Skip to main content
Erschienen in: Pediatric Radiology 2/2020

Open Access 06.11.2019 | Original Article

Two-dimensional ultrasound measurements vs. magnetic resonance imaging-derived ventricular volume of preterm infants with germinal matrix intraventricular haemorrhage

verfasst von: Casper Beijst, Jeroen Dudink, Rens Wientjes, Isabel Benavente-Fernandez, Floris Groenendaal, Margaretha J. Brouwer, Ivana Išgum, Hugo W. A. M. de Jong, Linda S. de Vries

Erschienen in: Pediatric Radiology | Ausgabe 2/2020

Abstract

Background

Post-haemorrhagic ventricular dilatation can be measured accurately by MRI. However, two-dimensional (2-D) cranial US can be used at the bedside on a daily basis.

Objective

To assess whether the ventricular volume can be determined accurately using US.

Materials and methods

We included 31 preterm infants with germinal matrix intraventricular haemorrhage. Two-dimensional cranial US images were acquired and the ventricular index, anterior horn width and thalamo-occipital distance were measured. In addition, cranial MRI was performed. The ventricular volume on MRI was determined using a previously validated automatic segmentation algorithm. We obtained the correlation and created a linear model between MRI-derived ventricular volume and 2-D cranial US measurements.

Results

The ventricular index, anterior horn width and thalamo-occipital distance as measured on 2-D cranial US were significantly associated with the volume of the ventricles as determined with MRI. A general linear model fitted the data best: ∛ventricular volume (ml) = 1.096 + 0.094 × anterior horn width (mm) + 0.020 × thalamo-occipital distance (mm) with R2 = 0.831.

Conclusion

The volume of the lateral ventricles of infants with germinal matrix intraventricular haemorrhage can be estimated using 2-D cranial US images by application of a model.
Hinweise

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Very-low-birth-weight (<1,500 g) infants are at increased risk of developing a germinal matrix intraventricular haemorrhage, which occurs in 20–25% of cases [1, 2]. Approximately 35% of the infants who have a germinal matrix intraventricular haemorrhage develop post-haemorrhagic ventricular dilatation [3]. In severe cases of post-haemorrhagic ventricular dilatation, there is an increased risk of raised intracranial pressure and subsequent brain damage, and an intervention such as a lumbar puncture, the placement of a subcutaneous reservoir or a ventriculo-peritoneal shunt might be required [4, 5].
To diagnose and monitor post-haemorrhagic ventricular dilatation and to assess whether intervention is required, two-dimensional (2-D) cranial US is used. The most important measurements are the ventricular index, anterior horn width and thalamo-occipital distance [6]. These measurements aid in managing post-haemorrhagic ventricular dilatation, i.e. in determining when to perform lumbar punctures, whether to insert a ventricular reservoir, how often and how much cerebrospinal fluid to withdraw and whether to insert a ventriculo-peritoneal shunt. Another important prognostic imaging marker is the ventricular volume. This, however, cannot be measured using 2-D cranial US. Magnetic resonance imaging (MRI) is generally considered to be the most accurate way to determine the ventricular volume in neonates. Not all premature babies undergo MR scanning. When they do, it is usually performed once or twice during the neonatal period and at most centers the MRI is performed around term-equivalent age.
Although MRI provides high-resolution and anatomical detail, MRI has limitations. It can be challenging to maintain physiological stability during an MRI examination. Moreover, there are potential risks for infants during transfer to the MR unit, sedation might be required, and the examination is time-consuming and often costly. Also, MRI cannot be easily repeated for monitoring and follow-up. Two-dimensional cranial US is routinely used on the neonatal intensive care unit to diagnose and assess the progression of post-haemorrhagic ventricular dilatation. Therefore, it would be advantageous if US could also be used for ventricular volume assessment [7].
The purpose of the study was to assess whether the ventricular volume can be determined accurately using US, to pave the way for a US-derived measurement of ventricular volume. The rationale is that the ventricular volume might be more directly linked to intracranial pressure, which in turn is considered to be the actual cause of additional damage to the surrounding white matter.

Materials and methods

Study population

For this study, we included preterm infants with a germinal matrix intraventricular haemorrhage and a gestational age less than 31 weeks who were admitted to the Level 3 neonatal intensive care unit between 2007 and 2013 and had an MRI examination at term-equivalent age along with coronal and sagittal 2-D cranial US examination within 3 days of the MRI. Because we used clinically obtained anonymized data, our medical ethics committee waived the requirement for written informed parental consent for participation in the study.

Ultrasonography

The infants routinely underwent cranial US for the assessment of ventricular dilatation. Two-dimensional cranial US imaging was performed by neonatologists and physician assistants with experience varying between 1 year and 25 years. The US images were reviewed by neonatologists with 10 years to 25 years of experience in performing cranial ultrasound. For comparison of 2-D cranial US and MR images, we included only the US images that were acquired within 3 days of the MRI examination. Ultrasound images were acquired using a Xario XG, Aplio MX or Aplio ultrasound machine (Toshiba Medical Systems, Tokyo, Japan) with PST65-AT (9S4) transducer operating in a frequency range between 4.2 MHz and 9 MHz, or an ATL-5000 ultrasound machine (Philips Medical Systems, Best, The Netherlands) with a C8–5 transducer operating in a frequency range between 5 MHz and 8 MHz.
In general, US imaging was performed within 6 h of admission, at least three times in the first week after birth, then weekly till discharge to a Level 2 hospital, and again at term-equivalent age. Criteria for post-haemorrhagic ventricular dilatation were a ventricular index >97th percentile according to Levene and Starte [8], anterior horn width >6 mm or thalamo-occipital distance >24 mm, according to Davies et al. [9].
For the assessment of ventricular volume, we acquired three linear measurements from the cranial US images: the ventricular index (defined as the distance between the falx and the lateral wall of the anterior horn in the coronal plane) [8], the anterior horn width (defined as the diagonal width of the anterior horn measured at its widest point in the coronal plane) [9], and the thalamo-occipital distance (defined as the distance between the outermost point of the thalamus at its junction with the choroid plexus and the outermost part of the occipital horn in the parasagittal plane) [6], as shown in Fig. 1.
The ventricular index, anterior horn width and thalamo-occipital distance were measured by multiple observers. When measurements of a single infant were performed by more than one observer, we used measurements of the most experienced observer for further analysis. We calculated the intraclass correlation coefficient for the images where multiple observer measurements were available.

Magnetic resonance imaging

MR images were acquired at term-equivalent age on an Achieva 3.0-tesla (T) MR system (Philips Healthcare, Best, The Netherlands). At our institution, all preterm infants with a germinal matrix intraventricular haemorrhage with a gestational age less than 28 weeks undergo MRI. In those with a gestational age >28 weeks, MRI is performed when a large intraventricular haemorrhage is diagnosed with US. In infants with US abnormalities, such as a large intraventricular haemorrhage, MRI is also performed to obtain more information on additional lesions in the white matter and the cerebellum. Different exam protocols were used for infants scanned before and after May 2008. Before May 2008, infants were scanned with an axial 3-D T1-weighted gradient echo sequence using a repetition time of 9.4 ms, echo time of 4.6 ms, a slice thickness of 2.0 mm without gap; and an axial T2-weighted spin-echo sequence with a repetition time of 6,293 ms, echo time of 120 ms and slice thickness of 2.0 mm. As of June 2008, T1-weighted images were acquired using a coronal 3-D gradient echo scan with a repetition time of 9.5 ms, echo time of 4.6 ms, slice thickness of 1.2 mm and no gap; and T2-weighted images were acquired using a coronal spin-echo scan with a repetition time of 4,847 ms, echo time of 150 ms, slice thickness of 1.2 mm and no gap. More details on the MRI protocol have been published [10].
We used a probabilistic brain segmentation algorithm as described by Anbeek et al. [11] to perform the automatic segmentation of the brain volume. The algorithm not only segments the ventricles, but also separately classifies the white matter, cortical and deep grey matter, brainstem, cerebellum and extracerebral cerebrospinal fluid. The algorithm does this by a supervised voxel classification considering the intensity of T1- and T2-weighted images as well as the location of the voxel within the brain. The resulting segmentations are obtained by summation of the individual voxels with sufficient calculated probability of belonging to a certain tissue type. A study was previously performed to validate the resulting segmentations against human observers, and found no significant differences in the segmented volumes in five preterm infants [12]. The volume of the automatically segmented ventricles (lateral, 3rd and 4th ventricles) was obtained by summation of the number of voxels weighted by their respective probabilities as obtained from the classification algorithm. Because the training of automatic segmentation algorithms for pathological anatomy remains a challenge, we reviewed all automatic segmentations and manually corrected wherever necessary. An MRI segmentation of the ventricles is shown in Fig. 2.

