Treatment of osteochondral defects: chondrointegration of metal implants improves after hydroxyapatite coating

Zeitschrift:
Knee Surgery, Sports Traumatology, Arthroscopy
Autoren:
Hanna Schell, Elisabeth Zimpfer, Katharina Schmidt-Bleek, Tobias Jung, Georg N. Duda, Leif Ryd
Wichtige Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Abstract

Purpose

The treatment of osteochondral defects in joint cartilage remains challenging due to its limited repair capacity. This study presents a metallic osteochondral plug with hydroxyapatite (HA)-coated cap edges for improved implant-tissue contact. The hypothesis was that improved attachment prevents from synovial fluid-influx and thereby avoids osteolysis and resulting implant instability.

Methods

In total, 24 female, adult sheep were randomized into three groups. All animals received an Episealer®-implant in the medial condyle of the right knee. The implants were coated with two different HA versions or uncoated (control group). After 12 weeks, the implant-tissue connections were analysed radiologically and histologically.

Results

In general, the groups with the coated cap edges showed a better quality of tissue connection to the implant. The occurrence of gaps between tissue and implant was more seldom, the binding of calcified and hyaline cartilage to the cap was significantly better than in the uncoated group. A histomorphometrically measured lower amount of void space in these groups compared to the group with the uncoated edges confirmed that.

Conclusions

The hypothesis of a tighter cartilage bone contact was confirmed. The HA coating of the implant’s cap edges resulted in better adherence of cartilage to the implant, which was not previously reported. In conclusion, this led to a better contact between implant and cartilage as well as neighbouring bone. In clinical routine, joint fluid is aggressive, penetrates through cartilage rifts, and promotes osteolysis and loosening of implants. The observed sealing effect will act to prevent joint fluid to get access to the implant–tissue interfaces. Joint fluid is aggressive, can cause osteolysis, and can, clinically cause pain. These effects are liable to decrease with these findings and will further the longevity of these osteochondral implants.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de. Zusätzlich können Sie eine Zeitschrift Ihrer Wahl in gedruckter Form beziehen – ohne Aufpreis.

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel
  1. Sie können e.Med Orthopädie & Unfallchirurgie 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.

Neu im Fachgebiet Orthopädie und Unfallchirurgie

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Orthopädie und Unfallchirurgie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise