. 1). A total of 2,830 of the identified 129,271 groins (2.2%) had insufficient data for analysis.
Of the included 49,029 hernia repairs, a total of 1,752 had reoperation for recurrence during the follow-up (FU) period (3.6%). The mean FU was 2101 days, or 5.76 years. The most commonly used method of fixation was tacks (23,285, 47.5%), followed by glue fixation (9,181, 18.7%), then clips (6,167, 12.6%), self-fixation (5,075, 10.4%), no fixation (3,065, 6.25%), and lastly sutures (694, 1.4%).
Male sex, sliding, medial, and pantaloon hernias, hernia size and center volume were found to have statistically significant association to reoperation, and subsequent multivariable Cox’ regression analyses corrected for these variables. No correlation was found to mesh length, width or area, nor whether the operation was supervised (Table 1).
Table 1
Patient and peroperative characteristics
Reoperation for recurrence
No known recurrence
P-value
n = 1,752
n = 47,277
Age at procedure, mean
56.54
56.92
p = 0.293
Sex
p < 0.001
Female
131 (1.6%)
7,910 (98.4%)
Male
1,621 (3.9%)
39,367 (96.1%)
Supervised operation
p = 0.208
Yes
63 (2.1%)
2,941 (97.9%)
No
695 (2.5%)
27,446 (97.5%)
Hernia side
p = 0.402
Right
954 (3.5%)
26,222 (96.5%)
Left
798 (3.7%)
21,055 (96.4%)
Hernia specification
p < 0.001
Femoral
63 (2.6%)
2,386 (97.4%)
Lateral
923 (3.4%)
26,112 (96.6%)
Medial
635 (3.8%)
16,143 (96.2%)
Pantaloon
117 (5.0%)
2,224 (95.0%)
Sliding hernia
p < 0.001
Yes
187 (5.7%)
3,112 (94.3%)
No
908 (3.0%)
29,454 (97.0%)
Avg. hernia size, fingers width
2.44
2.76
p < 0.001
Mesh length (mean)
12.65 cm
12.69 cm
p = 0.736
Mesh width (mean)
13.25 cm
13.44 cm
P = 0.202
Mesh area
196.83 cm2
206.02 cm2
P = 0.128
Center volume
Low
589 (5.3%)
10,471 (94.7%)
p < 0.001
Medium
200 (3.2%)
5,993 (96.8%)
High
1001 (3.6%)
26,868 (96.4%)
The data of this study are available on request from the corresponding author in anonymized form. The data are not publicly available due to concerns of participant privacy and restrictions of the utilized database
Reoperation by method of fixation
The reoperation rates at 5 years FU varied from 1.3% to 5.2%, with self-fixation and glue fixation having the lowest rates and tack fixation having the highest (Table 2). As tack fixation was the largest group, it was chosen as reference for hazard analysis. The univariate Cox regression computed HRs ranged from 0.25 (self-fixating) to 0.98 (sutures) and 1 (tack fixation, reference value). The unadjusted cumulative reoperation free survival is depicted as a Kaplan–Meier plot in Fig. 2.
Table 2
Fixation method comparison
Mesh
No
5y Reoperation
HR
aHR (95%-CI)
Glue adhesive
9,181 (18.7%)
1.8%
0.32
0.28 (0.23–0.38)
Smaller defect1
5,299 (23.8%)
1.4%
0.34
0.32 (0.23–0.43)
Larger defect2
2,877 (17.3%)
2.1%
0.22
0.19 (0.11–0.30)
Medial Hernia
2,597 (15.5%)
2.1%
0.33
0.15 (0.04–0.64)
Combined Hernia
275 (11.8%)
2.8%
0.46
0.33 (0.12–0.90)
Self-fixation
5,075 (10.3%)
1.3%
0.25
0.24 (0.18–0.32)
Smaller defect
2,218 (9.9%)
1.4%
0.30
0.24 (0.14–0.42)
Larger defect
2,812 (16.9%)
None
0.20
0.18 (0.11–0.31)
Medial Hernia
1,688 (10.0%)
1.8%
0.26
0.13 (0.03–0.54)
Combined Hernia
285 (1.22%)
None
0.16
0.16 (0.05–0.66)
No fixation
4,627 (9.4%)
4.0%
0.71
0.55 (0.39–0.71)
Smaller defect
2,303 (10.3%)
3.5%
0.83
0.70 (0.51–0.97)
Larger defect
1,964 (11.8%)
5.9%
0.53
0.43 (0.21–0.56)
Medial Hernia
1,455 (8.7%)
5.4%
1.05
1.01 (0.44–2.29)
Combined Hernia
183 (7.8%)
2.6%
0.48
0.58 (0.23–1.45)
Clips
6,167 (12.6%)
4.1%
0.82
0.85 (0.68–1.21)
Smaller defect
3,459 (15.5%)
4.2%
0.99
1.00 (0.79–1.27)
Larger defect
1,178 (7.1%)
5.9%
0.76
0.64 (0.45–1.22)
Medial Hernia
2,327 (13.9%)
4.0%
0.83
1.30 (0.59–2.88)
Combined Hernia
240 (10.3%)
7.1%
0.96
1.01 (0.51–1.99)
Sutures
694 (1.4%)
5.2%
0.98
0.77 (0.36–1.55)
Smaller defect
332 (1.5%)
4.4%
1.1
0.58 (0.22–1.57)
Larger defect
139 (0.8%)
2.5%
0.68
0.90 (0.34–2.4)
Medial Hernia
174 (1.0%)
6.6%
1.13
1.71 (0.23–12.6)
Combined Hernia
26 (1.1%)
6.7%
0.62
0.84 (0.12–6.16)
Tacks
23,285 (47.5%)
5.3%
1 (ref)
1 (ref)
Smaller defect
8,683 (38.9%)
4.8%
-
-
Larger defect
7,680 (46.1%)
7.2%
-
-
Medial Hernia
8,537 (50.9%)
5.1%
-
-
Combined Hernia
1,331 (56.9%)
6.9%
-
-
1: Defect size of 2 fingers width or smaller