Data analysis

For data analysis we used the SPSS statistical software package version 21 (IBM, Armonk, NY). We performed linear regression analysis to study the association between the ventricular volume and the US measurements. The ventricular volume on MRI was determined for the left and right separately, and compared to the ventricular index, anterior horn width and thalamo-occipital distance measurements of that respective side of the brain. Furthermore, we fitted a general linear model with multiple covariates using forward selection with a P-value <0.05. We performed a log transformation and a transformation with the cube root (∛) of the ventricular volume to improve homoscedasticity (the homogeneity of the variance) and obtain a normal distribution of the residuals. We performed the linear US measurements, the volume measurements on MRI, and the model fitting for both the left and right side separately. In this way, we could optimally incorporate the measurements of infants with asymmetrical ventricles in our model.

Model validation

For validation purposes we compared our model to the model of Benavente-Fernandez et al. [13]. This model predicts the ventricular volume as measured by 3-D cranial US, using linear 2-D cranial US measurements in infants with post-haemorrhagic ventricular dilatation.

Results

In total, 31 preterm infants with germinal matrix intraventricular haemorrhage were eligible for inclusion in this study, and 14 infants (45%) developed post-haemorrhagic ventricular dilatation (Table 1). T1- and T2-weighted MR images were obtained at term-equivalent age (median 41.1 weeks, range 40.3–43.1 weeks). Ventricular volumes were obtained by automatic segmentation. The automatic segmentation was manually reviewed and corrected where necessary.
Table 1
Patient characteristics
  
Mean
Standard Deviation
Minimum
Maximum
Count
Gestational age (weeks)
 
27.2
1.5
24.9
30.4
 
Birth weight (g)
 
1,039
288
670
1,775
 
Post-menstrual age MRI (weeks)
 
41.3
0.8
40.3
43.1
 
Weight at term-equivalent age (g)
 
3,496
480
2,285
4,735
 
Grade intra-ventricular haemorrhage
1
    
0
2
    
25
3
    
6
Ventriculo-peritoneal drain
Yes
    
2
No
    
29
Gender
Female
    
11
Male
    
20
Within 3 days of the MRI examination, coronal and sagittal cerebral cranial US images were acquired (Fig. 1); at this time no ventricular taps from a Rickham reservoir were performed. For each child the ventricular index, anterior horn width and thalamo-occipital distance were measured on both sides and used for further statistical analysis as separate measurements (Table 2). We determined the intraclass correlation coefficient (ICC) using the one-way random effects models, yielding ICCs of 0.924 (95% confidence interval [CI] 0.854–0.961), 0.988 (95% CI 0.975–0.994) and 0.984 (95% CI 0.951–0.995) for the ventricular index, anterior horn width and thalamo-occipital distance, respectively.
Table 2
MRI and two-dimensional cranial US measurements at term-equivalent age (n=31)
 
