Skip to main content
Erschienen in: Journal of Interventional Cardiac Electrophysiology 1/2020

26.10.2019

Transseptal puncture performed with the new needle-free ‘SafeSept’ guidewire: a multicentre experience

verfasst von: Anthony W. C. Chow, Vanessa Cobb, Ali Sepahpour, James W. McCready

Erschienen in: Journal of Interventional Cardiac Electrophysiology | Ausgabe 1/2020

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Increasing numbers of transseptal punctures (TSP) are performed for different cardiac interventions, especially for left atrial (LA) ablation of atrial fibrillation (AF). Significant TSP-related complications may occur even by experienced operators. Any strategies to mitigate these should be adopted. Previous reports have confirmed the safety and efficacy of the first generation nitinol guidewire for TSP. The objective was to assess the safety and efficacy of a new nitinol ‘needle free’ SafeSept guidewire (NFSG) for TSP. The distal J-shaped tip (0.014 in.) of the NFSG is unchanged in design; however, the shaft is thicker (0.0315 in.) and stiffer.

Methods

This is a multicentre study evaluating TSP without a transseptal needle, using the NFSG via a standard transseptal sheath-dilator, in patients with an indication for TSP.

Results

A total of 145 consecutive unselected patients were studied. The mean patient age was 59 ± 15. Thirty-one (21.4%) were female and 43 (29.7%) had a previous procedure requiring TSP. The mean transverse LA diameter was 4.0 ± 0.8 cm. AF or LA tachycardia comprised the major indication for TSP (n = 115). Transesophageal echocardiography for direct guidance of TSP was utilised in cases performed under general anesthesia (n = 66, 45.5%). LA access by TSP using the NFSG was achieved successfully in 100% of cases after 1.6 ± 2.3 passes. There were no TSP or other major procedural complications.

Conclusions

Use of the NFSG appears to be a safe and effective method in performing TSP. The major improvement in the design of the NFSG is that TSP can be performed without a transseptal needle and exchange for a standard guidewire is not required upon establishing LA access.
Literatur
1.
Zurück zum Zitat De Ponti R, Cappato R, Curnis A, Della Bella P, Padeletti L, Raviele A, et al. Trans-septal catheterization in the electrophysiology laboratory: data from a multicenter survey spanning 12 years. J Am Coll Cardiol. 2006;47(5):1037–42.CrossRef De Ponti R, Cappato R, Curnis A, Della Bella P, Padeletti L, Raviele A, et al. Trans-septal catheterization in the electrophysiology laboratory: data from a multicenter survey spanning 12 years. J Am Coll Cardiol. 2006;47(5):1037–42.CrossRef
3.
Zurück zum Zitat Holmes DR, Kar S, Price MJ, Whisenant B, Sievert H, Doshi SK, et al. Prospective randomized evaluation of the watchman left atrial appendage closure device in patients with atrial fibrillation versus long-term warfarin therapy: the PREVAIL trial. J Am Coll Cardiol. 2014;64(1):1–12.CrossRef Holmes DR, Kar S, Price MJ, Whisenant B, Sievert H, Doshi SK, et al. Prospective randomized evaluation of the watchman left atrial appendage closure device in patients with atrial fibrillation versus long-term warfarin therapy: the PREVAIL trial. J Am Coll Cardiol. 2014;64(1):1–12.CrossRef
4.
Zurück zum Zitat Rogers DP, Lambiase PD, Dhinoja M, Lowe MD, Chow AW. Right atrial angiography facilitates transseptal puncture for complex ablation in patients with unusual anatomy. J Interv Card Electrophysiol. 2006;17:29–34.CrossRef Rogers DP, Lambiase PD, Dhinoja M, Lowe MD, Chow AW. Right atrial angiography facilitates transseptal puncture for complex ablation in patients with unusual anatomy. J Interv Card Electrophysiol. 2006;17:29–34.CrossRef
10.
Zurück zum Zitat Katritsis GD, Siontis GC, Giazitzoglou E, Fragakis N, Katritsis DG. Complications of transseptal catheterization for different cardiac procedures. Int J Cardiol. 2013;15;168(6):5352–4.CrossRef Katritsis GD, Siontis GC, Giazitzoglou E, Fragakis N, Katritsis DG. Complications of transseptal catheterization for different cardiac procedures. Int J Cardiol. 2013;15;168(6):5352–4.CrossRef
Metadaten
Titel
Transseptal puncture performed with the new needle-free ‘SafeSept’ guidewire: a multicentre experience
verfasst von
Anthony W. C. Chow
Vanessa Cobb
Ali Sepahpour
James W. McCready
Publikationsdatum
26.10.2019
Verlag
Springer US
Erschienen in
Journal of Interventional Cardiac Electrophysiology / Ausgabe 1/2020
Print ISSN: 1383-875X
Elektronische ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-019-00608-w

Weitere Artikel der Ausgabe 1/2020

Journal of Interventional Cardiac Electrophysiology 1/2020 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.