2: Defect size larger than 2 fingers width
Fig. 2
Kaplan–Meier plot showing reoperations by fixation method
In the following multivariate correction, the differences between groups remained and proved statistically significant (Table 2). Adjusting for sex, hernia specification, center volume, sliding hernias and size, only the suture-group (aHR = 0.77) and the clips group (aHR = 0.85) had no significant difference to the tack fixation group (p = 0.356 and p = 0.146, respectively). The non-penetrative fixation groups had the lowest adjusted hazards (aHR = 0.28 for glue-fixation and aHR = 0.24 for self-fixation); the no fixation group proved closer to non-penetrative fixation in quality albeit with a slightly higher association to reoperation still, with a stastistically significant aHR of 0.55 (p < 0.001). Of note, the relative superiority of the no fixation method was not evident when looking only at medial or pantaloon hernias, with 5 year reoperation rates similar to tack fixation (5.4% and 5.1% for medial hernias, respectively).
Anzeige
Differences between meshes
The 5-year reoperation rates between meshes ranged from 1.4% (TiMesh) to 13.6% (Physiomesh). The UltraPro mesh was most commonly used and selected as reference for regression.
Of the 20 included meshes, eight had significantly lower aHR compared to the most widely used reference mesh, with none significantly higher: Bard 3D Light, Cousin, Galmesh, Galmesh Light, Optilene, Parietene and UltraPro Advanced. The lowest aHRs with significant difference was ProGrip (aHR = 0.23), which is obligatorily self-fixating, and Optilene (aHR = 0.38).
When analyzing by mesh properties (weight and pore size), there was no significant differences in aHR.
The entire range of included meshes with 5-year reoperation, HR and aHR with 95%-CI are given in Table 3 and for mesh properties in Table 4.
Table 3
Mesh type comparison
Mesh
No
5y Reoperation
HR
aHR (95%-CI)
Bard 3D
708 (1.7%)
2.5%
0.50
1.06 (0.52–1.93)
Bard 3D Light
3,367 (8.1%)
2.8%
0.53
0.59 (0.41–0.80)
Bard Soft
2,473 (6.0%)
4.8%
0.92
0.76 (0.56–1.10)
Cousin
2,274 (5.5%)
3.5%
0.64
0.62 (0.40–0.94)
Galmesh
1,996 (4.8%)
3.4%
0.60
0.56 (0.29–0.82)
Galmesh Light
2,114 (5.1%)
4.0%
0.70
0.65 (0.44–0.91)
Klas
113 (0.3%)
10.9%
2.06
1.87 (0.75–5.11)
Optilene
3,015 (7.3%)
None
0.45
0.38 (0.22–0.67)
Optilene Elastic
310 (0.7%)
None
0.49
0.58 (0.23–1.51)
Parietene
6,676 (16.1%)
3.4%
0.59
0.50 (0.40–0.64)
Parietex
2,261 (5.5%)
5.3%
0.93
0.76 (0.51–1.07)
ProGrip
5,875 (14.2%)
2.7%
0.32
0.23 (0.15–0.30)
Physiomesh
53 (0.1%)
13.6%
2.21
0.83 (0.12–5.91)
Prolene
1,500 (3.6%)
5.3%
0.97
0.72 (0.39–1.07)
SupraMesh
314 (0.8%)
4.4%
0.75
0.70 (0.18–3.14)
SurgiPro
189 (0.5%)
8.8%
1.50
1.90 (0.79–4.16)
TiMesh
78 (0.2%)
1.4%
0.79
0.32 (0.04–2.23)
UltraPro
7,941 (19.1%)
5.1%
1 (ref)
1 (ref)
UltraPro Adv
1,502 (3.6%)
2.8%
0.52
0.56 (0.36–0.76)
VyPro
141 (0.3%)
8.3%
1.57
1.43 (0.75–2.49)
Table 4
Mesh density and pore size
Mesh
No
5y Reoperation
HR
aHR (95%-CI)
Large/Light2
35,741 (77.2%)
4.3%
1 (ref)
1 (ref)
Small/Light1
78 (0.2%)
1.4%
1.13
0.48 (0.07–3.38)
Large/Heavy4
4,004 (8.6%)
5.2%
1.25
1.15 (0.90–1.47)
Small/Heavy3
6,428 (13.9%)
4.6%
1.05
0.89 (0.69–1.15)
1: Mesh with small pore size and lightweight properties
2: Mesh with large pore size and lightweight properties
3: Mesh with small pore size and heavyweight properties
4: Mesh with large pore size and heavyweight properties
Effect of mesh choice on risk of reoperation by method of fixation
For each fixation group, the risk of reoperation was evaluated based on choice of mesh. The most widely used meshes were the Parietene for the glue fixation group, Bard 3D Light for the no fixation group, Bard Soft for the clips fixation group, and UltraPro for the tack fixation group. These were set as reference for intragroup cox regression analysis. As the ProGrip was the only mesh in the self-fixation group, no analysis was done for this fixation method. For the suture fixation group, there was a complete match between reoperations and potential confounding variables, negating the cox regression and rendering the aHR uncomputable. Similar analyses by mesh properties were also carried out.