Mean
Median
Minimum
Maximum
SD
Ventricular volume, MRI (ml)
8.1
6.1
2.0
30.3
5.9
Ventricular index, cranial US (mm)
13.6
13.7
8.4
19.5
2.0
Anterior horn width, cranial US (mm)
4.4
3.7
0.6
12.1
2.6
Thalamo-occipital distance, cranial US (mm)
20.7
20.4
7.4
46.0
8.0
SD standard deviation
Figure 3 shows the ventricular volume plotted as a function of ventricular index, anterior horn width and thalamo-occipital distance. We performed a linear fit to add a trendline to the plot, which resulted in the following relations for each of the covariates: ventricular volume (ml) = −16.1+1.77 × ventricular index (mm) (R2 = 0.334); ventricular volume (mm) = −0.01+1.84 × anterior horn width (mm) (R2 = 0.658); and ventricular volume (ml) = −3.70 + 0.576 × thalamo-occipital distance (mm) (R2 = 0.574).
We fitted a general linear model to the data using forward variable selection (P<0.05) to obtain a model that can be used to predict the ventricular volume using 2-D cranial US measurements. For further analysis we chose to use a transformation with the cube root rather than a log transformation because this better explains the relationship between the physical quantities. That is, the model predicts volumes that scale as the third power of the linear measurements. Forward variable selection showed that the model with anterior horn width and thalamo-occipital distance had the highest adjusted R2 and provided a fit with the most symmetrical distribution of residuals, resulting in a model that can be used to predict ventricular volume (∛ventricular volume [ml] = 1.096+0.094 × anterior horn width + 0.020 × thalamo-occipital distance [mm] with R2 = 0.831). More details are given in Table 3. The residuals of our model are shown in Fig. 4.
Table 3
Parameter values of the univariate general linear model with the parameters anterior horn width and thalamo-occipital distance and the cube root of the ventricular volume as the independent variables
Parameter
Parameter value
P
95% confidence interval
Lower bound
Upper bound
a
1.096
<0.001
.969
1.224
b
0.094
<0.001
.072
.116
c
0.020
<0.001
.013
.028
R2 = 0.831 (Adjusted R2 = 0.825)
∛ventricular volume (ml) = a + b × anterior horn width (mm) + c × thalamo-occipital distance (mm)
The model as described here was validated against the model of Benavente-Fernandez et al. [13] (ventricular volume [ml] = −11.02+0.668 × ventricular index [mm] + 0.817 × anterior horn width [mm] + 0.256 × thalamo-occipital distance [mm]) [13]. Figure 5 shows the predicted ventricular volume as a function of the measured ventricular volume for both models and the residual error as a function of the measured ventricular volume for both models. The figure shows that larger discrepancies can be observed for ventricular volume greater than 20 ml, whereas smaller discrepancies of up to 5 ml were observed for ventricular volumes smaller than 20 ml.