The data is shown in Tables 5, 6, 7, 8, and 9. No significant difference between meshes was found in the glue fixation group. In no fixation, the Cousin mesh had significantly higher hazard than the reference mesh, despite being among the better performing meshes in the analysis including all fixation methods. In tack fixation, seven meshes had significantly lower hazards than the reference (UltraPro); these were also significantly better in the analysis including all fixation methods.
Table 5
Glue fixation group
Mesh
No
HR
aHR (95%-CI)
Bard 3D
64 (0.9%)
1.00
2.70 (0.39–20.18)
Bard Soft
85 (1.2%)
3.19
2.97 (0.70–12.62)
Cousin
71 (1.0%)
0.92
1.69 (0.23–12.50)
Galmesh
123 (1.8%)
0.62
1.13 (0.15–8.38)
Galmesh Light
408 (5.8%)
0.52
0.48 (0.11–2.02)
Optilene
624 (8.9%)
1.04
1.35 (0.47–3.93)
Parietene
3,472 (49.4%)
1 (ref)
1 (ref)
Parietex
368 (5.2%)
1.46
1.16 (0.54–2.44)
UltraPro
1,418 (20.2%)
1.81
1.15 (0.55–2.44)
UltraPro Adv
359 (5.1%)
0.94
1.28 (0.49–3.35)
Versatex Monofil
1,043 (14.8%)
0.38
0.34 (0.08–1.47)
Large/Light
6,750 (78.1%)
1 (ref)
1 (ref)
Small/Light
3 (0.03%)
-
-
Large/Heavy
453 (5.2%)
0.96
0.93 (0.33–2.59)
Small/Heavy
1,437 (16.6%)
0.66
0.49 (0.18–1.38)
Table 6
No fixation group
Mesh
No
HR
aHR (95%-CI)
Bard 3D Light
624 (30.7%)
1 (ref)
1 (ref)
Bard 3D
58 (2.9%)
3.01
2.64 (0.54–12.82)
Bard Soft
470 (23.1%)
2.21
1.99 (0.74–5.33)
Cousin
491 (24.2%)
2.03
2.81 (1.13–6.99)
Galmesh Light
389 (19.1%)
1.10
0.72 (0.18–2.89)
Large/Light
3,600 (95.9%)
1(ref)
1(ref)
Small/Light
0
-
-
Large/Heavy
14 (0.37%)
-
-
Small/Heavy
139 (3.7%)
1.26
0.96 (0.23–3.96)
Table 7
Clips fixation group
Mesh
No
HR
aHR (95%-CI)
Bard 3D
255 (5.1%)
0.73
1.88 (0.78–4.57)
Bard 3D Light
924 (18.4%)
0.56
0.75 (0.34–1.67)
Bard Soft
1,432 (28.5%)
1 (ref)
1 (ref)
Cousin
573 (11.4%)
0.77
0.85 (0.45–1.63)
Galmesh Light
678 (13.5%)
1.39
1.44 (0.80–2.61)
Optilene
407 (8.1%)
0.98
0.89 (0.36–2.24)
Parietene
513 (10.2%)
1.18
1.28 (0.70–2.35)
Prolene
68 (1.4%)
0.26
0.53 (0.07–3.89)
UltraPro
566 (11.3%)
0.71
0.83 (0.36–1.89)
UltraPro Adv
205 (4.1%)
0.61
0.68 (0.23–1.94)
Large/Light
5,386 (88.8%)
1 (ref)
1 (ref)
Small/Light
0
-
-
Large/Heavy
133 (2.2%)
0.75
0.36 (0.05–2.58)
Small/Heavy
548 (9.0%)
0.88
0.96 (0.42–2.20)
Table 8
Sutures group
Mesh
No
HR
aHR (95%-CI)
Bard Soft
61 (32.3%)
0.30
n/a
UltraPro
128 (67.7%)
1 (ref)
1 (ref)
Large/Light
498 (76.4%)
1 (ref)
1 (ref)
Small/Light
0
-
-
Large/Heavy
68 (10.4%)
1.28
5.20 (0.90–30.2)
Small/Heavy
86 (13.2%)
2.02
4.70 (0.47–46.6)
Table 9
Tacker fixation group
Mesh
No
HR
aHR (95%-CI)
Bard 3D
325 (1.7%)
0.24
0.31 (0.04–2.20)
Bard 3D Light
1,468 (7.8%)
0.52
0.55 (0.35–0.87)
Bard Soft
425 (2.3%)
0.80
1.49 (0.66–3.38)
Cousin
149 (0.8%)
1.06
1.39 (0.35–5.63)
Galmesh
1,775 (9.5%)
0.54
0.39 (0.25–0.61)
Galmesh Light
598 (3.2%)
0.74
0.69 (0.40–1.20)
Klas
106 (0.6%)
1.73
1.83 (0.80–4.15)
Optilene
1,937 (10.3%)
0.36
0.28 (0.17–0.44)
Optilene Elastic
244 (1.3%)
0.45
0.46 (0.17–1.25)
Optilene LP
207 (1.1%)
0.49
0.40 (0.13–1.25)
Parietene
2,674 (14.3%)
0.77
0.62 (0.47–0.83)
Parietex
2,161 (11.5%)
0.89
0.70 (0.51–0.97)
PhysioMesh
51 (0.3%)
1.97
0.79 (0.11–5.62)
Prolene
1,229 (6.6%)
0.93
0.59 (0.37–0.93)
SupraMesh
237 (1.3%)
0.64
1.26 (0.31–5.07)
SurgiPro
168 (0.9%)
1.37
1.89 (0.84–4.27)
TiMesh
75 (0.4%)
0.71
0.26 (0.04–1.89)
UltraPro
5,804 (31.0%)
1 (ref)
1 (ref)
UltraPro Adv
913 (4.9%)
0.53
0.49 (0.32–0.75)
Versatex Monofil
439 (2.3%)
0.81
0.62 (0.34–1.11)
VyPro
133 (0.7%)
1.44
1.17 (0.63–2.16)
Large/Light
14,486 (65.6%)
1 (ref)
1 (ref)
Small/Light
75 (0.3%)
0.86
0.34 (0.05–2.44)
Large/Heavy
3,327 (15.1%)
1.02
0.86 (0.66–1.12)
Small/Heavy
4,200 (19.0%)
0.90
0.77 (0.58–1.03)
There was no demonstrably significant difference in either crude HR or aHR when comparing meshes of various combinations of pore size and density for any fixation method, and the results was consistent with those listed in Table 4.
Supplementary analyses of tack fixation
Of the 23,285 patients in the tack fixation group, a total of 7,305 received fixation with absorbable tacks and 11,223 with non-absorbable tacks, with a data insufficiency in 4,757 cases. In the absorbable group, 342 had a reoperation for recurrence (4.7%) compared to 532 (4.7%) in the non-absorbable group (p = 0.854).
Discussion
This analysis was carried out on a large, diverse, and unselected patient population over a 12-year study period. We found that despite being widely used, the tissue-penetrating fixation methods, and tack fixation in particular, carried a higher risk of reoperation for recurrence, regardless of mesh properties and absorbability of tacks. Meanwhile, the less extensively used non-penetrating methods, and to a lesser extent no fixation, had better results (Table 2).
The apparent superiority of non-penetrative fixation may owe to the fact that self-adhesive meshes and fixation with fibrin sealant enables fixation of the entire mesh, as opposed to tack fixation which omits fixation on the inferolateral edge of the mesh as to avoid damage to nerves and vessels.
Our results are consistent with other studies, although to our knowledge, this is the only large study that includes no fixation in an analysis of multiple fixation methods [11, 12, 14‐16, 18‐22, 24‐31]. Previous studies have compared fixation to no fixation with no distinction in fixation methods [19, 22, 25], non-penetrative to penetrative fixation methods [10‐12, 16, 21, 24, 36‐42], or penetrative fixation to self-fixation [43] and have largely reported little or no differences. Several meta-analyses have been published; two compared tack fixation to no fixation [19, 25], four compared tack fixation to glue fixation [21, 23, 26, 28], two compared no fixation to all fixation [31], and one compared self-fixating mesh with conventional mesh [29], demonstrating non-superiority. One systematic review included trials comparing two methods of fixation or fixation to no fixation, but no trials compared multiple fixation methods to no fixation [44]. These studies have had smaller patient populations and shorter follow-up than the present study.
Our findings are in accordance with previous international recommendations of using non-penetrative fixation of mesh in an attempt to limit risk of chronic pain [45] and to use fixation on medial hernias [8]. More recent publications do not include this recommendation [8, 46, 47], only indicating a weak suggestion of atraumatic devices such as glue-fixation or fibrin sealant to reduce the risk of post-operative, acute pain. While concerns of pain development are not substantiated sufficiently to recommend non-penetrative mesh fixation (or no fixation for non-medial hernias), concerns of recurrence should be demonstrably weighted in these methods'favour.
We found little evidence of an independent impact of the interaction between choice of mesh and fixation method (Tables 5, 6, 7, 8, and 9). Of note, the Cousin mesh, which was among the meshes who had significantly lower hazard compared to the most widely used mesh (UltraPro) in the all-fixation method analysis (Table 3), faired significantly poorer in the no-fixation analysis (Table 6). There were numerous significant differences in tack fixation, though these matched the significant differences in the analysis including all fixation methods. Low-weight meshes (LWM) had no increased risk compared to high-weight meshes (HWM) regardless of pore size in the adjusted cox regression. This is largely consistent with previous studies: a large meta-analysis showed significant superiority of HWM when including non-fixated mesh repairs and no significant difference when analyzing only mesh fixation repairs [48].