Discussion

In this study, we found that 2-D cranial US ventricular measurements of infants with germinal matrix intraventricular haemorrhage were significantly associated with the volume of the ventricles as determined with MR images in a wide range of ventricular volumes. The proposed model provides us with an estimate of the ventricular volume that can be obtained by performing only a few linear measurements on US images, which enables bedside estimation of ventricular volume. This highlights the utility of cranial US in the diagnosis and monitoring of preterm infants with post-haemorrhagic ventricular dilatation, and should encourage neonatologists to perform linear cranial US measurements of the cerebral ventricles. Figure 5 shows us that larger deviations from the model seem to appear for volumes over 20 ml. Although more data are needed to confirm this hypothesis, it seems that the model struggles to predict the larger volumes accurately. This might be explained by the fact that the shape of the ventricles changes considerably for larger volumes, so the estimation of the volume based on two local linear measurements might be more likely to fail. We argue that this does not necessarily hamper the use of the model in clinical practice. Infants with ventricular volumes higher than 20 ml undergo MRI regardless of the exact results of the US examination. Especially for smaller ventricular volumes, the model might aid in the discrimination between healthy infants and those with mild cases of germinal matrix intraventricular haemorrhage.
Other relevant work on the prediction of ventricular volume using US has been performed by Benavente-Fernandez et al. [13]. They created a model to predict the ventricular volume as measured by 3-D cranial US using linear 2-D cranial US measurements for infants with post-haemorrhagic ventricular dilatation. There were two major differences between this model (ventricular volume = a + b × ventricular index + c × anterior horn width + d × thalamo-occipital distance) and our model (∛ventricular volume = a + b × anterior horn width + c × thalamo-occipital distance) that might explain the observed discrepancies. First, the Benavente-Fernandez model included the ventricular index, anterior horn width and thalamo-occipital distance, whereas we only included the anterior horn width and thalamo-occipital distance. Second, their model did not perform a cube root transformation of the ventricular volume (∛ventricular volume). Other differences might be explained by the fact that 3rd and 4th ventricles were not included in the 3-D volume measurements by Benavente-Fernandez et al. [13], whereas they were included in our segmentations. The 3rd ventricle is hard to visualize on US imaging and, hence, few reference values are known in the literature. Moreover, the relative overestimation of ventricular volume on MRI as compared to 3-D cranial US has also been reported by Kishimoto et al. [14], which might be attributed to the fact that the posterior horns are difficult to visualize.
Limitations of this retrospective study include errors in the measurement of the variables apart from the variance in the data owing to the anatomical variation among patients and the intraobserver variability. For example, errors in the measurement of the ventricular index, anterior horn width and thalamo-occipital distance with 2-D cranial US might have occurred because of the absence of contrast between the borders of the ventricles and the rest of the surrounding brain tissue, especially for the occipital horns. Errors in the measurement of the ventricular volume on MRI might be caused by susceptibility artifacts or partial volume effects. Finally, errors in the model parameters could have been introduced by differences between the actual volume of the ventricles at the time of the MRI and 2-D cranial US acquisition; to minimize this difference, we chose a maximum of 3 days between the MRI and 2-D cranial US acquisition.
The reliability of our model is affected by the variance of linear 2-D cranial US measurements and because few data points are available for ventricles with a large volume. The fact that linear measurements with large variance are used, and that the distribution of several of the measured parameters is asymmetrical, could impact the prediction. This could also explain the fact that larger discrepancies between our model and the Benavente-Fernandez model were observed for larger volumes.
Although this study aimed to improve ventricular volume determination in clinical practice, the relation between ventricular volume and intracranial pressure — which might be the actual cause of damage — remains unclear, and additional investigations performed simultaneously, such as near infrared spectroscopy (NIRS), amplitude integrated electroencephalogram (EEG) and Doppler measurements of cerebral blood flow might provide additional important information [15, 16].

Conclusion

With the use of 2-D cranial US an approximation of the ventricular volume can be provided, which might be helpful for better management of post-haemorrhagic ventricular dilatation. Validation of our model in a larger study group including different time points and ventricular sizes is warranted.