The study is not without limitations. The apparent superiority of the no fixation group is vulnerable to bias, as less difficult cases with easier approachability and few complicating factors may lend themselves to fixationless mesh implantation. We also had no way of correcting for surgeon experience, and more experienced surgeons may gravitate towards certain fixation methods. The size of this potential effect is presumably minor, as procedures carried out by supervised residents was not associated with higher reoperation rates in this study. There was some data incompleteness on whether operations were supervised and perioperative finding of sliding hernia, but previous studies indicate that these factors are not associated with higher risk of recurrence [49, 50]. The Danish Hernia Database uses fingers width rather than a centimeter estimate to measure hernia defect size. This is to align with former European Hernia Society standards; these have however since been changed. There is at this point no indications of hernia defect size in centimeters or inches in the Danish Hernia Database.
We used reoperation as a surrogate for true recurrence, as the dataset did not include the latter. Studies show true recurrence rates to be almost 40% higher than reoperation rates [51]. Some patient groups are less likely to seek reoperation, and if these are distributed unevenly among the three analyzed groups, results may shift. In addition, the 5-year reoperation rate did not seem to linearly correlate to reoperation hazard. This suggests that when comparing factors in inguinal hernia repair related to recurrence, an even longer scale of time is needed to provide a more correct estimate of total life-time risk of reoperation. This is consistent with a previous study [52]. While the self-fixating mesh had the best results in this study, it also had the shortest mean follow-up (Fig. 2); more data on long-term recurrence risk is therefore needed before recommending self-fixation over, for example, glue fixation.
The lack of significant differences in the analyses of mesh model hazard for various fixation methods (Tables 5, 6, 7, 8, and 9) may be attributable to a type II error, when looking at a dataset with this many variables and mesh types, and comparably few light-weight/small pore meshes were included.
Per the data, we cannot say that there is such a thing as a right mesh for a given fixation method in TAPP, and either mesh or fixation method may be chosen by the quality of each factor independently.
Ideally, a randomized controlled study would be better fitted to test our study’s endpoints. However, randomizing a sufficient patient population to six different fixation methods (as have been compared in this study) and following them for a period of many years, in addition to accounting for mesh type, would be an arduous and unfeasible task. This, we feel, is the next best option.
There are numerous options available in inguinal hernia repair and even so specifically in TAPP repair. These findings demonstrate that hernia repair with glue fixation and self-fixating meshes carry low risk of reoperation, and may be a preferable approach to other fixation methods.
Choice of mesh also matters, and the mesh-related risk of recurrence is not necessarily dependent on easily categorized factors such as mesh weight or pore size. Recent guidelines recommend using HWM in TAPP [47]. This recommendation is based on the textile characteristics and vulnerability to bursting in tack fixation as well as higher risk of failure in direct and large defects [48]. We did not find that HWM meshes in particular improved results significantly, although some types of mesh had significantly superior results. Arguments for certain mesh-types based on their feasibility with tissue-penetrating fixation methods should be weighted by the apparent higher risk of reoperation carried by these methods.
In the end, several patient and operative related factors affect risk of recurrence. A plethora of different meshes exist and are continuously developed, making it difficult to attain strong data even on a large patient population. In this multifactorial environment, there is a need to be guided by strong evidence. Large, national, or even international, databases of mesh, surgeon, and patient characteristics as well as outcomes are needed to provide actionable data. Future efforts may advantageously make extensive use of such databases as an essential guide to mesh selection and quality assurance, to lastingly improve outcomes for this substantial patient group.
Declarations
This study meets the requirements of the STROBE statement checklist.
Conflicts of interest
We, the authors, declare that we have no conflicts of interest.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Unsere Produktempfehlungen
Die Chirurgie + umfangreiches Online-Angebot
Print-Titel
Das Abo mit mehr Tiefe
Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.
Mit e.Med Chirurgie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Chirurgie, den Premium-Inhalten der chirurgischen Fachzeitschriften, inklusive einer gedruckten chirurgischen Zeitschrift Ihrer Wahl.
Transabdominal pre-peritoneal hernia repair: risk of operation for recurrence depends on choice of both mesh and fixation device. A study from the Danish Hernia Database
Verfasst von
Alexander Mortensen
Anne Bodilsen
Hans Friis-Andersen
Fortelny RH, Petter-Puchner AH, Glaser KS, Redl H (2012) Use of fibrin sealant (Tisseel/Tissucol) in hernia repair: a systematic review. Surg Endosc 26(7):1803–1812. https://doi.org/10.1007/s00464-012-2156-0CrossRefPubMed
4.
Li W, Sun D, Sun Y, Cen Y, Li S, Xu Q, Li Y, Qi Y, Lin Y, Yang T, Xu P (2017) The effect of transabdominal preperitoneal (TAPP) inguinal hernioplasty on chronic pain and quality of life of patients: mesh fixation versus non-fixation. Surg Endosc 31(10):4238–4243. https://doi.org/10.1007/s00464-017-5485-1CrossRefPubMed
5.