Compliance with ethical standards

Conflicts of interest

None
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Pädiatrie

Kombi-Abonnement

Mit e.Med Pädiatrie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Pädiatrie, den Premium-Inhalten der pädiatrischen Fachzeitschriften, inklusive einer gedruckten Pädiatrie-Zeitschrift Ihrer Wahl.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Murphy BP, Inder TE, Rooks V et al (2002) Posthaemorrhagic ventricular dilatation in the premature infant: natural history and predictors of outcome. Arch Dis Child Fetal Neonatal Ed 87:F37–F41CrossRef Murphy BP, Inder TE, Rooks V et al (2002) Posthaemorrhagic ventricular dilatation in the premature infant: natural history and predictors of outcome. Arch Dis Child Fetal Neonatal Ed 87:F37–F41CrossRef
2.
Zurück zum Zitat Volpe JJ (1989) Intraventricular hemorrhage in the premature infant-current concepts. Part I. Ann Neurol 25:3–11CrossRef Volpe JJ (1989) Intraventricular hemorrhage in the premature infant-current concepts. Part I. Ann Neurol 25:3–11CrossRef
3.
Zurück zum Zitat de Vries LS, Liem KD, van Dijk K et al (2002) Early versus late treatment of posthaemorrhagic ventricular dilatation: results of a retrospective study from five neonatal intensive care units in the Netherlands. Acta Paediatr 91:212–217CrossRef de Vries LS, Liem KD, van Dijk K et al (2002) Early versus late treatment of posthaemorrhagic ventricular dilatation: results of a retrospective study from five neonatal intensive care units in the Netherlands. Acta Paediatr 91:212–217CrossRef
4.
Zurück zum Zitat de Vries LS, Groenendaal F, Liem KD et al (2018) Treatment thresholds for intervention in posthaemorrhagic ventricular dilation: a randomised controlled trial. Arch Dis Child Fetal Neonatal Ed 104:F70–F75CrossRef de Vries LS, Groenendaal F, Liem KD et al (2018) Treatment thresholds for intervention in posthaemorrhagic ventricular dilation: a randomised controlled trial. Arch Dis Child Fetal Neonatal Ed 104:F70–F75CrossRef
5.
Zurück zum Zitat Leijser LM, Miller SP, van Wezel-Meijler G et al (2018) Posthemorrhagic ventricular dilatation in preterm infants: when best to intervene? Neurology 90:e698–e706CrossRef Leijser LM, Miller SP, van Wezel-Meijler G et al (2018) Posthemorrhagic ventricular dilatation in preterm infants: when best to intervene? Neurology 90:e698–e706CrossRef
6.
Zurück zum Zitat Brouwer MJ, de Vries LS, Groenendaal F et al (2012) New reference values for the neonatal cerebral ventricles. Radiology 262:224–233CrossRef Brouwer MJ, de Vries LS, Groenendaal F et al (2012) New reference values for the neonatal cerebral ventricles. Radiology 262:224–233CrossRef
7.
Zurück zum Zitat Horsch S, Bengtsson J, Nordell A et al (2009) Lateral ventricular size in extremely premature infants: 3D MRI confirms 2D ultrasound measurements. Ultrasound Med Biol 35:360–366CrossRef Horsch S, Bengtsson J, Nordell A et al (2009) Lateral ventricular size in extremely premature infants: 3D MRI confirms 2D ultrasound measurements. Ultrasound Med Biol 35:360–366CrossRef
8.
Zurück zum Zitat Levene MI, Starte DR (1981) A longitudinal study of post-haemorrhagic ventricular dilatation in the newborn. Arch Dis Child 56:905–910CrossRef Levene MI, Starte DR (1981) A longitudinal study of post-haemorrhagic ventricular dilatation in the newborn. Arch Dis Child 56:905–910CrossRef
9.
Zurück zum Zitat Davies MW, Swaminathan M, Chuang SL, Betheras FR (2000) Reference ranges for the linear dimensions of the intracranial ventricles in preterm neonates. Arch Dis Child Fetal Neonatal Ed 82:F218–F223CrossRef Davies MW, Swaminathan M, Chuang SL, Betheras FR (2000) Reference ranges for the linear dimensions of the intracranial ventricles in preterm neonates. Arch Dis Child Fetal Neonatal Ed 82:F218–F223CrossRef