Andresen K, Fenger A, Burcharth J, Pommergaard H, Rosenberg J (2017) Mesh fixation methods and chronic pain after transabdominal preperitoneal (TAPP) inguinal hernia surgery: a comparison between fibrin sealant and tacks. Surg Endosc 31(10):4077–4084. https://doi.org/10.1007/s00464-017-5454-8CrossRefPubMed
6.
Wirth U, Saller M, Ahnen T, Köckerling F, Schardey H, Schopf S (2020) Long-term outcome and chronic pain in atraumatic fibrin glue versus staple fixation of extra light titanized meshes in laparoscopic inguinal hernia repair (TAPP): a single-center experience. Surg Endosc 34(5):1929–1938. https://doi.org/10.1007/s00464-019-06965-xCrossRefPubMed
Mahon D, Decadt B, Rhodes M (2003) Prospective randomized trial of laparoscopic (transabdominal preperitoneal) vs open (mesh) repair for bilateral and recurrent inguinal hernia. Surg Endosc 17(9):1386–1390. https://doi.org/10.1007/s00464-002-9223-xCrossRefPubMed
10.
Aiolfi A, Cavalli M, Ferraro S, Manfredini L, Bonitta G, Bruni PG, Bona D, Campanelli G (2021) Treatment of inguinal Hernia: Systematic review and updated network meta-analysis of randomized controlled trials. Ann Surg 274(6):954–961. https://doi.org/10.1097/sla.0000000000004735CrossRefPubMed
Ceccarelli G, Casciola L, Pisanelli MC, Bartoli A, Di zitti L, Spaziani A, Biancafarina A, Stefanoni M, Patriti A (2008) Comparing fibrin sealant with staples for mesh fixation in laparoscopic transabdominal hernia repair: a case control-study. Surg Endosc 22(3):668–73. https://doi.org/10.1007/s00464-007-9458-7CrossRefPubMed
13.
Pommergaard HC, Burcharth J, Andresen K, Fenger AQ, Rosenberg J (2017) No difference in sexual dysfunction after transabdominal preperitoneal (TAPP) approach for inguinal hernia with fibrin sealant or tacks for mesh fixation. Surg Endosc 31(2):661–666. https://doi.org/10.1007/s00464-016-5017-4CrossRefPubMed
14.
Habeeb T, Mokhtar MM, Sieda B et al (2020) Changing the innate consensus about mesh fixation in trans-abdominal preperitoneal laparoscopic inguinal hernioplasty in adults: Short and long term outcome. Randomized controlled clinical trial. Int J Surg 83:117–124. https://doi.org/10.1016/j.ijsu.2020.09.013CrossRefPubMed
15.
Ferrarese A, Marola S, Surace A et al (2014) Fibrin glue versus stapler fixation in laparoscopic transabdominal inguinal hernia repair: a single center 5-year experience and analysis of the results in the elderly. Int J Surg 12(Suppl 2):S94–S98. https://doi.org/10.1016/j.ijsu.2014.08.371CrossRefPubMed
16.
Fenger AQ, Helvind NM, Pommergaard H, Burcharth J, Rosenberg J (2016) Fibrin sealant for mesh fixation in laparoscopic groin hernia repair does not increase long-term recurrence. Surg Endosc 30(3):986–992. https://doi.org/10.1007/s00464-015-4280-0CrossRefPubMed
17.
Christoffersen MW, Brandt E, Helgstand F, Westen M, Rosenberg J, Kehlet H, Strandfelt P, Bisgaard T (2015) Recurrence rate after absorbable tack fixation of mesh in laparoscopic incisional hernia repair. Br J Surg 102(5):541–547. https://doi.org/10.1002/bjs.9750CrossRefPubMed
18.
Shi Z, Fan X, Zhai S, Zhong X, Huang D (2017) Fibrin glue versus staple for mesh fixation in laparoscopic transabdominal preperitoneal repair of inguinal hernia: a meta-analysis and systematic review. Surg Endosc 31(2):527–537. https://doi.org/10.1007/s00464-016-5039-yCrossRefPubMed
19.
Eltair M, Hajibandeh S, Hajibandeh S, Balakrishnan S et al (2019) Meta-analysis of laparoscopic groin hernia repair with or without mesh fixation. Int J Surg 71:190–199. https://doi.org/10.1016/j.ijsu.2019.10.004CrossRefPubMed
20.
Antoniou SA, Köhler G, Antoniou GA, Muysoms FE, Pointner R, Granderath F (2016) Meta-analysis of randomized trials comparing nonpenetrating vs mechanical mesh fixation in laparoscopic inguinal hernia repair. Am J Surg 211(1):239-249 e2. https://doi.org/10.1016/j.amjsurg.2015.06.008CrossRefPubMed
21.
Sajid MS, Ladwa N, Kalra L, McFall M, Baig MK, Sains P (2013) A meta-analysis examining the use of tacker mesh fixation versus glue mesh fixation in laparoscopic inguinal hernia repair. Am J Surg 206(1):103–111. https://doi.org/10.1016/j.amjsurg.2012.09.003CrossRefPubMed
22.
Mayer F, Niebuhr H, Lechner M, Dinnewitzer A, Köhler G, Hukauf M, Fortelny RH, Bittner R, Köcklerling F (2016) When is mesh fixation in TAPP-repair of primary inguinal hernia repair necessary? The register-based analysis of 11,230 cases. Surg Endosc 30(10):4363–4371. https://doi.org/10.1007/s00464-016-4754-8CrossRefPubMedPubMedCentral
23.
Li J, Ji Z, Zhang W (2015) Staple fixation against adhesive fixation in laparoscopic inguinal Hernia repair: A meta-analysis of randomized controlled trials. Surg Laparosc Endosc Percutan Tech 25(6):471–477. https://doi.org/10.1097/sle.0000000000000214CrossRefPubMed
24.
Fortelny RH, Petter-Puchner AH, May C, Jaksch W, Benesch T, Khakpour Z, Redl H, Glaser KS (2012) The impact of atraumatic fibrin sealant vs. staple mesh fixation in TAPP hernia repair on chronic pain and quality of life: results of a randomized controlled study. Surg Endosc 26(1):249–54. https://doi.org/10.1007/s00464-011-1862-3CrossRefPubMed
25.
Sajid MS, Ladwa N, Kalra L, Hutson K, Sains P, Baig MK (2012) A meta-analysis examining the use of tacker fixation versus no-fixation of mesh in laparoscopic inguinal hernia repair. Int J Surg 10(5):224–231. https://doi.org/10.1016/j.ijsu.2012.03.001CrossRefPubMed
26.
Habib Bedwani NAR, Kelada M, Smart N, Szydlo R, Patten DK, Bhargava A (2021) Glue versus mechanical mesh fixation in laparoscopic inguinal hernia repair: meta-analysis and trial sequential analysis of randomized clinical trials. Br J Surg 108(1):14–23. https://doi.org/10.1093/bjs/znaa002CrossRefPubMed
27.
Alabi A, Haladu N, Scott NW, Imamura M, Ahmed I, Ramsay G, Brazelli M (2022) Mesh fixation techniques for inguinal hernia repair: an overview of systematic reviews of randomised controlled trials. Hernia 26(4):973–987. https://doi.org/10.1007/s10029-021-02546-xCrossRefPubMed
28.
Shah NS, Fullwood C, Siriwardena AK, Sheen AJ (2014) Mesh fixation at laparoscopic inguinal hernia repair: a meta-analysis comparing tissue glue and tack fixation. World J Surg 38(10):2558–2570. https://doi.org/10.1007/s00268-014-2547-6CrossRefPubMed
29.
Wang D, Jiang J, Fu Y, Qu P (2022) The comparison of self-gripping mesh and conventional mesh in laparoscopic inguinal hernia repair: the results of meta-analysis. Updates Surg 74(3):857–863. https://doi.org/10.1007/s13304-021-01218-wCrossRefPubMed
30.
Riemenschneider KA, Lund H, Pommergaard HC (2023) No evidence for fixation of mesh in laparoscopic transabdominal preperitoneal (TAPP) inguinal hernia repair: a systematic review and meta-analysis of randomized controlled trials. Surg Endosc 37(11):8291–8300. https://doi.org/10.1007/s00464-023-10237-0CrossRefPubMedPubMedCentral
31.
Kobayashi F, Watanabe J, Koizumi M, Sata N (2023) Efficacy and safety of mesh non-fixation in patients undergoing laparo-endoscopic repair of groin hernia: a systematic review and meta-analysis. Hernia 27(6):1415–1427. https://doi.org/10.1007/s10029-023-02919-4CrossRefPubMedPubMedCentral
Harslof S, Zinther NB, Harslof T, Danielsen C, Wara P, Friis-Andersen H (2017) Mesh shrinkage depends on mesh properties and anchoring device: an experimental long-term study in sheep. Hernia 21(1):107–113. https://doi.org/10.1007/s10029-016-1528-0CrossRefPubMed
34.
Novik B, Sandblom G, Ansorge C, Thorell A (2022) Association of mesh and fixation options with reoperation risk after laparoscopic groin Hernia surgery: A swedish Hernia registry study of 25,190 totally extraperitoneal and transabdominal preperitoneal repairs. J Am Coll Surg 234(3):311–325. https://doi.org/10.1097/xcs.0000000000000060CrossRefPubMedPubMedCentral
Boldo E (2008) Pain after laparascopic bilateral hernioplasty : Early results of a prospective randomized double-blind study comparing fibrin versus staples. Surg Endosc 22(5):1206–1209. https://doi.org/10.1007/s00464-007-9587-zCrossRefPubMed
37.
Olmi S, Scaini A, Erba L, Guaglio M, Croce E (2007) Quantification of pain in laparoscopic transabdominal preperitoneal (TAPP) inguinal hernioplasty identifies marked differences between prosthesis fixation systems. Surgery 142(1):40–46. https://doi.org/10.1016/j.surg.2007.02.013CrossRefPubMed
38.