10.
Zurück zum Zitat Brouwer MJ, de Vries LS, Kersbergen KJ et al (2016) Effects of posthemorrhagic ventricular dilatation in the preterm infant on brain volumes and white matter diffusion variables at term-equivalent age. J Pediatr 168:41–49.e1CrossRef Brouwer MJ, de Vries LS, Kersbergen KJ et al (2016) Effects of posthemorrhagic ventricular dilatation in the preterm infant on brain volumes and white matter diffusion variables at term-equivalent age. J Pediatr 168:41–49.e1CrossRef
11.
Zurück zum Zitat Anbeek P, Vincken KL, Groenendaal F et al (2008) Probabilistic brain tissue segmentation in neonatal magnetic resonance imaging. Pediatr Res 63:158–163CrossRef Anbeek P, Vincken KL, Groenendaal F et al (2008) Probabilistic brain tissue segmentation in neonatal magnetic resonance imaging. Pediatr Res 63:158–163CrossRef
12.
Zurück zum Zitat Blok CA, Kersbergen KJ, van der Aa NE et al (2014) Unmyelinated white matter loss in the preterm brain is associated with early increased levels of end-tidal carbon monoxide. PLoS One 9:e89061CrossRef Blok CA, Kersbergen KJ, van der Aa NE et al (2014) Unmyelinated white matter loss in the preterm brain is associated with early increased levels of end-tidal carbon monoxide. PLoS One 9:e89061CrossRef
13.
Zurück zum Zitat Benavente-Fernandez I, Lubián-Gutierrez M, Jimenez-Gomez G et al (2017) Ultrasound lineal measurements predict ventricular volume in posthaemorrhagic ventricular dilatation in preterm infants. Acta Paediatr 106:211–217CrossRef Benavente-Fernandez I, Lubián-Gutierrez M, Jimenez-Gomez G et al (2017) Ultrasound lineal measurements predict ventricular volume in posthaemorrhagic ventricular dilatation in preterm infants. Acta Paediatr 106:211–217CrossRef
14.
Zurück zum Zitat Kishimoto J, Fenster A, Lee DSC, de Ribaupierre S (2016) In vivo validation of a 3-D ultrasound system for imaging the lateral ventricles of neonates. Ultrasound Med Biol 42:971–979CrossRef Kishimoto J, Fenster A, Lee DSC, de Ribaupierre S (2016) In vivo validation of a 3-D ultrasound system for imaging the lateral ventricles of neonates. Ultrasound Med Biol 42:971–979CrossRef
15.
Zurück zum Zitat Klebermass-Schrehof K, Rona Z, Waldhör T et al (2013) Can neurophysiological assessment improve timing of intervention in posthaemorrhagic ventricular dilatation? Arch Dis Child Fetal Neonatal Ed 98:F291–F297CrossRef Klebermass-Schrehof K, Rona Z, Waldhör T et al (2013) Can neurophysiological assessment improve timing of intervention in posthaemorrhagic ventricular dilatation? Arch Dis Child Fetal Neonatal Ed 98:F291–F297CrossRef
16.
Zurück zum Zitat Norooz F, Urlesberger B, Giordano V et al (2015) Decompressing posthaemorrhagic ventricular dilatation significantly improves regional cerebral oxygen saturation in preterm infants. Acta Paediatr 104:663–669CrossRef Norooz F, Urlesberger B, Giordano V et al (2015) Decompressing posthaemorrhagic ventricular dilatation significantly improves regional cerebral oxygen saturation in preterm infants. Acta Paediatr 104:663–669CrossRef
Metadaten
Titel
Two-dimensional ultrasound measurements vs. magnetic resonance imaging-derived ventricular volume of preterm infants with germinal matrix intraventricular haemorrhage
verfasst von
Casper Beijst
Jeroen Dudink
Rens Wientjes
Isabel Benavente-Fernandez
Floris Groenendaal
Margaretha J. Brouwer
Ivana Išgum
Hugo W. A. M. de Jong
Linda S. de Vries
Publikationsdatum
06.11.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
Pediatric Radiology / Ausgabe 2/2020
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-019-04542-x