Lovisetto F, Zonta S, Rota E, Mazzilli M, Bardone M, Bottero L, Faillace G, Longoni M (2007) Use of human fibrin glue (Tissucol) versus staples for mesh fixation in laparoscopic transabdominal preperitoneal hernioplasty: a prospective, randomized study. Ann Surg 245(2):222–231. https://doi.org/10.1097/01.sla.0000245832.59478.c6CrossRefPubMedPubMedCentral
39.
Tolver MA, Rosenberg J, Juul P, Bisgaard T (2013) Randomized clinical trial of fibrin glue versus tacked fixation in laparoscopic groin hernia repair. Surg Endosc 27(8):2727–2733. https://doi.org/10.1007/s00464-012-2766-6CrossRefPubMed
40.
Brugger L, Bloesch M, Ipaktchi R, Kurmann A, Candinas D, Beldi G (2012) Objective hypoesthesia and pain after transabdominal preperitoneal hernioplasty: a prospective, randomized study comparing tissue adhesive versus spiral tacks. Surg Endosc 26(4):1079–1085. https://doi.org/10.1007/s00464-011-2003-8CrossRefPubMed
41.
Wang MG, Tian M, Zhao X, Nie Y, Chen J, Shen Y (2013) Effectiveness and safety of n-butyl-2-cyanoacrylate medical adhesive for noninvasive patch fixation in laparoscopic inguinal hernia repair. Surg Endosc 27(10):3792–3798. https://doi.org/10.1007/s00464-013-2970-zCrossRefPubMed
42.
Bittner R, Gmähle E, Gmähle B, Schwartz J, Aasvang E, Kehlet H (2010) Lightweight mesh and noninvasive fixation: an effective concept for prevention of chronic pain with laparoscopic hernia repair (TAPP). Surg Endosc 24(12):2958–2964. https://doi.org/10.1007/s00464-010-1140-9CrossRefPubMed
43.
Fumagalli Romario U, Puccetti F, Elmore U, Massaron S, Rosati R (2013) Self-gripping mesh versus staple fixation in laparoscopic inguinal hernia repair: a prospective comparison. Surg Endosc 27(5):1798–1802. https://doi.org/10.1007/s00464-012-2683-8CrossRefPubMed
44.
Lederhuber H, Stiede F, Axer S, Dahlstrand U (2017) Mesh fixation in endoscopic inguinal hernia repair: evaluation of methodology based on a systematic review of randomised clinical trials. Surg Endosc 31:4370–4381. https://doi.org/10.1007/s00464-017-5509-xCrossRefPubMed
45.
Bittner R, Montgomery MA, Arregui E et al (2015) Update of guidelines on laparoscopic (TAPP) and endoscopic (TEP) treatment of inguinal hernia (International Endohernia Society). Surg Endosc 29(2):289–321. https://doi.org/10.1007/s00464-014-3917-8CrossRefPubMed
46.
Bittner R, Bain K, Berrevoet F et al (2019) Update of Guidelines for laparoscopic treatment of ventral and incisional abdominal wall hernias (International Endohernia Society (IEHS))-Part A. Surg Endosc 33(10):3069–3139. https://doi.org/10.1007/s00464-019-06907-7CrossRefPubMedPubMedCentral
Bakker WJ, Aufenacker TJ, Boschman JS, Burgmans JPJ (2021) Heavyweight mesh is superior to lightweight mesh in laparo-endoscopic inguinal hernia repair: A meta-analysis and trial sequential analysis of randomized controlled trials. Ann Surg 273(5):890–899. https://doi.org/10.1097/sla.0000000000003831CrossRefPubMed
49.
Cristophersen C, Fonnes S, Andresen K, Rosenberg J (2023) Risk of reoperation for recurrence after elective primary groin and ventral hernia repair by supervised residents. JAMA Surg 158(4):359–367. https://doi.org/10.1001/jamasurg.2022.7502CrossRef
Kald A, Nilsson E, Anderberg B et al (1998) Reoperation as surrogate endpoint in hernia surgery. A three year follow-up of 1565 herniorrhaphies. Eur J Surg 164(1):45–50. https://doi.org/10.1080/110241598750004940CrossRefPubMed
Ist eine Speichelgangsblockade und die damit verbundene Sialadenitis nicht durch Steine bedingt, wird oftmals versucht, die Symptomatik zu lindern, indem man den Gang mechanisch weitet. Ein aktuelles Review kann den Eingriff als chancenreiches Verfahren bestätigen und deckt gleichzeitig Schwächen auf.
Mehr als eine Million Menschen in Deutschland leiden unter Hallux valgus – eine Fehlstellung des Großzehs, die je nach Schweregrad und Symptomen behandelt wird. Welche neuen Empfehlungen die aktualisierte S2e-Leitlinie bietet, erklärt der Orthopäde Prof. Sebastian Baumbach im MedTalk Leitlinie KOMPAKT der Zeitschrift Orthopädie und Unfallchirurgie.
Ein chirurgischer Eingriff kann für Patienten mit primärem Hyperparathyreoidismus gegenüber dem konservativen Management metabolisch von Vorteil sein. Denn wie eine Studie zeigt, senkt die Operation das Diabetesrisiko.